K Number
K012693
Date Cleared
2002-08-01

(352 days)

Product Code
Regulation Number
868.5895
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MISTRAL Universal Jet Ventilator is designed to use for short term application in bronchoscopy and laryngoscopy. Jet ventilation applied with the Mistral Universal Jet Ventilator is useful in airway surgery, as it is performed in thoracic surgery units and ENT surgery. Jet ventilation is the optimal ventilation technique during the application of LASER light, where the presence of an ETT bears the risk for ignition, airway fire and burn injuries. Jet ventilation is useful for the removal of foreign bodies from the airway (e.g. after accidental aspiration of foreign bodies) via rigid bronchoscopes, for supplemental oxygenation during lung surgery (when the operated lung must be recruited for additional oxygenation, but may not be ventilated conventionally for surgical reasons), for surgery of the lower trachea close to the carina (when during the reconstruction phase, no tight airway sealing can be achieved), for radiation therapy of lung metastasis (when the tidal movements of the chest during spontaneous respiration or conventional ventilation would preclude the focusing of the radiation beam onto the target.
The Mistral Universal Jet Ventilator is intended for use during bronchoscopy or laryngoscopy for applications up to 45 minutes duration on adults in a hospital or other clinical setting.

Device Description

The MISTRAL is a Universal Jet Ventilator. Key features include airway pressure measurement using a high pressure resistant sensor, a medium sized screen display for ventilation parameters and airway pressure graphics, a built-in air/oxygen blender allowing concentrations between 21 and 100% oxygen with an oxygen sensor, a bypass flow outlet for mask ventilation, a microprocessor (H8 532) and memory (27C512, DS1386), and various sensors (200 cmH2O pp-pressure sensor, 72.5 PSI driving-pressure sensor). It also includes fault detection for sensors and gas input, alarm functions for high PP pressure, system malfunctions, ventilation issues, FiO2 not adjustable, low O2 pressure, low air pressure, low battery, power failure, and maintenance required, and error functions for sensor malfunctions.

AI/ML Overview

The provided text is a 510(k) summary for the MISTRAL Universal Jet Ventilator. It focuses on describing the device, its functions, safety features, and how it compares to a predicate device (ACUTRONIC's VS-200s Universal Jet Ventilator) to establish substantial equivalence.

However, the summary does not contain information about specific acceptance criteria or an explicit study that proves the device meets such criteria. It details the device's design, features, and alarm functions, aiming to demonstrate its safety and effectiveness by comparison to a previously cleared device. There's no mention of performance metrics, sample sizes, expert ground truth, or adjudication methods for acceptance criteria for the MISTRAL device itself in the context of a dedicated performance study.

The document primarily acts as a detailed description for regulatory submission, highlighting technological characteristics and safety measures, rather than a report of a clinical or performance study with defined acceptance criteria.

Therefore, I cannot populate the requested table or answer the specific questions about acceptance criteria and a performance study based on the provided text. The document's purpose is to show technical equivalence, not to detail the results of a contemporary performance study against predefined criteria.

{0}------------------------------------------------

AUG01 2002
--------------

K012693

1 of 6

0074

510(k) Summary MISTRAL

ACUTRONIC
MEDICAL SYSTEMS

rev1

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Submitter's name, address, telephone number, contact person, and date summary 1. prepared:

  • ACUTRONIC Medical Systems AG a. Fabrik im Schiffli CH 8816 Hirzel Tel. 0041 1 729 70 80 Fax. 0041 1 729 70 81 e-mail: p.cavelti@acutronic-medical.ch
  • Contact Person: Pius Cavelti b. Director Sales / Marketing
  • Date Summary Prepared: 21.06.2002 C.

Name of device, including trade name and classification name: 2.

  • Trade/Proprietary Name: MISTRAL a.
  • Classification Name: Ventilator Continuous b.
  • ldentification of the predicate device or legally marketed device or devices to which 3. substantial equivalence is being claimed:
Company:ACUTRONIC Medical Systems AG
Device:VS-200s Universal Jet Ventilator
510(k):K900745C
Date Cleared:04/29/91
  • A description of the device that is the subject of the 510(k), including explanation of 4. how the device functions, basic scientific concepts, significant physical and performance characteristics (design, material, physical properties):

Airway pressure measurement:

One of the key features of the MISTRAL is the safety concept to avoid having excessive pressures occurring in the patient's airways. To reach this goal, a high pressure resistant sensor with excellent resolution in the working range of 1 to 200 mbar is used.

Display:

A medium sized screen displays the main parameters of ventilation and allows quick information of the anaesthetist about the status of the ventilation. A graphic indication of the airway pressure makes quantitative reading possible and allows quick access for parameter changes. Special functions are activated in menus and displayed on the screen.

File: 510(k) Summary MISTRAL rev1Date: 15.02.2001Index: 01
----------------------------------------------------------------

{1}------------------------------------------------

0075

Air / Oxygen blender:

The MISTRAL has a built in blender, which allows concentrations between 21 and 100 % oxygen. The access for changes is via a knob on the front panel. This parameter is accessible without going into menus, to allow a rapid change in case of emergency. The MISTRAL uses an oxygen sensor, which is calibrated each time the unit, is switched on. If the ventilator can't reach the selected FiO2, an alarm message appears on the screen, informing about a problem in the gas supply.

Bypass flow:

The MISTRAL has an additional outlet for gas, which is mechanically adjusted to 15 LPM and a concentration between 21 % and 100 %. This feature allows mask ventilation of the patient for induction in anaesthesia.

Microprocessor and Memory:

H8 532 (manufactured by Hitachi)

Memory devices:

27C512 (several types available) DS1386 (manufactured by Dallas) 64kB EPROM 32kB RAM timekeeper external watchdoq

Clock frequency 10MHz

Sensors:

200 cmH2O pp-pressure sensor (Honeywell) monitoring pressure condition in Jet tube sampling on jet frequency max 2.5 samples per second

72.5 PSI driving-pressure sensor (Honeywell) monitoring driving pressure sampling rate maximum 200 ms

Fault-Detection:

PP sensor is periodically checked for disconnection, shortcut and defective membrane (period: depending on frequency setting) failure results in error message

Driving pressure sensor is checked for disconnection & shortcut (period: depending on frequency setting) failure results in error message

Oxygen sensor is periodically checked for disconnection, shortcut and capacity (period: depending on actual requlated difference capacity only in auto calibration mode (startup routine)) failure results in error message

Full time gas input detection failure results in alarm message

fully regulated oxygen concentration with timeout function failure results in alarm message

Company of the consisted on the comments of the comments of the comments of the comments of
File: 510(k) Summary MISTRAL rev1Date: 15.02.2001Index: 01

{2}------------------------------------------------

Alarm functions

General information

In case of an alarm, the Mistral will display the alarm message on the screen and an acoustic alarm sounds. The acoustic alarm is ceased as soon as the alarm condition is cleared. However, the screen message remains until it is confirmed by pressing the alarm reset button. If you press alarm reset button while the alarm conditions is not cleared, there is an alarm mute for 1 minute activated, which depresses the acoustic alarm for this period. The screen inverses to make the user aware that the alarm mute is active.

The Mistral has an alarm concept to allow a safe use of the equipment for patient and operator. The following section shows the different alarm types and their sources:

PP too high:

This alarm is activated if the pressure in the Jet line remains above the set limit for the PP pressure. This alarm is a safety feature for the use of Jet Ventilation with single lumen catheters. It allows a security shut-off of the Jet Ventilator in case of an airway obstruction.

System alarms

General information

  • These alarms signal the user a state of improper function of the device. There is no immediate danger to the patient but the device should be replaced and serviced by authorised personal (exception: low level of pressure input).
    Specification and characteristics of the ventilation alarm

Application:

The ventilation device makes sure that max. oxygen level in the Jet ventilator is below a level of concern. The ventilation is periodically checked for ether disconnect (alarm level 1) or overload (alarm level 2)of the fan. This is an alarm which can not be disabled.

Specification and characteristics of the FIO2 not adjustable alarm

Application:

The Flo2 not adjustable alarm is evoked as soon as the O2 level can not be adjusted because of any other reason than missing ether O2 or airway pressure. There could be a mechanical problem or a system malfunction of the O2 requlation. If the O2 sensor could not be calibrated this alarm is also set active.

Specification and characteristics of the low O2 pressure alarm

Application:

The O2 pressure is continuously watched by the controller and goes on if there is not enough pressure on the O2 inlet. This is an alarm which can not be disabled. The alarm resets itself if pressure is sufficient.

File: 510(k) Summary MISTRAL rev1Date: 15.02.2001Index: 01
----------------------------------------------------------------

{3}------------------------------------------------

Specification and characteristics of the low air pressure alarm

Application:

The air pressure is continuously watched by the controller and goes on if there is not enough pressure on the air inlet. This is an alarm which can not be disabled. The alarm resets itself if pressure is sufficient.

Specification and characteristics of low bat alarm

Application:

Battery level is measured and in case of a worn out battery the message "low bat" is displayed instead of the time and date. There is no alarm going off because the battery is not used unless the power failure fault. This battery has not a data backup function. In case of "low bat" a replacement of the battery is sufficient.

Specification and characteristics of Power failure alarm

Application:

Main line is less than 95 VAC the beeper on the back panel sounds with interval supplied by battery power. This is an alarm, which can not be disabled.

Specification and characteristics of MAINTENANCE REQUIRED alarm

Application:

Mistral has an integrated timer function for servicing intervals. After one year, on the display the message "MAINTENANCE REQUIRED" appears. The unit may still be used, however, a service engineer must be called for servicing of the equipment to assure a proper function.

Error functions

Overview

The error message signals a system malfunction. The error messages can be reset by the alarm button but if the malfunction is still detected the message immediately returns to the screen. In any of these cases the Monsoon needs a service.

Type of errorcharacteristic
PP sensorthe PP sensor is out of range (probablydisconnected or defect)
DP sensorthe DP sensor is out of range (probablydisconnected or defect)
O2 sensor isweakO2 signal is getting to weak replace thesensor
File: 510(k) Summary MISTRAL rev1Date: 15.02.2001Index: 01
----------------------------------------------------------------

0077

{4}------------------------------------------------

న. Statement of intended use:

The MISTRAL Universal Jet Ventilator is designed to use for short term application in bronchoscopy and laryngoscopy.

Indication for Use

Jet ventilation applied with the Mistral Universal Jet Ventilator is useful in airway surgery, as it is performed in thoracic surgery units and ENT surgery. Jet ventilation is the optimal ventilation technique during the application of LASER light, where the presence of an ETT bears the risk for ignition, airway fire and burn injuries. Jet ventilation is useful for the removal of foreign bodies from the airway (e.g. after accidental aspiration of foreign bodies) via rigid bronchoscopes, for supplemental oxygenation during lung surgery (when the operated lung must be recruited for additional oxygenation, but may not be ventilated conventionally for surgical reasons), for surgery of the lower trachea close to the carina (when during the reconstruction phase, no tight airway sealing can be achieved), for radiation therapy of lung metastasis (when the tidal movements of the chest during spontaneous respiration or conventional ventilation would preclude the focusing of the radiation beam onto the target.

0078

File: 510(k) Summary MISTRAL rev 1Date: 15.02.2001Index: 01
-----------------------------------------------------------------

{5}------------------------------------------------

510(k) Summary MISTRAL

6 of 6

Statement of how the technological characteristics of the device compare to those 6. of the predicate or legally marketed device.

Technical featureVS-200SMISTRALComment
EEP LimitEnd ExpiratoryPressure LimitYes, measured in betweenthe pulses of Jetinsufflations. Valve opensduring pause to directpressure on pressure sensorYes, measured with a highpressure resistantpressure sensor.Resolution 1 mbarDue toavailability ofhigh pressureresistant sensor,quicker responseof the pressurelimit = increasedpatient safety
Display of EEPpressure levelNoYes, displayed after eachjet insufflationAllows trendinformation onEEP pressure.
Bypass flowNoYes, adjusted at factoryupon customer request.Allows driving of nebulizeror for continues flow foranaesthesia inductionNo need forconventionalanaesthesiamachine forinduction
Built-in air-oxygenblenderNo. Use of Bird blenderYes, electronic blenderCompact sizewithout need forexternal tubings
Oxygen measurementNo. Need of separateoxygen monitorYes, with automaticcalibration of sensorduring start-up procedureEase of use.
HumidificationYes, injection of distilledwater into jet nozzleNoExperiences inthe past haveshown, thatinjection of watermay create somelesions in thetrachea if dropsize is too big.
In-service informationNoYes, message appears ondisplay when maintenanceis neededFacilitates the inhospital controlof equipmentservicingconditions
Flexible software forfuture improvementsNo, digital electronic, nomicroprocessorYes, allows use of differentlanguage for operatorwithout change of frontlabel.Menu drivenoperatorsinterface
Inlet pressurecontrolled by servovalveNoYes, if pressure limit isexceeded for more then 3s, the inlet pressure valvesclose for patient safetyIn case of a jetvalve failure, theinlet pressure isshut offautomatically toavoidoverpressure inthe patients lung
File: 510(k) Summary MISTRAL rev1Date: 15.02.2001Index: 01007
---------------------------------------------------------------------

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with the profiles overlapping each other. The logo is black and white.

Public Health Service

AUG 0 1 2002

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Acutronic Medical Systems AG C/O Terry Torzala, R.A.C. Official Correspondent Medical Equipment Development, Incorporated P.O. Box 85820 Tucson, Arizona 85754-5820

  • K012693 Re:
    Trade/Device Name: Mistral Critical Care Jet Ventilator and Accessories Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: June 26, 2002 Received: June 26, 2002

Dear Mr. Torzala:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2 - Mr. Torzala

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely,

Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K012693

Device Name: Mistral Universal Jet Ventilator

. •

.

Indications For Use:

  1. 1

The Mistral Universal Jet Ventilator is intended for use during bronchoscopy or laryngoscopy for applications up to 45 minutes duration on adults in a hospital or other clinical setting.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

CAuha

Dental, Infection Control. neral Hospital Devi

510(k) Number K0612693

Prescription Use
(Per 21 CFR 801.109) √

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

A - 2

OR

§ 868.5895 Continuous ventilator.

(a)
Identification. A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.(b)
Classification. Class II (performance standards).