K Number
K012327

Validate with FDA (Live)

Manufacturer
Date Cleared
2001-12-20

(150 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CSIST DICOM Gateway and Image Manager™ is a device that captures 2 dimensional images and data or receives images and data from various medical imaging sources (i.e. ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data (2 dimensional or 3 dimensional) can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.

Device Description

CSIST DICOM Gateway and Image Manager™ consists of three modules: A DICOM Gateway, Image Manager Server, and Image Manager Client.

  • The DICOM Gateway captures (2D) images from a non-DICOM imaging modality, generates a DICOM medical image, and stores it in the CSIST Image Manager Server.
  • The CSIST Image Manager Server accepts DICOM medical images (2D and 3D) from different acquisition modalities, stores these in the Image Archive and then transfers the generated images to the requested Image Manager Client.
  • The Image Manager Server also has a "Performed Procedure Step Manager", which can issue a message to department system scheduler/order filler and the Image Manager Server, when the acquisition modality informs the Performed Procedure Step Manager that a specific procedure step has been started or completed.
AI/ML Overview

The provided document is a 510(k) summary for the CSIST DICOM Gateway and Image Manager™ System. It focuses on demonstrating substantial equivalence to predicate devices and describes the device's function and intended use. However, it does not contain information regarding specific acceptance criteria, a detailed study proving performance against those criteria, or the methodology for establishing ground truth for a test set.

Therefore, the requested information cannot be fully extracted from the provided text.

Here's a breakdown of what can and cannot be answered based on the document:

1. A table of acceptance criteria and the reported device performance

  • Cannot be provided. The document does not specify any quantitative acceptance criteria (e.g., accuracy, sensitivity, specificity, image quality metrics) or report performance metrics against such criteria. It states that "Any differences between the two devices will not affect safety and effectiveness" based on a finding of "substantial equivalence" to predicate devices.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Cannot be provided. The document does not describe any specific test set used to evaluate the device's performance. The submission relies on demonstrating substantial equivalence, not a standalone performance study with a test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Cannot be provided. Since no specific test set and ground truth establishment process is described, this information is absent.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Cannot be provided. No adjudication method is mentioned as there's no described test set.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Cannot be provided. The device is a Picture Archiving and Communications system (PACS) component, not an AI-assisted diagnostic tool. No MRMC study is mentioned, nor would it be typical for this type of device to evaluate AI reader improvement.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Cannot be provided. The document describes the device as a system for capturing, storing, communicating, processing, and displaying medical images. It's a foundational PACS component, not an algorithm with standalone diagnostic performance. The document explicitly states: "A physician, providing ample opportunity for competent human intervention interprets images and information being printed."

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

  • Cannot be provided. No ground truth definition is mentioned as no performance study is detailed.

8. The sample size for the training set

  • Cannot be provided. The device is not described as an AI/ML device that requires a training set. It's a data management and display system.

9. How the ground truth for the training set was established

  • Cannot be provided. No training set or ground truth establishment for a training set is mentioned.

Summary based on the document:

This 510(k) emphasizes "substantial equivalence" to existing predicate devices (OLICON 02 Workstation & PACS View Software K973959 and ALI 3D Tool Module for Medical Images K003762). The "proof" is the comparison against these already cleared devices, arguing that the CSIST system's design, general function, intended use, safety, and efficacy are similar, and any differences "will not affect safety and effectiveness."

The device's role is described as:

  • Capturing 2D images from non-DICOM modalities, generating DICOM images.
  • Accepting, storing, and transferring 2D and 3D DICOM images from various modalities.
  • Providing a "Performed Procedure Step Manager."

It is explicitly stated that a "physician, providing ample opportunity for competent human intervention interprets images and information being printed." This reinforces that the device is a tool for managing medical images, not an automated diagnostic system that requires performance metrics like sensitivity or specificity against a ground truth.

The document mentions "hazard analysis" with all potential hazards classified as "Minor," and adherence to "voluntary standards," which are common safety and effectiveness considerations for medical devices, but not specific performance metrics in the way requested for an AI/ML diagnostic algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a logo with a trident symbol inside a black circle. To the right of the circle, the text "CSIST" is written in a simple, sans-serif font. The trident is stylized with three points at the top and a central vertical line intersecting a horizontal bar. The overall design is simple and monochromatic.

510(k) Summary of Safety and Effectiveness

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

Date Prepared:

October 12, 2001

Submitter's Information: 21 CFR 807.92(a)(1)

CSIST (Chung-Shan Institute of Science and Technology) 481 Chia-An Sec. Chung-Cheng Road Chia-An Village, Lung Tan 32500; Taoyuan Taiwan, R.O.C.

Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)

Trade Name:CSIST DICOM Gateway and Image Manager™ System
Common Name:Frame Grabber and Viewing Workstation
Classification Name:Picture Archiving and Communications System
Device Classification:21 CFR 892.2050

Predicate Device: 21 CFR 807. 92(a)(3)

ManufacturerOLICON Imaging Systems, Inc.A.L.I. Technologies, Inc
Device NameOLICON 02 Workstation & PACSView SoftwareALI 3D Tool Module forMedical Images
K NumberK973959K003762
Decision Date12/22/972/28/01
Product Code90 LMD90 LLZ
ClassificationClass IIClass II

Device Description: 21 CFR 807 92(a)(4)

CSIST DICOM Gateway and Image Manager™ consists of three modules: A DICOM Gateway, Image Manager Server, and Image Manager Client.

  • 내 The DICOM Gateway captures (2D) images from a non-DICOM imaging modality, generates a DICOM medical image, and stores it in the CSIST Image Manager Server.
  • ( The CSIST Image Manager Server accepts DICOM medical images (2D and 3D) from different acquisition modalities, stores these in the Image Archive and then transfers the generated images to the requested Image Manager Client.

510(k) submission for CSIST DICOM Gateway and Image Management™

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a black and white logo. The logo consists of a circle with a trident-like symbol inside. The trident has three prongs pointing upwards. To the right of the circle, the text "CSIST" is written in a simple font, with a horizontal line underneath it.

  • The Image Manager Server also has a "Performed Procedure Step Manager", 해 which can issue a message to department system scheduler/order filler and the Image Manager Server, when the acquisition modality informs the Performed Procedure Step Manager that a specific procedure step has been started or completed.

Indications for Use: 21 CFR 807 92(a)(5)

The CSIST DICOM Gateway and Image Manager™ is a device that captures 2 dimensional images and data or receives images and data from various medical imaging sources (i.e. ultrasound systems, R/F Units, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways or other imaging sources). Images and data (2 dimensional or 3 dimensional) can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.

Typical users of this system are trained professionals, including but not limited to physicians, nurses, and technicians.

Technological Characteristics: 21 CFR 807 92(a)(6)

The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being printed.

Conclusion: 21 CFR 807 92(b)(1)

The two predicate devices (K003762 and K973959) and the new CSIST Gateway & Image Manager device are medical image PACS devices that are substantially equivalent in areas of design, general function, intended use, safety, and efficacy. Any differences between the two devices will not affect safety and effectiveness.

The CSIST DICOM Gateway & Image Manager is substantially equivalent to the OLICON 02 Workstation & PACS View Software (K973959) for image acquisition and image management for non-3D images and substantially equivalent to ALI 3D Tool Module (K003762) used with the ALI Ultra PACS for viewing and image management of 3D images.

The CSIST DICOM Gateway and Image Manager™ system has been and will be designed and manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey.

The submission contains the results of a hazard analysis and all potential hazards have been classified as Minor.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 2 0 2001

CSIST % Mr. Carl Alletto Otech, Inc. 1100 Lakeview Blvd. DENTON TX 76208 Re: K012327

Trade/Device Name: CSIST DICOM Gateway and Image Manager Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving and Communications system Regulatory Class: II Product Code: 90 LLZ Dated: October 14, 2001 Received: October 18, 2001

Dear Mr. Alletto:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enorosale) to regally atment date of the Medical Device Amendments, or to conniered pror to May 20, 1978, and the provisions of the Federal Food, Drug, de necs that have boon require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, dicrororo, market of the Act include requirements for annual registration, listing of general controls provisions of taxtice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your device is thankled (600 a00 10) . Existing major regulations affecting your device can be it may of Subjoct to additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act that I Dri has made a associations administered by other Federal agencies. You must or any I oderal battle Act's requirements, including, but not limited to: registration and listing compry with and the rece required to the Part 801); good manufacturing practice requirements as set (21 OF IC Fat 801), lacemis (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality by stems (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours.

Nancy C. Snogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows a logo with a trident symbol inside a black circle. To the right of the logo, the letters "CSIS" are printed in a simple font. A horizontal line is drawn beneath the logo and the letters.

(Indications for Use Form)

510(k) Number: Kol 2327

Device Name:

The CSIST DICOM Gateway/Image Manager™

Indications for Use:

The CSIST DICOM Gateway and Image Manager™ is a device that captures 2 dimensional images and data or receives images and data from various medical imaging sources (i.e. ultrasound systems, R/F Units, computed & direct radiographic imaging boardod (from anture devices, scanners, imaging gateways or other imaging sources). Images and data (2 dimensional or 3 dimensional) can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations.

Typical users of this system are trained professionals, including but not limited to physicians, nurses, and technicians.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

. Y Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

David A. Levenson

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Device
510(k) Number K012327

Page 1 of -1

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).