K Number
K021411

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-05-29

(26 days)

Product Code
Regulation Number
862.3280
Age Range
18 - 65
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Liquichek™ Urine Toxicology Control is intended for use as quality control urine to monitor the performance of laboratory urine toxicology screening procedures.

Device Description

Liquichek™ Urine Toxicology Control Levels S1, S2 and S3 are prepared from human urine with added constituents of animal origin, drugs, drug metabolites, preservatives, and stabilizers. The control is provided in liquid form for convenience.

AI/ML Overview

The provided document describes the safety and effectiveness of the Liquichek™ Urine Toxicology Control Levels S1, S2, and S3. This product is a quality control urine intended to monitor the performance of laboratory urine toxicology screening procedures.

The study primarily focuses on demonstrating the stability and shelf life of the device rather than a direct performance comparison against an external standard or clinical outcome. The acceptance criteria and "device performance" in this context refer to the product's ability to maintain its specified characteristics over time under various storage conditions.

Here's a breakdown of the requested information based on the provided text:


1. Table of Acceptance Criteria and Reported Device Performance

Note: The document does not present "acceptance criteria" in a typical quantitative clinical trial sense (e.g., sensitivity, specificity, accuracy against a gold standard) for a diagnostic device. Instead, it focuses on the stability of the control material over time. The "performance" is the successful demonstration of these stability claims.

Acceptance Criteria (Implied by Study Claims)Reported Device Performance
Open Vial Stability: All analytes stable for 30 days when stored tightly capped at 2-8°C.Successfully demonstrated stability for 30 days at 2-8°C.
Shelf Life: All analytes stable for 24 months when stored at 2-8°C.Successfully demonstrated stability for 24 months at 2-8°C (with ongoing real-time studies).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document does not explicitly state the number of individual control vials or batches tested in the stability studies. It broadly states "Stability studies have been performed."
  • Data Provenance: Not explicitly stated. The studies were conducted by Bio-Rad Laboratories, likely internally. The country of origin of the data is not specified, nor is whether the data is retrospective or prospective, though "real-time studies will be ongoing" for shelf-life suggests prospective data collection.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable. This device is a quality control material, not a diagnostic test requiring expert interpretation of results to establish ground truth. The "ground truth" for the stability studies would be the known concentration of analytes in the control material and their measured concentrations over time.
  • Qualifications of Experts: N/A.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. The studies performed are chemical stability studies of a control material rather than an evaluation requiring clinical adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No. This type of study is not relevant for a quality control material.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Was a standalone study done? This question is not directly applicable in the typical sense for an "algorithm." However, the stability studies can be considered "standalone" in that they evaluate the intrinsic characteristics of the control material (analyte concentration over time) without human interpretation being the primary variable. The performance measured is the physical/chemical stability of the product itself.

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for the stability studies is analytical measurements of analyte concentrations over time. The "ground truth" concentrations for the control material are established during its manufacture and characterized by Bio-Rad Laboratories. The stability studies then monitor if these known concentrations remain within acceptable limits. There is no "expert consensus," "pathology," or "outcomes data" involved in establishing the ground truth for this type of product's stability.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable. This product is a quality control material, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set Was Established: Not applicable, as there is no training set for this device.

{0}------------------------------------------------

MAY 29 2002

K02/411

Summary of Safety and Effectiveness Liquichek™ Urine Toxicology Control Levels S1, S2 and S3

1.0 Submitter

Bio-Rad Laboratories 9500 Jeronimo Road, Irvine, California 92618-2017 Telephone: (949) 598-1200 (949) 598-1555 Fax:

Contact Person

Ofelia Cachola Regulatory Affairs Specialist Telephone: (949) 598-1287

Date of Summary Preparation

May 1, 2002

2.0 Device Identification

Product Trade Name:Liquichek™ Urine Toxicology Control LevelsS1, S2 and S3
Common Name:Drug Mixture Controls
Classifications:Class I
Product Code:91DIF
Regulation Number:CFR 862.3280

Device to Which Substantial Equivalence is Claimed 3.0

Liquichek™ Urine Toxicology Control Bio-Rad Laboratories Irvine, California

Docket Number: K991558

4.0 Description of Device

Liquichek™ Urine Toxicology Control Levels S1, S2 and S3 are prepared from human urine with added constituents of animal origin, drugs, drug metabolites, preservatives, and stabilizers.

The control is provided in liquid form for convenience.

{1}------------------------------------------------

5.0 Statement of Intended Use

Liquichek™ Urine Toxicology Control is intended for use as quality control urine to monitor the performance of laboratory urine toxicology screening procedures.

Comparison of the new device with the Predicate Device 6.0

The new control claims substantial equivalence to the Liquichek Urine Toxicology Control (K991558).

Table 1. Similarities and Differences between new and predicate device.

CharacteristicsBio RadLiquichek™ Urine ToxicologyControl(New Device)Bio RadLiquichek™ Urine ToxicologyControl(Predicate Device)
Similarities
Intended UseLiquichek™ Urine ToxicologyControl is intended for use asquality control urine to monitor theperformance of laboratory urinetoxicology screening procedures.Liquichek™ Urine ToxicologyControl is intended for use asquality control urine to monitor theperformance of laboratory urinetoxicology screening procedures.
LevelsSame as predicate device.Level S1= Drugs added atconcentrations20-25% belowimmunoassay cutoffs.Level S2= Drugs added atconcentrations 20-25%above immunoassaycutoffs.Level S3= Elevated immunoassayControl.
FormLiquidLiquid
MatrixHuman urineHuman urine
Storage(Unopened)2-8° Cuntil expiration date2-8° Cuntil expiration date
Open VialClaim2-8° C for 30 days.2-8° C for 30 days.
Differences
AnalytesSame analytes as the predicatedevice with the additional claimsD-Amphetamine, Secobarbital,Nordiazepam, 11-Nor-Δ-9-THC-9-
for Creatinine, pH and SpecificGravity.COOH, Benzoylecgonine, Ethanol,Lysergic Acid Diethylamide (LSD),Methadone, Methaqualone,Nortriptyline, Morphine-(Free),Phencyclidine, Propoxyphene.

06_SSE461_S1UrnTx

{2}------------------------------------------------

2.0 Statement of Supporting Data

Stability studies have been performed to determine the open vial stability and shelf life for the Liquichek™ Urine Toxicology Control. Product claims are as follows:

  • 2.1 Open vial: Once the control is opened, all analytes will be stable for 30 days when stored tightly capped at 2-8°C.
  • 2.2 Shelf Life: 24 months when stored at 2-8°C

Real time studies will be ongoing to support the shelf life of this product.

All supporting data is retained on file at Bio-Rad Laboratories.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white emblem for the U.S. Department of Health & Human Services. The emblem features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Elizabeth Platt Regulatory Affairs Manager Bio-Rad Laboratories, QSD 9500 Jeronimo Road Irvine, California 92618-2017

MAY 2 9 2002

Re: K021411

Trade/Device Name: Liquichek™ Urine Toxicology Control Levels S1, S2 and S3 Regulation Number: 21 CFR § 862.3280 Regulation Name: Clinical Toxicology Control Material Regulatory Class: I Product Code: DIF Dated: May 1, 2002 Received: May 3, 2002

Dear Ms. Platt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Ko2141 510 (k) Number (if known):_

Device Name:

Liquichek™ Urine Toxicology Control Levels S1, S2 and S3

Indications for Use:

A quality control urine to monitor the performance of laboratory urine toxicology screening procedures.

Alan Cooper

ision of Clinical Laboratory Devices

510(k) Number

(PLEASE DO NOT WRITE BELOW THE LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription useuseX
or
Over-the Counter

§ 862.3280 Clinical toxicology control material.

(a)
Identification. A clinical toxicology control material is a device intended to provide an estimation of the precision of a device test system and to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. This generic type of device includes various single, and multi-analyte control materials.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.