K Number
K020889

Validate with FDA (Live)

Date Cleared
2002-06-05

(78 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Endoscopic Light Source XL202/L3 is intended to be used with fiber optic endoscopes to provide illumination of body cavities, hollow organs and canals during endoscopic procedures.

The device is classified as Cardiac Floating (CF) which allows the use in endoscopic cardiac procedures when used in conjunction with the proper instrumentation.

Device Description

The Endoscopic Light Source XL202/L3 uses a 180 W xenon lamp to provide illumination during endoscopic surgery through a fiber optic cable, which is connected to the device. Furthermore, the light source can be equipped with an additional 150 W halogen lamp for diagnostic applications in endoscopy. The 150 W halogen lamp also function as a backup up lamp to avoid interrupting a procedure in case of a malfunction of the xenon lamp. The color temperature of the xenon lamp and halogen lamp is approximately 6000 % and 3400 % respectively, the lamp life approximately 500 hours and 50 h respectively.

AI/ML Overview

The provided text is for a 510(k) summary for an Endoscopic Light Source. This type of document is for a medical device that provides illumination and is not an AI/ML powered device. Therefore, many of the requested categories related to AI/ML device studies and performance will not be applicable.

Here's an analysis of the provided text in relation to your request:

Acceptance Criteria and Device Performance Study for Endoscopic Light Source XL202/L3

The 510(k) summary for the Endoscopic Light Source XL202/L3 does not describe a study involving specific acceptance criteria and performance metrics in the way one would for an AI/ML algorithm or a new diagnostic device requiring clinical performance evaluation. Instead, the "performance data" section focuses on compliance with established electrical safety and general medical device standards, and the demonstration of "substantial equivalence" to predicate devices.

The underlying principle for the approval of this device is substantial equivalence, meaning it performs as safely and effectively as other legally marketed devices.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria CategorySpecific Criteria/StandardReported Device Performance
Electrical SafetyIEC 601-1Complies
Electromagnetic Compatibility (EMC)IEC 601-1-2Complies
General Medical Device DirectiveMedical Device Directive 93/42/EECConforms
Underwriters Laboratories StandardUL2601-1Meets requirements
Intended Use EquivalenceProvide illumination of body cavities, hollow organs and canals during endoscopic procedures.Similar in design and technological characteristics to predicate devices; intended use is identical.
Safety and EffectivenessNo new questions of safety and effectiveness compared to predicates.Differences between the XL202/L3 and predicate devices are minor and raise no new questions of safety and effectiveness.
Illumination Source (Xenon)180W Xenon lamp180W xenon lamp, color temperature approx. 6000K, lamp life approx. 500 hours.
Illumination Source (Halogen - Backup)150W Halogen lamp150W halogen lamp, color temperature approx. 3400K, lamp life approx. 50 hours. Also functions as backup.

2. Sample Size Used for the Test Set and Data Provenance:

This information is not applicable as the document describes a traditional medical device (light source) without a test set of data in the context of an AI/ML or diagnostic performance study. The "testing" referred to is compliance testing against established engineering and safety standards.

3. Number of Experts Used to Establish Ground Truth and Qualifications:

This information is not applicable. Ground truth is not established in the clinical or diagnostic sense for a light source. Device functionality is assessed against engineering specifications and safety standards.

4. Adjudication Method for the Test Set:

This information is not applicable. There is no "test set" in the context of clinical or diagnostic performance that would require an adjudication method.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

This information is not applicable. The device is an endoscopic light source, not a diagnostic or AI-assisted interpretation tool. Therefore, a study to measure human reader improvement with or without AI assistance is irrelevant.

6. Standalone (Algorithm Only) Performance Study:

This information is not applicable. This is a hardware device (light source), not an algorithm.

7. Type of Ground Truth Used:

This information is not applicable in the context of clinical or diagnostic ground truth. For this type of device, the "ground truth" is defined by compliance with established engineering standards (e.g., light output specifications, electrical safety parameters) and the device's ability to perform its intended function (providing illumination) as reliably as predicate devices.

8. Sample Size for the Training Set:

This information is not applicable. There is no "training set" for a hardware medical device like a light source.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable. As there is no training set, there is no ground truth establishment for it.

In summary: The provided 510(k) summary focuses on demonstrating that the Endoscopic Light Source XL202/L3 meets safety and performance standards equivalent to existing, legally marketed predicate devices through compliance testing and comparison of technical specifications. It is not an AI/ML-driven device or a device requiring a clinical performance study with a test set, ground truth, or expert adjudication.

{0}------------------------------------------------

K020889

Poglogy

510(K) SUMMARY Endoscopic Light Source XL202/L3

L Submitter:

WORLD OF MEDICINE Lemke GmbH Danziger Strasse 21 82194 Gröbenzell Germany

II. Device Names:

  • Classification Name: 1.
  • Common or Usual Name: 2.
  • Proprietary Name: 3.

Accessory to an Endoscope Endoscopic Light Source Endoscopic Light Source XL202/L3

III. Classification:

Class II. This device is described in 21 C.F.R. § 876.1500. The product code for the device is GCT.

IV. Predicate Devices:

  • Karl Storz Xenon Light Source Model 201320-20 (K934559) manufactured t by Karl Storz Imaging, Inc.
  • Xenon Light Source Auto Lip 5123. Model 2123.011 (K983628) . manufactured by Richard Wolf Medical Instruments Corp.
  • Battery Powered Endoscopic Light Source (K960081)manufactured by 0 Mitsubishi Cable America, Inc.
  • LH-150 manufactured by Pentax Precision Instrument Corp. .
  • LH SC Halogen Light Source manufactured by Olymous America, Inc. t

V. Intended Use:

The Endoscopic Light Source XL202/L3 is intended to be used with fiber optic endoscopes to provide illumination of body cavities, hollow organs and canals during endoscopic procedures.

The device is classified as Cardiac Floating (CF) which allows the use in endoscopic cardiac procedures when used in conjunction with the proper instrumentation.

VI. Device Description:

The Endoscopic Light Source XL202/L3 uses a 180 W xenon lamp to provide illumination during endoscopic surgery through a fiber optic cable, which is

{1}------------------------------------------------

connected to the device. Furthermore, the light source can be equipped with an additional 150 W halogen lamp for diagnostic applications in endoscopy. The 150 W halogen lamp also function as a backup up lamp to avoid interrupting a procedure in case of a malfunction of the xenon lamp. The color temperature of the xenon lamp and halogen lamp is approximately 6000 % and 3400 % respectively, the lamp life approximately 500 hours and 50 h respectively.

VII. Substantial Equivalence:

The Endoscopic Light Source XL202/L3 described in this notification is similar in design and technological characteristics to the Karl Storz Xenon Light Source Model 201320-20 (K934559) manufactured by Karl Storz Imaging, Inc. and the Xenon Light Source Auto Lip 5123, Model 2123.011 (K983628) manufactured by Richard Wolf Medical Instruments Corp., the Battery Powered Endoscopic Light Source (K960081)manufactured by Mitsubishi Cable America, Inc., the LH -150 manufactured by Pentax Precision Instrument Corp.and the CLH-SC Halogen Light Source manufactured by Olympus America, Inc.

Both the Endoscopic Light Source XL202/L3 and the predicate devices are intended to provide illumination of body cavities, hollow organs and canals during endoscopic procedures.

The differences between the Endoscopic Light Source XL202/L3 and predicate device are minor and raise no new questions of safety and effectiveness.

Accordingly. WORLD OF MEDICINE Lemke GmbH believes that the Endoscopic Light Source XL202/L3 is substantially equivalent to the predicate devices currently on the market.

VIII. Performance Data:

The Endoscopic Light Source XL202/L3 complies with the International Standard IEC 601-1, IEC 601-1-2 and conforms to the Medical Device Directive 93/42/EEC. In addition, the Endoscopic Light Source XL202/L3 meets the requirements of the Underwriters Laboratories standard UL2601-1.

Signed:

S. Raab

Susanne Raab Official Correspondent

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 0 5 2002

WORLD OF MEDICINE Lemke GmbH % Ms. Susanne Raab 91 Trowbridge Street CAMBRIDGE MA 02138

Re: K020889 Trade/Device Name: Endoscopic Light Source XL202/L3 Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: 78 GCT Dated: March 15, 2002 Received: March 19, 2002

Dear Ms. Raab:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/odrh/dsma/dsmamain.html.

Sincerely yours,

Nancy Cbrogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Image /page/3/Figure/9 description: The image contains a scattering of small, dark, irregular shapes against a bright white background. The shapes are distributed sparsely across the frame, with some clustering towards the bottom left. The contrast between the dark shapes and the white background is high, making the shapes stand out distinctly.

{4}------------------------------------------------

STATEMENT OF INDICATIONS FOR USE

0208889

APPLICANT:

WORLD OF MEDICINE Lemke GmbH

510(K) NUMBER (if known):

DEVICE NAME:

Endoscopic Light Source XL202/L3

INDICATIONS FOR USE:

The Endoscopic Light Source XL202/L3 is intended to be used with fiber optic endoscopes to provide illumination of body cavities, hollow organs and canals during endoscopic procedures.

The device is classified as Cardiac Floating (CF) which allows the use in endoscopic cardiac procedures when used in conjunction with the proper instrumentation.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Per 21 C.F.R. § 801.109)

(Optional Format 1-2-96)

Prescription Use
(Per 21 CFR 801.109)

Nancy C Brogdon

vision Signivision of Reproducti and Radiological Device 510(k) Number

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.