K Number
DEN230035

Validate with FDA (Live)

Device Name
DentalMonitoring
Date Cleared
2024-05-17

(379 days)

Product Code
Regulation Number
872.1770
Type
Direct
Panel
Dental
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

DentalMonitoring is a medical device software using image processing algorithms to analyze pictures of the oral cavity (hereinafter Scans). Scans are taken using the DM App, a smartphone, and the manufacturer's proprietary hardware products. Scans are taken by the patient, a non-healthcare professional, or a healthcare professional. The Scan is taken in healthcare facilities, such as a dental practice, or in a non-healthcare environment, such as the patient's own home. For some clinical parameters. DentalMonitoring requires a 3D Model. The product is designed to assist healthcare professionals in remotely monitoring orthodontic treatments and treatment progress. The results of DentalMonitoring are intended to be used as an aid in diagnosis and monitoring, not on a stand-alone basis for clinical decision-making. DentalMonitoring is indicated for use for patients over the age of 6 and reports results solely on permanent teeth.

DentalMonitoring can monitor the following clinical parameters:

  • . oral hygiene: dental plaque / food residue;
  • 트 soft tissue statement: gingival recession, black triangle;
  • . dental statement: closure of extraction space, tooth wear;
  • . alignment: closure of all anterior spaces; and
  • . dental occlusion:
    • o in 2D Monitoring: midline deviation, overbite/open bite, overjet
  • o in 3D Monitoring: canine class, midline deviation, overbite/open bite, overjet
    Additionally, the following clinical parameters specific to orthodontic treatment types or phases can be monitored using DentalMonitoring:
  • for aligner treatments: tracking (seat/unseat), attachment loss, button loss; .
  • . for braces: bracket debonding, tie loss, self-ligating clips, passive archwire and auxiliaries; and
  • for thermoformed retainers: tracking (seat/unseat). I
    Based on an initial 3D Model provided by a healthcare professional. DentalMonitoring can also provide healthcare professionals with 3D Models representative of the patient's dentition and treatment progress. This device is a prescription device and is not intended for over-the-counter use
Device Description

DentalMonitoring is a standalone software with embedded artificial intelligence comprising machine learning and locked neural network algorithms. The product is composed of these parts: the Dashboard, the DM App along with a DM Cheek Retractor and DM ScanBox, and the Data Analysis Platform.

The four steps are detailed below:

    1. Patient profile set-up: The healthcare professional sets up the patient profile through a web- based interface, Dashboard, accessible at www.dental-monitoring.com/doctor. During the patient profile set-up, the healthcare professional is prompted to set up a Protocol in order to select the clinical parameters they wish to monitor in accordance with the patient's treatment. During the set-up, the healthcare professional also determines the appropriate frequency at which the patient should use the DM App to take pictures of their intraoral cavity.
    1. Intraoral pictures acquisition: A set of pictures of the patient's intraoral cavity, hereinafter referred to as a Scan, is taken using the DM App along with a DM Cheek Retractor and DM ScanBox. The DM Cheek Retractor is an intraoral retraction device for the cheeks and lips to allow for sufficient space for image capture. The DM ScanBox is an extraoral holding apparatus for the smartphone that attaches to the DM Cheek Retractor. The DM App guides the user through the appropriate steps in order to obtain a complete Scan. This DM App is a mobile application installed on a smartphone; thus, the pictures are captured through the smartphone's main camera allowing the procedure to be totally non-invasive to the patient.
    1. Analysis of acquired intraoral pictures: The Scan is processed through the Data Analysis Platform. The Data Analysis Platform includes a technical processing phase and clinical processing phase, the latter being a clinical analysis to determine if any event has occurred within the clinical parameters the healthcare professional has set up to be monitored. The Data Analysis Platform uses AI comprising image processing algorithms and neural networks.
    1. Communication of results of the analysis performed through the Data Analysis Platform are communicated to the healthcare professional through the Dashboard on one hand, and to the patient through the DM App on the other hand. Results are shared with the healthcare professional in an exhaustive fashion, providing them with detailed information. As for the patient, the results are communicated in accordance with the rules defined by the healthcare professional in the Protocol applied to the patient.

DentalMonitoring enables HCPs to adapt the use of the product according to their needs depending on each patient's orthodontic treatment. DentalMonitoring comprises two types of monitoring: 2D Monitoring and 3D Monitoring. Some clinical parameters are specific to either 2D Monitoring, or 3D Monitoring.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the studies proving the device meets them, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are generally demonstrated through the clinical performance results (Sensitivity, Specificity, Slope, Intercept, and Bias) of each monitored parameter. The document doesn't explicitly state quantitative acceptance thresholds for each metric before the study, but rather presents the study results as adequate to demonstrate performance. The "Final Study Results" for each indication serve as the reported device performance.

Clinical ParameterMetricRequired Performance (Implicit Acceptance Criteria)Reported Device Performance (Result & 95% CI)
Dental plaque / food residue (2D)SensitivitySufficient for aid in diagnosis/monitoring82.5% [79.9%, 84.8%]
SpecificitySufficient for aid in diagnosis/monitoring83.2% [80.5%, 85.5%]
Gingival recession (2D)SensitivitySufficient for aid in diagnosis/monitoring79.9% [70.3%, 86.9%]
SpecificitySufficient for aid in diagnosis/monitoring97.4% [95.8%, 98.4%]
Black triangle (2D)SensitivitySufficient for aid in diagnosis/monitoring81.0% [73.9%, 86.5%]
SpecificitySufficient for aid in diagnosis/monitoring98.4% [96.9%, 99.2%]
Closure of extraction space (2D)SensitivitySufficient for aid in diagnosis/monitoring100.0% [94.2%, /] (Upper bound not provided)
SpecificitySufficient for aid in diagnosis/monitoring91.8% [87.2%, 94.9%]
Tooth wear (2D)SensitivitySufficient for aid in diagnosis/monitoring84.5% [74.1%, 91.2%]
SpecificitySufficient for aid in diagnosis/monitoring97.0% [94.9%, 98.2%]
Closure of all anterior spaces (2D)SensitivitySufficient for aid in diagnosis/monitoring98.3% [94.9%, 99.5%]
SpecificitySufficient for aid in diagnosis/monitoring83.3% [79.9%, 86.3%]
Tracking (seat/unseat) (2D)Sensitivity (presence/absence)Sufficient for aid in diagnosis/monitoring93.2% [91.3%, 94.7%]
Specificity (presence/absence)Sufficient for aid in diagnosis/monitoring86.2% [83.4%, 88.6%]
Sensitivity (level: slight/noticeable)Sufficient for aid in diagnosis/monitoring91.1% [85.9%, 94.5%]
Specificity (level: slight/noticeable)Sufficient for aid in diagnosis/monitoring90.5% [87.7%, 92.7%]
Attachment loss (2D)SensitivitySufficient for aid in diagnosis/monitoring98.2% [94.3%, 99.4%]
SpecificitySufficient for aid in diagnosis/monitoring100.0% [98.7%, /] (Upper bound not provided)
Button loss (2D)SensitivitySufficient for aid in diagnosis/monitoring98.4% [94.0%, 99.6%]
SpecificitySufficient for aid in diagnosis/monitoring99.0% [96.9%, 99.7%]
Bracket debonding (2D)SensitivitySufficient for aid in diagnosis/monitoring98.4% [93.8%, 99.6%]
SpecificitySufficient for aid in diagnosis/monitoring99.6% [98.5%, 99.9%]
Tie loss (2D)SensitivitySufficient for aid in diagnosis/monitoring93.3% [85.7%, 97.0%]
SpecificitySufficient for aid in diagnosis/monitoring96.5% [94.0%, 98.0%]
Self-ligating clips (2D)SensitivitySufficient for aid in diagnosis/monitoring91.1% [82.5%, 95.7%]
SpecificitySufficient for aid in diagnosis/monitoring88.3% [84.1%, 91.5%]
Canine class (3D)SlopeSufficient for aid in diagnosis/monitoring0.95 [0.92, 0.98]
InterceptSufficient for aid in diagnosis/monitoring-0.10 mm [-0.14, -0.04]
Bias at 0.00 mmSufficient for aid in diagnosis/monitoring-0.10 mm [-0.18, 0.00]
Bias at -2.00 mm +/- mmSufficient for aid in diagnosis/monitoring-0.3% [-4.9, 6.8]
Bias at 3.00 mm +/- mmSufficient for aid in diagnosis/monitoring-9.4% [-12.9, -5.3]
Bias at 5.00 mm +/- mmSufficient for aid in diagnosis/monitoring-7.9% [-11.0, -4.4]
Midline deviation (2D)SlopeSufficient for aid in diagnosis/monitoring0.93 [0.89, 0.97]
InterceptSufficient for aid in diagnosis/monitoring0.0 mm (-0.03) [0.0, 0.0] [-0.05, 0.01]
Bias at 0.0 mmSufficient for aid in diagnosis/monitoring0.0 mm (0.03) [0.0, 0.0] [-0.07, 0.03]
Bias at -2.0 mmSufficient for aid in diagnosis/monitoring5.3% [-0.8, 10.7]
Bias at 2.0 mmSufficient for aid in diagnosis/monitoring-8.9% [-13.6, -3.2]
Midline deviation (3D)SlopeSufficient for aid in diagnosis/monitoring0.98 [0.96, 1.00]
InterceptSufficient for aid in diagnosis/monitoring-0.01 mm [-0.02, 0.01]
Bias at 0.0 mmSufficient for aid in diagnosis/monitoring-0.01 mm [-0.03, 0.02]
Bias at -2.00 mmSufficient for aid in diagnosis/monitoring2.0 % [-0.5, 4.8]
Bias at 2.00 mmSufficient for aid in diagnosis/monitoring-2.7% [-5.2, -0.4]
Overbite / open bite (2D)SlopeSufficient for aid in diagnosis/monitoring0.95 [0.91, 0.99]
InterceptSufficient for aid in diagnosis/monitoring0.2 mm [0.1, 0.3]
Bias at 0.0 mmSufficient for aid in diagnosis/monitoring0.2 mm [0.0, 0.3]
Bias at 3.0 mmSufficient for aid in diagnosis/monitoring0.9% [-2.4, 3.7]
Bias at 5.0 mmSufficient for aid in diagnosis/monitoring-1.3 % [-4.1, 1.2]
Bias at 7.0 mmSufficient for aid in diagnosis/monitoring-2.3 % [-5.5, 0.5]
Overbite / open bite (3D)SlopeSufficient for aid in diagnosis/monitoring0.97 [0.96, 0.99]
InterceptSufficient for aid in diagnosis/monitoring0.01 mm [-0.02, 0.05]
Bias at 0.00 mmSufficient for aid in diagnosis/monitoring0.01 mm [-0.03, 0.05]
Bias at 3.00 mmSufficient for aid in diagnosis/monitoring-2.2% [-3.1, -1.5]
Bias at 5.00 mmSufficient for aid in diagnosis/monitoring-2.4% [-3.3, -1.4]
Bias at 7.00 mmSufficient for aid in diagnosis/monitoring-2.4% [-3.5, -1.3]
Overjet (2D)SlopeSufficient for aid in diagnosis/monitoring0.84 [0.78, 0.89]
InterceptSufficient for aid in diagnosis/monitoring-0.3 mm [0.1, 0.4]
Bias at 0.0 mmSufficient for aid in diagnosis/monitoring-0.3 mm [0.1, 0.5]
Bias at 3.0 mmSufficient for aid in diagnosis/monitoring-7.0% [-11.5 -3.5]
Bias at 5.0 mmSufficient for aid in diagnosis/monitoring-11.2% [-15.6, -7.1]
Bias at 7.0 mmSufficient for aid in diagnosis/monitoring-13.0% [-18.2, -8.3]
Bias at 9.0 mmSufficient for aid in diagnosis/monitoring-14.1% [-19.8, -8.8]
Overjet (3D)SlopeSufficient for aid in diagnosis/monitoring1.03 [1.01, 1.05]
InterceptSufficient for aid in diagnosis/monitoring0.14 mm [0.09, 0.19]
Bias at 0.00 mmSufficient for aid in diagnosis/monitoring0.14 mm [0.07, 0.23]
Bias at 3.00 mmSufficient for aid in diagnosis/monitoring7.2% [6.0, 8.9]
Bias at 5.00 mmSufficient for aid in diagnosis/monitoring5.4% [4.5, 6.7]
Bias at 7.00 mmSufficient for aid in diagnosis/monitoring4.7% [3.7, 6.0]
Bias at 9.00 mmSufficient for aid in diagnosis/monitoring4.3% [3.2, 5.6]
Passive archwire and auxiliaries (2D)SensitivitySufficient for aid in diagnosis/monitoring89.0% [84.9%, 92.1%]
SpecificitySufficient for aid in diagnosis/monitoring80.4% [75.1%, 84.8%]
Passive archwire and auxiliaries (3D)SensitivitySufficient for aid in diagnosis/monitoring90.4% [86.7%, 93.2%]
SpecificitySufficient for aid in diagnosis/monitoring85.5% [80.8%, 89.2%]
Updated 3D Model (Mean Absolute Error)MAESufficient similarity to reference 3D Model0.10 mm [0.093, 0.103]

2. Sample Sizes Used for the Test Set and Data Provenance

The studies are categorized as Retrospective and Prospective. The data provenance consistently states United States.

  • Retrospective Study (21-001):

    • Data Provenance: Retrospectively collected Scans from the US.
    • Sample Size (Test Set):
      • Dental plaque / food residue: 5,064 evaluations (derived from the 2x2 table)
      • Gingival recession: 870 evaluations
      • Black triangle: 1,137 evaluations
      • Closure of extraction space: 478 evaluations
      • Tooth wear: 1,066 evaluations
      • Closure of all anterior spaces: 713 evaluations
      • Tracking (seat/unseat): 3,323 evaluations
      • Attachment loss: 765 evaluations
      • Button loss: 659 evaluations
      • Bracket debonding: 659 evaluations
      • Tie loss: 653 evaluations
      • Self-ligating clips: 647 evaluations
  • Prospective Studies:

    • Occlusion Study (21-002):
      • Data Provenance: Prospectively collected from four sites located in the United States.
      • Sample Size (Test Set):
        • Canine class (3D): 215 patients, 297 results.
        • Midline deviation (2D): 277 patients, 277 results.
        • Midline deviation (3D): 291 patients, 294 results.
        • Overbite / open bite (2D): 285 patients, 301 results.
        • Overbite / open bite (3D): 287 patients, 298 results.
        • Overjet (2D): 208 patients, 245 results.
        • Overjet (3D): 263 patients, 292 results.
    • Archwire & Auxiliaries Study (21-003):
      • Data Provenance: Prospectively collected from six sites located in the United States.
      • Sample Size (Test Set):
        • Passive archwire and auxiliaries (2D & 3D): 730 evaluations.
        • Patient counts: 273 (2D) and 269 (3D).
    • Updated 3D Model Study (21-004):
      • Data Provenance: Prospectively collected from seven sites located in the United States.
      • Sample Size (Test Set): 250 patients, 536 results.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Retrospective Study (21-001):

    • Number of Experts: A panel of three independent orthodontists.
    • Qualifications: "orthodontists" (no specific years of experience or board certification mentioned, but implied clinical expertise in orthodontics).
    • Adjudication Method: "In case of non-dominant results, a consensus was established by the same group of three independent orthodontists."
  • Prospective Studies (21-002, 21-003, 21-004):

    • For occlusion (21-002), archwire & auxiliaries (21-003), and Updated 3D Model (21-004), the "Reference Method results were generated by measuring the occlusion parameters undergoing evaluation on 3D Models... Measurements were done manually and performed in three replicates using a CAD/CAM software." This implies that the ground truth was established through precise measurements using specialized software rather than direct human expert assessment on the dental images themselves for these specific parameters. The text doesn't specify if these measurements were performed by experts (e.g., orthodontists) or trained technicians, but the context of "Reference Method" implies a high standard of accuracy.

4. Adjudication Method for the Test Set

  • Retrospective Study (21-001): The initial assessment by three independent orthodontists was followed by consensus review for "non-dominant results" by the same group of three orthodontists.
  • Prospective Studies (21-002, 21-003, 21-004): The ground truth was established through manual measurements with CAD/CAM software, performed in three replicates. This is a technical measurement-based adjudication rather than human consensus on observations.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not reported in this document. The studies primarily focused on the standalone performance of the AI algorithm against a "Reference Method" (which was expert consensus or precise measurements). There is no mention of human readers with AI assistance being compared to human readers without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, extensive standalone performance testing of the algorithm only (without human-in-the-loop) was done. This is explicitly stated in the "AI/ML standalone testing" section and demonstrated by the clinical performance studies that compare the DentalMonitoring results directly to the "Reference Method" (ground truth). The device's indication for use also clarifies that it is "intended to be used as an aid in diagnosis and monitoring, not on a stand-alone basis for clinical decision-making," but the performance reported is of the algorithm itself.

7. The Type of Ground Truth Used

  • Retrospective Study (21-001): Expert Consensus by a panel of three independent orthodontists.
  • Prospective Studies (21-002, 21-003, 21-004): Precise Measurements using CAD/CAM Software on 3D models derived from intraoral scanners, performed in three replicates. This acts as a highly accurate, objective ground truth for quantitative parameters.

8. The Sample Size for the Training Set

The document does not explicitly state the sample size used for the training set. It only describes the "Clinical Investigation Program" which provides data to support performance, implying this is the test data. The "AI/ML standalone testing" section mentions that "Neural networks and algorithms were classified into three families... A unique protocol was established per family in order to perform the evaluation," but it doesn't detail the training phase or its dataset size.

9. How the Ground Truth for the Training Set Was Established

The document does not explicitly describe how the ground truth for the training set was established. The detailed descriptions of ground truth establishment are specifically for the test sets used in the retrospective and prospective clinical studies. It's generally assumed that similar rigorous methods are used for training data, but it's not detailed here.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR

DentalMonitoring

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Dental Image Analyzer. A dental image analyzer is a prescription home use device that uses software intended to collect and analyze patient-specific. optical camera-based, intraoral images. The analyses are provided to dental health care professionals as an aid to diagnosis of oral health conditions and/or to monitor treatment progress.

NEW REGULATION NUMBER: 872.1770

CLASSIFICATION: Class II

PRODUCT CODE: SBC

BACKGROUND

DEVICE NAME: DentalMonitoring

SUBMISSION NUMBER: DEN230035

DATE DE NOVO RECEIVED: 05/04/2023

SPONSOR INFORMATION:

Dental Monitoring SAS Isabelle Puraye 75 rue de Tocqueville Paris 75017

INDICATIONS FOR USE

DentalMonitoring is a medical device software using image processing algorithms to analyze pictures of the oral cavity (hereinafter Scans). Scans are taken using the DM App, a smartphone, and the manufacturer's proprietary hardware products. Scans are taken by the patient, a non-healthcare professional, or a healthcare professional. The Scan is taken in healthcare facilities, such as a dental practice, or in a non-healthcare environment, such as the patient's own home. For some clinical parameters. DentalMonitoring requires a 3D Model. The product is designed to assist healthcare professionals in remotely monitoring orthodontic treatments and treatment progress. The results of DentalMonitoring are intended to be used as an aid in diagnosis and monitoring, not on a stand-alone basis for clinical decision-making. DentalMonitoring is indicated for use for patients over the age of 6 and reports results solely on permanent teeth.

DentalMonitoring can monitor the following clinical parameters:

  • . oral hygiene: dental plaque / food residue;
  • 트 soft tissue statement: gingival recession, black triangle;
  • . dental statement: closure of extraction space, tooth wear;
  • . alignment: closure of all anterior spaces; and
  • . dental occlusion:
    • o in 2D Monitoring: midline deviation, overbite/open bite, overjet

{1}------------------------------------------------

  • o in 3D Monitoring: canine class, midline deviation, overbite/open bite, overjet
    Additionally, the following clinical parameters specific to orthodontic treatment types or phases can be monitored using DentalMonitoring:

  • for aligner treatments: tracking (seat/unseat), attachment loss, button loss; .

  • . for braces: bracket debonding, tie loss, self-ligating clips, passive archwire and auxiliaries; and

  • for thermoformed retainers: tracking (seat/unseat). I

Based on an initial 3D Model provided by a healthcare professional. DentalMonitoring can also provide healthcare professionals with 3D Models representative of the patient's dentition and treatment progress. This device is a prescription device and is not intended for over-the-counter use

LIMITATIONS

DentalMonitoring should not be used on patients under the age of 6.

Dental Monitoring should not be used by patients presenting systemic conditions affective connective tissues not allowing them to open/close their buccal cavity sufficiently to acquire acceptable Scans.

The following population should be assisted by a third party to perform Scans:

  • . Children up to 12;
  • . Adults or children visually impaired;
  • . Adults or children hearing impaired:
  • . Or any condition that might prevent the patient from adopting the right position to take a Scan.

DentalMonitoring is not intended to replace standard practices for treatment. Final clinical decisions remain the sole responsibility of the healthcare professional. In order to establish a diagnosis. further examinations are required according to the current standard of care, such as dental radiographs and/or tactile examinations with instrumentation.

DentalMonitoring results are limited only to elements visible in the input Scans. DentalMonitoring does not provide results on non-visible elements such as potential interproximal and lingual cavities. DentalMonitoring does not provide results on lingual surfaces.

PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS. PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

DentalMonitoring is a standalone software with embedded artificial intelligence comprising machine learning and locked neural network algorithms. The product is composed of these parts: the Dashboard, the DM App along with a DM Cheek Retractor and DM ScanBox, and the Data Analysis Platform.

{2}------------------------------------------------

Image /page/2/Figure/0 description: This image shows a diagram of a 4-step process. The first step is patient profile set-up, which is represented by a computer screen. The second step is intraoral pictures acquisition, which is represented by a smartphone. The third step is analysis of acquired intraoral pictures, which is represented by a tooth and magnifying glass. The fourth step is communication of results, which is represented by a smartphone.

The four steps are detailed below:

    1. Patient profile set-up: The healthcare professional sets up the patient profile through a web- based interface, Dashboard, accessible at www.dental-monitoring.com/doctor. During the patient profile set-up, the healthcare professional is prompted to set up a Protocol in order to select the clinical parameters they wish to monitor in accordance with the patient's treatment. During the set-up, the healthcare professional also determines the appropriate frequency at which the patient should use the DM App to take pictures of their intraoral cavity.
    1. Intraoral pictures acquisition: A set of pictures of the patient's intraoral cavity, hereinafter referred to as a Scan, is taken using the DM App along with a DM Cheek Retractor and DM ScanBox. The DM Cheek Retractor is an intraoral retraction device for the cheeks and lips to allow for sufficient space for image capture. The DM ScanBox is an extraoral holding apparatus for the smartphone that attaches to the DM Cheek Retractor. The DM App guides the user through the appropriate steps in order to obtain a complete Scan. This DM App is a mobile application installed on a smartphone; thus, the pictures are captured through the smartphone's main camera allowing the procedure to be totally non-invasive to the patient.

{3}------------------------------------------------

    1. Analysis of acquired intraoral pictures: The Scan is processed through the Data Analysis Platform. The Data Analysis Platform includes a technical processing phase and clinical processing phase, the latter being a clinical analysis to determine if any event has occurred within the clinical parameters the healthcare professional has set up to be monitored. The Data Analysis Platform uses AI comprising image processing algorithms and neural networks.
    1. Communication of results of the analysis performed through the Data Analysis Platform are communicated to the healthcare professional through the Dashboard on one hand, and to the patient through the DM App on the other hand. Results are shared with the healthcare professional in an exhaustive fashion, providing them with detailed information. As for the patient, the results are communicated in accordance with the rules defined by the healthcare professional in the Protocol applied to the patient.

DentalMonitoring enables HCPs to adapt the use of the product according to their needs depending on each patient's orthodontic treatment. DentalMonitoring comprises two types of monitoring: 2D Monitoring and 3D Monitoring. Some clinical parameters are specific to either 2D Monitoring, or 3D Monitoring.

SUMMARY OF NONCLINICAL/BENCH STUDIES

BIOCOMPATIBILITY/MATERIALS

The DM Cheek Retractor were evaluated per ISO 10993-1 (2020): "Biological evaluation of medical devices Part 1: evaluation and testing with a risk management" and ISO 7405-2018 - Dentistry - evaluation of biocompatibility of medical devices used in dentistry. The proposed device passed cytotoxicity tests according to ISO 10993-5, skin sensitization according to ISO 10993-10, and irritation or intracutaneous reactivity according to ISO 10993-23.

SOFTWARE

DentalMonitoring has a Modate Level of Concern (LOC). Appropriate software documentation for was provided in compliance with the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued in May 2005). Verification and validation (V&V) activities were performed and described at the unit and system level.

CYBER SECURITY

DentalMonitoring approach to cybersecurity aligns with FDA's 2020 guidance titled, "Guidance on Cybersecurity For Medical Devices" The device also conforms to the cybersecurity requirements identified in Section 524B to the FD&C Act.

PERFORMANCE TESTING - BENCH

AI/ML standalone testing

Standalone performance of each network and algorithm was evaluated. Neural networks and algorithms were classified into three families according to the outputs they provide. A unique protocol was established per family in order to perform the evaluation, with acceptance criteria adapted to each neural network and algorithm. DentalMonitoring comprises technical networks and algorithms as well as clinical neural networks and algorithms. The standalone testing showed acceptable performance of critical

{4}------------------------------------------------

features to demonstrate the device's ability to perform both it's clinical and technological functions.

SUMMARY OF CLINICAL INFORMATION

The Sponsor provided data from a Clinical Investigation Program to support the clinical performance of DentalMonitoring.

Study typeStudycodeStudy nameOverview of targetedindicationsList(indication;monitoring plan)couples targeted
Retrospective21-001RetrospectiveQualitative indicationswith Reference Methodcompatible withDentalMonitoring pictureset review(dental plaque / foodresidue: 2D)(gingival recession: 2D)(black triangle: 2D)(closure of extractionspace: 2D)(closure of all anteriorspaces: 2D)(tracking (seat/unseat):2D)(attachment loss: 2D)(button loss: 2D)(bracket debonding: 2D)(tie loss: 2D)(self-ligating clips: 2D)
Prospective21-002OcclusionOcclusion indications(canine class: 3D)(midline deviation: 2D)(midline deviation: 3D)(overbite/open bite: 2D)(overbite/open bite: 3D)(overjet: 2D)(overjet: 3D)
Prospective21-003Archwire&AuxiliariesPassive archwire andauxiliaries indication(passive archwire andauxiliaries: 2D)(passive archwire andauxiliaries: 3D)
Prospective21-004Updated 3DModelUpdated 3D Modelindication(Updated 3D Model: 3D)
DentalMonitoring Clinical Investigation Program
---------------------------------------------------

Retrospective study

DentalMonitoring evaluated the qualitative indications that were able to have the Reference Method used solely DentalMonitoring picture sets (= Scan). The study was performed using Scans retrospectively collected and involved a total of 15 US orthodontists.

The Reference Method results were generated as follow:

{5}------------------------------------------------

  • Scans were reviewed by a panel of three independent orthodontists with an assessment done per tooth ్ల / interdental space.
  • In case of non-dominant results, a consensus was established by the same group of three ਼ independent orthodontists. Dominant results and result of the consensus review were merged in order to generate the Reference Method initial outcome.
  • -Per indication, results presenting discrepancies between the Reference Method and the Candidate Method, i.e., False Positive results and False Negative results, were reviewed by one of the orthodontists among the 15 that took part in the study.

Data was collected in order to qualify the patient demographics.

{6}------------------------------------------------

Results per indication

Dental plaque / food residue

Characteristic# (%)
Age group
12- 22 years of age367 (49.5%)
≥ 22 years of age4 (0.6%)
≤12 years old369 (49.8%)
Unknown1 (0.1%)
Location
Central420 (56.7%)
East Coast209 (28.2%)
US - Other5 (0.7%)
West Coast60 (8.1%)
Western30 (4.0%)
Out of US17 (2.3%)
Treatment type
Aligners415 (56.0%)
Brackets - Ceramic9 (1.2%)
Brackets - Self-ligating128 (17.3%)
Brackets - Traditional135 (18.2%)
Mixed32 (4.3%)
None5 (0.7%)
Retainers17 (2.3%)

Final comparison between DentalMonitoring and Reference

Final comparison between DentalMonitoring and ReferenceReference Result
Positive (Noticeable or Slight)NegativeTotal
Positive (Noticeable orSlight)1,8743562,230
DMResultNegative3622,4722,834
Total2,2362,8285,064

Final study results for dental plaque / food residue

Result95% CI LowerBound95% CI UpperBound
Sensitivity82.5%79.9%84.8%
Specificity83.2%80.5%85.5%

Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, OnePlus, Huawei, Xiaomi, Essential Phone

{7}------------------------------------------------

Gingival recession

Characteristic# (%)
Age group
12- 22 years of age205 (51.3%)
≥ 22 years of age194 (48.5%)
≤12 years old1 (0.2%)
Location
Central236 (59%)
East Coast96 (24%)
West Coast30 (7.5%)
US - Other5 (1.2%)
Out of US12 (3%)
Western21 (5.3%)
Treatment type
Aligners205 (51.2%)
Retainers17 (4.2%)
Brackets - Ceramic7 (1.8%)
Brackets - Self-ligating68 (17%)
Brackets - Traditional82 (20.5%)
Mixed treatment20 (5%)
No treatment1 (0.3%)

Final comparison between DentalMonitoring and Reference

Reference Result
PositiveNegativeTotal
DMResultPositive771895
DMResultNegative19756775
DMResultTotal96774870

Final study results for gingival recession

Result95% CI LowerBound95% CI UpperBound
Sensitivity79.9%70.3%86.9%
Specificity97.4%95.8%98.4%

Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, Huawei, Essential Phone, Kyocera, OnePlus, and Wiko.

{8}------------------------------------------------

Black triangle

Characteristic# (%)
Age group
12 - 22 years of age233 (62%)
≥ 22 years of age140 (37.2%)
≤ 12 years old3 (0.8%)
Location
Central231 (61.4%)
East Coast97 (25.8%)
West Coast23 (6.1%)
Out of US7 (1.9%)
Western14 (3.7%)
US-Other4 (1.1%)
Treatment type
Aligners166 (44.2%)
Retainers12 (3.2%)
Brackets - Ceramic9 (2.4%)
Brackets - Self-ligating79 (21%)
Brackets - Traditional87 (23.1%)
Mixed treatment21 (5.6%)
No treatment2 (0.5%)

Final comparison between DentalMonitoring and Reference

Reference Result
DM ResultPositiveNegativeTotal
Positive17912191
DMResultNegative40906946
Total2199181,137

Final study results for black triangle

Result95% CI LowerBound95% CI UpperBound
Sensitivity81.0%73.9%86.5%
Specificity98.4%96.9%99.2%

Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, OnePlus, Huawei, and Xiaomi.

{9}------------------------------------------------

Closure of extraction space

Characteristic# (%)
Age group
12 -22 years of age105 (41.3%)
≥ 22 years of age5 (2.0%)
≤ 12 years old142 (55.9%)
Unknown2 (0.8%)
Location
Central137 (53.9%)
East Coast67 (26.4%)
US - Other6 (2.4%)
West Coast25 (9.8%)
Western11 (4.3%)
out of US8 (3.2%)
Treatment type
Aligners125 (49.2%)
Brackets - Ceramic3 (1.2%)
Brackets - Self-ligating55 (21.7%)
Brackets - Traditional44 (17.3%)
Mixed treatment16 (6.3%)
None1 (0.4%)
Retainers10 (3.9%)

Final comparison between DentalMonitoring and Reference

Reference Result
PositiveNegativeTotal
DMResultPositive19824222
Negative0256256
Total198280478

Final study results for closure of extraction space

Result95% CI Lower Bound95% CI Upper Bound
Sensitivity100.0%94.2%/
Specificity91.8%87.2%94.9%

Scans were taken using the following phone manufacturers: Apple, Samsung, LG, Google, Motorola, Essential Phone, Huawei, and Wiko.

{10}------------------------------------------------

Tooth wear

Characteristic# (%)
Age group
12 - 22 years of age200 (54.8%)
≥ 22 years of age164 (44.9%)
≤ 12 years old1 (0.3%)
Location
Central232 (63.6%)
East Coast83 (22.7%)
West Coast22 (6.0%)
Western14 (3.8%)
Out of US11 (3.0%)
US-Other3 (0.8%)
Treatment type
Aligners162 (44.4%)
Brackets - Ceramic5 (1.4%)
Brackets - Self-ligating74 (20.3%)
Brackets - Traditional88 (24.1%)
Mixed1 (0.3%)
Mixed treatment22 (6%)
None1 (0.3%)
Retainers12 (3.3%)

Final comparison between DentalMonitoring and Reference

Reference Result
PositiveNegativeTotal
DMResultPositive8921110
Negative13943956
Total1029641,066

Final study results for tooth wear

Result95% CILower Bound95% CIUpper Bound
Sensitivity84.5%74.1%91.2%
Specificity97.0%94.9%98.2%

Scans were taken using the following phone manufacturers: Apple, Samsung, LG, Google, Motorola, Huawei, and Essential Phone.

{11}------------------------------------------------

Closure of all anterior spaces

Characteristic# (%)
Age group
12 - 22 years of age297 (57.5%)
> 22 years of age213 (41.3%)
≤ 12 years old4 (0.8%)
Unknown2 (0.4%)
Location
Central307 (59.5%)
East Coast131 (25.4%)
West Coast38 (7.4%)
Out of US10 (1.9%)
US - Other4 (0.8%)
Western26 (5.0%)
Treatment type
Aligners248 (48.1%)
Brackets - Ceramic6 (1.2%)
Brackets - Self-ligating114 (22.1%)
Brackets - Traditional106 (20.5%)
Mixed2 (0.4%)
Mixed treatment21 (4.1%)
None6 (1.2%)
Retainers13 (2.5%)

Final comparison between DentalMonitoring and Reference

Reference Result
PositiveNegativeTotal
Positive18088268
DMResultNegative3442445
Total183530713

Final study results for anterior space closure

Result95% CI LowerBound95% CI UpperBound
Sensitivity98.3%94.9%99.5%
Specificity83.3%79.9%86.3%

Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, Huawei, T-Mobile, Essential Phone, and Xiaomi.

{12}------------------------------------------------

Tracking (seat/unseat)

Characteristic# (%)
[12; 22[ years of age279 (44.8%)
≥ 22 years of age332 (53.3%)
≤ 12 years old9 (1.4%)
Unknown3 (0.5%)
Location
Central286 (45.9%)
East Coast192 (30.8%)
out of US17 (2.7%)
US - Other8 (1.3%)
West Coast85 (13.7%)
Western35 (5.6%)
Treatment type
Aligners581 (93.3%)
Brackets - Traditional4 (0.6%)
Mixed1 (0.2%)
Mixed treatment9 (1.4%)
Retainers28 (4.5%)

Clinical performances are expressed as two different Sensitivities and Specificities since the parameter outputs three levels.

  • Sensitivity and Specificity of the product in its ability to detect people with the evaluated 미 condition from people without; and
  • Sensitivity and Specificity of the product in its ability to discriminate the evaluated condition l at a slight level from a noticeable level

{13}------------------------------------------------

Reference Result
Positive (Noticeableor Slight)NegativeTotal
Positive (Noticeable orSlight)1,3601651,525
DMResultNegative1011,6971,798
Total1,4611,8623,323

Detection of patients with the evaluated condition (presence/absence)

Result95% CI Lower Bound95% CI Upper Bound
Sensitivity93.2%91.3%94.7%
Specificity86.2%83.4%88.6%

Detection of patients with the level of the condition (slight/noticeable)

Reference Result
NoticeableSlightTotal
DMResultNoticeable19399292
Slight181,0501,068
Total2111,1491,360
Result95% CI LowerBound95% CI UpperBound
Sensitivity91.1%85.9%94.5%
Specificity90.5%87.7%92.7%

Scans were taken using the following phone manufacturers: Apple, Samsung, Google, LG, Motorola, OnePlus, Huawei, Essential Phone, Wiko, and Xiaomi.

{14}------------------------------------------------

Attachment loss

Characteristic# (%)
Age group
]12; 22[ years of age146 (41.9%)
> 22 years of age200 (57.5%)
≤ 12 years old1 (0.3%)
Unknown1 (0.3%)
Location
Central171 (49.1%)
East Coast109 (31.3%)
West Coast39 (11.2%)
US - Other4 (1.2%)
Out of US8 (2.3%)
Western17 (4.9%)
Treatment type
Aligners330 (94.8%)
Retainers8 (2.3%)
Brackets - Traditional4 (1.2%)
Mixed treatment6 (1.7%)

Final comparison between DentalMonitoring and Reference

Reference Result
PositiveNegativeTotal
DMResultPositive1830183
Negative3579582
Total186579765

Final study results for attachment loss

Result95% CI LowerBound95% CI UpperBound
Sensitivity98.2%94.3%99.4%
Specificity100.0%98.7%

Scans were taken using the following phone manufacturers: Apple, Samsung, Google, Motorola, LG, Xiaomi, OnePlus, and Essential Phone.

{15}------------------------------------------------

Button loss

Characteristic# (%)
Age group
]12; 22[ years of age200 (52.5%)
> 22 years of age158 (41.5%)
≤ 12 years old20 (5.2%)
Unknown3 (0.8%)
Location
Central138 (36.2%)
East Coast125 (32.8%)
West Coast92 (24.1%)
Out of US4 (1%)
Western16 (4.2%)
US - Other6 (1.6%)
Treatment type
Aligners370 (97.1%)
Retainers1 (0.3%)
Brackets - Ceramic1 (0.3%)
Brackets - Traditional3 (0.8%)
Mixed treatment6 (1.6%)

Final comparison between DentalMonitoring and Reference

Reference Result
PositiveNegativeTotal
DMResultPositive1275132
Negative2525527
Total129530659

Final study results for button loss

Result95% CI LowerBound95% CI UpperBound
Sensitivity98.4%94.0%99.6%
Specificity99.0%96.9%99.7%

Scans were taken using the following phone manufacturers: Apple, Samsung, LG, Motorola, Google, and Huawei.

{16}------------------------------------------------

Bracket debonding

Characteristic# (%)
Age group
]12; 22[ years of age270 (71.3%)
≥ 22 years of age107 (28.2%)
≤ 12 years old2 (0.5%)
Location
Central291 (76.8%)
East Coast60 (15.8%)
West Coast16 (4.2%)
Out of US1 (0.3%)
Western11 (2.9%)
Treatment type
Aligners1 (0.3%)
Brackets - Ceramic10 (2.5%)
Brackets - Self-ligating171 (45.1%)
Brackets - Traditional150 (39.6%)
Mixed treatment45 (11.9%)
None1 (0.3%)
Retainers1 (0.3%)

Final comparison between DentalMonitoring and Reference

Reference Result
PositiveNegativeTotal
DMResultPositive1272129
Negative2528530
Total129530659

Final study results for bracket debonding

Result95% CI LowerBound95% CI UpperBound
Sensitivity98.4%93.8%99.6%
Specificity99.6%98.5%99.9%

Scans were taken using the following phone manufacturers: Apple, Samsung, Google, LG, Motorola, One Plus.

{17}------------------------------------------------

Tie loss

Characteristic# (%)
Age group
]12; 22[ years of age240 (77.4%)
$\geq$ 22 years of age66 (21.3%)
$\leq$ 12 years old4 (1.3%)
Location
Central255 (82.2%)
East Coast39 (12.6%)
West Coast10 (3.2%)
Western3 (1.0%)
Out of US3 (1.0%)
Treatment type
Aligners1 (0.3%)
Brackets - Ceramic13 (4.2%)
Brackets - Self-ligating5 (1.6%)
Brackets - Traditional267 (86.2%)
Mixed1 (0.3%)
Mixed treatment23 (7.4%)

Final comparison between DentalMonitoring and Reference

Reference Result
PositiveNegativeTotal
DMResultPositive11115126
Negative7520527
Total118535653

Final study results for tie loss

Result95% CI LowerBound95% CI UpperBound
Sensitivity93.3%85.7%97.0%
Specificity96.5%94.0%98.0%

Scans were taken using the following phone manufacturers: Apple, Samsung, Motorola, LG, Google.

{18}------------------------------------------------

Self-ligating clips

Characteristic#(%)
Age group
]12; 22[ years of age217(66.8%)
≥ 22 years of age103(31.7%)
≤ 12 years old5(1.5%)
Location
Central225(69.3%)
East Coast69(21.2%)
Western16(4.9%)
West Coast12(3.7%)
Out of US3(0.9%)
Treatment type
Brackets - Ceramic5(1.5%)
Brackets - Self-ligating262(80.6%)
Brackets - Traditional21(6.5%)
Mixed2(0.6%)
Mixed treatment35(10.8%)

Final comparison between DentalMonitoring and Reference

Reference Result
PositiveNegativeTotal
DMResultPositive7148119
DMResultNegative7521528
DMResultTotal78569647

Final study results for self-ligating clip

Result95% CI LowerBound95% CI UpperBound
Sensitivity91.1%82.5%95.7%
Specificity88.3%84.1%91.5%

Scans were taken using the following phone manufacturers: Apple, Samsung, Google, LG, Motorola, Huawei, Kyocera, and OnePlus.

{19}------------------------------------------------

Occlusion study

DentalMonitoring evaluated the occlusion indications using 3D Models acquired with intraoral scanners as input data to the Reference Method. The study was performed using data collected prospectively. The study involved four sites located in the United States: three sites for patient enrollment and one site for generation of the Reference Method results.

The Reference Method results were generated by measuring the occlusion parameters undergoing evaluation on 3D Models of the patients enrolled in the study. Measurements were done manually and performed in three replicates using a CAD/CAM software.

Analysis was performed according to "Measurement Procedure Comparison and Bias Estimation Using Patient Samples; Clinical and Laboratory Standards Institute guideline", EP09c - Third Edition, 2018. Analysis comprised using a :Passing-Bablok regression analysis, including plots of Candidate Method verses Reference Method to calculate the slope, intercept, and correlation coefficient of the plot. Bias at levels of interest per study group were computed at the levels of interest and confidence intervals were calculated with the 95% Bootstrapping.

Data was collected in order to qualify the patient demographics.

{20}------------------------------------------------

Results per indication

Canine class - 3D Monitoring

A total of 215 patients rendering 297 results were included in the analysis. One patient could render multiple results corresponding to left canine class and right canine class.

Characteristic# (%)
Gender
Male153 (71.2%)
Female62 (28.8%)
Age group
[6; 12] years of age23 (10.7%)
]12; 22[ years of age75 (34.9%)
≥ 22 years of age117 (54.4%)
Race*
American Indian or Alaska Native13 (6.0%)
Asian10 (4.7%)
Black or African American12 (5.6%)
Native Hawaiian or Other Pacific Islander0 (0.0%)
White168 (78.1%)
Unknown race14 (6.5%)
Ethnicity
Spanish origin32 (14.9%)
Not Hispanic or Latino183 (85.1%)
Income of the household (annual)
Less than $19,9999 (4.2%)
$20,000-$37,99911 (5.1%)
$38,000-$44,9993 (1.4%)
$45,000-$54,9997 (3.3%)
$55,000-$64,99914 (6.5%)
$65,000-$74,99911 (5.1%)
$75,000-$99,99925 (11.6%)
$100,000 or more69 (32.1%)
Unknown66 (30.7%)
Level of education
< Highschool Diploma84 (39.1%)
Highschool Diploma or Equivalent38 (17.6%)
Associates Degree25 (11.6%)
Bachelors Degree44 (20.5%)
Masters Degree20 (9.3%)
Doctoral Degree4 (1.9%)

{21}------------------------------------------------

Slope0.95
[0.92, 0.98]
Intercept-0.10 mm
[-0.14, -0.04]*
Bias at 0.00 mm-0.10 mm
[-0.18, 0.00]*
Bias at -2.00 mm-0.3%
[-4.9, 6.8]
Bias at 3.00 mm-9.4%
[-12.9, -5.3]
Bias at 5.00 mm-7.9%
[-11.0, -4.4]

*Of note, the difference between the two 95% C1 is due to the fact that the bootstrap is used for bias C1 calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.

{22}------------------------------------------------

Midline deviation - 2D Monitoring

A total of 277 patients rendering 277 results were included in the analysis. One patient rendered a unique result.

Characteristic# (%)
Gender
Male191 (69.0%)
Female86 (31.0%)
Age group
[6; 12] years of age36 (13.0%)
]12; 22[ years of age94 (33.9%)
≥ 22 years of age147 (53.1%)
Race*
American Indian or Alaska Native12 (4.3%)
Asian11 (4.0%)
Black or African American26 (9.4%)
Native Hawaiian or Other Pacific Islander0 (0.0%)
White215 (77.6%)
Unknown race16 (5.7%)
Ethnicity
Spanish origin37 (13.4%)
Not Hispanic or Latino240 (86.6%)
Income of the household (annual)
Less than $19,99912 (4.3%)
$20,000-$37,99914 (5.1%)
$38,000-$44,9996 (2.2%)
$45,000-$54,9997 (2.5%)
$55,000-$64,99914 (5.1%)
$65,000-$74,99918 (6.5%)
$75,000-$99,99929 (10.5%)
$100,000 or more100 (36.1%)
Unknown77 (27.7%)
Level of education
< Highschool Diploma100 (36.1%)
Highschool Diploma or Equivalent54 (19.5%)
Associates Degree31 (11.2%)
Bachelors Degree58 (20.9%)
Masters Degree29 (10.5%)
Doctoral Degree5 (1.8%)

{23}------------------------------------------------

Slope0.93
[0.89, 0.97]
Intercept0.0 mm (-0.03)
[0.0, 0.0] [-0.05, 0.01]*
Bias at 0.0 mm0.0 mm (0.03)
[0.0, 0.0] [-0.07, 0.03]*
Bias at -2.0 mm5.3%
[-0.8, 10.7]
Bias at 2.0 mm-8.9%
[-13.6, -3.2]

*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard C1 calculation linked to Passing-Bablok regression that is used.

{24}------------------------------------------------

Midline deviation - 3D Monitoring

A total of 291 patients rendering 294 results were included in the analysis. One patient rendered a unique result.

Characteristic# (%)
Gender
Male200 (68.7%)
Female91 (31.3%)
Age group
[6; 12] years of age47 (16.2%)
]12; 22[ years of age98 (33.6%)
≥ 22 years of age146 (50.2%)
Race*
American Indian or Alaska Native12 (4.1%)
Asian6 (2.1%)
Black or African American26 (8.9%)
Native Hawaiian or Other Pacific Islander0 (0.0%)
White224 (77.0%)
Unknown race30 (10.3%)
Ethnicity
Spanish origin38 (13.1%)
Not Hispanic or Latino253 (86.9%)
Income of the household (annual)
Less than $19,99912 (4.1%)
$20,000-$37,99916 (5.5%)
$38,000-$44,9996 (2.1%)
$45,000-$54,9997 (2.4%)
$55,000-$64,99915 (5.2%)
$65,000-$74,99918 (6.2%)
$75,000-$99,99933 (11.3%)
$100,000 or more101 (34.7%)
Unknown83 (28.5%)
Level of education
< Highschool Diploma109 (37.5%)
Highschool Diploma or Equivalent55 (18.8%)
Associates Degree36 (12.4%)
Bachelors Degree61 (21.0%)
Masters Degree28 (9.6%)
Doctoral Degree2 (0.7%)

{25}------------------------------------------------

Slope0.98[0.96, 1.00]
Intercept-0.01 mm[-0.02, 0.01]*
Bias at 0.0 mm-0.01 mm[-0.03, 0.02]*
Bias at -2.00 mm2.0 %[-0.5, 4.8]
Bias at 2.00 mm-2.7%[-5.2, -0.4]

*Of note, the difference between the two 95% Cl is due to the fact that the bootstrap is used for bias Cl calculation, whereas for the intercept and slope it is the standard C1 calculation linked to Passing-Bablok regression that is used. Interpretation of bias signs

{26}------------------------------------------------

Overbite / open bite - 2D Monitoring

A total of 285 patients rendering 301 results were included in the analysis. One patient could render multiple results corresponding to both central incisor couples.

Characteristic# (%)
Gender
Male194 (68.1%)
Female91 (31.9%)
Age group
[6; 12] years of age36 (12.7%)
]12; 22[ years of age87 (30.5%)
≥ 22 years of age162 (56.8%)
Race*
American Indian or Alaska Native10 (3.5%)
Asian11 (3.9%)
Black or African American32 (11.2%)
Native Hawaiian or Other Pacific Islander0 (0.0%)
White216 (75.8%)
Unknown race21 (7.4%)
Ethnicity
Spanish origin36 (12.6%)
Not Hispanic or Latino249 (87.4%)
Income of the household (annual)
Less than $19,99911 (3.9%)
$20,000-$37,99916 (5.5%)
$38,000-$44,9998 (2.8%)
$45,000-$54,9997 (2.5%)
$55,000-$64,99912 (4.2%)
$65,000-$74,99919 (6.7%)
$75,000-$99,99933 (11.6%)
$100,000 or more103 (36.1%)
Unknown76 (26.7%)
Level of education
< Highschool Diploma102 (35.8%)
Highschool Diploma or Equivalent51 (17.9%)
Associates Degree31 (10.9%)
Bachelors Degree65 (22.8%)
Masters Degree32 (11.2%)
Doctoral Degree4 (1.4%)

{27}------------------------------------------------

Slope0.9595% CI = [0.91, 0.99]
Intercept0.2 mm95% CI = [0.1, 0.3]*
Bias at 0.0 mm0.2 mm95% CI = [0.0, 0.3]*
Bias at 3.0 mm0.9%95% CI = [-2.4, 3.7]
Bias at 5.0 mm-1.3 %95% CI = [-4.1, 1.2]
Bias at 7.0 mm-2.3 %95% CI = [-5.5, 0.5]

*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.

{28}------------------------------------------------

Overbite / open bite - 3D Monitoring

A total of 287 patients rendering 298 results were included in the analysis. One patient could render multiple results corresponding to both central incisor couples.

Characteristic# (%)
Gender
Male90 (31.4%)
Female197 (68.6%)
Age group
[6; 12] years of age42 (14.6%)
]12; 22[ years of age89 (31%)
≥ 22 years of age156 (54.4%)
Race*
American Indian or Alaska Native10 (3.5%)
Asian10 (3.5%)
Black or African American31 (10.8%)
Native Hawaiian or Other Pacific Islander0 (0.0%)
White217 (75.6%)
Unknown race26 (9.9%)
Ethnicity
Spanish origin37 (12.9%)
Not Hispanic or Latino250 (87.1%)
Income of the household (annual)
Less than $19,99911 (3.8%)
$20,000-$37,99917 (5.9%)
$38,000-$44,9998 (2.8%)
$45,000-$54,9997 (2.4%)
$55,000-$64,99912 (4.2%)
$65,000-$74,99920 (7.0%)
$75,000-$99,99932 (11.2%)
$100,000 or more102 (35.5%)
Unknown78 (27.2%)
Level of education
< High School Diploma101 (35.2%)
High School Diploma or Equivalent51 (17.8%)
Associates Degree33 (11.5%)
Bachelors Degree71 (24.7%)
Masters Degree29 (10.1%)
Doctoral Degree2 (0.7%)

{29}------------------------------------------------

Slope0.9795% CI = [0.96, 0.99]
Intercept0.01 mm95% CI= [-0.02, 0.05]*
Bias at 0.00 mm0.01 mm95% CI = [-0.03, 0.05]*
Bias at 3.00 mm-2.2%95% CI = [-3.1, -1.5]
Bias at 5.00 mm-2.4%95% CI = [-3.3, -1.4]
Bias at 7.00 mm-2.4%95% CI = [-3.5, -1.3]

*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.

{30}------------------------------------------------

Overjet - 2D Monitoring

A total of 208 patients rendering 245 results were included in the analysis. One patient could render multiple results corresponding to both central incisor couples.

Characteristic# (%)
Gender
Male69 (33.2%)
Female139 (66.8%)
Age group
[6; 12] years of age34 (16.3%)
]12; 22[ years of age70 (33.7%)
≥22 years of age104 (50.0%)
Race*
American Indian or Alaska Native9 (4.3%)
Asian8 (3.8%)
Black or African American23 (11.1%)
Native Hawaiian or Other Pacific Islander0 (0.0%)
White159 (76.5%)
Unknown race16 (7.7%)
Ethnicity
Spanish origin32 (15.4%)
Not Hispanic or Latino176 (84.6%)
Income of the household (annual)
Less than $19,99910 (4.8%)
$20,000-$37,99912 (5.8%)
$38,000-$44,9996 (2.9%)
$45,000-$54,9997 (3.4%)
$55,000-$64,9997 (3.4%)
$65,000-$74,99911 (5.2%)
$75,000-$99,99921 (10.1%)
$100,000 or more77 (37.0%)
Unknown57 (27.4%)
Level of education
< Highschool Diploma87 (41.8%)
Highschool Diploma or Equivalent35 (16.8%)
Associates Degree20 (9.6%)
Bachelors Degree41 (19.8%)
Masters Degree21 (10.1%)
Doctoral Degree4 (1.9%)

{31}------------------------------------------------

Slope0.8495% CI = [0.78, 0.89]
Intercept-0.3 mm95% CI = [0.1, 0.4]*
Bias at 0.0 mm-0.3 mm95% CI = [0.1, 0.5]*
Bias at 3.0 mm-7.0%95% CI = [-11.5 -3.5]
Bias at 5.0 mm-11.2%95% CI = [-15.6, -7.1]
Bias at 7.0 mm-13.0%95% CI = [-18.2, -8.3]
Bias at 9.0 mm-14.1%95% CI = [-19.8, -8.8]

*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.

{32}------------------------------------------------

Overjet - 3D Monitoring

A total of 263 patients rendering 292 results were included in the analysis. One patient could render multiple results corresponding to both central incisor couples.

Characteristic# (%)
Gender
Male83 (31.6%)
Female180 (68.4%)
Age group
[6; 12] years of age48 (18.3%)
]12; 22[ years of age80 (30.4%)
≥ 22 years of age135 (51.3%)
Race*
American Indian or Alaska Native10 (3.8%)
Asian9 (3.4%)
Black or African American29 (11.0%)
Native Hawaiian or Other Pacific Islander0 (0.0%)
White202 (77.2%)
Unknown race17 (6.5%)
Ethnicity
Spanish origin36 (13.7%)
Not Hispanic or Latino227 (86.3%)
Income of the household (annual)
Less than $19,99911 (4.2%)
$20,000-$37,99915 (5.7%)
$38,000-$44,9997 (2.7%)
$45,000-$54,9997 (2.7%)
$55,000-$64,99912 (4.5%)
$65,000-$74,99917 (6.5%)
$75,000-$99,99925 (9.5%)
$100,000 or more90 (34.2%)
Unknown79 (30.0%)
Level of education
< High School Diploma100 (38.0%)
Highschool Diploma or Equivalent45 (17.1%)
Associates Degree28 (10.6%)
Bachelors Degree61 (23.2%)
Masters Degree24 (9.2%)
Doctoral Degree5 (1.9%)

{33}------------------------------------------------

Slope1.03[1.01, 1.05]
Intercept0.14 mm[0.09, 0.19]*
Bias at 0.00 mm0.14 mm[0.07, 0.23]*
Bias at 3.00 mm7.2%[6.0, 8.9]
Bias at 5.00 mm5.4%[4.5, 6.7]
Bias at 7.00 mm4.7%[3.7, 6.0]
Bias at 9.00 mm4.3%[3.2, 5.6]

*Of note, the difference between the two 95% CI is due to the fact that the bootstrap is used for bias CI calculation, whereas for the intercept and slope it is the standard CI calculation linked to Passing-Bablok regression that is used.

{34}------------------------------------------------

Archwire and auxiliaries study

DentalMonitoring evaluated the passive archwire and auxiliaries indication using 3D Models acquired with intraoral scanners as input data to the Reference Method. The study was performed using data collected prospectively. The study involved six sites located in the United States: five sites for patient enrollment and one site for generation of the Reference Method results.

The Reference Method results were generated by performing a best fit on the acquired 3D Models. The best fit was performed per arch and per patient between the 3D Models acquired at two different timepoints. The best fits were done using a CAD/CAM software.

Data was collected in order to qualify the patient demographics.

{35}------------------------------------------------

Characteristic# (%)
Gender
Female172 (63%)
Male101 (37%)
Age group
[6; 12] years of age24 (9%)
]12; 22[ years of age197 (72%)
≥22 years of age52 (19%)
Race*
American Indian or Alaska Native8 (3%)
Asian11 (4%)
Black or African American48 (18%)
Native Hawaiian or Other Pacific Islander1 (0.3%)
White190 (70%)
Refused to disclose31 (11%)
Ethnicity
Spanish origin64 (23%)
Not Hispanic or Latino209 (77%)
Income of the household (annual)
Less than $19,99916 (6%)
$20,000-$37,99917 (6%)
$38,000-$44,99916 (6%)
$45,000-$54,99916 (6%)
$55,000-$64,99913 (5%)
$65,000-$74,99910 (4%)
$75,000-$99,99931 (11%)
$100,000 or more76 (28%)
Refused to disclose78 (28%)
Level of education
< Highschool Diploma175 (64%)
Highschool Diploma or Equivalent31 (11%)
Associates Degree18 (7%)
Bachelors Degree32 (12%)
Masters Degree11 (4%)
Doctoral Degree5 (2%)

Passive archwire and auxiliaries - 2D Monitoring

{36}------------------------------------------------

Reference Method
PassiveActiveTOTAL
Candidate MethodPositive class: Passive35763420
Negative class: Active43267310
TOTAL400330730
Result95% CI LowerBound95% CI UpperBound
Sensitivity89.0%84.9%92.1%
Specificity80.4%75.1%84.8%

{37}------------------------------------------------

Characteristic# (%)
Gender
Female166 (62%)
Male103 (38%)
Age group
[6; 12] years of age27 (10%)
]12; 22[ years of age196 (73%)
≥ 22 years of age46 (17%)
Race*
American Indian or Alaska Native9 (3%)
Asian11 (4%)
Black or African American43 (16%)
Native Hawaiian or Other Pacific Islander1 (0.4%)
White189 (70%)
Refused to disclose31 (12%)
Ethnicity
Spanish origin64 (24%)
Not Hispanic or Latino209 (76%)
Income of the household (annual)
Less than $19,99913 (5%)
$20,000-$37,99914 (5%)
$38,000-$44,99912 (4%)
$45,000-$54,99913 (5%)
$55,000-$64,99915 (6%)
$65,000-$74,9999 (3%)
$75,000-$99,99934 (13%)
$100,000 or more81 (30%)
Refused to disclose78 (29%)
Level of education
< Highschool Diploma174 (65%)
Highschool Diploma or Equivalent28 (10%)
Associates Degree18 (7%)
Bachelors Degree30 (12%)
Masters Degree12 (4%)
Doctoral Degree4 (1%)

Passive archwire and auxiliaries - 3D Monitoring

{38}------------------------------------------------

Reference Method
PassiveActiveTOTAL
Candidate MethodPositive class: Passive36148409
Negative class: Active39282321
TOTAL400330730
Result95% CI LowerBound95% CI UpperBound
Sensitivity90.4%86.7%93.2%
Specificity85.5%80.8%89.2%

Updated 3D Model study

DentalMonitoring evaluated the Updated 3D Model indication using 3D Models acquired with intraoral scanners as the Reference Method. For each case, they compared the 3D Model generated by Dental Monitoring to the baseline 3D Model using a best fit between the two 3D objects. Each best fit renders a deviation value and this value is then used to quantify the absolute difference between the two 3D objects, which was used to calculate the mean absolute error (MAE). The study was performed using data collected prospectively. The study involved seven sites located in the United States: six sites for patient enrollment and one site for generation of the study results.

The study results were generated by performing a best fit between the acquired 3D Models and the 3D Model generated by DentalMonitoring, i.e. Updated 3D Model. The best fit was performed per arch and per patient. The best fits were done using a CAD/CAM software.

Data was collected in order to qualify the patient demographics.

Characteristic# (%)
Gender
Female80 (32.0%)
Male170 (68.0%)
Age group
]12; 22[ years of age114 (45.6%)
≥ 22 years of age136 (54.4%)
Race*
American Indian or Alaska Native14 (5.6%)
Asian8 (3.2%)
Black or African American12 (4.8%)
Native Hawaiian or Other Pacific Islander0 (0.0%)
White221 (88.4%)
Refused to disclose4 (1.6%)
Ethnicity
Spanish origin22 (8.8%)
Not Hispanic or Latino226 (90.4%)
Unknown2 (0.8%)
Income of the household (annual)
Less than $19.9999 (3.6%)

{39}------------------------------------------------

$20,000-$37,99918 (7.2%)
$38,000-$44,9997 (2.8%)
$45,000-$54,99912 (4.8%)
$55,000-$64,99912 (4.8%)
$65,000-$74,99913 (5.2%)
$75,000-$99,99932 (12.8%)
$100,000 or more81 (32.4%)
Refused to disclose66 (26.4%)
Level of education
< Highschool Diploma106 (42.1%)
Highschool Diploma or Equivalent34 (13.5%)
Associates Degree29 (11.6%)
Bachelors Degree53 (21.0%)
Masters Degree24 (9.5%)
Doctoral Degree5 (2.0%)
Unknown1 (0.4%)

*Of note, the race does not correspond to the total number of patients as some patients with mixed races were included in the analysis.

# ofpatients# ofresultsMeanAbsoluteError(MAE)95% CI
2505360.10[0.093; 0.103]

Human Factors Testing

A Human factors validation testing was performed to demonstrate the usability of the device. Considering DentalMonitoring has two external user-facing interfaces, i.e. the Dashboard and the DM App, one study was performed for each of these interfaces.

Dashboard study

The Dashboard study enrolled two user groups:

  • . doctors; and
  • . practice employees.

Doctors represented 17 participants while practice employees represented 16 participants.

Testing took place in the United States in three cities, each located in different states.

The testing shows that the Dashboard can be used safely and effectively by the user population. Most of the users used the device properly and there is no significative difference between the different groups of population. Even if some use errors and use difficulties were observed, they cannot cause any harm to the users based on the performed analysis of the root cause. Furthermore, in the Dashboard, no tasks are considered as critical.

DM App study

The DM App study was divided into two parts: validation of the healthcare professional-facing interface

De Novo Summary (DEN230035)

{40}------------------------------------------------

which offers some of the features of the Dashboard, and validation of the patient-facing interface.

The DM App healthcare professional-facing interface study enrolled one user group constituted of doctors and practice employees. A total of 17 doctors along with 16 practice employees composed a testing group of 33 participants.

Testing took place in the United States in three cities, each located in different states.

The testing shows that the DM App healthcare professional-facing interface can be used safely and effectively by the user population. Most of the users used the device properly and there is no significative difference between the different groups of population. Furthermore, even if some use errors and use difficulties were observed, they cannot cause any harm to the users based on the performed analysis of the root cause.

The DM App patient-facing interface study enrolled five user groups:

  • . adults 22 years of age or older:
  • . adolescents between 13 and 21 years old;
  • children: pairs of pediatric patients (6 to 12 years of age) and lay caregivers (22 years of age or . older):
  • adults with disabilities: pairs of adult patients with disabilities who cannot independently use . the product (22 years of age or older) and lay caregivers (22 years of age or older); and
  • healthcare professionals (includes doctors and practice employees). .
Population groupadultsadolescentschildrenadults withdisabilitieshealthcareprofessionals
# of participants1615151717

Testing took place in the United States in five cities, each located in different states. This study included use of the DM Check Retractor and DM ScanBox which are mandatory hardware products to be used with the DM App in order to appropriately acquire Scans.

The testing shows that the DM ScanBox, the DM Cheek Retractor and the DM App, can be used safely and effectively by the user population. Most of the users used the device properly and there is no significative difference between the different groups of population. Furthermore, some use errors and use difficulties were observed; however, they are not expected to cause any harm to the users based on an analysis of the root cause that was conducted. Users were able to execute the relevant tasks directly or in workarounds within the patient profiles. The task errors identified were found to be non-critical and the testing was able to demonstrate that the device can be adequately used by the health care professional and patient according to the instructions for use.

LABELING

Labeling has been included. It comprises a user manual dedicated to healthcare professionals on one hand and a user manual dedicated to patients on the other hand.

Labeling also includes the two distinct labels embedded in the Dashboard, and in the DM App.

RISKS TO HEALTH

{41}------------------------------------------------

The table below identifies the risks to health that may be associated with use of and the measures necessary to mitigate these risks.

Identified RiskPotential Mitigation Measure
Adverse tissue reactionBiocompatibility evaluation
Use error/ improper device use leading toincorrect oral health information conveyed tohealthcare professionalLabelingUsability testing
Software malfunction leading to inaccurate patientmonitoring and diagnosisSoftware verification, validation, and hazardanalysis.
Software failure to identify or monitor the correctoral health condition leading to delayed treatmentClinical validationSoftware verification, validation, and hazardanalysis
Device failure to adequately capture imageleading to incorrect identification or monitoring oforal health conditionNon-clinical performance testingTechnological specifications

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act. the Dental Image Analyzer is subject to the following special controls:

(1) The patient-contacting components of the device must be demonstrated to be biocompatible.

(2) Non-clinical performance testing data must demonstrate that the device performs as intended under anticipated conditions of use and must include full specifications of the hardware requirements and testing to demonstrate the specified hardware ensures adequate data for validated and accurate measurements; (3) Clinical performance data must demonstrate the accuracy and precision of the diagnostic and monitoring device algorithms to perform as intended under anticipated conditions of use.

(4) Usability testing must demonstrate that the user can correctly use the device, based solely on reading the directions for use.

(5) Software verification, validation, and hazard analysis must be performed.

(6) Labeling must include:

(i) Information on the patient population for which the device has been demonstrated to be effective.

(ii) Information on how the device operates during the typical course of treatment

(iii) A detailed summary of the device technical parameters, including compatible operating systems, hardware, and/or accessories.

(iv) Instructions for cleaning of any reusable components.

(v) Warning that the device is not intended to be used independently for diagnosis and that the device does not replace clinical decision making.

  • (7) Patient labeling must be provided and must include:
    (i) Relevant contraindications, warnings, precautions, adverse effects/complications, and when to seek help from a dental healthcare professional.

(ii) Information on how the device operates during the typical course of treatment.

(iii) A detailed summary of the device technical parameters, including compatible operating systems, hardware, and/or accessories.

(iv) Information on the patient population for which there is clinical evidence of effectiveness.

  • (v) Instructions for cleaning of any reusable components.
    De Novo Summary (DEN230035)

{42}------------------------------------------------

BENEFIT-RISK DETERMINATION

The risks of the device are based on data collected in a clinical studies described above. Both the clinical and usability risks of the device have been evaluated and use of the device itself presents no significant direct risks to the user. The key clinical risk of this proposed device are the inability to identify or monitor the correct oral health condition leading to delayed treatment. The risks were quantified by sensitivity and specificity in the clinical study, which demonstrated the device's function. The risk of the proposed device is low.

The probable benefits of the device are also based data collected in a clinical studies as described above. This device can be a benefit to the clinician to monitor orthodontic progress between scheduled appointments and facilitate additional interaction between patient and clinician, as needed. Additionally, the device can provide benefit by specifically identifying potential clinical scenarios and areas of concern that would require further evaluation/intervention. The device is not intended to replace regularly scheduled appointments and would be considered low risk, as it ultimately facilitates more interaction between the patient and clinician during treatment.

PATIENT PERSPECTIVES

This submission did not include specific information on patient perspectives for this device.

BENEFIT/RISK CONCLUSION

In conclusion, given the available information above, for the following indication statement:

DentalMonitoring is a medical device software using image processing algorithms to analyze pictures of the oral cavity (hereinafter Scans). Scans are taken using the DM App. a smartphone, and the manufacturer's proprietary hardware products. Scans are taken by the patient, a non-healthcare professional, or a healthcare professional. The Scan is taken in healthcare facilities, such as a dental practice, or in a non-healthcare environment, such as the patient's own home. For some clinical parameters, DentalMonitoring requires a 3D Model.

The product is designed to assist healthcare professionals in remotely monitoring orthodontic treatments and treatment progress. The results of DentalMonitoring are intended to be used as an aid in diagnosis and monitoring, not on a stand-alone basis for clinical decision-making.

DentalMonitoring is indicated for use for patients over the age of 6 and reports results solely on permanent teeth.

DentalMonitoring can monitor the following clinical parameters:

  • oral hygiene: dental plaque / food residue; I
  • . soft tissue statement: gingival recession, black triangle;
  • . dental statement: closure of extraction space, tooth wear;
  • . alignment: closure of all anterior spaces; and
  • dental occlusion:
    • · in 2D Monitoring: midline deviation, overbite/open bite, overjet
    • o in 3D Monitoring: canine class, midline deviation, overbite/open bite, overjet

Additionally, the following clinical parameters specific to orthodontic treatment types or phases can be monitored using DentalMonitoring:

De Novo Summary (DEN230035)

{43}------------------------------------------------

  • . for aligner treatments: tracking (seat/unseat), attachment loss, button loss;
  • . for braces: bracket debonding, tie loss, self-ligating clips, passive archwire and auxiliaries; and
  • . for thermoformed retainers: tracking (seat/unseat).

Based on an initial 3D Model provided by a healthcare professional, DentalMonitoring can also provide healthcare professionals with 3D Models representative of the patient's dentition and treatment progress. This device is a prescription device and is not intended for over-the-counter use.

The probable benefits outweigh the probable risks for the DentalMonitoring. The device provides benefits and the risks can be mitigated by the use of general and the identified special controls.

CONCLUSION

The De Novo for the DentalMonitoring is granted and the device is classified as follows:

Product Code: SBC

Device Type: Dental Image Analyzer

Regulation Number: 21 CFR 872.1770

Class: Class II

N/A