K Number
K022129

Validate with FDA (Live)

Date Cleared
2002-07-25

(24 days)

Product Code
Regulation Number
866.1645
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram negative and Gram- positive bacteria of human origin.

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.

This premarket notification is for the addition of the antimicrobial Moxifloxacin at concentrations of 0.125-2 ug/mL to Gram negative ID/AST or AST only Phoenix panels. Moxifloxacin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobic.

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software.
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
  • . BD Phoenix AST Broth used for performing AST tests only.
  • . BD Phoenix AST Indicator solution added to the AST broth to aid in bacterial growth determination.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a gram-negative or gram-positive isolate. For each isolation equivalent to 0.5 McFarland standard is prepared in Phoenix ID broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S. I, or R (sensitive, intermediate, and resistant).

AI/ML Overview

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state quantitative acceptance criteria in a table format, but it implies them through the "Essential Agreement (EA)" and "Category Agreement (CA)" metrics. For intra- and inter-site reproducibility, the acceptance criteria are implicitly greater than 90% and greater than 95%, respectively.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance
Intra-site Reproducibility (Moxifloxacin, Gram-negative)> 90%> 90%
Inter-site Reproducibility (Moxifloxacin, Gram-negative)> 95%> 95%
Essential Agreement (EA)Not explicitly statedSummarized in Table 1 (Table 1 is incomplete in the provided text, but states "1000" for "Moxitloxacın" which implies good performance.)
Category Agreement (CA)Not explicitly statedSummarized in Table 1 (Table 1 is incomplete in the provided text, but states "1000" for "Moxitloxacın" which implies good performance.)

2. Sample Size Used for the Test Set and Data Provenance:

  • Type of Isolates: Clinical, stock, and challenge isolates.
  • Provenance: Across multiple geographically diverse sites across the United States.
  • Retrospective/Prospective: Not explicitly stated, but the mention of "clinical isolates" being compared to "reference broth microdilution method" suggests these were likely prospective or at least collected for this specific study. "Stock and challenge isolates" are by nature pre-defined.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This information is not provided in the document. The ground truth for clinical isolates was established by the "NCCLS reference broth microdilution method," which is a standardized laboratory method, not reliant on individual human experts in the same way as, for example, a radiologist's interpretation. For "challenge set isolates," performance was compared to "expected results," implying predefined ground truth based on known characteristics of those isolates.

4. Adjudication Method for the Test Set:

This information is not applicable as the ground truth was established by comparing to a reference method (NCCLS broth microdilution) or expected results for challenge isolates, not through human expert consensus needing adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

This was not a MRMC comparative effectiveness study involving human readers with and without AI assistance. The study assessed the performance of an automated microbiology system (BD Phoenix) against a reference laboratory method (NCCLS broth microdilution). It did not evaluate human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

Yes, a standalone study was performed. The entire evaluation focuses on the performance of the "BD Phoenix™ Automated Microbiology System" (an algorithm/automated device) at determining antimicrobial susceptibility. Its results are compared directly to the NCCLS reference method, without human intervention in the interpretation phase being evaluated.

7. The Type of Ground Truth Used:

  • For Clinical Isolates: The ground truth was established by the NCCLS reference broth microdilution method. This is an established laboratory standard method.
  • For Challenge Set Isolates: The ground truth was based on "expected results," implying pre-defined, known results for these specific isolates, often used to test the limits or specific characteristics of a system.

8. The Sample Size for the Training Set:

The document does not specify a training set size. This is a premarket notification for an existing system (BD Phoenix) with the addition of a new antimicrobial agent (Moxifloxacin). The "clinical studies" described are likely validation/test sets, not a training set in the machine learning sense. The system itself was likely developed and "trained" (in a broader engineering sense) prior to this specific submission.

9. How the Ground Truth for the Training Set Was Established:

As no specific training set is outlined in this document for the addition of Moxifloxacin, the method for establishing ground truth for any underlying training of the overall BD Phoenix system is not provided.

{0}------------------------------------------------

KC022129

JUL 2 5 2002

CONFIDENTIAL AND PROPRIETARY

510(k) SUMMARY

SUBMITTED BY:Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152 Phone: 410-316-4260 Fax: 410-316-4499
CONTACT NAME:Kathryn Babka Powers Regulatory Affairs Specialist
DATE PREPARED:June 27, 2002
DEVICE TRADE NAME:BD Phoenix™ Automated Microbiology System - Moxifloxacin 0.125 - 8 µg/mL
DEVICE COMMON NAME:Antimicrobial susceptibility test system-short incubation
DEVICE CLASSIFICATION:In accordance with FDA's reclassification order issued December 28, 2001, Docket # 97P-0313, the BD Phoenix™ Automated Microbiology System has been classified as a Class II device, Automated Antimicrobial Susceptibility System Test, short incubation (Product Code LON)
PREDICATE DEVICES:VITEK® System (PMA No. N50510) and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
INTENDED USE:The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram negative and Gram- positive bacteria of human origin.

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software.
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
  • . BD Phoenix AST Broth used for performing AST tests only.
  • . BD Phoenix AST Indicator solution added to the AST broth to aid in bacterial growth determination.

{1}------------------------------------------------

BD PHOENIX™ Automated Microbiology System Moxifloxacın 0.125 - 8 ug/mL

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a gram-negative or gram-positive isolate. For each isolation equivalent to 0.5 McFarland standard is prepared in Phoenix ID broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S. I, or R (sensitive, intermediate, and resistant).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram negative ID/AST or AST only Phoenix panels with Moxifloxacin.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices". March 8, 2000. : 上一篇:

Site Reproducibility

Intra- and inter-site reproducibility of Moxifloxacin in the BD Phoenix System was evaluated at three sites using a panel of gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram Negative Phoenix panels with Moxifloxacin and associated reagents.

The results of the study demonstrate for the antimicrobial Moxifloxacin an overall intra-site reproducibility greater than 90% and an overall inter-site reproducibility greater than 95% for the gram-negative isolates tested.

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with

{2}------------------------------------------------

the Gram Negative Phoenix Panel format containing Moxifloxacin. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + on two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Table 1 summarizes the performance for the isolates tested in this study.

Table 1: Performance of BD Phoenix System for Gram-Negative Organisms by Drug

Antimicrobialoncentration114. 11mAn An Antony Property Courtes
.10000000no on1000nr
Moxitloxacın110/m- I hand don't load .Dodhad had find the first and the first and the first and the first and the first and the first and the first and the first and the first and the first and the forme the forme t

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with Moxifloxacin is substantially equivalent as outlined in the FDA draft guidance document. "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002). Ofloxacin (K020323. April 14, 2002), and Levofloxacin (K020322, March 27, 2002).

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight. The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Kathryn Babka Powers, RAC Regulatory Affairs Specialist Becton Dickinson and Company 7 Loveton Circle Sparks, MD 21152

JUL 2 5 2002

K022129 Re:

Trade/Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial Moxifloxacin (0.125-8 ug/mL) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: June 28, 2002 Received: July 1, 2002

Dear Ms. Powers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number: TBD

BD Phoenix™ Automated Microbiology System for use with the antimicrobial Moxifloxacin Device Name: (0.125-8 ug/mL) Gram negative ID/AST or AST only Phoenix panels.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.

This premarket notification is for the addition of the antimicrobial Moxifloxacin at concentrations of 0.125-2 ug/mL to Gram negative ID/AST or AST only Phoenix panels. Moxifloxacin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobic.

Active In Vitro and in Clinical Infections Against:

Active In Vitro Against:

Aerobic Gram-negative microorganisms Klebsiella pneumoniae

Aerobic Gram-negative microorganisms Citrobacter freundii Enterobacter cloacae Escherichia coli Klebsiella oxytoca Proteus mirabilis

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Fuedda tu-Pooly

sion Sian-Off vision of Clinica

(Optional Format 3-10-98)

510(k) Number

Prescription Use #
Page 4

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”