(63 days)
The Fixion™ Interlocking Intramedullary Nail is intended for use in fixation of various types of fractures, including diaphyseal fractures and short distal or proximal fragments in the long bones.
The Fixion Interlocking Intramedullary Nailing System consist of the following main components: 1. The Nail Implant is an expandable non-slotted stainless steel cylindrical tube, with a cap protected, female threaded proximal end with holes for interlocking screws to lock the nail in the bone. The Nail Implant may also have both proximal and distal ends with holes for interlocking screws. 2. The Instrument Set consists of a few accessories used during insertion and removal (if required) of the Nail Implant. 3. The Inflation Device is a single-use, manual, plastic pump to be filled with sterile inflation liquid. Once the nail is positioned within the medullary canal, rotation of the pump handle allows for nail diameter increase to its intended diameter under x-ray and controlled pressure. Interlocking screws may be used to lock the nail inside the bone.
The provided text K013449 describes the Fixion Interlocking Intramedullary Nailing System. However, this document does not contain information about acceptance criteria or a study proving device performance in the context of diagnostic accuracy, which is typically associated with AI/ML devices.
Instead, the document focuses on:
- Materials and Mechanical Performance for a physical medical device (an intramedullary nail).
- Substantial equivalence to a predicate device.
- Intended use and system description.
Therefore, I cannot provide the requested information in the format of a diagnostic performance study, as the provided input does not pertain to such a device or study.
Here's a breakdown of why each requested point cannot be addressed with the provided text:
- A table of acceptance criteria and the reported device performance: The document states the device meets ASTM F138 (material spec), ASTM F565 (care/handling), and ASTM F1264-99 (4-point bending mechanical testing). These are standards the device adheres to, not specific performance metrics or acceptance criteria as would be defined for a diagnostic AI model (e.g., sensitivity, specificity, AUC). There are no reported numerical performance metrics from a study in this context.
- Sample sized used for the test set and the data provenance: Not applicable. This document describes a physical implant, not a diagnostic algorithm.
- Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical implant's mechanical properties is established through material testing and engineering standards, not expert clinical review of data.
- Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
- If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML diagnostic device.
- If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
- The type of ground truth used (expert concensus, pathology, outcomes data, etc): For a physical device, the "ground truth" would be adherence to material specifications (ASTM F138) and mechanical performance standards (ASTM F1264-99).
- The sample size for the training set: Not applicable. This device does not use a training set in the context of AI/ML.
- How the ground truth for the training set was established: Not applicable.
In summary, the provided document describes a traditional medical device (an intramedullary nailing system) and its compliance with established material and mechanical testing standards, not an AI/ML-driven diagnostic device with associated performance studies.
{0}------------------------------------------------
K0134/9
510(K) Summary
12-19-01
Disc-O-Tech Medical Technologies, Ltd. Fixion Interlocking Intramedullary Nailing System
Company Name
Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St., Herzelia, 46728, Israel
Submitter's Name and Contact Person
Elad Magal Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot St., Herzelia, 46728, Israel Tel: +972 9 9511511 Fax: +972 9 9548939
Date Prepared
October 2001
Trade/Proprietary Name
Fixion 1 1 Interlocking Intramedullary Nailing System (Fixion IL Nail)
Classification Name
Intramedullary Fixation Rod 21 CFR & 888.3020 Class II
Predicate Devices
-
- Fixion Interlocking Intramedullary Nailing System (Fixion IL Nail) by the company -K002783
-
- Unreamed Humeral Nail by Synthes K933518
-
- Unreamed Tibial Nail by Synthes K914453
-
- Unreamed Femoral Nail by Synthes K923580
-
- Ti Distal Femoral Nail by Synthes K970733
-
- True/Flex Upper Extremity IM Nail by Encore (Applied Osteo Systems, Inc.) -K902264
Performance Standards
The following standards were used:
-
- The Fixion IL Nail is manufactured from 316L Stainless Steel, which meets the requirements of ASTM F138 - Standard Specification for Wrought 18 Chromium - 14 Nickel - 2.5 Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS
Page 23 of 38
- The Fixion IL Nail is manufactured from 316L Stainless Steel, which meets the requirements of ASTM F138 - Standard Specification for Wrought 18 Chromium - 14 Nickel - 2.5 Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS
{1}------------------------------------------------
S31673).
-
- The Fixion IL Nailing System accessories incorporate surgical grade stainless steel and silicone.
-
- The Fixion IL Nail is designed to meet the requirements of ASTM F565 Standard practice for Care and Handling of Orthopedic Implants and Instruments.
-
- The 4 point bending mechanical testing was performed according to ASTM F1264-99 - Standard for Mechanical Performance Considerations for Intramedullary Fixation Devices.
Intended Use
The Fixion™ Interlocking Intramedullary Nail is intended for use in fixation of various types of fractures, including diaphyseal fractures and short distal or proximal fragments in the long bones.
System Description
The Fixion Interlocking Intramedullary Nailing System consist of the following main components:
-
The Nail Implant is an expandable non-slotted stainless steel cylindrical tube, with a cap protected, female threaded proximal end with holes for interlocking screws to lock the nail in the bone. The Nail Implant may also have both proximal and distal ends with holes for interlocking screws.
-
The Instrument Set consists of a few accessories used during insertion and removal (if required) of the Nail Implant.
-
The Inflation Device is a single-use, manual, plastic pump to be filled with sterile inflation liquid.
Once the nail is positioned within the medullary canal, rotation of the pump handle allows for nail diameter increase to its intended diameter under x-ray and controlled pressure. Interlocking screws may be used to lock the nail inside the bone.
Substantial Equivalence
The Fixion IL Nail is substantially equivalent to the Fixion IL Nail currently cleared for marketing under 510(k) K002783.
The modified Fixion IL Nail has the following similarities to that which previously received 510(k) concurrence:
-
Has an equivalent intended use
-
Has basically the same operating principles
-
Incorporates similar design
-
Incorporates the same materials and processes
-
Is sterilized and packed, basically, in the same manner.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the department's name around the perimeter. Inside the circle is a stylized symbol that resembles a bird or a flame, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20050
DEC 1 9 2001
Mr. Elad Magal General Manager Disc-O-Tech Medical Technologies, Ltd. 3 Hasadnaot Street Herzelia, 46728 Israel
Re: K013449
Trade/Device Name: Fixion Interlocking Intramedullary Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: II Product Code: HSB Dated: October 15, 2001 Received: October 17, 2001
Dear Mr. Magal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. Elad Magal
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
h
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
page 1 of 1
Indication for Use
510(K) Number (if known): K013449
Device Name: Fixion™ Interlocking Intramedullary Nailing System (Fixion™ II. Nail)
Indication for Use: The Fixion™ Interlocking Intramedullary Nail is intended for use in fixation of various types of fractures, including diaphyseal fractures and short distal or proximal fragments in the long bones.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use 1 (per 21 CFR 801.109)
OR
Over the Counter Use No
.
Division Sign-Off) Division of General, Restorative and Neurological Devices
510(K) Number K013449
Page 26 of 38
3
- CONFIDENTIAL -
N/A