K Number
K020804

Validate with FDA (Live)

Date Cleared
2002-11-19

(252 days)

Product Code
Regulation Number
878.4750
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Auto Suture* Premium Plus CEEA* Disposable Stapler has application throughout the anal canal to perform surgical treatment of hemorrhoidal disease.

Device Description

The Auto Suture* PREMIUM PLUS CEEA* Disposable Stapler is a single patient use device which places a double staggered row of titanium staples. Immediately after staple formation, the instrument's knife blade resects the excess tissue, creating a circular anastomosis.

AI/ML Overview

This looks like a 510(k) premarket notification for a surgical stapler, which is a physical medical device and not an AI/ML powered device. As such, the requested information (acceptance criteria and study details related to AI/ML device performance) is not applicable to this submission.

Therefore, I cannot provide the requested table and study details as they pertain to AI/ML device evaluation.

{0}------------------------------------------------

K020804

IX. 510(k) Summary of Safety and Effectiveness

NOV 1 9 2002

SUBMITTER:United States Surgical150 Glover AvenueNorwalk, CT 06856
CONTACT PERSON:Sarah Hubeny
DATE PREPARED:March 8, 2002
CLASSIFICATION NAME:Staple, Implantable
COMMON NAME:Staple, Implantable
PROPRIETARY NAME:Auto Suture* PREMIUM PLUS CEEA*Disposable Stapler
DEVICE DESCRIPTION:The Auto Suture* PREMIUM PLUS CEEA* DisposableStapler is a single patient use device which places a doublestaggered row of titanium staples. Immediately after stapleformation, the instrument's knife blade resects the excesstissue, creating a circular anastomosis.
INTENDED USE:The Auto Suture* Premium Plus CEEA* DisposableStapler has application throughout the anal canal to performsurgical treatment of hemorrhoidal disease.
MATERIALS:All component materials of the Auto Suture* PremiumPlus CEEA* Disposable Stapler are comprised ofmaterials which are in accordance with ISO Standard#10993-1.

:

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administra 9200 Corporate Boulevard Rockville MD 20850

United States Surgical Sarah Hubeny Regulatory Affairs Associate 150 Glover Avenue Norwalk, Connecticut 06856

Re: K020804

Trade/Device Name: Auto Suture PREMIUM PLUS CEEA Disposable Stapler Regulation Number: 872.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: September 19, 2002 Received: September 19, 2002

Dear Ms. Hubeny:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,

{2}------------------------------------------------

Page 2 – Ms. Sarah Hubeny

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Purret

for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

VC208C4

IV. Indications For Use:

510(k) Number (if known):

Name: Auto Suture* PREMIUM PLUS CEEA* Disposable Stapler

Indications For Use:

The Auto Suture* Premium Plus CEEA* Disposable Stapler has application throughout the anal canal to perform surgical treatment of hemorrhoidal disease.

(Please do not write below this line - continue on another page if needed) Concurrence of CDRH, Office of Evaluation (ODE)

ﺮ ﺍﻟﻤﺮﺍﺟﻊ OR Over-The-Counter Use: Prescription Use: (Per 21 CFR §801.109)

Muriane C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number_________________________________________________________________________________________________________________________________________________________________

§ 878.4750 Implantable staple.

(a)
Identification. An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.