K Number
K021839

Validate with FDA (Live)

Date Cleared
2002-07-16

(42 days)

Product Code
Regulation Number
874.3620
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the AART Gluteal Implant (Silicone Carving Block) is for augmentation and reconstructive surgery where additional shaping by the surgeon may be necessary. It may be used in body contouring to minimize muscular defects.

Device Description

The AART Gluteal Implants (Silicone Carving Blocks) are manufactured from a medical grade silicone elastomer that has been molded into various convex oval or round shapes. They are provided in four styles with dimensions varying in length, width, height and initial shape. The AART Gluteal Implant (Silicone Carving Block) is intended to be used for augmentation and reconstructive surgery where additional shaping by the surgeon may be necessary.

AI/ML Overview

Here's an analysis of the provided text regarding the AART Gluteal Implant, focusing on acceptance criteria and study details.

Based on the provided document, there is no information about specific acceptance criteria or an associated study that proves the device meets such criteria.

The document is a 510(k) Summary and an FDA clearance letter for a medical device called the "AART Gluteal Implant (Silicone Carving Block)." This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria through a dedicated study.

Therefore, most of the requested information cannot be extracted from this document. However, I can still fill in what's available and explain why other sections are blank.


Analysis of Acceptance Criteria and Device Performance

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Not specified in the document. The 510(k) submission focuses on substantial equivalence rather than explicit performance criteria for this specific device.Not specified in the document. Performance is inferred to be similar to the predicate device due to substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample size for test set: Not applicable. The document describes a 510(k) submission based on substantial equivalence, not a clinical study involving a test set of data.
  • Data provenance: Not applicable. No clinical data is presented for this device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of experts: Not applicable. No test set requiring ground truth establishment is described.
  • Qualifications of experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication method: Not applicable. No test set requiring adjudication is described.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC study performed: No. The document does not describe any MRMC comparative effectiveness study. The submission is for substantial equivalence.
  • Effect size of human readers with vs. without AI assistance: Not applicable. This device is a physical implant, not an AI-assisted diagnostic or therapeutic tool.

6. Standalone (Algorithm Only) Performance Study

  • Standalone study performed: No. This device is a physical implant, not an algorithm.

7. Type of Ground Truth Used for the Test Set

  • Type of ground truth: Not applicable. No test set is described.

8. Sample Size for the Training Set

  • Sample size for training set: Not applicable. As this is a physical implant and not an AI-based system, there is no concept of a "training set."

9. How the Ground Truth for the Training Set Was Established

  • Method of ground truth establishment: Not applicable.

Summary Explanation:

The provided document is a 510(k) clearance for a Class II medical device (ear, nose, and throat synthetic polymer material, although the device is a gluteal implant, it was classified under this regulation number for some reason, as noted in the FDA letter). The core of a 510(k) submission is to demonstrate that the new device is "substantially equivalent" to a legally marketed predicate device. This typically involves comparing material properties, design, intended use, and performance characteristics to show they are sufficiently similar, meaning the new device is as safe and effective as the predicate.

For devices cleared through the 510(k) pathway, especially those like physical implants, extensive clinical trials with strict acceptance criteria and performance data (as would be seen in a PMA submission or for AI/diagnostic devices) are often not required if substantial equivalence can be adequately demonstrated. The document clearly states: "The AART Gluteal Implant is substantially equivalent in function, design, performance and materials to the Silimed Gluteal Implant marketed by Silimed LLC." This statement, along with the device description and comparison to the predicate, forms the basis of the FDA's clearance, not performance against specific, pre-defined acceptance criteria.

{0}------------------------------------------------

JUL 16 2002

ATTACHMENT 7

510(k) Summary

K02 1839

May 31, 2002

Applicant:Aesthetic and Reconstructive Technologies, Inc. (AART)
3545 Airway Drive, Suite 108
Reno, NV 89511
(775) 853-6800 / FAX (775) 853-6805
Contact Person:Catherine Riple
Consultant for AART, Inc.
(805) 239-1059
Proprietary Name:AART Gluteal Implant
Common Name:Silicone Carving Block
Classification Name:Elastomer, Silicone Block

Substantial Equivalence: The AART Gluteal Implant (Silicone Carving Block) is substantially equivalent in function, design, performance and materials to the Silimed Gluteal Implant marketed by Silimed LLC.

Device Description: The AART Gluteal Implants (Silicone Carving Blocks) are manufactured from a medical grade silicone elastomer that has been molded into various convex oval or round shapes. They are provided in four styles with dimensions varying in length, width, height and initial shape. The AART Gluteal Implant (Silicone Carving Block) is intended to be used for augmentation and reconstructive surgery where additional shaping by the surgeon may be necessary.

Intended Use: The intended use for the AART Gluteal Implant (Silicone Carving Block) is for augmentation and reconstructive surgery where additional shaping by the surgeon may be necessary. It may be used in body contouring to minimize muscular defects.

Predicate Device: The AART Gluteal Implant (Silicone Carving Block) is substantially equivalent in material, design, function, and performance to the Silimed Gluteal Implant marketed by Silimed LLC. of Crofton, MD. All products have identical intended uses and are offered in similar shapes and sizes.

Packaging: The AART Gluteal Implant (Silicone Carving Block) will be offered non-sterile, The implants will be individually packaged in an autoclavable peel pouch and labeled on the plastic side with appropriate identification for traceability. They will then be boxed in a SSB chipboard box along with a package insert and labeled for inventory and shipment.

Sterilization: The AART Gluteal Implants (Silicone Carving Blocks) are offered non-sterile. Recommended autoclave cycles are stated in the Package Insert which is included with each product.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a staff with two snakes entwined around it. The logo is black and white.

JUL 16 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Aesthetic and Reconstructive Technologies, Inc. c/o Ms. Catherine Riple 3545 Airway Drive, Suite 108 Reno. Nevada 89511

Re: K021839 Trade Name: AART Gluteal Implant Regulation Number: 874.3620 Regulation Name: Ear, nose and throat synthetic polymer material Regulatory Class: II Product Code: MIB Dated: May 31, 2002 Received: June 4, 2002

Dear Ms. Riple:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Catherine Riple ... ...

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Stupt Rhodes

lia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

ATTACHMENT 1

AART Gluteal Implant (Silicone Carving Block) Ko21839

Page

DEVICE NAME:

INDICATIONS FOR USE:

SIO (k) NUMBER (IF KNOWN) :

The intended use of the AART Gluteal Implant (Silicone Carving Block) is for augmentation and reconstructive surgery where additional shaping by the surgeon may be necessary. It may be used in body contouring to minimize muscular defects.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) .

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109

O

Over-The-Counter-Use_ (Optional Format 1-2-

Stephen Rurder

OR

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number_KOZ1839

§ 874.3620 Ear, nose, and throat synthetic polymer material.

(a)
Identification. Ear, nose, and throat synthetic polymer material is a device material that is intended to be implanted for use as a space-occupying substance in the reconstructive surgery of the head and neck. The device is used, for example, in augmentation rhinoplasty and in tissue defect closures in the esophagus. The device is shaped and formed by the suregon to conform to the patient's needs. This generic type of device is made of material such as polyamide mesh or foil and porous polyethylene.(b)
Classification. Class II.