K Number
K014063

Validate with FDA (Live)

Date Cleared
2002-02-07

(58 days)

Product Code
Regulation Number
870.2770
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Task Force® Monitor 3040 (TFM) is intended to noninvasively measure and display a patient's hemodynamic parameters using Impedance Cardiography (ICG), Electrocardiography (ECG), and photoplethysmography (continuous), oscillometric Blood Pressure (oscBP) and continuous blood pressure. The TFM continuously monitors the subject's hemodynamic parameters without reporting any diagnosis. Every measurement must be supervised by a medical trained professional. The Task Force® Monitor 3040 is a diagnoses aiding device and therefore not designed for vital sign monitoring or self-monitoring of patients.

Device Description

The device architecture of the TFM is equivalent to the BioZ.pc. The TFM system consists of a Patient Biosignal Electronic System (PBES) and a PC for calculating and displaying the hemodynamic parameters. ICG signals from a patient are assessed using the same method as the predicate device (8 electrodes, injecting a minimal current). The user inputs the patient's data into the TFM software on the PC including gender, age, height, weight, hematocrit and the distance between the inner ICG electrodes. The PC runs Windows 2000 operating system. The parameters are displayed on a full color screen and measurement data can be stored on the PC's hard disk. The TFM calculates and displays the same hemodynamic parameters as the predicate device.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the CNSystems Task Force® Monitor 3040, based on the provided text:

Acceptance Criteria and Reported Device Performance

The documents primarily focus on demonstrating substantial equivalence to predicate devices rather than directly stating specific acceptance criteria with quantified performance targets for the Task Force® Monitor 3040 itself. Instead, the performance specifications of the TFM are compared against its predicate devices.

However, based on the "Performance specification" section, we can infer the following:

Acceptance Criteria (Inferred from Performance Specification)Reported Device Performance (Task Force® Monitor 3040)Predicate Device (BioZ.PC) PerformancePredicate Device (Ohmeda 2350 Finapres) Performance
ICG Measurement
- Measuring Current400µA eff., 40 kHz2.5mA rms, 70kHz (no further information on temporal derivative of impedance)Not applicable (Finapres does not perform ICG)
- Temporal derivative of impedance (dZ/dt)±10 Ohm/sNot explicitly statedNot applicable
ECG Measurement
- Measuring Range±5mVNot explicitly statedNot applicable
- Indication Range (Heart Rate)30 - 150 beats/minNot explicitly statedHeart rate measuring range: 12 to 175bpm, accuracy: ± 5bpm or ± 5% of reading (whichever is higher)
- Sampling Frequency1000Hz (Europe), 1200Hz (USA)1000Hz (in Europe)Not explicitly stated for Finapres ECG, but it measures pulse rate.
Oscillometric Blood Pressure (oscBP)
- Measuring Accuracy±5mmHgNot explicitly stated (BioZ.PC uses manual user input or optional oscillometric BP, no accuracy specified for its oscillometric function)BP measuring range: 20 to 260mmHg, accuracy: pressure transducer Linearity: + 2mmHg to -4.5mmHg (+ 0.25kPa to - 0.55kPa)
- Measuring Range50 to 250mmHgNot explicitly stated20 to 260mmHg
Continuous Blood Pressure (continuous BP, based on Penaz method and corrected by oscBP)
- Measuring Accuracy±5mmHg (due to oscBP)Not explicitly statedBP measuring range: 20 to 260mmHg, accuracy: pressure transducer Linearity: + 2mmHg to -4.5mmHg (+ 0.25kPa to - 0.55kPa) for systolic and diastolic blood pressure, and also calculates mean arterial blood pressure. (Finapres itself does not state a correction against an oscillometric device like TFM).
- Measuring Range50 to 250mmHgNot explicitly stated20 to 260mmHg
Patient Accommodation LimitsWeight: 66.14 - 440.9 lbs (30 - 200 kg) Height: 1.64 - 8.20 feet (50 - 250 cm)Weight: 60 lb. (27.3 kg) to 340 lb. (164 kg) Mean arterial pressure greater than 130mmHg (not accommodated) Late sepsis (e.g., stage 3 and 4) (not accommodated) Aortic insufficiency (not accommodated)Not explicitly stated

Study Information

  1. Sample Size used for the test set and the data provenance:

    • The document mentions "A clinical study shows, that both devices follow the arterial blood pressure very well with the difference, that the TFM has no interruptions in the signal." However, no specific sample size for this clinical study is provided.
    • Data provenance is not explicitly stated. Given the manufacturer is CNSystems Medizintechnik GmbH in Austria, it is likely the study was conducted in Europe, possibly Austria, but this is an inference. It's also not stated if the study was retrospective or prospective.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not specified. The document does not provide details on the establishment of ground truth or the involvement of experts for the clinical study mentioned.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not specified. There is no mention of an adjudication process for the clinical study.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study involving human readers and AI assistance was not mentioned. The document describes a comparison between the new device (Task Force® Monitor 3040) and two predicate devices, not a study comparing human performance with and without AI assistance. The TFM is a measurement device, not an AI-assisted diagnostic tool in the typical sense implied by an MRMC study.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • The primary study mentioned is a "clinical study" comparing the TFM and the Finapres device's blood pressure measurement. This suggests a standalone performance evaluation of the device's measurement capabilities. The device is intended "to noninvasively measure and display a patient's hemodynamic parameters" and states "Every measurement must be supervised by a medical trained professional." This implies it's a device that provides data to a professional, rather than an "algorithm only" in the sense of a fully automated diagnostic AI. The "standalone" performance here refers to the accuracy of its measurements against a reference.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the clinical study, the ground truth for comparing continuous blood pressure is implied to be through direct comparison with the Finapres device and its established methods, and the TFM uses an oscillometric device for "absolute BP values." Therefore, the ground truth is likely based on established physiological measurement techniques, with the oscillometric device serving as a reference for absolute BP values.
  7. The sample size for the training set:

    • Not applicable / Not specified. The document does not describe the Task Force® Monitor 3040 as an AI-driven device requiring a training set in the modern sense. It uses "methods and algorithms for calculating the patient's hemodynamic parameters as both predicate devices." The development would have been based on physiological principles and existing algorithms rather than machine learning training on a large dataset.
  8. How the ground truth for the training set was established:

    • Not applicable / Not specified. As the device is not described as based on machine learning, there is no mention of a training set or ground truth establishment for it. Its validation is based on demonstrating substantial equivalence to predicate devices using established measurement principles.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. Inside the circle is a symbol that resembles a stylized caduceus, with three parallel lines that curve and merge into a single line at the bottom.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 0 7 2002

CNSystems Medizintechnik GmbH c/o Mr. Mark Job TÜV Product Service 1775 Old Highway 8 New Brighton, MN 55112-1891

Re: K014063

Trade Name: Task Force® Monitor 3040 Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II (two) Product Code: DSB Dated: January 31, 2002 Received: January 31, 2002

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your Section > IQ(x) premanted is substantially equivalent (for the indications
referenced above and have determined the devices marketed in interstate referenced above and nave delemined the devices marketed in interstate
for use stated in the enclosure) to legally marketed to a fevice. Device Amendments, for use stated in the enclosure) to regally inanced teched Medical Device Amendments, or to
commerce prior to May 28, 1976, the enactment of the Federal Food. Drue, commerce prior to May 28, 1976, the enactinence with the provisions of the Federal Food, Drug, devices that have been recuassified in accordance was areas approval application (PMA).
and Cosmetic Act (Act) that do not require approval assured a provisions of the Act. and Cosmetic Act (Act) that do not require approvate of the general controls provisions of the Act. The You may, therefore, market the device, subject to the gentlation annual registration, listing of
general controls provisions of the Act include requirements for annual misbr general controls provisions of the Act merate requirement.
devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (see above) more in the endering your device can
may be subject to such additional controls. Birts 800 to 808 In addition FDA may may be subject to such additional controlist Existing major roganto 898 In addition, FDA may
be found in the Code of Federal Regulations, Title 21, Parts 800 to 898 In a be lound in the Code of Federal Regeralistian, and the Federal Register.

{1}------------------------------------------------

Page 2 - Mr. Mark Job

Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised allat I DA 3 issuation of a basevice with other requirements of the Act that I DA has made a decommanding administered by other Federal agencies. You must of ally receral statutes and regulations and limited to: registration and listing (21 comply with an the Act 3 requirements, not 801); good manufacturing practice requirements as set CI K I at 807), idocing (21 CFR Part 820); and if applicable, the electronic forul in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to oogin mains of substantial equivalence of your device to a legally premarked notineation: "The PDF interession for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific arrive tor your avitro diagnostic devices), please contact the Office of additionally 21 CF N Fat 007.10 101 11 12:11 Paally, for questions on the promotion and advertising of Compliance at (301) 591 1010. Finance at (301) 594-4639. Also, please note the your do roo, prease connect more to premarket notification" (21CFR Part 807.97). regulation entitied, "Misoranung of responsibilities under the Act may be obtained from the Other general information on your reportional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Dork Tith

L Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

of T Page I

510(k) Number (if known): K014063

Task Force® Monitor 3040 Device Name:

Indications For Use:

Indications For Use:
The Task Force® Monitor 3040 (TFM) is intended to noninvasively measure and display a patient's The Task Force® Mouitor 3040 (TFM) is intended to nominastry Incerverer diography (BCG),
hemodynamic parameters using Impedaace Cardiography (ICG), Electrocrapity (BCG), ( The Marce parameters using Impelane Cardingson (ICC), phocores (ontinental), The TFM (1), The TFM (1), The TFM (1), The TFM (10), The TFM (10), The TFM monitors oscillometric Blood Pressure (oscBP) and continuous silova I reporting any diagnosis.
continuously the subject's hemodynamic parameters without reporting any diagnosis. ostinuously the subject's hemodynamic parameters without reportung any the Castil Porce® Monito!
Every measure of the supervised by a medical trained professional. The Task F Every measurement must be supervised by a medical trained processor.
3040 is a diagnoses aiding device and therefore not designed for vital sign monitoring or self-monitoring of patients.

(Please do not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

(Optional Format 3-10-98)

Respiratory Devices Division of Cardiova 510(k) Number

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for CNSystems. The logo consists of a stylized human figure on the left, followed by the text "CNSystems" in a bold, sans-serif font. Below the company name, there is a tagline that reads "The Brain & Heart Company" in a smaller, regular font.

1. Summary and Certification

1. 1 Premarket Notification 510(k) Summary

SUBSTANTIAL EQUIVALENCE

SUBSTANTIAL EQUI▼ALLENOE
Identification of predicate devices, model and manufacturer:

Predicate device:CardioDynamics BioZ.pc
Model:BZ-500/BZ-501
Manufacturer:CardioDynamics International Corporation
Predicate Device 510(k):K001081
Reason for Submission:New device
Predicate device:Ohmeda Finapres
Model:2350 Finapres

Manufacturer: Predicate Device 510(k): Reason for Submission:

Ohmeda Medical K880572 New device

The Task Force® Monitor 3040 is substantially equivalent to the BioZ.pc in terms of design, in-The Task Force® Monitor 3040 is substantally equivated to the Diozalia 1000 is substantially
tended use and principle of operation. Furthermore the Task Force® Monitor 3040 i tended use and principle of operation. Futurerinon interest and principle of operation.
equivalent to the Finapres 2350 in terms of intended use and principle of open

equivalent to the Finapres 2350 in tems of intended use openamic patient monitor-The Task Force® Monitor 3040 simply combines inc nomination of the measurement
ing principle of the BioZ.po system with the nominuous blood pressure measurement ing principle of the BioZ.pc system with the nomination electronics (which provides all data acof the Finapres 2350 device. The patient instrumentation electronics with the of the Finapres 2350 device. The patent instrumention the PC which interacts with the user.
quited isolation and defib protection) are separated from the PC which interacts w quisition, isolation and defib protection) are separated notes and the software is enclosed.
The software is pre-installed on the PC, a backup copy of the software is enclose

The software of the Task Force® Monitor 3040 controls the hardware, displays the hemodynamic
. The software of the supportunities atore the measurement data on the hard disk The software of the Task Force® Monitor 3040 ontrols the naturent data on the hard disk (like in
parameters and given the possibility to store the measurement data on the bir parameters and gives the possibility to store the measurent and on and and announce integrity
the BioZ.pc). Both devices (Task Force® Monitor 3040 and BioZ.pc) use a socured the BioZ.pc). Both devices (Tast Force" Niomie 10-0 and Divenove that no common has occurred of check when the monitoring software is first activated, to filsure mail and other will be the beginning of each measurement.

A difference is the software installation kit: while CardioDynamics encloses a software installation
ics allows only and on the supported block on the PC . Furthermore Cardio A difference is the software installetion kit: While Cardiobynamics allows only
kit diskette, the TFM software is pre-installed on the PC. Furthermore CardioDynamics, kit diskette, the TFM software is pre-installed on the P. Futush by CardioDynamics,
particular models of notebook PC's which have been previously validated by Cardio the TFM particular models of notebook PC's which have been previously valuation of the first of the TFM
due to the short life cycle of state-of-the-art PC's, CNSystems ships a fully system.

The Task Force® Monitor 3040, the BioZ.pc system and the Finapres device are portable in design The Task Force® Monitor 3040, the BioZ.pc system and clinical see of the Task Force® and for use in the hospital, outpatient and climical secures. Inc. monthers using Impedance Monitor 3040 is to noninvasively measure a patient s lientodyname parament. Monitoring is accom-

Chapter 6: Page 1 (13)

Image /page/3/Picture/15 description: The image contains a handwritten text that appears to be "UB". The letters are written in a simple, slightly slanted style. The "U" is on the left, and the "B" is on the right, with the two letters connected.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CNSystems, which includes a stylized figure of a person next to the company name in bold, sans-serif font. Below the company name, in a smaller font, is the tagline "The Brain & Heart Company." The logo is simple and professional, conveying a sense of expertise in the fields of neuroscience and cardiology.

CNSystems Medizintechnik GmbH urope, Austria, 8020 Graz umkircherstrasse t +43/316/7 23456-0, Fax: -2 nail: office@cnsystems.at

plished by attaching 3 double electrodes and a neutral electrode for ICG (1 in the neck and 2 on each side of the thorax), 4 electrodes for ECC, one finger cuff for continuous blood pressure mad 2 on
surement and one upper arm cuff for assillements blog pressure measurement and one upper arm cuff for oscillonetic blood pressure measurement. ICG is injecting a minimal current through the oscholietive blood pressurement. ICG is injecting
the inner electrodes. the inner electrodes.

The Task Force® Monitor 3040 uses the same methods and algorithms for calculating the patient's hemodynamic parameters as both predicate devices.

Blood pressure is measured in two ways: the absolute blood pressure values are measured with an oscillometric device and the continuous blood pressure values are measured with an
(same method as the Finances device). The TEM a (same method as the Finances device). The TFM system automatically corrects the continuous blood pressure trend to the absolute values of the oscillometric device. Due to the state-of-the-ar electronic components, the blood pressure is monitored contiguously without any internuptions while the Finances device has o interupt the measurement for resetting the set point from time to
time.

Both the Task Force® Monitor 3040 and the BioZ.pc are IBM PC-based products, which differ
only in the version of the operation system only in the version of the operation system.

The Task Force® Monitor 3040 (TFM) is substantially equivalent to its predicate devices, the BioZ.pc currently marketed by CardioDynamics International Corporation and the Finances, the The TFM has the same intended use and no technological differences which would rise new questions concerning safety and effectiveness.

The justification for this substantial equivalence determination is presented below.

Substantial equivalence is shown in the following table (on the next page).

{5}------------------------------------------------

CNSystems
The Brain & Heart Company

CNSystems Medizintechnik GmbH
Europe, Austria, 8020 Graz
Baumkircherstria, 8020 Graz
Tel.: +43/315/77 234566-0 Fax: -2
Email: office@cnsystems.at

ubstantial Equivalenc

Iरे
6
AttributeTask Force Monitor 3040 (new)BioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
ManufacturerCNSystems Medizintechnik GmbHBaumkircherstrasse 1A - 8020 Graz, AustriaCardioDynamics International Corp.6175 Nancy Ridge DriveSan Diego, CA 92121, USAOhmeda Medical355 Inverness Dr. SouthEnglewood, CO 80112 5810, USA
510(k) numbernoneK001081K880572
Indication for useThe TFM has the same intended use/indi-cations for use as the BioZ.PC device.Additionally it combines the ICG methodand the continuous blood pressure methodof the Finapres device.The device is for use in the hospital, out-patient and clinical setting. The BioZ.PCis to noninvasively measure a patient'shemodynamic parameters using Imped-ance Cardiography (ICG). Monitoring isaccomplished by attaching 8 electrodes tothe patient, injecting a minimal currentthrough the upper electrodes, and readingthe returning voltage waveform from theinner electrodes. Each measurement hasto be supervised by a medical profes-sional.The Finapres device is to noninvasivelyand continuously measure a patient's arte-rial blood pressure using the photopl-ethysmographic method first described byPenaz. The device uses a finger cuff forsystolic and diastolic blood pressure mea-surement and calculates also mean arterialblood pressure.
AttributeTask Force Monitor 3040 (new)BioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
Physical description / functioningThe device architecture of the TFM is equivalent to the BioZ.pc. The TFM system consists of a Patient Biosignal Electronic System (PBES) and a PC for calculating and displaying the hemodynamic parameters. ICG signals from a patient are assessed using the same method as the predicate device (8 electrodes, injecting a minimal current). The user inputs the patient's data into the TFM software on the PC including gender, age, height, weight, hematocrit and the distance between the inner ICG electrodes. The PC runs Windows 2000 operating system. The parameters are displayed on a full color screen and measurement data can be stored on the PC's hard disk. The TFM calculates and displays the same hemodynamic parameters as the predicate device.The device consists of a Patient Interface Module and a separate PC. The PC software provides the user interface for the BioZ.pc system and checks the hardware when the software is activated. The user inputs patient parameters into the user software, including patient gender, height, body frame size, weight, age and blood pressure. The Monitor then utilizes these parameters and measures the ICG signals to determine the hemodynamic properties of that particular patient. The PC runs Windows 98 operating system. The parameters are displayed on a full color screen and measurement data can be stored on the PC's hard disk.This device measures noninvasively arterial blood pressure in the finger using a method originally devised ba Dr. Jan PENAZ. The monitor displays the pressure wave form, digital values of systolic, diastolic and mean pressure as well as pulse rate and a time annotated trend display. The device uses a cathode ray tube (CRT) for displaying the signals and values. To follow the arterial blood pressure the Finapres has a servo valve to control the cuff pressure, the pressure transducer are located in the patient interface module. The finger cuff contains the photo electronic components for measuring a blood Plethysmograph and a bladder for applying pressure to the finger. The bladder is wrapped around the patient's finger and connected to the patient interface module. A pump in the monitor unit provides the finger cuff with required air pressure.

{6}------------------------------------------------

/ CNSystems
The Brain & Heart Company

CNSystems Medizintechnik GmbH
Europe, Austria, 8020 Graz
Baumkircherstria, 8020 Graz
Tel.: +43/316/77 234566-0 Fax: -2
Email: +43//316/77 234566-07 2ystems.at

{7}------------------------------------------------

CNSystems Medizintechnik GmbHEurope, Austria, 8020 GrazBaumkircherstrasse 1Tel.: +43/316/7 23456-0, Fax: -2Email: office@cnsystems.at
Table
CNSystemsThe Brain & Heart CompanyAttributeTask Force Monitor 3040 (new)A difference between the TFM and theBioZ.pc is the blood pressure measure-ment. The predicate device uses manualuser input of blood pressure values oroptional an oscillometric BP measure-ment. The interval of measurement can beset by the user. The TFM system includesthe continuous blood pressure method ofthe Finapres device for assessing onlineand real-time blood pressure values.BioZ.PC (Predicate)Monitoring is accomplished by attaching8 electrodes to the patient, injecting aminimal current through the upper elec-trodes, and reading the return voltagewaveform from the inner electrodes.The device calculates and displays 12hemodynamic parameters includingStroke Volume (SV), Cardiac Output(CO), System Vascular resistance, Veloc-ity Index (VI), Thoracic Fluid Content(TFC), Systolic Time Ratio, Left Ventric-ular Ejection Time (LVET), Pre-EjectionPeriod (PEP), Left Cardiac Work/Index(LCWI), Heart rate (HR), Systolic andDiastolic Blood Pressure (SBP, DBP).Each PC provided will be fully testedwith its BioZ.pc device prior to shipment.Ohmeda 2350 Finapres(Predicate)

ન્ડ 7

{8}------------------------------------------------

AttributeTableBioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
Task Force Monitor 3040 (new)
The Penaz-method for beat-to-beat blood pressure measurement (same as the Finapres device) follows very good the arterial blood pressure trend but the absolute values of BP are not accurate due the physiology of finger arteries. Therefore TFM uses a oscillometric blood pressure device for assessing absolute BP values and continuos BP for trend recording. A difference between the TFM and the Finapres device is the interruption of measurement. While the Finapres interrupts from time to time for re-setting the pressure set point, the TFM has no interruptions because of state-of.the-art electronic components. A clinical study shows, that both devices follow the arterial blood pressure very well with the difference, that the TFM has no interruptions in the signal.Because of the combination of ICG and continuous BP further analysis can be made: spectra of blood pressure variability and Baroreceptor Reflex Sensitivity (BRRS). Both methods have been published many times.
AttributeTask Force Monitor 3040 (new)BioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
Patient Circuitry PackageInternal to TFM instrumentInternal to BioZ.PC instrumentInternal to Finapres instrument
CPUIntel Pentium III of higherIntel 586 or equivalentnone
CPU PackagingExternal PCExternal PCnone
PC Operating SystemWindows® 2000DOS within Windows®98none
PC SoftwarePre-installed and fully tested on every TFM prior to shipment.Provided on a CD-ROMnone
User Interface SoftwareTFM program V1.1 Revision 3BioZ ver 1.52not known
User DisplayExternal PC VGA ScreenExternal PV VGA ScreenInternal cathode ray tube (CRT)
Blood Pressure ElectronicsInternal to TFM instrumentInternal to BioZ.PC instrumentInternal to Finapres instrument
CNSystems Medizintechnik GmbHEurope, Austria, 8020 GrazBaumkircherstrasse 1Tel.: +43/316/7 23456-0, Fax: -2Email: office@cnsystems.at
Table
CNSystemsThe Brain & Heart CompanyAttributeTask Force Monitor 3040 (new)BioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
Performance specificationICG measuring current: 400µA eff., 40 kHz, temporal derivative of impedance dZ/dt: ±10 Ohm/sECG measuring range ±5mV, indication range 30 - 150 beats/min, sampling frequency: 1000Hz (Europe), 1200Hz (USA)oscillometric BP measuring accuracy: ±5mmHg, measuring range: 50 to 250mmHgcontinuous BP measuring accuracy: ±5mmHg (due to oscBP), measuring range: 50 to 250mmHgThe inaccurateness of the ICG signal can raise under following conditions:* sepsis* severe heart insufficiency* aortic valve insufficiency* aortic valve stenosis*patient movement artefactsICG measuring current: 2.5mA rms, 70kHz, no further information available; ECG sampling frequency: 1000Hz (in Europe)The device does not accommodate patients who:* weight more than 340 lb. (164 kg)* weight less than 60 lb. (27.3 kg)* have mean arterial pressure greater than 130mmHg* have late sepsis (e.g., stage 3 and 4)* have aortic insufficiencyBP measuring range: 20 to 260mmHg, measuring accuracy: pressure transducer Linearity: + 2mmHg to -4.5mmHg (+ 0.25kPa to - 0.55kPaHeart rate measuring range: 12 to 175bpm, measuring accuracy: ± 5bpm or ± 5% of reading, whichever is higher
Table
Task Force Monitor 3040 (new)BioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
AttributeThe inaccurateness of the oscillometricblood pressure signal can raise under fol-lowing conditions:* weak pulses* irregular pulses* patient movement artefacts* tremor artefacts* respiratory artefacts* gravel road artefactsThe device does accommodate patientswithin the following limits:* weight: 66.14 - 440.9 lbs (30 - 200 kg)* height: 1.64 - 8.20 feet (50 - 250 cm)
Power Consumption115V versionnominal voltage: 115V / 60Hzpower consumption: 240mA, 25W230V versionnominal voltage: 230V / 50Hzpower consumption: 120mA, 25W115V / 60Hz230V / 50Hz115V versionnominal voltage: 115V / 60Hzpower consumption: 1,5A230V versionnominal voltage: 230V / 50Hzpower consumption: 0,75A

.

:

{9}------------------------------------------------

CNSystems
The Brain & Heart Company 3

CNSystems Medizintechnik GmbH
Europe, Austria, 8020 Graz
Baumkircherstria, 8020 Graz
Tel.: +43/315/77 234565-0 Frax: -2
Email: +43/315/77 234565-0 Frax: -22
Email: office@cns

{10}------------------------------------------------

U60

{11}------------------------------------------------

u (a t

{12}------------------------------------------------

AttributeTask Force Monitor 3040 (new)BioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
Dimensions / weightMain monitor(height x width x depth)4.13in x 18.31in x 11.42in(10.5cm x 46.5 cm x 29cm)Continuous device(height x width x depth)1.38in x 3.31in x 6.46in(3.5cm x 8.4cm x 16.4cm)Oscillometric device(height x width x depth)1.38in x 3.15in x 5.12in(3.5cm x 8cm x 13cm)Weight (including all TFM devices andcables without PC and screen:15.4 lbs (7 kg)Dimensions(width x height x depth)14.0in x 12.1in x 7.0in(35.6cm x 30.8cm x 17.8cm)weight: 12 lbs (5.5kg)Monitor unit(height x width x depth)6.1in x 11.8in x 16.7in(15.5cm x 30cm x 42.5cm)Patient Interface Module1.63in x 2.75in x 3.9in(4.14vm x 6.98cm x 9.9cm)Weight: 26.2 lbs (11.9 kg)
AttributeTask Force Monitor 3040 (new)BioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
Materials / AccessoryPatient cables: 8.2 ft. (2.5 m)Electrodes: all disposable electrodes are for single use, tested and certified according to ISO 10993-1ICG: 4 electrodes* Polyethylene foam plastic with a medical adhesive made of polyacrylate* composite film made of 38µm polyester film and 12µm aluminium film* the electro conductive adhesive consists of polyacrylate, polyole, water and electrolytethe electrodes are latex free and PVC freeECG: 4 electrodes(510(k) number: K853939)polyethylene foam plastic, 1mm with a medical adhesive made of polyacrylate sensor system: the stud (upper part) is made of stainless steel, the eyelet (lower part) is made of ABC plastics with 0,0025mm Ag/AgCl-coating a quick-recovery conductive gel containing 3% KCl* a little sponge drenched in gelPatient cable: 10ft (3m)Electrodes: Four Pre-Gelled, Disposable Dual SensorsPatient cable: 16ft (5m)Finger cuff: plastic material
Task Force Monitor 3040 (new)BioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
Attribute* a label made of polyethylene* the cover foil is made of polyethylene, 40 - 70µm* the cover for the sponge is made of polyethyleneThe ECG electrodes are latex free.
Finger cuff and fixing cuffs:The fabric is tested according to ISO 10993 and is used by Borch Textile Group for their Surgical Gowns (FDA 510(k): K954674).Oscillometric cuff: The TFM uses BP cuffs from Rudolf Riester GmbH & Co. KG, Jungingen, Germany. These cuffs are generally used by Riester BP devices (for example K002954)
Table
AttributeTask Force Monitor 3040 (new)BioZ.PC (Predicate)Ohmeda 2350 Finapres (Predicate)
Standards* IEC 60601-1, Medical electrical equipment, General requirements for safety* IEC 60601-1-1, 1. Collateral Standard: Safety requirements* IEC 60601-1-2, 2. Collateral Standard: Electromagnetic compatibility - Requirements and tests* IEC 60601-1-4, 4. Collateral Standard: Programmable electrical medical systems* IEC60601-2-25, Part 2: Particular requirements for the safety of electrocardiographs* IEC 60601-2-30, Part 2: Particular requirements for the safety of automatic cycling indirect blood pressure monitoring equipment.* EN 1060-1, Non-invasive sphygmomanometers - Part 1: General requirements* EN 1060-3, Non-invasive sphygmomanometers - Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems* ANSI/AAMI SP 10, Electronic or automated sphygmomanometers* ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and* IEC 601-1* UL2601* CSA 622.2

Chapter 6: Page 10 (13)

UK

{13}------------------------------------------------

2

ל לוח

{14}------------------------------------------------

stems The Brain & Heart Company

在线

CNSystems Medizintechnik GmbH
Europe, Austria, B020 Graz
Baumkircherstria, B020 Graz
Tel.: +43/3156/7 2345640 Frax: -2
Email: office@cnsystems.at


{15}------------------------------------------------

:

Chapter 6: Page 13 (13)

§ 870.2770 Impedance plethysmograph.

(a)
Identification. An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.(b)
Classification. Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.