K Number
K020670

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-10-08

(221 days)

Product Code
Regulation Number
882.1320
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Physiometrix PSArray² EEG Electrode is applied directly to the patient's skin to enable recording of electrophysiologic signals (such as EEG).

Device Description

The Physiometrix Model 4310, PSArray² EEG Electrode, is a single-use, disposable, pre-gelled electrode array. The PSArray² is comprised of six (6) electrodes located at F7, Fp1, FpZ', FpZ, Fp2 and F8. The electrode is packaged with 1 electrode per pouch, 25 pouches per box and 4 boxes per case. The electrodes are mounted in a polyethylene foam coated with a pressure sensitive adhesive. The electrode pad is an area of silver/silver chloride ink that is connected to cable connector by a silver ink trace. An electrolyte gel is held in place over the electrode pad by a open-cell polyurethane sponge located in wells created by the foam basepad. The electrolyte gel was selected for its low impedance properties. No pre-application prepping is required to achieve adequate impedance levels for operation of the PSA4000. The PSArray² EEG Electrode is connected to the PSA4000 System via a patient cable that attaches to the substrate connector. The PSArray² is designed for use with the PSA4000 System.

AI/ML Overview

The provided text is a 510(k) summary for the Physiometrix Model 4310 PSArray² EEG Electrode. This submission focuses on establishing substantial equivalence to a predicate device, rather than proving performance against specific acceptance criteria in the way a new, high-risk device might. The "acceptance criteria" here are largely met by demonstrating compliance with recognized standards and similarity to the predicate.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria (Standard / Test)Reported Device Performance
Electrical Testing (AAMI/ANSI EC-12 (1991) for Disposable ECG Electrodes)The PSArray² device exceeded requirements indicated in the standard.
Biocompatibility Testing (ISO-10993, Biological Evaluation of Medical Devices Part I & FDA Guidance "Protocol for Dermal Toxicity Testing")The skin contact materials were determined to be safe for contact with skin.

2. Sample Size Used for the Test Set and Data Provenance:

The document does not explicitly state the sample sizes for the electrical or biocompatibility testing. The nature of these tests (e.g., electrical impedance, irritation assays) typically involves a sample of the manufactured device.

Data provenance is not specified (e.g., country of origin). The studies appear to be laboratory-based tests conducted by the manufacturer, rather than clinical studies with patient data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

This type of information is generally not applicable to the tests performed for this device. Electrical and biocompatibility testing rely on objective measurements against established standard thresholds, not expert consensus on interpreting complex medical data.

4. Adjudication Method for the Test Set:

Not applicable. The tests are designed to provide objective, quantitative results against a standard, not subjective interpretations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance:

Not applicable. This device is an EEG electrode, a hardware component for signal acquisition. It does not involve AI or human interpretation of complex medical images, and therefore, an MRMC study is not relevant.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done:

Not applicable. The device is a passive electrode, not an algorithm.

7. The Type of Ground Truth Used:

  • Electrical Testing: The "ground truth" is defined by the objective performance requirements outlined in the AAMI/ANSI EC-12 (1991) standard for disposable ECG electrodes. The device performance (e.g., impedance, offset voltage) is measured against these numerical targets.
  • Biocompatibility Testing: The "ground truth" is established by the accepted scientific methodologies and criteria defined in ISO-10993 Part I and FDA Guidance documents for evaluating the biological safety of materials in contact with skin (e.g., absence of cytotoxicity, irritation, sensitization).

8. The Sample Size for the Training Set:

Not applicable. This device is a hardware component and does not utilize a "training set" in the context of machine learning or AI.

9. How the Ground Truth for the Training Set Was Established:

Not applicable, as there is no training set for this device.

{0}------------------------------------------------

Pase page v

510(K) SUMMARY

Date:February 25, 2002
OCT 08 2002
Company:Physiometrix, Inc.Five Billerica Park101 Billerica AvenueN. Billerica, MA 01862
Contact:Dawn E. FrazerVice PresidentRegulatory Affairs & Quality Assurance(978) 670-2422 x243dfrazer@physiometrix.com
Subject Device:Model 4310 PSArray² EEG Electrode
Classification:Class II, CFR 21 Part 882.1320, Cutaneous Electrodes
Intended Use:The Physiometrix PSArray² EEG Electrode is applied directly to the patient's skin to enable recording of electrophysiologic signals (such as EEG).
Description:The Physiometrix Model 4310, PSArray² EEG Electrode, is a single-use, disposable, pre-gelled electrode array. The PSArray² is comprised of six (6) electrodes located at F7, Fp1, FpZ', FpZ, Fp2 and F8. The electrode is packaged with 1 electrode per pouch, 25 pouches per box and 4 boxes per case. The electrodes are mounted in a polyethylene foam coated with a pressure sensitive adhesive. The electrode pad is an area of silver/silver chloride ink that is connected to cable connector by a silver ink trace. An electrolyte gel is held in place over the electrode pad by a open-cell polyurethane sponge located in wells created by the foam basepad. The electrolyte gel was selected for its low impedance properties. No pre-application prepping is required to achieve adequate impedance levels for operation of the PSA4000. The PSArray² EEG Electrode is connected to the PSA4000 System via a patient cable that attaches to the substrate connector. The PSArray² is designed for use with the PSA4000 System.
Predicate Device:K001055, Model 4300, PSArray EEG Electrode
Similarities:The PSArray² is similar to the predicate device in the following ways:a. The intended use is same.b. The technology is same.c. The materials are the same.
Differences:The PSArra'y is different from the predicate device in the following way:a. The PSArray² has 6 electrodes while the predicate device has 7 electrodes.
Test Results:Electrical testing was conducted in accordance with the American National Standard for Disposable ECG Electrodes, AAMI/ANSI EC-12 (1991). The PSArray² device exceeded requirements indicated in the standard.

{1}------------------------------------------------

K020670 page 2 of 2
page vi

Biocompatibility testing was conducted in accordance with the International Standard ISO-10993, Biological Evaluation of Medical Devices Part I: Evaluation and Testing and FDA Guidance, "Protocol for Dermal Toxicity Tosting for Medical
Devices In Contact with Skin." The skin contact materials were determined to be safe for contact with skin.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three human profiles incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

PHYSIOMETRIX INC. Dawn E. Frazer Vice President, Regulatory Affairs & Quality Assurance Five Billerica Park 101 Billerica Avenue North Billerica,Massachusetts & 862

APR - 9 2012

Re: K020670

Trade/Device Name: PSArray2 EEG Electrode Model 4310 Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous electrode Regulatory Class: II Product Code: GXY Dated (Date on orig SE ltr): October 7, 2002 Received (Date on orig SE ltr): October 7, 2002

Dear Ms. Frazer:

This letter corrects our substantially equivalent letter of October 8, 2002.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Kesia Alexander

Sincerely yours,

Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

for

{4}------------------------------------------------

Model 4310, PSArray2 EEG Electrode Special 510(k): Device Modification

.

the

INDICATIONS FOR USE STATEMENT

Applicant:Physiometrix, Inc.
510(k) Number (if known)Not assigned
Device NameModel 4310 PSArray $2$ EEG Electrode
Indications For UseThe Physiometrix PSArray $2$ EEG Electrode is applied directly topatient's skin to enable recording of electrophysiologic (such asEEG) signals.

.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use _ (Per 21 CFR 801.109)

OR

Over-The-Counter Use _

Hyt Rurd

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number_KOZO670

§ 882.1320 Cutaneous electrode.

(a)
Identification. A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.(b)
Classification. Class II (performance standards).