K Number
K020429

Validate with FDA (Live)

Device Name
AB BELT PRO
Manufacturer
Date Cleared
2002-09-13

(217 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ab Belt Pro device is indicated for use for the improvement of abdominal muscle tone, for strengthening of the abdominal muscles and for developing a firmer abdomen.

Device Description

The Ab Belt Pro is a single channel battery operated muscle stimulation system specifically designed to exercise the abdominal muscles for the purposes of improving abdominal muscle tone, strengthening the abdominal muscles and developing a firmer abdomen. It comprises an electronic stimulation controller module which generates the required electrical stimulation signals and an abdominally worn electrode belt which connects the signals from the stimulator to the skin electrodes located on the inner surface of the belt.

AI/ML Overview

This document is a 510(k) premarket notification for a medical device called the Ab Belt Pro, a powered muscle stimulator. The purpose of this notification is to demonstrate that the Ab Belt Pro is substantially equivalent to a legally marketed predicate device, the Slendertone™ Flex (K010335).

Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not contain a table of acceptance criteria or reported device performance in the way typically expected for a detailed clinical or performance study. Instead, it focuses on demonstrating substantial equivalence to a predicate device.

The "acceptance criteria" here are implied by the criteria for substantial equivalence to the predicate device:

Acceptance Criteria (Implied)Reported Device Performance (Ab Belt Pro)
Same indications for use as predicate device"The Ab Belt Pro device has the same indications for use as the predicate device." (Indications listed as: improvement of abdominal muscle tone, strengthening of abdominal muscles, developing a firmer abdomen.)
Equivalent technological characteristics and instructions for use as predicate device"The Ab Belt Pro device has equivalent technological characteristics and instructions for use, as compared to the predicate device."
Meets mandatory performance standard identified in 21 CFR 898"The device meets the mandatory performance standard identified in 21 CFR 898."
Biocompatibility of electrodes established"The biocompatibility of the electrodes has been established."
Conductive gel is legally-marketed"The conductive gel is a legally-marketed gel cleared through 510(k) number K983964."

2. Sample size used for the test set and the data provenance

Not applicable. This document is a 510(k) submission focused on substantial equivalence, not a clinical trial report with a test set of patients. There is no mention of a test set, sample size, or data provenance from a prospective or retrospective study on human subjects for the Ab Belt Pro's performance in achieving its indicated uses. The testing mentioned refers to technical compliance and biocompatibility.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. As there is no clinical test set described, there are no experts used to establish ground truth for a test set in the context of device efficacy. The "experts" involved would be those performing testing for technical standards compliance or biocompatibility, but these are not clinical experts establishing ground truth for treatment outcomes.

4. Adjudication method for the test set

Not applicable. No clinical test set or efficacy study on human subjects is described, so no adjudication method is mentioned.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a muscle stimulator, not an AI-powered diagnostic or imaging interpretation tool. Therefore, an MRMC study related to readers and AI assistance is not relevant or described.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This device is a physical muscle stimulator and does not involve algorithms in the context of diagnostic or interpretive performance.

7. The type of ground truth used

For the purpose of substantial equivalence, the "ground truth" implicitly refers to:

  • The established regulatory requirements and standards (e.g., 21 CFR 898).
  • The characteristics and established safety/efficacy profile of the predicate device (Slendertone™ Flex: K010335).
  • Biocompatibility standards and prior clearance of the conductive gel.

There is no mention of ground truth established through expert consensus on muscle toning, pathology results, or patient outcomes data for the Ab Belt Pro itself to demonstrate its effectiveness in the 510(k) application. This is because 510(k) focuses on equivalence, not de novo demonstration of safety and efficacy.

8. The sample size for the training set

Not applicable. As this is not an AI/ML device, there is no concept of a "training set" in the context of an algorithm.

9. How the ground truth for the training set was established

Not applicable. See point 8.

{0}------------------------------------------------

SEP 1 3 2002

2.0 510(K) SUMMARY

2.1 Sponsor Information

SponsorThe Dezac Group
54-56 Bath Road
Cheltenham, Glos.
GL53 7HG
United Kingdom
Registrationin England No. 2186341

Mr. Kevin Herbert, Project Engineer Contact Person Phone +44 1242 702300 Fax +44 1242 702301 E-mail kherbert@dezac.co.uk

2.2 Device Name

Trade Name of DeviceAb Belt Pro
Common NameMuscle Stimulator
Classification namePowered Muscle Stimulator
Product CodeNGX
Regulation ClassII
Regulation Number890.5850

2.3 Indications for Use

The Ab Belt Pro device is indicated for use for the improvement of abdominal muscle tone, for strengthening of the abdominal muscles and for developing a firmer abdomen.

Device Description 2.4

The Ab Belt Pro is a single channel battery operated muscle stimulation system specifically designed to exercise the abdominal muscles for the purposes of improving abdominal muscle tone, strengthening the abdominal muscles and developing a firmer abdomen.

10F 3

{1}------------------------------------------------

20F=3/KOLOSI

It comprises an electronic stimulation controller module which generates the required electrical stimulation signals and an abdominally worn electrode belt which connects the signals from the stimulator to the skin electrodes located on the inner surface of the belt.

The device is supplied with a set of three identical electrodes, a tube of conductive gel, an instruction manual, a set of batteries and a fabric belt extension for fuller figures. Power is derived from three (3) 1.5V AAA batteries located in a compartment protected by a removal battery cover at the rear (belt side) of the device. The central umbilical electrode is common to each of the left and right stimulation circuits.

The electrodes connect mechanically to 'press-studs' mounted on the inner face of the abdominal belt. The studs are presented in such a way that each of the outer electrodes may be rotated about a single common stud to accommodate both inner and outer muscle groups. Only these single common studs of the outer electrodes and the central umbilical electrically connected to the stimulator unit. This prevents stimulation to the user through a stud, which is not covered by an electrode pad.

The user extends the belt and puts it in a wrapping motion from front to back, closing it at the back using Velcro patches. When the belt is in place the central electrode locates over the umbilicus and the two outer electrodes locate on either side of the body towards the mid axillary line, between the pelvis and the rib cage. It is well known in the art that this electrode positioning is particularly useful for stimulating the abdominal muscles.

The pulsed stimulation current passes between the outer and center electrodes only. There is no current passed from outer electrode to outer electrode. The user has no access to the wiring or connectors as they are stored internally within the belt structure. As a result he or she cannot alter the current path and so the possibilities for misuse are greatly reduced.

2.5 Basis for Substantial Equivalence

Predicate Device

Slendertone™ Flex: K010335 (Class II) Bio-Medical Research Ltd c/o Mr. Robert Dormer

{2}------------------------------------------------

KO20439

3013

Hyman, Phelps & McNamara, P.C. 700 13th Street NW, Suite 1200 Washington, D.C. 20005

  • The Ab Belt Pro device has the same indications for use as the predicate device. .
  • The Ab Belt Pro device has equivalent technological characteristics and instructions for use, . as compared to the predicate device.
  • . The device meets the mandatory performance standard identified in 21 CFR 898.
  • The biocompatibility of the electrodes has been established. .
  • The conductive gel is a legally-marketed gel cleared through 510(k) number K983964. .

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other. The profiles are black and the background is white.

SFP 1 3 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Kevin Herbert The Dezac Group 54-56 Bath Road Cheltenham, Glos. GL53 7HG United Kingdom

Re: K020429

Trade/Device Name: Ab Belt Pro Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: September 9, 2002 Received: September 10, 2002

Dear Mr. Herbert:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice

{4}------------------------------------------------

Page 2 -- Mr. Kevin Herbert

requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html

Sincerely vours.

Sincerely yours,

Mark N Milburn

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for use statement

510(k) number (if known):

K020429

.

Device Name:

AbBelt Pro

Sponsor Name:

The Dezac Group

Indications for use:

The AbBelt Pro is indicated for use for improvement of abdominal muscle tone, for strengthening of the abdominal muscles, and for the development of a firmer abdomen.

Mark N. Milliman

(Division Sign-Off) (Division Sigil-On), Restorative Divis on on on of cess 510(k) Number --

N/A