(81 days)
CP Medical's synthetic, absorbable placement sleeve accessory is intended to be used with absorbable seeding spacers to facilitate the introduction of radionuclide seeds and spacers into the body during Brachytherapy procedures. It is used to orient, hold, carry and maintain spacing of the radionuclide seeds and the spacer component.
The CP-Sleeve is indicated for use in soft tissues or organ tissue but should not be used during cardiovascular or neurological procedures.
The CP Medical CP-SLEEVE consists of absorbable polymer or copolymer material, braided and non-braided, which is used with absorbable seeding spacers to facilitate the introduction of radionuclide seeds and spacers into the body during Brachytherapy procedures.
This document is a 510(k) premarket notification for a medical device called the "CP-Sleeve," an accessory for brachytherapy procedures. It declares the device to be substantially equivalent to previously marketed predicate devices.
Here's an analysis based on your requested information:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not contain a table of acceptance criteria or reported device performance data in the typical sense of quantitative metrics for device functionality or clinical outcomes. This submission is a 510(k), which focuses on demonstrating substantial equivalence to a predicate device, rather than providing extensive de novo performance data.
The "performance" described is in terms of the device's functional characteristics and intended use being similar to the predicate:
| Acceptance Criteria (Implied by 510(k)) | Reported Device Performance (Summary) |
|---|---|
| Material Composition | Comprised of absorbable polymer or copolymer material, braided and non-braided. |
| Functionality | Used with absorbable seeding spacers to facilitate the introduction of radionuclide seeds and spacers into the body. |
| Mechanism of Action | Used to orient, hold, carry, and maintain spacing of radionuclide seeds and the spacer component. |
| Intended Use Compatibility | Intended for use in brachytherapy procedures, specifically in soft tissue or organ tissue (not cardiovascular or neurological). |
| Sterility/Use | Supplied non-sterile as a single-use device. |
| Substantial Equivalence | Demonstrated to be substantially equivalent to the I-125 Rapid Strand's accessory Sleeve (K940632 and K010821) and the CP Medical Absorbable Seeding Spacer component (K010621). |
2. Sample Size Used for the Test Set and Data Provenance
The document does not describe a specific "test set" or provide sample sizes for performance studies. 510(k) submissions typically rely on a comparison to predicate devices, and sometimes include bench testing or limited clinical data if necessary to prove equivalence. This document primarily focuses on the description of the device and its intended use in relation to predicates. Therefore, there's no information on:
- Sample size for a test set.
- Country of origin of data.
- Retrospective or prospective nature of data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Since there is no described test set or performance study involving human interpretation of data, there is no information about experts used to establish ground truth.
4. Adjudication Method for the Test Set
As no test set involving human interpretation is described, there is no information on an adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study, and consequently any associated effect size of human readers improving with AI assistance, was not performed or reported in this 510(k) submission. This is a traditional medical device (an accessory sleeve), not an AI/software device.
6. Standalone (Algorithm Only) Performance Study
As this is a physical medical device and not an algorithm/software, a standalone AI algorithm performance study was not conducted.
7. Type of Ground Truth Used
Given the nature of this 510(k) submission for a physical accessory device, the "ground truth" is established through bench testing and comparison to the known performance and characteristics of the predicate devices. There isn't a "pathology, outcomes data, etc." ground truth in the context of an algorithm or diagnostic device. The ground truth here is the established safety and effectiveness of the legally marketed predicate devices, which the CP-Sleeve claims to be substantially equivalent to in material, function, and intended use.
8. Sample Size for the Training Set
There is no mention of a training set or its sample size. This type of information is relevant for AI/machine learning models, not for a physical brachytherapy sleeve.
9. How the Ground Truth for the Training Set Was Established
As there is no training set, there is no information on how its ground truth was established.
{0}------------------------------------------------
and
Summary of Safety and Effectiveness
In accordance with the Food and Drug Administration Rule to implement provisions of the in dood dails with the t of 1990 and in conformance with 21 CFR 807, this is to serve oale Medical Donseer of Effectiveness for the proposed CP Medical CP-SLEEVE component device.
| Manufacturer: | CP Medical, Inc.2414 NE Pacific AvenuePortland, OR 97232PHONE: (503) 232-1555FAX: (503) 230-9993 |
|---|---|
| Contact Person: | Mary Ann Greenawalt, DirectorRegulatory and Quality |
| Device Name: | |
| Trade Name: | Brachytherapy Sleeve Accessory |
| Common Name: | Accessory to seed and spacer componentsradionuclide brachytherapy Source device |
| Proprietary name: | CP-Sleeve |
System, applicator, radionuclide, Classification: manual & Source, brachytherapy, radionuclide (accessory to)
November 29, 2001 Date Prepared:
Predicate Device: The predicate device to the CP Medical CP-SLEEVE accessory device is the I-125 Rapid Strand 's accessory Sleeve (K940632 and K010821) for the brachytherapy application system and the CP Medical Absorbable Seeding Spacer component (K010621)
Device Description: The CP Medical CP-SLEEVE consists of absorbable polymer or copolymer material, braided and non-braided, which is used with absorbable seeding spacers to facilitate the introduction of radionuclide seeds and spacers into the body during Brachytherapy procedures.
Intended Use: CP Medical's synthetic, absorbable placement sleeve accessory is intended to be used with absorbable seeding spacers to facilitate the introduction of radionuclide seeds and spacers into the body during Brachytherapy procedures. It is used to orient, hold, carry and maintain spacing of the radionuclide seeds and the spacer component.
{1}------------------------------------------------
Indications: The CP Medical CP-SLEEVE component device is indicated for use as a accessory in brachytherapy procedures. It is supplied non-sterile as a singleuse device. The CP-Sleeve is indicated for use in soft tissue or organ tissue but should not be used during cardiovascular or neurological procedures.
Comparison of Technological Characteristics: The proposed device, the CP Medical CP-SLEEVE is comprised of an absorbable suture material and is intended to facilitate the loading, delivery and implantation of radioactive seeds and spacers. Similarly, the nredicate device is an absorbable suture material intended to facilitate the loading, delivery and implantation of radioactive seeds and spacers.
end
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal is circular and contains the department's name around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it and a pair of wings at the top.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 2 2002
Ms. Mary Ann Greenawalt Director Regulatory and Quality CP Medical P.O. Box 6724 PORTLAND OR 97208
Re: K013975
Trade/Device Name: CP-Sleeve™ Placement Sleeve for Brachytherapy Procedures Regulation Number: 21 CFR 892.5730 Regulation Name: Radionuclide brachytherapy source Regulatory Class: II Product Code: 90 KXK Dated: November 29, 2001 Received: December 3, 2001
Dear Ms. Greenawalt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 8xx.1xxx | (301) 594-4591 |
|---|---|
| 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
| Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of
510(k) Number: K013975
Device Name(s): CP-Sleeve
Intended Use(s) of the Device:
CP Medical's synthetic, absorbable placement sleeve accessory is intended to be used with absorbable seeding spacers to facilitate the introduction of radionuclide seeds and spacers into the body during Brachytherapy procedures. It is used to orient, hold, carry and maintain spacing of the radionuclide seeds and the spacer component.
The CP-Sleeve is indicated for use in soft tissues or organ tissue but should not be used during cardiovascular or neurological procedures.
Please do not write below this line - continue on another page if
necessary
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K013975
510(k) Number
Prescription Use
Or
Over-The-Counter Use
(per 21 CFR 801.109)
N/A