(90 days)
PROstate Spectroscopy and imaging Exam (PROSE) is an image guided, clinical imaging and spectroscopy package which acquires high resolution anatomical images and volume localized, water/lipid suppressed hydrogen spectra and/or multi-voxel spectroscopic images of the prostate gland using an endo-rectal coil with phased-array coils. PROSE option can be used in conjunction with a Magnetic Resonance Scanner to permit non-invasive acquisition of high resolution images and spectral information about relative concentrations of metabolites of prostate gland that can be interpreted by a trained physician, and yield information that may assist in diagnosis of prostate diseases.
PROSE is a version of the PRESS (Point RESolved Spectroscopy), "double" spin echo pulse sequence that uses a 90° and two slice selective refocusing RF pulses to generate a spin echo from a localized volume. PROSE utilizes the standard Graphic Prescription tools and one or more sets of localizer images to determine the size and location of the spectroscopic volume.
The provided document is a 510(k) summary for the PROstate Spectroscopy and imaging Exam (PROSE) software option. It primarily focuses on demonstrating substantial equivalence to a predicate device and safety, rather than providing detailed performance acceptance criteria and a study proving those criteria.
Therefore, many of the requested sections about specific performance metrics and study details cannot be fully answered from the given text.
Here's a breakdown of what can be extracted and what cannot:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not explicitly stated in terms of quantitative performance metrics for PROSE. The document mainly focuses on safety and substantial equivalence to a predicate device. | "Evaluation testing was done to verify the performance of the option in the clinical environment." (No specific results or metrics are provided). |
Explanation: The document states that "Evaluation testing was done to verify the performance of the option in the clinical environment," but it does not specify what performance metrics were evaluated or what the acceptance criteria for those metrics were. It does not provide any numerical results to report against acceptance criteria.
2. Sample size used for the test set and the data provenance
- Sample size for the test set: Not specified.
- Data provenance (e.g., country of origin of the data, retrospective or prospective): Not specified.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of experts: Not specified.
- Qualifications of experts: Not specified. The document mentions "interpreted by a trained physician," but this is for clinical use, not for establishing ground truth in a study.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Adjudication method: Not specified.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC study done: Not mentioned.
- Effect size of human reader improvement: Not applicable, as no such study is reported.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone performance study done: Not mentioned. The device is described as "clinical imaging and spectroscopy package... can be interpreted by a trained physician," implying human involvement.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of ground truth: Not specified.
8. The sample size for the training set
- Training set sample size: Not specified. The document does not describe a machine learning model that would typically have a separate training set. It describes a software option for an MR scanner.
9. How the ground truth for the training set was established
- How ground truth for training set was established: Not applicable, as no training set is described.
Summary of available information:
The document is a 510(k) premarket notification, which primarily focuses on demonstrating substantial equivalence to a predicate device (GE Medical Systems PROBE software option, K930265) and ensuring safety.
- Key points mentioned:
- PROSE is a software option for GE Magnetic Resonance Scanners.
- It's an image-guided, clinical imaging and spectroscopy package for the prostate gland.
- It acquires high-resolution anatomical images, volume-localized, water/lipid suppressed hydrogen spectra, and/or multi-voxel spectroscopic images.
- It uses an endo-rectal coil with phased-array coils.
- The information can be interpreted by a trained physician to assist in diagnosing prostate diseases.
- It was evaluated for safety according to IEC 601-2-33.
- Evaluation testing was done to verify performance in the clinical environment, but no specific performance data or acceptance criteria are provided.
- The manufacturer concludes it "does not result in any new potential hazards."
{0}------------------------------------------------
AUG 2 2 2001
GE Medical Systems
page 1 of 2
PO Box 414, W-709 Milwaukee, WI 53201 USA
SUMMARY OF SAFETY AND EFFECTIVENESS
- This 510(k) summary of safety and effectiveness information is submitted in o accordance with the requirements of 21 CFR Part 807.87(h).
- o Identification of Submitter Larry A. Kroger, Ph.D., 262-544-3894
- o Identification of the Product PROstate Spectroscopy and imaging Exam (PROSE) software option
| Manufactured by: | GE Medical Systems |
|---|---|
| 3200 N. Grandview Blvd. | |
| Waukesha, WI 53188 |
-
0 Device Description
PROSE is a version of the PRESS (Point RESolved Spectroscopy), "double" spin echo pulse sequence that uses a 90° and two slice selective refocusing RF pulses to generate a spin echo from a localized volume. PROSE utilizes the standard Graphic Prescription tools and one or more sets of localizer images to determine the size and location of the spectroscopic volume. -
0 Indications for Use
PROstate Spectroscopy and imaging Exam (PROSE) is an image guided, clinical imaging and spectroscopy package which acquires high resolution anatomical images and volume localized, water/lipid suppressed hydrogen spectra and/or multi-voxel spectroscopic images of the prostate gland using an endo-rectal coil with phased-array coils. PROSE option can be used in conjunction with a Magnetic Resonance Scanner to permit non-invasive acquisition of high resolution images and spectral information about relative acquilities of metabolites of prostate gland that can be interpreted by a trained physician, and yield information that may assist in diagnosis of prostate diseases.
Image /page/0/Picture/14 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined in a stylized script. The letters are enclosed within a circular frame that has decorative swirls or flourishes around the perimeter. The logo is presented in black and white.
{1}------------------------------------------------
GE Medical Systems
Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circle. There are decorative swirls around the letters, adding a touch of elegance to the design.
page 2 of 2
o
PO Box 414, W-709 Milwaukee, WI 53201 USA
Comparison with Predicate
The Prostate Spectroscopy and Imaging Exam (PROSE) software option is substantially equivalent to existing scanning, processing and display Substantially oquiration to E Medical Systems PROBE software option (K930265). The PROBE software is modified to deliver improved spatial (1000200). The PRESS volume and extend the use of the PROBE coverage through the Prostate Gland. PROBE is largely limited to the Brain.
0 Summary of Studies
The Prostate Spectroscopy and Imaging Exam (PROSE) software option was evaluated to the IEC 601-2-33 International medical equipment safety was Uraldatod to the Resonance Systems. Evaluation testing was done to verify the performance of the option in the clinical environment.
0 Conclusions
It is the opinion of GE Medical Systems that the Prostate Spectroscopy and Imaging Exam (PROSE) software option does not result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 2 2 2001
Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical Systems P.O. Box 414, W-709 MILWAUKEE WI 53201
Re: K011604
Prostate Spectroscopy and Imaging Exam (PROSE) (Magnetic Resonance Diagnostic System) Dated: May 23, 2001 Received: May 24, 2001_ Regulatory Class: II 21 CFR 892.1000/Procode: 90 LNI
Dear Dr. Kroger:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the we have reviewed your Section J10(K) noutleation of misen to maxione) to legally marketed predicate devices device is substantally equivacing tor the nations by the enactment date of the Medical Device Amendments, or to devices marketed in interstale commerce provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, that like been reclassified in accordance with the provisions of the Act. The general controls provisions of the Act therefore, market the device to the general estates pro not manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject If your device is classified (SC above) into blance electing your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing I tue 21, I als 800 to 092. 11 substantian (QS) for Medical Devices: General regulation (21 CFR Part Practice requirements, as set four in the Quality Dresson on Drug Administration (FDA) will verify such assumptions. Failure 020) and may with the GMP regulation artion. In regulators action. In addition, FDA may publish further announcements to compy with the Unit Togallion ing Teach . Please note: this response to your premarket notification submission does not concenting your dorine in nie i bave under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of I his tetter whi anow your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro If you don't sponne arroo to your your e Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number the receiner of 00 00000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
Sincerely yours.
Nanci brondon
U.S.D.A.
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):_ Kol 160 4 _________________________________________________________________________________________________________________________________________
Device Name: PROstate Spectroscopy and imaging Exam (PROSE) Package for GE Signa Magnetic Resonance Scanners
Indications For Use:
PROstate Spectroscopy and imaging Exam (PROSE) is an image guided, clinical I KOState Spectroscopy and "magmigh acquires high resolution anatomical images and volume localized, water/lipid suppressed hydrogen spectra and/or multi-voxel spectroscopic images of the prostate gland using an endo-rectal coil with phased-array specifosophe images of the probate gainmention with a Magnetic Resonance Scanner to colls. PROSE option can be used in colljances. In spectral information about penint non-mvasive acquisition of ingil reast gland that can be interpreted by a relative concentrations of invacomes of prosumes of prostate diseases.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancyc Broglon
| (Division Sign-Off) |
|---|
| --------------------- |
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K011604 |
|---|---|
| --------------- | --------- |
Prescription Use______________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801.109) OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
N/A