K Number
K011660

Validate with FDA (Live)

Date Cleared
2001-08-21

(84 days)

Product Code
Regulation Number
870.3680
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Weck Cardiac Pacing Wires are intended for use in temporary cardiac pacing following cardiac surgery in adult and pediatric patients.

Device Description

Weck cardiac pacing wires consist of a small curved stainless steel needle attached to fluoropolymer coated stainless steel braided wire which is attached on the opposite end to a Keith or Milner breakaway needle (see picture below). The wire is bare at the distal end to allow conduction of pacing signals to the heart. This is a standard temporary epicardial pacing wire design.

There are three design variations included in this submission that are offered as a convenience to the user:

  1. Color coding White and orange color coding of the insulation allows the user to differentiate between atrial and ventricular wires after chest closure.
  2. Heart needle Three heart needle sizes are intended to allow for variations in physiology between patients. Another variation has no curved heart needle for surgeons who prefer to suture the wire to the epicardium rather than implanting it.
  3. Retention feature A "wing" version wire is formed by peeling back the insulation in three places, 120° apart, for a distance of 1/2" near the curved heart needle. The "wing" is intended as a convenience feature to provide a suture-free method of securing the electrode to the heart.
AI/ML Overview

The provided text describes a 510(k) summary for Weck Cardiac Pacing Wires, which aims to demonstrate substantial equivalence to predicate devices rather than conducting a full de novo study with strict acceptance criteria and performance evaluations as might be found for a novel device. Therefore, much of the requested information regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the traditional sense of a clinical trial or algorithm performance study is not explicitly detailed.

However, based on the document, we can infer the "acceptance criteria" as ensuring the device performs equivalently to predicate devices on key functional attributes. The "study" refers to the engineering and bench testing performed to confirm these characteristics.

Here's a breakdown of the requested information based on the provided text, with explanations where information is not explicitly stated or is inferred from the regulatory context:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Inferred from reported evaluations)Reported Device Performance
Biocompatibility according to FDA G95-1 / ISO 10993-1 requirementsConfirmed biocompatibility.
Electrical continuityConfirmed to be equivalent and perform acceptably.
Strength of the needle swages and ability to withstand tensile forces during useConfirmed to be equivalent and perform acceptably.
Corrosion resistanceConfirmed to be equivalent and perform acceptably. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Heart needle insertion forceConfirmed to be equivalent and perform acceptably. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Extraction forceConfirmed to be equivalent and perform acceptably. ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Fatigue resistanceConfirmed to be equivalent and perform acceptably. ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Compatibility with pacemakersConfirmed to be equivalent and perform acceptably.
Lead pacing impedanceConfirmed to be equivalent and perform acceptably.
Lead sensing impedanceConfirmed to be equivalent and perform acceptably.
Substantial equivalence to predicate devices (Weck's preamendment Temporary Cardiac Electrodes, Ethicon TPW 32 and TPW 92 pacing wires, and Medtronic Model 6500 and 6491 temporary pacing wires) for stated intended use in temporary cardiac pacing following cardiac surgery in adult and pediatric patients.The FDA reviewed the 510(k) and determined the device is substantially equivalent to legally marketed predicate devices for the stated indications for use.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify sample sizes for "test sets" in terms of patient data or clinical trials. The "testing" refers to bench and engineering evaluations. The provenance of this engineering testing data (e.g., country of origin, retrospective/prospective) is not provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and not provided. The "ground truth" for this type of device (a medical wire) would be based on established engineering standards and material science, not expert consensus on interpretations of images or patient data.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable and not provided. Adjudication methods are typically used in clinical trials or studies where there is subjective assessment by multiple reviewers. This submission focuses on objective engineering and performance characteristics.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device, nor is it a device where "human readers" would be involved in interpreting its output. No MRMC study was conducted or mentioned.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for the performance evaluations (electrical continuity, strength, corrosion resistance, etc.) would be established against engineering specifications, material standards, and industry-accepted benchmarks for temporary cardiac pacing wires, based on the performance of the predicate devices. This is implied by the statement "Weck evaluated each characteristic of the wires important to proper functioning and confirmed that each of the wires is equivalent and performs acceptably."

8. The sample size for the training set

Not applicable. This is not a machine learning or AI device that requires a "training set."

9. How the ground truth for the training set was established

Not applicable. No training set was used.

{0}------------------------------------------------

AUG 21 2001

K011660 p 1/2

Premarket Notification [510(k)] Summary

Submitter Name, Address, and Date of Preparation 1.

Brian J. Young Sr. Regulatory Affairs Manager Weck Closure Systems One Weck Drive Research Triangle Park, NC 27709

(919) 361-4041 Telephone: Date Prepared: December 13, 2000

Name of the Device, Common, Proprietary (if known), and Classification 2.

Classification Name:Temporary pacemaker electrode
Common Name:Temporary epicardial pacing wire
Proprietary Name:Weck Cardiac Pacing Wires

Identification of the legally marketed device to which the submitter claims 3. equivalence

Weck Temporary Cardiac Pacing Wires are substantially equivalent to Weck's preamendment Temporary Cardiac Electrodes, Ethicon TPW 32 and TPW 92 pacing wires, and Medtronic Model 6500 and 6491 temporary pacing wires.

Description of the Device 4.

Weck cardiac pacing wires consist of a small curved stainless steel needle attached to fluoropolymer coated stainless steel braided wire which is attached on the opposite end to a Keith or Milner breakaway needle (see picture below). The wire is bare at the distal end to allow conduction of pacing signals to the heart. This is a standard temporary epicardial pacing wire design.

Image /page/0/Picture/12 description: The image shows a freehand drawing of a spiral. The spiral is made up of multiple lines that curve around a central point. The lines are drawn in a dark color, and the background is white. The spiral is not perfectly symmetrical, and the lines are not perfectly smooth.

{1}------------------------------------------------

llbbo p 2/2

Page 2 / Weck Cardiac Pacing Wire 510(k) Summary

There are three design variations included in this submission that are offered as a convenience to the user:

  • Color coding White and orange color coding of the insulation allows the user to 1. differentiate between atrial and ventricular wires after chest closure.
    1. Heart needle Three heart needle sizes are intended to allow for variations in physiology between patients. Another variation has no curved heart needle for surgeons who prefer to suture the wire to the epicardium rather than implanting it.
    1. Retention feature A "wing" version wire is formed by peeling back the insulation in three places, 120° apart, for a distance of 1/2" near the curved heart needle. The "wing" is intended as a convenience feature to provide a suture-free method of securing the electrode to the heart.

ડ. Intended Use of the Device

Weck Cardiac Pacing Wires are intended for use in temporary cardiac pacing following cardiac surgery in adult and pediatric patients.

Summary of Technological Characteristics 6.

The technological characteristics of the modified device are the same as or equivalent to the predicate device.

7. Performance Data

Weck confirmed biocompatibility of the wires according to FDA G95-1 / ISO 10993-1 requirements.

Weck evaluated each characteristic of the wires important to proper functioning and confirmed that each of the wires is equivalent and performs acceptably with respect to: (1) electrical continuity; (2) strength of the needle swages and ability to withstand tensile forces that may be encountered during use; (3) corrosion resistance; (4) heart needle insertion force: (5) extraction force; (6) fatigue resistance: (7) compatibility with pacemakers; (8) lead pacing impedence; and (9) lead sensing impedence.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white circular seal for the Department of Health & Human Services - USA. The seal features an abstract image of an eagle with three stripes for wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top half of the circle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 1 2001

Mr. Brian J. Young Sr. Regulatory Affairs Manager Weck Closure Systems One Weck Drive, P.O. Box 12600 Research Triangle Park, N.C. 27709

Re: K011660

Trade Name: Weck Temporary Pacing Wires Regulation Number: 21 CFR 870.3680 Regulatory Class: Class II (two) Product Code: 74 LDF Dated: May 25, 2001 Received: May 29, 2001

Dear Mr. Young:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this

{3}------------------------------------------------

Page 2 - Mr. Brian J. Young

response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance (DSMICA) at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely vours.

Dake Mella

James E. Dillard III Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

6 Statement of indications for use

510(k) Number (if assigned): Device Name:

RD11440
Weck Cardiac Pacing Wires

INDICATIONS FOR USE

Weck Cardiac Pacing Wires are intended for use in temporary cardiac pacing following cardiac surgery in adult and pediatric patients.

Concurrence of CDRH, Office of Device Evaluation (ODE)

KOLLCA

Division of Cardiovascular & Respiratory Devices
510(k) Number

Prescription Use
(Per 21 CFR 801.109)

§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.

(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.