K Number
K013757

Validate with FDA (Live)

Date Cleared
2002-01-08

(56 days)

Product Code
Regulation Number
862.3030
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For the quantitative determination of toxic levels of acetaminophen in human serum or plasma on automated clinical chemistry analyzers. Measurements obtained by this device are used in the diagnosis and treatment of acctaminophen overdose.

Device Description

The Roche Acetaminophen assay contains an in vitro diagnostic reagent system indicated for the quantitative determination of toxic levels of acetaminophen in human serum or plasma on automated clinical chemistry analyzers. The proposed labeling indicates the Roche/Hitachi 911, 912, 917, and Modular P analyzers can be used with the Roche Acetaminophen reagent kits.

AI/ML Overview

Here's an analysis of the provided text regarding the Roche Acetaminophen Assay, focusing on acceptance criteria and supporting studies:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" for the Roche Acetaminophen Assay. However, it does present performance characteristics for comparison to the predicate device. For the purpose of this response, I will infer the acceptance criteria are implicitly defined by the acceptable performance compared to the predicate and the general requirements for such assays (e.g., adequate precision and method correlation).

Performance CharacteristicAcceptance Criteria (Implied)Reported Roche Acetaminophen Performance
PrecisionCV% comparable to or better than predicate device.Level 1: Total CV% = 5.7% (Predicate Level 1: 7.5%)
Level 2: Total CV% = 1.4% (Predicate Level 2: 4.4%)
Level 3: Total CV% = 4.9% (Predicate Level 3: 4.9%)
Method ComparisonGood correlation (R value) with predicate device. Slope and intercept indicating substantial equivalence.Y = -0.31 + 0.987x (vs. Roche COBAS INTEGRA Acetaminophen)
R = 0.999 (vs. Roche COBAS INTEGRA Acetaminophen)
Range = 1.2 to 160.6 ug/mL
Lower Detection LimitAcceptable for clinical use of toxic levels.Not explicitly stated for the device itself, but the range of method comparison starts at 1.2 ug/mL.
SpecificityNot explicitly detailed in the summary."All of the evaluation studies gave acceptable results compared to the predicate device."
Interfering SubstancesNot explicitly detailed in the summary."All of the evaluation studies gave acceptable results compared to the predicate device."

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size:
    • For comparison against the Roche COBAS INTEGRA Acetaminophen assay on the COBAS Integra 700: N = 150
    • For comparison of the predicate device (Roche COBAS INTEGRA Acetaminophen) against the Abbott TDx Acetaminophen assay: N = 87 (This refers to data for the predicate, not the new device's primary test set)
  • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable to this type of device. The Roche Acetaminophen Assay is an in vitro diagnostic (IVD) quantitative assay. Its "truth" is established by comparing its measurements to a reference method (the predicate device) or a gold standard method, not by expert interpretation.

4. Adjudication Method for the Test Set

This is not applicable as there is no human interpretation or subjective assessment involved that would require adjudication for this IVD assay. The comparison is based on quantitative analytical results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This is not applicable. The Roche Acetaminophen Assay is an IVD device for quantitative chemical analysis, not an AI-assisted diagnostic imaging or classification tool that would involve human "readers."

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The performance of the Roche Acetaminophen Assay, as described, is inherently a "standalone" performance in the context of an IVD. It generates a quantitative result without human intervention in the analytical process. The study described (method comparison) assesses this standalone analytical performance against a known method.

7. The Type of Ground Truth Used

The "ground truth" for the performance evaluation was established by comparing the results of the new Roche Acetaminophen Assay against a predicate device (Roche COBAS INTEGRA Acetaminophen Assay), which is itself an already validated and legally marketed device. In essence, the predicate serves as the reference method or "gold standard" for this comparative study.

8. The Sample Size for the Training Set

This information is not provided as there is no mention of a "training set" in the context of this chemical assay. Chemical assays are developed and validated through analytical studies, not typically machine learning training protocols.

9. How the Ground Truth for the Training Set Was Established

This is not applicable as there is no training set described for this type of device.

{0}------------------------------------------------

JAN 8 2002

Roche Acetaminophen Assay

510(k) Summary
----------------

510(k) Summary

K013757

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

1) Submitter name, address, contactRoche Diagnostics Corporation 9115 Hague Rd. Indianapolis, IN 46250 (317) 845-2000
Contact Person:Mike Flis
Date Prepared:November 9, 2001
2) Device nameRoche Acetaminophen
--------------------------------------------
3) Predicate deviceWe claim substantial equivalence to the COBAS INTEGRA Acetaminophen Assay.
--------------------------------------------------------------------------------------------------------
4) Device DescriptionThe Roche Acetaminophen assay contains an in vitro diagnostic reagent system indicated for the quantitative determination of toxic levels of acetaminophen in human serum or plasma on automated clinical chemistry analyzers. The proposed labeling indicates the Roche/Hitachi 911, 912, 917, and Modular P analyzers can be used with the Roche Acetaminophen reagent kits.
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Continued on next page

{1}------------------------------------------------

510(k) Summary, Continued

Y = -0.31 + 0.987x

Range = 1.2 to 160.6 ug/mL

Level 1

7.9

5.6

5.7

5.7

R =0.999

Precision:

Mean (μg/mL)

CV% (withinnun)

CV% (day to day)

CV% (total)

)

  1. Intended use For the quantitative determination of toxic levels of acetaminophen in human serum or plasma on automated clinical chemistry analyzers.

  2. Comparison to predicate device The Roche COBAS INTEGRA Acetaminophen was evaluated for several performance characteristics, including precision, lower detection limit, method comparison, specificity, and interfering substances. All of the evaluation studies gave acceptable results compared to the predicate device. These experiments provide evidence that the Roche Acetaminophen Assay is substantially equivalent to the currently marketed Roche COBAS INTEGRA Acetaminophen Assay. The following table presents the precision and method comparison results.

Roche AcetaminophenRoche COBAS INTEGRA Acetaminophen, (Predicate)
Versus Roche COBAS INTEGRA Acetaminophen assay on the COBAS Integra 700Versus Abbott TDx Acetaminophen assay
N = 150N=87
Y=1.047x – 6.689

R=0.984

Level 1

9.9

న.8

3.7

7.5

Level 3

102.13

0:5

1.4

1.4

Level 2

34.83

1.2

1.0

ાં .ડ

Range = 0-234.7μg/mL

Level 2

32.9

0.9

4.0

4.4

Level 3

97.4

0.7

4.3

4.9

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms. The symbol is composed of three curved lines that suggest movement or flight.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. Mike Flis Regulatory Affairs Principal Roche Diagnostics Corporation 9115 Hague Road P.O. Box 50457 Indianapolis, IN 46250-0457

JAN 8 2002

Re: K013757

Trade/Device Name: Roche Acetaminophen Assay Regulation Number: 21 CFR 862.3030 Regulation Name: Acetaminophen test system Regulatory Class: Class II Product Code: LDP Dated: November 9, 2001 Received: November 13, 2001

Dear Mr. Flis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Roche Diagnostics Corporation

510(k) Number (if known): K013757 Device Name: Roche Acetaminophen Assay Indications for Use:

For the quantitative determination of toxic levels of acetaminophen in human serum or plasma on automated clinical chemistry analyzers. Measurements obtained by this device are used in the diagnosis and treatment of acctaminophen overdose.

-Off)
Division. Clinical Laboratory Devices
510(k) Number K013757

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

Prescription Use V (Per 21 CFR 801.109) Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

§ 862.3030 Acetaminophen test system.

(a)
Identification. An acetaminophen test system is a device intended to measure acetaminophen, an analgestic and fever reducing drug, in serum. Measurements obtained by this device are used in the diagnosis and treatment of acetaminophen overdose.(b)
Classification. Class II.