K Number
K012776

Validate with FDA (Live)

Date Cleared
2001-11-16

(88 days)

Product Code
Regulation Number
888.3560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Duracon® Constrained Posterior Stabilizer (PS) Tibial Inserts are intended to be used with legally marketed Duracon® tibial baseplates and wedges, Duracon® PS and TS femoral components, and patellar components as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be present. More specific indications/contraindications are listed below:

Indications

  • Noninflammatory degenerative joint disease including osteoarthritis, traumatic arthritis or avascular necrosis,
  • Rheumatoid arthritis,
  • Correction of functional deformity,
  • Revision procedures where other treatments or devices have failed,
  • Post-traumatic loss of joint anatomy, particularly when there is patello-femoral erosion, dysfunction or prior patellectomy; and,
  • Irreparable fracture of the knee
Device Description

The Duracon® Constrained PS tibial inserts share critical design features as the predicate Duracon® TS tibial inserts. Both designs incorporate an intercondylar post which fits into the intercondylar cavity of the Duracon® TS or PS femoral component. Both designs have a metallic support bracket imbedded into the tibial insert at the time of manufacture. This bracket has two metallic feet which engage the posterior tabs of the baseplate during assembly. The insert-baseplate locking mechanism is augmented by a hex head locking screw in both designs.

AI/ML Overview

The provided text describes a 510(k) summary for the Duracon® Constrained PS Tibial Inserts. However, it does not contain the specific information required to complete the table and answer all questions regarding acceptance criteria and a study demonstrating device performance. The document focuses on regulatory approval based on "substantial equivalence" to a predicate device, rather than detailed performance studies with acceptance criteria.

The "Performance Data" section merely states: "Mechanical testing has been performed to demonstrate the substantial equivalence of the subject inserts to the predicate inserts." It does not provide any specifics about the type of tests, the acceptance criteria, or the results.

Therefore, many of the requested details cannot be extracted from the provided text.

Here's a breakdown of what can and cannot be answered:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Not specified in the document.Mechanical testing was performed to demonstrate substantial equivalence to the predicate device. Specific results and parameters are not provided.

2. Sample size used for the test set and the data provenance

  • Sample size: Not specified.
  • Data provenance: Not specified (e.g., retrospective or prospective. It's mechanical testing, not a clinical study on patients, so concepts like country of origin for data aren't directly applicable in the same way).

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This is not applicable as the studies mentioned are mechanical testing, not a clinical assessment requiring expert ground truth.

4. Adjudication method for the test set

  • Not applicable for mechanical testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, an MRMC comparative effectiveness study was not done. This device is a knee prosthesis, not an AI diagnostic tool for human readers.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable as this is a physical medical device, not an algorithm.

7. The type of ground truth used

  • For mechanical testing, the "ground truth" would typically be established engineering standards or performance characteristics of the predicate device. However, the document doesn't specify these.

8. The sample size for the training set

  • Not applicable. This is not a machine learning model, so there's no "training set."

9. How the ground truth for the training set was established

  • Not applicable.

Summary of available information from the document:

  • Device Name: Duracon® Constrained PS Tibial Inserts
  • Predicate Device: Duracon® TS tibial inserts (K973164)
  • Performance Data: Mechanical testing confirmed "substantial equivalence" to the predicate device. No details of the tests, criteria, or results are provided in this summary.
  • Intended Use & Indications: Clearly defined, similar to the predicate device, for total knee arthroplasty in specific conditions where cruciate ligaments are inadequate or absent and stability is impaired.
  • Contraindications: Clearly defined absolute and relative contraindications.

{0}------------------------------------------------

KC012776 ા of 3

NOV 1 6 2001

510(k) Summary Duracon® Constrained PS Tibial Inserts

510(k) Summary Duracon® Constrained PS Tibial Insert

Submission Information

Name and Address of the Sponsor of the 510(k) Submission:Howmedica Osteonics Corp. 59 Route 17 Allendale, NJ 07401-1677
Contact Person:Margaret F. Crowe Regulatory Affairs Consultant (201) 934-4359
Date of Summary Preparation:August 10, 2001

Date of Summary Preparation

August 10, 2001

Device Identification

Proprietary Name:Duracon® Constrained PS Tibial Insert
Common Name:Knee Prosthesis
Classification Name and Reference:Knee Joint, Patellofemorotibial,Polymer/Metal/Polymer, Semi-Constrained, Cemented Prosthesis21 CFR '888.3560

Predicate Device Identification

The Duracon® Constrained PS Tibial Inserts are substantially equivalent to the Duracon® TS tibial inserts found substantially equivalent in premarket notification K973164.

Device Description

The Duracon® Constrained PS tibial inserts share critical design features as the predicate Duracon® TS tibial inserts. Both designs incorporate an intercondylar post

1a

{1}------------------------------------------------

KQ12776 2 of 3

510(k) Summary Duracon® Constrained PS Tibial Inserts

which fits into the intercondylar cavity of the Duracon® TS or PS femoral component. Both designs have a metallic support bracket imbedded into the tibial insert at the time of manufacture. This bracket has two metallic feet which engage the posterior tabs of the baseplate during assembly. The insert-baseplate locking mechanism is augmented by a hex head locking screw in both designs.

Intended Use

The intended use of the Duracon® Constrained PS tibial inserts is identical to that of the predicate Duracon® TS tibial bearing inserts. As with the predicate inserts, the new inserts are single use devices for use as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be present.

Indications

  • Noninflammatory degenerative joint disease including osteoarthritis, traumatic t arthritis or avascular necrosis,
  • . Rheumatoid arthritis,
  • Correction of functional deformity, 을
  • Revision procedures where other treatments or devices have failed, .
  • Post-traumatic loss of joint anatomy, particularly when there is patello-femoral t erosion, dysfunction or prior patellectomy; and,
  • . Irreparable fracture of the knee

{2}------------------------------------------------

KO12776 3073

510(k) Summary Duracon® Constrained PS Tibial Inserts

Contraindications

Absolute contraindications include:

  • 프 Overt infection,
  • Distant foci of infections (which may cause hematogenous spread to the implant . site).
  • Rapid disease progression as manifested by joint destruction or bone absorption apparent on roentgenogram,
  • . Skeletally immature patients
  • Cases where there is poor bone stock which would make the procedure . unjustifiable.

Conditions presenting an increased risk of failure include:

  • Uncooperative patient or patient with neurological disorders who is incapable of 트 following instructions,
  • . Osteoporosis,
  • 트 Metabolic disorders which may impair bone formation,
  • 트 Osteomalacia, and
  • 트 Previous arthrodesis

Performance Data

Mechanical testing has been performed to demonstrate the substantial equivalence of the subject inserts to the predicate inserts.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 6 2001

Ms. Margaret Crowe Regulatory Affairs Consultant Howmedica Osteonics Corporation 59 Route 17 Allendale, New Jersey 07401-1677

Re: K012776

Trade Name: Duracon® Constrained Posterior Stabilized (PS) Tibial Inserts Regulation Number: 888.3560 Regulation Name: Prothesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer Regulatory Class: II Product Code: JWH Dated: August 17, 2001 Received: August 20, 2001

Dear Ms.Crowe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Ms. Margaret Crowe

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

fo Mark n Millburn

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

1012776

510(k) Number (if known): K

NOV 1 6 2001

Duracon® Constrained PS Tibial Insert Device Name:

The Duracon® Constrained Posterior Stabilizer (PS) Tibial Inserts are intended to be used with legally marketed Duracon® tibial baseplates and wedges, Duracon® PS and TS femoral components, and patellar components as part of a total knee system in primary or revision cemented total knee arthroplasty. These inserts are intended to be used in cases where there is destruction of the joint surfaces with or without bone deformity, where the cruciate ligaments are inadequate, not present, or cannot be preserved during the operative procedure, especially when anterior-posterior stability is impaired due to absence of the patella. The collateral ligaments may or may not be present. More specific indications/contraindications are listed below:

Indications

  • Noninflammatory degenerative joint disease including osteoarthritis, traumatic 트 arthritis or avascular necrosis,
  • Rheumatoid arthritis, ■
  • Correction of functional deformity, .
  • Revision procedures where other treatments or devices have failed, 사
  • Post-traumatic loss of joint anatomy, particularly when there is patello-femoral 트 erosion, dysfunction or prior patellectomy; and,
  • в Irreparable fracture of the knee

to Mark N. Mulkerson

" ><<<<<<rive

K012776
510(k) Number

N/A