K020577

Validate with FDA (Live)

Date Cleared
2002-05-22

(90 days)

Product Code
Regulation Number
876.5130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Applied Ureteral Catheter is indicated for use in the urinary tract where drainage or irrigation of the surgical site is desired. It may be used to introduce fluids such as radiopaque contrast media and saline solutions.

Device Description

The Applied Ureteral Catheter has a 5 Fr body, 70 cms long. It incorporates a retentive pigtail coil at one end. The coil aids in retention of the catheter in the renal pelvis during use. Drainage holes are arranged in a spiral pattern extending over the length of the coil to allow for fluid to be irrigated/aspirated through the catheter. The tip of coil has a smooth, atraumatic design for ease of passage through the urinary tract. A luer lock connector is attached at the other end, which may be connected to a drainage or irrigation system as desired. The catheter body is radiopaque and is coated with a hydrophilic coating, which is activated when wet.

The ureteral catheter is designed for placement over a guidewire of up to .038 inches in diameter that is pre-positioned through the urological tract. The catheter coil is temporarily straightened as it is slid over the guidewire. It is pushed up the urological tract until it is placed in a desired position. The luer lock hub is then connected to a source of irrigation or drainage fluid to attain the desired irrigation/aspiration of the urological site.

The ureteral catheter will be supplied in a tyvek/mylar peel pouch (one per pouch) and placed in an outer product carton (one to ten pouches per carton). The ureteral catheter may also be supplied along with a guidewire of up to .038 inches in diameter as a kit.

AI/ML Overview

The provided text describes a 510(k) summary for a Ureteral Catheter (K020577) and includes a statement about the tests performed to demonstrate substantial equivalence to a predicate device. However, it does not provide specific acceptance criteria or detailed study results in numerical form or statistical analyses for the device's performance.

Based on the provided text, I can infer the following about the acceptance criteria and the "study" (which appears to be a series of tests) that aimed to prove the device met these criteria:

1. Table of Acceptance Criteria and Reported Device Performance:

The document states that the Applied Ureteral Catheter was evaluated for "function and performance" similar to currently marketed devices and that the following tests were performed. It implies that the acceptance criteria for these tests were met if the device performed comparably to predicate devices. However, no specific numerical targets or results are provided in the document.

Acceptance Criteria (Inferred from tests)Reported Device Performance
Irrigation Flow Rate: Must be comparable to predicate devices.The test was performed. (Specific flow rate not provided).
Drainage Flow Rate: Must be comparable to predicate devices.The test was performed. (Specific flow rate not provided).
Coil Retention Strength: Must be comparable to predicate devices to ensure retention in the renal pelvis.The test was performed. (Specific strength not provided).
Biocompatibility: Materials must be biocompatible per ISO 10993-1.Materials were "successfully tested to verify biocompatibility of the materials per ISO 10993-1."

2. Sample Size for the Test Set and Data Provenance:

The document does not specify the sample size used for the described tests (Irrigation Flow Rate, Drainage Flow Rate, Coil Retention Strength, Biocompatibility). It also does not mention the data provenance (e.g., country of origin, retrospective or prospective nature). These appear to be bench tests or material tests, not clinical studies involving human populations or geographically diverse data.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

This information is not applicable as the evaluation described consists of engineering/bench tests and biocompatibility testing, not studies requiring expert interpretation or ground truth establishment in a clinical context.

4. Adjudication Method for the Test Set:

This information is not applicable for the same reasons as point 3.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

An MRMC study was not done. The document focuses on demonstrating substantial equivalence through performance tests, not through comparative effectiveness for human readers with or without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

A standalone study was not done. This device is a physical medical device (catheter), not an algorithm or AI system. The performance described refers to its physical and functional characteristics.

7. Type of Ground Truth Used:

The "ground truth" for the performance tests would be established by validated measurement methods and standards applicable to medical device testing (e.g., flow measurement standards for flow rate, materials testing standards for strength, ISO 10993-1 for biocompatibility). For biocompatibility, the ground truth is whether the materials meet the specified safety standards.

8. Sample Size for the Training Set:

This information is not applicable. The device is a physical ureteral catheter, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set Was Established:

This information is not applicable for the same reasons as point 8.

In summary: The provided document is a 510(k) summary for a physical medical device (ureteral catheter). It describes a series of engineering and biocompatibility tests conducted to demonstrate substantial equivalence to a predicate device. It does not provide detailed quantitative acceptance criteria or raw performance data, nor does it involve clinical studies, AI algorithms, or human reader performance evaluations.

{0}------------------------------------------------

MAY 2 2 2002

PAGE 1 OF 2

510(k) SUMMARY

510(k) NUMBER:K020577
SUBMITTED BY:Applied Medical Resources Corporation22872 Avenida EmpresaRancho Santa Margarita, CA-92688(949) 713-8000
CONTACT PERSON:Anil BhalaniVice President of Regulatory Affairs and Clinical Programs
DATE OF PREPARATION:May 10, 2002
NAME OF DEVICE:Ureteral Catheter
CLASSIFICATION NAME:Catheter, Ureteral, Gastro-Urology(Regulation Number 21CFR 876.5130, Urological catheterand accessories)
TRADE NAME:Applied Ureteral Catheter
PREDICATE DEVICE:Pigtail Ureteral Catheter (Ureteral Catheter, K923729),Cook Urological Inc., Spencer, IN.

INTENDED USE STATEMENT: The Applied Ureteral Catheter is indicated for use in the urinary tract where drainage or irrigation of the surgical site is desired. It may be used to introduce fluids such as radiopaque contrast media and saline solutions. The catheter is radiopaque and has a hydrophilic coating to facilitate insertion.

SUMMARY STATEMENT: The Applied Ureteral Catheter has a 5 Fr body, 70 cms long. It incorporates a retentive pigtail coil at one end. The coil aids in retention of the catheter in the renal pelvis during use. Drainage holes are arranged in a spiral pattern extending over the length of the coil to allow for fluid to be irrigated/aspirated through the catheter. The tip of coil has a smooth, atraumatic design for ease of passage through the urinary tract. A luer lock connector is attached at the other end, which may be connected to a drainage or irrigation system as desired. The catheter body is radiopaque and is coated with a hydrophilic coating, which is activated when wet.

The ureteral catheter is designed for placement over a guidewire of up to .038 inches in diameter that is pre-positioned through the urological tract. The catheter coil is temporarily straightened as it is slid over the guidewire. It is pushed up the urological tract until it is placed in a desired position. The luer lock hub is then connected to a source of irrigation or drainage fluid to attain the desired irrigation/aspiration of the urological site.

{1}------------------------------------------------

K020577
PAGE 2 OF 2

Page 2 of 2 510(k) Summary Applied Ureteral Catheter

The ureteral catheter will be supplied in a tyvek/mylar peel pouch (one per pouch) and placed in an outer product carton (one to ten pouches per carton). The ureteral catheter may also be supplied along with a guidewire of up to .038 inches in diameter as a kit.

The Applied Ureteral Catheter is similar in function and performance to ureteral catheters currently marketed for the same application. The following tests were performed on the Applied Ureteral Catheter.

  • a. Irrigation Flow Rate
  • b. Drainage Flow Rate
  • c. Coil Retention Strength

Materials used in the manufacture of the Applied Ureteral Catheter were successfully tested to verify biocompatibility of the materials per ISO 10993-1.

Based on the above testing it was concluded that the Applied Ureteral Catheter is substantially equivalent to predicate devices and introduces no new safety and effectiveness issues when used as instructed.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The text is centered and the background is white.

Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem that appears to be an abstract representation of an eagle or other bird-like figure.

MAY 2 2 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Anil Bhalani Vice President of Regulatory Affairs and Clinical Programs Applied Medical Resources Corp. 22872 Avenida Empresa RANCHO SANTA MARGARITA CA 92688

Re: K020577

Trade/Device Name: Applied Ureteral Catheter with SL-6 Hydrophilic Coating Regulation Number: 21 CFR 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: 78 EYB Dated: February 19, 2002 Received: February 21, 2002

Dear Mr. Bhalani: ·

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for abo barrent to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmaldsmamain.html.

Sincerely yours,

Nancy C Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR USE

Applied Medical Resources is providing this separate cover page for the Applied Ureteral Catheter "Indications for Use" as required.

Not assigned 510(k) Number:

Ureteral Stent Device Name:

Indications for Use: The Applied Ureteral Catheter is indicated for use in the urinary tract where drainage or irrigation of the surgical site is desired. It may be used to introduce fluids such as radiopaque contrast media and saline solutions.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Over-The -Counter Use OR

(Optional Format -2-96)

David C. Segerson

(Division Sign-Off) Division of Reproductive and Radiological Dev 510(k) Number

§ 876.5130 Urological catheter and accessories.

(a)
Identification. A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).(b)
Classification. (1) Class II (performance standards).(2) Class I for the ureteral stylet (guidewire), stylet for gastrourological catheter, ureteral catheter adapter, ureteral catheter connector, and ureteral catheter holder. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.