Intended Use

The Medcomp Pediatric Ash Split Cath™ XL is indicated for use in attaining longterm vascular access for hemodialysis and apheresis in pediatric, adolescent, and adult patients as determined by the prescribing physician. It may be inserted percutaneously and is primarily placed in the internal jugular vein. Alternate insertion sites include the subclavian and femoral vein.

Device Description

The Medcomp Pediatric Ash Split Cath™ XL is a 10F polyurethane, double lumen catheter used to remove and return blood through two-segregated lumen passages. Both lumens are "D" shaped, tapered at the distal tip, with six side holes. The distal venous lumen extends beyond the arterial lumen to reduce recirculation. The fixed polvester cuff allows for tissue ingrowth for long term placement. A polyurethane sleeve is positioned between the hub to cuff location. The lumens are connected to the extensions via a soft pliable hub with suture wing. Red and blue luer connectors and clamps identify the arterial and venous extensions. The clamps incorporate I.D. Rings which indicate priming volume and site care information.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Medcomp Pediatric Ash Split Cath™ XL:

The provided document (K020936) is a 510(k) summary for a medical device. In the context of a 510(k), the "acceptance criteria" and "device performance" are typically demonstrated through substantial equivalence to a legally marketed predicate device. This means the new device is as safe and effective as the predicate based on performance data.

Key takeaway: This submission relies on in-vitro performance data to establish substantial equivalence to predicate devices, and clinical studies were not deemed necessary. Therefore, many of the requested categories related to clinical studies, expert-established ground truth, and human reader performance are not applicable or cannot be answered from the provided text.


1. Table of Acceptance Criteria and Reported Device Performance

Given that this is a 510(k) for a product line extension relying on substantial equivalence, the "acceptance criteria" are implied to be "meeting or exceeding the performance of the predicate device" in relevant in-vitro tests. The "reported device performance" is that it does meet these criteria.

Acceptance Criteria (Implied)Reported Device Performance
Performance equivalent to legally marketed predicate devices in:Device demonstrates substantial equivalence to legally marketed devices in:
- Tensile strength- Tensile strength
- Joint strength- Joint strength
- Leakage- Leakage
- Recirculation- Recirculation
- Flow performance- Flow performance
- Mechanical hemolysis- Mechanical hemolysis

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document refers to "in-vitro performance data," which implies laboratory testing of various samples of the device. The exact number of units tested for each in-vitro parameter (tensile strength, leakage, etc.) is not provided.
  • Data Provenance: The data is from in-vitro testing and not from human subjects. The country of origin for the testing itself is not specified, but the submitter (Medcomp) is based in the USA (Harleysville, PA). The data is considered prospective in the sense that it was generated specifically for this 510(k) submission, even though it's laboratory data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not Applicable. This device did not involve human clinical studies or the establishment of ground truth by medical experts in the way that, for example, an AI diagnostic algorithm would. The "truth" for the in-vitro tests is based on measurable physical properties and engineering standards relative to the predicate device.

4. Adjudication Method for the Test Set

  • Not Applicable. See point 3. No adjudication method for expert consensus was needed as there were no clinical studies where expert opinions would need to be reconciled.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study was not done. The submission explicitly states: "Clinical studies were not deemed necessary since in-vitro testing was sufficient to demonstrate safety and effectiveness by way of comparison to legally marketed predicate devices." Therefore, there is no information on how human readers might improve with or without AI assistance, as AI is not a component of this device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • Not Applicable. This is a physical medical device (catheter), not a software algorithm. Therefore, "standalone" algorithm performance is not relevant.

7. The Type of Ground Truth Used

  • Engineering/Performance Standards and Predicate Device Data: The "ground truth" for this device's performance is established by demonstrating that its in-vitro performance characteristics (tensile strength, joint strength, leakage, recirculation, flow performance, mechanical hemolysis) are substantially equivalent to those of legally marketed predicate devices. This relies on established scientific and engineering principles for material and device performance.

8. The Sample Size for the Training Set

  • Not Applicable. As this is not a machine learning or AI device, there is no concept of a "training set."

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. See point 8.

{0}------------------------------------------------

KO2 ©936

510(k) SUMMARY

A. Submitter Information:

Submitter:MEDCOMP®1499 Delp DriveHarleysville, PA 19438(215) 256-4201 Telephone(215) 256-1787 Fax
Contact:Florence A. Caikoski
Date Prepared:June 25, 2002
OCT 11 2002
Trade Name:Medcomp Pediatric Ash Split Cath™ XL
Common Name:Hemodialysis Catheter, Implanted
Classification:78 MSD
C.F.R. Section:876.5540
Predicate Device:K013162 Medcomp Ash Split Cath™W/Strain ReliefK984022 Arrow Pediatric Two-Lumen

D. Device Description:

B.

C.

The Medcomp Pediatric Ash Split Cath™ XL is a 10F polyurethane, double lumen catheter used to remove and return blood through two-segregated lumen passages. Both lumens are "D" shaped, tapered at the distal tip, with six side holes. The distal venous lumen extends beyond the arterial lumen to reduce recirculation. The fixed polvester cuff allows for tissue ingrowth for long term placement. A polyurethane sleeve is positioned between the hub to cuff location.

Hemodialysis Catheter

The lumens are connected to the extensions via a soft pliable hub with suture wing. Red and blue luer connectors and clamps identify the arterial and venous extensions. The clamps incorporate I.D. Rings which indicate priming volume and site care information.

E. Intended Use:

The Medcomp Pediatric Ash Split Cath™ XL is indicated for use in attaining longterm vascular access for hemodialysis and apheresis in pediatric, adolescent, and adult patients as determined by the prescribing physician. It may be inserted percutaneously and is primarily placed in the internal jugular vein. Alternate insertion sites include the subclavian and femoral vein.

{1}------------------------------------------------

K02 0936

F. Comparison to Predicate Device:

The proposed device is a product line extension to the Medcomp legally marketed device, and is identical in design and materials.

The modifications include:

  • Pediatric population indication for use ●
  • Change from 14F to 10F outer diameter ●
  • 18cm length .
  • Femoral insertion site .

G. Performance Data:

In-vitro performance data for the proposed device including tensile strength, joint strength, leakage, recirculation, flow performance, and mechanical hemolysis demonstrates this device is substantially equivalent to legally marketed devices intended for hemodialysis and apheresis treatments.

Clinical studies were not deemed necessary since in-vitro testing was sufficient to demonstrate safety and effectiveness by way of comparison to legally marketed predicate devices.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Florence Caikoski Regulatory Affairs Associate MedComp® 1499 Delp Drive HARLEYSVILLE PA 19438

Re: K020936

Trade/Device Name: 10F x 18cm Pediatric Ash Split-Cath™ XL, Model ASPC18P-XL, and 10F x 24cm Pediatric Ash Split-Cath™ XL, Model ASPC24P-XL Regulation Number: 21 CFR 8876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: 78 MSD Dated: July 15, 2002 Received: July 16, 2002

Dear Ms. Caikowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Florence Caikoski

· Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours.

David li. Seymon

hr

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

510(k) Number: K020936

Device Name: Medcomp Pediatric Ash Split Cath™ XL

Indications for use:

The Medcomp Pediatric Ash Split Cath™ XL is indicated for use in attaining longterm vascular access for hemodialysis and apheresis in pediatric, adolescent, and adult patients as determined by the prescribing physician. It may be inserted percutaneously and is primarily placed in the internal jugular vein. Alternate insertion sites include the subclavian and femoral vein.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use(Per 21 CFR 801.109)
OROver-The-Counter

(Optional Format 1-2-96)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K0209361 03

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.