(85 days)
The MFCD 118i Medical Flat Color Display is intended to be used as a tool in displaying and viewing digital images for review and analysis by trained medical practitioners. The device is suitable for use in the MRI environment.
The MFCD 118i device is a digital image display designed for use in strong magnetic environments
The provided document describes a 510(k) premarket notification for the Barco MFCD 118i, a medical flat color display intended for use in MRI environments. It does not contain information about acceptance criteria or a study proving the device meets specific performance metrics in the way one would describe for an AI/CADe device.
The 510(k) summary emphasizes the device's intended use as a tool for displaying and viewing digital images for review and analysis by trained medical practitioners, specifically in an MRI environment. The primary "proof" for this type of device in a 510(k) submission is typically its substantial equivalence to an already legally marketed predicate device, rather than a clinical performance study with specific acceptance criteria and outcome metrics.
Therefore, many of the requested sections (acceptance criteria table, sample sizes, expert ground truth, adjudication, MRMC study, standalone performance, ground truth type for training, and training set sample size) are not applicable to the information provided in this 510(k) summary for a display device.
However, I can extract the relevant information regarding the device's purpose and its basis for market approval:
1. Table of Acceptance Criteria and Reported Device Performance
- Acceptance Criteria: For a display device like the MFCD 118i, the "acceptance criteria" in the context of a 510(k) would generally relate to its ability to function safely and effectively within its intended environment (MRI) and display images adequately for medical review, demonstrating substantial equivalence to a predicate device. Specific quantitative performance metrics (e.g., contrast ratio, luminance uniformity, viewing angle, resolution) are usually covered in technical specifications and predicate comparisons, but are not detailed in this summary.
- Reported Device Performance:
- Intended Use: To display and view digital images for review and analysis by trained medical practitioners.
- Key Feature: Suitable for use in the MRI environment (compatible with strong magnetic environments).
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Suitable for use in strong magnetic environments (MRI compatibility) | The MFCD 118i device incorporates materials that are compatible for use in strong magnetic environments. |
| Functions as a digital image display for medical review and analysis | Intended to be used as a tool in displaying and viewing digital images for review and analysis by trained medical practitioners. |
| Substantially equivalent to a legally marketed predicate device | Reviewed by FDA and determined to be substantially equivalent to Barco NV Display Systems MFCD 118 Medical Flat Color Display (K000291). |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable for this device type within this document. This document does not describe a clinical performance study with a test set of data/images. The approval is based on substantial equivalence to a predicate device's design and intended use.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. No ground truth establishment is described as this is not an AI/CADe device being evaluated for diagnostic accuracy.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. No test set or adjudication process is described.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This is a display device, not an AI/CADe system. No MRMC study was performed or is relevant to its approval.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This is a display device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable. No ground truth is established for this type of device's 510(k) submission.
8. The sample size for the training set
- Not applicable. This is a display device, not an AI/CADe system. There is no training set mentioned in relation to this device.
9. How the ground truth for the training set was established
- Not applicable. No training set or ground truth establishment is described.
Summary of "Study" that Proves the Device Meets Acceptance Criteria:
The "study" or justification for the Barco MFCD 118i meeting its implied acceptance criteria is the 510(k) Substantial Equivalence Determination process conducted by the FDA. The submission likely included technical specifications, design documents, and possibly electromagnetic compatibility (EMC) testing reports to demonstrate its MRI compatibility and performance characteristics. However, these specific documents are not part of the provided summary.
The FDA determined that the MFCD 118i is substantially equivalent to Barco NV Display Systems MFCD 118 Medical Flat Color Display (K000291). This means the FDA concluded that the new device is as safe and effective as the predicate device (which was already legally marketed) and does not raise new questions of safety or effectiveness. The core "proof" lies in this comparison to an existing, approved device, rather than a de novo clinical trial with defined performance metrics and ground truth.
{0}------------------------------------------------
APR 0 3 2002
510(K) SUMMARY
Manufacturer:
Submitted By:
Contact Information:
Classification Name:
Common/Usual Name:
Classification Number:
Substantial Equivalence:
Proprietary Name:
Device Description:
Intended Use:
Barco NV Display Systems Theodoor Sevenslaan 106 8500 Kortriik Belgium
Ferguson Medical Consultant to Barco NV
Phone: +32(0) 56 23 32 11 FAX: +32(0) 56 23 3 74
System, image processing
In-room MRI display, IMRI, flat panel for MRI applications, and others
MFCD 118i
21 CFR 892.2050/Procode 90LLZ
Barco NV Display Systems MFCD 118 Medical Flat Color Display (K000291)
The MFCD 118i device is a digital image display designed for use in strong magnetic environments
The Barco MFCD 118i device is intended to be used as a tool in displaying and viewing digital image for review and analysis by trained medical practitioners. The device is suitable for use in the MRI environment
The MFCD 118i device incorporates materials Technological Characteristics: that are compatible for use in strong magnetic environments
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping heads suggesting a sense of community or interconnectedness. The figure is enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 0 3 2002
Barco NV Display Systems % Mr. Frank Ferguson Official Correspondent Ferguson Medical P.O. Box 12038 LA JOLLA CA 92039-2038 Re: K020059
Trade/Device Name: MFCD 118i Flat panel display Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: 90 LLZ Dated: December 21, 2001 Received: January 8, 2002
Dear Mr. Ferguson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate 10 mmerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|---|---|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (If known): KU20059
Device Name: MFCD 118i
Indications For Use:
The MFCD 118i Medical Flat Color Display is intended to be used as a tool in displaying and viewing digital images for review and analysis by trained medical practitioners. The device is suitable for use in the MRI environment.
PLEASE DO NOT WRITE BELOW THIS LINE ~ CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
David L. Stevenson
Prescription Use __ ్ల (Per 21 CFR 801.109)
OR
Over-The- Counter Use ________________________________________________________________________________________________________________________________________________________
N/A