K Number
K013529

Validate with FDA (Live)

Date Cleared
2002-01-09

(78 days)

Product Code
Regulation Number
872.3690
Panel
Dental
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Point 4 Translucent Modified is a dental composite restorative material intended to be used in all classes of cavities

Device Description

The device is a micro-hybrid light cured resin-based dental restorative which contains approximately 79% by weight (59% by volume) inorganic filler but with a particle size nearly half that of Kerr's Herculite XRV and Prodigy. This breakthrough in filler technology provides a higher, longer-lasting polish similar to a microfill, without any negative effects to the physical properties.

AI/ML Overview

This looks like a 510(k) summary for a dental composite restorative material, Point 4 Translucent Modified. The provided text does not contain information about acceptance criteria or a study that proves the device meets those criteria in the way you've outlined for a medical device with performance metrics like sensitivity, specificity, or accuracy.

Instead, this document focuses on establishing substantial equivalence to an already legally marketed predicate device (Kerr Corporation, Prodigy 4 Translucent Shades). For devices seeking substantial equivalence, the "acceptance criteria" are generally met by demonstrating that the new device has the same intended use and similar technological characteristics (e.g., composition, physical properties) as the predicate device, or that any differences do not raise new questions of safety and effectiveness.

Therefore, I cannot fill out your requested table and study details as the information is not present in the provided text. The document does not describe a clinical study or a performance study with specific metrics like those typically associated with diagnostic or AI-driven medical devices.

Here's why the information you're looking for isn't here:

  • Device Type: This is a dental restorative material (a composite filling), not a diagnostic device or a device with algorithms that generate performance metrics like sensitivity or specificity. Its "performance" is generally evaluated through its physical and chemical properties, biocompatibility, and clinical handling, rather than algorithmic accuracy.
  • Regulatory Pathway: This is a 510(k) submission, which aims to demonstrate substantial equivalence to an existing device. This pathway often relies on comparison to a predicate device's characteristics and established clinical use, rather than requiring novel clinical trials with specific performance endpoints and ground truth adjudication as you've described.
  • Missing Information: The provided text only includes a 510(k) summary, a cover letter from the FDA, and an "Indications for Use" statement. It does not include the detailed technical data, test results, or study reports that would typically contain the information you're asking for. These would usually be in other sections of the full 510(k) submission.

If this were a submission for a diagnostic device or an AI-driven medical device, the following would be applicable, but it's not for this particular product based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance:

  • This table would contain metrics like sensitivity, specificity, accuracy, positive predictive value, negative predictive value, AUC, etc., along with the predetermined acceptance thresholds for these metrics. The reported device performance would be the actual results from a clinical or analytical validation study.

2. Sample size used for the test set and the data provenance:

  • For an AI device, this would detail the number of images/cases in the independent test set, and whether they were collected retrospectively or prospectively, and from which geographic regions or institutions.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • This refers to the clinicians (e.g., board-certified radiologists, pathologists) who reviewed the test set data to establish the definitive diagnosis or outcome for each case. Their experience level is crucial.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • Describes how discrepancies among experts in establishing ground truth were resolved (e.g., two experts agree, or a third expert adjudicates if the first two disagree).

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • This would describe a study where human readers interpret cases both with and without AI assistance, allowing for a comparison of their diagnostic performance (e.g., sensitivity, specificity, reading time) and quantifying the AI's impact.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • This would detail the performance metrics of the AI algorithm operating independently, without any human input or override.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Explains the "gold standard" used to verify the correctness of the device's outputs (e.g., confirmed pathological diagnosis, long-term patient outcomes, or the consensus of highly qualified experts).

8. The sample size for the training set:

  • The number of cases/data points used to train the AI model.

9. How the ground truth for the training set was established:

  • Describes the methodology for annotating and confirming labels for the data used to train the AI (similar to point 7, but for the training phase).

In summary, for the provided document about "Point 4 Translucent Modified," the described information is not applicable because it's a materials science submission based on substantial equivalence, not a performance study for a diagnostic or AI-driven device.

{0}------------------------------------------------

4013529

sds

SYBRON DENTAL SPECIALTIES

JAN 0 9 2002

Section III - 510(k) Summary of Safety and Effectiveness

Submitter:

Sybron Dental Specialties, Inc. 1717 W. Collins Avenue Orange, California 92867 (714) 516-7484 - Phone (714) 516-7488 - Facsimile Colleen Boswell - Contact Person

Date Summary Prepared: October 2001

Device Name:

  • Trade Name Point 4 Translucent Modified .
  • Common Name -- Dental Composite Restorative Material .
  • Classification Name Tooth Shade Resin Material, per 21 CFR § 872.3690 .

Devices for Which Substantial Equivalence is Claimed:

  • Kerr Corporation, Prodigy 4 Translucent Shades .

Device Description:

The device is a micro-hybrid light cured resin-based dental restorative which contains approximately 79% by weight (59% by volume) inorganic filler but with a particle size nearly half that of Kerr's Herculite XRV and Prodigy. This breakthrough in filler technology provides a higher, longer-lasting polish similar to a microfill, without any negative effects to the physical properties.

Intended Use of the Device:

The intended use of Point 4 Translucent Modified is for use in all classes of cavities.

Substantial Equivalence:

Point 4 Translucent Modified is substantially equivalent to other legally marketed devices in the United States. The dental composite restorative material marketed by Kerr Corporation functions in a manner similar to and is intended for the same use as the product manufactured by Kerr Dental Materials Center.

{1}------------------------------------------------

Public Health Service

Image /page/1/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is an abstract image of a bird.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Colleen Boswell Sybron Dental Specialties, Incorporated 1717 West Collins Avenue Orange, California 92867

JAN 0 9 2002

Re: K013529

Trade/Device Name: Point 4 Translucent Modified Regulation Number: 872.3690 Regulation Name: Dental Composite Restorative Material Regulatory Class: II Product Code: EBF Dated: October 18, 2001 Received: October 23, 2001

Dear Ms. Boswell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{2}------------------------------------------------

Page 2 - Ms. Boswell

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Section I


Indications for Use Statement

Ver/3 - 4/24/96
-----------------
Applicant:Kerr Dental Material Center
510(k) Number (if known):K013529
Device Name:Point 4 Translucent Modified
Indications For Use:

Point 4 Translucent Modified is a dental composite restorative material intended to be used in all classes of cavities

Susan Humon

(Division Sign-Off)

Division of Dental, Infection Control,

General Hospital Devices

NumberK018309
-----------------

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Per 21 CFR 801.109)

(Optional Format 1-2-96)

§ 872.3690 Tooth shade resin material.

(a)
Identification. Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.(b)
Classification. Class II.