K Number
K012197

Validate with FDA (Live)

Date Cleared
2001-11-16

(126 days)

Product Code
Regulation Number
870.5310
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Powerheart Cardiac Rhythm Module (CRM) is intended to acquire the ECG rhythm for the detection of, and to provide treatment for, ventricular tachyarrhythmias of patients who are at risk of sudden cardiac arrest. The device is intended to be used in medically supervised environments by trained personnel, in which patients are under the direct care of physicians and/or medical persons authorized by the state, province or country regulations in which they practice. The Powerheart CRM can be used in three modes: fully automatic mode, advisory mode or manual mode and includes a pacer which allows an operator to deliver external stimuli for temporary transcutaneous cardiac pacing. The Powerheart has previously undergone clinical testing on adults which support the safety and effectiveness of using the device on patients at risk of sudden cardiac arrest who are at least 8 years of age. The device is not classified as a wearable external defibrillator and is not intended for use in the home environment.

Device Description

The Powerheart Cardiac Rhythm Module (CRM) is a compact, lightweight, automatic external cardioverter defibrillator. It incorporates essential circuitry and software to provide a detection/analysis/defibrillation/pacing system, in a single device, for monitoring and providing therapy to patients at risk for Sudden Cardiac Arrest.

AI/ML Overview

The provided text does not contain the detailed study information required to fill out all the requested fields for acceptance criteria and the study proving the device meets those criteria.

The submission is a 510(k) summary for the Powerheart® Cardiac Rhythm Module™ (CRM™), an Automatic External Cardioverter Defibrillator. While it describes the device's intended use and technological characteristics, it largely references compliance with standards and past clinical testing rather than presenting new study data with acceptance criteria and results.

Here's what can be extracted and what information is missing:

1. Table of acceptance criteria and the reported device performance:

Acceptance CriteriaReported Device Performance
Not explicit in the provided text. The text states: "The performance tests conducted on the Powerheart Cardiac Rhythm Module will be subjected to pre-determined pass criteria that will support claims of substantial equivalence." This indicates that such criteria exist, but they are not detailed here.Not explicit in the provided text. The text states: "The performance tests conducted on the Powerheart Cardiac Rhythm Module will follow the applicable portions of the ANSIAAMI DF2 and DF39 requirements for cardiac defibrillator devices and automatic external defibrillators to demonstrate the safety and efficacy of the CRM device." This implies the device performed adequately against these standards, but specific performance metrics are not given.

Missing Information:

  • Specific numerical or statistical thresholds for performance (e.g., sensitivity, specificity, accuracy for arrhythmia detection; energy delivery capabilities, pacing parameters).
  • Actual results of the performance tests against these criteria.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

Missing Information. The document mentions: "The Powerheart has previously undergone clinical testing on adults which support the safety and effectiveness of using the device on patients at risk of sudden cardiac arrest who are at least 8 years of age." However, it does not provide details about:

  • The sample size of this previous clinical testing.
  • Whether the data was prospective or retrospective.
  • The country of origin for the data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience):

Missing Information. This level of detail is not mentioned in the provided text, as the document primarily refers to compliance with performance standards (ANSIAAMI DF2 and DF39) rather than a specific clinical validation study with expert-adjudicated ground truth.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

Missing Information. Similar to point 3, this detail is not provided.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable/Missing Information. The device described is an Automatic External Cardioverter Defibrillator (AED) and a pacing device. While it uses "FDA-cleared and patented arrhythmia detection software," it's not a diagnostic aid where human readers or their improvement with AI assistance would typically be measured. The device automatically detects and delivers therapy or advises the operator to shock, or is used in manual mode. Therefore, an MRMC study related to human reader improvement with AI assistance is not described or relevant for this type of device based on the provided text.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Partially Addressed / Implied.

  • The "Fully Automatic Mode" implies standalone algorithm performance as it "automatically provides defibrillation therapy."
  • The document states: "The Powerheart CRM utilizes an FDA-cleared and patented arrhythmia detection software that continuously monitors, detects and delivers therapy..."
  • The performance tests follow "applicable portions of the ANSIAAMI DF2 and DF39 requirements," which would include testing the device's automated detection and therapy delivery algorithms. However, a specific "standalone study" with detailed results (like sensitivity/specificity for arrhythmia detection) is not presented.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

Missing Information. For the "previously undergone clinical testing," the method for establishing ground truth is not specified. For the performance tests against ANSIAAMI standards, these standards typically involve specific test signals and simulated conditions, which serve as the "ground truth" for evaluating the device's response.

8. The sample size for the training set:

Missing Information. The document mentions "FDA-cleared and patented arrhythmia detection software," which indicates development and likely a training phase. However, no details about the training set size are provided.

9. How the ground truth for the training set was established:

Missing Information. No information is given regarding the training set or its ground truth establishment.

In summary: The provided text is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device and compliance with relevant standards (ANSIAAMI DF2 and DF39). It explicitly states that performance tests were conducted and will be subjected to pre-determined pass criteria, but it does not present the detailed results, acceptance criteria, or study methodologies (like sample sizes, ground truth establishment, or expert involvement) that would be found in a comprehensive clinical or performance study report.

{0}------------------------------------------------

510(k) SUMMARY

Submitter's Name and Address 10.1

NOV 1 6 2001

CARDIAC SCIENCE, INC. 16931 Millikan Avenue Irvine, CA. 92606

Contact person:

Sew-Wah Tay, PhD Vice President Cardiac Science, Inc. 5420 Feltl Road Minneapolis, MN 55343 Tel: 952-939-2942 Fax: 952-939-4191 Email: stay@cardiacscience.com

10.2 Date Summary was Prepared

October 21, 2001

10.3 Device Name

Proprietary: Powerheart® Cardiac Rhythm Module™ (CRMTM)

Common: Automatic External Cardioverter Defibrillator

10.4 Classification

    1. 870.5300 DC-defibrillator
    1. 870.5550 External Transcutaneous Cardiac Pacemaker (Noninvasive)
    1. Class III 74 MJK

10.5 Predicate Device

Powerheart® AECD®, K993533

10.6 Description of Device

The Powerheart Cardiac Rhythm Module (CRM) is a compact, lightweight, automatic external cardioverter defibrillator. It incorporates essential circuitry and software to provide a detection/analysis/defibrillation/pacing system, in a single device, for monitoring and providing therapy to patients at risk for Sudden Cardiac Arrest.

{1}------------------------------------------------

10.7 Intended Use of Device

Indication for Use:

The Powerheart Cardiac Rhythm Module (CRM) is intended to acquire the ECG rhythm for the detection of, and to provide treatment for, ventricular tachyarrhythmias of patients who are at risk of sudden cardiac arrest. The device is intended to be used in medically supervised environments by trained personnel, in which patients are under the direct care of physicians and/or medical persons authorized by the state, province or country regulations in which they practice. The Powerheart CRM can be used in three modes: fully automatic mode, advisory mode or manual mode and includes a pacer which allows an operator to deliver external stimuli for temporary transcutaneous cardiac pacing. The Powerheart has previously undergone clinical testing on adults which support the safety and effectiveness of using the device on patients at risk of sudden cardiac arrest who are at least 8 years of age. The device is not classified as a wearable external defibrillator and is not in the home environment.

Fully Automatic Mode

In this mode, the device continuously monitors the patient's ECG rhythm, detects shockable ventricular tachyarrhythmias and automatically provides defibrillation therapy for the treatment of sudden cardiac arrest.

Advisory (AED) Mode

In this mode, the device continuously monitors the patient's ECG rhythm, detects shockable ventricular tachyarrhythmias and advises the operator to press the shock button to deliver defibrillation therapy for the treatment of sudden cardiac arrest.

Manual Mode

In this mode, as is the case with all standard manual defibrillators, the operator manually selects the energy setting, manually selects the charge button, determines if the patient has a shockable ventricular tachyarrhythmia and decides when to manually deliver defibrillation therapy.

Pacing

The device includes a non-invasive external pacer that allows a physician and/or trained medical person to provide temporary external cardiac pacing for patients with symptomatic bradycardia or asystole.

10.8 Summary of Technological Characteristics

The Powerheart CRM utilizes an FDA-cleared and patented arrhythmia detection software that continuously monitors, detects and delivers therapy, according to physician-prescribed parameters, to patients experiencing ventricular tachyarrhythmias. The delivery output is an FDA-cleared, biphasic truncated exponential waveform, which the Powerheart CRM can transmit in one of three modes: Automatic, Advisory and Manual. The Powerheart CRM also provides non-invasive temporary cardiac pacing.

10.9 Summary of Performance Data

The performance tests conducted on the Powerheart Cardiac Rhythm Module will follow the applicable portions of the ANSIAAMI DF2 and DF39 requirements for cardiac defibrillator devices and automatic external defibrillators to demonstrate the safety and efficacy of the CRM device.

{2}------------------------------------------------

10.10 Conclusion

The performance tests conducted on the Powerheart CRM will be subjected to pre-determined pass criteria that will support claims of substantial equivalence.

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 6 2001

Sew-Wah Tay, Ph.D. Vice President Cardiac Science, Inc. 5420 Feltl Road Minneapolis, MN 55343

Re: K012197

Trade Name: Powerheart® Cardiac Rhythm Module Regulation Number: 21 CFR 870.5300 Regulation Name: Defibrillator, Automatic, External Regulatory Class: Class III (three) Product Code: MKJ, LDD, DRO Dated: October 29, 2001 Received: October 30, 2001

Dear Ms. Tay:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Sew-Wah Tay, Ph.D.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

signature

James H. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows a heavily edited or obscured piece of text. Due to the extensive alterations, it is impossible to discern the original content or meaning of the text. The image appears to be a close-up of the text, with the modifications applied directly to the characters.

510(k) Number: K012197

NOV 1 6 2001

Device Name: Powerheart® Cardiac Rhythm Module

Indication for Use:

The Powerheart Cardiac Rhythm Module (CRM) is intended to acquire the ECG rhythm for the detection of, and to provide treatment for, ventricular tachyarrhythmias of patients who are at risk of sudden cardiac arrest. The device is intended to be used in medically supervised environments by trained personnel, in which patients are under the direct care of physicians and/or medical persons authorized by the state, province or country regulations in which they practice. The Powerheart CRM can be used in three modes: fully automatic mode, advisory mode or manual mode and includes a pacer which allows an operator to deliver external stimuli for temporary transcutaneous cardiac pacing. The Powerheart has previously undergone clinical testing on adults which support the safety and effectiveness of using the device on patients at risk of sudden cardiac arrest who are at least 8 years of age. The device is not classified as a wearable external defibrillator and is not intended for use in the home environment.

Fully Automatic Mode

In this mode, the device continuously monitors the patient's ECG rhythm, detects shockable ventricular tachyarthythmias and automatically provides defibrillation therapy for the treatment of sudden cardiac arrest.

Advisory (AED) Mode

In this mode, the device continuously monitors the patient's ECG rhythm, detects shockable ventricular tachyarrhythmias and advises the operator to press the shock button to deliver defibrillation therapy for the treatment of sudden cardiac arrest.

Manual Mode

In this mode, as is the case with all standard manual defibrillators, the operator manually selects the energy setting, manually selects the charge button, determines if the patient has a shockable ventricular tachyarrhythmia and decides when to manually deliver defibrillation therapy.

Pacing

The device includes a non-invasive external pacer that allows a physician and/or trained medical person to provide temporary external cardiac pacing for patients with symptomatic bradycardia or asystole.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular & Respiratory Devices
K012/197
Prescription UseOROver-The Counter Use
(Per 21 CFR 801.109)

§ 870.5310 Automated external defibrillator system.

(a)
Identification. An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.(b)
Classification. Class III (premarket approval)(c)
Date PMA or notice of completion of PDP is required. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED that was in commercial distribution before May 28, 1976. A PMA will be required to be submitted to the Food and Drug Administration by April 29, 2015, for any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976, or that has, by April 29, 2015, been found to be substantially equivalent to any AED accessory described in paragraph (a) that was in commercial distribution before May 28, 1976. Any other AED and AED accessory described in paragraph (a), shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.