K Number
K021825

Validate with FDA (Live)

Date Cleared
2002-07-01

(27 days)

Product Code
Regulation Number
888.3050
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The components of the Spiral Radius 90-DTM SST System are indicated for spinal fixation.

When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The Indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachments are the lumbar, thoracic and cervical spine and sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, The Spiral Radius 90-D™ SST System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and (d) who are having the device removed after the development of solid fusion mass.

In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Spiral Radius 90-DTM SST System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

Device Description

The Spiral Radius 90-DJM SST System consists of stainless steel rods attached to the spinal column through the use of hooks. Screws may be used to attach rods to the spinal column in certain select patients and certain select levels (See Indications). Cross-bars may be used to connect rods to provide a more rigid construct, as well as screws to rods, and hooks to rods. Screws and hooks are provided in several sizes to accommodate varying patient morphology.

AI/ML Overview

The provided document is a 510(k) summary for a medical device called the "Surgical Dynamics Spiral Radius 90-DTM SST System," an orthopedic spinal fixation system.

It does not describe any acceptance criteria or studies proving the device meets those criteria in the way typically associated with performance evaluations of AI/ML or diagnostic devices. Instead, this document is a regulatory submission demonstrating substantial equivalence to a previously marketed device. This means the device is deemed safe and effective because it is similar to another legally marketed device, not through direct performance studies against specific criteria.

Therefore, most of the requested information cannot be extracted from this document, as it pertains to a different type of regulatory submission and evaluation.

Here's what can be inferred or explicitly stated based on the provided text, and what cannot be answered:

1. Table of Acceptance Criteria and Reported Device Performance:

  • Cannot be provided. This document does not contain acceptance criteria for device performance (e.g., accuracy, sensitivity, specificity, or engineering performance metrics with thresholds). It is a substantial equivalence claim, not a performance study report.

2. Sample Size Used for the Test Set and Data Provenance:

  • Cannot be provided. There is no "test set" in the context of this 510(k) submission. The substantial equivalence claim relies on comparing the new device to a predicate device, not on testing the device against a defined dataset of patients/cases.

3. Number of Experts Used to Establish Ground Truth and Their Qualifications:

  • Cannot be provided. Since there's no test set or performance evaluation against a ground truth, there were no experts used in this capacity.

4. Adjudication Method:

  • Cannot be provided. Not applicable due to the reasons above.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Cannot be provided. There is no mention of an MRMC study. This type of study is common for diagnostic devices where human reader performance is being evaluated, often with or without AI assistance, which is not the subject of this 510(k) submission.

6. Standalone Performance Study (Algorithm Only):

  • Cannot be provided. This document is for a mechanical spinal fixation system, not an algorithm or AI-driven device. Therefore, a "standalone algorithm performance" study is not relevant.

7. Type of Ground Truth Used:

  • Cannot be provided. Not applicable. The "ground truth" for this submission is the established safety and effectiveness of the predicate device to which the Spiral Radius 90-D™ SST System is compared.

8. Sample Size for the Training Set:

  • Cannot be provided. This device is hardware; it does not involve training data or algorithms in the context of this 510(k) submission.

9. How Ground Truth for the Training Set Was Established:

  • Cannot be provided. Not applicable.

Summary based on the document:

This 510(k) submission (K021825) outlines the Surgical Dynamics Spiral Radius 90-DTM SST System, a spinal fixation system consisting of rods, screws, hooks, and cross-bars. The device's components, indicated uses (anterolateral/anterior, nonpedicle posterior, and pedicle screw systems for various spinal segments and conditions), and the claim of substantial equivalence to an existing marketed device are described. This regulatory pathway means that the device's safety and effectiveness are established by its similarity to a legally marketed predicate device, rather than through independent studies evaluating specific performance criteria against a ground truth.

Key takeaway for your request: This document is not the type that would contain the specific performance metrics, study designs, and ground truth information you're asking for, as it's a "substantial equivalence" claim for a hardware device, not a performance study report for a diagnostic or AI-enabled device.

{0}------------------------------------------------

Attachment 4

1 2002 JUL -

510(k) Summary

Q21825

Surgical Dynamics Spiral Radius 90-DTM SST System

page 1 of 2

United States Surgical 150 Glover Avenue Norwalk, CT 06856 USA

DEVICE DESCRIPTION

The Spiral Radius 90-DJM SST System consists of stainless steel rods attached to the spinal column through the use of hooks. Screws may be used to attach rods to the spinal column in certain select patients and certain select levels (See Indications). Cross-bars may be used to connect rods to provide a more rigid construct, as well as screws to rods, and hooks to rods. Screws and hooks are provided in several sizes to accommodate varying patient morphology.

INDICATIONS FOR USE

The components of the Spiral Radius 90-D™ SST System are indicated for spinal fixation.

When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The Indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachments are the lumbar, thoracic and cervical spine and sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, The Spiral Radius 90-D™ SST System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and (d) who are having the device removed after the development of solid fusion mass.

In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Spiral Radius 90-DTM SST System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

{1}------------------------------------------------

SUBSTANTIAL EQUIVALENCE*

K021825
page 2 of 2

The modified components of the Surgical Dynamics Spiral Radius 90-DTM vstem was claimed to be substantially equivalent* to the currently marketed version of the device. Information pertaining to this device was provided in the submission. ...

  • Any claim of substantial equivalence is made exclusively in regard to the U.S. Food, Drug and Cosmetic Act and should not be viewed in any other light.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of a bird or dove, composed of three curved lines that also suggest a human profile.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

1 2002 JUL

Ms. Tina Rideout Regulatory Affairs Senior Associate United States Surgical Surgical Dynamics 150 Glover Avenue Norwalk, Connecticut 06856

Re: K021825

Trade/Device Name: Spiral Radius 90-D™ SST System Regulation Number: 21 CFR 888.3050, 888.3060 and 888.3070 Regulation Name: Spinal interlaminal fixation orthosis, Spinal intervertebral body fixation orthosis, Pedicle screw system, Spondylolisthesis spinal fixation device system Regulatory Class: II Product Code: KWP. KWO. MNH, MNI Dated: June 3, 2002 Received: June 4, 2002

Dear Ms. Rideout:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{3}------------------------------------------------

. Page 2 – Ms. Tina Rideout

(21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594- . Additionally, for questions on the promotion and advertising of vour device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Stephen Rhodes

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Attachment 2

Indications for Use Statement

510(k) Number

Device Name: Spiral Radius 90-DTM SST System

Indications For Use

The components of the Spiral Radius 90-DTM SST System are indicated for spinal fixation.

When used as an anterolateral/anterior system consisting of rods and screws, the levels of attachment are the lumbar, thoracic and cervical spine. The points of attachment and methods are screw fixation to the anterolateral vertebral bodies of the lumbar and thoracic spine (T6-L5) and the anterior vertebral bodies of the cervical spine. The Indications are degenerative disc disease defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

When used as a nonpedicle posterior system consisting of hooks, crosslinks and sacral/iliac screws the levels of attachments are the lumbar, thoracic and cervical spine and sacrum and ilium. Intended uses include hook and sacral screw fixation to the lumbar spine, noncervical spine and to the T1-S1 spine; and hook and sacral/iliac screw fixation to the noncervical spine. The indications are degenerative disc disease defined as back pain of discogenic origin with degenerative of the disc confirmed by history and radiographic studies, spondylolisthesis, fracture, spinal stenosis, deformities (i.e., scoliosis, kyphosis, lordosis), tumors, pseudoarthrosis, or failed previous fusion (pseudoarthrosis).

When used as a pedicle screw system, in the non-cervical spine of skeletally mature patients. The Spiral Radius 90-D™ SST System is indicated for patients: (a) having severe spondylolisthesis (Grades 3 and 4) at the L5-S1 joint: (b) who are receiving fusions using autogenous bone graft only: (c) who are having the device fixed or attached to the lumbar and sacral spine (L3 and (d) who are having the device removed after the development of solid fusion mass.

In addition, when used as a pedicle screw system, in the non-cervical spine of skeletally mature patients, the Spiral Radius 90-D™ SST System is intended to provide immobilization and stabilization of spinal segments, as an adjunct to fusion, in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudoarthrosis).

(Please do not write below this line - continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)
--------------------------------------------------------
Prescription Use:(Per 21 CFR 801.109)XOR Over-The-Counter Use:
------------------------------------------------------------------------
Special 510 (k): Premarket NotificationJune 3, 2002Division of General, RestorativePage 1.8and Neurological Devices
-------------------------------------------------------------------------------------------------------------------------------------
510(k) NumberK021875
------------------------

§ 888.3050 Spinal interlaminal fixation orthosis.

(a)
Identification. A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.(b)
Classification. Class II.