Intended Use

The 2120is Maternal/Fetal Monitoring System is intended for monitoring fetal and maternal vital signs: fetal heart rate; fetal heart rate high/low/poor signal quality alarms; maternal uterine activity; heart/pulse rate, blood pressure, nondiagnostic maternal ECG and %SpO2. Optional fetal movement detection is available.

A Corometrics 2120is Maternal/Fetal Monitoring System is used for non-invasive and invasive monitoring of the fetus during the antepartum period as well as throughout labor and delivery (i.e. fetal heart rate and uterine activity monitoring). Fetal movement detection and fetal heart rate alarm options (user selectable high/low and poor signal quality alarms) are available.

A Corometrics 2120is Maternal/Fetal Monitoring System is intended for monitoring maternal vital signs to help assess maternal well-being. The vital signs which can be measured with these monitor configurations are summarized as follows:

NOTE: Maternal vital signs provided by the monitor should only be used as an adjunct in patient assessment in conjunction with clinical signs and symptoms.

Blood Pressure: The monitor is intended for use in the non-invasive monitoring of maternal blood pressure (NBP). This monitor is not intended for use in the neonatal or pediatric blood pressure monitoring.

Pulse Oximetry. The monitor is intended for use in the non-invasive monitoring of maternal functional oxygen saturation of arterial hemoglobin (MSpO2).

Heart/Pulse Rate. The monitor is intended for use in the non-invasive monitoring of the maternal heart/pulse rate. Additionally, an MECG waveform "snapshot" may be displayed and printed.

NOTE: Only the maximum configuration provides both maternal heart rate and pulse rate data.

Bed-to-bed surveillance is available when 2120is Series Monitors are networked together.

Device Description

The 2120is Maternal/Fetal Monitoring System consists of the following features/options that can be available in multiple configurations:

  • fetal heart rate (via Doppler Ultrasound of FECG)
  • maternal uterine activity (via intrauterine pressure catheter or tocotransducer
  • fetal movement detection
  • maternal non-invasive blood pressure (clinician prompted or automatic)
  • maternal pulse oximetry
  • maternal heart/pulse rate (MECG) and ECG waveform "snapshot"

The 2120is is a full-featured maternal/fetal monitor. The device is also capable of providing bed-to-bed surveillance when multiple 2120is monitors are connected via an Ethernet network. The monitor is also capable of acting as a bedside terminal to the QS system. In this capacity, the 2120is monitor provides the features of a standard PC QS bedside terminal.

AI/ML Overview

The provided text describes the 510(k) summary for the Corometrics Model 2120is Maternal/Fetal Monitor. It outlines the device description, intended use, and indicates that the device's technology is the same as its predicate devices. However, the document does not include specific acceptance criteria or a detailed study proving the device meets those criteria in the way typically found for AI/ML-based medical devices.

Instead, the submission focuses on demonstrating substantial equivalence to a predicate device through:

  • Compliance with voluntary standards: This is a common approach for traditional medical devices.
  • Internal validation processes: Requirements specification review, code inspections, software and hardware testing, safety testing, environmental testing, and final validation.

Therefore, the requested information cannot be fully provided as the submission does not contain details about:

  • A table of acceptance criteria with reported performance metrics like sensitivity, specificity, or AUC.
  • Sample sizes for test sets, data provenance, or expert ground truth establishment for performance evaluation.
  • Adjudication methods, MRMC studies, or standalone algorithm performance.
  • Training set details or ground truth establishment for a machine learning model.

This is consistent with the nature of a 510(k) submission for a non-AI/ML device where the focus is on demonstrating substantial equivalence to a legally marketed predicate through adherence to standards and robust internal development/testing processes, rather than a prospective clinical performance study with specific metrics.

Summary of what can be extracted/inferred:

Information RequestedDetails from Document
1. Table of Acceptance Criteria and Reported Device PerformanceNot explicitly stated in terms of quantitative performance metrics (e.g., sensitivity, specificity, accuracy). The acceptance criteria are implicitly met by: - Compliance with voluntary standards. - Successful completion of internal validation steps: Requirements specification review, Code inspections, Software and hardware testing, Safety testing, Environmental testing, Final validation. The reported performance is that the device "is as safe, as effective, and performs as well as the predicate device" (Corometrics Model 2120is Maternal/Fetal Monitor and Masimo SET® Pulse Oximetry).
2. Sample size for test set and data provenanceNot applicable/not provided. The submission focuses on device engineering, safety, and functional testing rather than a clinical performance study with a test set of patient data for an algorithm.
3. Number of experts and qualifications for ground truth (test set)Not applicable/not provided.
4. Adjudication method (test set)Not applicable/not provided.
5. MRMC comparative effectiveness study and effect sizeNo. This is not an AI/ML device, so such a study is not relevant or performed.
6. Standalone (algorithm only) performance studyNo. This is a hardware monitor, not an algorithm.
7. Type of ground truth usedFor the functional and safety testing, the "ground truth" would be established by engineering specifications, regulatory standards, and expected physiological responses, verified through direct measurement and comparison to known accurate instruments (e.g., during calibration and validation for SpO2 or NIBP).
8. Sample size for the training setNot applicable/not provided. This is not an AI/ML device that requires a "training set" in the machine learning sense.
9. How the ground truth for the training set was establishedNot applicable/not provided.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness

Date:April 8, 2002
Submitter:GE Medical Systems Information Technologies61 Barnes Park Road NorthWallingford, CT 06492 USA
Contact Person:Joelle NeiderRegulatory Affairs SpecialistGE Medical Systems Information TechnologiesPhone: 203-949-8232Fax: 414-918-8113
Device: Trade Name:Corometrics Model 2120is Maternal/Fetal Monitor
Common/Usual Name:Maternal/Fetal Monitor
Classification Names:21 CFR 884.2740 System, Monitoring, Perinatal
Predicate Devices:K012718 Corometrics Model 2120is Maternal/Fetal MonitorK990966 Masimo SET® 2000 Pulse Oximeter and LNOP® series ofsensors and cables
Device Description:The 2120is Maternal/Fetal Monitoring System consists of the followingfeatures/options that can be available in multiple configurations:fetal heart rate (via Doppler Ultrasound of FECG)maternal uterine activity (via intrauterine pressure catheter ortocotransducerfetal movement detectionmaternal non-invasive blood pressure (clinician prompted orautomatic)maternal pulse oximetrymaternal heart/pulse rate (MECG) and ECG waveform "snapshot"
The 2120is is a full-featured maternal/fetal monitor. The device is alsocapable of providing bed-to-bed surveillance when multiple 2120ismonitors are connected via an Ethernet network. The monitor is alsocapable of acting as a bedside terminal to the QS system. In thiscapacity, the 2120is monitor provides the features of a standard PC QSbedside terminal.
Intended Use:The 2120is Maternal/Fetal Monitoring System is intended for monitoringfetal and maternal vital signs: fetal heart rate; fetal heart ratehigh/low/poor signal quality alarms; maternal uterine activity;heart/pulse rate, blood pressure, nondiagnostic maternal ECG and%SpO2. Optional fetal movement detection is available.
Technology:The Corometrics Model 2120is Maternal/Fetal Monitor employs thesame fundamental scientific technology as the predicate devices.

8

{1}------------------------------------------------

KOZ1135 page 4/2

The Model 2120is Maternal/Fetal Monitor complies with the voluntary Test Summary: The Model 212016 Maternair to of this submission. The following Standards as detailou in Society of the development of the Model 2120is:

  • Requirements specification review .
  • Code inspections .
  • Software and hardware testing .
  • Safety testing .
  • Environmental testing .
  • Final validation .

Conclusion:

The results of these measurements demonstrated that the Model The Tesults of these Thousarentions SET Pulse Oximetry) is as safe, as effective, and performs as well as the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the department's mission to promote health, well-being, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 0 9 2002

Mr. Joelle Neider Corporate Regulatory Affairs GE Medical Systems Information Technologies 61 Barnes Park Road North P.O. Box 333 WALLINGFORD CT 06492-0333 Re: K021135

Trade/Device Name: Corometrics Model 2120is Maternal/Fetal Monitor Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II

Product Code: 85 HGM Dated: April 8, 2002 Received: April 9, 2002

Dear Mr. Neider:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known):

510(k) filed on April 8, 2002

Device Name:

Corometrics Model 2120is Maternal/Fetal Monitoring System

Indications For Use:

Fetal Surveillance | .

A Corometrics 2120is Maternal/Fetal Monitoring System is used for non-invasive and invasive monitoring of the fetus during the antepartum period as well as throughout labor and delivery (i.e. fetal heart rate and uterine activity monitoring). Fetal movement detection and fetal heart rate alarm options (user selectable high/low and poor signal quality alarms) are available.

11 Maternal Monitoring

A Corometrics 2120is Maternal/Fetal Monitoring System is intended for monitoring maternal vital signs to help assess maternal well-being. The vital signs which can be measured with these monitor configurations are summarized as follows:

NOTE: Maternal vital signs provided by the monitor should only be used as an adjunct in patient assessment in conjunction with clinical signs and symptoms.

Blood Pressure: The monitor is intended for use in the non-invasive monitoring of maternal blood pressure (NBP). This monitor is not intended for use in the neonatal or pediatric blood pressure monitoring.

Pulse Oximetry. The monitor is intended for use in the non-invasive monitoring of maternal functional oxygen saturation of arterial hemoglobin (MSpO2).

Heart/Pulse Rate. The monitor is intended for use in the non-invasive monitoring of the maternal heart/pulse rate. Additionally, an MECG waveform "snapshot" may be displayed and printed.

NOTE: Only the maximum configuration provides both maternal heart rate and pulse rate data.

Bed-to-bed surveillance is available when 2120is Series Monitors are networked together.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use/OROver-The-Counter Use
(Per 21 CFR 801.109)(Optional Format 1-2-96)

Nancy Broadon

(Division Sign-Off)

Division of Reproductive, Abdominal,

and Radiological Devices

510(k) NumberK021135
------------------------

31

§ 884.2740 Perinatal monitoring system and accessories.

(a)
Identification. A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.(b)
Classification. Class II (performance standards).