K Number
K020313

Validate with FDA (Live)

Date Cleared
2002-05-08

(98 days)

Product Code
Regulation Number
862.3100
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACON® One Step Multi-Drug Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays for the qualitative and simultaneous detection of two to six drugs in a variety of combinations in human urine. The designated cut-off concentrations for these drugs are as follows: Amphetamine at 1,000 ng/ml, Cocaine at 300 ng/ml, Methamphetamine at 1,000 ng/ml, Opiates at 2,000 ng/ml, Marijuana at 50 ng/ml and Phencyclidine at 25 ng/ml. They are intended for healthcare professionals including professionals at the point of care sites.

Device Description

The ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative and simultaneous detection of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine in urine samples. The test is based on the principle of antigen-antibody immunochemistry. It utilizes mouse monoclonal antibodies to selectively detect elevated levels of Amphetamine, Methamphetamine, Cocaine, Opiates, THC and PCP in urine at the cut-off concentrations of 1,000 ng/ml (AMP), 1,000 ng/ml (mAMP), 300 ng/mL (COC), 2,000 ng/ml (OPI), 50 ng/ml (THC) and 25 ng/ml (PCP). These tests can be performed without the use of an instrument.

A positive urine specimen will not generate a colored-line for the specific drug tested in the designated test region. A negative urine specimen or a urine specimen containing of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine at the concentrations below the designated cut-off levels will generate a colored-line in the designated test region for the drug. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device appear to be based on achieving high agreement rates (positive, negative, and overall) with both previously FDA-cleared single drug tests and Gas Chromatography/Mass Spectrometry (GC/MS) analysis. While specific numeric acceptance thresholds are not explicitly stated as "acceptance criteria," the consistently high percentage agreements demonstrated in the results (mostly 95% and above, often 99%) with narrow confidence intervals imply that such high agreement was the de facto criterion for acceptance.

Device Performance vs. Predicate Device (ACON Single Tests):

DrugDeviceAcceptance Criteria (Implied: High Agreement)Reported Performance (Positive Agreement)Reported Performance (Negative Agreement)Reported Performance (Overall Agreement)
AMPTest CardHigh Agreement (>95% for all)99% (97% - 99%)>99% (98% - 99%)>99% (98% - 99%)
COCTest CardHigh Agreement (>95% for all)>99% (97% - 99%)99% (97% - 99%)99% (98% - 99%)
mAMPTest CardHigh Agreement (>95% for all)99% (96% - 99%)>99% (98% - 99%)99% (98% - 99%)
OPITest CardHigh Agreement (>95% for all)98% (95% - 99%)>99% (98% - 99%)>99% (97% - 99%)
THCTest CardHigh Agreement (>95% for all)>99% (97% - 99%)99% (97% - 99%)99% (98% - 99%)
PCPTest CardHigh Agreement (>95% for all)99% (93% - 99%)>99% (98% - 99%)99% (97% - 99%)
AMPTest DeviceHigh Agreement (>95% for all)>99% (97% - 99%)>99% (98% - 99%)>99% (98% - 99%)
COCTest DeviceHigh Agreement (>95% for all)>99% (97% - 99%)>99% (97% - 99%)>99% (98% - 99%)
mAMPTest DeviceHigh Agreement (>95% for all)>99% (97% - 99%)>99% (98% - 99%)>99% (98% - 99%)
OPITest DeviceHigh Agreement (>95% for all)>99% (96% - 99%)>99% (98% - 99%)>99% (98% - 99%)
THCTest DeviceHigh Agreement (>95% for all)>99% (97% - 99%)>99% (98% - 99%)>99% (98% - 99%)
PCPTest DeviceHigh Agreement (>95% for all)99% (95% - 99%)>99% (98% - 99%)99% (97% - 99%)

Device Performance vs. GC/MS Analysis:

DrugDeviceAcceptance Criteria (Implied: High Agreement)Reported Performance (Positive Agreement)Reported Performance (Negative Agreement)Reported Performance (Overall Agreement)
AMPTest CardHigh Agreement (>90% generally)95% (90% - 98%)99% (97% - 99%)97% (95% - 99%)
COCTest CardHigh Agreement (>90% generally)95% (89% - 98%)99% (97% - 99%)98% (95% - 99%)
mAMPTest CardHigh Agreement (>90% generally)90% (84% - 94%)>99% (98% - 100%)96% (93% - 98%)
OPITest CardHigh Agreement (>90% generally)99% (96% - 99%)99% (96% - 99%)99% (97% - 99%)
THCTest CardHigh Agreement (>90% generally)95% (90% - 98%)95% (91% - 98%)95% (92% - 97%)
PCPTest CardHigh Agreement (>90% generally)90% (81% - 95%)99% (96% - 99%)96% (93% - 98%)
AMPTest DeviceHigh Agreement (>90% generally)94% (89% - 97%)99% (97% - 99%)97% (94% - 99%)
COCTest DeviceHigh Agreement (>90% generally)95% (89% - 98%)99% (97% - 99%)98% (95% - 99%)
mAMPTest DeviceHigh Agreement (>90% generally)90% (83% - 94%)>99% (98% - 100%)95% (92% - 97%)
OPITest DeviceHigh Agreement (>90% generally)99% (96% - 100%)99% (96% - 99%)99% (97% - 99%)
THCTest DeviceHigh Agreement (>90% generally)95% (90% - 98%)96% (91% - 98%)95% (92% - 97%)
PCPTest DeviceHigh Agreement (>90% generally)90% (81% - 95%)99% (96% - 99%)96% (92% - 98%)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Over 1,000 clinical urine specimens were employed. Approximately 10% of these samples had drug concentrations in the -25% to +25% cut-off range.
  • Data Provenance: The data is described as "clinical urine specimens," implying they were collected from real-world patients. The country of origin is not explicitly stated, but the submission is to the FDA (USA), suggesting the data is relevant to US clinical settings, though not necessarily collected in the US. The study is retrospective, as it uses existing clinical specimens to compare the new device against established methods.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the number or qualifications of experts used to establish the ground truth.

  • For the comparison with predicate devices, the "ground truth" seems to be the result from the previously FDA-cleared ACON single drug tests. These are also immunochromatographic tests.
  • For the comparison with GC/MS analysis, the GC/MS result is considered the gold standard "ground truth." GC/MS analysis is a laboratory-based method. The qualifications of the personnel performing or interpreting the GC/MS results are not specified, although it is a recognized analytical standard in toxicology.

4. Adjudication Method for the Test Set

The document does not describe any specific adjudication method (e.g., 2+1, 3+1). The comparison is direct: the new device's result is compared to the predicate device's result and to the GC/MS result. Discrepancies are noted in the agreement percentages, but a formal adjudication process for discordant results is not detailed.

5. If a Multi-reader Multi-case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC Study: This is not an AI-assisted diagnostic device. It is a rapid diagnostic test (immunochromatographic assay) designed for qualitative, visual interpretation. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study as typically understood for AI in medical imaging, and the concept of human readers improving with AI assistance, is not applicable here.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable as this is a physical immunoassay device, not an algorithm. The interpretation of the test result (presence or absence of a colored line) is inherently human-in-the-loop, though it's a straightforward visual inspection, not a complex diagnostic image interpretation task. The "standalone" performance here refers to the device's ability to produce a result that aligns with the ground truth when interpreted as intended. The performance tables do represent the standalone performance of the device when read by a human.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Two types of ground truth were used:

  1. Predicate Device Results: The results from previously FDA-cleared ACON single drug test strips.
  2. GC/MS Analysis (Gas Chromatography/Mass Spectrometry): This is a highly accurate and widely accepted laboratory analytical method for drug detection and quantification in toxicology, considered the "gold standard" for this application.

8. The Sample Size for the Training Set

  • The document does not mention a training set. This is because the device is a chemical/immunological assay, not a machine learning or AI algorithm that requires training data. The development of such devices relies on chemical formulation, antibody specificity, and physical design, not data training in the AI sense.

9. How the Ground Truth for the Training Set was Established

  • As there is no training set in the context of an AI/ML algorithm, this question is not applicable. The device's design and performance are established through laboratory R&D and then validated using clinical samples against established methods (predicate device and GC/MS) as described above.

{0}------------------------------------------------

8. SUMMARY OF 510(k) SAFETY AND EFFECTIVENESS

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

KO20313 The Assigned 510(k) number is

Submitter:

ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121

Tel.: 858-535-2030 Fax: 858-535-2038

Date: January 28, 2002

Contact Person: Edward Tung, Ph.D.

Product Names:

ACON One Step Multi-Drug Multi-Line Screen Test Card

ACON One Step Multi-Drug Multi-Line Screen Test Device

Common Name:

Immunochromatographic test for the simultaneously qualitative detection of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine in urine.

Device Classification:

The ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are similar to other FDA-cleared devices for the qualitative and simultaneous detection of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine in urine specimens. These tests are used to provide a preliminary analytical result only. Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine test systems have been classified as Class II devices with moderate complexity.

Classification Name:

Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine test system

{1}------------------------------------------------

Intended Use:

The ACON® One Step Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays for the qualitative and simultaneous detection of two to six drugs in a variety of combinations in human urine samples. The designated cut-off concentrations for these drugs are as follows: Amphetamine at 1,000 ng/ml, Cocaine at 300 ng/ml, Methamphetamine at 1,000 ng/ml, Opiates at 2,000 ng/ml, Marijuana at 50 ng/ml and Phencyclidine at 25 ng/ml. They are intended for healthcare professionals including professionals at the point of care sites.

Description:

The ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative and simultaneous detection of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine in urine samples. The test is based on the principle of antigen-antibody immunochemistry. It utilizes mouse monoclonal antibodies to selectively detect elevated levels of Amphetamine, Methamphetamine, Cocaine, Opiates, THC and PCP in urine at the cut-off concentrations of 1,000 ng/ml (AMP), 1,000 ng/ml (mAMP), 300 ng/mL (COC), 2,000 ng/ml (OPI), 50 ng/ml (THC) and 25 ng/ml (PCP). These tests can be performed without the use of an instrument.

A positive urine specimen will not generate a colored-line for the specific drug tested in the designated test region. A negative urine specimen or a urine specimen containing of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine at the concentrations below the designated cut-off levels will generate a colored-line in the designated test region for the drug. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.

Predicate Devices:

ACON™ of Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single Drug Test strips were used as the predicate devices for the ACON® One Step Multi-Drug Multi-Line Screen Test Card and Test Device to compare their performance with clinical urine specimens.

510(k) Number for these predicate devices are:

ACON AMP One Step Amphetamine Test StripK011673
ACON mAMP One Step Methamphetamine Test StripK011672
ACON COC One Step Cocaine Test StripK010841
ACON OPI One Step Opiates Test StripK011353
ACON THC One Step Marijuana Test StripK003557
ACON PCP One Step Phencyclidine Test StripK011730

{2}------------------------------------------------

Comparison to a Predicate Device:

A comparison of the features of the ACON™ One Step Multi-Line Screen Test Card and Test Device versus the ACON™ Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single Tests is shown below:

  • Both tests are assays intended for the qualitative detection of Amphetamine, Cocaine, . Methamphetamine, Opiates, Marijuana, and Phencyclidine in urine samples.
  • Both tests are intended as a screening method that provides a preliminary analytical test result. .
  • Both tests are immunochromatographic, lateral flow assays for the rapid detection of . Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine with a visual, qualitative end result, while ACON Multi-Drug Multi-Line Test detects 2 to 6 of the above drugs simultaneously.
  • Both tests utilize the same basic immunoassay principles that rely on antigen/ antibody ● interactions to indicate a positive or negative result.
  • Both tests have the same cut-off for each drug test. .

Safety and Effectiveness Data:

Accuracy

A clinical evaluation was conducted using clinical urine specimens. This evaluation compared the test results between ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device versus previously FDA-cleared Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single tests; as well as against data obtained from the customary GC/MS analysis. Over 1,000 clinical specimens were employed including approximately 10% of the samples with drug concentrations in the -25% to +25% cut-off range. The comparisons of data obtained from this study yielded the following results:

{3}------------------------------------------------

ACON One Step Multi-Drug Multi-Line Screen Test Card

vs.

ACON Test CardPositive AgreementNegative AgreementOverall Agreement
AMP129/129=>99%(97% - 99%)*172/172=>99%(98% - 99%)*301/301=>99%(98% - 99%)*
COC112/112=>99%(97% - 99%)*186/187=99%(97% - 99%)*298/299=99%(98% - 99%)*
mAMP121/122=99%(96% - 99%)*174/174=>99%(98% - 99%)*295 / 296=99%(98%-99%)
OPI131/133=98%(95% - 99%)*164/164=>99%(98% - 99%)295/297 =>99%(97%-99%)
THC124/124=>99%(97% - 99%)*175/176=99%(97% - 99%)*299/300=99%(98%-99%)*
PCP71/72=99%(93% - 99%)*160/160=>99%(98%-99%)*231/232=99%(97% - 99%)*

Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single Tests:

  • 95% Confidence Interval

ACON One Step Multi-Drug Multi-Line Screen Test Device vs.

Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine Single Test:

ACON Test DevicePositive AgreementNegative AgreementOverall Agreement
AMP129/129=>99%(97%-99%)*172/172=>99%(98%-99%)301/301=>99%(98%-99%)*
COC112/112=>99%(97%-99%)*186/187=>99%(97%-99%)*298/299=>99%(98%-99%)*
mAMP121/121=>99%(97% - 99%)*175/175=>99%(98%-99%)*296 / 296 =>99%(98%-99%)*
OPI132/133=>99%(96% - 99%)*164/164=>99%(98%-99%)296/297=>99%(98%-99%)*
THC124/124 =>99%(97% - 99%)*175/176=>99%(98%-99%)*299 / 300=>99%(98%-99%)*
PCP71/72 = 99%(95% - 99%)*160/160=>99%(98%-99%)*231/232 = 99%(97% - 99%)*
  • 95% Confidence Interval

{4}------------------------------------------------

Clinical study results of ACON One Step Multi-Line Screen Test Card and Test Device are compared to GC/MS analysis data:

The GC/MS cut-off levels for each of the six drugs tested are as following:

Amphetamine1,000 ng/ml
Cocaine300 ng/ml
Methamphetamine1,000 ng/ml
Opiate2,000 ng/ml
Marijuana50 ng/ml
Phencyclidine25 ng/ml

Samples with drug concentration above the cut-off level were considered presumptive positive and concentrations below the cut-off are considered negative.

ACON One Step Multi-Drug Multi-Line Screen Test Card vs. GC/MS Analysis

ACON Test CardPositive AgreementNegative AgreementOverall Agreement
AMP129/136=95%(90% - 98%)*164/165=99%(97% - 99%)*293/301 = 97%(95% - 99%)*
COC112/118 = 95%(89% - 98%)*180/181 = 99%(97% - 99%)*292/299 = 98%(95% - 99%)*
mAMP121/134 = 90%(84% - 94%)*162/162 = >99%(98% - 100%)*283 / 296 = 96%(93% - 98%)*
OPI130/131 = 99%(96% - 99%)*164/166 = 99%(96% - 99%)*294/297 = 99%(97% - 99%)*
THC116/122 = 95%(90% - 98%)*169/178 = 95%(91% - 98%)*285 / 300 = 95%(92% - 97%)
PCP70/78 = 90%(81% - 95%)*153/154 = 99%(96% - 99%)*223/232 = 96%(93% - 98%)*
  • 95% Confidence Interval

{5}------------------------------------------------

ACON Test DevicePositive AgreementNegative AgreementOverall Agreement
AMP128/136=94%(89% - 97%)164/165=99%(97% - 99%)292/301 = 97%(94% - 99%)*
COC112/118 = 95%(89% - 98%)*180/181 = 99%(97% - 99%)*292/299 = 98%(95% - 99%)*
mAMP120/134 = 90%(83% - 94%)*162/162 =>99%(98% - 100%)282 / 296 = 95%(92% - 97%)*
OPI130/131 = 99%(96% - 100%)164/166 = 99%(96% - 99%)*294/297 = 99%(97% - 99%)*
THC116/122 = 95%(90% - 98%)*170/178 = 96%(91% - 98%)*286 / 300 = 95%(92% - 97%)*
PCP70/78 = 90%(81% - 95%)*152/154 = 99%(96% - 99%)*222/232 = 96%(92% - 98%)*

ACON One Step Multi-Drug Multi-Line Screen Test Device vs. GC/MS Analysis

  • 95% Confidence Interval

Conclusion:

Clinical study results demonstrate the substantial equivalency between the ACON One Step Multi-Drug Multi-Line Screen Test Card and Test Device and the Amphetamine, Methamphetamine, Cocaine, Opiates, THC and PCP single tests, which has already being cleared by FDA and marketed in the United States. It is also demonstrated that these tests are safe and effective in detecting Amphetamine, Cocaine, Methamphetamine, Opiates, Marijuana, and Phencyclidine at the following cut-off concentrations: Amphetamine 1,000 ng/ml, Cocaine 300 ng/ml, Methamphetamine 1,000 ng/ml, Opiates 2,000 ng/ml, THC 50 ng/ml and PCP 25 ng/ml. The physician's office laboratory POL study demonstrated that these tests are also suitable for use by professional at point-of-care site.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three parallel lines forming the staff and a snake winding around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Re:

MAY 0 8 2002

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Edward Tung, Ph.D. Director of Regulatory Affairs ACON Laboratories, Inc. 4108 Sorrento Valley Blvd. San Diego, CA 92121

K020313 Trade/Device Names: ACON® One Step Multi-Line Screen Test Card ACON® One Step Multi-Drug Multi-Line Screen Test Device Regulation Number: 21 CFR 862.3100; 21 CFR 862.3250; 21 CFR 862.3870; 21 CFR 862.3650; 21 CFR 862.3610 Regulation Name: Amphetamine test system; Cocaine and cocaine metabolite test System; Cannabinoid test system; Opiate test system; Methamphetamine test system Regulatory Class: Class II; Class II; Class II; Class II, Class II Product Code: DKZ; DIO; LDJ; DJG; LCM; LAF Dated: April 8, 2002 Received: April 12, 2002

Dear Dr. Tung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2 - .

This letter will allow you to begin marketing your device as described in your 510(k) premarket I nis lotton . The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, (Joase contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

INDICATIONS FOR USE 10.

510(k) Number:

K020313

ACON® One Step Multi-Drug Multi-Line Screen Test Card Device Name:

ACON® One Step Multi-Drug Multi-Line Screen Test Device

Indications for Use:

The ACON® One Step Multi-Drug Multi-Line Screen Test Card and Test Device are rapid chromatographic immunoassays for the qualitative and simultaneous detection of two to six drugs in a variety of combinations in human urine. The designated cut-off concentrations for these drugs are as follows: Amphetamine at 1,000 ng/ml, Cocaine at 300 ng/ml, Methamphetamine at 1,000 ng/ml, Opiates at 2,000 ng/ml, Marijuana at 50 ng/ml and Phencyclidine at 25 ng/ml. They are intended for healthcare professionals including professionals at the point of care sites.

Lean Cooper

(Division Sign-Off)
Division of Clinical L
510(k) Number. K020313

(Please do not write below this point) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

Or

Over-The-Counter Use

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).