K Number
K014083

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-01-08

(28 days)

Product Code
Regulation Number
868.5830
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cell Saver 5 Autologous Blood Recovery System is intended for use as an autotransfusion apparatus in conjunction with the Cell Salvage and Sequestration Protocols and single use sterile disposable sets. The Cell Saver 5 Autologous Blood Recovery System is intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient. The intended use of the Sequestration Protocol is to collect an autologous preoperative platelet rich plasma product.

LN291A Haemonetics Basic Mini Volume (70mL) Cell Saver 5 Bowl Set is a functionally closed systems intended to be used to perform cell salvage using the CS5 Autologous Blood Recovery Svstem.

LN242 10L Waste Bag is intended to be used as an alternate waste bag with the CS5 Autologous Blood Recovery System.

Device Description

This Special 510(k) premarket notification describes a modification to Haemonetics' currently legally marketed CELL SAVER 5 SYSTEM and the Cell Saver 5 Protocols and associated disposable sets. The proposed modifications involve the addition of a new smaller volume, blow molded bowl disposable set, the LN291A Basic Mini Volume (70mL) Cell Saver 5 Bowl Set and the associated software and hardware changes required to accommodate the smaller volume set. It also includes the addition of an alternate 10L waste bag, LN242. The intended use of the modified device is the same as for the predicate device and has not changed as result of the changes in software for the Cell Salvage and Sequestration Protocols.

Additionally, the design configuration, material composition, manufacturing methods and operational principles for the changed device are equivalent to those of the predicate device.

AI/ML Overview

This document is a 510(k) premarket notification for a modification to an existing medical device, the Haemonetics Cell Saver 5 Autologous Blood Recovery System. The modification involves adding a new, smaller volume bowl disposable set (LN291A) and an alternate waste bag (LN242), along with associated software and hardware changes.

Therefore, the submission is focused on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with detailed acceptance criteria and reported performance metrics in the way a novel device might. The document explicitly states: "These changes do not affect the intended use or alter the fundamental scientific technology of the device."

Given this context, many of the requested sections (e.g., sample size for test set, number of experts for ground truth, MRMC study, how ground truth for training set was established) are not applicable because the submission is about demonstrating equivalence of a modified, existing device, not a de novo clearance requiring extensive clinical performance studies.

Here's an attempt to answer the questions based on the provided text, while acknowledging the limitations of the document's purpose:


1. A table of acceptance criteria and the reported device performance

The document does not explicitly state quantitative acceptance criteria or detailed reported device performance metrics in a table format. The basis for acceptance is a demonstration of substantial equivalence to the predicate device. The "performance" assessment is qualitative, focusing on whether the modifications adversely affect performance.

The key claim relating to performance is:

  • "Verification testing has been performed and demonstrated that the performance of the modified device is not adversely affected by the software changes." (Section: DESIGN CONTROL ACTIVITIES)
  • "The Cell Saver 5 Autologous Blood Recovery System, the Cell Saver 5 Protocols, software revision K, and its associated disposable set LN291A are substantially equivalent to legally marketed devices." (Section: CONCLUSION)

There are no specified numerical targets (e.g., specific recovery rates, wash efficiencies, or platelet counts) or measurements presented for the modified device to compare against. The "acceptance criteria" can be inferred as maintaining the established performance characteristics of the predicate device, which is considered satisfied by the "substantial equivalence" determination.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify a "test set" in the context of a clinical study or data-driven performance evaluation. The "verification testing" mentioned is likely internal engineering and quality assurance testing related to the software and hardware changes, rather than a clinical trial. Therefore, sample size, data provenance (country of origin), and retrospective/prospective nature are not described.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not Applicable. The submission does not detail a study involving expert-established ground truth. The assessment is based on design control activities and verification testing to ensure the modified device functions as intended and is equivalent to the predicate.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not Applicable. There is no "test set" or adjudication process described for clinical performance.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not Applicable. This device is an autotransfusion apparatus, not an AI-driven diagnostic or imaging device used with human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not Applicable. This is a hardware and software modification to a medical device, not an algorithm being evaluated in isolation. The "software changes" are integrated into the device's operation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not Applicable. The submission focuses on substantial equivalence based on engineering and performance characteristics relative to a predicate device, not on specific clinical endpoints validated by ground truth methods like pathology or expert consensus. The "ground truth" (if one could even apply the term loosely) would be the established performance and safety profile of the predicate device.

8. The sample size for the training set

Not Applicable. This document describes a modification to an existing device, not the development of a machine learning model that requires a training set.

9. How the ground truth for the training set was established

Not Applicable. As per point 8, there is no training set mentioned or implied in this 510(k) submission.

{0}------------------------------------------------

8 2002 jan

Image /page/0/Picture/1 description: The image shows a sequence of handwritten alphanumeric characters. The sequence reads 'K014083'. The characters are written in a simple, clear style, with each digit and letter easily distinguishable.

510(K) SUMMARY

Date:November 30, 2001
Sponsor:Haemonetics Corporation400 Wood RoadBraintree, MA 02184
Contact:John Sokolowski
Tel:(781) 356-9488
Fax:(781) 356-3558
Proprietary Name:Haemonetics Cell Saver 5 Autologous Blood Recovery System
Haemonetics List No. 291A - Haemonetics Corporation Basic MiniVolume (70mL) Cell Saver 5 Bowl Set
Haemonetics List No. 242 10L Waste Bag
Classification Name:Autotransfusion Apparatus (21 CFR 868.5830- Class II)
Empty container for the collection and processing of blood andblood components (21 CFR 864.9100)
Common Name:Cell Saver 5LN291A Disposable Set with 70mL bowl
LN242 10L Replacement Waste Bag
List NoPredicate DeviceReference
CS5 with Cell Salvage andSequestration ProtocolSoftware Revision KCS5 with Cell Salvage andSequestration Protocol SoftwareRevision -K932890
LN291A Haemonetics BasicMini Volume (70 mL) CellSaver 5 Bowl Set• LN261 Haemonetics Basic LowVolume High Speed Cell Saver 5Bowl Set (125mL bowl)• LN263 Haemonetics Basic HighSpeed Cell Saver 5 Bowl Set(225mL bowl)K932890
LN242 10L Waste Bag• LN246 10L Waste Bag• Pre-amendment

:

: 上一篇:

:

{1}------------------------------------------------

DEVICE DESCRIPTION

Modification to an Existing Device

This Special 510(k) premarket notification describes a modification to Haemonetics' currently legally marketed CELL SAVER 5 SYSTEM and the Cell Saver 5 Protocols and associated disposable sets. The proposed modifications involve the addition of a new smaller volume, blow molded bowl disposable set, the LN291A Basic Mini Volume (70mL) Cell Saver 5 Bowl Set and the associated software and hardware changes required to accommodate the smaller volume set. It also includes the addition of an alternate 10L waste bag, LN242. The intended use of the modified device is the same as for the predicate device and has not changed as result of the changes in software for the Cell Salvage and Sequestration Protocols.

Additionally, the design configuration, material composition, manufacturing methods and operational principles for the changed device are equivalent to those of the predicate device.

Intended Use

The Cell Saver 5 Autologous Blood Recovery System is intended for use as an autotransfusion apparatus in conjunction with the Cell Salvage and Sequestration Protocols and single use sterile disposable sets. The Cell Saver 5 Autologous Blood Recovery System is intended for use to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient. The intended use of the Sequestration Protocol is to collect an autologous preoperative platelet rich plasma product.

LN291A Haemonetics Basic Mini Volume (70mL) Cell Saver 5 Bowl Set is a functionally closed systems intended to be used to perform cell salvage using the CS5 Autologous Blood Recovery Svstem.

LN242 10L Waste Bag is intended to be used as an alternate waste bag with the CS5 Autologous Blood Recovery System.

DESIGN CONTROL ACTIVITIES

For the production, design, manufacturing and worldwide marketing of automated blood component collection systems, Haemonetics has established and is operating under a quality system that is based upon the requirements of the US Food and Drug Administration's Quality System Regulation, International Organization's ISO 9001, the European Committee for Standardization's EN 46001, and the Medical Device Directive 93/42/EEC.

In accordance with Haemonetics' Quality System, potential risks associated with the software modifications were identified. Verification testing has been performed and demonstrated that the performance of the modified device is not adversely affected by the software changes.

{2}------------------------------------------------

CONCLUSION

The Cell Saver 5 Autologous Blood Recovery System, the Cell Saver 5 Protocols, software revision K, and its associated disposable set LN291A are substantially equivalent to legally marketed devices. The proposed modifications include the addition of the LN291A Basic Mini Volume (70mL) Cell Saver 5 Disposable Bowl Set and the associated software and hardware changes to the Cell Saver 5 System.

The LN242 10L Waste Bag is substantially equivalent to the LN246 10L Waste Bag.

These changes do not affect the intended use or alter the fundamental scientific technology of the device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of a bird or other winged creature.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

8 2002 JAN

Mr. John Sokolowski Vice President, Regulatory Affairs Haemonetics Corporation 400 Wood Road Braintree, MA 02184-9144

Re: K014083

Trade Name: Haemonetics® Cell Saver® 5 Autologous Recovery System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II (two) Product Code: CAC Dated: December 10, 2001 Received: December 11, 2001

Dear Mr. Sokolowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. John Sokolowski

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

. V. Dale Miller

ram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication for Use Statement 2.8

K014083 510(k) Number:

Haemonetics Cell Saver 5 Autologous Blood Recovery System Device Names:

Haemonetics List No. 291A Disposable Set: Haemonetics Corporation Basic Mini Volume (70 mL) Cell Saver 5 Bowl Set

Haemonetics List No. 242 10L Waste Bag

  • The Cell Saver 5 Autologous Blood Recovery System is intended for use Indications for Use: to recover blood shed during or subsequent to an operation or as a result of trauma, processing the blood by a centrifugation and washing procedure, and pumping this processed red cell product to either a bag for gravity reinfusion into the patient or to the arterial line of an extracorporeal circuit for reinfusion into the patient. The intended use of the Sequestration Protocol is to collect an autologous, preoperative, platelet rich plasma product for reinfusion to the same patient within 6 hours of collection.
    The LN291A is intended to be used with the CS5 Autologous Blood Recovery System for cell salvage protocols.

The LN242 is intended to be used with the CS5 Autologous Blood Recovery System as a replacement waste bag.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use_ _ حال_ OR Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109)

510(k) Number

8

§ 868.5830 Autotransfusion apparatus.

(a)
Identification. An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.(b)
Classification. Class II (performance standards).