K Number
DEN230067

Validate with FDA (Live)

Device Name
Chronos®
Manufacturer
Date Cleared
2024-08-28

(335 days)

Product Code
Regulation Number
880.6511
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Chronos® is indicated for use in a healthcare environment to achieve a high-level disinfection of surfaces of external, transvaginal, and transrectal ultrasound probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.

Device Description

The Germitec Chronos® device is intended to provide Ultraviolet C (UV-C) chemical-free highlevel disinfection (HLD) of ultrasound probes within an enclosed disinfection chamber. As defined in FDA Guidance - Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, found at https://www.fda.gov/media/80265/download, HLD is a lethal process utilizing a sterilant under less than sterilizing conditions. The process kills all forms of microbial life except for large numbers of bacterial spores. The subject device is intended for use in health care facilities by trained personnel. The Chronos® is an automated medical device consisting of a sealed disinfection chamber, incorporating multiple UV-C lamps and reflective walls. UV-C dose is identified as the only critical determinant of the effectiveness of the disinfection cycle.

The software is used to determine achievement of the disinfection total dose and acceptance of the automated HLD cycle. The total dose delivered to the surface of each ultrasound probe undergoing. the disinfection process is continuously monitored by two optical sensors, UV-C sensitive photodiodes. The control system continuously takes readings from the sensors, tracks progress of the ongoing disinfection cycle, and terminates the disinfection cycle when the cycle-controlling photodiodes indicate that the pre-defined UV-C dose has been achieved. Additionally, the device utilizes an independent cycle monitoring step through the use of a third UV-C photodiode. This independent probe is used to verify that the pre-defined UV-C HLD dose has been delivered.

Onto each ultrasound probe for which compatibility has been validated, a ring is placed onto the ultrasound cable. This ring is used to set the probe into the proper position within the chamber, as well as to include an identifier barcode. This barcode is read by the Chronos® software, and provides identification of both the appropriate disinfection dose as well as traceability for the ultrasound probe undergoing disinfection. The encoding of the barcode to each specific probe model and installation of the ring on each probe can only be performed by an authorized service representative from Germitec.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Important Note: The provided text describes the regulatory information, device description, and a summary of various bench and in-use studies. It details what performance characteristics were tested and what standards were met but does not provide specific numerical acceptance criteria for each test or the precise quantitative results achieved (e.g., exact log reduction numbers for all microorganisms, specific UV-C dose levels, or exact exposure times). It states that the device "can reliably and consistently achieve a > 6 log reduction" for simulated use and that "results support the determination of microbicidal potency." This indicates the device met the implicit criteria for HLD.


1. Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Performance Characteristic Tested (as outlined in Special Controls)Reported Device Performance and Confirmation of Meeting Criteria
Non-clinical performance testing:
(i) Resistance testing (hierarchy of resistance to UV radiation using a panel of clinically relevant and UV-resistant microorganisms)A hierarchy of UV-C resistance was determined. The most resistant microorganism tested was Aspergillus brasiliensis (Trichomaceae family). This indicates the device's efficacy was tested against a range of increasingly resistant microbes to establish its limits and inform subsequent testing. The implication is that the device is effective against organisms less resistant than A. brasiliensis.
(ii) Potency testing (against bacterial spores, vegetative bacteria, mycobacteria, yeast/molds (spores), spore forming molds and virus, demonstrating microbial log reduction appropriate for HLD)Microbicidal potency testing was conducted for sporicidal, bactericidal, fungicidal, mycobactericidal, and virucidal efficacy against a broad panel of organisms (B. subtilis, B. pumilus, C. sporogenes, D. radiodurans, S. enterica, S. aureus, P. aeruginosa, P. fungorum, M. terrae, M. bovis, M. avium, C. auris, C. albicans, T. mentagrophyte, A. brasiliensis, P. polonicum, Poliovirus Type 1, Adenovirus Type 5, Murine Norovirus, Vaginal Herpes, Polyomavirus, HPV16/HPV18). The documentation provided "support[s] the determination of microbicidal potency for the device," suggesting that the required log reductions for HLD were achieved for these categories.
(iii) Simulated use testing (under worst-case use and device conditions, including challenge microorganism, inoculum, device complexity, exposure dose, and time)Conducted over a range of worst-case ultrasound probes, demonstrating that the Chronos® can reliably and consistently achieve a > 6 log reduction of Bacillus subtilis. B. subtilis was identified as the most resistant bacterial spore and most appropriate for simulated use. Three worst-case locations on each probe (based on UV-C penetration difficulty) were identified, and a > 6 log microbial reduction was demonstrated for these challenging geometries. This confirms the device's ability to perform under stringent, real-world-simulated conditions to the HLD standard.
(iv) In-use testing (under real-world clinical use conditions)Performed on clinically used transvaginal, transrectal (with sheath and gel), and external (dry wipe only) probes. "In-Use testing demonstrated the achievement of HLD on these clinical used ultrasound probes." This validates the device's performance in a true clinical setting, accounting for factors like residual soil and clinical handling.
(v) Photobiological safety of lamps/systems (Performance testing)This refers to the UV-C Leak Test, which demonstrated that UV-C leakage remained < [b)( nW/cm2, indicating that the device prevents harmful UV-C exposure from escaping the chamber.
(vi) Safety features intended to prevent exposure (Performance testing)Covered by the UV-C Leak Test and Door Interlock Test. The Door Interlock Test was conducted in accordance with IEC 61010-1 requirements and implies successful operation preventing access during operation.
(vii) Long-term material compatibility of UV radiation on clinically relevant surfacesAn evaluation of the surface materials of representative ultrasound probes following exposure to the UV-C disinfection cycle was provided. While specifics on "long-term" are not detailed, it satisfies the requirement for material compatibility assessment.
Biocompatibility evaluationSafe residual levels of chemicals on medical device surfaces and/or gaseous byproducts in air.Biocompatibility risk evaluation and testing conducted in accordance with FDA 2020 Biocompatibility Guidance document (ISO 10993-1). Since UV-C does not remain on the probe, the focus was on the evaluation of probe surface materials after UV-C exposure. This indicates that no harmful residuals or byproducts affecting biocompatibility are left.
Software verification, validation, and hazard analysisFor any software components.Software testing conducted as per FDA guidance document, "Content of premarket submissions for Device Software Functions." This confirms that the software controlling the HLD cycle, monitoring, error reporting, and maintenance reminders underwent appropriate V&V and hazard analysis. Independent Cycle Monitoring System Performance Testing demonstrated accurate identification of successful/unsuccessful cycles under various conditions. Probe 3D Scanning Validation confirmed accurate capture of geometric complexity, and Optical Simulation Verification confirmed the Minimum Effective Dose (MED) at worst-case locations.
Performance data for EMC, thermal, and electrical safetyElectromagnetic compatibility (EMC), thermal and electrical safety of the device.EMC: Evaluated in accordance with IEC 61326-1 and IEC 60601-1-2. Electrical Safety: Completed in accordance with IEC 61010-1 and IEC 61010-2-040. Thermal Safety: Probe Surface Temperature testing conducted in accordance with IEC 60601-2-37, demonstrating the ability to control temperatures to prevent burn risk. Ventilation: Verified control of excessive heat and mitigation of contamination from outside air. "Testing has been successfully completed per the referenced standards."
Labeling RequirementsAll specified warnings, instructions, user procedures, hazard warnings, safety features, and material compatibility information.The labeling includes a product label and User Manual. It is states to be "in accordance with the special controls listed below," implying all specific labeling requirements (warnings, instructions, maintenance, preparation, UV hazard, safety features, material compatibility) have been met.

Study Details:

  1. Sample size used for the test set and the data provenance:

    • Simulated Use Testing: Not explicitly stated as a numerical sample size of probes per se, but conducted "over a range of worst-case ultrasound probes," with "Representative transvaginal and transrectal (mucous membrane contacting), and external (skin contacting) ultrasound probes" selected for their geometric complexity. For each probe, "three worst-case locations were identified."
    • In-Use Testing: Not explicitly stated as a numerical sample size. Used "Ultrasound probes used in a clinical setting." Transvaginal, transrectal, and external probes were used.
    • Potency Testing: The number of organisms challenged for each microbicidal activity is listed (e.g., 3 for sporicidal, 5 for bactericidal, etc.).
    • Data Provenance: The text states the sponsor is Germitec, located in Bordeaux, France. The studies appear to be retrospective bench studies and prospective in-use studies conducted by the sponsor to validate the device's performance. No specific information about varying countries of origin for the data is provided; it's implicit the data originates from Germitec's testing.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not explicitly state the number or qualifications of experts used to establish a "ground truth" for the test sets in the traditional sense of clinical interpretations.
    • For the performance testing, the "ground truth" is established by laboratory methods of microbial reduction, minimum effective UV-C dose, and adherence to engineering/safety standards. The experts involved would be microbiologists, engineers, and quality assurance personnel conducting these validated tests, following established protocols (e.g., those from the FDA and IEC).
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • This concept (e.g., 2+1, 3+1 adjudication for conflicting diagnoses) is typically applicable to studies involving human interpretation of medical images or clinical decisions where consensus is needed.
    • For the type of bench and in-use performance testing described here (microbial reduction, dose measurement, safety checks), an adjudication method is not applicable. The "ground truth" is determined by objective, measurable outcomes using validated laboratory and engineering techniques, not subjective interpretations requiring multiple expert opinions for consensus.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done.
    • This device is an automated disinfection chamber, not an AI-powered diagnostic or interpretive tool that assists human "readers." Therefore, a study assessing how human readers improve with AI assistance is not relevant to this product.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, the primary performance studies are essentially standalone. The device itself is an automated system.
    • The "performance testing" (potency, simulated use, optical verification, independent cycle monitoring, etc.) quantifies the device's ability to achieve HLD without human intervention during the disinfection cycle. The software controls the process and determines success or failure based on sensor readings and programmed parameters. While human operators initiate the cycle and load/unload probes, the disinfection itself is fully automated and "algorithm only."
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Microbiological Ground Truth: For the efficacy studies (potency, simulated use, in-use), the ground truth is established by quantitative microbiological reduction. This involves culturing and counting microorganisms before and after disinfection to determine log reduction, against pre-defined thresholds for High-Level Disinfection.
    • Physical/Engineering Ground Truth: For aspects like UV-C dose delivery, safety features (leakage, interlock), temperature control, and electrical/EMC, the ground truth is based on direct physical measurements and adherence to established engineering standards (e.g., IEC standards, pre-defined UV-C doses like the Minimum Effective Dose - MED).
  7. The sample size for the training set:

    • The document does not explicitly describe a "training set" in the context of machine learning for an AI algorithm that would be trained on a dataset.
    • While the device has software, its function is largely control, monitoring, and verification based on pre-programmed parameters and sensor feedback, not a learning algorithm that would require a distinct training set in the AI sense.
    • The "training" of the system would be its initial design, programming, and rigorous testing against engineering specifications and biological standards.
  8. How the ground truth for the training set was established:

    • As there is no explicitly defined "training set" for an AI model, this question is not directly applicable in the AI sense.
    • If one were to consider the "ground truth" that guided the design and programming of the device's control system and its internal thresholds (e.g., what constitutes an HLD cycle, the required UV-C dose), this ground truth would have been established through:
      • Microbiological research and industry standards for High-Level Disinfection.
      • Experimental determination of UV-C resistance of various microorganisms.
      • Optical simulations and empirical measurements to understand UV-C dose distribution within the chamber and on different probe geometries.
      • Adherence to relevant medical device standards (e.g., IEC standards for safety, FDA guidance for reprocessing).

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR GERMITEC CHRONOS

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Ultraviolet radiation disinfection chamber device. An ultraviolet radiation disinfection chamber device is intended to disinfect patient contacting medical devices using UV radiation after the device has been cleaned. Disinfection of the medical device is achieved within an enclosed chamber through the exposure to UV radiation.

NEW REGULATION NUMBER: 21 CFR 880.6511

CLASSIFICATION: Class II

PRODUCT CODE: SCS

BACKGROUND

DEVICE NAME: Germitec Chronos®

SUBMISSION NUMBER: DEN230067

DATE DE NOVO RECEIVED: October 3, 2023

SPONSOR INFORMATION:

Germitec 19 Rue Vauban 33000 Bordeaux, France

INDICATIONS FOR USE

Chronos® is indicated for use in a healthcare environment to achieve a high-level disinfection of surfaces of external, transvaginal, and transrectal ultrasound probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.

LIMITATIONS

The Germitec Chronos® is intended to be marketed for Over-The-Counter (OTC) use.

PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS

{1}------------------------------------------------

DEVICE DESCRIPTION

The Germitec Chronos® device is intended to provide Ultraviolet C (UV-C) chemical-free highlevel disinfection (HLD) of ultrasound probes within an enclosed disinfection chamber. As defined in FDA Guidance - Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, found at https://www.fda.gov/media/80265/download, HLD is a lethal process utilizing a sterilant under less than sterilizing conditions. The process kills all forms of microbial life except for large numbers of bacterial spores. The subject device is intended for use in health care facilities by trained personnel. The Chronos® is an automated medical device consisting of a sealed disinfection chamber, incorporating multiple UV-C lamps and reflective walls. UV-C dose is identified as the only critical determinant of the effectiveness of the disinfection cycle.

The software is used to determine achievement of the disinfection total dose and acceptance of the automated HLD cycle. The total dose delivered to the surface of each ultrasound probe undergoing. the disinfection process is continuously monitored by two optical sensors, UV-C sensitive photodiodes. The control system continuously takes readings from the sensors, tracks progress of the ongoing disinfection cycle, and terminates the disinfection cycle when the cycle-controlling photodiodes indicate that the pre-defined UV-C dose has been achieved. Additionally, the device utilizes an independent cycle monitoring step through the use of a third UV-C photodiode. This independent probe is used to verify that the pre-defined UV-C HLD dose has been delivered.

Image /page/1/Picture/3 description: The image shows a medical device with a user interface panel on the left side. On the right side, there is a probe holder and barcode reader opening, a disinfection chamber, and a door. The device appears to be designed for disinfecting medical probes and reading barcodes.

Pictures of Chronos® - Door Closed (left) and Door Open (right):

Onto each ultrasound probe for which compatibility has been validated, a ring is placed onto the ultrasound cable. This ring is used to set the probe into the proper position within the chamber, as well as to include an identifier barcode. This barcode is read by the Chronos® software, and

{2}------------------------------------------------

provides identification of both the appropriate disinfection dose as well as traceability for the ultrasound probe undergoing disinfection. The encoding of the barcode to each specific probe model and installation of the ring on each probe can only be performed by an authorized service representative from Germitec.

Image /page/2/Picture/1 description: The image shows a barcode reader and its components. The top left image shows the front view of the barcode reader, with labels pointing to the window, barcode sticker, ring, and probe cable. The top right image shows a side view of the barcode reader, with a label pointing to the barcode reader opening. The bottom image shows a close-up view of the ring receptacle and the ring, with labels pointing to each component.

Barcode ring location:

SUMMARY OF BENCH STUDIES

REPROCESSING, STERILITY AND SHELF-LIFE

The Germitec Chronos® is provided non-sterile and does not require additional reprocessing or sterilization. The device requires regular maintenance in order to continue to function as intended.

Annual preventive maintenance activities are defined for general system checks, device cleaning (internal and external components), calibration of optical sensors (UV-C photodiodes) and replacement of the ventilation filter. Additionally, the UV-C lamps must be replaced on a two-year interval, or every 100 hours of cumulative use, whichever comes first. The software provides an indication to the operator of UV-C lamp usable life remaining.

BIOCOMPATIBILITY

The Germitec Chronos® device is designed to provide liquid-free, high-level disinfection of reusable ultrasound probes using UV-C as the germicidal agent. Since UV-C does not remain on the ultrasound probe following completion of an HLD cycle, there is no indirect contact with the patient or personnel conducting HLD reprocessing. The sponsor has

{3}------------------------------------------------

provided an evaluation of the surface materials of representative ultrasound probes following exposure to the UV-C disinfection cycle. Biocompatibility risk evaluation and testing has been conducted in accordance with the FDA 2020 Biocompatibility Guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," found at https://www.fda.gov/media/142959/download

SOFTWARE & CYBERSECURITY

The Germitec Chronos® device incorporates a software-based control system used to control and monitor the delivery of UV-C dose required for high level disinfection of reusable ultrasound probes. The control system includes an ultrasound probe identifying system that captures key information about the probe, using a barcode reader, to link it to each disinfection. In addition to controlling the HLD cycle, the Chronos® device software controls system operation, ensures that the device is operating as expected, and communicates potential cycle errors to the user, as well as providing warnings about system status, condition, and reminders regarding preventive maintenance. The software testing was conducted as per FDA guidance document, "Content of premarket submissions for Device Software Functions, " found at https://www.fda.gov/media/153781/download.

Cybersecurity testing was conducted as per FDA guidance documents. "Cybersecurity in Medical Devices: Ouality System Considerations and Content of Premarket Submissions" found at https://www.fda.gov/media/119933/download, and "Postmarket Management of Cybersecurity in Medical Devices, " at https://www.fda.gov/media/95862/download.

ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY

Electromagnetic Compatibility (EMC) of the Chronos® device has been evaluated in accordance with IEC 61326-1, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements, and IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.

Electrical safety testing of the Chronos® device was completed in accordance with IEC 61010-1. Safety requirement for electrical equipment for measurement, control and laboratory use. Part 1: General requirements and IEC 61010-2-040, Safety requirements for electrical equipment for measurement, control and laboratory use. Part 2-040: Particular requirement for sterilizers and washer-disinfectors used to treat medical materials. Testing has been successfully completed per the referenced standards. including electrical, mechanical resistance to shock/impact, and mechanical safety testing.

PERFORMANCE TESTING - BENCH

The sponsor conducted the following performance tests to support that the device can achieve its intended use:

{4}------------------------------------------------

UV-C Hierarchy of Resistance .

Performance testing has been conducted to support a side-by-side evaluation of microorganism UV-C resistance. Each microorganism was tested at a series of varying UV-C doses in the Chronos® device. For the purposes of hierarchy testing, a panel of microorganisms was selected which represented a combination of organisms currently used as surrogates for other HLD processes or organisms reported to be particularly resistant to UV-C in literature. The panel of microorganisms included in UV-C resistance hierarchy testing included bacterial spores, bacteria, mycobacteria, yeast/molds (spores) and Trichomaceae family (spore forming molds). A hierarchy of UV-C resistance was determined for the Chronos® device. Based on the results obtained, the most resistant microorganism tested was Aspergillus brasiliensis, of the Trichomaceae family (spore forming molds).

Potency Testing .

The Chronos® device was tested for potency using the standard array of microbicidal testing to support device efficacy. Modified testing was developed to demonstrate microbicidal potency for the device as there exists no current standard method in place for UV-C disinfection processes. Potency testing was conducted to demonstrate sporicidal, bactericidal, fungicidal, mycobactericidal and virucidal efficacy of the device. Microbicidal potency testing included the following microorganisms.

Microbicidal ActivityOrganisms Challenged
SporicidalBacillus subtilis
Bacillus pumilus
Clostridium sporogenes
BactericidalDeinococcus radiodurans
Salmonella enterica
Staphylococcus aureus
Pseudomonas aeruginosa
Paraburkholderia fungorum
MycobactericidalMycobacterium terrae
Mycobacterium bovis
Mycobacterium avium
FungicidalCandida auris
Candida albicans
Trichophyton mentagrophyte (spore form)
VirucidalPoliovirus Type 1
Adenovirus Type 5
Murine Norovirus
Vaginal Herpes
Polyomavirus
HPV16 / HPV18

{5}------------------------------------------------

Sporicidal ActivityOrganisms Challenged
Trichomaceae family (spore forming molds)Aspergillus brasiliensisPenicillium polonicum (spore form)

Based on the testing documentation provided, the results support the determination of microbicidal potency for the device.

● Simulated Use Testing

Simulated use testing was conducted over a range of worst-case ultrasound probes to demonstrate that the Chronos® can reliably and consistently achieve a > 6 log reduction of the surrogate microorganism (Bacillus subtilis) following exposure to the disinfection cycle. Based on the completed hierarchy of resistance testing and the results of potency testing, Bacillus subtilis was identified as the most resistant bacterial spore and determined to be the most appropriate for use in Simulated Use testing. Representative transvaginal and transrectal (mucous membrane contacting), and external (skin contacting) ultrasound probes were selected based on their geometric complexity and difficulty to consistently achieve UV HLD.

For each probe, the three worst-case locations were identified based on difficulty for UV-C exposure to penetrate and therefore posed the greatest challenge to microbial inactivation. For each of these worst-case locations, a > 6 log microbial reduction was demonstrated for the worst-case ultrasound probes over a range of challenging geometries.

Optical Verification on Worst-Case Probes .

Worst-case ultrasound probes and associated "cold spot" locations on each probe, underwent optical verification studies to demonstrate that each location received the minimum effective dose (MED) of (b)(4) |mJ/cm2 UV-C exposure. All optical measurements at these geometrically worst-case locations were shown to receive the MED following exposure to the Chronos® device automated disinfection cycles.

In-Use Testing .

Ultrasound probes used in a clinical setting were subjected to the Chronos® automated high level disinfection cycles. Transvaginal and transrectal probes were used with single use probe protective sheath and gel. Transvaginal and transrectal probes were cleaned using cleaning solutions routinely used by hospital. External probes were used without a sheath and only underwent a dry wipe prior to exposure in the Chronos® system. In-Use testing demonstrated the achievement of HLD on these clinical used ultrasound probes.

{6}------------------------------------------------

UV-C Leak Test .

The sponsor has conducted performance testing to support to demonstrate that the chamber door prevents UV-C leakage from the disinfection chamber during a routine disinfection cycle. Documentation was provided to show that the amount of UV-C leakage remained < [b)( nW/cm2.

. Door Interlock Test

Door interlock performance testing was conducted in accordance with IEC 61010-1 requirements.

. Verification of ventilation contamination risk control

Performance of the Chronos® device ventilation was conducted to support control of excessive heat build up while mitigating potential for probe contamination from outside air. Testing provided documented evidence verifying the ability of the ventilation system to limit microbial contamination entering through the ventilation system.

. Probe Surface Temperature

Performance testing was conducted to demonstrate the ability of the Chronos® device to control temperatures within the chamber to ensure ultrasound probes do not pose burn risk to the operator or patient. Testing was conducted in accordance with IEC 60601-2-37.

. Independent Cycle Monitoring System Performance Testing

Performance testing was conducted to demonstrate the ability of the Chronos® device's independent cycle monitoring system to accurately to determine if a completed disinfection cycle is accurately identified as a success or failure. Testing was conducted to simulate conditions both where the target dose was reached and conditions where the target dose was not actually reached. In all test runs the independent monitoring system was able to identify successful and unsuccessful cycles.

Probe 3D Scanning Validation .

Validation testing was completed to demonstrate that the optical scanning process used is able to accurately capture the geometric complexity of ultrasound probes. Testing included a range of ultrasound probe geometries to challenge the system. Document provided demonstrated that the method used to acquire probe 3D scanned files was sufficiently robust to support the optical simulation verification process.

{7}------------------------------------------------

Optical Simulation Verification .

Optical Simulation verification was conducted to evaluate the interaction between the geometric variations of the ultrasound probes and the UV-C dose distribution when exposed to the Chronos® disinfection chamber. Validation of the simulation included the three worst-case probes identified for each dose group. This validation testing confirmed that the MED was systematically received at the three worst-case-location of the probe when exposed to the automated pre-defined UV-C disinfection cycle.

LABELING

The labeling consists of a product label and User Manual.

Labeling for this device is in accordance with the special controls listed below.

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of an ultraviolet radiation disinfection chamber device and the measures necessary to mitigate these risks.

Risks to HealthMitigation Measures
Patient cross-contamination due todevice failure or operator errorleading to inadequate microbialreductionNon-clinical performance testingSoftware verification, validation, and hazard analysisLabeling
Exposure to UV radiation, leadingto skin and eye damageNon-clinical performance testingBiocompatibility evaluationSoftware verification, validation, and hazard analysisLabeling
Incompatibility of the materialsintended to be disinfected leadingto reduced performance orpremature failureNon-clinical performance testingBiocompatibility analysisSoftware verification, validation, and hazard analysisLabeling
Electrical shock or thermalhazardsNon-clinical performance testingElectrical safety testingThermal safety testing
Interference with other devicesElectromagnetic compatibility testingElectrical safety testing

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the ultraviolet radiation disinfection chamber device is subject to the following special controls:

{8}------------------------------------------------

  • (1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
    • (i) Resistance testing must demonstrate the hierarchy of resistance to UV radiation using a panel of microorganisms that are clinically relevant and those known to be resistant to UV radiation;
    • (ii) Potency testing against bacterial spores, vegetative bacteria, mycobacteria, yeast/molds (spores), spore forming molds and virus, must demonstrate that the UV disinfection device is able to achieve microbial log reduction of a panel of clinically relevant and UV resistant microorganisms, as appropriate for the intended level of disinfection;
    • (iii) Simulated use testing must evaluate device performance under worst-case use and device conditions, including selection of appropriate challenge microorganism(s), microbial inoculum with residual soil, device complexity, exposure dose, and time, as appropriate for the intended level of disinfection;
    • (iv) In-use testing must evaluate device performance under real-world clinical use conditions:
    • Performance testing must demonstrate the photobiological safety of any lamps or (v) lamp systems;
    • Performance testing must validate safety features intended to prevent exposure; (vi) and
    • (vii) Performance testing must characterize the long-term material compatibility of UV radiation on clinically relevant surfaces and/or devices following repeated UV disinfection cycle exposure.
  • (2) Biocompatibility evaluation must demonstrate safe residual levels of chemicals on medical device surfaces and/or gaseous byproducts in air.
  • Software verification, validation, and hazard analysis must be performed for any (3) software components.
  • (4) Performance data must demonstrate the electromagnetic compatibility (EMC), thermal and electrical safety of the device.
  • (5) The labeling must include:
    • (i) Warnings and instructions to ensure the device is used for UV disinfection purposes, and is not to be used as a sterilant or as part of a sterilization process for medical devices:
    • (ii) User setup, operating and maintenance procedures;
    • Description of the required preparation of equipment for disinfection in the UV (iii) chamber device;
    • (iv) UV hazard warning labels:
    • Explanation of device safety features; and (v)
    • Information regarding material compatibility. (vi)

{9}------------------------------------------------

BENEFIT-RISK DETERMINATION

Benefit:

The sponsor demonstrated that the probable benefits include the following:

  • . Chemical-free, automated. High Level Disinfection (HLD) device for the surfaces of external, transvaginal, and transrectal probes. HLD reprocessing is performed following routine cleaning activities of the reprocessed ultrasound.
  • Reduced disinfection duration. .
  • . Traceability of the disinfection process.

Risks:

The risks of the device are based on nonclinical laboratory testing described above.

The main probable risk for the subject device is that a failure in the automated reprocessing cycle does not achieve the required microbial reduction, and high level disinfection is not achieved. Probes which have not been adequately disinfected may lead to patient crosscontamination, and an increased health risk associated with infections. Further, additional risks associated with UV exposure, electrical shock and interference with other devices, as well as thermal burns associated with probes following the disinfection cycle have also been identified. Additionally, cybersecurity risks have been identified for the device. These risks have all been mitigated through completion of performance testing, warnings, precautions, and special controls. Human factors testing was deemed not required based on an evaluation of potential high-risk use scenarios and the anticipated level of user interface needed to support routine operation of the device. In addition, In-Use testing was provided to demonstrate real world use conditions.

Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

Benefit/Risk Conclusion

In conclusion, given the testing provided and available information summarized above, for the following indication statement:

Chronos® is indicated for use in a healthcare environment to achieve a high-level disinfection of surfaces of external, transvaginal, and transrectal ultrasound probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.

The probable benefits outweigh the risks for the Germitec Chronos®. The device provides benefits, and the risks can be mitigated by the use of general controls and the identified special controls.

{10}------------------------------------------------

CONCLUSION

The De Novo request for the Germitec Chronos® is granted and the device is classified as follows:

Product Code: SCS Device Type: Ultraviolet radiation disinfection chamber device Regulation Number: 21 CFR 880.6511 Class: II

N/A