(362 days)
The Galen ES is intended to assist in the movement of the specified Integra Jako Micro Laryngeal Alligator Forceps to a surgeon-selected position within a surgical field and remain in place until moved elsewhere by the operator. The Galen ES is indicated to be used in rigid microlaryngeal procedures where the specified Jako Micro Laryngeal Alligator Forceps would be utilized by an experienced trained otolaryngologist performing microlaryngeal surgery in an operating room environment in patients at least 18 years old.
The Galen ES (see Figure 1 and Figure 2) is a cooperatively-controlled surgical assistant platform (i.e., designed to be controlled by direct user interaction with the surgical instrument), intended to assist in the manipulation of the Integra Jako Micro Laryngeal Alligator Forceps (Integra Part# 3731042 serrated 9.0mm, straight 9.25 inch (235 mm)) (see Figure 3).
NOTE: The Integra Jako Micro Laryngeal Alligator Forceps will not be provided with the system.
The Galen ES is designed to work with the surgeon by sensing the physical forces and movements applied by the surgeon to the Integra Jako Micro Laryngeal Alligator Forceps connected to the Galen ES and successively moving the instrument in a manner consistent with the forces applied.
Here's a detailed breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:
Acceptance Criteria and Device Performance
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Simulated Use Testing:1. Performs as intended for indicated surgery.2. Tested in a simulated hospital environment with anatomically relevant model.3. Compatibility testing of all indicated instruments.4. Human factors/usability evaluation.5. Validation of device use by surgeons, including: - User interface and controller(s) - Compatibility with surgeon-applied forces, torques, speeds, and motion - Time required for emergency removal of device and associated instruments in the event of power loss or device failure. | Human Factors Engineering Study (Cadaver): 1. "Human factors testing demonstrated that all critical tasks were completed, and acceptance criteria were met by all users." 2. Testing in a simulated operating room environment. 3. The device is intended to be used with Integra Jako Micro Laryngeal Alligator Forceps (Integra Part# 3731042). Compatibility is implicitly demonstrated by the use of this specific instrument in the study.4. Usability validation study performed.5. Validation of surgeon use demonstrated: - Surgeons performed tasks including selecting instruments on UI, working ergonomically, completing workflow, recognizing critical info. - The device is designed to work with the surgeon by sensing the physical forces and movements applied. - "On average, surgeon participants (n=15) in the followup usability study required 14 seconds to remove the Jako forceps from the airway... This is considered clinically acceptable because these time frames would be consistent with removal of a manual surgical forceps tool during manual surgery." |
| Non-clinical Performance Testing:1. Hardware and system verification.2. Verification of critical parameters (including minimum and maximum forces, torques, speeds, and range of motion). | "The following bench tests were performed to mitigate the risks... Force sensor accuracy, Torque sensor accuracy, System performance verification: Degrees of freedom, o Motion actuation, Limit sensing, Passive instrument holding, Device immobilization, Recovery from unintended power loss." "All bench testing met acceptance criteria." |
| Software Verification, Validation, and Hazard Analysis:1. Performed for all software components. | "The Galen ES Surgical System software was developed in accordance with the following FDA guidance documents and standards... All testing and results met the acceptance criteria and/or standards." (References: FDA Guidance Documents on Software Validation & Content of Premarket Submissions; IEC 62304; ISO 14971) |
| Performance Testing:1. Demonstrates electromagnetic compatibility (EMC).2. Demonstrates electrical safety.3. Demonstrates thermal safety of the device. | "Electromagnetic compatibility (EMC) and electrical safety of the device was evaluated... The following Electrical/ Mechanical/Thermal Safety, and EMC testing has been performed: IEC 60601-1:2005 Ed.3+ A1, IEC 60601-1-2:2014 Ed.4, IEC 60601-1-6:2010 Ed.3+A1, IEC 80601-2-77:2019 Ed.1. All testing and results met the acceptance criteria of the above standards." |
| Sterile Components:1. All parts or components entering the sterile field must be demonstrated to be sterile. | "The device's RSA cleaning instructions were validated to achieve a Sterility Assurance Level (SAL) of 10-6. The ISA and surgical drapes are single-use accessories, which are individually packaged and sterilized via Ethylene Oxide (EO) to a SAL of 10-6." "All testing and results met the acceptance criteria and/or standards." |
| Shelf Life:1. Performance testing supports shelf life of sterile components by demonstrating continued sterility, package integrity, and device functionality. | "The purpose of the sterility, reprocessing, packaging, and shelf-life evaluations were to mitigate the risk of infection for the patient... The packaging was validated via a protocol compliant with the requirements of FDA-recognized consensus standards. All testing and results met the acceptance criteria and/or standards." |
| Biocompatibility:1. All patient-contacting components of the device must be demonstrated to be biocompatible. | "Biocompatibility testing was not performed because the device is not intended to have either direct or indirect contact with the body." (This implies no patient-contacting parts, thus the criteria is met by not needing testing.) |
| Labeling:1. Identification of compatible instruments.2. Statement about needed training.3. Summary of relevant performance testing.4. Reprocessing instructions for reusable components. | "Labeling for the Galen ES device includes the following: Identification of compatible surgical instruments, Recommended training for the safe use of the device, Summary of all relevant performance testing, including simulated use testing, Reprocessing instructions for reusable device components." (The document states labeling includes these, implying compliance.) |
2. Sample Size Used for the Test Set and Data Provenance
The primary test set for demonstrating clinical usability and safety was the Human Factors Engineering (HFE) study:
- Sample Size: A total of 47 individuals participated:
- 16 setup operators
- 31 operating surgeons (split into an initial usability study with 16 surgeons and a confirmatory usability study with 15 surgeons).
- Data Provenance: The study was conducted in a simulated operating room environment using cadaveric tissue. This indicates prospective data collection for the purpose of the study. The location or country of origin for the study itself is not specified but is implied to be within the regulatory jurisdiction relevant to the De Novo request (likely US).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The concept of "ground truth" for this device is primarily derived from successful task completion and expert judgment in the usability study, rather than a diagnostic accuracy assessment against a specific pathology.
- Number of Experts:
- 31 operating surgeons participated in the HFE study. Their successful completion of tasks was central to establishing that the device could be used safely and effectively.
- These surgeons likely double as the experts whose performance, when using the device, constitutes the "truth" of its usability.
- Qualifications of Experts:
- Setup operators' clinical experience ranged from 0.5 to 41 years.
- Operating surgeons' clinical experience ranged from 1 to 35 years. This implies a range from relatively less experienced to highly experienced otolaryngologists.
4. Adjudication Method for the Test Set
The document does not explicitly describe an adjudication method in the sense of multiple independent reviewers agreeing on a diagnostic label. Instead, the "evaluation" section states:
- "Successful performance of a task was defined as follows: Actions performed by the participant were in agreement with the details listed in the expected results column of the study protocol. Said actions were completed on the first attempt performed by the participant."
This suggests a pre-defined protocol-driven assessment of task completion, likely observed and recorded by study proctors or researchers, rather than an expert consensus/adjudication process post-hoc on the "ground truth" of performance. The "ground truth" here is the expected correct action for each task.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No MRMC study evaluating human diagnostic reading improvement with AI assistance was performed.
- This device is a cooperative surgical assistant robot, not an AI-powered diagnostic imaging tool. Its primary function is to assist in instrument manipulation during surgery, not to interpret medical images or data.
- The study design focused on human factors, usability, and the device's ability to maintain instrument positioning and stability, which are performance characteristics for a surgical tool, not diagnostic accuracy.
- The document mentions "reducing user fatigue and tremor as demonstrated in bench testing" as a benefit, but this is not quantified in the Human Factors study in terms of comparative effectiveness with and without the device in a clinical setting with human subjects. The HFE study focused on safe and effective use of the device.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No standalone "algorithm only" performance study was done in the context of diagnostic performance or decision-making.
- The device is explicitly a "cooperative surgical assistant platform" where the "device works in conjunction with the surgeon's movements." It is fundamentally designed for human-in-the-loop operation.
- However, extensive "Non-clinical Performance Testing - Bench" was conducted to characterize hardware and system performance (e.g., force/torque sensor accuracy, degrees of freedom, motion actuation, limit sensing). These can be considered standalone technical performance evaluations of the device's physical and computational capabilities without a human actively operating it in a surgical context, but they are not "algorithm-only" performance in the sense of an AI model making independent decisions.
7. The Type of Ground Truth Used
For the Human Factors Engineering (HFE) study, the "ground truth" was established by pre-defined protocol-driven expectations for task completion.
- It was not expert consensus on an existing medical condition or pathological finding.
- It was based on expected safe and effective actions for device setup, diagnostic testing, procedure execution, device shutdown, and emergency procedures.
- For critical tasks, "successful performance of a task was defined as: Actions performed by the participant were in agreement with the details listed in the expected results column of the study protocol AND Said actions were completed on the first attempt performed by the participant."
8. The Sample Size for the Training Set
The document describes training for the users of the device, not a training set for a machine learning algorithm.
- The users (surgeons and setup operators) underwent a "training program that mirrored the intended user training that will be provided to customers." This involved a didactic module and a hands-on training module.
- No specific sample size for a machine learning training set is mentioned, as the device is not described as an AI/ML-driven diagnostic or decision-making system in the traditional sense. It's a robotic assist platform for physical manipulation guided by the surgeon's forces.
9. How the Ground Truth for the Training Set Was Established
As noted above, there is no mention of a machine learning "training set" with ground truth in the context of a diagnostic AI. The "training" described is for human users.
- The "ground truth" for ensuring human users were trained properly was through the training program's curriculum and practice sessions. Participants had to practice tasks to gain familiarity before the study, emphasizing critical tasks. The effectiveness of this training was then assessed by their performance in the validation study against the predefined successful task completion criteria.
{0}------------------------------------------------
DE NOVO CLASSIFICATION REQUEST FOR GALEN ES COOPERATIVE SURGICAL ASSISTANT PLATFORM, GALEN ES
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Cooperative powered surgical assist device for ENT surgery. A cooperative powered surgical assist device for ear, nose, and throat (ENT) surgery is a device that facilitates ENT surgical procedures, including instrument placement. The device works in conjunction with the surgeon's movements to assist with precise and stable positioning of an instrument while maintaining the surgeon's direct physical control of the instrument.
NEW REGULATION NUMBER: 21 CFR 874.4460
CLASSIFICATION: Class II
PRODUCT CODE: QXG
BACKGROUND
DEVICE NAME: Galen ES Cooperative Surgical Assistant Platform, Galen ES
SUBMISSION NUMBER: DEN220047
DATE DE NOVO RECEIVED: July 22, 2022
SPONSOR INFORMATION:
Galen Robotics 1100 Wicomico St., Ste. 725 Baltimore, Maryland 21230
INDICATIONS FOR USE
The Galen ES Cooperative Surgical Assistant Platform, Galen ES is indicated as follows:
The Galen ES is intended to assist in the movement of the specified Integra Jako Micro Laryngeal Alligator Forceps to a surgeon-selected position within a surgical field and remain in place until moved elsewhere by the operator. The Galen ES is indicated to be used in rigid microlaryngeal procedures where the specified Jako Micro Laryngeal Alligator Forceps would be utilized by an experienced trained otolaryngologist performing microlaryngeal surgery in an operating room environment in patients at least 18 years old.
LIMITATIONS
{1}------------------------------------------------
The sale, distribution, and use of the Galen ES are restricted to prescription use in accordance with 21 CFR 801.109.
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
DEVICE DESCRIPTION
The Galen ES (see Figure 1 and Figure 2) is a cooperatively-controlled surgical assistant platform (i.e., designed to be controlled by direct user interaction with the surgical instrument), intended to assist in the manipulation of the Integra Jako Micro Laryngeal Alligator Forceps (Integra Part# 3731042 serrated 9.0mm, straight 9.25 inch (235 mm)) (see Figure 3).
NOTE: The Integra Jako Micro Laryngeal Alligator Forceps will not be provided with the system.
The Galen ES is designed to work with the surgeon by sensing the physical forces and movements applied by the surgeon to the Integra Jako Micro Laryngeal Alligator Forceps connected to the Galen ES and successively moving the instrument in a manner consistent with the forces applied.
Image /page/1/Figure/6 description: The image shows a medical robot with several labeled components. The robot has a manipulator arm (1) connected to a robot side adapter (RSA) (2) and an instrument side adapter (ISA) (3). The robot also features an arm immobilization mechanism (AIM) (4), a monitor and touchscreen UI (5), a mobile base (6), and casters (7) for mobility.
Figure 1: Galen ES Cooperative Surgical Assistant Platform (Front View)
{2}------------------------------------------------
Image /page/2/Figure/0 description: This image shows a medical device with various labeled parts. The device includes a monitor (1), a handle (2), and a power cord (13). Other labeled components include numbers 3, 4, 5, 6, 7, 8, 9, 10, 11, and 12, which likely correspond to specific features or controls of the device.
-
- Monitor
-
- Monitor arm
-
- Foot pedal and diagnostic tool storage baskets
-
- Rear panel
-
- Base immobilizing pedal
-
- Emergency Power Off (ESTOP) buttons
-
- Robot Side Adapter (RSA)
-
- Instrument Side Adapter (ISA)
-
- Manipulator arm
-
- Immobilization release lever 11. Arm Immobilization Mechanism (AIM)
-
- Handle
-
- Foot pedal and cable
Figure 2: Galen ES Cooperative Surgical Assistant Platform (Rear View)
The following accessories are included with the Galen ES:
-
- Two (2) sterile, single-use surgical drapes
- a. One (1) to cover the device's manipulator arm and mobile base,
- b. One (1) to cover the device's monitor and touchscreen
-
- One (1) Robot Side Adapter (RSA)
-
- One (1) RSA Sterilization Tray
- One (1) Instrument Side Adapter (ISA). The ISA is a sterile, single-use accessory, which 4. connects to the device's manipulator arm via the Robot Side Adapter (RSA), with the intended use of holding the Jako forceps
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a robotic arm with an instrument attached to it. The instrument is a pair of scissors. The robotic arm is labeled as "Arm", and the instrument is labeled as "Instrument". Other labels include "ISA", "RSA", and "QCM".
Figure 3: Jako forceps which have been attached to the ISA. The ISA is coupled to the RSA which is attached to the Robotic Arm. The ISA (with attached forceps) can be manually removed from the arm (if necessary).
Table 1 provides a high-level description of the critical system components and accessories of the device.
Table 1: Critical System Components and Accessories Relevant to Treatment of a Disease or Condition
{4}------------------------------------------------
| System Component or Accessory | Description |
|---|---|
| 1. Manipulator Arm(Figure 1-1) | The robotic element containing several components that allow the device to function as a robotic assistant for the surgeon:The Force-Torque sensor reads the forces and torques applied to the forceps attached to the device, allowing the device to sense the actions of the surgeon. The motors drive the motion of the arm, which allows the device to physically move the instrument in a manner consistent with the forces applied. The encoders sense the position of the arm, allowing the device to know its current location and make movements. These elements work in concert, allowing the surgeon to experience improved instrument stability and control, as they manipulate the instrument in tight, narrow-corridor procedures. |
| 2. Robot Side Adapter (RSA)(Figure 1-2) | A removable and sterilizable accessory attached to the manipulator arm that serves as the attachment point for the ISA. |
| 3. Instrument Side Adapter (ISA)(Figure 1-3) | The ISA is a removable one-time use accessory that attaches to the RSA.It serves as the attachment point for the forceps that are used during the surgical procedure.The ISA allows the forceps to be securely attached to the device and ensures a secure attachment to the RSA.The ISA is intended as a one-time-use disposable, simplifying post-procedural clean up and sterilization. |
| 4. Arm Immobilization Mechanism (AIM)(Figure 1-4) | Protects the manipulator arm and force sensor from damage during transportation and storage. |
| System Component or Accessory | Description |
| 5. Monitor with touchscreen(Figure 1-5) | Displays information and receives inputsGuides the supporting operators through the set up and shut downproceduresAllows the surgeon to change device settingsInforms the surgeon when the device encounters an error. |
| 6. Mobile Base(Figure 1-6) | Contains the device's components and serves as a rigid frame toprotect them from physical damage. |
| 7. Casters(Figure 1-7) | Attach to the bottom of the chassis and allow the device to bemoved. |
| 8. Emergency (ESTOP) buttons(Figure 2-6) | When an ESTOP is engaged, the power to the manipulator armmotors is removed, and brakes on the delta stage portion of themanipulator arm are engaged. This serves to lock the manipulatorarm in place. |
| 9. Base-immobilizing pedal(Figure 2-5) | When the Base Immobilizing Pedal is depressed, it activates theBase Immobilizing Actuator, locking the device in a fixedposition, preventing it from being moved on its casters.The immobilization of the Device is hydraulically-actuated anddoes not require electrical power to be engaged or disengaged.This ensures the device remains in the same position during theprocedure. |
| 10. Immobilization release lever(Figure 2-10) | Disengages the Base-immobilizing actuator, allowing the device tobe moved on its casters. |
| 11. Foot pedal and cable(Figure 2-13) | When the foot pedal is engaged, the power to the manipulator armis enabled, allowing movement of the manipulator arm. Thisallows the surgeon to use the device to manipulate the attachedforceps. |
{5}------------------------------------------------
{6}------------------------------------------------
Third-party Devices
The Galen ES is intended to be used with the following third-party devices (Table 2), which are not provided with the Galen ES System, to enable the subject device to perform its intended purpose.
| Component | Description |
|---|---|
| AlligatorForceps | Jako Micro Laryngeal Alligator Forceps Serrated 9.0mm, straight 9.25 inch (235 mm) – Integra Part# 3731042 |
| Table 2: Third-Party Devices Not Provided with the Galen ES System | ||
|---|---|---|
SUMMARY OF NONCLINICAL/BENCH STUDIES
Non-clinical performance tests were conducted to demonstrate that the Galen ES will perform as anticipated for its intended use and to mitigate the risks to health as outlined below.
BIOCOMPATIBILITY/MATERIALS
Biocompatibility testing was not performed because the device is not intended to have either direct or indirect contact with the body.
SHELF LIFE/STERILITY
The purpose of the sterility, reprocessing, packaging, and shelf-life evaluations were to mitigate the risk of infection for the patient. The device's RSA cleaning instructions were validated to achieve a Sterility Assurance Level (SAL) of 10-6. The ISA and surgical drapes are single-use accessories, which are individually packaged and sterilized via Ethylene Oxide (EO) to a SAL of 106. The packaging was validated via a protocol compliant with the requirements of FDA-recognized consensus standards.
All testing and results met the acceptance criteria and/or standards.
ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY
Electromagnetic compatibility (EMC) and electrical safety of the device was evaluated to mitigate the risk of electrical fault resulting in injury to patient or user.
The following Electrical/ Mechanical/Thermal Safety, and EMC testing has been performed:
| Identifier | Title |
|---|---|
| IEC 60601-1:2005 Ed.3+ A1 | General Requirements for Basic Safety and Essential |
| Performance |
{7}------------------------------------------------
| Identifier | Title |
|---|---|
| IEC 60601-1-2:2014 Ed.4 | General Requirements for Basic Safety and EssentialPerformance – Collateral Standard: ElectromagneticDisturbances |
| IEC 60601-1-6:2010Ed.3+A1 | General Requirements for Basic Safety and EssentialPerformance – Collateral Standard: Usability |
| IEC 80601-2-77:2019 Ed.1 | Particular Requirements for the Basic Safety and EssentialPerformance of Robotically Assisted Surgical Equipment |
All testing and results met the acceptance criteria of the above standards.
SOFTWARE
The software was evaluated to mitigate the risks of electrical, and mechanical faults associated with the device not working as intended due to the programming, or due to system malfunctions.
The Galen ES Surgical System Software consists of two (2) separate modules - the Galen ES Control software and the Galen ES GUI software. The Galen ES GUI software presents the main user interface for the Galen ES Surgical System and runs on the touchscreen computer of the Galen ES Surgical System. The remainder of the software modules control the functions of their respective hardware.
The Galen ES Surgical System software was developed in accordance with the following FDA guidance documents and standards:
| Identifier | Title |
|---|---|
| FDA Guidance Documents | |
| General Principles of Software Validation, issued January 11, 2002 | |
| Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices,issued May 11, 2005 | |
| Off-the-Shelf Software Use in Medical Devices, issued September 27, 2019 | |
| Software | |
| IEC 62304:2006 Ed.1+A1 | Medical device software - Software life-cycle processes |
| Risk Management | |
| ISO 14971:2019 | Medical devices - Application of risk management to medical devices |
Galen considers the software to be a major level of concern (LOC) because failure or latent flaw could directly result in death or serious injury to the patient or operator. Furthermore, a failure or latent flaw could indirectly result in death or serious injury of
{8}------------------------------------------------
the patient or operator through incorrect or delayed information or through the action of a healthcare provider.
All testing and results met the acceptance criteria and/or standards.
CYBERSECURITY
The Galen ES Surgical System was evaluated for cybersecurity risks consistent with the FDA guidance document titled, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, issued October 2, 2014.
Acceptable information was provided per the above referenced cybersecurity guidance about software malfunctions. malicious or otherwise. which could lead to patient harm, or the ability to control/manipulate/affect the device remotely, whether by design or not.
PERFORMANCE TESTING - BENCH
The following bench tests were performed to mitigate the risks of electrical, and mechanical fault resulting in injury to patient or user, tissue perforation and/or injury due to system malfunction, user error resulting in patient injury:
- . Force sensor accuracy
- Torque sensor accuracy
- System performance verification: ●
- Degrees of freedom O
- o Motion actuation
- Limit sensing O
- Passive instrument holding O
- Device immobilization O
- Recovery from unintended power loss O
The bench tests characterize device performance and design verification for the Galen ES.
These bench tests were conducted using verification protocols including objective acceptance criteria. Bench testing was also conducted on a preliminary version of the device to support reduction of both surgeon fatigue and tremor. All bench testing met acceptance criteria.
PERFORMANCE TESTING - CADAVER
Human Factors Engineering
Human Factors Engineering Testing was performed to mitigate the risks of devicespecific use error or of mechanical or electrical faults that result in injury to the patient or user. Usability testing was performed in accordance with FDA guidance document titled,
{9}------------------------------------------------
Applying Human Factors and Usability Engineering to Medical Devices, issued February 3, 2016, to demonstrate the device usability.
Purpose
The usability validation study was conducted to evaluate whether the Galen ES System, as designed, supports safe use by representative users when performing rigid microlaryngeal procedures in cadaver. The validation testing consisted of the following objectives:
-
- Evaluate whether typical users can safely and effectively interact with the device
-
- Confirm that use-related risks have been appropriately captured and assessed in the Use Related Risk Analysis (URRA) or uncovering any previously unforeseen use errors
-
- Assess the effectiveness of the risk mitigation measures that have been applied during the Galen ES' design stage
-
- Identify and validate the critical tasks and sub-tasks that were a part of the device use training program provided in the field.
Study Design
A total of 47 individuals participated in the human factors testing: 16 were setup operators and 31 were operating surgeons. The human factors testing spanned two studies: a usability study (with 16 surgeons) and a followup confirmatory usability study (with 15 surgeons). The confirmatory usability study was conducted to more directly test for and ensure device usability from a safety perspective, including in emergent situations (simulated hemorrhage) requiring immediate device removal. The setup operators' years of clinical experience ranged from 0.5 to 41 years, while the operating surgeons' clinical experience ranged from 1 to 35 years.
The testing was conducted in a simulated operating room environment and involved device setup, diagnostic testing, procedure execution, and device shutdown, as well as emergency procedures involving critical tasks.
Participants underwent a training program that mirrored the intended user training that will be provided to customers, except on a condensed schedule. This program emphasized practicing and performing critical tasks to gain familiarity with the device before entering the study room. An enforced 1-hour training decay period occurred between completion of the training program and commencement of the study session.
Twenty seven (27) tasks were defined for surgeons in a simulated hospital environment. Tasks included: Selecting correct instrument on user interface, Working ergonomically with the device without problems, Completing workflow, Recognizing critical information, e.g., if the device moved to the edge of its bounds. Three clinically relevant laryngoscopy tasks were also included in a confirmatory usability study with 15 surgeon participants: 1) Anatomical mapping and fiducial Identification, 2) Vocal cord
{10}------------------------------------------------
manipulation, and 3) Hemorrhage simulation and emergency Galen ES removal and logging of device removal time.
Evaluation
Successful performance of a task was defined as follows:
- Actions performed by the participant were in agreement with the details listed in ● the expected results column of the study protocol
- . Said actions were completed on the first attempt performed by the participant
Results
Human factors testing demonstrated that all critical tasks were completed, and acceptance criteria were met by all users. The Galen ES device has been found to be safe and effective for the intended users, uses, and use environment. The participants provided valuable feedback through their execution of the tasks set out by the validation protocol, as well as the subsequent critiques provided during debriefing interviews and questionnaires. Results showed participants were able to perform all aspects of a microlaryngeal procedure appropriate for use with the Galen ES system. Although some tasks were determined to be difficult, analyses revealed that no new hazardous use scenarios were identified, and no critical task was modified.
The emergency removal of the forceps and Galen ES from the surgical field and logging the removal time in a simulated emergency (bleeding) event was conducted in the confirmatory usability study. On average, surgeon participants (n=15) in the followup usability study required 14 seconds to remove the Jako forceps from the airway after initiation of a simulated bleeding event began; 9 participants took less than 10 seconds, 3 participants took between 10 and 20 seconds, and the remaining 3 participants took longer than 20 seconds (longest time was 61 seconds). This is considered clinically acceptable because these time frames would be consistent with removal of a manual surgical forceps tool during manual surgery, i.e., without the use of a cooperative powered surgical assist device.
Pediatric Extrapolation
In this De Novo request, existing data were leveraged to support the of the proposed device in the 18-21 years old pediatric subpopulation. It was appropriate to indicate the device for individuals 18 and older because patients aged 18 to 21 do not carry additional differences or risks relative to the adult patient population for this device.
TRAINING PROGRAM DESIGN
The training program for the Galen ES System is comprised of a series of modules to ensure that each user understands each component of the system and thoroughly comprehends the various workflows involved in operating the device.
The training program includes the following sections:
{11}------------------------------------------------
-
- Didactic Training Module: A slideshow presentation that includes an introduction to the device's technology and operation. This includes a technical overview of the robot, functional aspects of the system, and troubleshooting tips.
-
- Hands-on Training Module:
- Specific training for nurses and technicians this will include practicing tasks a. such as using the touchscreen, drapes, ISA, and RSA.
- b. Specific training for surgeons this will include practicing Galen-assisted procedures using a simulated patient anatomy bench model and cadaveric tissue.
-
- Proctored Device Use: Following completion of the training the company will engage in 1 to 2 proctored procedures with the user where a company expert will be available to address any device-related questions during the procedure.
LABELING
Labeling for the Galen ES device includes the following:
- Identification of compatible surgical instruments
- Recommended training for the safe use of the device
- Summary of all relevant performance testing, including simulated use testing
- Reprocessing instructions for reusable device components ●
RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of a cooperative powered surgical assist device for ENT surgery and the measures necessary to mitigate these risks.
| Risks to Health | Mitigation Measures |
|---|---|
| Injury to anatomic structures resultingin bleeding or loss of function | Simulated use testing |
| Human factors evaluation | |
| Non-clinical performance testing | |
| Software verification, validation and hazard analysis | |
| Labeling | |
| Electromagnetic compatibility (EMC) testing | |
| Electrical safety testing | |
| Thermal safety testing | |
| Biocompatibility evaluation | |
| Inability to remove the deviceexpeditiously (e.g., if device becomessluggish or frozen) resulting in:• Lack of sufficient control ofbleeding | Simulated use testing |
| Non-clinical performance testing | |
| Software verification, validation and hazard analysis | |
| Labeling | |
| Electromagnetic compatibility (EMC) testing | |
| Electrical safety testing |
{12}------------------------------------------------
| Risks to Health | Mitigation Measures |
|---|---|
| • Prolonged time for removal anddelayed access to criticalstructures | |
| Infection | LabelingSimulated use testingSterilization validationShelf life testing |
SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the cooperative powered surgical assist device for ENT surgery is subject to the following special controls:
- (1) Simulated use testing must demonstrate that the device performs as intended under anticipated conditions of use to successfully assist in the indicated surgery, including:
- Testing in a simulated hospital environment with an anatomically relevant model; (i)
- (ii) Compatibility testing of all indicated instruments;
- Human factors/usability evaluation; and (iii)
- Validation of device use by surgeons, including: (iv)
- The user interface and controller(s); (A)
- Compatibility with the ranges of surgeon-applied forces, torques, speeds, and (B) motion; and
- (C) Time required for emergency removal of the device and associated instruments in the event of power loss or device failure.
- (2) Non-clinical performance testing must demonstrate hardware and system verification, including verification of critical parameters (including minimum and maximum forces, torques, speeds, and range of motion).
- Software verification, validation, and hazard analysis must be performed for any software (3) components of the device.
- (4) Performance testing must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device.
- All parts or components of the device that enter the sterile field must be demonstrated to be (ર) sterile.
- Performance testing must support the shelf life of device components provided sterile by (6) demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.
- All patient-contacting components of the device must be demonstrated to be (7) biocompatible.
- (8) Labeling must include:
- Identification of compatible instruments; (i)
- A statement about any training needed prior to use of the device; (ii)
- A summary of all relevant performance testing, including simulated use testing; and (iii)
- Reprocessing instructions for reusable device components. (iv)
{13}------------------------------------------------
BENEFIT-RISK DETERMINATION
Risks
The risks of the Galen ES device include injury to anatomic structures resulting in bleeding or loss of function, inability to remove the device expeditiously (e.g., if device becomes sluggish or frozen) resulting in lack of sufficient control of bleeding or prolonged time for removal and delayed access to critical structures, and infection. General risks of the Galen ES include injury to patients or users due to use error or electrical/mechanical faults. Use-related errors may also occur if surgeons or setup operators are insufficiently trained on safe and effective use of the device. These risks are mitigated (in part) with usability testing and non-clinical performance testing, labeling (including warnings and precautions) and training.
Benefits
The following probable benefits are identified for the Galen ES based on the data collected in the human factors studies or as demonstrated in bench testing:
-
- Assist a surgeon in maneuvering and positioning the intended Jako forceps within the narrow surgical corridor presented during rigid microlaryngeal procedures,
-
- Enables prolonged positioning of the Jako forceps in a specific position that is determined by the surgeon,
-
- Moves in concert with the applied forces. Provides stability in the procedure by reducing user fatigue and tremor as demonstrated in bench testing, and
-
- When used as indicated, frees hand of surgeon to utilize other instrument(s).
The usability studies conducted in support of this device provided data from 31 operating surgeons who performed simulated microlaryngeal procedures on cadaver. During these simulated procedures, all critical tasks were successfully completed and no significant events were encountered.
Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
The Galen ES is intended to assist in the movement of the specified Integra Jako Micro Laryngeal Alligator Forceps to a surgeon-selected position within a surgical field and remain in place until moved elsewhere by the operator. The Galen ES is indicated to be used in rigid microlaryngeal procedures where the specified Jako Micro Laryngeal Alligator Forceps would be utilized by an experienced trained otolaryngologist
{14}------------------------------------------------
performing microlaryngeal surgery in an operating room environment in patients at least 18 years old.
The probable benefits outweigh the probable risks for the Galen ES Cooperative Surgical Assistant Platform, or Galen ES. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
CONCLUSION
The De Novo request for the Galen ES Cooperative Surgical Assistant Platform, Galen ES is granted and the device is classified as follows:
Product Code: QXG Device Type: Cooperative powered surgical assist device for ENT surgery Regulation Number: 21 CFR 874.4460 Class: II
N/A