K Number
K021823

Validate with FDA (Live)

Date Cleared
2002-06-17

(13 days)

Product Code
Regulation Number
892.1200
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the CardioSPECT SC and the CardioSPECT D90 is to detect the location and distribution of gamma ray emitting radionuclides in the human body and store the data for analysis as the same equivalent device, K013353.
Intended use is to detect & image the distribution of radionuclides in the body, when the radiopharmaceutical is administered in the body.

Device Description

The SC and D90 are scintillation detector, Nal, systems with PMT a) light gathering capacity. The D90 has two (2) detectors and the SC has one detector both with the ability to rotate these detectors around the patient. The system uses a collimator on each detector to permit spatial image resolution of the gamma photons emitted from the patient. The collimators lock in place. There is no built-in radioactive scanning source and none is required. The SC and D90 are capable of both Planar (Class I) and SPECT or ECT (Class ii) operation. The acquisition/processing computer is a presently approved system sold and supported by NC Systems, Inc. but manufactured by Segami Corporation. This application is for the ECT device-qamma camerasystem(s) and not the previously approved computer.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study details:

1. Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly list numerical acceptance criteria in terms of performance metrics (e.g., sensitivity, specificity, accuracy) for the CardioSPECT SC and D90 devices. Instead, the primary acceptance criterion is substantial equivalence to the predicate device, the NC Systems, Inc. QUAD (K013353).

The reported device performance is that the CardioSPECT SC and D90:

  • "perform well or better than the legally marketed device identified in paragraph 3.(a) of this section." (paragraph 6.g)
  • are "safe and as effective" as the legally marketed predicate device (paragraph 6.g).

Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criterion (Implicit)Reported Device Performance
Substantial Equivalence to Predicate Device (QUAD - K013353)Demonstrated through comparison of specifications, hazard analysis, labeling, technology, engineering, performance, and materials.
Safety equivalent to predicateDevice demonstrated to be "safe."
Effectiveness equivalent to predicateDevice demonstrated to be "as effective."
Performance equivalent or superior to predicateDevice demonstrated to "perform well or better than the legally marketed device."
Clinical utility for detecting and imaging radionuclide distributionIntended use confirmed: "to detect & image the distribution of radionuclides in the body."

2. Sample Size for the Test Set and Data Provenance:

  • Sample Size: The document states that "The clinical tests submitted were performed on, and clearly identified, the SC and D()." It further specifies that "The subjects were adults who were under the care of a physician licensed to perform these procedures." However, the exact number of subjects or cases used in the clinical test set is not provided.
  • Data Provenance: The document does not specify the country of origin for the clinical data. It states the manufacturing site is in Budapest, Hungary, but this doesn't directly indicate where the clinical studies were conducted. The study was conducted on "adults who were under the care of a physician licensed to perform these procedures," implying a prospective clinical study rather than retrospective data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

The document does not provide information on the number of experts used to establish ground truth or their qualifications. It only mentions that the subjects were "under the care of a physician licensed to perform these procedures," but this doesn't detail the "ground truth" establishment process for evaluating the device's performance against clinical reality.

4. Adjudication Method for the Test Set:

The document does not specify any adjudication method (e.g., 2+1, 3+1, none) used for the test set.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • Was it done? The document does not describe a multi-reader multi-case (MRMC) comparative effectiveness study. The comparison is between the new devices (SC and D90) and a predicate device (QUAD) in terms of their overall safety and effectiveness, rather than evaluating human reader performance with and without AI assistance.
  • Effect size: Since an MRMC study with AI assistance was not described, there is no information provided on the effect size of how much human readers improve with AI vs. without AI assistance.

6. Standalone Performance Study (Algorithm only without human-in-the-loop):

The document describes the device as an "Emission Computed Tomography System" (Gamma Camera-SPECT), which is inherently a medical imaging device rather than an AI algorithm. The tests were performed on the device itself. Therefore, a standalone algorithm-only performance study in the context of AI is not applicable to this type of medical device as presented. The "clinical tests" described represent the device's standalone performance in acquiring and imaging radionuclide distribution.

7. Type of Ground Truth Used:

The document does not explicitly state the type of ground truth used for the clinical tests. Given that it's an imaging device for detecting radionuclide distribution, it's highly probable that the ground truth would have been established through clinical assessment/diagnosis by physicians, potentially combined with other diagnostic imaging modalities or patient outcomes, but this is not detailed. It mentions subjects "under the care of a physician licensed to perform these procedures," suggesting clinical correlation.

8. Sample Size for the Training Set:

The document does not specify a training set sample size. This is consistent with the type of device (SPECT system) which doesn't typically undergo machine learning training in the same way an AI algorithm would. Its performance is based on the physics of gamma detection and image reconstruction.

9. How the Ground Truth for the Training Set Was Established:

As there's no mention of a traditional "training set" for a machine learning model, the document does not provide information on how ground truth for a training set was established. The device's performance relies on its engineering specifications and physical characteristics tested through non-clinical and clinical evaluations, not on learning from a labeled dataset.

{0}------------------------------------------------

K021823

JUN 1 7 2002

510(K) SUMMARY

21CFR 807.92

INTRODUCTION

This 510(K) Premarket Notification has been prepared to demonstrate that the CardioSPECT SC and CardioSPECT D90 Gamma ECT, manufactured by NC Systems, Inc. is substantially equivalent to the NC Systems, Inc. QUAD which has previously under went the 510(K) premarket notification process and approved (K013353).

SUBMITTER IDENTIFICATION 1.

Applicant's Name and Street Address:Nuclear Cardiology Systems, Inc.
dba NC Systems, Inc.
5660 Airport Blvd., Suite 101
Boulder, Colorado 80301

Charles H. Rose Contact Person:

T-(303) 541-0044 Telephone and Fax Numbers of Contact Person: 800-548-4024 F-(303) 541-0066

Address of Manufacturing Site: Alsotorokvesz 14 H-1022 Budapest, Hungary

Date of Submission:February 4, 2002

{1}------------------------------------------------

DEVICE NAME 2.

Gamma Camera-SPECT Gamma Camera Device Name (common): CardioSPECT SC and CardioSPECT D90 Proprietary Name: Emission Computed Tomography System Classification Name: (ECT)

Regulatory Class:II
Product Code:90KPS
Regulation Number:21CFR 892.1200
Classification PanelRadiological

IDENTIFICATION OF EQUIVALENT DEVICE ന്

The equivalent device is the:

Re: K013353

Trade/Device Name: NeuroSpect QUAD ECT System Requlation Number: 21CFR 892.1200 Requlation Name: Emission computed tomography system Regulatory Class: II Product Code: 90 KPS

This device, the QUAD, is also manufactured by the applicant, NC Systems, Inc. and has the same technology characteristics as the applicant device.

The equivalent device, the QUAD, is a scintillation detector. Nal, system with PMT light gathering capacity. The system configuration is a four (4) detector system with the mechanical ability to rotate the detectors around the patient. The system uses a collimator on each detector to permit spatial image resolution of the gamma photons emitted from the patient. The collimators lock in place. There is no built-in radioactive scanning source and none is required. The QUAD is capable of both Planar (Class I) and SPECT (ECT) (Class II) operation. The acquisition/processing computer is a presently approved system sold and supported by NC Systems, Inc. but manufactured by the Segami Corporation.

{2}------------------------------------------------

4、DEVICES SUBJECT TO PRE-MARKET NOTIFICATION (APPLICANT DEVICE)

The applicant device subject to this pre-market notification submission are the SC and D90 ECT systems which are submitted under a single application as they are equivalent devices with substantially equivalent specifications, safety and intended use.

  • The SC and D90 are scintillation detector, Nal, systems with PMT a) light gathering capacity. The D90 has two (2) detectors and the SC has one detector both with the ability to rotate these detectors around the patient. The system uses a collimator on each detector to permit spatial image resolution of the gamma photons emitted from the patient. The collimators lock in place. There is no built-in radioactive scanning source and none is required. The SC and D90 are capable of both Planar (Class I) and SPECT or ECT (Class ii) operation. The acquisition/processing computer is a presently approved system sold and supported by NC Systems, Inc. but manufactured by Segami Corporation. This application is for the ECT device-qamma camerasystem(s) and not the previously approved computer.
  • The electrical and mechanical components as well as the b) manufacturing specifications of the applicant systems, the SC and D90 are substantially equivalent to the QUAD, the legally marketed device.
  • Specifications: Are provided in Attachment 2. c)

{3}------------------------------------------------

INTENDED USE 5.

The intended use of the CardioSPECT SC and the CardioSPECT D90 is to detect the location and distribution of gamma ray emitting radionuclides in the human body and store the data for analysis as the same equivalent device, K013353.

DETERMINE OF SUBSTANTIAL EQUIVALENCE ર.

The SC and D90 systems have been compared to the legally marketed QUAD system and determined to be substantially equivalent.

  • a) The specifications are compared on Enclosure G.a.
  • b) The Hazard Analysis is compared on Enclosure 6.b pgs 1-2.
  • c) The labeling, technology, engineering, performance and materials are compared on Enclosure 6.c.
  • d) References to Mediso have been previously addressed as referring to the manufacturing of NC Systems.
  • e) The non-clinical tests submitted were performed on the applicant systems, the SC and D90 in the same manner as the pre-approved QUAD system. These tests were performed under the same standards as the pre-approved QUAD systems.
  • The clinical tests submitted were performed on, and clearly identified, f) the SC and D(). The subjects were adults who were under the care of a physician licensed to perform these procedures. These procedures were performed as requested for the application.

{4}------------------------------------------------

  • g) The conclusion of the non-clinical and clinical tests demonstrate that the device (SC & D90) is safe, as effective and performs well or better than the legally marketed device identified in paragraph 3.(a) of this section.
  • h) The differences between the legally marketed system QUAD and the applicant systems SC and D() are in the number of detectors used for data acquisition. The change in the number of detectors:

QUAD 4

SC 1

D90 2

permit changes in the area imaged and the image acquisition time. These changes do not alter the safety and effectiveness of the systems.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem that features a stylized image of an eagle with three human profiles incorporated into its design. The eagle's head and neck are clearly visible, while the bodies of the three human profiles are suggested by the curves and lines of the eagle's body.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 1 7 2002

Nuclear Cardiology Systems, Inc. % Mr. Wolfram Gmelin Technical Manager, Third Party Review TÜV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470

Re: K021823

Trade/Device Name: CardioSPECT SC and CardioSPECT D90 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system

Regulatory Class: II Product Code: 90 KPS Dated: May 30, 2002 Received: June 4, 2002

Dear Mr. Gmelin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{6}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Page 1 of 2

U.S. Food and Drug Administration - Genter for Devices and Radiological Health

Page _ of _ of _

510(k) Number (if known): KO21923

Device Name: CARDIOSPECT SC AND CARDIOSPECT D 90

Indications for Use:

Intended use is to detect & image the distribution of radionuclides in the body, when the radiopharmaceutical is administered in the body .

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurence of CDRH, Office of Device Evaulation (ODE)
(Division Sign-Off)
Deproductive Abdominal

prescription use $\checkmark$

Division of Reproductive, Abdomin and Radiological Devices 510(k) Number -

§ 892.1200 Emission computed tomography system.

(a)
Identification. An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.(b)
Classification. Class II.