K Number
DEN220023

Validate with FDA (Live)

Device Name
Procise ADL
Manufacturer
Date Cleared
2023-09-29

(543 days)

Product Code
Regulation Number
862.3115
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Procise ADL assay is a time-resolved fluorescence energy transfer immunoassay for the quantitative determination of adalimumab (ADL) levels in venous serum in patients undergoing adalimumab therapy, using the ProciseDx Analyzer.

Measurements obtained by this assay can be used to detect adalimumab as an aid in the management of patients with inflammatory bowel diseases (IBD): Crohn's disease and ulcerative colitis being treated with adalimumab. The test is intended for use in a clinical laboratory.

Device Description

Each Procise ADL assay kit includes Procise ADL reagent cartridges, buffer bulbs, and Procise ADL low and high assay controls as follows:

  • . Twenty pouched Procise ADL cartridges each containing a lyophilized test-specific reagent bead located in the cartridge cap comprised of test-specific conjugates (monoclonal Fab anti-ADL/TNFa complex labeled with acceptor fluorophore and TNFa protein labeled with donor fluorophore)
  • Twenty 1.5 mL buffer bulbs .
  • . Two pouched assay ADL Low controls
  • . Two pouched assay ADL High controls
  • Product Insert .
  • . Quick Reference Guide

The Procise ADL assay requires the ProciseDx Analyzer. The ProciseDx Analyzer is designed to detect time-resolved fluorescent signal from both the donor and FRET acceptor emission within the Procise ADL assay.

AI/ML Overview

Acceptance Criteria and Study Proving Device Meets Criteria for Procise ADL Assay

The Procise ADL assay is a quantitative, time-resolved fluorescence energy transfer immunoassay for measuring adalimumab (ADL) levels in patient serum, intended to aid in the management of inflammatory bowel disease (IBD) patients undergoing adalimumab therapy. As the provided document is a De Novo Decision Summary from the FDA, it outlines the review of the analytical performance studies conducted by the manufacturer, rather than a clinical trial with acceptance criteria for clinical efficacy. Therefore, the "acceptance criteria" here refer to the predefined performance goals the analytical studies needed to meet for FDA authorization.

1. Table of Acceptance Criteria and Reported Device Performance

Performance Metric CategorySpecific MetricAcceptance Criteria (Implicit from Study Design/Results Presented)Reported Device Performance and How it Meets Criteria
Precision/ReproducibilityTotal %CV for within-laboratory precision (20-Day)Generally, %CV below a certain threshold (e.g., typically <10-15% for quantitative assays, varying with concentration). Specific thresholds not explicitly stated but inferred from passing results.Sample 1 (2.47 µg/mL): Total %CV = 10.6% Sample 2 (5.79 µg/mL): Total %CV = 6.8% Sample 3 (10.22 µg/mL): Total %CV = 6.1%Sample 4 (12.71 µg/mL): Total %CV = 7.3%Sample 5 (45.79 µg/mL): Total %CV = 5.2% All total %CVs are within acceptable limits for a quantitative assay.
Total %CV for inter-site reproducibilitySimilar to within-laboratory precision, acceptable %CV across multiple sites.QC Low (3.20 µg/mL): Total CV (%) = 7.0QC High (22.77 µg/mL): Total CV (%) = 7.2Sample 1 (5.18 µg/mL): Total CV (%) = 6.2Sample 2 (9.77 µg/mL): Total CV (%) = 6.1Sample 3 (44.32 µg/mL): Total CV (%) = 8.9 All total %CVs are within acceptable limits across multiple sites.
LinearityPercent deviation from linearityWithin ±5% across the reportable range.Native Samples: Percent deviation from linearity was within ±5% across the range tested (0.5 - 74.9 µg/mL). For the claimed measuring range of 1.0-50 µg/mL, the linear regression yielded a Slope of 0.98, Intercept of 0.28, and R^2 of 0.9995. This demonstrates excellent linearity across and beyond the claimed reportable range.
Analytical Specificity/InterferenceBias due to interferentsBias ≤10% for each potential interferent and concentration level tested compared to the nominal condition.None of the 21 tested substances (including common drugs, endogenous substances, and anti-drug antibodies to ADL and infliximab) showed significant interference (bias ≤10% at relevant concentrations, often at or above physiological/therapeutic levels). The assay demonstrates good analytical specificity, with limitations appropriately noted for highly hemolyzed or icteric samples.
High-Dose Hook EffectAbsence of hook effectResults unimpacted by a hook effect for concentrations up to 300 µg/mL.Serum samples tested at ADL concentrations of approximately 25, 50, 100, 200, and 300 µg/mL. For results above the upper limit of quantification (50 µg/mL), the analyzer displayed ">50 µg/mL". Device results were unimpacted by a hook effect up to 300 µg/mL. This confirms that very high ADL levels will not cause a falsely low reading within the tested range, providing a crucial safety feature.
Assay Reportable RangeDefined measuring rangeEstablish a range from 1.0 to 50.0 µg/mL based on linearity and other analytical performance.The assay reportable range was established as 1.0 to 50.0 µg/mL. This is supported by linearity, precision, and detection limit studies. The studies confirm the device's ability to accurately quantify ADL within this range.
TraceabilityTraceability to WHO IS StandardCalibration standards traceable to the ADL WHO International Standard (NIBSC code: 17/236). Recovery study to demonstrate functional accuracy.Calibration standards are traceable to the ADL WHO International Standard. A recovery study with the WHO IS Standard demonstrated good agreement: Slopes of 0.99-1.03 and R^2 of 1.00 for individual replicates. Individual percent biases ranged from -4% to 21%, with most at lower percentages, particularly away from the lower detection limits. This validates the accuracy and standardization of the assay's measurements.
Sample StabilityStability at various conditionsSupport for labeling claims regarding stability at room temperature, refrigerated, frozen (-80°C), and multiple freeze-thaw cycles.Room Temp/Refrigerated: Supported 3 days stability. Frozen (-80°C): Supported up to 142 weeks stability. Freeze-Thaw: Supported up to 5 freeze-thaw cycles. These studies ensure the reliability of results given various sample handling conditions in a clinical laboratory setting.
Detection LimitsLoB, LoD, LoQLoB, LoD, and LoQ determined based on CLSI EP17-A2 guidelines. For LoQ, total error <35%.LoB: 0.08 µg/mL LoD: 0.25 µg/mL LoQ: 0.96 µg/mL (with <35% total error). These values define the lower limits of reliable measurement for the assay.
Method ComparisonCorrelation with comparator methodsStrong correlation (e.g., Pearson r > 0.95 or 0.98) and acceptable agreement (slope close to 1, intercept close to 0) with validated comparator methods.N=61 samples compared to a representative comparator method. Slope = 0.91, Intercept = 1.08 µg/mL, Correlation Coefficient (r) = 0.994. Sample range tested: 2.15-42.7 µg/mL. This demonstrates excellent agreement and correlation with existing validated methods, indicating the new assay provides comparable results.
Biosimilar CompatibilityQuantitative equivalence in measuring Amjevita®Demonstrate quantitative equivalence in measuring the biosimilar drug Amjevita®. Linear regression coefficient of determination (R2) of 1.0.The assay demonstrated quantitative equivalence in measuring Amjevita® across seven concentrations (2.8 to 53.3 µg/mL). Individual % Bias ranged from -1% to 10% (excluding the highest point) and the linear regression coefficient of determination (R2) was 1.0. This confirms that the assay can accurately measure a key adalimumab biosimilar, broadening its clinical utility.

2. Sample Sizes and Data Provenance

  • Test Set (Analytical Studies):

    • Precision (20-Day): 240 measurements per sample level (5 levels), resulting from 2 replicates/day x 2 times/day x 20 days x 3 lots/analyzers.
    • Reproducibility (Multi-site): Replicates of 3, 2 times/day, for 5 days by 2 operators using 2 instruments at each of 3 external sites, for 5 serum samples. Total number of data points not explicitly summed but substantial.
    • Linearity (Spiked Samples): 13 different ADL levels.
    • Linearity (Native Samples): 11 serum levels.
    • Analytical Specificity/Interference: Each of 21 interferents tested at 2 ADL concentrations (5 and 25 ug/mL).
    • High-Dose Hook Effect: 5 ADL concentrations, tested in replicates of five using three assay lots and three buffer bulb lots.
    • Procise ADL Assay Recovery Study (WHO IS Traceability): 11 dilutions of WHO IS Standard, tested in triplicates.
    • Sample Stability:
      • Room Temperature & Refrigerated: 3 patients (low, mid, high range ADL), tested in duplicate at baseline, Day 3, Day 5, Day 7; used 3 lots.
      • Frozen Serum: 33 patients, tested within a day of draw, then retested in duplicate after storage at -80°C.
      • Freeze-Thaw: 5 frozen serum samples (~3-40 ug/mL ADL), tested after 1, 2, 3, and 5 freeze-thaw cycles.
    • Detection Limits (LoB, LoD, LoQ):
      • LoB: 4 native serum specimens (no ADL), 5 replicates over 3 days using 2 lots (total 60 measurements per lot).
      • LoD: 4 ADL serum samples (0.1-0.5 ug/mL), 5 replicates across 3 days using 2 lots (total 60 measurements per lot).
      • LoQ: 4-5 samples, 5 replicates across 3 days using 2 lots (total 60-75 measurements per lot).
    • Method Comparison: 62 deidentified serum samples.
    • Biosimilar Compatibility: 7 concentrations of Amjevita® and ADL control, tested in replicates of six.
  • Data Provenance: The document does not specify the country of origin for the patient samples. All studies are described as analytical performance studies, which are typically retrospective in nature, using banked or collected human serum samples rather than real-time prospective clinical trials.

3. Number of Experts and Qualifications for Ground Truth

  • Ground Truth for Analytical Studies: Not applicable in the same way as clinical studies. For quantitative in vitro diagnostic (IVD) devices like this, the "ground truth" for analytical performance tests is typically established by:
    • Known concentrations: For linearity, spike-in studies, and detection limit studies, precise dilutions of known standards (e.g., ADL WHO International Standard) are used.
    • Reference methods/comparators: For method comparison studies, the "ground truth" is established by results from already validated and established comparator methods.
    • Pooled samples / neat samples: For precision, stability, and interference, the "ground truth" is the established concentration of the pooled or neat human serum samples as measured by a reference method or confirmed by the manufacturer's internal processes.
  • Experts: No external experts (e.g., radiologists for imaging) are mentioned for establishing ground truth for these analytical performance studies. The studies are performed by laboratory personnel following established CLSI guidelines, implying the "experts" are the qualified laboratory scientists and statisticians who conduct and analyze the studies.

4. Adjudication Method for the Test Set

  • Adjudication methods (e.g., 2+1 consensus) are typically used in clinical studies, especially those involving subjective interpretation of data (e.g., image reading).
  • For quantitative analytical performance studies of an IVD, explicit "adjudication" of results is not generally performed in the same manner. The data is quantitative, and performance is assessed statistically against predefined criteria based on CLSI guidelines. Any discrepancies would be investigated through root cause analysis rather than expert consensus on individual results.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was done.
  • This device is a quantitative in vitro diagnostic assay for a measurand (adalimumab levels), not an AI-assisted diagnostic imaging device that involves human reader interpretation. Therefore, a MRMC study or assessment of human reader improvement with AI assistance is not applicable. The device provides a direct quantitative measurement.

6. Standalone (Algorithm Only) Performance

  • This is a standalone quantitative in vitro diagnostic device. Its performance is the algorithm's (assay's) performance. It does not involve human interpretation within its primary function where a human-in-the-loop scenario would be relevant for separate evaluation. The "ProciseDx Analyzer" is the instrument platform that runs the assay and displays the results.

7. Type of Ground Truth Used

The ground truth for the various analytical studies included:

  • Known concentrations/Spiking: For linearity, detection limits, high-dose hook effect, and biosimilar compatibility, samples were prepared with precisely known concentrations of ADL or ADL biosimilars by spiking into negative serum or by serial dilution from a concentrated stock.
  • Reference Standards: Specifically, the ADL WHO International Standard (NIBSC code: 17/236) served as the ultimate reference for traceability and accuracy verification.
  • Established Comparator Methods: For method comparison, previously validated and accepted clinical laboratory methods for measuring adalimumab served as the ground truth.
  • Native Patient Samples: Used for precision, stability, and interference studies, with their ADL concentrations determined by the assay itself or an established method as a baseline.

8. Sample Size for the Training Set

  • The document describes analytical performance studies for assay validation, not a machine learning model that requires a distinct "training set" of data.
  • For an IVD assay, the "training" aspect is inherent in the assay's development and optimization process (e.g., reagent formulation, instrument calibration, algorithm for calculating concentration from raw signal), which is an iterative process often using a combination of characterized biological samples, synthetic samples, and controls. The specific sample size for this development phase is not detailed in a regulatory summary focused on validation.

9. How the Ground Truth for the Training Set Was Established

  • Since there isn't a defined "training set" as in a machine learning context, the concept of establishing ground truth for it is not directly applicable.
  • The "ground truth" used during the development and optimization of such an assay would typically be established by:
    • Pharmacopoeial standards and reference materials: Highly characterized ADL raw materials and formulated drug products.
    • Gravimetric and volumetric preparations: Precisely prepared solutions and dilutions using analytical balances and calibrated pipettes to create samples with known ADL concentrations.
    • Well-characterized biological matrices: Pooled human serum from healthy donors or patients, characterized for background interference or spiked with known amounts of ADL.
    • Comparison to established research methods: During early development, comparison to gold standard research-level assays or mass spectrometry might be used to confirm initial assay performance.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine. The text is written in a clear, sans-serif font, with "FDA" in a blue square and the rest of the text in blue.

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR Procise ADL DECISION SUMMARY

I Background Information:

A De Novo Number

DEN220023

B Applicant

ProciseDx Inc.

C Proprietary and Established Names

Procise ADL

D Regulatory Information

ProductCode(s)ClassificationRegulationSectionPanel
QYDClass II21 CFR 862.3115 - Anti-tumornecrosis factor alpha monoclonalantibody test system forinflammatory bowel diseaseTX - ClinicalToxicology

II Submission/Device Overview:

A Purpose for Submission:

De Novo request for evaluation of automatic class III designation for Procise ADL

B Measurand:

Adalimumab (ADL)

C Type of Test:

Quantitative, Time-resolved fluorescence energy transfer immunoassay

{1}------------------------------------------------

III Indications for Use:

A Intended Use(s):

See Indications for Use below.

B Indication(s) for Use:

The Procise ADL assay is a time-resolved fluorescence energy transfer immunoassay for the quantitative determination of adalimumab (ADL) levels in venous serum in patients undergoing adalimumab therapy, using the ProciseDx Analyzer.

Measurements obtained by this assay can be used to detect adalimumab as an aid in the management of patients with inflammatory bowel diseases (IBD): Crohn's disease and ulcerative colitis being treated with adalimumab. The test is intended for use in a clinical laboratory.

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

For in vitro diagnostics use only

D Special Instrument Requirements:

ProciseDx Analyzer

IV Device/System Characteristics:

A Device Description:

Each Procise ADL assay kit includes Procise ADL reagent cartridges, buffer bulbs, and Procise ADL low and high assay controls as follows:

  • . Twenty pouched Procise ADL cartridges each containing a lyophilized test-specific reagent bead located in the cartridge cap comprised of test-specific conjugates (monoclonal Fab anti-ADL/TNFa complex labeled with acceptor fluorophore and TNFa protein labeled with donor fluorophore)
  • Twenty 1.5 mL buffer bulbs .
  • . Two pouched assay ADL Low controls
  • . Two pouched assay ADL High controls
  • Product Insert .
  • . Quick Reference Guide

The Procise ADL assay requires the ProciseDx Analyzer. The ProciseDx Analyzer is designed to detect time-resolved fluorescent signal from both the donor and FRET acceptor emission within the Procise ADL assay.

B Principle of Operation

The Procise ADL assay is a sandwich immunoassay that uses time-resolved fluorescence to detect the presence and quantity of ADL in patient serum specimens. It is a homogenous assay

{2}------------------------------------------------

that uses an energy transfer between a terbium cryptate acceptor fluorophore labeled anti-ADL/TNFa complex Fab' antibody and a terbium cryptate donor fluorophore labeled to TNFa protein. When ADL is present in a sample and tested with the Procise ADL assay, it binds to the donor labeled TNFa protein allowing the anti-ADL/TNFa Fab' antibody bound to an acceptor to bind.

Once the labeled TNFa protein and anti-ADL/TNFa Fab' antibody are bound together within a complex, their close proximity allows for fluorescence resonance energy transfer (FRET) to occur. The acceptor fluorophore emission created from FRET is measured along with the donor signal. The ratio of the two emissions is used by the ProciseDx Analyzer to determine the concentration of ADL within the sample. The acceptor to donor ratio is proportional to the amount of ADL in the sample.

V Standards/Guidance Documents Referenced:

CLSI EP05-A3, 3rd Edition, 2015, Evaluation of Precision of Quantitative Measurement Procedures, Approved Guideline

CLSI EP06 2nd Edition, 2021, Evaluation of the Linearity of Quantitative Measurement Procedures

CLSI EP07 3rd Edition, 2018, Interference Testing in Clinical Chemistry

CLSI EP09c 3rd Edition, 2020, Measurement Procedure Comparison and Bias Estimation Using Patient Samples

CLSI EP17-A2, 2013, Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline

CLSI EP25-A, 2013, Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline.

CLSI EP37 1st Edition, 2018, Supplemental Tables for Interference Testing in Clinical Chemistry

CLSI EP32-R, 2014, Metrological Traceability and Its Implementation; A Report

VI Performance Characteristics:

A Analytical Performance:

    1. Precision/Reproducibility:
      Precision studies were conducted following the recommendations in CLSI EP05-A3 guideline.

Precision studies were performed using five pools of native ADL serum samples with targeted ADL concentrations at approximately 2.8, 6, 10, 14, and 45 ug/mL. Three lots of Procise ADL assay reagents were paired with three lots of buffer bulbs using three different

{3}------------------------------------------------

instruments. The paired assay lot and instruments were changed from the first daily run compared to the second. Samples were tested in replicates of two, two times per day, for twenty days. The precision results characterize the precision of the Procise ADL assay across the measurement range and are summarized below for all lots.

SampleNMean($\mu$g/mL)BetweenLotBetweenAnalyzerBetweenDayBetween RunWithinRunTotal
SD%CVSD%CVSD%CVSD%CVSD%CVSD%CV
12402.470.124.8%0.124.9%0.052.2%0.010.6%0.197.8%0.2610.6%
22405.790.264.6%0.111.9%0.071.2%0.020.3%0.264.5%0.406.8%
324010.220.262.6%0.212.0%0.100.9%0.040.4%0.515.0%0.626.1%
424012.710.292.3%0.171.3%0.141.1%0.060.5%0.856.7%0.937.3%
524045.790.491.1%0.250.5%0.180.4%0.170.4%2.315.0%2.395.2%

Procise ADL Assay Precision (20-Day) for All Reagent Lots and Analyzers

A reproducibility study was performed at three external sites. Reproducibility studies were performed using five serum samples consisting of three pools of native serum samples with ADL concentrations approximately 5, 10, and 40 ug/mL and two serum quality control samples with ADL concentrations of 3.2 and 22.8 ug/mL. The same Procise ADL assay lot was used at all three sites. At each site, samples were tested in replicates of three, two times per day, for five days by two operators using two instruments. The reproducibility results characterize the precision of the Procise IFX assay across the measurement range and are summarized below for all lots.

SampleMean(µg/mL)WithinRunBetweenRunBetweenDayBetweenInstrumentBetweenOperatorBetweenSiteTotal
SDCV (%)SDCV (%)SDCV (%)SDCV (%)SDCV (%)SDCV (%)SDCV (%)
QC Low3.200.206.20.041.30.080.30.031.00.061.90.072.10.227.0
QC High22.771.345.90.492.10.351.50.251.10.522.30.411.81.637.2
15.180.244.60.061.20.061.20.091.60.132.50.122.30.326.2
29.770.535.40.101.00.111.10.111.10.161.60.151.50.606.1
344.323.277.30.801.80.761.70.581.31.052.41.503.43.958.9

Procise ADL Assay Reproducibility for All Sites

2. Linearity:

Linearity studies were conducted following the recommendations in CLSI EP06 2nd edition guideline.

{4}------------------------------------------------

A linearity study was conducted to evaluate linearity across the measuring range of the Procise ADL assay. Serum samples with 13 different ADL levels were evaluated: 0.5, 0.8, 1.4. 2.1. 3.1. 4.6. 6.9. 10.2. 15.9. 23.0. 34.3. 51.6 and 74.9 ug/mL. A native serum sample with ADL concentrations of 18.1 ug/mL was spiked with ADL Intermediate Dilution (PN4707) ug/mL) to obtain a concentration of 84.0 ug/mL. This was mixed with a native serum pool with no ADL to achieve the 13 different levels of ADL tested. The linear regression results are shown below. The percent deviation from linearity was within ±5% across the reportable range of the assay.

Claimed MeasuringRangeSample RangeTestedSlopeInterceptR2
1.0-50 µg/mL0.5-74.9µg/mL0.980.280.9995

These results support the claimed measuring range of 1.0 to 50 ug/mL for ADL.

A second linearity study was conducted with native samples. A pool of negative native serum was mixed in known ratios with a single ADL native serum sample with an ADL concentration of 28.1 ug/mL measured by the Procise ADL assay to create eleven serum levels known relative to one another. The percent deviation from linearity was within ±5% across the range of the assay tested.

3. Analytical Specificity/Interference:

Interference studies were conducted following the recommendations in CLSI EP37 guideline.

Each potentially interfering substance was prepared at twice the CLSI recommended level in pooled ADL negative serum which was then combined at a 1:1 ratio with native ADL serum sample pools to obtain ADL serum concentrations at approximately 5 and 25 ug/mL plus the interferent. The control samples without interferent were made by combining the native ADL serum at each level with ADL negative serum at a 1:1 ratio. None of the substances in the tables below showed significant interference, defined as bias ≤10% for each potential interferent and concentration level tested when compared to the nominal condition.

List of interferents at their concentration up to which no interference was observed.

CompoundTested concentration
5-aminosalicylate2.04 mg/dL
6-mercaptopurine0.148 mg/dL
Acetaminophen15.6 mg/dL
Acetylsalicylic Acid3 mg/dL
Antidrug antibodies to ADL200 ng/mL
Ascorbic Acid5.25 mg/dL
Azathioprine0.258 mg/dL
Bilirubin Conjugated20 mg/dL
Bilirubin Unconjugated40 mg/dL
Budesonide0.0146 µmol/L
Ciprofloxacin1.2 mg/dL

{5}------------------------------------------------

CompoundTested concentration
Hemolysate800 mg/dL
Human Anti-Mouse Antibody200x
Infliximab20 ug/mL
Methotrexate2000 umol/L
Metronidazole12.3 mg/dL
Prednisone0.276 umol/L
Rheumatoid Factors1285 IU/mL
Sulfasalazine7.5 mg/dL
Total Protein15 g/dL
Triglycerides1500 mg/dL
Vitamin D300 ng/mL

The sponsor included the following limitation in the labeling:

Note: Due to possibility of introducing error, highly hemolyzed (>800 mg/dL hemolysate) or highly icteric (>20 mg/dL conjugated bilirubin) samples should not be tested in the Procise ADL assay.

High-Dose Hook Effect

To evaluate the potential for a high dose hook effect in the Procise ADL assay, serum samples were tested at ADL concentrations of approximately 25, 50, 100, 200 and 300 ug/mL. The samples were tested in replicates of five using three assay lots and three buffer bulb lots. For results above the Procise ADL assay upper limit of quantification (50 ug/mL), the ProciseDx Analyzer displayed >50 ug/mL as a result. Device results are unimpacted by a hook effect at ADL concentrations of up to 300 ug/mL.

    1. Assay Reportable Range:
      The assay reportable range is from 1.0 to 50.0 ug/mL
    1. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

Traceability

The Procise ADL assay calibration standards are traceable to the ADL WHO International Standard: 1st International Standard for Adalimumab (NIBSC code: 17/236).

Procise ADL Assay Recovery Study

A study was performed to assess the functional accuracy of the Procise ADL assay calibration and control methodology. 50 ug of the WHO IS Standard for ADL was reconstituted using 1.0 mL of pooled native negative serum to achieve a known starting concentration of 50 ug/mL. Native ADL negative serum was used to dilute the starting sample to create a total of 11 samples, each one-third less than the previous. These 11 dilutions were tested in triplicates using one lot of the Procise ADL assay components in singlicate. The results are summarized below individually for three replicates.

{6}------------------------------------------------

AnalyteReplicateSlopeY-InterceptR2
ADL11.030.111.00
20.990.291.00
31.030.051.00

Procise ADL Result vs. WHO IS Standard for ADL Linear Regression Results

Procise ADL Result Summary When Testing the WHO IS Standard for ADL

WHO IS[ADL](µg/mL)ReplicateProcise[ADL](µg/mL)% Bias%CV
50.0151.33%3%
248.6-3%
350.71%
33.3135.05%1%
234.54%
335.46%
22.2122.62%2%
223.45%
322.62%
14.8115.55%2%
214.91%
314.80%
9.919.4-4%4%
210.01%
310.34%
6.616.71%3%
26.94%
36.5-2%
4.415.013%6%
24.41%
34.52%
2.913.04%2%
23.17%
33.04%
2.012.214%8%
22.04%
31.9-3%
1.311.621%11%
21.33%
31.30%

{7}------------------------------------------------

Sample Stability

Room Temperature and Refrigerated

To test the stability of serum specimens at room temperature or at 4℃, the sponsor performed a stability study using freshly drawn serum from three patients (targeting low, mid, and high range ADL concentrations) receiving ADL therapy. Refrigerated (~4°C) and room temperature (~22°C) specimens were tested in duplicate at Baseline, defined as 2-4 hours after draw (Day 0). Day 3. Day 5. and Day 7 using three different lots of Procise ADL reagents. The results support the labeling claim of three days of stability at room temperature or refrigerated (4℃).

Frozen Serum Stability

The sponsor performed a frozen serum stability study to demonstrate that serum samples stored at -80% that contain ADL are stable when measured with the Procise ADL assay. The sponsor performed the stability study using freshly drawn serum from 33 patients, tested by using the Procise ADL assay within a day of draw. Samples were stored frozen at -80℃ then thawed and retested in duplicate using one lot of Procise ADL assay reagents. The results show that serum samples with ADL stored at -80°C are stable up to 142 weeks.

Freeze-Thaw Serum Stability

The sponsor performed a freeze-thaw study to test the stability of frozen clinical serum specimens stored at -80°C that contain ADL when the samples undergo multiple freeze-thaw cycles. The sponsor tested five frozen serum samples with ADL concentrations of ~3, 5, 7, 20, and 40 ug/mL, previously measured using the Procise ADL assay when freshly collected. Each sample was tested after 1, 2, 3, and 5 freeze-thaw cycles and the ADL results were compared back to the original pre-frozen value. The results show that ADL is stable in serum samples frozen at -80°C that have undergone up to five freeze-thaw cycles.

6. Detection Limit:

Detection limits were assessed following the recommendations in CLSI EP17-A2 guideline.

The limit of blank (LoB) was determined using four native serum specimens containing no ADL. Each of the four serum samples were tested in replicates of five over three days using two Procise ADL assay lots for a total of 60 measurements per reagent lot. The 95th percentile from each reagent lot was calculated, multiplied by the standard deviation of the blank and the result was added to the mean of the blank to calculate the LoB for each lot. The sponsor determined the LoB to be 0.08 ug/mL which was the highest LoB calculation between the two Procise ADL assay lots.

The limit of detection (LoD) was determined using four different ADL serum samples with the concentrations: 0.1. 0.2, 0.4 and 0.5 ug/mL. Each of the four serum samples were tested in replicates of five across three days using two Procise ADL assay reagent lots for a total of 60 measurements per lot. Two operators performed the testing. A one-sided 95% confidence t-distribution table (with a sample size of 60) was used to calculate a multiplier. To calculate the LoD, this multiplier was multiplied by the pooled standard deviations for 60 data points for each assay lot then the result was added to the LoB. The sponsor determined the LoD to be 0.25 ug/mL using the assay lot with the highest LoD determination.

{8}------------------------------------------------

The limit of quantitation (LoQ) was determined using four to five (depending on the Procise ADL assay reagent lot) samples above the LoD with ADL concentrations at: 0.7, 0.9, 1.1 and 1.4 ug/mL (Lot 1) and 0.4, 0.5, 0.7, 0.9 and 1.1 ug/mL (Lot 2). Each of the four serum samples were tested in replicates of five across three days using two Procise ADL assay reagent lots for a total of 60-75 measurements per lot. Two operators performed the testing. The LoQ was determined to be 0.96 ug/mL at which they observed <35% total error.

    1. Assay Cut-Off:
      See Assay Reportable Range section above.
    1. Accuracy (Instrument):
      See Traceability and Method comparison sections.
    1. Carry-Over:
      Not applicable, the Procise ADL assay cartridge is single use.

B Comparison Studies:

    1. Method Comparison:
      A method comparison study was conducted comparing the Procise ADL assay to two validated comparator methods. A total of 62 deidentified serum samples were tested with the Procise ADL assay and the results were compared to results from Comparators 1 and 2. Data was analyzed using weighed Deming regression and Pearson correlation. The results for a representative comparator method are summarized below.

Procise ADL vs. Comparator

NSlopeIntercept(µg/mL)CorrelationCoefficient (r)Sample range tested(µg/mL)
610.911.080.9942.15-42.7

2. Matrix Comparison:

Not applicable. Not applicable. Serum is the only matrix claimed for the Procise ADL assay.

C Clinical Studies:

    1. Clinical Sensitivity:
      Not applicable
    1. Clinical Specificity:
      Not applicable

{9}------------------------------------------------

3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Clinical therapeutic drug monitoring (TDM) for adalimumab is based on the understanding that there is a relationship between drug exposure and clinical outcomes when treating patients with IBD with therapeutic biologics. Inter-individual variability exists in the rate of clearance of therapeutic biologics, including for adalimumab; therefore, having information that helps clinicians understand whether a patient is clearing the drug faster than expected provides information that aids clinicians in the management of their patients. The Procise ADL assay is intended to quantitatively detect the presence (or absence) of adalimumab in patients with IBD receiving adalimumab therapy. Published literature and professional society practice guidelines on the use of TDM in patients with IBD receiving adalimumab support that the device output (i.e., detecting the presence of adalimumab) provides information that is clinically meaningful to aid healthcare providers in the management of patients with IBD.

In patients with IBD who have active disease, detecting the presence or absence of circulating drug is useful to aid clinicians in clinical decision-making. As noted in the U.S. clinical practice guideline "American Gastroenterological Association Institute Guideline on Therapeutic Drug Monitoring in Inflammatory Bowel Disease" (2017), failure of treatment with adalimumab is generally due to one of two possibilities: 1) mechanistic failure or 2) pharmacokinetic failure. Pharmacokinetic failure occurs when therapeutic levels of drug are not achieved/maintained. Pharmacokinetic failure can occur via either immune or nonimmune mediated pathways. In immune-mediated PK failure, anti-drug antibody formation results in increased drug clearance and reduced or undetectable levels of the drug.

Regardless of the cause of the undetectable drug level, understanding whether a patient has circulating drug present is informative, as therapeutic proteins can exert beneficial clinical effects only when circulating at concentrations that allow interaction between the antibody and the target receptor, leading to downstream pharmacodynamic effects.

In summary, there is available evidence to support that the "detectable" or "not detectable" quantitative output from the Procise ADL assay will provide meaningful information to clinicians to aid in determining whether a lack of therapeutic response may be due, at least in part, to lack of circulating drug. Including the measured quantitative level can help the physician put the result into context.

D Clinical Cut-Off:

Not applicable

E Expected Values/Reference Range:

Not applicable.

{10}------------------------------------------------

Procise ADL Assay Biosimilars

To demonstrate that the Procise ADL assay can accurately detect an Adalimumab biosimilar drug (Amjevita), a pool of ADL negative human serum was used to create seven different levels each of Adalimumab and Amjevita: 2.5, 5, 10, 20, 30, 40, and 50 µg/mL. Each biosimilar concentration, as well as the ADL control, was tested in replicates of six. The results are summarized below.

ADLBiosimilarExpected [ADL]µg/mLObserved[ADL biosimilar]µg/mL (n=6)% Bias
Amjevita®53.358.710%
44.445.42%
33.132.8-1%
22.023.46%
10.811.46%
5.55.5-1%
2.82.91%
Procise ADL Serum Biosimilar Drug Results
-------------------------------------------------------

The data show that the Procise ADL assay demonstrates quantitative equivalence in measuring Amjevita®. The sponsor included the following statements in the labeling:

Procise ADL is validated to monitor drug levels of biological drugs which contains the active substance Adalimumab, that is the original drug Humira® and biosimilar drug Amjevita®. Additional biosimilar drugs have not been validated.

Compatibility with the Procise ADL Assay was confirmed for biosimilar drug Amjevita® with at seven (7) concentrations ranging from 3 to 50 ug/mL. Serum has a linear regression coefficient of determination (R2) of 1.0 for the biosimilar.

VII Proposed Labeling:

The labeling supports the decision to grant the De Novo request for this device.

{11}------------------------------------------------

VIII Identified Risks and Mitigations:

Risks to HealthMitigation Measures
Incorrect test resultsCertain design verification and validationactivities and documentation, includingcertain studies.Certain labeling information, includingcertain limiting statements.
Incorrect interpretation of test resultsCertain design verification and validationactivities and documentation, includingcertain studies.Certain labeling information, includingcertain limiting statements.

IX Benefit/Risk Assessment:

A Summary of the Assessment of Benefit:

There are currently no devices with FDA marketing authorization for determining adalimumab blood concentrations. The accuracy of the device is adequate to support clinical benefit when the device is used solely as noted in the indications for use.

In general, there is a risk of anti-drug antibody (ADA) development during treatment with therapeutic proteins, including adalimumab. Although the approved dosing regimen is intended to maintain an acceptable level of circulating drug, development of anti-drug antibodies, or other intrinsic factors affecting clearance, may lead to a reduced drug level or absence of detectable drug level during treatment. The clinical benefit of the Procise ADL assay is that it can quantitatively detect the presence (or absence) of adalimumab. This can help determine whether a lack of therapeutic response is due, at least in part, to lack of detectable drug. Providing the "detectable" or "not detectable" quantitative output will provide useful information to physicians and including the measured level can help them put this result into context.

B Summary of the Assessment of Risk:

The risks with use of the device are associated with misclassification due to incorrect test results (i.e., falsely high and falsely low test results) and incorrect interpretation of results such as interpreting a particular target trough concentration as optimal when there is not sufficient data to support this claim.

C Patient Perspectives:

This submission did not include specific information on patient perspectives for this device.

D Summary of the Assessment of Benefit-Risk:

The clinical benefit of the Procise ADL assay is having a determination of the presence or absence of adalimumab in IBD patients, with the quantitative level provided for context. In

{12}------------------------------------------------

patients with active disease, understanding whether they have detectable trough levels is informative for clinicians to aid in clinical management. Development of anti-drug antibodies can occur at any time during treatment and can increase drug clearance and lead to an undetectable trough level, which may impact the efficacy of the drug.

When the device is used as intended, the risks are mitigated by the Procise ADL assay indications for use, which do not propose therapeutic or reference ranges or refer to any action a clinician should take, such as dose adjustment, and the special controls. The output of the Procise ADL assay is intended to be used as one piece of information, taking into consideration many other factors such as the patient's clinical status and other laboratory test values. The device provides information regarding whether a patient has circulating drug at the time of testing. It is not intended to be used to make treatment decisions/change in medical management in isolation The indications for use, along with special controls which require appropriate device verification and validation testing to support all clinical and analytical claims, are sufficient to mitigate the risks associated with misclassification due to incorrect test results (i.e., falsely high and falsely low test results) and incorrect interpretation of results.

Device design verification and validation activities, including studies to support all analytical claims, clinical claims, and testing environments to ensure acceptable clinical and analytical performance, as well as certain labeling information will help ensure that the device functions as intended and mitigate the risk of incorrect test results (i.e., falsely low or falsely high test results) or the incorrect interpretation of the test results.

While general controls are insufficient to mitigate the risks of the device, the probable benefits outweigh the probable risks for the Procise ADL assay, considering the indications for use and the mitigation of the risks provided by the special controls. Overall, the probable benefits outweigh the probable risks of incorrect test results or incorrect interpretation of test results for the proposed indications for use, in light of the special controls and general controls.

X Conclusion:

The De Novo request is granted and the device is classified under the following and subject to the special controls identified in the letter granting the De Novo request:

Product Code(s): OYD Device Type: Anti-tumor necrosis factor alpha monoclonal antibody test system for inflammatory bowel disease Class: II (Special Controls) Regulation: 21 CFR 862.3115

N/A