K Number
K022570

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-10-31

(90 days)

Product Code
Regulation Number
876.5540
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

THE MEDCOMP REPAIR KIT IS INDICATED FOR USE IN REPLACING DAMAGED FEMALE LUER CONNECTORS, CLAMPS, OR REPAIRING EXTENSIONS WHERE THERE IS A MINIMUM OF 4.5cm VIABLE EXTENSION TUBING.

REPAIRS MEDCOMP CATHETERS: 10F Ash Split Cath® XL, 10F Ash Split Cath® II, 14F Ash Split Cath®, 14F Ash Split Cath® XL, 14F Ash Split Cath® II, 16F Ash Split Cath® II

Device Description

The Medcomp Repair Kit is two-piece replacement luer adaptor and cap used to repair damaged female luer connectors on Medcomp Hemodialysis Catheters. The Medcomp Repair Kit also consists of replacement arterial and venous clamps.

AI/ML Overview

Acceptance Criteria and Device Performance Study for Medcomp Repair Kit (K022570)

The Medcomp Repair Kit is a two-piece replacement luer adaptor and cap intended for repairing damaged female luer connectors on Medcomp Hemodialysis Catheters. It also includes replacement arterial and venous clamps. The device is designed to repair extensions where there is a minimum of 4.5cm viable extension tubing on specific Medcomp Catheter models.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Simulated Use PerformanceDemonstrated substantial equivalence to predicate device
Tensile StrengthDemonstrated substantial equivalence to predicate device
LeakageDemonstrated substantial equivalence to predicate device
(General): Safety and EffectivenessSufficiently demonstrated through in-vitro testing
(General): Substantial Equivalence to Predicate DeviceIn-vitro performance data supports substantial equivalence

2. Sample Size for Test Set and Data Provenance

No specific test set sample size is explicitly stated in the provided document. The performance data relied on in-vitro testing only. Therefore, there is no information regarding the country of origin or whether the data was retrospective or prospective, as these aspects are typically relevant for clinical studies.

3. Number of Experts and Qualifications for Ground Truth Establishment

Not applicable. The study involved in-vitro testing, not clinical evaluation requiring expert consensus for ground truth establishment.

4. Adjudication Method for Test Set

Not applicable. As the study was an in-vitro performance evaluation, there was no human adjudication process involved.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. The submission explicitly states that "Clinical studies were not deemed necessary."

6. Standalone (Algorithm Only) Performance Study

Not applicable. The Medcomp Repair Kit is a physical medical device, not an algorithm or AI-driven system. Therefore, a standalone (algorithm only) performance study is not relevant.

7. Type of Ground Truth Used

The ground truth for this device's performance was established through in-vitro testing against the performance characteristics of a legally marketed predicate device (K011015 Bard Catheter Repair Kit). This implies that the 'ground truth' was defined by the established performance benchmarks and regulatory compliance of the predicate device.

8. Sample Size for the Training Set

Not applicable. There was no "training set" as this device is not an AI/ML algorithm requiring training data. The assessment was based on comparing the repair kit's physical and functional performance to a predicate device.

9. How the Ground Truth for the Training Set Was Established

Not applicable. As there was no training set, there was no ground truth established for one. The evaluation focused on demonstrating that the performance of the Medcomp Repair Kit was substantially equivalent to the predicate device through in-vitro testing.

{0}------------------------------------------------

K022570

OCT 31 2002

510(k) SUMMARY

A. Submitter Information:

Submitter:

MEDCOMP® 1499 Delp Drive Harleysville, PA 19438 (215) 256-4201 Telephone (215) 256-1787 Fax Florence A. Caikoski Contact: Date Prepared: October 30, 2002 Medcomp Repair Kit Trade Name: Common Name: Kit, Repair, Catheter, Hemodialysis Classification: NFK 876.5540 C.F.R. Section:

C. Predicate Device:

в.

K011015 Bard Catheter Repair Kit

D. Device Description:

The Medcomp Repair Kit is two-piece replacement luer adaptor and cap used to repair damaged female luer connectors on Medcomp Hemodialysis Catheters. The Medcomp Repair Kit also consists of replacement arterial and venous clamps.

E. Intended Use:

The Medcomp Repair Kit is indicated for use in replacing damaged female luer connectors, clamps, or repairing extensions where there is a minimum of 4.5cm viable extension tubing.

Repairs Medcomp Catheters: 10F Ash Split Cath® XL, 10F Ash Split Cath® II, 14F Ash Split Cath®, 14F Ash Split Cath® XL, 14F Ash Split Cath® II, 16F Ash Split Cath® II

F. Comparison to Predicate Device:

The technological characteristics of the Medcomp Repair Kit are substantially equivalent to the predicate device in terms of intended use, design, material type, performance, and method of sterilization.

G. Performance Data:

In-Vitro performance data for the Medcomp Repair Kit including simulated use, tensile strength and leakage demonstrates that this device is substantially equivalent to legally marketed device.

Clinical studies were not deemed necessary since in-vitro testing was sufficient to demonstrate safety and effectiveness by way of comparison to the legally marketed predicate device.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administrat 9200 Corporate Boulevaro Rockville MD 20850

Ms. Florence A. Caikoski Regulatory Affairs Associate MEDCOMP 1499 Delp Drive HARLEYSVILLE PA 19438

Re: K022570

Trade/Device Name: MEDCOMP Repair Kit, Model # ASPCRPK, for repairing 10F and 14F Ash Split Cath® XL; 14F Ash Split Cath®; and 10F, 14F and 16F Ash Split Cath® II

Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: 78 NFK Dated: July 31, 2002 Received: August 2, 2002

Dear Ms. Caikoski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

Page 2 – Ms. Florence Caikoski

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807): listing (21 CFR Part 807). labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594- _ . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours.

Demid C. digson

fr

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

INDICATIONS FOR I

510(k) Number: K022570

Device Name: MEDCOMP REPAIR KIT

Indications for use:

THE MEDCOMP REPAIR KIT IS INDICATED FOR USE IN REPLACING DAMAGED FEMALE LUER CONNECTORS, CLAMPS, OR REPAIRING EXTENSIONS WHERE THERE IS A MINIMUM OF 4.5cm VIABLE EXTENSION TUBING.

REPAIRS MEDCOMP CATHETERS: 10F Ash Split Cath® XL, 10F Ash Split Cath® II, 14F Ash Split Cath®, 14F Ash Split Cath® XL, 14F Ash Split Cath® II, 16F Ash Split Cath® II

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter

Edwin A. Symon

(Division Sign-510(k) Number

(Optional Format 1-2-96)

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.