K Number
K011826

Validate with FDA (Live)

Date Cleared
2002-01-24

(227 days)

Product Code
Regulation Number
870.1750
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EPS320 Stimulator System is an electrical stimulus generator for diagnostic cardiac stimulation during electrophysiological testing of the human heart.

Device Description

The Micropace EPS320 Cardiac Stimulator (Clinical Stimulator) is an external programmable computerised diagnostic cardiac stimulator. It is used in specialised hospital electrophysiological diagnostic laboratories by specialist cardiologists to electrically stimulate the heart to initiate and terminate tachyarrhythmias and allow measurement of refractory and conduction properties of the heart by third party equipment. The EPS320 system consists of a Stimulus Generator Unit (SGU) comprising of a manufactured metal instrument case and a standard IBM clone type Personal Computer (PC), the two interconnected by a serial data connection. The device has two independent fully programmable and isolated constant current pulse generator channels intended for temporary programmacro and locativa third party transvenous intracardiac electrodes. The pacing stimulus is output via a Stimulus Connection Box and may typically be routed to the intravenous pacing electrodes via third party electronic switching equipment. During normal operation of the EPS320, custom software on the PC provides keyboard input device and a graphical user interface which the operator uses to control the stimulation process in real time. The custom PC software interprets the user instructions and sends specific real-time commands to the I he ousehi I O software and and to control its function and generate appropriate pacing stimulus pulses. The two isolated outputs can deliver stimulus pulse widths from 0.5ms to 10ms, adjustable in 0.5 to 1ms steps with current amplitudes adjustable from 0.1mA to 25mA in steps of 0.1mA. Current delivery is limited within the specified range by a maximum available output voltage of 26V. The EPS320 is not a life support device and may be used only in the presence of a backup cardiac THE LE SES IS .In case of failure of mains power supply, malfunction of the PC or its software, or loss of the data link, the SGU can operate independently of the PC in the Manual Backup mode, powered by an internal backup battery. This operational mode can provides regular pacing stimuli at intervals and pulse current amplitudes programmable from and displayed on the SGU's front panel. It may be used to briefly support a bradycardic patient until temporary external pacing is established using an approved temporary pacemaker. A secondary electrically self-contained Emergency Fixed Pace approved willing the front of the SGU may be used for the same purpose in case of loss of normal SGU function.

AI/ML Overview

The provided text describes the Micropace Model EPS320 Clinical Stimulator and seeks 510(k) clearance based on its substantial equivalence to the predicate device, the DCI Model EP-2 Clinical Stimulator. The document primarily focuses on technical specifications and functional comparisons rather than a study demonstrating performance against specific acceptance criteria in a clinical setting.

Here's an analysis based on the provided input:

Acceptance Criteria and Device Performance

The document does not explicitly state "acceptance criteria" in the traditional sense of a performance study with numerical thresholds. Instead, it aims to demonstrate that the EPS320 is substantially equivalent to the predicate device (DCI Model EP-2) and is as safe and effective. The "acceptance criteria" are implicitly met if the FDA agrees that the EPS320's specifications and functionality are comparable or improved, and its differences do not raise new questions of safety or effectiveness.

The "reported device performance" is presented through a direct comparison with the predicate device's specifications in Tables 5-1 and 5-2.

Table of Acceptance Criteria (Implicit) and Reported Device Performance

Feature/ParameterImplicit Acceptance Criteria (Comparable to Predicate Device)Micropace Model EPS320 Performance (Reported Values)Predicate Device (DCI Model EP-2) Performance (Reported Values)Rationale for Equivalence (as per submission)
1. Isolated Stimulus ChannelsSame number of channels.22Same number of Channels as the EP-2.
2. Pulse AmplitudeRange sufficient for therapeutic purposes; comparable to EP-2.0.1 to 25mA into 800Ω load Increment: 0.1 mA up to 1mA Amplitude Accuracy: ± 2% or ± 0.2 mA (whichever is greater)0.1-20 mA (2500Ω load) Accuracy: 5% or 0.2 mA (whichever is larger)EPS320 output current is sufficient for therapeutic purposes and substantially the same as the EP-2.
3. Pulse DurationWithin accepted clinical requirements; comparable to EP-2.0.5 to 10 msec Increment: 0.5 – 10msec, increments of 1 – 10 msec Accuracy: ± 0.15ms0.5-10 msec (maximum 10% of ISI) Increment: 0.1 msec Accuracy: 0.05 msec (Pulse amplitude > 2 mA)Pulse Duration is within the profession's accepted requirements as attested by 4 years of post-market surveillance of the EPS320.
4. Interstimulus Interval (ISI)Substantially the same as EP-2.180 msec to 9990 msec ± 1msec or 0.1% (whichever is greater) (Range) 30 - 9900 msec ± 1msec or 0.1% (whichever is greater) (Burst) Increment: 1 msec100-10000 ± 1 msec (Range) 10-1000 ± 1 msec (Burst) Increment: 1 msecSubstantially the same as the EP-2.
5. Sequential (AV) DelaySubstantially the same as EP-2, with enhanced safety.Min: 25 - 980msec (maximum ISI – 120 msec) Increment: 1 msec10-1000 ± 1 msec (maximum ISI - 50 msec) Increment: 1 msecA-V delay is substantially the same as the EP-2. EPS320 is limited to ISI less 120 msec which is considered an enhanced safety feature whilst retaining equivalent operational effectiveness.
6. Reprogrammed ProtocolsEquivalent functionality to EP-2, covering common clinical needs.Equivalent protocol to all but 1 protocol in the EP-2 (Refractory Threshold omitted as no longer in common clinical use). Contains additional protocols like Multi-Sx, Nodal ERP / RSync_S2, Overdrive Burst / Load ATP.Threshold, Refractory study, Wenckebach block, SA Node recovery, Pace, Burst, Overdrive, Underdrive, Arrhythmia Induction, Refractory Threshold.EPS320 contains an equivalent protocol to all but 1 protocol in the EP-2, providing substantially equivalent functionality. One protocol, the Refractory Threshold protocol present in the EP-2 is no longer in common clinical use in the experience of the expert designer and was not included in the EPS320.
9. Sensing (ECG synchronization)External ECG input range appropriate; enhanced safety features.External: 50 - 2000mV Internal: Pacing catheter tip Trigger lockout (refractory time): 50 - 5000msec ECG delay: 50-5000msecAutomatic or manual trigger setting - Sensitivity: 1-500 mV Automatic or manual trigger setting - Trigger lockout (refractory time): 10-1000 msec (100 Hz max) Automatic or manual trigger setting - ECG delay: 100-5000 msecExternal ECG input range (mV) for the EPS320 is appropriate for common external ECG sources. Presence of pacing catheter tip sensing in EPS320 is considered an enhanced safety feature. Longer minimum lockout (50 vs. 10msec) is an enhanced safety feature reducing susceptibility to noise, without significant loss of functionality. Lockout period and ECG delay are substantially equivalent to the EP-2.
Hardware PlatformSimilar architecture; safe and effective.Standard "PC" Personal Computer / custom "Stimulus Generator Unit ""Control Terminal" custom computer / custom "Stimulus Module"Both devices have similar architectures. The EPS320's 'open PC architecture' uses verification of software environment and resource integrity and is based on extensive risk analysis. The absolute safety and efficacy of the EPS320 hardware platform has been demonstrated in clinical practice and is thus at least equivalent to that of the predicate EP-2 device.
Backup Pacing CircuitsAs safe or more safe than EP-2.Yes (2) – Manual Backup operation mode and Emergency Fixed Pace circuit.NoThese functions are safety enhancements over the predicate device EP-2.
Device Feedback to UserAs safe or more safe than EP-2.Open Pacing Circuit Warning: Yes Stimulation impedance display: Yes Graphical display of ECG Trigger: YesNoThese features are safety enhancements over the predicate device EP-2.
Power SourceComparable reliability and safety.Mains operated with 2 backup battery systems (12V lead acid, 9V Lithium).Internal rechargeable 6V battery.Mains source of power was considered safer than a battery as it removes the possibility of lack of device availability due to depleted batteries. Batteries in EPS320 are for fault conditions.

Breakdown of Study Information:

  1. Sample Size Used for the Test Set and Data Provenance:

    • The submission does not describe a specific "test set" or a formal clinical study with a defined sample size to prove the device meets acceptance criteria in a quantitative manner.
    • The provided information states that the EPS320 is an "expert-designed product, based upon the electrophysiological expertise of the cardiologist Managing Director and on extensive post market experience with an earlier EPS220 model in service in eight (8) Australian hospitals since 1995." It also mentions that "The current and earlier EPS320 models have been in clinical use in more than 40 clinical centres predominantly in Australia, Europe and India, as well as in Turkey, Vietnam and China for up to 4 years without significant adverse effects." This "post-market experience" serves as informal evidence of performance and safety rather than a structured clinical test set.
    • Data Provenance: Australia, Europe, India, Turkey, Vietnam, and China.
    • Retrospective/Prospective: The "post-market experience" suggests a retrospective review of observational data from clinical use, not a prospective study designed to evaluate the EPS320 against specific endpoints.
  2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    • Since there's no formal "test set" or ground truth establishment process described for a specific study, this information is not directly applicable.
    • However, the device design itself is based on "electrophysiological expertise of the cardiologist Managing Director." This individual serves as a key "expert" influencing the device's design and functionality. The submission does not detail specific qualifications of this Managing Director beyond "cardiologist."
  3. Adjudication Method for the Test Set:

    • Not applicable, as no formal test set and ground truth establishment with adjudication is described.
  4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a cardiac stimulator, not an AI-driven diagnostic imaging device that uses multi-reader studies.
  5. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. The EPS320 is a medical device that requires human operation by specialist cardiologists to apply electrical stimulation. It is not an autonomous algorithm.
  6. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.):

    • No formal ground truth for a test set is described. The closest equivalent to "ground truth" for the device's design and effectiveness claims comes from:
      • Expert experience: "electrophysiological expertise of the cardiologist Managing Director."
      • Clinical outcomes/safety data: "4 years of post-market surveillance of the EPS320" and "clinical use in more than 40 clinical centres... for up to 4 years without significant adverse effects." This implicitly refers to positive clinical outcomes and lack of adverse events as evidence of safety and effectiveness in real-world scenarios.
      • Comparison to predicate: The EP-2's established safety and effectiveness serve as the baseline "ground truth" for the EPS320's substantial equivalence claim.
  7. The Sample Size for the Training Set:

    • Not applicable. This submission does not describe an AI or machine learning algorithm that requires a "training set." The extensive clinical use of "current and earlier EPS320 models" in "more than 40 clinical centres" acts more as an extended real-world validation of the device's design and safety.
  8. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set described for an AI/ML algorithm.

{0}------------------------------------------------

Version 5.5

5. 510(k) Summary

K011826

5.1 21 CFR 807.92(a)(1) - Submitter details

Submitter's name: Address:

Phone number: Fax number:

Contact person:

Date the summary was prepared:

JAN 2 4 2002

Dr. Michael Cejnar 571 Canterbury Road, Campsie, NSW. 2194 Australia. 61+ 2 9787 6166

61+ 2 9787 6144

Dr. Michael Cejnar

25th May 2001

5.2 21 CFR 807.92(a)(2) - Device name

Trade name: Common name: Classification name:

Micropace model EPS320 Clinical Stimulator Programmable Cardiac Electrophysiological Stimulator External programmable pacemaker pulse generator (Per 21 CFR section 870.1750)

5.3 21 CFR 807.92 (a)(3) - Identification of Substantially Equivalent device

Predicate Company name

Predicate Device name

Digital Cardiovascular Instruments Inc. Berkeley CA

DCI Model EP-2 Clinical Stimulator

{1}------------------------------------------------

5.4 21 CFR 807.92 (a)(4) - Description of the Micropace model EPS320

The Micropace EPS320 Cardiac Stimulator (Clinical Stimulator) is an external programmable computerised diagnostic cardiac stimulator. It is used in specialised hospital electrophysiological diagnostic laboratories by specialist cardiologists to electrically stimulate the heart to initiate and terminate tachyarrhythmias and allow measurement of refractory and conduction properties of the heart by third party equipment.

The EPS320 system consists of a Stimulus Generator Unit (SGU) comprising of a manufactured metal instrument case 12.25"(W) x 13.25"(D) x 3.5"(H) and a standard IBM clone type Personal Computer (PC), the two interconnected by a serial data connection. The device has two independent fully programmable and isolated constant current pulse generator channels intended for temporary programmacro and locativa third party transvenous intracardiac electrodes. The pacing stimulus is output via a Stimulus Connection Box and may typically be routed to the intravenous pacing electrodes via third party electronic switching equipment.

During normal operation of the EPS320, custom software on the PC provides keyboard input device and a graphical user interface which the operator uses to control the stimulation process in real time. The custom PC software interprets the user instructions and sends specific real-time commands to the I he ousehi I O software and and to control its function and generate appropriate pacing stimulus pulses. The two isolated outputs can deliver stimulus pulse widths from 0.5ms to 10ms, adjustable in 0.5 to 1ms steps with current amplitudes adjustable from 0.1mA to 25mA in steps of 0.1mA. Current delivery is limited within the specified range by a maximum available output voltage of 26V.

The EPS320 is not a life support device and may be used only in the presence of a backup cardiac THE LE SES IS .In case of failure of mains power supply, malfunction of the PC or its software, or loss of the data link, the SGU can operate independently of the PC in the Manual Backup mode, powered by an internal backup battery. This operational mode can provides regular pacing stimuli at intervals and pulse current amplitudes programmable from and displayed on the SGU's front panel. It may be used to briefly support a bradycardic patient until temporary external pacing is established using an approved temporary pacemaker. A secondary electrically self-contained Emergency Fixed Pace approved willing the front of the SGU may be used for the same purpose in case of loss of normal SGU function.

The EPS320 is an expert-designed product, based upon the electrophysiological expertise of the cardiologist Managing Director and on extensive post market experience with an earlier EPS220 model in service in eight (8) Australian hospitals since 1995. The current and earlier EPS320 models have been in clinical use in more than 40 clinical centres predominantly in Australia, Europe and India, as well as in Turkey, Vietnam and China for up to 4 years without significant adverse effects.

5.5 21 CFR 807.92 (a)(5) - Statement of the Intended Use

INTENDED USE:

The EP320 stimulator is intended to be used for diagnostic electrical stimulation of the heart for the purpose of I it is 220 alimitation of tachyarrhythmias, refractory measurements and measurements of electrical conduction.

{2}------------------------------------------------

5.6 21 CFR 807.92 (a)(6) - Summary of Technological differences

This section compares the specifications and other relevant performance characteristics of the EPS320 and the predicate device, EP-2. It tabulates the compared items side by side and then specifies the rationale for claims of substantial equivalence.

5.6.1 Definition of terms used in the following tables

TERMDEFINITIONTERMDEFINITION
ISIERPInterstimulus IntervalEffective RefractoryPeriodSNRTATPSinus Node Recovery TimeAnti Tachycardia Pacing
PCECGIECGPersonal ComputerElectrocardiographIntracardiac ECGSGUIECAICDStimulus Generator UnitInternational Electrotechnical CommissionAutomatic Implantable CardioverterDefibrillator

5.6.2 Technological differences between devices.

A comparison of published performance specifications in the "Specifications" sections of the EPS320 and in the equivalent section of the predicate device, EP-2 documentation is shown in Table 5-1 below.

ComparativeelementDCI Model EP - 2(Predicate Device)Micropace model EPS320
A. ManufacturerDigital CardiovascularInstruments, Inc. Berkeley, CAMicropace Pty Ltd. Campsie, N.S.WAustralia
B. FDA ConcurrenceYes (K854196/B)Pending
1. Isolated StimulusChannels22
2. Pulse Amplitude –Range0.1-20 mA ( $2500 \Omega$ load)0.1 to 25mA into $800 \Omega$ load
Increment0.1 mA up to 1mA Amplitude
Accuracy5% or 0.2 mA (whichever islarger)$\pm$ 2% or $\pm$ 0.2 mA, (whichever is greater)
3. Pulse durationRange0.5-10 msec (maximum 10% ofISI)0.5 to 10 msec
Increment0.1 msec0.5 – 10msec, increments of 1 – 10 msec
Accuracy0.05 msec (Pulse amplitude > 2mA)$\pm$ 0.15ms
4. InterstimulusInterval (ISI)Range100-10000 $\pm$ 1 msec180 msec to 9990 msec $\pm$ 1msec or 0.1%(whichever is greater)
ComparativeelementDCI Model EP - 2(Predicate Device)Micropace model EPS320
Range (Burst)10-1000 ± 1 msec30 - 9900 msec ± 1msec or 0.1% (whicheveris greater)
Increment1 msec1 msec
5. Sequential (AV)Delay
Range10-1000 ± 1 msec (maximum ISI- 50 msec)Min: 25 - 980msec (maximum ISI – 120msec)
Increment1 msec1 msec
6. ReprogrammedThresholdThreshold,
ProtocolsRefractory studyNodal ERP / RSync_S2
Wenckebach blockWenckebach
SA Node recoverySNRT
PacePace
BurstBurst Pace
OverdriveOverdrive Burst / Load ATP
Underdrive-
Arrhythmia InductionMulti-Sx
Refractory Threshold-
7. Programmable55 programmable stimulator complete set-ups
Protocol keys
8. Number of6 (s2-s7)Max: 6 (s2-s7)
Extrastimuli
9. Sensing (ECG
synchronisation)
Automatic or manual1-500 mVExternal: 50 - 2000mV
trigger setting -Internal: Pacing catheter tip
Sensitivity
Automatic or manual10-1000 msec (100 Hz max)50 - 5000msec
trigger setting -
Trigger lockout
(refractory time)
Automatic or manual100-5000 msec50-5000msec
trigger setting -
ECG delay
10. AdditionalStimulus Channel A MarkerSync 3 Input Marker
Outputs:Stimulus Channel B MarkerSync 3 Input Marker
Programmable Auxiliary/PaperadvanceSync 1 Input Marker
ECG trigger markerSync 2 Input MarkerN/A
11. Power SourceInternal rechargeable 6V batteryMain power source: 2201-240VAC or 110-120VAC to 14.5 VDC, 750mA Low VoltagePower Supply Transformer
Backup power: 1. 12V 2.1Ah lead acidbattery
Backup power: 2. 9V PP3 Lithium battery
Battery operating time:17 hours (typical)Indefinite (Mains operated)
12. Stimulation
Module
Physical Dimensions3.0 x 12.0 x 14.03.3 x 12.25 x 13.25
(inches)
ComparativeelementDCI Model EP - 2(Predicate Device)Micropace model EPS320
Weight (lbs)11.218. 7
13. Control Terminal
ListingsULUL & CE
Type14 in diagonal green phosphorCRTPC with LCD display
Display dimensions(inches)13.5 x 14 x 1512 wide x 9 high x 3 deep (flat Screen)
Footprint dimensions(inches)12.25 x 10.2510.6wide x 7.5deep x 15high (includingLCD panel)
Keyboard dimensions(inches)2.25 x 17.25 x 7.6Modified QWERTY keyboard1.6 high x 19.0 wide x 8.4 deepModified QWERTY keyboard
Weight (lbs)31.5Approx. 12.0 lb.
14. IsolationTransformer
Leakage currentLess than 100 μALess than 100 μA
Power rating500 VA400VA
Physical dimensions(inches)8.12 x 9.00 x 6.757.75 x 5 x 4
Weight (lbs)2215.4
15. Charger
ListingsUL544 3-prong wall plug-in,automatic cut-offInternal SGU 'on-board' circuitry
Charge/Discharge timerationLess than 1Less than 1
Charge time from lowbattery indication tofull charge< 16 hours<16 hours
Dimensions (inches)2.55 x 2.89 x 1.88N/A
16. Environment
Operating Temperature+10 to +40 degrees Celsius+10 to +40 degrees Celsius
Storage Temperature-20 to +60 degrees Celsius-20 to +60 degrees Celsius
Relative Humidity25% to 90% non-condensing25% to 90% non-condensing
Operating Altitude0 to 4572 m0 to 4572 m
Storage Altitude0 to 7620 m0 to 7620 m

{3}------------------------------------------------

510(k) Pre-market Submission Micropace Pty Ltd.

Version 5.5

1 Input electrical mains voltage will be dependant on geographical location/available mains voltage

:

:

j

,

{4}------------------------------------------------

510(k) Pre-market Submission Micropace Pty Ltd.

Version 5.5

Table 5-1 Substantial Equivalence Comparison -1

Table 5-2 below lists technological differences in performance characteristics between the EPS320 and the predicate EP-2 device not included in the "Specifications" section of either device.

Comparative itemDCI Model EP - 2(Predicate Device)Micropace model EPS320
1. Hardwarecomputer platform"Control Terminal" customcomputer /custom "Stimulus Module"Standard "PC" Personal Computer /custom "Stimulus Generator Unit "
2. User interface
OutputText Windowed DisplayGraphical Windowed Display
3. ECG Sensing
Catheter-Tip IECGsensingNoYes
Balanced ChargePacingNoYes

.

{5}------------------------------------------------

Version 5.5

4. Device feedback touser
Open Pacing CircuitWarningNoYes
Stimulation impedancedisplayNoYes
Graphical display ofECG TriggerNoYes
5. Backup PacingCircuitsNoYes (2)
6. StimulationProtocolImplementation
A-V Joined PacingStimulationLimitedYes
Load_ATP - AICD-Style Overdrive PacingNoYes
ECG trigger time outNoYes
On the flyprogramming of allparametersLimitedYes
Shortcut KeysYesYes (different)

Table 5-2 Substantial Equivalence Comparison -2

5.6.3 Rationale for claims of equivalence of devices

The EPS320 and the predicate device are equivalent in their published specifications (see Table 5-1) for the following reasons:

Table 5.1SectionReducedsafety oreffectiveness issues?Is the EPS320 as safe and effective as the EP-2?(Rationale)
1 IsolatedStimulusChannelsNoSame number of Channels as the EP-2.
2 PulseAmplitudeNoEPS320 output current is sufficient for therapeutic purposes and substantiallythe same as the EP-2.
3 PulsedurationNoPulse Duration is within the profession's accepted requirements as attested by4 years of post-market surveillance of the EPS320.
4InterstimulusInterval (ISI)NoSubstantially the same as the EP-2.
5 Sequential(AV) DelayNoA-V delay is substantially the same as the EP-2. EPS320 is limited to ISI less120 msec which is considered an enhanced safety feature whilst retainingequivalent operational effectiveness.
6Reprogrammed ProtocolsNoEPS320 contains an equivalent protocol to all but 1 protocol in the EP-2,providing substantially equivalent functionality. One protocol, the RefractoryThreshold protocol present in the EP-2 is no longer in common clinical use inthe experience of the expert designer and was not included in the EPS320.
Table 5.1SectionReducedsafety oreffectiveness issues?Is the EPS320 as safe and effective as the EP-2?(Rationale)
7Programmable ProtocolkeysNoEPS320 implements programmability by storing and recalling completestimulator set ups rather than individual dual protocols. The two approachesare basically functionally similar.
8 Number ofExtrastimuliNoThe number of selectable extra stimuli is the same as the EP-2, dependant onthe protocol selected.
9 Sensing(ECGsynchronisationNoExternal ECG input range (mV) for the EPS320 is appropriate for commonexternal ECG sources found in contemporary laboratories.
Presence of pacing catheter tip sensing in EPS320 is considered an enhancedsafety feature in providing an additional source of ECG trigger.
Longer minimum lockout in EPS320 (50 vs. 10msec) is an enhanced safetyfeature reducing susceptibility to noise, without significant loss offunctionality. Lockout period and ECG delay are substantially equivalent tothe EP-2
10 AdditionalOutputs:NoSync 3 output on the EPS320 provides essentially the same function asStimulus Ch 'A' and Stimulus Ch 'B' outputs of the EPS320
There is no requirement for an auxiliary (printer) output on the EPS320 as it isconnected to recording external systems.
11 PowerSourceNoAll components of the EPS320 system are powered from electrical mainspower via an isolation transformer. The SGU is powered from this isolatedmains by a 14.5V DC Power Supply Unit containing a transformer. TheEPS320 system, including these accessories (isolation transformer and DCPower Supply Unit, have been subjected to appropriate electrical and safetytests (See appendix A1 and A2).
Mains source of power was considered safer than a battery as it removes thepossibility of lack of device availability due to depleted batteries.
Batteries contained in the EPS320 are used solely under fault conditions.
12StimulationModuleNoPhysical attributes of the EPS320 are substantially the same as the EP-2, whilstmaintaining operational functionality in similar environment
13 ControlTerminalYesThe Computer used with the EPS320 can have software files added by theoperator creating potential for corruption of operating environment. To controlthis possibility, the EPS320 software checks for integrity of operatingenvironment during startup. Other parameters are essentially equivalent as theEP-2 terminal.
14 IsolationTransformerNoThe Isolation transformer is essentially equivalent to that described in the EP-2documentation.
15 ChargerNoThe EPS320 does not require a separate battery charger. Batteries are notrequired for normal operation.
16EnvironmentNoEnvironmental conditions are essentially equivalent as the EP-2.
Table 5.2SectionReducedsafety oreffectiveness issues?Is the EPS320 as safe and effective as the EP-2?(Rationale)
1. HardwareplatformYesBoth devices have similar architectures, in that a microprocessor is used tocontrol output pulse generation circuitry, and a physically separate computingdevice is used to provide user interface via a video display unit and akeyboard.The two devices differ predominantly in using different implementations toachieve similar functional requirements.The EP-2 used propriety microprocessor software whereas the EPS320 uses an'open PC architecture' - a standard PC loaded with custom software. Therewill also be differences in software architectures and interface designs andimplementations.The EPS320 system removes the danger of corruption of software environmentfrom unauthorised addition of third party software or incorrect operatingsystems by verification of software environment and resource integrity prior tostartup. The custom software is implemented on a stable DOS platform and itsdesign is based on extensive risk analysis and subject to appropriate designcontrol, verification and validation.The absolute safety and efficacy of the EPS320 hardware platform has beendemonstrated in clinical practice and is thus at least equivalent to that of thepredicate EP-2 device.
2. UserinterfaceNoThe user interface displays and input keyboards of the two devices aresubstantially functionally equivalent technologies.Software performance of the EPS320 is designed for user flexibility andergonomy and while these performance features probably represent the mainfunctional difference between the EPS320 and the predicate device they havelimited impact on device safety.
3. ECGSensingNoCatheter-Tip IECG sensing: The predicate EP-2 can trigger only on anexternal ECG source, which is usually derived from third party equipmentwhich amplifies the patient's intracardiac ECG (IECG) or surface ECG.The EPS320 contains an ECG amplifier connected to the pacing channel andcan thus trigger on this IECG source or on an external ECG source.Balanced Charge Pacing: This feature refers to a configuration of IECGamplifier sensing between stimulation pulses to minimise electrical chargebuild up on the pacing electrodes. It has no impact on the stimulation pulseshape, amplitude or efficacy and thus has no performance or safety impact.
4. Devicefeedback touserNoOpen Pacing Circuit Warning: This feature indicates to the user whenprogrammed stimulation current fails to be delivered to the patient.Stimulation impedance display: This feature displays to the user approximatestimulus-to-stimulus impedance of the stimulation circuit.Graphical display of ECG Trigger: This feature conveys to the user thequality of the ECG source used for Triggering.These features are safety enhancements over the predicate device EP-2.
5. BackupPacingCircuitsNoThe EPS320 has a Manual Backup operation mode which operates in theabsence of a correctly functioning computer and allows user to perform simplestimulation tasks.The EPS320 also contains a separate, self-powered Emergency Fixed Pacecircuit, capable of delivering pacing stimuli in the case of loss of function ofthe Stimulus Generator Unit.These functions are safety enhancements over the predicate device EP-2.
6. StimulationProtocolOptionsNoBoth devices provide predominantly equivalent stimulation protocols withdifferences confined to issues hotkey and menu layout, ergonomy and userconvenience. Below are representative examples.A-V Joined Pacing Stimulation: This stimulation allows simultaneousstimulation of both output channels for only some pacing stimuli in a pacingprotocol and not others.Load_ATP - AICD-Style Overdrive Pacing: This additional protocol in theEPS320 is an automated combination of the Burst and Overdrive protocols inthe predicate device.ECG trigger time out: In case of failure of arrival of a Trigger event, pacingwill commence in any case after this period of time.On the fly programming of all parameters: All pacing parameters can bealtered by hotkeys without stopping stimulation.Shortcut Keys: Various shortcut keys, different to the predicate device, areimplemented to enhance efficiency of use.

{6}------------------------------------------------

Table 5-3 Rationale of the data in Table 5-1

{7}------------------------------------------------

The EPS320 and the predicate device are equivalent in performance characteristics other than specifications (see Table 5-2) for the following reasons:

{8}------------------------------------------------

Version 5.5

Table 5-4 Rationale of the data in Table 5-2

5.6.4 Conclusion

The discussion above demonstrates that the EPS320 and the predicate device EP-2 posses similar architectural and functional characteristics, although differences exist in the specific technological implementation of some of the functional requirements. The EPS320, has demonstrably greater number of safety characteristics than the EP-2.

5.7 21 CFR 807.92 (b)(2) - Clinical comparison to the predicate device

Not Applicable

{9}------------------------------------------------

5.8 21 CFR 807.92 (b)(3) - Clinical & non-clinical test conclusions

Having analysed the characteristics that differ and those which are similar between the Model EP-2 and Model EPS320, we have concluded that, in our opinion:

  • The intended use of the EPS320 and EP-2 are the same. .
  • Safety of the EPS320 is as effective if not more effective than the predicate device. .
  • Differences between the systems are limited to the technological implementation of . equivalent functions, layout design and ergonomy of the user interface
  • The EPS320 is substantially equivalent to the EP-2 . .

5.9 21 CFR 807.92 (c) - 510(k) Summary

Presentation of the 510[k] summary is considered compliant with the requirements of this section.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the circumference. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human profiles, with three wavy lines forming the central design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 2 4 2002

Mr. Michael Cejnar Micropace Pty. Ltd. 571 Canterbury Road Campsie, NSW 2194 Australia

Re: K011826 EPS320 Cardiac Stimulator Regulation Number: 870.1750 Regulation Name: External Programmable Pacemaker Pulse Generator Regulatory Class: Class II (two) Product Code: JOQ Dated: October 23, 2001 Received: October 26, 2001

Dear Mr. Cejnar:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. In addition, if you wish to change or expand the current indications for use to include non-military environments, you will need to submit a new 510(k) premarket notification, and receive FDA clearance prior to marketing the device.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{11}------------------------------------------------

Page 2 - Mr. Michael Cejnar

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Dela Teh

Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{12}------------------------------------------------

4. Statement of Indications for Use

Indications for Use Statement

Ver/ 3 - 4/24/96

Applicant: Micropace Pty Ltd.

510(k) Number (if known):

Device Name: EPS320

Indications For Use:

The EPS320 Stimulator System is an electrical stimulus generator for diagnostic cardiac stimulation during electrophysiological testing of the human heart.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

oncurrence of CDRH, Office of Device Evaluation (ODE)

(Per 21 CFR 801.109)

1-2-96)

Division of Cardiovascular & Respiratory Devices

Division of Cardiovascular & Respiratory Devices
510(k) Number K611826

§ 870.1750 External programmable pacemaker pulse generator.

(a)
Identification. An external programmable pacemaker pulse generators is a device that can be programmed to produce one or more pulses at preselected intervals; this device is used in electrophysiological studies.(b)
Classification. Class II (performance standards).