K Number
K012796

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-03-14

(205 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Amluck automatic digital wrist blood pressure monitor, Model AK-3000 / AK-4000, is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to be 5.3" - 8.5".

Device Description

AMLUCK AK-3000 / AK-4000 uses the Oscillometric method to measure the blood pressure. The Oscillometric method is adopted clinically to measure the blood pressure recently. It is not needed to use the stethoscope, as in the traditional measuring method, to monitor the Korotkov sound when deciding the systolic or diastolic pressure. The Oscillometric method senses the vibrating signal via the closed air pipe system and utilizes the microcomputer to automatically sense the characteristics of the pulse signal. Through simple calculation, the reading can reflect the accurate real blood pressure, and the systolic pressure is defined as the pressure when the cuff pressure oscillating amplitude begins to increase and the diastolic pressure as the pressure when the cuff pressure oscillating amplitude stops decreasing.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the AMLUCK BLOOD PRESSURE MONITOR AK-3000 / AK-4000, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / TestStandard ReferenceReported Device Performance
Electric SafetyEN 60601-1:1990+A1+A2+A11+A12+A13PASS
EMC ConformityEN 60601-1-2: 1993PASS
Woven Cotton SheetingJIS L 1096 6.39.1.2 Method B 2Certified by SGS UK Ltd.
Performance & ClinicalAAMI / ANSI SP10Implied to meet (no specific performance metrics reported in summary, but indicated as the standard applied)

2. Sample Size Used for the Test Set and Data Provenance

The provided 510(k) summary does not explicitly state the sample size used for the AAMI / ANSI SP10 clinical test or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions that the test was conducted.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the 510(k) summary. For blood pressure monitors tested against AAMI/ANSI SP10, ground truth is typically established by trained observers (experts) using a reference sphygmomanometer, but the number and qualifications are not detailed here.

4. Adjudication Method for the Test Set

This information is not provided in the 510(k) summary. For clinical validation of blood pressure monitors, an adjudication method (such as independent observers and specific error tolerance) is part of the AAMI/ANSI SP10 standard, but the specific details of its application are not included in this document.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not conducted or reported in this 510(k) summary. This type of study is relevant for imaging devices where human readers interpret data with and without AI assistance to assess diagnostic performance. This device is an automated blood pressure monitor, not an imaging device.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study Was Done

Yes, implicitly. The AAMI/ANSI SP10 standard defines the accuracy requirements for automated sphygmomanometers, which involves comparing the device's readings against a reference standard. This is a standalone performance assessment of the algorithm's ability to measure blood pressure. The summary indicates that the device underwent "PERFORMANCE & CLINICAL TEST 3. AAMI / ANSI SP10."

7. The Type of Ground Truth Used

The ground truth for the "Performance & Clinical Test" (AAMI/ANSI SP10) would be established by simultaneous measurements from a trained observer using a calibrated, auscultatory reference sphygmomanometer. The AAMI/ANSI SP10 standard outlines rigorous procedures for this to ensure accurate reference measurements.

8. The Sample Size for the Training Set

The provided 510(k) summary does not provide any information about a training set size. Blood pressure monitors using the oscillometric method are typically developed and calibrated based on physiological models and empirical data, but the specific details of a "training set" in the context of machine learning (as often implied by this question) are not applicable or disclosed for this type of device and submission. The device's underlying algorithm is likely calibrated against known pressure measurements.

9. How the Ground Truth for the Training Set Was Established

As no "training set" in the machine learning sense is explicitly mentioned or relevant for this type of device in the provided summary, information on how its ground truth was established is not available. The oscillometric method itself is based on well-understood physiological principles and extensive validation in clinical settings, with ground truth established through comparative measurements against auscultatory methods.

{0}------------------------------------------------

AMLUCK Enterprises Co. Ltd.

No. 41, Ching Hsiang St., Tu-Cheng, Taipei, Taiwan, ROC 236 Fax: 886-2-22695115 Telephone: 886-2-226955555 Email: amluck@ms7.hinet.net

KO12796

MAR 1 4 2002

"510(k) Summary"

Submitter's Name: AMLUCK ENTERPRISES CO. LTD.

No. 41, Ching Hsiang St., Tu-Cheng, Taipei, Address: Taiwan, ROC 236

Telephone: 886-2-22695555

886-2-22695115 FAX:

Mr. THOMAS CHEN Contact Person:

Date Summary 12/15/2001 Prepared:

AMLUCK BLOOD PRESSURE Proprietary Name: MONITOR AK-3000 / AK-4000

BLOOD PRESSURE MONITOR Common Name:

NON-INVASIVE BLOOD-PRESSURE Classification Name: MEASUREMENT SYSTEM

( per 21CFR section 870.1130)

Device Class: Class II (performance standards)

Specialty: CARDIOVASCULAR

Product code: DXN

WRIST MICROLIFE WATCH BLOOD Legally Marketed PRESSURE MONITOR, MODEL BP-3BU1 ( Predicate ) Device : 510(k) No: K001182

{1}------------------------------------------------

Description of the new device:

AMLUCK AK-3000 / AK-4000 uses the Oscillometric method to measure the blood The Oscillometric method is adopted clinically to measure the blood pressure pressure. It is not needed to use the stethoscope, as in the traditional measuring method, recently. to monitor the Korotkov sound when deciding the systolic or diastolic pressure. The Oscillometric method senses the vibrating signal via the closed air pipe system and utilizes the microcomputer to automatically sense the characteristics of the pulse signal. Through simple calculation, the reading can reflect the accurate real blood pressure, and the systolic pressure is defined as the pressure when the cuff pressure oscillating amplitude begins to increase and the diastolic pressure as the pressure when the cuff pressure oscillating amplitude stops decreasing.

Technological Characteristics of our new device compared to the predicate device:

The technological characteristics of AMLUCK AK-3000 / AK-4000 are substantially equivalent to MICROLIFE BP 3BU1. AMLUCK AK-3000 / AK-4000 is of generally the same form and intended to be used in the same manner as the substantially equivalent products, MICROLIFE BP 3BU1.

{2}------------------------------------------------

Email: amluck(@ms7.hinet.net

Test Summary:

ELECTRIC SAFETY and EMC test reports, 1.

General safetyEN 60601-1:1990+A1+A2+A11+A12+A13PASS
EMC conformityEN 60601-1-2: 1993PASS

WOVEN COTTON SHEETING 2.

JIS L 1096 6.39.1.2 Method B 2, certified by SGS UK Ltd.

PERFORMANCE & CLINICAL TEST 3. AAMI / ANSI SP10

Amluck Enterprises Co. Ltd. believes this information and referred document to be sufficient for the FDA to find our proposed device substantially equivalent to the predicate product and other products currently in distribution.

Thomas Chen

Submitter, 01/11/2002 General Manager AMLUCK ENTERPRISES CO., LTD.

E

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, positioned to the right of a circular seal. The seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 1 4 2002

Amluck Enterprises Co., Ltd. c/o Dr. Yang, Tien-Hsing ROC Chinese-European Industrial Research Society No. 58, Fu-Chiun Street Hsin-Chu City Taiwan, ROC

Re: K012796

Trade Name: Amluck Automatic Digital Wrist Blood Pressure Monitor AK-3000/AK-4000 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: January 11, 2002 Received: January 15, 2002

Dear Dr. Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Dr. Yang, Tien-Hsing

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

N. Dala Miller

Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular And Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

AMLUCK Enterprises Co. Ltd. No. 41, Ching Hsiang St., Tu-Cheng, Taipei, Taiwan, ROC 236 Telephone: 886-2-226955555 Fax: 886-2-22695115 Email: amluck@ms7.hinet.net

Applicant:AMLUCK ENTERPRISES CO., LTD.
510(k) Number (if known):TBA K012796
Device Name:AMLUCK AUTOMATIC DIGITAL WRIST BLOOD PRESSUREMONITOR AK-3000 / AK-4000

. Indications for use:

The Amluck automatic digital wrist blood pressure monitor, Model AK-3000 / AK-4000, is a noninvasive blood pressure measurement system intended to measure the systolic and diastolic blood pressures and pulse rate of an adult individual, over age 18, at home by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to be 5.3" - 8.5".

( PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE

Concurrence of CDRH, office of Device Evaluation (ODE )

Division of Cardiovascular & Respiratory Devices
510(k) Number K012796

Prescription Use __ OR

( Per 21 CFR 801.109 )

Over - The - Counter - Use_ \

( Optional Format 1-2-96 )

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).