K Number
K012533

Validate with FDA (Live)

Manufacturer
Date Cleared
2001-08-22

(16 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The OSI Medical Dolphin Stand-Alone Pulse Oximeter and Accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The OSI Medical Dolphin Pulse Oximeter and Accessories are intended for use with additional monitoring equipment in clinical facilities and home environments.

The OSI Medical's sensor(s) are intended/indicated for use only with the OSI Medical Dolphin Stand-Alone Pulse Oximeter, and consist of the following:

Digital Dolphin Disposable Sensors intended for adults / pediatrics greater than 66.14 Ibs. (30kg).
Digital Dolphin Reusable Sensors Intended for adults / pediatrics greater than 66.14 Ibs. (30kg).
Digital Dolphin 110 Extension Cable

The OSI Medical extension cable is intended/indicated for use only with the OSI Medical sensor(s) and the OS/ Medical Dolphin Stand-Alone Pulse Oximeter.

Device Description

The OSI Medical Dolphin Pulse Oximeter and Accessories is a portable stand-alone device, connecting cable, and oximetry sensor(s) to noninvasively calculate the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate.

The monitor consists of a screen that displays the pulse plethysmographic waveform, the pulse rate, SpO2 value, the high and low SpO2 and pulse rate alarm limits, alarms, trends and status messages. It contains the electronic hardware and software that receives and calculates the signals from the LEDs within the sensor to determine the functional oxygen saturation of arterial hemoglobin (SpQz) and pulse rate, and provide for the connection to the connecting cable.

The OSI Medical Dolphin Pulse Oximeter is available in one configuration as a portable stand-alone oximeter that is 10 cm / 3.94 inches high, 27.5 cm / 10.83 inches wide, 25 cm / 9.84 inches deep and weighs about 4 kg / 8.8 lbs. The unit is powered either with a voltage input of 100-240 Vac, 50-60 Hz or with a sealed lead-acid battery with an operating time of approximately 4 hours based upon 2 Ampere hour battery (200mA OEM Module, 300mA System Module) and a charge time of about 4.5 hours to 80% capacity.

The extension cable connects between the monitor and oximetry sensor(s) and transfers LED drive power and the calibration drive to the oximetry sensor from the monitor receives the detector signal from the oximetry sensor.

The extension cable is available in one configuration and is approximately 8 feet / 2.44m in length, and the sensor(s) are approximately 18 inches / 45.72 cm in length.

The sensor(s) measure light absorption of blood from two light emitting diodes (LED's). Oxygen saturated blood absorbs light differently as compared to unsaturated blood. The amount of light absorbed by the blood is used to calculate the ratio of oxygenated hemoglobin to total hemoglobin in arterial blood.

The oximetry sensor is available in a disposable or reusable configuration, and with one configuration for the extension cable (8 feet).

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and the study details for the OSI Medical Dolphin Stand-Alone Pulse Oximeter:

This 510(k) submission primarily details the substantial equivalence of the OSI Medical Dolphin Disposable Optical Sensor, Model 510 by referring to an earlier 510(k) for the OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100 (K002036). The focus seems to be on demonstrating that the new disposable sensor works equivalently with the previously cleared oximeter. Therefore, the information regarding acceptance criteria and performance is largely indirect and based on the equivalence to the predicate device.

1. Table of Acceptance Criteria and Reported Device Performance

The main acceptance criteria for pulse oximeters are typically accuracy (SpO2 and pulse rate) and functionality, often assessed against a reference standard. Since this document is a 510(k) summary for a disposable sensor that relies on a previously cleared oximeter, explicit performance data for the entire system (oximeter + disposable sensor) is not detailed with specific accuracy numbers in the provided text. Instead, the document states that "Additional testing has been performed for the disposable sensors including biocompatibility and performance validation which has been included in this submission."

Given the information provided, we can infer the acceptance criteria and performance based on common pulse oximeter standards and the implicit claim of equivalence.

Acceptance Criteria CategorySpecific Criteria (Inferred from standards and product type)Reported Device Performance (Inferred/Stated)
Accuracy (SpO2)SpO2 Accuracy (Arms) within generally accepted limits (e.g., +/- 2-3% in range 70-100% SpO2) per ISO 9919 / EN 865 and FDA guidance."Performance validation" has been performed, implying it meets the accuracy requirements of the predicate device and relevant standards. No specific SpO2 accuracy (e.g., Arms) is provided in this summary.
Accuracy (Pulse Rate)Pulse Rate Accuracy (e.g., +/- 1-2 bpm)"Performance validation" has been performed. No specific pulse rate accuracy is provided in this summary.
BiocompatibilityCompliance with ISO 10993 standards for patient contact.Biocompatibility testing for the disposable sensors has been performed and included in the submission.
Electrical SafetyCompliance with IEC 60601-1 and CSA C22.2 No. 601.1The device complies with IEC 601-1, CSA C22.2 No. 601.1, and UL2601-1.
Electromagnetic Compatibility (EMC)Compliance with IEC 601-1-2The device complies with IEC 601-1-2.
Functionality/DesignMeets design and operational specifications for a pulse oximeter."Substantially equivalent in design concepts, technologies and materials" to the predicate device.
Environmental/PackagingCompliance with relevant environmental testing (e.g., IEC 6068-2 series, ISTA Procedure 2A).Complies with IEC 6068-2-6, IEC 6068-2-27, IEC 6068-2-64, and ISTA Procedure 2A.

2. Sample Size Used for the Test Set and Data Provenance

The summary does not explicitly state the sample size for the test set used for performance validation of the disposable sensors.

  • Data Provenance: The document does not specify the country of origin for the data. The study type (retrospective or prospective) is also not mentioned. However, "performance validation" for a new sensor usually implies a prospective study involving human subjects or controlled laboratory testing.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not provide information on the number of experts or their qualifications used to establish ground truth. For pulse oximeter performance validation, ground truth (arterial oxygen saturation) is typically established through a co-oximeter analyzing arterial blood samples, not directly by expert consensus on visual assessment.

4. Adjudication Method for the Test Set

The document does not describe any adjudication method. This is generally not applicable in the context of pulse oximeter accuracy studies, where comparison is made against a physiological reference standard rather than subjective expert interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No, an MRMC comparative effectiveness study was not done or described in this summary. This type of study is more relevant for diagnostic imaging devices where human interpretation is a key component. Pulse oximeters are typically evaluated for standalone accuracy.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, implicitly. The "performance validation" for the disposable sensors, in conjunction with the oximeter, would be a standalone performance assessment. The device calculates SpO2 and pulse rate automatically, so the relevant performance is its accuracy in comparison to a reference standard, independent of human interpretation of the values.

7. The Type of Ground Truth Used

The ground truth for pulse oximeter accuracy studies is typically established by:

  • Co-oximetry of arterial blood samples: This is the gold standard for directly measuring functional arterial oxygen saturation (%HbO2) in blood.
  • Controlled hypoxia studies: For SpO2 accuracy, subjects are typically desaturated under controlled conditions, and simultaneous arterial blood samples are taken and analyzed by a co-oximeter.

While not explicitly stated in this summary, "performance validation" for a pulse oximeter sensor would undoubtedly rely on one of these methods, most likely co-oximetry.

8. The Sample Size for the Training Set

The document does not specify a training set size. Pulse oximeters, especially those from 2001, typically do not employ machine learning models in the same way modern AI-powered diagnostic devices do. Their algorithms are usually based on spectrophotometry and signal processing principles, which are designed rather than "trained" on large datasets. If any form of calibration or algorithm optimization was done using data, it's not described as a "training set" in the modern AI sense.

9. How the Ground Truth for the Training Set Was Established

As explained in point 8, the concept of a "training set" and associated ground truth establishment for a pulse oximeter, as understood in contemporary AI/ML contexts, is not applicable or described in this 2001 510(k) summary. The device's underlying principles rely on known physiological properties of light absorption by blood, rather than statistical learning from a diverse training dataset.

{0}------------------------------------------------

K012533

AUG 2 2 2001

August 1, 2001

Summary of Safety and Effectiveness Information for the OSI Medical Dolphin Stand-Subject: Alone Pulse Oximeter and Accessories. OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100 and Accessories. Proprietary: Common: Oximeter. Oximeter Class II - 21 CFR 870.2700 - 74 DQA Classification:

The 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the SMDA 1990 and 1992.

The OSI Medical Dolphin Pulse Oximeter and Accessories is substantially equivalent to the following currently marketed device(s):

  • OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100 and Accessories #K002036. .
    The OSI Medical Dolphin Pulse Oximeter and Accessories is a portable stand-alone device, connecting cable, and oximetry sensor(s) to noninvasively calculate the functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate.

The monitor consists of a screen that displays the pulse plethysmographic waveform, the pulse rate, SpO2 value, the high and low SpO2 and pulse rate alarm limits, alarms, trends and status messages. It contains the electronic hardware and software that receives and calculates the signals from the LEDs within the sensor to determine the functional oxygen saturation of arterial hemoglobin (SpQz) and pulse rate, and provide for the connection to the connecting cable.

The OSI Medical Dolphin Pulse Oximeter is available in one configuration as a portable stand-alone oximeter that is 10 cm / 3.94 inches high, 27.5 cm / 10.83 inches wide, 25 cm / 9.84 inches deep and weighs about 4 kg / 8.8 lbs. The unit is powered either with a voltage input of 100-240 Vac, 50-60 Hz or with a sealed lead-acid battery with an operating time of approximately 4 hours based upon 2 Ampere hour battery (200mA OEM Module, 300mA System Module) and a charge time of about 4.5 hours to 80% capacity.

The extension cable connects between the monitor and oximetry sensor(s) and transfers LED drive power and the calibration drive to the oximetry sensor from the monitor receives the detector signal from the oximetry sensor.

The extension cable is available in one configuration and is approximately 8 feet / 2.44m in length, and the sensor(s) are approximately 18 inches / 45.72 cm in length.

The sensor(s) measure light absorption of blood from two light emitting diodes (LED's). Oxygen saturated blood absorbs light differently as compared to unsaturated blood. The amount of light absorbed by the blood is used to calculate the ratio of oxygenated hemoglobin to total hemoglobin in arterial blood.

{1}------------------------------------------------

The oximetry sensor is available in a disposable or reusable configuration, and with one configuration for the extension cable (8 feet).

The OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100 and Accessories have been designed to comply with the following standards:

    1. CSA C22.2 No. 601.1
  • IEC 601-1. Part 1 and Amendments 1 and 2 2.
    1. IEC 601-1-1, Part 1
    1. IEC 601-1-2, Part 1
    1. ISO 9919: 1992
    1. EN 865: 1997
    1. FDA Guidance Document for Pulse Oximeters: 9/7/1992
    1. ASTM 1415:1992, and Draft 10.1
    1. UL2601-1: Second Edition, 1997
    1. IEC 6068-2-6
    1. IEC 6068-2-27
    1. IEC 6068-2-64
    1. ISTA Procedure 2A

The OSI Medical Dolphin Stand-Alone Pulse Oximeter with disposable sensors is substantially equivalent in design concepts, technologies and materials to the OSI Medical Dolphin Stand-Alone Pulse Oximeter, with reusable sensors. Additional testing has been performed for the disposable sensors including biocompatibility and performance validation which has been included in this submission.

000012

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three heads, each head facing a different direction.

AUG 2 2 2001

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Jon Werner OA Manager OSI Medical, Inc. 13801 McCormick Drive Tampa, FL 33626

Re: K012533

OSI Medical Digital Dolphin Disposable Optical Sensor, Model 510 Regulation Number: 870.2700 Regulatory Class: II (two) Product Code: 74 DQA Dated: August 3, 2001 Received: August 6, 2001

Dear Mr. Werner:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce stired in May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Sourisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this

{3}------------------------------------------------

Page 2 - Mr. Jon Werner

response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance (DSMICA) at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Deck Teh

. Dillard III ames Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

INDICATIONS FOR INTENDED USE FOR 2100 PULSE OXIMETER

K012533 510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name: OSI Medical Dolphin Stand-Alone Pulse Oximeter, Model No. 2100

August 1, 2001

Indications for Use:

Indications for Use:
The OSI Medical Dolphin Stand-Alone Pulse Oximeter and Accessories are indicated for the continuous The OSI Medical Dolphin Stand-Alone Califical Che Aberoglobin (SpO-) and pulse rate (measured noninvasive monitoring of lunctional oxygen saturalion of united for use with addition ase with addition and home by an SpO2 sensor). The OST Medical Dolphill Pulse Unition and Noscother Pacilities and home environments.

The OSI Medical's sensor(s) are intended/indicated for use only with the OSI Medical Dolphin Stand-Alone Pulse Oximeter, and consist of the following:

Digital Dolphin Disposable Sensors intended for adults / pediatrics greater than 66.14 Ibs. (30kg). Digital Dolphin Reusable Sensors Intended for adults / pediatrics greater than 66.14 Ibs. (30kg). Digital Dolphin 110 Extension Cable

The OSI Medical extension cable is intended/indicated for use only with the OSI Medical sensor(s) and the OS/ Medical Dolphin Stand-Alone Pulse Oximeter.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number: K012533

Prescription Use
(Per 21 CFR 801.109)
OR
over-the-counter Use__

000009

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).