K Number
K014195

Validate with FDA (Live)

Date Cleared
2002-03-19

(88 days)

Product Code
Regulation Number
862.3550
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Boston Scientific Corporation (BSC) Kayak™ Hydrophilic Guide Wires are intended to facilitate the placement of a catheter during diagnostic or interventional procedures. The wire can be torqued to facilitate navigation through tortuous arteries and/or avoid unwanted side branches. This device is not intended for use in the coronary arteries or the neurovasculature.

Device Description

The Kayak™ Hydrophilic Guide Wires are steerable guide wires available in a nominal diameter of 0.035 and 0.038 inches and nominal lengths of 80, 150 and 180 centimeters. The available tip flexibilities will be • Standard, • Stiff • Long Taper. The distal three centimeters of all models are available in straight and pre-formed angled tip design. The Kayak™ Hydrophilic Guide Wire consists of the following three components: • nickel titanium core • polyurethane jacket • hydrophilic coating. The full length of the guide wire has a polyurethane jacket containing tungsten encasing the core wire. The full length of the guide wire has a hydrophilic coating for lubricity.

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and study information for the Kayak™ Hydrophilic Guide Wires:

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Tensile StrengthNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
Torque StrengthNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
Cantilever StiffnessNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
Tip ProlapseNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
Peel TestNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
Lubricity and DurabilityNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
BiocompatibilityNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"
Product Shelf LifeNot explicitly stated, implied to be sufficient for intended use."met all evaluation acceptance criteria"

Note: The document states that the "Test results demonstrate that the Kayak™ Hydrophilic Guide Wires met all evaluation acceptance criteria and are adequate for their intended use." However, the specific quantitative or qualitative acceptance criteria for each test are not detailed in this summary.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample size used for each non-clinical test. It only lists the types of tests performed. The data provenance is described as "in vitro testing and evaluation," meaning the tests were conducted in a laboratory setting, not on human subjects. There is no information regarding the country of origin of the data provided beyond Boston Scientific being located in Maple Grove, Minnesota, USA.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable as the described study is a non-clinical, in vitro performance test of a medical device (guide wire). There is no "ground truth" established by human experts in the context of diagnostic performance or clinical outcomes. The "ground truth" here is the physical and mechanical properties of the device, measured against predefined engineering specifications and standards.

4. Adjudication Method for the Test Set

This information is not applicable as the study is a non-clinical, in vitro performance test. Adjudication methods are typically relevant in clinical studies or diagnostic evaluations where expert consensus is needed to resolve discrepancies in interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted. This document describes non-clinical, in vitro testing of a medical device's physical and functional properties, not a study evaluating human reader performance with or without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

No, a standalone algorithm-only performance study was not conducted. This document describes the performance of a physical medical device (guide wire), not a software algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical tests described is based on engineering specifications, material science properties, and performance standards for guide wires. For example, tensile strength would be measured and compared against a specified minimum force the wire must withstand before breaking. Lubricity would be measured against a standard for friction.

8. The Sample Size for the Training Set

This information is not applicable as the study is a non-clinical performance test of a physical device. There is no "training set" in the context of machine learning or AI algorithm development.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

MAR 1 9 2002

K014195

Summary of Safety and Effectiveness

Section 6

Submitter's Information

Name and Address Boston Scientific Scimed, Inc. One Scimed Place Maple Grove, Minnesota 55311

Contact Person Mark Murphy 763-494-2377

Date December 20, 2001

Device Name

Proprietary Name Boston Scientific Corporation Kayak™ Hydrophilic Guide Wires

Common or Usual Name Kayak™ Guide Wires

Classification Name Catheter Guide Wire

Predicate Devices:

The following predicate devices are currently marketed.

Product510(k)Clearance Date
Terumo Radifocus GlidewireK96144507/09/1996
Cordis Hydronol SteerableGuidewireK97384501/05/1998

6-1

{1}------------------------------------------------

Summary of Safety and Effectiveness, Continued

ModelsAvailableThe Kayak™ Hydrophilic Guide Wires are steerable guide wiresavailable in a nominal diameter of 0.035 and 0.038 inches and nominallengths of 80, 150 and 180 centimeters. The available tip flexibilitieswill be• Standard,• Stiff• Long Taper.The distal three centimeters of all models are available in straight andpre-formed angled tip design.
DeviceComponentsThe Kayak™ Hydrophilic Guide Wire consists of the following threecomponents:• nickel titanium core• polyurethane jacket• hydrophilic coating
PolyurethaneJacketThe full length of the guide wire has a polyurethane jacket containingtungsten encasing the core wire.
HydrophilicCoatingThe full length of the guide wire has a hydrophilic coating for lubricity.

{2}------------------------------------------------

Summary of Safety and Effectiveness, Continued

The Boston Scientific Corporation (BSC) Kayak™ Hydrophilic Guide Intended Use Fires are intended to facilitate the placement of a catheter during diagnostic or interventional procedures. The wire can be torqued to diaghoods or ination through tortuous arteries and/or avoid unwanted side branches. This device is not intended for use in the coronary arteries or the neurovasculature.

Summary of Technological Characteristics The Kayak™ Hydrophilic Guide Wires are manufactured in a similar manner to legally marketed BSC hydrophilic guide wires and manner to 10gaily including the Renegade Braided Microcather (K973645) and the predicate device -- Terumo Radifocus Glidewire (K961445).

Non-Clinical Test Summary Testing and evaluation of the guide wires included:

  • · tensile strength
  • · torque strength
  • · cantilever stiffness
  • · tip prolapse
  • peel test
  • · lubricity and durability
  • · biocompatibility
  • · product shelf life

Results

Test results demonstrate that the Kayak™ Hydrophilic Guide Wires met all evaluation acceptance criteria and are adequate for their intended use.

Summary

The Kayak™ Hydrophilic Guide Wires are considered to be substantially equivalent to guide wires currently marketed by BSC and for Terumo Corporation based on a comparison of intended use, design and the results of in vitro testing and evaluation.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like figure with three distinct profiles visible within its form. The bird is oriented to the right, and the profiles suggest a sense of forward movement or progress. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 9 2002

Mr. Mark Murphy Regulatory Affairs Specialist Boston Scientific Scimed, Inc. One Scimed Place Maple Grove MN 55311

Re: K014195

Trade/Device Name: Kayak™ Hydrophilic Guide Wires Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire. Regulatory Class: Class II Product Code: DOX Dated: December 20, 2001 Received: December 21, 2001

Dear Mr. Murphy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Mr. Mark Murphy

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quarty by systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter wification. The FDA finding of substantial equivalence of your device to a legally premarket notification. The PDF micing of casion for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific acrise for for invitro diagnostic devices), please contact the Office of additionally 21 CH Far 8071 or additionally, for questions on the promotion and advertising of Compliance at (301) 291 1016. Timber of Compliance at (301) 594-4639. Also, please note the your do roo, product of market notification" (21CFR Part 807.97). If guilation chittived, "Nisoftananing of responsibilities under the Act may be obtained from the Other general mionmation on your neernational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Dork Teen

Bram Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Notification BSC Kayak Guide Wires

.

Indications for Use

Section 2

510(k) NumberK014195
Device Name:Kayak™ Hydrophilic Guide Wires
Indications forUse:The Boston Scientific Corporation (BSC) Kayak™ Hydrophilic Guide Wires are intended to facilitate the placement of a catheter during diagnostic or interventional procedures. The wire can be torqued to facilitate navigation through tortuous arteries and/or avoid unwanted side branches. This device is not intended for use in the coronary arteries or the neurovasculature.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109) ✓

OR

Over The Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

Division of Cardiovascular & Respiratory Devices
510(k) Number K014195

§ 862.3550 Lead test system.

(a)
Identification. A lead test system is a device intended to measure lead, a heavy metal, in blood and urine. Measurements obtained by this device are used in the diagnosis and treatment of lead poisoning.(b)
Classification. Class II.