K Number
K013021

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-05-30

(265 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LightLance™ Laser Skin Perforator is intended to be used for ablation of skin tissue to establish capillary blood access. The LightLance™ Laser Skin Perforator general indication for use is for the perforation of skin to draw capillary blood for screening purposes. The device is specifically indicated for obtaining capillary blood samples for subsequent analysis of blood glucose concentration in both institutional and home settings.

Device Description

The LightLance™ Laser Skin Perforator is a pulsed erbium doped yittrium-aluminumgamet (Er: YAG) laser. It outputs a single pulse of laser light with a wavelength of 2940 nm, which ablates a small hole in the skin to a depth sufficient to access capillary blood.

AI/ML Overview

The LightLance™ Laser Skin Perforator is a device designed for ablating skin tissue to access capillary blood, specifically for obtaining blood samples for glucose concentration analysis.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance
SafetyThe device must be safe for its intended use, comparable to legally marketed predicate devices."Thoroughly tested according to international standards equivalent to the standard IEC 601. Results from the testing indicate that the device is as safe... as the predicate device."
EffectivenessThe device must be effective for its intended use (establishing capillary blood access), comparable to legally marketed predicate devices."Thoroughly tested according to international standards equivalent to the standard IEC 601. Results from the testing indicate that the device is as ... effective as the predicate device." "The LightLance™ Laser Skin Perforator has the same intended use and general and specific indications as both of predicated Lasette devices."
Substantial EquivalenceThe device must be substantially equivalent to legally marketed predicate devices (LASETTE™ K983673 and K981746)."The LightLance™ Laser Skin Perforator substantially equivalent to the current, legally marketed LASETTE™ (K983673) and (K981746)." "The minor technological differences do not raise any new questions of safety and effectiveness."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document mentions "Performance data demonstrates that LightLance™ is as safe and effective as the Lasette" and "Results from the testing indicate that the device is as safe and effective as the predicate device." However, the specific number of subjects or tests involved in this performance data is not provided.
  • Data Provenance: Not explicitly stated. The document refers to "international standards equivalent to the standard IEC 601," implying a standardized testing methodology, but does not specify the country of origin of the data or whether the study was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • This information is not provided in the given text. The document refers to "testing" and "results" related to safety and effectiveness but does not detail how ground truth was established for these tests, nor the involvement or qualifications of experts.

4. Adjudication Method for the Test Set

  • This information is not provided in the given text.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC comparative effectiveness study was not done (or at least not described in the provided text). The study focused on demonstrating substantial equivalence to a predicate device through performance testing against international standards, not on comparing human reader performance with and without AI assistance.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done

  • The device is a physical medical device (laser skin perforator), not an AI algorithm. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" does not apply to this product. The device is used by a human to perform its function.

7. The Type of Ground Truth Used

  • Since the device is a physical instrument for producing a physical effect (skin perforation for blood access), the ground truth would likely be established by direct observation and measurement of the device's physical performance characteristics. This would include parameters such as:
    • Depth of perforation.
    • Consistency of perforation.
    • Ability to access capillary blood successfully.
    • Safety parameters (e.g., thermal damage to surrounding tissue, sterility, electrical safety).
  • The document indicates "thoroughly tested according to international standards equivalent to the standard IEC 601," which suggests a technical, objective ground truth based on measurable physical properties and compliance with safety and performance specifications.

8. The Sample Size for the Training Set

  • The device is a physical medical device, not an AI/machine learning algorithm. Therefore, the concept of a "training set" in the context of machine learning does not apply to this product.

9. How the Ground Truth for the Training Set was Established

  • As the device is not an AI/machine learning algorithm, there is no "training set" or "ground truth for the training set" in that context.

{0}------------------------------------------------

MAY 3 0 2002

K013021 1/3

Attachment 11

510(k) Summary

LightLance™ Laser Skin Perforator

Submitted by:

Innotech USA, Inc. Laser Devices Division 2975 Westchester Avenue, Suite 401 Purchase, NY 10577

September 4, 2001

This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

Contact Person:

Howard M. Holstein, Partner Hogan & Hartson L.L.P. Columbia Square 555 Thirteenth Street, NW Washington, DC 20004-1109 Phone: (202) 637-5813 Telefax: (202) 637-5910 E-mail: HMHOLSTEIN@HHLAW.COM

510(k) for the LightLance™ Laser Skin Perforator

0001400

{1}------------------------------------------------

KO13021 2/3

2. Device Name and Classification:

Trade Name:LightLance™ Laser Skin Perforator
Classification Name:Laser instrument, surgical, powered
Common Name:Laser Skin Perforator
Classification Panel:General and Plastic Surgery: Panel 79
CFR Section:21 CFR §878.4810
Device Class:Class II

Substantial Equivalence: 3.

The LightLance™ Laser Skin Perforator substantially equivalent to the current, legally marketed LASETTE™ (K983673) and (K981746). The LightLance™ Laser Skin Perforator has the same intended use and general and specific indications as both of predicated Lasette devices. These devices have very similar principles of operation and technological characteristics. The minor technological differences do not raise any new questions of safety and effectiveness. Performance data demonstrates that LightLance™ is as safe and effective as the Lasette.

Device Description: 4.

The LightLance™ Laser Skin Perforator is a pulsed erbium doped yittrium-aluminumgamet (Er: YAG) laser. It outputs a single pulse of laser light with a wavelength of 2940 nm, which ablates a small hole in the skin to a depth sufficient to access capillary blood.

510(k) for the LightLance™ Laser Skin Perforator

000161

{2}------------------------------------------------

01304 3/3

5. Intended Use/Indications for Use:

The LightLance™ Laser Skin Perforator is intended to be used for ablation of skin tissue to establish capillary blood access, The LightLance™ Laser Skin Perforator general indication for use is for the perforation of skin to draw capillary blood for screening purposes. The device is specifically indicated for obtaining capillary blood samples for subsequent analysis of blood glucose concentration in both institutional and home settings.

Performance Data: 6.

The LightLance™ Laser Skin Perforator has been thoroughly tested according to international standards equivalent to the standard IEC 601. Results from the testing indicate that the device is as safe and effective as the predicate device.

Conclusion: 7.

The LightLance™ Laser Skin Perforator is substantially equivalent to the current legally marketed Cell Robotics Lasette (K983673, K981746).

510(k) for the LightLance™ Laser Skin Perforator

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Innotech USA, Inc. c/o Mr. Howard Holstein Hogan and Hartson L.L.P. 555 Thirteenth Street, NW Washington, D.C. 20004-1109

Re: K013021

Trade/Device Name: LightLance™ Laser Skin Perforator Regulation Number: 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: February 22, 2002 Received: March 5, 2002

Dear Mr. Holstein:

We have reviewed your Section 510(k) premarket notification of intent to market the device wo nave roviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for two based in the May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Howard Holstein

This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ocgin marketing your device of your device to a legally prematket notification. The PDA intellig of successification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific advice for your devices di success, please contact the Office of additionally 21 CFC For 807.10 for in Title undersons on the promotion and advertising of Compliance at (301) 594-4637. Thumbilany for quest (301) 594-4639. Also, please note the your device, picase comace the Office of remarket notification" (21 CFR Part 807.97). regulation entitled, "Misbranding by reference to premarket notifical from the Other general information on your responsibilities under the Act may be obtained from the Other general information on your responsiver and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Miriam C. Provost

for Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INNOTECH US A. II

INDICATION FOR USE STATEMENT

510(k) Number:New Submission K013021
-----------------------------------------

LightLance™ Laser Skin Perforator Device Name:

Indication for Use:

The LightLance™ Laser Skin Perforator is intended to be used for ablation of skin tissue to establish capillary blood access. The LightLance™ Laser Skin Perforator general indication for use is for the perforation of skin to draw capillary blood for screening purposes. The device is specifically indicated for obtaining capillary blood samples for subsequent analysis of blood glucose concentration in both institutional and home settings

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muriam C. Provost

(Division Sign-Off) Division of General, Restorative and Neurological Devices

Prescription Use

510(k) Number K013021
OR Over-the-Counter Use

(per 21 CFR 801.109)

000008

INNOTECH USA, INC., 2975 WESTCHESTER AVENUE, SUITE 401, PURCHASE NY 10577 E-MAIL: innotec@ix.netcom.com TELEPHONE: (914) 251-0700 FAX: (914) 251-0951

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.