K Number
K020984

Validate with FDA (Live)

Date Cleared
2002-08-14

(149 days)

Product Code
Regulation Number
870.5310
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AccessAED™ semi-automatic external defibrillator is intended to be used for the treatment of cardiac arrest. The AccessAED should be applied to victims who are unconscious, with absence of breathing, and absence of a detectable pulse.

The AccessALS™ has the capability for manual operation by a user trained in advanced cardiac life support (ALS). In manual mode, the user can select the time for rhythm analysis, device charging, energy level, synchronization, and shock delivery. In the semiautomatic mode, AccessALS functions identically to the AccessAED.

The AccessAED and AccessALS are not currently indicated for children less than 8 years old (less than approximately 25 kgs. or 55 lbs.) The use of the AccessAED and AccessALS should be reviewed after each event and all adverse events should be reported.

Device Description

The AccessAED™ Automatic External Defibrillator has been designed specifically for the treatment of ventricular tachyarrhythmias in cardiac arrest. The AccessAED delivers a high-energy 200 or 360J biphasic waveform to patients in cardiac arrest resulting from ventricular fibrillation or high rate ventricular tachycardia.

The AccessAEDTM features include:

  • LED ready indicator .
  • power button .
  • shock delivery button .
  • LCD display with text messaging and ECG display option .
  • defibrillation pads housed in a tray for rapid application to patient .
  • . disposable Lithium battery system
  • rhythm analysis software .
  • energy selection button (manual mode only) ●
  • synchronization button (manual mode only) .
  • . analysis button (manual mode only)
  • charge button (manual mode only) .

When the AccessAED™ algorithm software identifies a rhythm that requires defibrillation, the device charges to the required energy level algorithm (200 J or 360 J) as dictated by the internal algorithm. When the device indicates a full charge and instructs the operator to push the shock button, the operator delivers the shock by pressing the shock delivery button.

AI/ML Overview

The AccessAED Automatic External Defibrillator's acceptance criteria and the study proving it meets these criteria are detailed below. It's important to note that the provided document is a 510(k) summary, which generally focuses on substantial equivalence to predicate devices rather than providing detailed clinical trial results with specific performance metrics against pre-defined acceptance criteria as would be seen in a PMA or extensive clinical study report.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state quantitative acceptance criteria for the AccessAED's performance in a clinical setting (e.g., sensitivity, specificity for rhythm analysis, or first-shock success rates against specific thresholds). Instead, the performance is primarily evaluated based on its compliance with industry standards and the demonstrated equivalence of its waveform to predicate devices.

Performance AspectAcceptance Criteria (Implicit/Standard)Reported Device Performance (as stated or implied)
Biphasic Waveform EfficacySimilar characteristics and results to predicate devices (Heartstream FR2 AED, Physio-Control Biphasic LifePak 500, Zoll M Series Biphasic Defibrillator) and demonstrably effective for defibrillation."The efficacy of the biphasic truncated exponential waveform in this device was demonstrated in a study of swine. The results of this study demonstrate the substantial equivalence of the AccessAED biphasic truncated exponential waveform.""Though there are minor differences in the characteristics of the AccessAED biphasic waveform and its predicate device waveforms, these differences do not raise new questions of safety and efficacy."
Compliance with Industry StandardsCompliance with AAMI DF2-1989 (Cardiac Defibrillator Devices) and AAMI DF39-1993 (Automatic External Defibrillators)."Performance test data is submitted with the 510 (k) documents. These data demonstrate that the device complies with the applicable sections of AAMI DF2-1989 (Cardiac Defibrillator Devices) and AAMI DF39-1993 (Automatic External Defibrillators)."
Design Control & Hardware TestingDeveloped under design control and hardware tested in accordance with established industry standards."The device was developed under design control, and the hardware was tested in accordance with established industry standards."
Rhythm Analysis Software FunctionalityIdentifies rhythms requiring defibrillation and instructs device to charge to appropriate energy (200 J or 360 J).The device charges to 200 J or 360 J "as dictated by the internal algorithm" when the software identifies a rhythm requiring defibrillation. This is a functional description rather than a performance metric.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: The document specifies a "study of swine" for waveform efficacy. However, it does not provide the sample size (i.e., number of swine) used in this study.
  • Data Provenance: The study was conducted on swine (an animal model), not human subjects. The country of origin for the data is not specified. The study appears to be prospective in the context of animal research for device validation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not mention the use of human experts to establish ground truth for the "study of swine." The ground truth for waveform efficacy in an animal study would typically be based on physiological measurements and outcomes directly observed during the experiment. Since it's an animal study focusing on waveform equivalence, the concept of "expert ground truth" as applied to diagnostic image interpretation by clinicians is not directly applicable here.

4. Adjudication Method for the Test Set

Not applicable. As described above, the ground truth was inherently established through the experimental design and physiological measurements in the swine study, not through a consensus of human experts.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, an MRMC comparative effectiveness study involving human readers with and without AI assistance was not conducted or reported in this 510(k) summary. The device is an automated external defibrillator, where the "AI" (rhythm analysis software) makes decisions independently, instructing the user when to deliver a shock, rather than assisting human readers in interpreting data.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the core of the device's function, particularly its rhythm analysis software and energy delivery algorithm, is inherently standalone. The device "identifies a rhythm that requires defibrillation" and "charges to the required energy level algorithm." The human operator's role is to apply the pads, power on the device, and press the shock button when instructed. The "study of swine" for waveform efficacy would be considered a standalone evaluation of the device's fundamental function. The performance data also indicates compliance with AAMI standards, which would include testing of the algorithm's performance in analyzing arrhythmias and delivering shocks.

7. The Type of Ground Truth Used

For the "study of swine" demonstrating waveform efficacy, the ground truth would have been established through physiological measurements and observed outcomes in the swine model, such as successful defibrillation, return of spontaneous circulation, or specific cardiac electrical activity. This is akin to direct experimental observation and measurement.

8. The Sample Size for the Training Set

The document does not provide information about a "training set" sample size. Given this is a 510(k) premarket notification in 2002 for an AED, typical machine learning "training sets" as understood today for AI algorithms were not routinely disclosed or applied in the same manner. The rhythm analysis software would have been developed and validated using a dataset of ECGs, but this is not explicitly described as a "training set" with a specified size.

9. How the Ground Truth for the Training Set Was Established

The document does not provide information on how the ground truth for any potential "training set" for the rhythm analysis software was established. If such a set existed, it would typically involve annotated ECG waveforms by cardiologists or other experts, but this is speculative given the information provided.

{0}------------------------------------------------

Access CardioSystems 510(k) K020984

AUG 1 4 2002

AccessAED Automatic External Defibrillator

510(k) Summary

Submitter's Name and Address:Access CardioSystems150 Baker Avenue ExtensionConcord, MA 01842
Contact Person:David Barash, M.D.978-371-4985
Device Name:AccessAED™ Automatic External Defibrillator

Classification: Automatic External Defibrillator Class III 21 CFR 870.1025

Predicate Devices:

Heartstream FR2 AED (K003565) Physio-Control Biphasic LifePak 500 (K983393) Zoll M Series Biphasic Defibrillator (K990762)

Indications for Use:

The AccessAED™ semi-automatic external defibrillator is intended to be used for the treatment of cardiac arrest. The AccessAED should be applied to victims who are unconscious, with absence of breathing, and absence of a detectable pulse.

The AccessALS™ has the capability for manual operation by a user trained in advanced cardiac life support (ALS). In manual mode, the user can select the time for rhythm analysis, device charging, energy level, synchronization, and shock delivery. In the semiautomatic mode, AccessALS functions identically to the AccessAED.

The AccessAED and AccessALS are not currently indicated for children less than 8 years old (less than approximately 25 kgs. or 55 lbs.) The use of the AccessAED and AccessALS should be reviewed after each event and all adverse events should be reported.

{1}------------------------------------------------

Access CardioSystems 510(k) K020984

AccessAED Automatic External Defibrillator Contraindications for Use

The AccessAED™ semi-automatic external defibrillator is contraindicated for patients who:

  • . are conscious
  • are breathing ●
  • have a detectable pulse ●

Training Requirements

The Access AED is intended for use by emergency responders specifically trained in the operation of the Access AED. Users should have, at the minimum, BCLS/AED training equivalent to that recommended by the American Heart Association or American Red Cross or training in other physician-authorized emergency medical response.

Device Description:

The AccessAED™ Automatic External Defibrillator has been designed specifically for the treatment of ventricular tachyarrhythmias in cardiac arrest. The AccessAED delivers a high-energy 200 or 360J biphasic waveform to patients in cardiac arrest resulting from ventricular fibrillation or high rate ventricular tachycardia.

The AccessAEDTM features include:

  • LED ready indicator .
  • power button .
  • shock delivery button .
  • LCD display with text messaging and ECG display option .
  • defibrillation pads housed in a tray for rapid application to patient .
  • . disposable Lithium battery system
  • rhythm analysis software .
  • energy selection button (manual mode only) ●
  • synchronization button (manual mode only) .
  • . analysis button (manual mode only)
  • charge button (manual mode only) .

When the AccessAED™ algorithm software identifies a rhythm that requires defibrillation, the device charges to the required energy level algorithm (200 J or 360 J) as dictated by the internal algorithm. When the device indicates a full charge and instructs the operator to push the shock button, the operator delivers the shock by pressing the shock delivery button.

{2}------------------------------------------------

Access CardioSystems 510(k) K020984

AccessAED Automatic External Defibrillator Substantial Equivalence:

The AccessAED™ Automatic External Defibrillator is substantially equivalent to other marketed devices delivering a biphasic waveform, specifically the Zoll M Series Biphasic Defibrillator and the Physio-Control Biphasic Lifepak 500, and the HeartStream FR2. The biphasic waveform in the AccessAED™ delivers a biphasic waveform with characteristics and results similar to both the predicate devices. The operation of this device and any other subtle or minor differences between this device and its predicate devices do not raise any new questions regarding safety and efficacy.

Performance Data:

Performance test data is submitted with the 510 (k) documents. These data demonstrate that the device complies with the applicable sections of AAMI DF2-1989 (Cardiac Defibrillator Devices) and AAMI DF39-1993 (Automatic External Defibrillators). The device was developed under design control, and the hardware was tested in accordance with established industry standards.

The efficacy of the biphasic truncated exponential waveform in this device was demonstrated in a study of swine. The results of this study demonstrate the substantial equivalence of the AccessAED biphasic truncated exponential waveform. Though there are minor differences in the characteristics of the AccessAED biphasic waveform and its predicate device waveforms, these differences do not raise new questions of safety and efficacy.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is centered on the image and is the only element present. The words are arranged in three lines.

Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the first profile being the most prominent.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 1 4 2002

David Barash, M.D. Executive Vice President and Medical Director Access CardioSystems, Inc. 150 Baker Avenue Extension, Suite 108 Concord, MA 01742

Re: K020984

Trade Name: AccessAED™ Semi-Automatic External Defibrillators (Models Access AED™ and Access ALSTM) Regulation Number: 21 CFR 870.1025, 21 CFR 870.5300 Regulation Name: Automatic External Defibrillator Regulatory Class: Class III (three) Product Code: MKJ, LDD Dated: July 2, 2002 Received: July 30, 2002

Dear Dr. Barash:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - David Barash, M.D

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Alise Doty M.D.

Bram D. Zuckerman, M.D) Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): __K020984

Device Name:

Indications For Use:

Indications for Use

The AccessAED™ semi-automatic external defibrillator is intended to be used for the treatment of cardiac arrest. The AccessAED should be applied to victims who are unconscious, with absence of breathing, and absence of a detectable pulse.

The AccessALS™ has the capability for manual operation by a user trained in advanced cardiac life support (ALS). In manual mode, the user can select the time for rhythm analysis, device charging, energy level, synchronization, and shock delivery. In the semiautomatic mode, AccessALS functions identically to the AccessAED.

The AccessAED and AccessALS are not currently indicated for children less than 8 years old (less than approximately 25 kgs. or 55 lbs.) The use of the AccessAED and AccessALS should be reviewed after each event and all adverse events should be reported.

Contraindications for Use

The AccessAED™ semi-automatic external defibrillator is contraindicated for patients who:

  • are conscious ●
  • are breathing ●
  • have a detectable pulse .

Training Requirements

The Access AED is intended for use by emergency responders specifically trained in the operation of the Access AED. Users should have, at the minimum, BCLS/AED training equivalent to that recommended by the American Heart Association or American Red Cross or training in other physician-authorized emergency medical response.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Prescription UseX
(Per 21 CFR 801.109)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular and Respiratory Devices
510(k) NumberK020984

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.