K Number
DEN220027

Validate with FDA (Live)

Date Cleared
2023-05-18

(381 days)

Product Code
Regulation Number
862.1602
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The B.R.A.H.M.STM sFlt-1/PIGF KRYPTORTM Test System is comprised of the B . R . H . M S PIGF plus KRYPTOR assay and the B . A . H . M . S sFit-1 KRYPTOR assay.

The B R . A . H . M . S PIGF plus KRYPTOR is an automated immunofluorescent assay using Time-Resolved Amplified Cryptate Emission (TRACE™) technology for the quantitative determination of the concentration of Placental Growth Factor (PIGF) in human serum and plasma (K2 EDTA) on the B.R.A.H.M.S KRYPTOR analyzer.

The B · A · H · M · S s F t - 1 K R PTOR is an automated immunofluorescent assay using Time-Resolved Amplified Cryptate Emission (TRACE) technology for the quantitative determination of the concentration of soluble fms-like tyrosine kinase-1 (sFit-1), also known as VEGF receptor-1, in human serum and plasma (K2 EDTA) on the B.R.A.H:M:S KRYPTOR analyzer.

The B · A · H · M · S PIGF plus KRYPTOR is to be used in conjunction with the B · A · M · S sFlt-1 KRYPTOR along with other laboratory tests and clinical assessments to aid in the risk assessment of pregnant women (singleton pregnancies between gestational age 23+0 to 34+6/7weeks) hospitalized for hypertensive disorders of pregnancy (preeclampsia, chronic hypertension with or without superimposed preeclampsia, or gestational hypertension) for progression to preeclampsia with severe features (as defined by American College of Obstetricians and Gynecologists (ACOG) guidelines) within 2 weeks of presentation.

Device Description

The B · A · H · M · S · F · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · PIGF plus KRYPTOR assay and the B.R.A.H.M.S sFlt-1 KRYPTOR assay.

B R . A H M . S PIGF plus KRYPTOR and B . R . H . M . STM sFit-1 KRYPTOR assays are supplied as two reagent kits that are run on the B-R-A-H-M-S KRYPTOR compact PLUS. Each kit is sufficient for 75 determinations, and is comprised of two bottled reagents:

PIGF
Cryptate conjugate: Lumi4®-Tb labeled, anti-human PlGF antibody, buffer, bovine albumin, rat immunoglobulins, bovine immunoglobulins, goat immunoglobulins, trehalose, mannitol.
XL conjugate: Cyanin 5.5 labeled, anti-human PlGF antibody, buffer, bovine albumin, rat immunoglobulins, bovine immunoglobulins, goat immunoglobulins.

sFlt-1
Cryptate conjugate: Lumi48-Tb labeled, anti-human sFlt-1 antibody, buffer, bovine albumin, bovine immunoglobulins, mouse immunoglobulins, dextran, preservative. XL conjugate: Cyanin 5.5 labeled, anti-human sFlt-1 antibody, buffer, bovine albumin, bovine immunoglobulins, mouse immunoglobulins, dextran, preservative.

The B · A · H · M · S KRYPTOR compact PLUS is an automated test platform. The analyzer consumables are:

B · R · A · H · M · S KRYPTOR compact Solution 1: ProClinTM 150 Solution
B · R · A · H · M · S KRYPTOR compact Solution 2: Potassium fluoride solution
B · R · A · H · M · S KRYPTOR compact Solution 3: Active chlorine and sodium hydroxide solution
B · R · A · H · M · S KRYPTOR compact Solution 4: Sodium hydroxide solution
B · R · A · H · M · S KRYPTOR BUFFER: Phosphate Buffer Saline
B · R · A · H · M · S KRYPTOR compact REACT: 60 plates

Also supplied and required are calibrators - single level B.R.A.H.M.S PIGF plus KRYPTOR CAL and two levels B. R. A. H. M.S sFlt-1 KRYPTOR CAL. The instrument is calibrated on a 15 day cycle, and re-calibrated with each new lot of reagents.

Also supplied and required are three levels of controls - B.R.A.H.M.S PIGF plus KRYPTOR OC Controls and B.R.A.H.M.S sFlt-1 KRYPTOR OC.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the B.R.A.H.M.S sFlt-1/PIGF KRYPTOR Test System, based on the provided text:

Acceptance Criteria and Device Performance

The primary clinical performance acceptance criteria revolve around the device's ability to aid in the risk assessment of pregnant women for progression to preeclampsia with severe features. This is evaluated through its positive predictive value (PPV), negative predictive value (NPV), sensitivity, and specificity.

Table of Acceptance Criteria and Reported Device Performance:

Performance MetricAcceptance Criteria (Implied by positive De Novo grant)Reported Device Performance (95% CI)
Positive Predictive Value (PPV)Sufficiently high to indicate a clinically useful risk assessment.65% (59.3, 70.6)
Negative Predictive Value (NPV)Sufficiently high to indicate a clinically useful risk assessment.96% (92.9, 97.6)
SensitivitySufficiently high to identify true positives.94% (89.1, 96.3)
SpecificitySufficiently high to identify true negatives.75% (70.2, 79.0)
Positive Likelihood Ratio (PLR)Implied to be clinically useful.3.72 (3.11, 4.46)
Negative Likelihood Ratio (NLR)Implied to be clinically useful.0.09 (0.05, 0.15)

Note: The document does not explicitly state numerical acceptance criteria prior to the study results. The granting of the De Novo request implies that the reported performance metrics were deemed acceptable by the FDA for the device's intended use.


Study Proving Device Meets Acceptance Criteria

The study described is the PRAECIS (Preeclampsia Risk Assessment: Evaluation of Cut-offs to Improve Stratification, Identification and Validation of a Cut-off for the Ratio of Soluble Fms-like Tyrosine Kinase-1 to Placental Growth Factor (sFlt-1/PlGF) to Stratify Risk in Pregnant Women with Hypertensive Disorders of Pregnancy) clinical study.

1. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: 556 enrollments (from 520 patients, with some patients enrolled multiple times).
  • Data Provenance: Prospective, multicenter, blinded, non-interventional study conducted across 18 sites in the US.

2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Number of Experts: A panel of 3 independent obstetricians.
  • Qualifications: "Obstetricians" is the only qualification mentioned. No specific experience levels (e.g., years of experience) are provided.

3. Adjudication Method for the Test Set:

  • Method: A central adjudication committee (panel of 3 independent obstetricians) was assembled to review data and arrive at consensus regarding each patient's diagnoses for the development of sPE following the 2013 ACOG clinical practice guidelines. This indicates a consensus-based adjudication.

4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

  • No, an MRMC comparative effectiveness study was not done. This study evaluates an in vitro diagnostic (IVD) test system (measurement of sFlt-1 and PlGF levels and their ratio) rather than an AI-assisted imaging diagnosis system where human readers would be involved in interpreting images. Therefore, the concept of "human readers improving with AI vs. without AI assistance" does not apply here.

5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

  • Yes, in essence, a standalone evaluation was performed for the device's output. The "algorithm" here is the immunofluorescent quantitative assay that measures sFlt-1 and PIGF concentrations and calculates their ratio. The clinical performance (PPV, NPV, sensitivity, specificity) is based on comparing the device's calculated ratio (using a defined cutoff of >40 for 'high risk' and <40 for 'low risk') directly against the ground truth established by the clinical outcome adjudicated by the expert panel. There isn't a "human-in-the-loop" step in the device's output determination influencing the sFlt-1/PIGF ratio result itself, though the result is then used by a clinician for risk assessment.

6. The Type of Ground Truth Used:

  • Expert Consensus + Outcomes Data: The ground truth for the clinical outcome ("progression to preeclampsia with severe features (sPE) within 2 weeks of presentation") was established by a central adjudication committee (panel of 3 independent obstetricians) who reviewed data and arrived at a consensus diagnosis following the 2013 ACOG guidelines. This relies on clinical outcomes and expert interpretation of those outcomes.

7. The Sample Size for the Training Set:

  • The document implies a two-part study: "identification (part 1) and validation (part 2) of the cut-off".
  • No explicit sample size is given for "Part 1" (the identification phase) for establishing the sFlt-1/PlGF ratio cutoff. The 556 enrollments refer to "Part 2" (the validation phase). It states the interpretation of results "determined in part 1".
  • The PRAECIS study description focuses on the validation cohort (Part 2).

8. How the Ground Truth for the Training Set Was Established:

  • As mentioned above, the sample size for the training/identification set (Part 1) is not explicitly detailed. However, the cut-off was "identified" in Part 1 of the PRAECIS study. It's reasonable to infer that the ground truth for establishing the cut-off in Part 1 would have been established similarly to Part 2, i.e., through clinical outcome assessment (development of sPE) adjudicated by experts, but on a separate cohort of patients used for discovery/derivation. The document doesn't provide details on the specific characteristics of the Part 1 cohort or its ground truth establishment process beyond stating it was used to derive the cutoff.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR B.R.A.H.M.S sFlt-1/ PIGF KRYPTOR Test System DECISION SUMMARY

I Background Information:

A De Novo Number

DEN220027

B Applicant

BRAHMS GmbH, Part of Thermo Fisher Scientific

C Proprietary and Established Names

B.R.A.H.M.S sFlt-1/ PIGF KRYPTOR Test System.

D Regulatory Information

ProductCode(s)ClassificationRegulationSectionPanel
QWHClass II withspecial controls21 CFR 862.1602 –Prognostic test fordevelopment orprogression ofpreeclampsiaClinical Chemistry andToxicology Devices

Submission/Device Overview: II

A Purpose for Submission:

De Novo request for evaluation of automatic class III designation for B. R. A. H. M. S sFlt-1/ PIGF KRYPTOR Test System.

B Measurand:

Placental growth factor (PlGF) Soluble fms-like tyrosine kinase-1 (sFlt-1)

C Type of Test:

Immunofluorescent quantitative assay

III Indications for Use:

{1}------------------------------------------------

A Indication(s) for Use:

The B.R.A.H.M.STM sFlt-1/PIGF KRYPTORTM Test System is comprised of the B . R . H . M S PIGF plus KRYPTOR assay and the B . A . H . M . S sFit-1 KRYPTOR assay.

The B R . A . H . M . S PIGF plus KRYPTOR is an automated immunofluorescent assay using Time-Resolved Amplified Cryptate Emission (TRACE™) technology for the quantitative determination of the concentration of Placental Growth Factor (PIGF) in human serum and plasma (K2 EDTA) on the B.R.A.H.M.S KRYPTOR analyzer.

The B · A · H · M · S s F t - 1 K R PTOR is an automated immunofluorescent assay using Time-Resolved Amplified Cryptate Emission (TRACE) technology for the quantitative determination of the concentration of soluble fms-like tyrosine kinase-1 (sFit-1), also known as VEGF receptor-1, in human serum and plasma (K2 EDTA) on the B.R.A.H:M:S KRYPTOR analyzer.

The B · A · H · M · S PIGF plus KRYPTOR is to be used in conjunction with the B · A · M · S sFlt-1 KRYPTOR along with other laboratory tests and clinical assessments to aid in the risk assessment of pregnant women (singleton pregnancies between gestational age 23+0 to 34+6/7weeks) hospitalized for hypertensive disorders of pregnancy (preeclampsia, chronic hypertension with or without superimposed preeclampsia, or gestational hypertension) for progression to preeclampsia with severe features (as defined by American College of Obstetricians and Gynecologists (ACOG) guidelines) within 2 weeks of presentation.

B Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

B.R.A.H.M.S sFit-1 KRYPTOR is not indicated to be used as a stand-alone diagnostic assay.

B . R . A . H . M S PIGF plus is not indicated to be used as a stand-alone diagnostic assay.

B.R.A.H.M.S sFlt-1 KRYPTOR must be run in conjunction with B.R.A.H.M.S PIGF plus KRYPTOR and the same patient sample must be used to run both assays.

For in vitro diagnostic use only.

Patient samples should not be diluted.

C Special Instrument Requirements:

B.R.A.H.M.S KRYPTOR compact Plus

IV Device/System Characteristics:

  • A Device Description:

{2}------------------------------------------------

The B · A · H · M · S · F · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · PIGF plus KRYPTOR assay and the B.R.A.H.M.S sFlt-1 KRYPTOR assay.

B R . A H M . S PIGF plus KRYPTOR and B . R . H . M . STM sFit-1 KRYPTOR assays are supplied as two reagent kits that are run on the B-R-A-H-M-S KRYPTOR compact PLUS. Each kit is sufficient for 75 determinations, and is comprised of two bottled reagents:

PIGF

Cryptate conjugate: Lumi4®-Tb labeled, anti-human PlGF antibody, buffer, bovine albumin, rat immunoglobulins, bovine immunoglobulins, goat immunoglobulins, trehalose, mannitol.

XL conjugate: Cyanin 5.5 labeled, anti-human PlGF antibody, buffer, bovine albumin, rat immunoglobulins, bovine immunoglobulins, goat immunoglobulins.

sFlt-1

Cryptate conjugate: Lumi48-Tb labeled, anti-human sFlt-1 antibody, buffer, bovine albumin, bovine immunoglobulins, mouse immunoglobulins, dextran, preservative. XL conjugate: Cyanin 5.5 labeled, anti-human sFlt-1 antibody, buffer, bovine albumin, bovine immunoglobulins, mouse immunoglobulins, dextran, preservative.

The B · A · H · M · S KRYPTOR compact PLUS is an automated test platform. The analyzer consumables are:

B · R · A · H · M · S KRYPTOR compact Solution 1ProClinTM 150 Solution
B · R · A · H · M · S KRYPTOR compact Solution 2Potassium fluoride solution
B · R · A · H · M · S KRYPTOR compact Solution 3Active chlorine and sodium hydroxide solution
B · R · A · H · M · S KRYPTOR compact Solution 4Sodium hydroxide solution
B · R · A · H · M · S KRYPTOR BUFFERPhosphate Buffer Saline
B · R · A · H · M · S KRYPTOR compact REACT60 plates

Also supplied and required are calibrators - single level B.R.A.H.M.S PIGF plus KRYPTOR CAL and two levels B. R. A. H. M.S sFlt-1 KRYPTOR CAL. The instrument is calibrated on a 15 day cycle, and re-calibrated with each new lot of reagents.

Also supplied and required are three levels of controls - B.R.A.H.M.S PIGF plus KRYPTOR OC Controls and B.R.A.H.M.S sFlt-1 KRYPTOR OC.

B Principle of Operation

B R . A H . M . S PIGF plus KRYPTOR and B . R . A . H . STM sFit - 1 KR Y PTOR assays are immunofluorescent quantitative assays. The assays are formatted as homogenous immunoassays not requiring a free-bound separation step because of the detection method: referred to as Time-Resolved Amplified Cryptate Emission (TRACE) technology. For each assay, when antigen present in the patient sample is bound between the two antibodies forming an immunocomplex, and the sample subjected to a fluorescence excitation, a long lived fluorescent emission proportional to the analyte concentration is measurement of the fluorescence.

{3}------------------------------------------------

the intensity is compared against the standard calibration curve to derive the concentrations of sFlt-1 and PIGF in the sample. The concentration ratio for sFlt-1/PIGF is calculated and reported by the test system.

V Standards/Guidance Documents Referenced:

CLSI EP05-A3. Evaluation of Precision Performance of Quantitative Measurement Methods: Approved Guideline - Third Edition.

CLSI EP06, Evaluation of Linearity of Quantitative Measurement Procedures, 2nd Edition.

CLSI EP07, Interference Testing in Clinical Chemistry, 3rd edition.

CLSI EP17-A2, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline - Second Edition.

CLSI EP25-A, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline.

CLSI EP37. Supplemental Tables for Interference Testing in Clinical Chemistry, 1st edition.

EN ISO 17511:2020, In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples.

VI Performance Characteristics:

A Analytical Performance:

    1. Precision/Reproducibility:
      The precision performance of the B · R · A · H · M · S PIGF plus KR YPTOR assay, B · R · A · H · M · S sFlt-1 KRYPTOR assay, and the sFlt-1/PlGF ratio was evaluated following the guidelines of CLSI EP05-A3, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Third Edition.

PIGF

Study #1 - Repeatability and Within-Laboratory Precision: Single-Site Study

The purpose of this study was to determine the repeatability and within laboratory precision of the B.R.A.H.M.S PIGF plus KRYPTOR assay using the B.R.A.H.M.S KRYPTOR compact PLUS instrument. In the study, 17 serum samples (12 native and five spiked) and three controls were assayed with PIGF concentrations ranging from 8.7 to 6.040 pg/mL. Each sample was assayed in duplicates per run, two runs per day across 20 days for a total of 80 replicates: using three lots of PIGF reagent kits, three lots of calibrators, and one lot of controls. The testing was conducted by two operators using four B R A H M S KRYPTOR compact PLUS instruments. Repeatability and within laboratory precision were calculated by

{4}------------------------------------------------

SamplePIGF Mean(pg/mL)Repeatability%CVBetween Day%CVBetweenInstrument%CVWithinLaboratory%CV
Serum 18.714.60.04.315.2
Serum 210.612.30.06.013.7
Serum 310.612.30.04.613.1
Serum 411.211.80.04.012.5
Serum 517.27.60.04.08.6
Serum 620.76.00.03.67.0
Control 135.14.11.32.85.1
Serum 742.83.10.03.95.0
Serum 859.42.80.03.54.5
Serum 979.82.32.82.34.3
Control 21032.11.20.72.5
Serum 101471.72.62.54.0
Serum 113641.12.52.23.5
Control 34261.00.61.51.9
Serum 128170.93.32.34.1
Serum 139730.80.02.02.2
Serum 141,9060.80.02.22.4
Serum 153,3821.10.02.12.3
Serum 165,3791.00.72.52.8
Serum 176,0400.90.02.82.9

a two-way nested ANOVA with factors day and instrument as variance components. The following is a summary of the results:

The lot-to-lot precision was calculated by a two-way nested ANOVA with factors of lot and instruments as variance components. The following is a summary of the results:

SamplePIGF Mean (pg/mL)Between lot %CV
Serum 18.73.3
Serum 210.67.1
Serum 310.62.4
Serum 411.25.1
Serum 517.22.4
Serum 620.72.3
Control 135.10.0
Serum 742.83.0
Serum 859.42.5
Serum 979.83.9
Control 21030.8
Serum 101473.6
Serum 113643.0

{5}------------------------------------------------

Control 34260.0
Serum 128173.8
Serum 139730.0
Serum 141,9060.0
Serum 153,3820.0
Serum 165,3790.0
Serum 176,0400.4

Study #2 - Reproducibility: Multisite Study

A three site reproducibility study was conducted testing five samples (four native and one spiked), and using one lot of reagent kit, one lot calibrator, and one lot of controls. In the study, at each site, using one instrument, each sample was assayed in replicates of five per run, one run per day over five days, for a total of 25 replicates. Reproducibility was analyzed by two-way nested ANOVA with factors days and sites as variance components. The results are summarized below:

SamplePIGF MeanRepeatabilityWithin-LaboratoryReproducibility
(pg/mL)% CV% CV% CV
Serum 115.29.510.911.1
Serum 219.77.27.27.4
Serum 32381.41.72.6
Serum 48941.02.04.6
Serum 55,9391.52.23.6

Study #3 Reproducibility with intended use samples

A three instrument reproducibility study was conducted with samples representative of the intended use patient population. In the study, four samples of pooled serum obtained from pregnant women were assayed in replicates of five in one run per day over five days using three instruments for a total of 75 measurements per sample. The testing used one lot of B.R.A.H.M.S PIGF plus KRYPTOR. The results were analyzed for %CV for repeatability. within-laboratory, and reproducibility via two-way nested ANOVA with factors instrument and days as variance components. The results are summarized below:

SamplePlGF MeanRepeatabilityWithin-LaboratoryReproducibility
(pg/mL)%CV% CV% CV
Serum 1159.89.99.9
Serum 2552.93.74.5
Serum 31411.71.82.8
Serum 42351.42.12.9

sFit-1

Study #1 - Repeatability and Within-Laboratory Precision: Single-Site Study

The purpose of this study was to determine the repeatability and within laboratory precision of the B.R. A.H.M.S sFit-1 KRYPTOR assay using the B.R.A.H.M.S KRYPTOR compact

{6}------------------------------------------------

PLUS instrument. In the study, 16 samples (13 native and three spiked) were assayed with sFlt-1 concentration ranging from 48.1 to 68,944 pg/mL. Each sample was assayed in duplicates per run, two runs per day across 20 days for a total of 80 replicates; using three lots of PIGF reagent kits, calibrators, and one lot of controls. The testing was conducted by two operators using four B.R.A.H.M.S KRYPTOR compact PLUS instruments. Repeatability and within laboratory precision were calculated by a two-way nested ANOVA with factors day and instrument as variance components. The following is a summary of the results:

SamplesFlt-1Mean(pg/mL)Repeatability%CVBetween Day%CVBetweenInstrument%CVWithinLaboratory%CV
Serum 148.19.40.06.911.6
Serum 275.95.90.72.36.3
Serum 31024.71.72.25.5
Serum 42471.81.43.34.0
Serum 55180.91.42.53.0
Serum 61,0400.51.42.32.7
Control 11,5590.72.33.54.2
Serum 71,6050.61.42.22.7
Serum 82,9320.51.41.92.4
Control 23,0550.41.83.33.7
Serum 94,6480.61.31.92.4
Control 39,7720.62.32.63.5
Serum 1010,4530.61.22.42.8
Serum 1122,8320.72.41.63.0
Serum 1250,4520.92.52.33.5
Serum 1368,9440.72.33.03.8

The lot-to-lot precision was calculated by a two-way nested ANOVA with factors lot and instruments as variance components. The following is a summary of the results:

SamplesFlt-1Mean(pg/mL)Between lot%CV
Serum 148.10.0
Serum 275.91.6
Serum 31022.4
Serum 42473.3
Serum 55182.7
Serum 61,0402.3
Control 11,5594.4
Serum 71,6052.1
Serum 82,9321.6
Control 23,0553.7

{7}------------------------------------------------

Serum 94,6480.5
Control 39,7723.2
Serum 1010,4530.0
Serum 1122,8320.0
Serum 1250,4520.0
Serum 1368,9440.0

Study #2 - Reproducibility: Multisite Study

A three site reproducibility study was conducted testing five serum samples, and using one lot of reagent kit, one lot calibrator, and one lot of controls. In the study, at each site, using one instrument, each sample was assayed in replicates of five per run, one run per day over five days, for a total of 25 replicates. Reproducibility was analyzed by two-way nested ANOVA with factors days and sites as variance components. The results are summarized below:

SamplesFlt-1 Mean (pg/mL)Repeatability % CVWithin-Laboratory % CVReproducibility % CV
Serum 120.928.328.538.4
Serum 22632.33.43.5
Serum 31,6970.82.33.0
Serum 411,4011.73.23.3
Serum 540,0670.73.85.1

Study #3 Reproducibility with intended use samples

A three instrument reproducibility study was conducted with samples representative of the intended use patient population. In the study, five samples of pooled serum obtained from pregnant women were assayed in replicates of five in one run per day over five days using three instruments for a total of 75 measurements per sample. The testing used one lot of B.R. A.H.M.S sFit-1 KRYPTOR. The results were analyzed for %CV for repeatability, within-laboratory, and reproducibility via two-way nested ANOVA with factors instrument and days as variance components. The results are summarized below:

SamplesFlt-1 Mean(pg/mL)Repeatability% CVWithin-Laboratory% CVReproducibility% CV
Serum 11,2440.61.22.6
Serum 25,2250.61.02.0
Serum 36,8480.51.02.1
Serum 432,3050.92.42.6
Serum 528,6250.81.81.8

sFlt-1/PIGF ratio

Study #1 - Repeatability and Within-Laboratory Precision: Single-Site Study

The purpose of this study was to determine the precision and reproducibility performance of the sFlt-1/PlGF ratio. In the study, 14 samples were selected from the individual 20-day precision data of the sFlt-1 and PIGF assays (see above) with concentrations such that the

{8}------------------------------------------------

range of ratios reasonably encompassed that expected in the intended use population; which based on the clinical study ranged from 1.5 to 1146. Repeatability and within laboratory precision were calculated via a two-way nested ANOVA software with factors day and instruments as variance components. The following is a summary of the analysis:

Mean sFlt-1conc., pg/mLMean PIGFconc., pg/mLMeanRatioRepeatability%CVWithin-Laboratory%CV
16058172.01.2%5.7%
464819062.41.1%3.7%
5181473.51.7%5.4%
46488175.71.0%5.4%
1605147111.9%5.4%
22832817281.1%5.4%
4648147321.8%5.1%
160542.8383.4%6.3%
68944817851.2%6.0%
464842.81093.3%6.1%
228321471561.9%5.1%
689441474711.8%6.0%
2283242.85343.2%5.7%
6894442.816133.2%6.3%

Study #2 - Reproducibility: Multisite Study

The reproducibility of the sFit-1/PIGF ratio was evaluated by selecting 6 samples from the individual multisite data of the sFlt-1 and PIGF assays (see above). Reproducibility was analyzed via two-way nested ANOVA software with factors days and sites as variance components. The results are summarized below:

Mean sFlt-1conc., pg/mLMean PlGFconc., pg/mLMeanRatioRepeatability%CVWithin-Lab%CVReproducibility%CV
16972387.11.3%2.4%3.4%
1140189412.71.9%3.5%4.5%
4006789444.61.3%4.1%4.1%
1140123847.91.7%3.0%3.3%
169719.786.67.5%7.5%8.4%
400672381681.5%3.4%3.5%

2. Linearity:

The linearity performance of the B.R.A.H.M.S PIGF plus KRYPTOR and B.R.A.H.M.STM sFit-1 KRYPTOR run on the B.R.A.H.M.S KRYPTOR compact PLUS was established following the guidelines of CLSI EP06 Evaluation of Linearity of Quantitative Measurement Procedures, 2nd Edition.

PIGF

{9}------------------------------------------------

In the study, three separate panels that together cover the full measuring range were evaluated. A high concentration for each panel was prepared with recombinant PIGF antigen. The high concentration samples had expected concentrations at 4,424 pg/mL. 943 pg/mL. and 101 pg/mL for panels 1, 2, and 3, respectively. Known volumes of each respective high concentration sample in the panel were mixed with a low concentration sample (expected concentration of zero) to prepare different levels for linearity testing in each panel. With each linearity panel, the samples were measured in four replicates with one lot of reagent kit using one instrument. The results for each linearity panel were combined and analyzed by weighted linear regression to find the best fit straight line to the data with zero intercept; with expected concentrations based on the dilutions on the x-axis and average observed concentration on the y-axis. At each concentration, the deviation from linearity was calculated based on the difference between the observed concentration and the predicted concentration of the best fit straight line. The deviation from linearity at each level supports the claimed PlGF measuring range of 7.6 to 4.000 pg/mL.

sFlt-1

In the study. five separate panels that together covered the full measuring range were evaluated. A high concentration for each panel was prepared with recombinant sFlt-1 antigen. The high concentration samples had expected concentrations at 100,000 pg/mL, 50,000 pg/mL, 1,000 pg/mL, 1,000 pg/mL, and 200 pg/mL for panels 1, 2, 3, 4 and 5 respectively. Known volumes of each respective high concentration sample in the panel were mixed with a non-zero sample with an expected concentration of 19.2 pg/mL. With each linearity panel, the samples were measured in five replicates with one lot of reagent kit using one instrument. The results for each linearity panel were combined and analyzed by weighted linear regression to find the best fit straight line to the data with an intercept; with expected concentrations based on the dilutions on the x-axis and average observed concentration on the v-axis. At each concentration, the deviation from linearity was calculated based on the difference between the observed concentration and the predicted concentration of the best fit . straight line. The deviation from linearity at each level supports the claimed sFlt-1 measuring range of 315 to 90,000 pg/mL.

sFlt-1/PIGF Ratio

Not applicable

High dose hook effect

A study was conducted to assess whether PIGF and sFit-1 concentrations above the upper limit of the measuring range are reported with a value within the measuring range because of a hook effect when the immunoassay reagents are under conditions of antigen excess.

PIGF

In the study, 16 samples were prepared with recombinant PIGF antigen at target PIGF concentrations ranging from 40 to 1,600,000 pg/mL. Each sample was measured in duplicate with one reagent lot and one instrument. The results of the study support the sponsor's labeling claim that the B.R.A.H.M.S PIGF plus KRYPTOR exhibited no high dose hook effect at PIGF concentrations up to 800,000 pg/mL.

{10}------------------------------------------------

sFit-1

In the study, 17 samples were prepared with recombinant human sFlt-1 at target sFlt-1 concentrations ranging from 10 to 400,000 pg/mL. Each sample was measured in duplicate with one reagent lot and one instrument. The results of the study support the sponsor's labeling claim that the B.R.A.H.M.S sFit-1 KRYPTOR exhibited no high dose hook effect at sFlt-1 concentrations up to 400,000 pg/mL.

sFlt-1/PIGF Ratio

Not applicable

    1. Analytical Specificity/Interference:
      The analytical specificity performance of the B.R.A.H.M.S PlGF plus KRYPTOR and B · R · A · H · M · S · F · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · evaluated by testing for interference from endogenous and exogenous substances, and crossreactivity.

Interference from endogenous and exogenous substances

PIGF

In the study, two samples of either: (1) pooled human serum from healthy pregnant women spiked with PlGF at 40 and 500 pg/mL (50 and 60 pg/mL PlGF for folic acid testing, 50 and 130 pg/mL PIGF for iron testing) or (2) recombinant human PIGF diluted in defibrinated normal human plasma with PIGF at 30 and 400 pg/mL were prepared and divided into two aliquots; i.e. test sample (with added interferent) and control sample (with no added interferent). Sample testing was conducted with two lots of the reagent kit using three instruments (except for folic acid testing which used one lot and one instrument). Each test and control sample was assayed in replicates of three (except folic acid and iron testing which included five replicates). The data showed that the % interference for endogenous substances was between -9.3% and +10.8%, and between -6.1% and 15.4% for exogenous substances with the B.R.A.H.M.S PIGF plus KRYPTOR assay.

sFlt-1

In the study, two samples of pooled of human serum from healthy pregnant women spiked with sFlt-1 at 1500 and 10,000 pg/mL (2400 and 11,000 pg/mL sFit-1 for folic acid testing, 50 and 130 pg/mL sFlt-1 for iron testing) were prepared, and divided into two aliquots, i.e., test sample (with added interferent) and control sample (with no added interferent). Sample testing was done with one lot of reagent kit and using one instrument. Each test and control sample was assaved in replicates of three (except folic acid and iron testing which included five replicates). The data showed that the % interference for endogenous substances was between -10.8% and -0.6%, and between -1.8% and +2.2% for exogenous substances with the B.R.A.H.M.S PIGF plus KRYPTOR assay.

For each substance, the results were analyzed for the highest concentration of that substance tested that did not show significant interference. The results are summarized below for both analytes.

{11}------------------------------------------------

Interfering SubstanceHighest concentration of interferenttested that did not show significantinterference
Endogenous *
Albumin6.0 g/dL
Bilirubin25 mg/dL
Hemoglobin500 mg/dL
Triglycerides3000 mg/dL
Exogenous
Acetaminophen20 mg/dL
Acetylsalicylic Acid65.2 mg/dL
Ascorbic acid3.0 mg/dL
Caffeine5.9 mg/dL
Calcium20 mg/dL
Dihydralazin (Dihydralazine)20 mg/dL
Ethanol5% (v/v)
Folic Acid2.4 mg/dL
Gentamicin1.0 mg/dL
Heparin (intravenous)3000 U/L
Ibuprofen50 mg/dL
Iron (ferrous)72 mg/dL
α-Methyldopa1.5 mg/dL
Metoprolol0.5 mg/dL
  • Concentration based on the amount of substance added to the samples; actual concentrations of endogenous substances in the samples were higher since the native concentrations of these substances in the samples were not known.

sFlt-1/PIGF Ratio

In the study, four samples from healthy pregnant women spiked sFlt-1 were selected from the individual data from sFlt-1, and the same samples spiked with PIGF were selected to analyze for interference from each of albumin, bilirubin, hemoglobin, and triglycerides based on the sFlt-1/PlGF ratio.

SubstanceConcentrationTest sampleaverage ratioBlank sampleaverage ratio% Interference
Albumin6.0 g/dL38.644.0-12.4
21.121.5-1.7
3.23.2-0.3
252291-13.7
Bilirubin25 mg/dL44.443.81.3
21.420.82.9
3.23.20.1
2952834.1
Hemoglobin500 mg/dL38.64.7-11.8
20.521.5-4.8

{12}------------------------------------------------

3.13.3-5.3
256289-11.2
Triglycerides3000 mg/dL36.238.5-5.8
17.318.5-6.3
2.62.8-7.9
242253-4.2

Interference from potential cross-reactive substances

In the study, two samples of pooled human serum from pregnant women were prepared and divided into two aliquots, i.e., test sample (with added substance) and control sample (with no added substance). The concentrations tested for each substance were 10.000 pg/mL, 5,000 pg/mL, 1,000 pg/mL and 100 pg/mL. Sample testing was conducted with one lot of reagent kit for each analyte using two instruments. Each test and control sample was assayed in replicates of five replicates. For each substance, the results were analyzed to identify the highest concentration of that substance that did not show significant interference. The results are summarized below.

Interfering substanceHighest concentration of interferenttested that did not show significantinterference
PIGF-2100 pg/mL
PIGF-3100 pg/mL
VEGF-A (VEGF-165)5,000 pg/mL
VEGF-B10,000 pg/mL
VEGF/PIGF heterodimer10,000 pg/mL

sFit-1

Interfering substanceHighest concentration of interferent tested that did not show significant interference
PIGF-210,000 pg/mL
PIGF-310,000 pg/mL
VEGF-A (VEGF-165)5,000 pg/mL
VEGF-B10,000 pg/mL
VEGF/PIGF heterodimer10,000 pg/mL

sFlt-1/PIGF Ratio

Interfering substanceHighest concentration of interferenttested that did not show significantinterference
PIGF-2100 pg/mL
PIGF-3100 pg/mL
VEGF-A (VEGF-165)5,000 pg/mL
VEGF-B5,000 pg/mL

{13}------------------------------------------------

VEGF/PlGF heterodimer10,000 pg/mL
-------------------------------------

Cross-reactivity

PIGF

A cross-reactivity study was conducted to quantify the level of cross-reactivity of the PIGF assay with certain substances structurally similar to the antigen; endogenous growth factors that share amino acid sequences with PIGF. In the study, two test samples were prepared by spiking individually various cross-reactive substances into native serum samples containing PIGF. These samples were paired with two reference samples of the same serum that were not spiked and with the same PIGF concentration. Each sample was assayed in replicates of three using two reagent kit lots and three instruments. The results from each of the test samples and reference samples were averaged. The difference in results between test and reference samples was calculated for each of the PIGF concentrations, and each analyzed for % cross-reactivity using the below equation. The table summarizes the cross-reactivities from the study.

% Cross-Reactivity = (PIGF result from test sample - PIGF result from reference sample) x 100% / Concentration of cross-reactive substance

SubstanceSubstanceConcentrationTest sampleaverage PIGF,pg/mLControl sampleaverage PIGF,pg/mL% Cross-Reactivity
PIGF-210,000 pg/mL1,10136.110.6%
PIGF-210,000 pg/mL1,63649811.4%
PIGF-310,000 pg/mL37336.13.4%
PIGF-310,000 pg/mL8014983.0%
VEGF-A (VEGF 165)10,000 pg/mL35.434.60.01%
VEGF-A (VEGF 165)10,000 pg/mL4804740.06%
VEGF/PIGFHeterodimer10,000 pg/mL13134.60.96%
VEGF/PIGFHeterodimer10,000 pg/mL5344740.60%

sFit-1

SubstanceSubstanceConcentrationTest sampleaverage sFlt-1,pg/mLControl sampleaverage sFlt-1,pg/mL% Cross-Reactivity
PIGF-1 (PIGF)10,000 pg/mL1,5111,526-0.2%
10,06410,0610.0%
PIGF-210,000 pg/mL1,5071,526-0.2%
10,06510,0610.0%
PIGF-310,000 pg/mL1,5151,526-0.1%
10,05510,061-0.1%
VEGF-A (VEGF-165)10,000 pg/mL1,5151,526-0.1%
10,05710,0610.0%
VEGF-B10,000 pg/mL1,5011,526-0.3%
VEGF-B10,000 pg/mL1,501

{14}------------------------------------------------

9,97510,061-0.9%
VEGF/PIGFheterodimer10,000 pg/mL1,5151,526-0.1%
9,96810,061-0.9%

4. Assay Reportable Range:

The reportable range for the sFit-1/PIGF ratio is based on the analytical measuring range for the individual assays for sFlt-1 and PlGF. From the data collected during the clinical study (see section C.12), the range in sFlt-1/PIGF was found to be between 0.4 and 7367. The results of the individual assays for PIGF and sFlt-1 are not used for interpretation of results.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

Metrological Traceability

The metrological traceability of the B.R.A.H.M.S PIGF plus KRYPTOR and B.R.A.H.M.STM sFit-1 KRYPTOR run on the B.R.A.H.M.S KRYPTOR compact PLUS was established following the guidelines in EN ISO 17511:2020 In vitro diagnostic medical devices - Requirements for establishing metrological traceability of values assigned to calibrators, trueness control materials and human samples.

PIGF

The metrological traceability of the PIGF assay, through the calibrators was established using a commercially available reference preparation of PIGF that is considered by the sponsor as the highest reference for the assay. The sponsor's detailed metrological traceability was reviewed and found to be acceptable.

sFlt-1

The metrological traceability of the sFlt-1 assay, through the calibrators was established using a commercially available reference preparation of sFlt-1 that is considered by the sponsor as the highest reference for the assay. The sponsor's detailed metrological traceability was reviewed and found to be acceptable

sFlt-1/PIGF Ratio Not applicable

Stability

In support of specimen storage stability, two studies were conducted.

    1. Specimen stability: One day stability at room temperature and 2-8 ºC

PIGF and sFit-1

The product labeling describes that samples may be stored up to 1 day at room temperature and at 2 - 8°C. To validate sample specimen stability claims in the labeling, a short-term stability study was conducted with both serum and K2EDTA plasma. In the study, 25 fresh serum and 25 K2EDTA plasma samples collected from pregnant women

{15}------------------------------------------------

were stored under two conditions of 18-25℃ (room temperature) and 2-8℃. For each storge condition, the samples were tested in singlicate at the following time points:

Fresh (within 3 hours of blood collection) 24 hours (± 4.5 hours from sample collection) 48 hours (± 3.5 hours from sample collection)

The data at time points 24 and 48 hours was analyzed for stability by comparison to the reference time point of fresh. The stability results showed that PIGF and sFlt-1 are stable in serum and K2EDTA plasma up to 48 hours at 2-8°C and 48 hours at room temperature, which supports the labeling claim of one day stability at room temperature and 2-8 °C.

2. Specimen stability: Six month stability at -70°C

PIGF and sFit-1

To support the testing of clinical samples that were stored for up to six months in the PRAECIS clinical study, a separate stability study was conducted to establish that the storage conditions (-70℃) yield stable samples, and therefore the results of the clinical study are representative of intended use samples. In the study, 20 fresh serum and 20 K2EDTA plasma samples collected from pregnant women (representing the intended use population) were assaved for reference (time zero), and then stored at -70°C. The stored samples were tested in singlicate at the following time points: 6 months and 15 to 19 months (depending on the sample).

At each time point, each sample was tested in singlicate using one instrument and one reagent lot. The results for each sample were analyzed by linear regression across all three time points to derive an estimate percentage of change in concentration at month 6 versus time zero. A second analysis was conducted by directly calculating the percentage change in concentration between time zero and the different time points. The stability data showed that PIGF and sFlt-1 are stable in serum and EDTA plasma up to six months at -70°C which supports the use of the stored samples in the clinical study.

6. Detection Limit:

The detection capability of the B R . A . H . M . S PIGF plus KR Y PTOR and B . R . A . H . M . STM sFlt-1 KRYPTOR run on the B.R.A.H.M.S KRYPTOR compact PLUS was evaluated for limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) following the recommendations in CLSI EP17-A2.

PIGF

LoB

The LoB was evaluated using four zero-analyte samples. In the study, for each of three reagent lots, the four samples were each assayed in replicates of five per run over four runs with four different instruments per run for a total of 80 replicates per lot. The LoB was

{16}------------------------------------------------

analyzed for each lot by the non-parametric method of the CLSI guidance. The claimed LoB of the assay, selected from the highest value obtained across the three lots, is 2.5 pg/mL.

LoD

The LoD was evaluated using four pooled serum samples obtained from pregnant women representative of the intended use population with PIGF concentrations near the estimated detection limit. In the study, for each of three reagent lots, the four samples were each assayed in replicates of five per run over four runs with four different instruments per run for a total of 80 replicates per lot. The LoD was analyzed for each lot by the method of the CLSI guidance, whereby LoD is equal to LoB + Cp (SDL) where Cp is a multiplier to give the 95th percentile of the normal distribution and SDL is the pooled standard deviation of all lowlevel samples. The claimed LoD of the assay, selected from the highest value obtained across the three lots, is 4.9 pg/mL.

LoO

The LoQ was evaluated using 9 pooled serum samples with PIGF concentrations ranging from 5.3 to 18 pg/mL. In the study, for each of two instruments with two separate reagent lots, the 9 samples were each assayed in replicates of 2 per run, one run per day over 9 days for a total of 18 measurements per sample per instrument/lot. Across each of the two instruments/lots, for each sample the within-laboratory CV% was calculated. The LoQ was analyzed per instrument/lot to be the lowest analyte concentration with a within-laboratory precision no greater than 20%. The claimed LoQ for PlGF was based on the highest value from the two instruments/lots, which as 7.6 pg/mL.

sFlt-1

LoB

The LoB was evaluated using four zero-analyte samples. In the study, for each of three reagent lots, the four samples were each assaved in replicates of five per run over three runs with three instruments over six days for a total of 60 replicates per lot. The LoB was analyzed for each lot by the non-parametric method of the CLSI guidance. The claimed LoB of the assay, selected from the highest value obtained across the three lots, is 17.8 pg/mL.

LoD

The LoD was evaluated using four samples comprised of pooled human serum or diluted human serum, and sFlt-1 concentrations near the estimated detection limit. In the study, for each of three reagent lots, the four samples were each assayed in replicates of five per run over three runs with three instruments over six days for a total of 60 replicates per lot. The LoD was analyzed for each lot by the method of the CLSI guidance, whereby equal to LoB + Cp (SDL) where Cp is a multiplier to give the 95th percentile of the normal distribution and SDL is the pooled standard deviation of all low-level samples. The claimed LoD of the assay, selected from the highest value obtained across the three lots, is 28.5 pg/mL.

LoO

The LoQ was evaluated using five pooled normal human serum samples with sFlt-1 concentrations ranging from 12.6 to 47.2 pg/mL. In the study, for each of two instruments

{17}------------------------------------------------

with two separate reagent lots, the five samples were each assayed in replicates of 2 per run, one run per day over 7 days for a total of 14 measurements per sample per instrument/lot. Across each of the two instruments/lots, for each sample the within-laboratory CV% was calculated. The LoQ was analyzed per instrument/lot to be the lowest analyte concentration with a within-laboratory precision no greater than 20%. The claimed LoO for PlGF was based on the highest value from the two instruments/lots, which as 29.4 pg/mL pg/mL.

The summary results for LoB, LoD and LoQ are shown below.

AnalyteLoBLoDLoQClaimed Measuring Range
PIGF (pg/mL)2.54.97.67.6 to 7,000 pg/mL
sFlt-1 (pg/mL)17.828.529.475.3 to 90,000 pg/mL

sFlt-1/PIGF Ratio

Not applicable

7. Assay Cut-Off:

See Other Clinical Supportive Data in section C.12.

B Comparison Studies:

    1. Method Comparison:
      Not applicable.
    1. Matrix Comparison:
      A matrix equivalency study to serum was conducted to support use of the B.R. A. H. M.S PIGF plus KRYPTOR and B.R.A.H:M.S™ sFlt-1 KRYPTOR assays with K2EDTA plasma, as claimed in the product labeling.

PIGF and sFlt-1

In the study, the equivalence between serum and K2EDTA plasma samples was established by assaying 46 matched samples collected from pregnant women. Each sample was assaved in singlicate using one lot of reagent kit and one instrument on two runs over two days. The PIGF and sFit-1 concentration data obtained from the serum and K2EDTA plasma samples were plotted and analyzed by Passing-Bablok regression analysis. The slope and intercept of the regression line were calculated, and summarized as follows:

AnalyteRange (serum)SlopeInterceptCorrelation coefficient(Spearman)
PIGF88 - 1757 pg/mL0.9437.00.976
sFlt-1166 - 3400 pg/mL0.97131.00.974

{18}------------------------------------------------

The study results support the sponsor's claim that human serum and K2EDTA plasma are acceptable sample types to be used with the PIGF and sFit-1 assays.

sFlt-1/PIGF Ratio Not applicable.

C Clinical Studies:

    1. Clinical Sensitivity:
      See section C.12.

11. Clinical Specificity:

See section C.12.

12. Other Clinical Supportive Data:

The clinical performance of the B.R.A.H.M.S sFlt-1/ PIGF KRYPTOR Test System was evaluated in a clinical study titled PRAECIS: Preeclampsia Risk Assessment: Evaluation of Cut-offs to Improve Stratification. Identification and Validation of a Cut-off for the Ratio of Soluble Fms-like Tyrosine Kinase-1 to Placental Growth Factor (sFlt-1/PlGF) to Stratify Risk in Pregnant Women with Hypertensive Disorders of Pregnancy.

The study was a prospective, multicenter, blinded, non-interventional study conducted across 18 sites in the US. The purpose of the two-part study was the identification (part 1) and validation (part 2) of the cut-off for the ratio of sFlt-1 to PlGF used to prognose the risk of developing preeclampsia with severe features (sPE) in women hospitalized for a hypertensive disorder of pregnancy.

In the study (part 2), 520 women with singleton pregnancies between 23+0 and 34+6/7 weeks gestation hospitalized (or being hospitalized) for a hypertensive disorder of pregnancy were enrolled. Hypertensive disorders included: preeclampsia (PE), chronic hypertension (HTN) with or without superimposed PE or gestational hypertension, as defined by American College of Obstetricians and Gynecologists (ACOG) 2013 guidelines (Hypertension in Pregnancy. Obstet Gynecol. 2013 Nov; 122(5):1122-31.) and the 2019 update (Chronic Hypertension in Pregnancy, OBSTETRICS & GYNECOLOGY, VOL. 133, NO. 1, JANUARY 2019). Women who met the criteria for sPE at the time of enrollment were excluded. Women that met the following criteria were also not eligible for the study: withdrew patient consent, multiple fetus pregnancy, outside the gestational age range, or delivered before specimens could be collected. Patients with multiple hospitalizations were not excluded; 34 patients were enrolled twice, one patient was enrolled three times. The study population comprised 556 enrollments of 520 patients, and with the following races:

Race% (n)
American Indian or Alaskan Native0.38 (2)
Asian5.96 (31)

{19}------------------------------------------------

Black/African American29.62 (154)
Hawaii or Other Pacific Islander0.58 (3)
Other4.23 (22)
Unknown or Undeclared5.19 (27)
White/Caucasian54.04 (281)

For each patient enrolled into the study, samples (serum and K2EDTA plasma) were collected for determination of sFIt-1 and PIGF levels at baseline, and derivation of the sFlt-1/PlGF ratio. The samples were stored frozen (-70℃) and sent to a central laboratory for later testing as a batch.

All women enrolled in the study were followed for two weeks or until delivery within two weeks. The development of sPE within a two-week window of enrollment was the study endpoint/clinical outcome. A central adjudication committee (panel of 3 independent obstetricians) was assembled to review data and arrive at consensus regarding each patient's diagnoses for the development of sPE following the 2013 ACOG clinical practice guidelines.

The interpretation of results of the subject device based on the sFlt-1/PIGF ratio determined in part 1 were as follows:

High risk if sFlt-1/PlGF ratio > 40 Low risk if sFlt-1/PlGF ratio < 40

The severe preeclampsia status per adjudication panel and subject device serum results were analyzed based on the number of enrollments (n=556). The following is a summary of serum results.

Severe preeclampsia status
YesNoTotal
sFlt-1/PlGFHigh risk174True positive93False positive267
Low risk12False negative277True negative289
Sum186370556

Prevalence for developing sPE within two weeks in the clinical study = 33% (186/556).

Of the 556, results, there were 93 false positives and 12 false negatives.

The positive predictive value (PPV) and negative predictive value (NPV) are summarized below:

PPV (95% CI)NPV (95% CI)
65% (59.3, 70.6)96% (92.9, 97.6)

{20}------------------------------------------------

The positive likelihood ratio (PLR) and negative likelihood ratio (NLR) are summarized below:

PLR (95% CI)NLR (95% CI)
3.72 (3.11, 4.46)0.09 (0.05, 0.15)

The sensitivity and specificity are summarized below:

Sensitivity (95% CI)Specificity (95% CI)
94% (89.1, 96.3)75% (70.2, 79.0)

The clinical results were stratified into subgroups based on the four hypertensive disorders of pregnancy, gestation age (<30 weeks and > 30 weeks), and maternal age (18-34 years and 35-50 years). There was no significant difference in clinical performance across the hypertensive disorders of pregnancy, gestation age, or maternal age.

There was no significant difference in clinical performance when testing K2EDTA plasma samples in the same study.

D Clinical Cut-Off:

See section C.12 above.

E Expected Values/Reference Range:

In a population of 166 healthy pregnant women between week 23+0 up to week 34+6 gestation, the sFit-1/PIGF ratio median was established at 3.2, the 2.5th percentile at 0.8 and the 97.5% percentile at 31.

Other Supportive Performance Characteristics Data: F

Not applicable.

Proposed Labeling:

The labeling supports the decision to grant the De Novo request for this device.

Training:

The sponsor's training program documentation was reviewed and found acceptable as part of risk management activities.

Identified Risks and Mitigations:

{21}------------------------------------------------

Risks to HealthMitigation Measures
Incorrect performance of the test leading tofalse positive results.Certain design verification and validationactivities and documentation.
Certain labeling information, includingcertain limiting statements and performancecharacteristics.
Incorrect interpretation of test results.Certain design verification and validationactivities, including certain licensedpractitioner training as part of riskmanagement activities.
Certain labeling information, includingcertain limiting statements.

Benefit/Risk Assessment:

A Summary of the Assessment of Benefit:

Clinical benefit may be foreseen for the following reasons:

  • A true positive test could help risk stratify a patient who may benefit from increased . surveillance and preparedness for the possibility of developing preeclampsia with severe features (sPE).
  • Because progression to sPE can occur rapidly, heightened level of suspicion with a positive . test could aid in risk stratification that would allow for preparation and subsequent rapid intervention in the event that clinical circumstances dictate such intervention is necessary.
  • True positive patients may benefit from increased frequency of monitoring of their . condition by their healthcare provider and earlier step-up care (e.g .. referral to higher level hospital with level III or IV NICU). Such advanced knowledge of risk could be especially helpful in rural or community settings where anticipation of clinical deterioration could help in planning for transfers to higher levels of care.
  • With a higher level of suspicion, a clinician may keep a patient in the hospital for a longer . period of observation, as a positive test could indicate a high risk to go on to sPE.

B Summary of the Assessment of Risk:

The probable risks associated with the use of the B R . A . H . M . S sFlt-1 / PIGF KRYPTOR Test System are from undetected instances of false positive test results and/or misinterpretation.

Based on the reported clinical performance of the device in the clinical validation study, the positive predictive value is 65%, indicating that 35% of the time a patient with a positive result did not go on to develop preeclampsia with severe features, i.e., false positive result. The risk of

{22}------------------------------------------------

harm to the patient caused by a false positive result classifying the patient as high likelihood of progressing to sPE are due to inappropriate clinical decisions. Patients with a positive test result would be more closely monitored and preparation for possibility of progression to sPE could be initiated. However, no active intervention/treatment is to be undertaken based on a positive test result.

When used as intended, the device is a risk predictor for the future progression to sPE, and should not be misinterpreted to be used for the diagnosis of sPE, used for decisions of delivery or hospital discharge, or used for monitoring the patient's condition by repeated testing. The risk of harm to the patient caused by misinterpretation arises from appropriate clinical management interventions, such as health complications to the mother and baby, or lowering vigilant expectant management.

C Patient Perspectives:

This submission did not include specific information on patient perspectives for this device.

D Summary of the Assessment of Benefit-Risk:

The risks of false positive results are mitigated by the requirement of certain design verification and validation activities, including certain studies to ensure clinical and analytical performance. Certain labeling information, including limiting statements in the patient test reports and labeling, serve to reduce the risk of false positive results and/or incorrect interpretation of results.

When used as intended, general controls are insufficient, but in light of the special controls, the probable benefits of this device outweigh the probable risks of this device.

VII Conclusion:

The De Novo request is granted, and the device is classified under the following and subject to the special controls identified in the letter granting the De Novo request:

Product Code: OWH

Device Type: Prognostic test for development or progression of preeclampsia

Device Class: Class II with special controls.

Regulation: 21 CFR 862.1602

N/A