K Number
K022426

Validate with FDA (Live)

Date Cleared
2002-08-08

(14 days)

Product Code
Regulation Number
862.1675
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

BD Preset™ and BD A-Line™ Blood Collection Syringe are intended to collect whole blood specimens for diagnostic testing which may include: pH, blood gases, electrolytes (including ionized calcium), metabolytes, co-oximetry, and other tests

Device Description

The BD Preset™ and BD A-Line™ Blood Collection Syringes are sterile, single use devices designed to collect whole blood specimens for diagnostic testing.

BD A-Line™ Blood Collection Syringe is specifically designed for aspiration of blood samples from arterial lines.

BD Preset™ Blood Collection Syringe is a specifically designed (may include needle) that can be preset to a desired volume, but permits aspiration when necessary. Includes a venting system that expels residual air through the selfventing membrane (as blood fills the syringe), which ensures rapid filling.

AI/ML Overview

The medical device in question is the BD Preset™ and BD A-Line™ Blood Collection Syringes.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state specific quantitative acceptance criteria for the device. Instead, it relies on a qualitative statement of "equivalence" to a predicate device.

Acceptance Criteria (Implied)Reported Device Performance
Clinical performance and equivalence to predicate devices"All results from the studies show equivalence between the principal devices and the predicate devices."
Intended Use: Collect whole blood for diagnostic testing (pH, blood gases, electrolytes, metabolites, co-oximetry, and other tests)The device is "designed to collect whole blood specimens for diagnostic testing" as listed in the intended use. The equivalence study supports this implicitly.
Sterility"sterile, single use devices" (stated in device description)

2. Sample Size Used for the Test Set and Data Provenance

  • The document does not specify the sample size used for the test set in the performance studies.
  • The document does not specify the data provenance (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts

  • The document does not provide information regarding the number of experts used or their qualifications for establishing ground truth. Since the studies focus on device performance and equivalence to a predicate, it's likely that analytical results from the collected blood specimens were compared, rather than expert evaluation of images or clinical diagnoses.

4. Adjudication Method

  • The document does not describe any adjudication method. This is consistent with a performance study focused on analytical equivalence rather than diagnostic interpretation.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • A MRMC comparative effectiveness study was not performed or reported. This type of study is more relevant for diagnostic AI tools where human readers interpret patient data (e.g., images) with and without AI assistance to measure improvements in diagnostic accuracy or efficiency. The BD Preset™ and BD A-Line™ Blood Collection Syringes are collection devices, not diagnostic interpretation tools.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop Performance)

  • A standalone performance study was implicitly done in the sense that the "principal devices" (the new syringes) were tested directly to assess their performance and compare them to the predicate devices. There is no "algorithm" in the context of these blood collection syringes. The performance study evaluates the physical and functional attributes of the syringe itself.

7. Type of Ground Truth Used

  • The document implies that the ground truth for the performance studies was based on the results of diagnostic tests performed on the collected blood specimens. The study aimed to show that the new syringes collected blood specimens that yielded comparable diagnostic test results to those collected by the predicate device. This would likely involve:
    • Analytical data: Measurements of pH, blood gases, electrolytes, metabolites, co-oximetry, etc., from blood samples.
    • Comparison to predicate device: The predicate device's performance, which is presumably well-established and accepted, serves as a de-facto "ground truth" or reference point for equivalence.

8. Sample Size for the Training Set

  • The concept of a "training set" is not applicable to this device. This is a physical medical device (blood collection syringe), not an AI algorithm that requires a training set of data.

9. How Ground Truth for the Training Set Was Established

  • As a training set is not applicable, this question is not relevant for this medical device.

{0}------------------------------------------------

AUG 0 8 2002

1022426

Attachment 4

510(k) Summary of Safety and Effectiveness

  • I. General Information
    This Summary of Safety and Effectiveness information is being submitted in accordance with the requirements of the SMDA of 1990 and 21 § 807.92

Establishment:

Address:BD Vacutainer Systems, PreAnalytical Solutions1 Becton DriveFranklin Lakes, NJ 07417-1885
Registration Number:Contact Person:2243072M. Wendy BosshardtRegulatory Affairs SpecialistTelephone no.: 201-847-6280Fax No. 201-847-4858
DeviceDate of Summary:July 22, 2002
Trade Name:BD Preset™ and BD A-Line™ BloodCollection Syringes
Classification Name:Tubes, Vials, Systems, SerumSeparators, Blood Collection
Classification:Performance Standards:Class IINone Established under 514 of theFood, Drug and Cosmetic Act

Special 510(k): Device Modification
BD Preset™ and BD A-Line™

{1}------------------------------------------------

  • II. Safety and Effectiveness Information Supporting Substantial Equivalence
    • · Device Description

The BD Preset™ and BD A-Line™ Blood Collection Syringes are sterile, single use devices designed to collect whole blood specimens for diagnostic testing.

BD A-Line™ Blood Collection Syringe is specifically designed for aspiration of blood samples from arterial lines.

BD Preset™ Blood Collection Syringe is a specifically designed (may include needle) that can be preset to a desired volume, but permits aspiration when necessary. Includes a venting system that expels residual air through the selfventing membrane (as blood fills the syringe), which ensures rapid filling.

  • Intended Use
    The BD Preset™ and BD A-Line™ Blood Collection Syringes are intended to collect whole blood specimens for diagnostic testing which may include: pH. blood gases, electrolytes (including ionized calcium), metabolytes, co-oximetry, and other tests.

  • · Synopsis of Performance Study Results
    Performance studies were done the performance and equivalence of the principal devices to the predicate devices currently marketed in the United States.

All results from the studies show equivalence between the principal devices and the predicate devices. Therefore, the BD Preset™ and BD A-Line™ Blood Collection Syringes are substantially equivalent to the predicate devices.

{2}------------------------------------------------

III. Predicate Device Summary Table

  • · Substantial Equivalence
    Based on comparison of the device features, materials, intended use and performance, the BD Preset™ and BD A-Line™ Blood Collection Syringes are shown to be substantially equivalent to the commercially available predicate devices indicated in the table below. The predicate devices, K number, and clearance date are also identified in the table below.
ManufacturerPredicate DeviceK-NumberClearance Date
BectonDickinsonVACUTAINER™K982922Sept. 22, 1998
Brand BloodCollection Syringes

M. Wendy Booshardt

M. Wendy Bosshardt Regulatory Affairs Specialist Becton Dickinson VACUTAINER Systems Becton Dickinson and Company

July 22, 2002
Date

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" written around it. Inside the circle is a stylized image of an eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. M. Wendy Bosshardt Regulatory Affairs Specialist Becton, Dickinson and Co. 1 Becton Drive Franklin Lakes, NJ 07417

Re: K022426

Trade/Device Name: BD Preset™ and BD A-Line™ Blood Collection Syringe Regulation Number: 21 CFR 862.1675 Regulation Name: Blood specimen collection device Regulatory Class: Class II Product Code:. JKA Dated: July 22, 2002 Received: July 25, 2002

Dear Ms. Bosshardt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed "**** predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Attachment 2

Indications for Use Statement

.

. :

100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000

··············································································································································································

510(k) Number (if known)

:

Device NameBD Preset™ and BD A-Line™ Blood Collection Syringe
Indications for UseBD Preset™ and BD A-Line™ Blood Collection Syringeare intended to collect whole blood specimens fordiagnostic testing which may include: pH, blood gases,electrolytes (including ionized calcium), metabolytes,co-oximetry, and other tests

PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Device.
510(k) NumberK0224266
Prescription Use(per 21 CFR 801.109)OROver-The Counter Use
--------------------------------------------------------------------

§ 862.1675 Blood specimen collection device.

(a)
Identification. A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.(b)
Classification. Class II.