K Number
K013407

Validate with FDA (Live)

Manufacturer
Date Cleared
2001-12-06

(52 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A home drug screening, and if needed, confirmation service. This kit provides a / chome arage coroening, and intion/presence of the following drugs of abuse in urine: Opiates.

Device Description

Immunoassay for the qualitative detection of opiates in urine.

AI/ML Overview

Here's an analysis of the provided text regarding the "At Home Drug Test (Model 9084)":

Acceptance Criteria and Device Performance Study for At Home Drug Test (Model 9084)

The "At Home Drug Test (Model 9084)" is an immunoassay for the qualitative detection of opiates in urine, intended for home use to assist in preventing drug abuse, with a mechanism for anonymous confirmation testing at a SAMHSA certified laboratory.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of specific thresholds for sensitivity, specificity, or accuracy that the device had to meet. Instead, it describes performance relative to predicate devices and established laboratory methods. The reported performance suggests that the implied acceptance criteria were likely based on demonstrating high correlation and overall accuracy comparable to these established methods and devices.

MetricAcceptance Criteria (Implied/Relative)Reported Device Performance
Opiates Cut-offN/A (Defined as a test parameter)300 ng/ml
CorrelationSubstantially equivalent to Behring EMIT II and GC/MS>99% correlation when compared to Behring EMIT II and GC/MS methodology (using clinical specimens)
Overall AccuracyExcellent, comparable to commercially available tests (professional use)>98% overall accuracy (in the hands of professional users)
Overall AccuracyExcellent (consumer use)Excellent overall accuracy; 97.3% accuracy for cocaine (Note: This specific data point is for cocaine, not opiates, but cited as example of consumer study. General "excellent overall accuracy" implied for opiates in consumer use.)

2. Sample Size and Data Provenance

  • Test Set Sample Size: The document does not explicitly state the exact sample size for the clinical sample correlation study or the blind-labeled spiked study. It mentions "clinical specimens" and "clinical studies." For the consumer study, it provides a specific number for a cocaine test: 943 (918/943 or 97.3%).
  • Data Provenance: The studies were performed at "two independent laboratories" and involved "clinical specimens." The manufacturer is based in San Diego, California, USA, suggesting the studies likely took place in the USA, although this is not explicitly stated. The studies were retrospective or prospective is not specified.

3. Number of Experts and Qualifications for Ground Truth

The document does not specify the number or qualifications of experts used to establish the ground truth. It mentions that clinical studies were performed at "two independent laboratories" and that expert methods like "Behring EMIT II" and "GC/MS methodology" were used as comparators for ground truth. These are established laboratory techniques, implying that the personnel operating them would be qualified laboratory professionals, but no specific details are provided.

4. Adjudication Method

The document does not describe any specific adjudication method (e.g., 2+1, 3+1). The ground truth appears to be established by comparison with standard laboratory methods (Behring EMIT II and GC/MS) rather than a consensus of human reviewers for each case.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being performed. The studies focus on the standalone performance of the device against established laboratory methods and, in a separate consumer study, its accuracy when interpreted by lay users. There is no information about human readers improving with or without AI assistance, as this is a non-AI, immunoassay-based drug test.

6. Standalone Performance Study (Algorithm Only Without Human-in-the-Loop)

Yes, a standalone performance study was done. The "product performance characteristics... were evaluated in a clinical sample correlation study and a blind labeled spiked study." The results (e.g., >99% correlation and >98% accuracy) reflect the device's performance when used by professionals, likely in a controlled setting, and its direct comparison to "Behring EMIT II and GC/MS methodology"—these are all measures of standalone performance. The consumer study also measures standalone performance when interpreted by lay users.

7. Type of Ground Truth Used

The primary ground truth used was established laboratory methods:

  • Behring EMIT II: A widely accepted immunoassay for drug screening.
  • GC/MS (Gas Chromatography-Mass Spectrometry): Considered a gold standard for drug confirmation due to its high specificity and sensitivity.

8. Sample Size for the Training Set

The document does not mention a "training set" in the context of an algorithm or machine learning. This is a conventional immunoassay, not an AI or machine learning device. Therefore, a training set in that sense is not applicable. The "clinical specimens" and "spiked samples" used in the performance evaluation serve as the data for validating the device's accuracy.

9. How the Ground Truth for the Training Set was Established

As there is no "training set" in the context of machine learning, this question is not directly applicable. For the performance evaluation, the ground truth was established by the established laboratory methods: Behring EMIT II and GC/MS methodology. These methods are themselves highly reliable and considered accurate for detecting opiates.

{0}------------------------------------------------

DEC 0 6 2001

510 (k) SUMMARY AS REQUIRED BY SECTION 807.92(C)

K013407

Identification: At Home Drug Test (Model 9084)

Description: Immunoassay for the qualitative detection of opiates in urine.

Name Of Manufacturer:

Phamatech 9530 Padgett Street, Suite #101 San Diego, California 92126, USA

Intended Use: The At Home Drug Screening Test is a rapid, qualitative immunoassay for the detection of target drugs/metabolites in urine. The cut-off concentration for this test is as follows: Opiates; 300 ng/ml. This assay is intended for use in the home to assist in preventing drug abuse. This kit is designed to incorporate a mechanism for anonymous confirmation testing to be performed at a SAMHSA certified laboratory.

The At Home Drug Test, like many commercially available drug screening test Technology: kits, qualitatively measures the presence of target drug and metabolites by visual color sandwich one step immunoassay technology. Examples of such predicate devices include the ABMC (Ancramdale, NY) and the Applied Biotech SureStep Test (San Diego, CA 92121). All of the above devices rely on the basic immunochemical sandwich assay principle of recognition and formation of specific antibody / target drug / antibody / complexes.

The product performance characteristics of the At Home Drug Test were Performance: evaluated in a clinical sample correlation study and a blind labeled spiked study. The results of these studies demonstrate the Phamatech at Home Drug Test to be substantially equivalent to the reported performance characteristics of other commercially available tests for the qualitative detection of the stated target drugs in urine. Correlation studies, using clinical specimens, produced a >99% correlation when compared to the Behring EMIT II (Cupertino, CA 95014) and GC/MS methodology. Clinical studies, performed at two independent laboratories, were also performed. In them the Phamatech At Home Drug Test exhibited excellent overall accuracy (>98%) in the hands of professional users. A consumer study was also performed, in it the At Home Drug Test exhibited excellent overall accuracy. Consumer interpretation of the cocaine test showed accuracy to be 918/943 or 97.3%

For the reasons mentioned above, it may be concluded that the Phamatech At Conclusion: Home Drug Test is substantially equivalent to a variety of detection tests currently in commercial distribution and is safe in the hands of the lay user.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or flowing lines.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 0 6 2001

Mr. Carl Mongiovi Vice President Phamatech, Inc. 9530 Padgett Street, Suite 101 San Diego, CA 92126

Re: K013407

Trade/Device Name: At Home Drug Test (Model 9084) Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: Class II Product Code: MVO Dated: November 15, 2001 Received: November 19, 2001

Dear Mr. Mongiovi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed notification. THE I Drivin a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at additionally 607.10 for in The Samestions on the promotion and advertising of your device, (201) 59 + 15 the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small mionmation on Jour respond and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

INDICATIONS FOR USE

Applicant: Phamatech

510 (k) Number (if known): __KO13407

Device Name: At Home Drug Test (Model 9084)

Indications for Use:

A home drug screening, and if needed, confirmation service. This kit provides a / chome arage coroening, and intion/presence of the following drugs of abuse in urine: Opiates.

Thomas C. Duda for Jeane Cooper
(Division Sign-Off)

sion of Clinical Laborator 510(k) Number

PLEASE DO NOT WRITE BELOW THIS LINE

Concurrence of the CDRH Office of Device Evaluation (ODE)

Division Sign-off Division of Clinical Laboratory Devices 510 (k) Number:

Prescription Use: Per 21 CFR 801.109 OR

Over the Counter: X

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).