K Number
K020743

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-05-28

(83 days)

Product Code
Regulation Number
870.5310
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PadPro 2502 Sterile radiotransparent external electrodes are indicated for use in external pacing, defibrillation and monitoring applications as a sterile, disposable device for single patent use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patient's skin. Intended for use on adult patients.

When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (360 joule maximum).

Device Description

The electrodes are multifunction because they can be used for defibrillation, pacing, cardioversion, and monitoring. PadPro electrodes can be used on all makes and models of defibrillator, including all of the Bi-Phasic units. Radio transparent. "One Pad System" enables the pads to stay with the patient as they move through departments. PadPro has an electrode for any clinical need or patient situation. The high tack adhesive gel allows PadPro electrodes to be repositioned multiple times. PadPro can provide onsite conversion of current cables to accept the PadPro electrodes. The polymer adhesive gel allows superior contact for uniform current distribution and more effective defibrillation and pacing. PadPro's adapter system simply plugs into the OEM cable. Standardization of products - One connector can be used throughout the institution, no matter what brand or model of defibrillation/pacing unit is being used. All PadPro products are Latex free.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the "PadPro" 2502 Sterile Multifunction Electrodes. This submission claims substantial equivalence to an existing device, the "PadPro" 2001 Electrode (K014209). The primary difference for the 2502 model is that it is sterilized and labeled as such.

The information primarily focuses on comparing the new device to its predicate device to establish substantial equivalence rather than detailing an independent study to prove specific performance against acceptance criteria in the traditional sense of a clinical trial. The "study" here is essentially the comparison to the predicate device and adherence to recognized standards.

Here's an attempt to extract and infer the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (from referenced standards or implied from predicate)Reported Device Performance (for PadPro 2502)
Functional EquivalenceIndicated for use in external pacing, defibrillation, and monitoring applications as a disposable device for single patient use. Provides a conductive interface for defibrillator/pacemaker. Intended for adult patients. Compatible with defibrillators up to 360 joule maximum.SAME as predicate device ("PadPro" 2001). Can be used for defibrillation, pacing, cardioversion, and monitoring. Compatible with all makes/models of defibrillators, including Bi-Phasic units. Intended for adult patients, low power (360 joule max) defibrillators.
Physical CharacteristicsRadiotransparent, non-sterile (for predicate), latex-free, single patient use, self-adhesive, sealed in foil pouch. Size: 12 x 7 cm.Radiotransparent, sterile, single patient use, latex-free, sealed in pouch designed for ETO sterilization. Size: SAME 12 x 7 cm.
Safety StandardsIEC 601-1 (1988), Amendment No. 1 (1991), Amendment No. 2 (1995) - Medical Electrical Equipment - Part 1: General requirements for safety. Sec. 898.12 Performance standard (implied electrical safety/performance). ANSI/AAMI DF-39 (3.3.19) - self-adhesive electrodes for monitoring and defibrillation (implied performance in terms of adhesion, conductivity, current distribution).SAME standards met as predicate device. Implies that the sterilized device continues to meet these safety and performance standards.
SterilityNot applicable to the predicate device (non-sterile).Sterile. (This is the primary modification and implicitly, the sterilization process itself would need to be validated to ensure sterility without compromising performance).

2. Sample Size Used for the Test Set and Data Provenance

The text does not explicitly describe a separate "test set" or a formal clinical study with a specified sample size for the PadPro 2502. The core of the submission is a claim of substantial equivalence to the predicate device, PadPro 2001. The "study" largely consists of demonstrating that the modification (sterilization) does not alter the fundamental safety and effectiveness when compared against established standards.

There is no mention of data provenance (e.g., country of origin, retrospective/prospective). This type of information is typically associated with clinical trials, which are not detailed here.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not provided. As there's no described independent test set or clinical study with patient data requiring expert review for ground truth, this aspect is not applicable to the summary provided. The "ground truth" for this submission relies on the established performance and safety of the predicate device and the adherence to recognized industry standards.

4. Adjudication Method for the Test Set

This information is not provided and is not applicable given the nature of the submission (substantial equivalence claim based on predicate device and standards, not a new clinical study with expert adjudication).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. The document focuses on device characteristics and regulatory standards, not on human reader performance with or without AI assistance. The device in question is a medical electrode, not an AI-powered diagnostic tool.

6. If a Standalone (algorithm only without human-in-the-loop performance) was done

No, a standalone algorithm performance study was not done. The device is an electrode, not an algorithm.

7. The Type of Ground Truth Used

For the PadPro 2502, the "ground truth" is established through:

  • Performance of the predicate device (PadPro 2001): Its history of safe and effective use.
  • Adherence to recognized industry standards: IEC 601-1 and ANSI/AAMI DF-39. These standards define the expected performance and safety characteristics of such electrodes.
  • Sterilization validation (implied): While not explicitly detailed, the process of sterilization itself would have to be validated to ensure it achieves sterility without compromising the device's function or material integrity, thereby implicitly establishing a ground truth for its "sterile" claim.

8. The Sample Size for the Training Set

There is no mention of a "training set" in the context of this device submission. This concept typically applies to machine learning or AI models, which is not relevant for this medical electrode.

9. How the Ground Truth for the Training Set was Established

As there is no training set, this question is not applicable.

{0}------------------------------------------------

K020743

EXHIBIT 2

PadPro LLC. 5643 Plymouth Rd. Ann Arbor, Mi 48105 Phone: 734-663-0132 Fax: 734-663-1306 Contact: Cliff Poppy, President January 30, 2002 510(k) Summary of Safety and Effectiveness

    1. Identification of the Device: Proprietary-Trade Name: "PadPro" 2502 Sterile Multifunction Electrodes Classification Name: Electrode, Electrocardiograph, Multi-Function; MLN Common/Usual Name: Defibrillator Electrode
    1. Equivalent legally marketed device: This device identical in function and in design to the PadPro 2001 Electrode (K014209). The only difference is the modified device has been sterilized and is labeled as such.
    1. Indications for Use: The PadPro sterile radiotransparent external electrodes are indicated for use in external pacing, defibrillation and monitoring applications as a sterile, disposable device for single patent use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patient's skin. The electrode is intended for use on adult patients. When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (360 joule maximum).
    1. Description of the Devices: Features & Benefits:

The electrodes are multifunction because they can be used for defibrillation, pacing, cardioversion, and monitoring. PadPro electrodes can be used on all makes and models of defibrillator, including all of the Bi-Phasic units. Radio transparent. "One Pad System" enables the pads to stay with the patient as they move through departments. PadPro has an electrode for any clinical need or patient situation. The high tack adhesive gel allows PadPro electrodes to be repositioned multiple times. PadPro can provide onsite conversion of current cables to accept the PadPro electrodes. The polymer adhesive gel allows superior contact for uniform current distribution and more effective defibrillation and pacing. PadPro's adapter system simply plugs into the OEM cable. Standardization of products - One connector can be used throughout the institution, no matter what brand or model of defibrillation/pacing unit is being used. All PadPro products are Latex free.

{1}------------------------------------------------

ී. Safety and Effectiveness, comparison to predicate device:
-------------------------------------------------------------------------

..

Comparison Areas"PadPro" 2001 DefibrillatorElectrodes (K014209)."PadPro" 2502Defibrillator Electrodes
Indications for useFor use as disposable electrodes forautomatic and manual externaldefibrillators for monitoring, pacing,cardioversion, and defibrillation.SAME
Where usedHospitals and Paramedic situationsSAME
Basic featuresRadiotransparent, non sterile, latexfree, single patient use, self adhesive,in sealed foil pouch.Radiotransparent, sterilesingle patient use, latex free,sealed in a pouch designedfor ETO sterilization
Size12 x 7 cmSAME
Standard metInternational ElectrotechnicalCommission (IEC) 601-1: MedicalElectrical Equipment 601-1 (1988)Part 1: General requirements forsafety Amendment No. 1 (1991)Amendment No. 2 (1995 andSec.898.12 Performance standard;ANSI/AAMI DF-39 (3.3.19)standard, self adhesive electrodes formonitoring and defibrillationSAME
    1. Conclusion In all respects, the PadPro System Defibrillator Electrodes are substantially equivalent to other electrodes that are legally marketed for this purpose. The device meets the standards referenced above.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are rendered in black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profiles.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 8 2002

PadPro LLC. c/o Mr. Daniel Kamm, P.E. Regulatory Engineer Kamm & Associates P.O. Box 7007 Deerfield, IL 60015

Re: K020743

Trade Name: "PadPro" 2502 Sterile Multifunction Electrodes Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm Regulatory Class: Class III (three) Product Code: MKJ Dated: March 6, 2002 Received: March 6, 2002

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Daniel Kamm, P.E.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours,

Mepel Myna

Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

j) Indications for Use

510(k) Number KO20743

Device Name: "PadPro" 2502 Sterile Multifunction Electrodes

Indications for Use:

The PadPro 2502 Sterile radiotransparent external electrodes are indicated for use in external pacing, I he I add to 2502 blorito ranswarsparent as a sterile, disposable device for single patent use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patient's skin. Intended for use on adult patients.

When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (360 joule maximum).

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseOROver the Counter Use
(Per 21 CFR 801.109)

for Danna-Bea Tillman

Division of Cardiovascular & Respiratory Devices

510(k) NumberK020743
------------------------

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.