K Number
DEN200040

Validate with FDA (Live)

Device Name
OMNIgene GUT Dx
Manufacturer
Date Cleared
2021-11-03

(506 days)

Product Code
Regulation Number
866.2952
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

OMNIgene. GUT Dx is intended for the non-invasive collection of human fecal samples and the stabilization of DNA from the bacterial community for subsequent assessment of the microbiome profile by an assay validated for use with OMNIgene GUT Dx.

Device Description

The OMNIgene . GUT Dx device consists of a collection tube with a tube top and pusher cap with a screw seal, along with a spatula or spoon for transferring fecal specimen into the collection tube. The tube contains 2 mL of the stabilizing liquid and a stainless-steel mixing ball. These components are intended to stabilize bacterial DNA in human fecal specimens. notably to preserve the relative abundances of bacterial organisms, for potential downstream analysis of the fecal microbiome. The collection device is designed for storage of fecal specimens at room temperature (20-26°C/68-79°F) for up to 30 days.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the studies conducted to prove the OMNIgene.GUT Dx device meets these criteria, based on the provided text:

Acceptance Criteria and Reported Device Performance

The device's performance was evaluated through several analytical studies focusing on DNA yield, microbial community neutrality (preservation of relative abundances), and stability at room temperature, as well as reproducibility.

Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance CriteriaReported Device PerformanceStudy Section
DNA Yield≥ 120 ng (0.12 µg) per extraction aliquot in ≥ 95% of samples (Neutrality Study)100% of specimens met the DNA yield acceptance criteria of 120 ng (0.12 µg) for both adult and pediatric cohorts. (Neutrality Study Results)L.1.b, Neutrality results; L.1.c, Reproducibility results
≥ 120 ng (0.12 µg) per extraction aliquot in ≥ 95% of samples (Room Temperature Storage Stability Study)100% of specimens met the DNA yield acceptance criteria of 120 ng (0.12 µg) for both adult and pediatric cohorts. (Room Temperature Storage Stability Results)
≥ 120 ng (0.12 µg) per extraction aliquot in ≥ 95% of samples (Reproducibility Study)100% of specimens for both lot-to-lot and aliquot-to-aliquot reproducibility met the DNA yield acceptance criteria of 120 ng (0.12 µg).
Microbial Community Neutrality (Relative Abundance Preservation)Per donor, each Microbiome Panel species in the stored sample has read counts within the TeSaRCol (Test Sample Read Count Limit) range as calculated using the unstabilized Control Baseline. Neutrality demonstrated in ≥ 90% of donors that were successfully screened to contain that species.All species examined as part of the Microbiome Panel demonstrated neutrality in ≥ 90% of donors between unstabilized and OMD-200 fecal samples. Whole Microbiome Analysis: Group wise comparison of Aitchison distance showed change in microbial profile was significantly lower between OMD-200 collected fecal samples at baseline and unstabilized control fecal samples than observed donor-to-donor differences (P < 0.0001). No taxa had a differential abundance exceeding measured within-group variance (effect size < 1, adjusted P-value > 0.1). PCA and hierarchical clustering showed intra-donor clustering.L.1.b, Neutrality Acceptance Criteria; Neutrality Results; L.1.e, Neutrality (Whole Microbiome) Results
Room Temperature Storage StabilityPer donor, each Microbiome Panel species in the stored sample has read counts within the TeSaRCol range as calculated using the OMD Baseline. Stability demonstrated in ≥ 90% of donors that were screened to contain that species.All species examined as part of the Microbiome Panel demonstrated stability in ≥ 90% of donors when stored at room temperature for up to 30 days either with or without the addition of OM-LQR. Whole Microbiome Analysis: Group wise comparison of Aitchison distance showed no difference between stability time points (P > 0.05) and change was significantly lower than observed donor-to-donor differences (P < 0.0001). No taxa had a differential abundance exceeding measured within-group variance (effect size < 1, adjusted P-value > 0.1).L.1.b, Room Temperature Storage Stability Acceptance Criteria; Stability Results; L.1.e, Stability (Whole Microbiome) Results
ReproducibilityCV of classified read counts post-rarefaction < 30% for each Microbiome Panel Species between all samples per donor (lot-to-lot assessment) or between all aliquots per donor (aliquot-to-aliquot assessment). Consistency demonstrated in ≥ 90% of donors per Microbiome Panel species.Consistency of microbial composition demonstrated by a CV ≤ 30% for each Microbiome Panel Species of all donors investigated as part of both lot-to-lot and aliquot-to-aliquot assessments. Lot-to-lot: Max CV 21.95%, Overall pass rate 100%. Aliquot-to-aliquot: Max CV 32.48% (with 99.4% overall pass rate). Whole Microbiome Analysis: Aitchison distance showed lot-to-lot and aliquot-to-aliquot differences were significantly lower than stability time points; magnitude and variability much less than between donor changes (P < 0.0001). PCA showed intra-donor clustering.L.1.c, Reproducibility Acceptance Criteria; Reproducibility Results; L.1.e, Reproducibility (Whole Microbiome) Results

Study Details

1. Sample Sizes and Data Provenance

  • Test Set Sample Sizes:
    • Neutrality Study: 30 total donors (Adult Cohort: minimum 30 donors, Pediatric Cohort: minimum 30 donors initially, but final dataset included 30 total after screening; tables indicate "45" collected, "30" in final dataset for adults and 30 total for pediatrics with specific sub-cohort counts).
    • Room Temperature Storage Stability Study: 30 total donors (same 30 OMD-200 baseline samples from Neutrality Study, plus additional samples for different time points/conditions).
    • Reproducibility Study (Lot-to-lot & Aliquot-to-aliquot): 14 donors in the final dataset (adult cohort, screened).
  • Data Provenance:
    • Human fecal specimens collected from adult and pediatric donors.
    • Specimens were collected under an Institutional Review Board (IRB) approved protocol and informed consent was obtained.
    • Retrospective/Prospective: Primarily prospective or collected for the purpose of these studies (e.g., "Fecal collection occurred at either the donor's home or a collection site").
    • Country of Origin: Not explicitly stated, but implies a US-based study given the FDA regulatory context and IRB mention.

2. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: Not applicable. These studies are analytical performance assessments of a collection device, not diagnostic interpretation by human experts. The "ground truth" is derived from the WGS assay results of unstabilized (fresh) fecal samples and the established quantification limits for bacterial species.
  • Qualifications: N/A

3. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. This is an analytical performance study, not one requiring adjudication by human readers/experts for a diagnostic outcome. The primary measurements are DNA yield and relative read counts of specific bacterial species.

4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study: No. This type of study is not relevant for evaluating the performance of a sample collection and stabilization device. MRMC studies are typically used for evaluating the effectiveness of diagnostic imaging tools or AI algorithms that assist human interpretation.

5. Standalone Performance (Algorithm Only)

  • Standalone Performance: Not applicable in the traditional sense of an "algorithm." The device itself is a collection and stabilization tool. Its "performance" is assessed by its ability to preserve the biological sample's integrity for downstream analysis. The "algorithms" used are bioinformatics analyses (WGS, CLR transformation, Aitchison distance, ALDEx2, Shannon's index, PCA) to quantify changes in the microbial community, not algorithms that provide a diagnostic output on their own.

6. Type of Ground Truth Used

  • Ground Truth:
    • Analytical Ground Truth: Unstabilized (fresh) fecal samples collected from the same donors, processed immediately, serving as the "baseline" or "in vivo state" for comparison of microbial community composition.
    • Quantitative Ground Truth: Established Limit of Blank (LoB), Limit of Detection (LoD), Lower Limit of Quantitation (LLoQ), Upper Limit of Quantitation (ULoQ), and corresponding Baseline Read Count Limits (BRCLs) and Test Sample Read Count Limits (TeSaRCol) for a defined Microbiome Panel of bacterial species. These limits provide quantitative thresholds for acceptable performance.

7. Sample Size for the Training Set

  • Training Set Sample Size:
    • The document implies that the WGS Assay Validation section (L.1.a) describes the process of establishing the LoQ values and BRCLs for the Microbiome Panel species. This can be considered analogous to a "training" or "assay establishment" phase.
    • The specific sample size for that validation is not explicitly stated in terms of number of donors or samples for each bacterial species to define these limits. However, it mentions using "contrived sample sequencing and analysis" and calculations based on the LoQ values.

8. How the Ground Truth for the Training Set was Established

  • Ground Truth for Training/Assay Validation:
    • The "ground truth" for defining the WGS assay limits (LoB, LoD, LLoQ, ULoQ) and subsequently the BRCLs and TeSaRCols was established through analytical validation methods. This involved:
      • Defining a Microbiome Panel (MP) of bacterial species based on criteria like prevalence, abundance, phylogenetic representation, and GC content, plus specific pathogenic/pediatric species.
      • Using "Background DNA Mixture (BDM)" as a diluent for determining LoD, LoQ, and Range.
      • Establishing these limits through quantitative experimental procedures (e.g., "contrived sample sequencing and analysis") to determine the lowest and highest amounts of analyte that can be determined with acceptable precision and trueness.
      • The BRCLs were calculated based on these LLoQ and ULoQ values for each species, ensuring that baseline samples fell within the quantifiable range of the WGS assay.

In summary, the device's performance was rigorously evaluated through analytical studies using well-defined quantitative acceptance criteria related to DNA preservation, microbial community neutrality, and stability, with comparisons made against immediate processing of unstabilized samples from the same donors to establish "ground truth" for its performance claims.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR OMNIgene.GUT Dx DECISION SUMMARY

A. DEN Number:

DEN200040

B. Purpose for Submission:

De Novo request for evaluation of automatic class III designation for the OMNIgene . GUT Dx

C. Measurands:

Storage and stability of bacterial DNA from human fecal specimens.

D. Type of Device:

Device to preserve and stabilize relative abundances of microbial nucleic acids in clinical samples

E. Applicant:

DNA Genotek Inc.

F. Proprietary and Established Names:

OMNIgene.GUT Dx

OMD-200

G. Regulatory Information:

    1. Regulation section:
      21 CFR 866.2952
    1. Classification:
      Class II

{1}------------------------------------------------

    1. Product code(s):
      OPO
    1. Panel:
      83- Microbiology

H. Indications for Use:

    1. Indications for use:
      OMNIgene. GUT Dx is intended for the non-invasive collection of human fecal samples and the stabilization of DNA from the bacterial community for subsequent assessment of the microbiome profile by an assay validated for use with OMNIgene GUT Dx.
    1. Special conditions for use statement(s):
      For in vitro diagnostic use only For prescription use only
    1. Special instrument requirements:
      None

I. Device Description:

The OMNIgene . GUT Dx device consists of a collection tube with a tube top and pusher cap with a screw seal, along with a spatula or spoon for transferring fecal specimen into the collection tube. The tube contains 2 mL of the stabilizing liquid and a stainless-steel mixing ball. These components are intended to stabilize bacterial DNA in human fecal specimens. notably to preserve the relative abundances of bacterial organisms, for potential downstream analysis of the fecal microbiome. The collection device is designed for storage of fecal specimens at room temperature (20-26°C/68-79°F) for up to 30 days.

J. Standard/Guidance Document Referenced (if applicable):

Not applicable

K. Test Principle:

Not applicable

{2}------------------------------------------------

L. Performance Characteristics:

1. Analytical performance:

a. Whole Genome Sequencing (WGS) Assay Validation

To assess collection device performance, a whole genome sequencing assay workflow was validated. This validation included defining a microbiome panel of bacterial species and quantitating the Limit of Blank, Limit of Detection, Lower Limit of Quantitation, Upper Limit of Quantitation, Lower Baseline Read Count Limit, Upper Baseline Read Count Limit, and Range (see definitions below) for each of these bacterial species in the microbiome panel (MP). Using these established limits and ranges for this panel, the device performance was quantitated in terms of stabilization of bacterial DNA as well as specifically the preservation of relative abundances of these specific bacterial members within fecal specimens, described as "neutrality".

Definitions of Terms:

Background DNA Mixture (BDM) - Since WGS enumerates all DNA molecules present in a library preparation, the ability to detect targets of interest is directly proportional to the abundance of the target and the presence of other non-target or 'background' DNA molecules. A mixture of non-target DNA (bacterial in origin) has been created to serve as a diluent, for the determination of LoD, LoQ and Range for targets in the Microbiome Panel.

Limit of Blank (LoB) - The highest measurement result that is likely to be observed for a blank sample.

Limit of Detection (LoD) - The Lowest amount of analyte in a sample that can be detected above the defined Limit of Blank.

Lower Limit of Quantitation (LLoQ) - The lowest amount of analyte in a sample that can be quantitatively determined with acceptable precision and trueness.

Upper Limit of Quantitation (ULOQ) - The highest amount of analyte in a sample that can be quantitatively determined with acceptable precision and trueness.

Range - Dynamic range for reliable quantification of WGS reads will be established through contrived sample sequencing and analysis. This per species range falls between the ULOQ and LoD determined in the assay validation for each species and will be applied across all sequencing studies.

Lower Baseline Read Count Limit (I-BRCL) - The minimum read count a species can have within a baseline sample (e.g. an unstabilized (fresh) sample for neutrality tests; a T0 sample for stability tests) in order for the pair of samples from a donor to be included in testing for that species. I-BRCL is a species-specific value, which will be calculated for each species during assay validation, based on the LLoO for that species.

Upper Baseline Read Count Limit (u-BRCL) - The maximum read count a species can have within a baseline sample (e.g. an unstabilized (fresh) sample for neutrality tests; a T0 sample for stability tests) in order for the pair of samples from a donor to be included in testing for that species. u-BRCL is a species-specific value, which will

{3}------------------------------------------------

be calculated for each species during assay validation, based on the ULOO for that species.

Species IDFamily
(b)(4)

The members of this panel were selected based on the following criteria:

  • · High prevalence | (014) in representative donor population
  • · Abundance (b)(4) in donors where species is present
  • · Representative species from major bacterial families on the gut microbial community phylogenetic tree: (D)(-)

(b) Fry

  • Gram-negative and Gram-positive bacteria are represented o
  • (DR4) · Genomes range in GC content from

Additionally, a pathogenic species from the (54) family was included in the panel as well as three pediatric relevant species.

Lower and Upper Baseline Read Count Limits:

The purpose of the lower and upper Baseline Read Count Limits (I-BRCL and u-BRCL) was to define a read count range for baseline samples that was contained within the overall quantifiable range of the WGS assay. This was done to ensure that if the read counts of a microbiome panel species changed over time or in response to an external challenge, the change from baseline was quantifiable and did not immediately exceed the lower and upper quantification limits.

The LoQ values (LLoQ and ULoQ) were used to calculate the lower and upper Baseline Read Count Limit (I-BRCL and u-BRCL). The calculated upper and lower BRCL for each species was used in the screening of donors for OMNIgene.GUT Dx sample collections.

{4}------------------------------------------------

The per-panel species defined Limits and Ranges are summarized below.

Summary of Final Limits and Range for the WGS assay and the upper and lower limits for Baseline samples (BRCLs)

Test Sample Read Count Limits (TeSaRCol):

The above defined WGS assay limits are designed to define the read count range in which a Microbiome Panel species can be tested with confidence. The OMNIgene GUT Dx analytical testing used these limits in the context of the Test Sample Read Count Limit (TeSaRCol) method (Figure 1).

Image /page/4/Picture/4 description: The image is a gray rectangle with a red border. The text '(b)(4)' is in the upper center of the rectangle. The text is in red and is small in size.

5

{5}------------------------------------------------

Image /page/5/Figure/0 description: The image shows a gray rectangle with the text "Test Sample" on the left side. There is also the text "(b)(4)" at the top of the rectangle. On the right side of the rectangle, there is a legend that says "Limit" with "Max" and "Min" listed below.

  • b. Stability and Neutrality Studies
    This section describes two studies evaluating device performance in terms of:

    1. Microbiome Relative Abundance Preservation ("Neutrality")
    1. Room Temperature Storage Stability

These studies were conducted in two population cohorts, which followed identical study designs:

{6}------------------------------------------------

  • Adult Cohort 30 minimum donors, ages 18+ years .
  • · Pediatric Cohort 30 minimum donors, ages 3-46 months

Fecal collection occurred at either the donor's home or a collection site following provided user instructions. For the pediatric cohort, the actual collection was performed by a parent/guardian. Specimens were collected under an institutional review board (IRB) approved protocol and informed consent was obtained prior to collection. Collection was performed using a container (not the candidate device) without the stabilization liquid to a fill-to line to acquire bulk specimen, which was sent to the processing lab on ice within 24 hours of collection (adult cohort) or 48 hours of collection (pediatric cohort).

Donor Screening and Inclusion:

A sample from each donor was screened to ensure that each donor met set inclusion criteria for these studies. For adult donors to be included, screened fecal specimens had to contain a minimum of 10 species from the Microbiome Panel ( ( (6)(4) within Baseline Read Count Limits determined during the WGS assay validation study. Additionally, each adult relevant species of the Microbiome Panel ( on () (b)(4) ) had to be represented in the screened sample within BRCLs by a minimum of (0) donors. The clinically relevant species in the Microbiome Panel ( DKA) was evaluated as it appeared and had no criteria for minimum number of donors.

For pediatric donors to be included, due to inherent low prevalence and lack of species diversity in the youngest pediatric donors, each pediatric sub-cohort had different inclusion criteria. which increased with age to reflect the change to an adultlike microbial diversity. Specifically, the screened sample from the pediatric donor must have the minimum number of species (listed below) from the Microbiome Panel within their BRCLs as determined during assay validation:

  • 3-14 months: Minimum of 2 species total, with a minimum of 1 species from . (0)(4) specifically
  • · 15-30 months: Minimum of 6 species total, no restrictions on which species
  • · 31-46 months: Minimum of 8 species total, no restrictions on which species

Each pediatric relevant species of the Microbiome Panel ( @10) must be represented in the screened sample within BRCLs by a minimum of 10 donors.

(0)(4)

{7}------------------------------------------------

(b)(4)

Control samples included 1 positive specimen control (PSC)
(b)(4)

(b)(4)
One (1) negative specimen control (NSC) sample was included (b)(4)

(b)(4)
One (1) no template control (NTC) was included
(b)(4)

(b)(4)
(b)(4)

(b)(4)

Normalized libraries were pooled
together
(b)(4)

loaded onto the cartridge and the sequencing run is started. Sequencing

{8}------------------------------------------------

was done on the Illumina NextSeq 550 [

(0)(4)

(0)147 Note that each step (extraction,

quantification, library prep, sequencing) was individually validated using the same principles as described above for the whole genome sequencing assay.

(b)(4)

WGS Bioinformatics Analysis Procedure:

{9}------------------------------------------------

Microbiome Relative Abundance Preservation ("Neutrality"):

The objective of this "Neutrality" study was to validate the ability of the OMD-200 device to not interfere with the microbial community composition at the point of collection such that it maintains an unbiased representation of the in vivo state, i.e. preserves the relative abundances of bacterial organisms in the specimen. This study compared fecal specimens collected in OMD-200 devices containing the stabilization liquid ("OMD-200 baseline") to fecal specimens from the same donor collected without the stabilization liquid contained within the OMNIgene GUT Dx device ("unstabilized control baseline").

15141

While this study initially included 45 donors, only 30 total donors were included in the final dataset after screening donors for specified inclusion criteria identified in the

{10}------------------------------------------------

donor screening section above. Importantly, the OMD-200 baseline specimen data from this study were also used for the room temperature storage stability study described below.

Number of Donors45
Control SpecimenswithoutStabilizationLiquid# specimens collected per donor1
# of DNA extractions per specimen1
Total # extracted control specimens45
Total # of control specimens in final dataset30
OMD-200CollectedSpecimens# specimens collected per donor1
# of DNA extractions per specimen (OMD Baseline)1
Total # extracted OMD-200 specimens45
Total # extracted OMD-200 specimens in final dataset30
Total # of Extracted Samples90
Total # of Samples for Final Data Set60

Number of Donor Specimens and Extracted Samples in Neutrality Study

Representation of Microbiome Panel species in the adult cohort of the neutrality

Image /page/10/Picture/4 description: This image shows a table with two columns. The first column is labeled "Species ID" and the second column is labeled "Total Number of Donors". The table appears to be empty, with no data present in either column. The table is part of a study.

Representation of Microbiome Panel species in the pediatric cohort of the neutrality study

Total Number of Donors Species ID

{11}------------------------------------------------

Species IDTotal Number of Donors
Pediatric Sub cohort representation in the neutrality study
-----------------------------------------------------------------------
Pediatric Sub-Cohort# DonorsRepresentation
3-14 months1033.3%
15-30 months1240.0%
31-46 months826.7%

Microbial community consistency between control and test samples was evaluated based on whether rarefied species-level read counts for each Microbiome Panel species in the test samples fell within TeSaRCol range calculated from the unstabilized control sample:

Image /page/11/Picture/4 description: The image is a gray rectangle with a red border. In the top center of the rectangle is the text "(b)(4)". The text is also red.

Neutrality Acceptance Criteria:

The acceptance criteria for this study included a total DNA yield ≥ 120 ng 0 (b)(4) per extraction aliquot in ≥ 95% of samples. Additionally, microbial community neutrality is demonstrated as follows:

  • Per donor, each Microbiome Panel species in the stored sample has read . counts within the TeSaRCol range as calculated above using the unstabilized Control Baseline
  • Neutrality per Microbiome Panel species will be demonstrated in ≥ 90% of . donors that were successfully screened to contain that species

Neutrality Results: For the neutrality study, microbial DNA They

{12}------------------------------------------------

(6)(5) The results of this study showed that all species examined as part of the Microbiome Panel demonstrated neutrality in ≥ 90% of donors between unstabilized and OMD-200 fecal samples. Furthermore, for both the adult and pediatric cohorts evaluated, 100% of specimens met the DNA yield acceptance criteria of 120 ng ( (0)(4)

Colley

Statistical summary of OMD-200 device DNA yields (ug) for neutrality samples of both cohorts

Adult CohortPediatric Cohort
(b)(4)(b)(4)(b)(4)
Mean ± SD$11.98 \pm 4.37$$9.69 \pm 5.65$
95% CI of Mean10.35, 13.627.58, 11.81
Median11.458.51
Min, Max5.10, 26.472.41, 27.81
% ≥100%100%

Room Temperature Storage Stability:

The objectives of this stability study were to validate the ability of the OMD-200 device to stabilize the microbial community composition during room temperature storage and to validate that the addition of the OM-LQR processing reagent does not impact the ability of the OMD-200 device to stabilize samples during room temperature storage (20℃ - 26℃).

Number of Donor Specimens and Extracted Samples in Room Temperature Stability Study

Stability Study
Number of Donors
RT StorageWith OM-LQR# specimens collected per donor# of DNA extractions per specimenTotal # extracted specimens(b)(4)
RT StorageWithout OM-LQR# specimens collected per donor# of DNA extractions per specimenTotal # extracted specimens
Total # of Extracted Samples
Total # of Samples for Final Data Set

*30 samples are the shared OMD Baseline samples extracted as part of neutrality testing.

{13}------------------------------------------------

Representation of Microbiome Panel species in the adult cohort of room temperature stability study

Species IDTotal Number of Donors
2274

Pediatric Sub cohort representation in the room temperature storage study

Pediatric Sub-Cohort# DonorsRepresentation
3-14 months826.7%
15-30 months1756.7%
31-46 months516.7%

Representation of Microbiome Panel species in the pediatric cohort of room temperature storage stability study

Image /page/13/Picture/5 description: This image shows a table with two columns, "Species ID" and "Total Number of Donors". The table contains one row with the value "(DK-1)" in the "Species ID" column. The "Total Number of Donors" column is empty.

{14}------------------------------------------------

Image /page/14/Figure/0 description: The image is a gray rectangle with a thin red border. In the center of the rectangle, there is the text "(b)(4)" in a small, light-colored font. The background is a uniform gray color, and the red border is a simple, thin line that outlines the entire rectangle.

Room Temperature Storage Stability Acceptance Criteria:

The acceptance criteria for this study included a total DNA yield ≥ 120 ng ( per extraction aliquot in ≥ 95% of samples. Additionally, microbial community stability is demonstrated as follows:

  • . Per donor, each Microbiome Panel species in the stored sample has read counts within the TeSaRCol range as calculated using the OMD Baseline
  • Stability per Microbiome Panel species will be demonstrated in ≥ 90% of . donors that were screened to contain that species

Room Temperature Storage Stability Results:

For both the adult and pediatric cohorts evaluated, 100% of specimens met the DNA yield acceptance criteria of 120 ng ( b | Furthermore, all species examined as part of the Microbiome Panel demonstrated stability in ≥ 90% of donors when stored at room temperature for up to 30 days either with or without the addition of OM-LQR (auxiliary liquefaction reagent).

Statistical summary of DNA yields (ug) for room temperature stability samples of both cohorts.

Image /page/14/Figure/8 description: The image shows a table with two columns. The first column is labeled "Adult Cohort" and the second column is labeled "Pediatric Cohort". The table appears to be empty, with no data present in either column.

Reproducibility Study C.

The primary objectives of this study were to evaluate: 1) performance reproducibility of the OMD-200 collection device across multiple device lots when used to collect from the same fecal sample, and 2) microbiome profile reproducibility across multiple aliquots of the same OMD-200 collected fecal sample. (0)(4) 10/4)

{15}------------------------------------------------

Three different lots of OMD-200 devices, each containing a different lot of stabilizing solution, were used. The same sample was collected into one device from each lot to evaluate the reproducibility between device lots. To evaluate the reproducibility between aliquots, five aliquots were sequentially taken from the same collected OMD-200 sample. This study included 14 donors in the final dataset. These 14 were selected after screening for the inclusion criteria ========================================================================================================================== (0)(4)

15/4

section of the adult cohort.

Number of Donor Specimens and Extracted Samples in the Lot-to-lot Reproducibility Study

Total Number of Donors14
Number of specimens collected per donor3
Number of specimens collected per lot14
Number of DNA extractions per specimen1
Total # of Extracted Samples for Final Data Set42

{16}------------------------------------------------

Number of Donor Specimens and Extracted Samples in the Aliquot-to-aliquot Reproducibility Study

Total Number of Donors14
Number of specimens collected per donor1
Number of DNA extractions per specimen5*
Total # of Extracted Samples56
Total # of Samples for Final Data Set70*
  • One extraction was performed during the Lot-to-lot Reproducibility Study

Reproducibility Study Procedures:

From a single specimen, each donor collected (b)(4) aliquots into (b)(4) OMD-200devices, representing (b)(4) unique device lots. Specimens were received andprocessed by lab personnel.
(b)(4)
(b)(4)
(b)(4) All

aliquots from one donor were extracted by a single operator. DNA was extracted, quantified, sequenced, and analyzed. Consistency in microbial community composition across samples was assessed based on the coefficient of variation (CV) of classified read counts post-rarefaction.

Representation of Microbiome Panel species in the donor cohort

Image /page/16/Picture/7 description: The image shows a table with two column headers: "Species ID" and "Number of donors". The table itself is empty, with no data populated beneath the headers. The image is a close-up of the table headers, with a gray background filling the rest of the frame. The table is outlined in a light green color.

Reproducibility Acceptance Criteria:

The acceptance criteria for this study included a total DNA yield ≥ 120 ng (0.12 ug) per extraction aliquot in ≥ 95% of samples. Additionally, microbial community composition consistency is shown as follows:

  • CV of classified read counts post-rarefaction < 30% for each Microbiome . Panel Species between all Tour samples per donor (lot-to-lot assessment) or between all proaliquots per donor (aliquot-to-aliquot assessment).
  • . Consistency is demonstrated in ≥ 90% of donors per Microbiome Panel species

{17}------------------------------------------------

Reproducibility Results:

All specimens for reproducibility assessment both lot-to-lot and aliquot-to-aliquot met the DNA yield acceptance criteria of 120 ng (0.12 ug). Furthermore, consistency of the microbial composition was demonstrated by a CV ≤ 30% for each Microbiome Panel Species of all donors investigated as part of both the lot-to-lot and aliquot-toaliquot assessments

Statistical summary of DNA yields (ug) from samples used to assess lot-to-lot
reproducibility
Lot ALot BLot COverall
Total DNAYield(µg/aliquot)AcceptableMinimum =0.12 µgN14141442
Mean ± SD$8.06 \pm 4.97$$7.56 \pm 4.66$$7.50 \pm 5.15$$7.71 \pm 4.81$
95% CI of Mean5.19, 10.934.87, 10.254.53, 10.486.21, 9.21
Median6.906.096.016.40
Min, Max3.23, 19.692.73, 19.241.66, 21.471.66, 21.47
% ≥ 0.12 µg100%100%100%100%

Statistical summary of DNA yields (ug) from samples used to assess aliquot-toaliquot reproducibility

N70
Total DNA Yield(µg/aliquot)AcceptableMinimum =0.12 µgMean ± SD7.97 ± 4.07
95% CI of Mean7.00, 8.94
Median7.29
Min, Max2.61, 19.24
% ≥ 0.12 µg100%

Statistical summary of lot-to-lot microbial community consistency results

Min CV (%)0.22
Max CV (%)21.95
Mean CV (%)5.84
Median CV (%)4.92
Total data points156
Overall pass rate (CV ≤ 30%)100%

Statistical summary of aliquot-to-aliquot microbial community consistency

results
Min CV (%)0.62
Max CV (%)32.48
Mean CV (%)5.69
Median CV (%)4.74
Total data points156
Overall pass rate (CV ≤ 30%)99.4%

{18}------------------------------------------------

e. Whole Microbiome Analysis Studies

The primary objectives of the study were to:

  • · Demonstrate that the whole microbiome is stable during storage at room temperature in the range of 20-26°C in the OMD-200 device for up to 30 days.
  • · Demonstrate that the OMD-200 device preserves the representation of the in vivo state of the whole microbiome (i.e., "snap-shot" or neutrality), where the in vivo state is characterized as unstabilized fecal samples collected and processed immediately.
  • Demonstrate the performance reproducibility of the OMD-200 device for the . whole microbiome across multiple device lots (lot-to-lot) when used to collect from the same fecal sample and across multiple aliquots (aliquot-to-aliquot) of the same OMD-200 collected fecal sample.

Sequencing data generated during the execution of the Stability and Neutrality Study and the Reproducibility Study were leveraged for this analysis. The results presented in this report are intended to expand on the Microbiome Panel results from the OMD-200 validation to include the whole microbiome. The use of the Microbiome Panel in establishing the device claims allowed assessment of the performance of the OMD-200 collection device using clearly defined limits for Microbiome Panel species. which were established during the Whole Genome Sequencing (WGS) Assay Verification Study.

The samples were analyzed to obtain classified read counts for each Microbiome Panel species from control samples and test samples. Classified read counts were rarefied to 2 million reads per sample to account for technical variability in loading samples for sequencing. After rarefaction, the read counts from test samples were then compared, species-by-species, to the control samples to ensure that the test sample read counts were within a defined, acceptable range. The acceptable range was determined based on Test Sample Read Count Limits (TeSaRCol). The use of such an assay was needed to validate the candidate device in order to detect changes between test and control conditions with very specific and well-defined thresholds. Data for adult and pediatric cohorts were evaluated separately.

Data Analysis:

Counts data were transformed using the center-log ratio (CLR) for analysis. Differences in the microbiome between test conditions were evaluated using the Aitchison distance for group wise comparisons, and ALDEx2 differential analysis to evaluate changes between individual taxa. All data analysis, visual representations, and statistical analyses were performed using R. For each analysis objective, the differences in the microbiome were evaluated with three primary criteria:

  • The Aitchison distance was used to measure the difference in total bacterial . microbiome composition and relative abundance between paired samples (e.g., between samples collected with OMD-200 at baseline and the same samples at T60), and between comparison groups of paired samples. (D (4)

{19}------------------------------------------------

Additionally, the variability of the distance values (674) within groups were compared. In each case, magnitude and variability of change were compared to donor-to-donor differences. For comparisons between groups, a Kruskal-Wallis H test (non-parametric) was applied. Two group comparisons were compared with a t-test from baseline to the test group.

00741

  • · Differential relative abundance analysis was performed on every pass-filter species-level taxon within the sample using ALDEx2. Taxa were considered differentially abundant if the fold-change exceeded the maximum withincondition variance (effect size > 1), and statistically significant (Benjamini-Hochberg adjusted P-value < 0.1) by Welch's paired t-test. Data are represented in an effect plot for visualization of differential relative abundance (between condition) vs. dispersion (within-condition variability).
  • Alpha diversity was measured by Shannon's index, and number of observed . taxa was quantified after rarefaction to 2 million reads per sample. Differences in alpha diversity between conditions were measured with a paired t-test.

{20}------------------------------------------------

Stability and Neutrality Studies (Whole Microbiome):

Image /page/20/Picture/1 description: The image contains two gray rectangles stacked vertically. The top rectangle is larger than the bottom rectangle and is outlined in teal. The bottom rectangle is outlined in red and has the text '(b)(4)' in the top right corner. The image does not contain any other objects or text.

Stability (Whole Microbiome) Results:

Group wise comparison of the Aitchison distance using the Kruskal-Wallis H test and a two-group t-test showed no difference between each stability time point (P > 0.05), and that the change in microbial profile was significantly lower than observed donorto-donor differences (P < 0.0001; both adult and pediatric cohorts). The observed magnitude of change across time points (Aitchison distance) and the variability within each time point was much less than between donor changes, suggesting that the whole microbiome was stable for up to 30 days at room temperature with or without OM-LQR.

{21}------------------------------------------------

A group wise comparison of Aitchison distance for the stability validation dataset was conducted for an adult cohort and pediatric cohort. Each donor sample was compared at baseline (T0) to the same sample after storage at room temperature with (RTL) or without (RT) liquefaction reagent (OM-LQR) for 30 days (T30). The magnitude of change (distance) and the variability for each time point group is far lower than the between-donor change, suggesting that the microbiome profiles are stable over time. A Kruskal-Wallis non-parametric test was applied between groups, and two-group comparisons using a t-test from RT-T30 vs the remaining groups.

Differential abundance analysis was performed across all storage conditions in comparison to the baseline (OMDB) samples. For all conditions, no taxa had a differential abundance exceeding the measured within group variance and therefore no significant changes in taxa were detected (effect size < 1, adjusted P-value > 0.1). This indicates that the differences between these conditions did not exceed the inferred sample variability within the group.

{22}------------------------------------------------

A: Adult Cohort

CHECK

B: Pediatric Cohort

{23}------------------------------------------------

Image /page/23/Figure/0 description: The image is a gray rectangle with a red border. The text '(b)(4)' is located at the top center of the rectangle. The text is small and in black font.

Neutrality (Whole Microbiome) Results:

Group wise comparison of the Aitchison distance using the Kruskal-Wallis H test and a two-group t-test showed that the change in microbial profile was significantly lower between OMD-200 collected fecal samples at baseline (OMDB) and unstabilized control fecal samples at baseline (CB) than that observed for donor-to-donor differences at baseline (P < 0.0001; shown below for both adult and pediatric cohorts). This suggested that the OMD-200 device maintained the representation of the fecal sample at the point of collection.

A: Adult Cohort

B: Pediatric Cohort

{24}------------------------------------------------

Differential abundance analysis was performed between OMD-200 collected fecal samples (OMDB) and unstabilized control fecal samples (CB). No taxa had a differential abundance exceeding the measured within-group variance and therefore no significant changes in taxa were detected (effect size < 1, adjusted P-value > 0.1) (shown below for both adult and pediatric cohorts). This indicated that the differences between these conditions did not exceed the inferred sample variability within the group.

(b)(4)

A: Adult Cohort

B: Pediatric Cohort

PCA analysis (both adult and pediatric cohorts) and hierarchical cluster analysis of the data showed intra-donor clustering, regardless of collection and storage condition.

{25}------------------------------------------------

Reproducibility (Whole Microbiome) Study:

Reproducibility (Whole Microbiome) Results:

Group wise comparison of the Aitchison distance using the Kruskal-Wallis H test and a two-group t-test showed that, as expected, the difference between the same fecal sample collected into | OMD-200 collection devices (lot-to-lot) was significantly lower than the difference between stability time points (30 days). Furthermore, the difference between (1) aliquots extracted from a single OMD-200 collected fecal sample (aliquot-to-aliquot) also showed smaller changes in Aitchison distance compared to the stability time points (30 days) (Figure below).

Significant change (P < 0.05) was observed between the same fecal sample collected into | OMD-200 collection devices (lot-to-lot) and maliquots extracted from a single OMD-200 device (aliquot-to-aliquot) and the observed magnitude of change and the variability for the lot-to-lot and aliquot-to-aliquot cohorts was much less than. between donor changes, indicating that the OMD-200 device provided a reproducible sample collection (P < 0.0001).

{26}------------------------------------------------

(0)(4)

{27}------------------------------------------------

(b)(4)

A compositional principal component analysis of the reproducibility dataset was generated. The first two components explaining the most variance in the data were PC1 23.28% and PC2 with 17.94% variance explained (41.22% in total). Data showed intra-donor clustering on the first two components, regardless of condition (aliquot-to-aliquot replicates and lot-to-lot replicates).

    1. Comparison Studies:
    • a. Method Comparison:

Method comparison is not applicable for a nucleic acid collection and stabilization device. The device itself does not provide a result that can be used in making a clinical decision. Bench testing studies were done to determine the ability of OMNIgene. GUT Dx to stabilize bacterial DNA from fecal samples.

  • b. Matrix Comparison:
    Not applicable

    1. Clinical Studies:
      Not applicable
    1. Clinical cut-off:
      Not applicable
    1. Expected values/Reference range:
      Not applicable.

M. Instrument Name

Not applicable

N. System Descriptions:

    1. Modes of Operation:
      Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?

or No X Yes Yes

{28}------------------------------------------------

Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes of Nor No X or X and

    1. Software:
      FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes or No or No X ---

Hazard Analysis and software development are not applicable to this class of device:

    1. Specimen Identification:
      Collection devices are not intended to identify specimens. The device itself does not provide a result that can be used in making a clinical decision. Collection devices are intended to preserve and stabilize nucleic acids.
    1. Specimen Sampling and Handling: See section L.1.b regarding specimen stability.
    1. Calibration: Not applicable.
    1. Quality Control: Not applicable.
  • O. Other Supportive Instrument Performance Characteristics Data Not Covered in the "Performance Characteristics" Section above:

Not applicable

P. Proposed Labeling:

The labeling supports the decision to grant the De Novo request for this device.

{29}------------------------------------------------

Q. Identified Risks to Health and Mitigation Measures:

Identified Risks to HealthMitigation Measures
Failure to correctly operate thedevice leading to inadequate samplecollection.Certain labeling information, including warnings anddevice descriptions.
Failure to stabilize microbial nucleicacid resulting in an inaccurate assayresult.Certain design verification and validation studies.
Device use with unvalidated orincompatible assays leading toinaccurate assay results andimproper patient management.Certain labeling information, including warnings,device descriptions, and study information.
Malfunction of the collection devicemay lead to possible exposure toinfectious pathogens bylaboratorians or individualscollecting fecal samples.Certain labeling information, including warnings anddevice descriptions.

R. Benefit/Risk Analysis:

Summary
Summary ofBenefit(s)OMNIgene•GUT Dx is intended for the non-invasive collection of human fecalsamples and the stabilization of DNA from the bacterial community for subsequentassessment of the microbiome profile by an assay validated for use withOMNIgene•GUT Dx. OMNIgene•GUT Dx facilitates accurate analysis of the relative distribution ofdifferent bacteria in the fecal microbiome by standardizing sample collection andstabilization of bacterial nucleic acids during sample storage and transport to thelaboratory. This may facilitate the development of fecal microbiome assays byeliminating the need for assay manufacturers to independently develop separatedevices for sample collection/nucleic acid stabilization, as well as potentiallyallowing assay developers to leverage the validation studies performed forOMNIgene•GUT Dx in the development of their assay.
Summary ofRisk(s)Inaccurate assay results may occur due to failure to stabilize bacterial nucleic acids. OMNIgene•GUT Dx use with assays that have not been validated for use with thisdevice may lead to inaccurate test results with the potential for patient harm,depending on the assay being used with OMNIgene•GUT Dx. Malfunction of the collection device may lead to possible exposure to infectiouspathogens by laboratorians or individuals collecting fecal samples. Failure to operate the device correctly may lead to inadequate sample collection,which may lead to inaccurate test results with the potential for patient harm,depending on the assay being used with OMNIgene•GUT Dx.
ConclusionsDo theprobableThe probable benefits of the OMNIgene•GUT Dx device outweigh the potential risks inlight of the listed special controls and applicable general controls.

{30}------------------------------------------------

benefitsoutweigh theprobablerisks?The proposed special controls will help to reduce the risk of device errors and failures.Failure to correctly operate the device, leading to inadequate sample collection, ismitigated by special controls for labeling and certain design verification and validationstudies, as well as by current laboratory practices. The risk of the device not sufficientlystabilizing microbial nucleic acid, resulting in inaccurate assay results, is mitigated byspecial controls requiring certain design verification and validation studies duringdevice development. The labeling special controls, including device description andinclusion of appropriate warnings in the device labeling, will reduce the risk of use withincompatible or unvalidated assays and that malfunction of the collection device maylead to possible exposure to infectious pathogens by laboratorians or individualscollecting fecal samples. The OMNIgene•GUT Dx validation studies provided,including those required by the special controls, suggest that the device will be safe andeffective if used as directed by the package insert.
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

S. Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

T. Conclusion:

The De Novo request is granted and the device is classified under the following and subject to the special controls identified in the letter granting the De Novo request:

Product Code:QPO
Device Type:Device to preserve and stabilize relative abundances of microbial nucleicacids in clinical samples.
Class:II
Regulation:21 CFR 866.2952

N/A