K Number
DEN200026

Validate with FDA (Live)

Device Name
EndeavorRx
Date Cleared
2020-06-15

(60 days)

Product Code
Regulation Number
882.5803
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EndeavorRx is a digital therapeutic indicated to improve attention function as measured by computer-based testing in children ages 8-12 years old with primarily inattentive or combinedtype ADHD, who have a demonstrated attention issue. Patients who engage with EndeavorRx demonstrate improvements in a digitally assessed measure Test of Variables of Attention (TOVA) of sustained and selective attention and may not display benefits in typical behavioral symptoms, such as hyperactivity. EndeavorRx should be considered for use as part of a therapeutic program that may include: clinician-directed therapy, medication, and/or educational programs, which further address symptoms of the disorder.

Device Description

EndeavorRx is a digital therapeutic indicated to improve attention as measured by computerbased testing in children ages 8-12 years old with primarily inattentive or combined-type ADHD, who have a demonstrated an attention issue. The EndeavorRx program is a software-as-medical device (SaMD) that resides on the user's mobile device and can be executed at home. EndeavorRx is engineered as a therapeutically active treatment for attention in pediatric affected by ADHD. EndeavorRx is built on Akili's proprietary, patented, technology platform and uses adaptive algorithms (also known as Selective Stimulus Management Engine. SSME™) to deliver stimuli that are designed to engage the patient in a manner that improves their attention function. In a closed-loop system, the adaptive SSME™ algorithms automatically adjust the difficulty level for a personalized treatment experience that is tailored to the needs of each individual patient. EndeavorRx is delivered through a video game experience which leverages art, music, storytelling, and reward cycles to keep patients engaged. The adaptive algorithm constantly pushes patients precisely at predefined performance bounds relative to each individual, such that they are continuously encouraged to exceed their historic performance. The basic program inputs are steering, which is accomplished by using the internal accelerometer to measure the degree to which the mobile device is tilted, and tapping, which is accomplished using the touch screen to measure correct and incorrect targeting. The basic outputs are visual display of the game progression along with audio, which is accomplished by using the internal high-resolution display and internal speaker.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study that proves the EndeavorRx device meets those criteria, based on the provided text:

Acceptance Criteria and Reported Device Performance

The acceptance criteria for the EndeavorRx device are primarily driven by clinical performance demonstrating effectiveness and safety. The text doesn't explicitly state a table of "acceptance criteria" with specific thresholds prior to reporting results; rather, the clinical studies presented here serve as the evidence to establish that the device does meet the necessary criteria for regulatory approval. Based on the "Special Controls" section, the key criteria are demonstrated effectiveness using a validated measure and capture of adverse events.

Here's a table summarizing the implicit acceptance criteria (derived from "Special Controls" and the results presented) and the reported performance:

Acceptance Criterion (Implicit)Reported Device Performance
1. Clinical Performance Testing:
a. Use a validated measure to evaluate effectiveness of device to provide therapy for ADHD or any of its individual symptoms.STARS-ADHD Study (Pivotal):- Primary Endpoint Met: Statistically significant improvement from baseline in the TOVA API compared to digital control (0.93 vs 0.03; p=0.006).- Secondary Endpoints: Numerically favored EndeavorRx over control for ADHD-RS Total, ADHD-RS Inattention, and IRS. Trend observed in favor of EndeavorRx on IRS.- Clinically Meaningful Improvement: 34.5% of EndeavorRx treated participants moved into the normative range on at least one objective measure of attention across all studies.STARS-Adjunct Study:- IRS (Parent/Clinician-rated impairment): Significant improvement for both On Stimulants (-0.7, p<0.001) and No Stimulants (-0.5, p<0.001) groups compared to baseline after 1 month. Improvements stable after 1-month pause, further increased after second month.- ADHD-RS Total, ADHD-RS Inattentive, ADHD-RS Hyperactive, CGI-I: Significantly improved for both cohorts compared to baseline at day 28.- Correlation with Academic Performance: TOVA ACS and TOVA API performance significantly correlated to academic performance measures (TOSREC and MFaCTS) at each time point.ADHD Proof-of-Concept (POC) Study:- TOVA API: Significant improvements in the ADHD group (mean improvement 1.43, p=0.03).
b. Capture all adverse events.STARS-ADHD Study:- No deaths or serious adverse events (SAEs).- 15/348 (4.3%) participants experienced treatment-related AEs (either definitely or possibly related). 12/180 (6.7%) in EndeavorRx group.- Most common AEs in EndeavorRx group (>1%): decreased frustration tolerance (2.8%), headache (1.7%), emotional reaction (1.1%). Mostly transient and mild (3 moderate). No AEs led to discontinuation.STARS-Adjunct Study:- 37 (18%) participants experienced device-related AEs across two treatment months.- Most common AEs: decreased frustration tolerance (13.1%), headache (1.9%), irritability (1.5%). All mild or moderate. No SAEs.- 3 participants discontinued due to treatment-related AE (all decreased frustration tolerance).ADHD POC Study:- 9 AEs reported over all study phases, none judged related to EndeavorRx.Overall: Of 538 EndeavorRx participants across all studies, 50 (9.3%) experienced treatment-related AEs. No SAEs. AEs generally transient. Only 3 led to discontinuation.
2. Software Verification, Validation, and Hazard AnalysisFDA review of the software documentation found to be acceptable. (MODERATE level of concern, prior to risk mitigation).
3. Labeling RequirementsLabeling found to be sufficient and consistent with clinical data, covering hazards and clinically relevant information (e.g., instructions for use, OS requirements, not standalone, not a substitute for medication, summary of clinical performance, warnings for photosensitive epilepsy, color blindness, physical limitations, frustration, seizure, potential screen addiction, decreased sleep quality).

Study Details

Given the multiple studies described, I'll focus primarily on the STARS-ADHD study as it's identified as the "pivotal" study, and briefly mention key aspects of the others where relevant.

1. Sample Sized Used for the Test Set and Data Provenance:

  • Pivotal Study (STARS-ADHD):

    • Total Screened: 857 children
    • Test Set (Randomized): 348 children
      • EndeavorRx group (treatment): 180 children
      • Digital Control group: 168 children
    • Data Provenance: Multi-center randomized controlled trial conducted at 20 sites in the USA.
    • Retrospective/Prospective: Prospective. This was a randomized, double-blind, digital controlled study.
  • STARS-Adjunct Study:

    • Total Enrolled: 206 participants (130 "On Stimulants", 76 "No Stimulants")
    • Data Provenance: Not explicitly stated, but likely multi-center and prospective, similar to STARS-ADHD.
    • Retrospective/Prospective: Prospective.
  • ADHD Proof of Concept (POC):

    • Analyzed Sample: 40 children in ADHD group, 40 Healthy Controls.
    • Data Provenance: 3 sites in the US.
    • Retrospective/Prospective: Prospective (open-label study).

2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

The concept of "ground truth" for a diagnostic accuracy study (e.g., classifying an image) doesn't directly apply here, as EndeavorRx is a therapeutic device, not a diagnostic one. The "test set" in this context refers to the participant cohorts in the clinical trials.

  • Diagnosis of ADHD: Inclusion criteria for ADHD diagnoses were based on the DSM V (Diagnostic and Statistical Manual of Mental Disorders, Fifth Edition), and confirmed with a MINI-KID (Mini International Neuropsychiatric Interview for Children and Adolescents) and ADHD-RS (ADHD Rating Scale) Total score. TOVA API score was also used to identify a "demonstrated attention issue."
  • Experts involved: Clinicians at each of the 20 sites (STARS-ADHD) or 3 sites (ADHD POC) would have made the formal ADHD diagnoses. The document does not specify the number or specific qualifications (e.g., years of experience) of the individual diagnosing clinicians, but presumably, they were qualified medical professionals (e.g., pediatricians, child psychiatrists, clinical psychologists) experienced in diagnosing ADHD.

3. Adjudication Method for the Test Set:

  • STARS-ADHD: This was a double-blind study. This means neither the participants nor the investigators/clinicians interacting with them (who would be part of the "adjudication" of symptoms if it were a diagnostic study) knew whether the child was receiving EndeavorRx or the digital control. This inherent blinding serves as a robust form of "adjudication" against bias in assessing primary (TOVA) and secondary (ADHD-RS, IRS, CGI-I, BRIEF) outcomes. The TOVA is a computer-based, objective measure, which reduces subjective interpretation. Secondary measures like ADHD-RS and IRS are parent/clinician-reported, but the blinding helps mitigate bias.
  • ADHD POC: This was an open-label study, meaning participants and investigators knew who was receiving EndeavorRx. This is a less rigorous design in terms of blinding.
  • STARS-Adjunct: The summary does not specify blinding, but given the nature of the interventions (different game stimuli or no stimulants), full blinding might have been challenging for participants, though clinician assessment could have been blinded.

There is no mention of a formal adjudicated consensus process by an independent panel of experts reviewing cases, as might be done in studies of diagnostic imaging AI, but rather relies on the study design's blinding and validated assessment tools.

4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

  • No, an MRMC comparative effectiveness study was not done. MRMC studies are typically used to compare the performance of multiple human readers (e.g., radiologists) with and without AI assistance on a set of cases, primarily in diagnostic imaging. EndeavorRx is a therapeutic device for direct patient use, so this type of study design is not applicable.
  • The studies demonstrated the therapeutic effect of the device compared to a control (e.g., digital control or baseline), not a human reader's improvement with AI.

5. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

  • Yes, the primary evaluation was standalone. EndeavorRx is a "Software-as-Medical Device (SaMD)" that resides on the user's mobile device and is executed at home. The therapeutic effect (improving attention function) is largely driven by the device's adaptive algorithms (Selective Stimulus Management Engine - SSME™) acting directly on the patient. While it's used "as part of a therapeutic program that may include: clinician-directed therapy," the measured improvements in attention function (TOVA API) are attributed directly to the device's use. The studies measured the outcome of the algorithm's therapeutic delivery.

6. The Type of Ground Truth Used:

Again, "ground truth" as a diagnostic reference standard doesn't perfectly fit a therapeutic device. Instead, the "ground truth" for evaluating the effectiveness of EndeavorRx was:

  • Objective Computer-Based Testing: The Test of Variables of Attention (TOVA) Attention Performance Index (API) / Attention Composite Score (ACS). This is an FDA-cleared continuous performance test measuring attention. This is considered an objective measure.
  • Validated Rating Scales (Parent/Clinician-Reported):
    • ADHD-RS (Total, Inattentive, Hyperactive subscales)
    • ADHD Impairment Rating Scale (IRS)
    • CGI-I (Clinical Global Impression-Improvement)
  • Academic Performance Measures: TOSREC (Test of Silent Reading Efficiency and Comprehension) and MFaCTS (Measured Functional Activities of Daily Living - for children).

These measures served as the "truth" against which the device's therapeutic impact was assessed.

7. The Sample Size for the Training Set:

The document does not specify a "training set" in the sense of a dataset used to train the adaptive algorithms (SSME™) of EndeavorRx.

  • The device uses "adaptive algorithms (also known as Selective Stimulus Management Engine. SSME™) to deliver stimuli that are designed to engage the patient in a manner that improves their attention function." It "automatically adjust the difficulty level for a personalized treatment experience."
  • This implies that the individual patient's performance during gameplay serves as the dynamic "training set" for the adaptive algorithm's real-time adjustments for that specific patient.
  • The core algorithms themselves were developed at the University of California, San Francisco (by Adam Gazzaley, M.D.), likely drawing on neuroscience research; the initial development would have involved a form of "training" or optimization, but this wouldn't be referred to as a "training set" in the context of general clinical trial reporting for an AI/ML device like this. The effectiveness studies (STARS-ADHD, STARS-Adjunct, etc.) are evaluating the final, deployed algorithm's therapeutic effect.

8. How the Ground Truth for the Training Set Was Established:

Since there isn't a described "training set" in the conventional sense for the algorithm's initial development, the concept of "ground truth for the training set" as it applies to a supervised machine learning model isn't directly applicable here.

  • The SSME™ adapts to individual player performance. Therefore, the "ground truth" for its adaptive mechanism is the real-time performance data of the player (e.g., accuracy, reaction time, consistency) which the algorithm uses to adjust difficulty and optimize the therapeutic challenge. This is an internal, dynamic "ground truth" established during each play session.
  • The science behind EndeavorRx was developed academically (UCSF), which suggests a foundation in cognitive neuroscience principles guiding the design of the adaptive stimuli and feedback loops. The "truth" for this underlying scientific design would be established through decades of cognitive psychology and neuroscience research on attention and learning.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR

EndeavorRx

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Digital therapy device for Attention Deficit Hyperactivity Disorder. A digital therapy device for Attention Deficit Hyperactivity Disorder (ADHD) is a software intended to provide therapy for ADHD or any of its individual symptoms as an adjunct to clinician supervised treatment.

NEW REGULATION NUMBER: 21 CFR 882.5803

CLASSIFICATION: Class II

PRODUCT CODE: QFT

BACKGROUND

DEVICE NAME: EndeavorRx

SUBMISSION NUMBER: DEN200026

DATE DE NOVO RECEIVED: April 16, 2020

SPONSOR INFORMATION:

Akili Interactive Labs Inc. 125 Broad Street, 4th Floor Boston, MA 02110

INDICATIONS FOR USE

EndeavorRx is a digital therapeutic indicated to improve attention function as measured by computer-based testing in children ages 8-12 years old with primarily inattentive or combinedtype ADHD, who have a demonstrated attention issue. Patients who engage with EndeavorRx demonstrate improvements in a digitally assessed measure Test of Variables of Attention (TOVA) of sustained and selective attention and may not display benefits in typical behavioral symptoms, such as hyperactivity. EndeavorRx should be considered for use as part of a therapeutic program that may include: clinician-directed therapy, medication, and/or educational programs, which further address symptoms of the disorder.

LIMITATIONS

The sale, distribution, and use of EndeavorRx are restricted to prescription use in accordance with 21 CFR 801.109.

The device is not intended to be used as a stand-alone therapeutic device.

{1}------------------------------------------------

EndeavorRx may not be appropriate for patients with photo-sensitive epilepsy, color blindness, or physical limitations that restrict use of a mobile device; parents should consult with their child's healthcare provider.

PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

EndeavorRx is a digital therapeutic indicated to improve attention as measured by computerbased testing in children ages 8-12 years old with primarily inattentive or combined-type ADHD, who have a demonstrated an attention issue. The EndeavorRx program is a software-as-medical device (SaMD) that resides on the user's mobile device and can be executed at home.

Image /page/1/Picture/4 description: The image shows two different ways to play a game on a tablet. In image a), the person is using two hands to control the game, and there are arrows indicating that the person is swiping the screen. In image b), the person is using one hand to control the game, and there are concentric circles indicating that the person is tapping the screen.

Figure 1. EndeavorRx Medical Device Software

EndeavorRx is engineered as a therapeutically active treatment for attention in pediatric affected by ADHD. EndeavorRx is built on Akili's proprietary, patented, technology platform and uses adaptive algorithms (also known as Selective Stimulus Management Engine. SSME™) to deliver stimuli that are designed to engage the patient in a manner that improves their attention function. In a closed-loop system, the adaptive SSME™ algorithms automatically adjust the difficulty level for a personalized treatment experience that is tailored to the needs of each individual patient.

EndeavorRx is delivered through a video game experience which leverages art, music, storytelling, and reward cycles to keep patients engaged. The adaptive algorithm constantly pushes patients precisely at predefined performance bounds relative to each individual, such that they are continuously encouraged to exceed their historic performance. The science behind EndeavorRx was developed at the University of California, San Francisco by Adam Gazzaley, M.D., Founding Director of the University of California San Francisco's Neuroscape and Akili's Chief Science Advisor.

The basic program inputs are steering, which is accomplished by using the internal accelerometer to measure the degree to which the mobile device is tilted (Figure 1a), and tapping, which is accomplished using the touch screen to measure correct and incorrect targeting (Figure 1b). The basic outputs are visual display of the game progression along with audio, which is accomplished by using the internal high-resolution display and internal speaker.

4 J.A. Anguera et al., "Video Game Training Entrol in Older Adults," Nature 501, no. 7465 (September 5, 2013): 97–101, https://doi.org/10.1038/nature12486.

{2}------------------------------------------------

The program includes features to ensure it is used per the prescribed regimen (approximately 25 minutes per day for 5 days per week), including lock-out after the allocated gameplay, as well as reminders to the user (and parents) to maximize compliance with the regimen.

SUMMARY OF NONCLINICAL/BENCH STUDIES

SOFTWARE

EndeavorRx Proprietary Software as a Medical Devices (SaMD) was reviewed according to the FDA Guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," issued May 11, 2005. The software was found to have a MODERATE level of concern because a failure of the device may result in a minor injury, either to a patient or to a user of the device prior to risk mitigation. FDA review of the software documentation provided in support of EndeavorRx was found to be acceptable.

SUMMARY OF CLINICAL INFORMATION

Note: In early clinical evaluations, EndeavorRx was referred to as "ProjectEvo" and in later evaluations as "AKL-T01." Therefore "ProjectEVO" and "AKL-T01" will appear in the cited publications.

Five clinical studies have been conducted with EndeavorRx in over 600 children with ADHD including: the STARS-ADHD, STARS-Adjunct, and ADHD POC studies, or with comorbid ADHD, including: the SPD+ADHD Pilot and ASD+ADHD Pilot studies. Each trial is summarized below.

STARS-ADHD

Background and Trial Design

Patient population and duration of device use:

Children with ADHD in STARS-ADHD used EndeavorRx for 4 weeks.

Overview of trial design:

The STARS-ADHD pivotal study2 of EndeavorRx was a multi-center, randomized, double-blind, digital controlled study in 348 children aged 8-12 years. who were diagnosed with ADHD and had a demonstrated attention issue. Children were randomized 1:1 to EndeavorRx or a digital control. Both groups used the treatment or digital control at home on a tablet device for four weeks. The digital control mimicked the reward and engagement of EndeavorRx but deployed different stimuli and did not include Akili's SSME adaptive algorithms.

The primary outcome measure was, change in the Test of Variables of Attention (TOVA) Attention Performance Index (API). TOVA is an FDA-cleared continuous performance test measuring attention, which also aids in the evaluation of ADHD treatments. TOVA Attention Performance Index (TOVA ACS/APP) is an attentional composite measure that captures attentional functioning at a broad scale. Two of its subcomponents measure more specific attentional processes as follows: Reaction Time Mean First

2 Scott H. Kollins et al., "A Novel Digital Intervention for Actively Reducing Severity of Paediatric ADHD (STARS-ADHD): A Randomised Controlled Trial," The Lancet Digital Health 2, no. 4 (April 1, 2020): e168-78, https://doi.org/10.1016/S2589-7500(20)30017-0.

3 TOVA Attention Composite Score (TOVA ACS) is exactly the same as the TOVA API above and was simply a name change in the measure between TOVA V8 and TOVA V9.

{3}------------------------------------------------

Half (RT Mean H1) measures selective and sustained attention, and Reaction Time Variability (RT Var) measures attentional consistency. 4.5

Secondary endpoints included: mean changes in ADHD-RS (Total, Inattentive, Hyperactive), ADHD Impairment Rating Scale (IRS). CGI-I, and BRIEF (working memory scale, inhibit scale). The study was managed by the Duke Clinical Research Institute and was conducted at 20 sites in the USA.

Inclusion and exclusion criteria included: children with an ADHD diagnosis based on the DSM V, with a MINI-KID and ADHD-RS Total score ≥28 and who demonstrated an attention issue based on a TOVA API baseline score < - 1.8 at screening. Participants had to be stable and off ADHD medication or therapy (3-7 days before baseline), and not present other significant comorbid psychiatric diagnoses. Nonverbal IQ was required to be ≥80.

Enrollment

857 children were screened and 509 were excluded as follows:

  • 482 did not meet inclusion criteria primarily because they did not have a demonstrated attention . issue as assessed with TOVA API
  • 32 met exclusion criteria .
  • 4 enrolled late .

The remaining 348 children were allocated to EndeavorRx (n=180) and digital control (n=168).

Demographics

The mean age of the participants in this study was 9.6 years and mostly male (b) (4)(Table 1). The majority of children presented with the combined (72%) or inattentive (26%) ADHD subtype. All children were currently off of any ADHD medication.

Table 1. STARS-ADHD: Demographics / Baseline Characteristics
--------------------------------------------------------------------
EndeavorRxDigital ControlTotal
N(b) (4)
Age Mean (SD)
Sex (Male)
Race (White)

Accountability

97% of participants completed the study. Reasons for discontinuation were lost to follow-up (5/9). withdrawal by parent or guardian (3/9) and investigator (1/9). Both groups showed high engagement completing >80% of recommended sessions.

4 Gordon B. Forbes, "Clinical Utility of the Test of Variables of Attention (TOVA) in the Diagnosis of Attention-Deficit/Hyperactivity Disorder," Journal of Clinical Psychology 54, no. 4 (June 1998): 461-476, https://doi.org/10.1002/(SICI)1097-4679(199806)54:461):1410-JCLP8>3.0.CO;2-Q.

Macela P Henriquez et al., "Low Senum Sphingolipids in Children with Attention Deficit-Hyperactivity Disorders," Frontiers in Neuroscience 9 (2015): 300, https://doi.org/10 3389/fnins.2015.00300.

{4}------------------------------------------------

Results

Safety

EndeavorRx was shown to be safe in this study, with no deaths or other serious adverse events (AEs) reported. Fifteen (15/348, 4.3%) participants experienced treatment-related AEs, either definitely related or possibly related, across both EndeavorRx and the digital control groups. Twelve of the AEs (12/180, 6.7%) were within the EndeavorRx group. Three categories of AEs were experienced by more than 1% of study participants within the EndeavorRx group: decreased frustration tolerance (5/180, 2.8%), headache (3/180. 1.7%), and emotional reaction (2/180, 1.1%). AEs experienced by the EndeavorRx participants were mostly transient and considered mild except 3 which were moderate. No AEs led to treatment discontinuation.

Effectiveness

The predefined primary endpoint was achieved in that EndeavorRx showed a statistically significant. improvement from baseline in the TOVA API compared to the digital control (0.93 vs 0.03; p=0.006) (Figure 2).

Figure 2 Primary Endpoint: TOVA API, STARS-ADHD Pivotal Trial

Image /page/4/Figure/6 description: This image shows a bar graph comparing the API score of AKL-T01 and a control group. The API score for AKL-T01 is approximately 0.9, while the API score for the control group is approximately 0.0. The p-value is 0.006, and the ITT population for AKL-T01 is n=180, while the control group is n=168.

Of the five secondary Clinical Outcomes, mean improvements in three of these (ADHD-RS Total, ADHD-RS Inattention and IRS) numerically favored EndeavorRx over the control, and a trend was observed in favor of EndeavorRx on IRS(D) (4) Notably, of these eleven response rate outcomes, all (4) the IRS response rate(D)(4) and the parent perspective of improved favored EndeavorR responder rates (al (b) attention response rate (b) (4). There was no statistically meaningful difference in a non-parametric analysis of the 7 secondary parental or clinical rating scales (D)

STARS-Adjunct

Background and Trial Design

Patient population and duration of device use:

Children with ADHD in the STARS-Adjunct study used EndeavorRx for 1 month, followed by a 1-nonth treatment pause, and then a second treatment month.

{5}------------------------------------------------

Overview of study design:

The goal of the STARS-Adjunct trial was to study the effects of EndeavorRx on symptoms and impairments in children either on or off stimulant medicating the durability of effects, one month after treatment, and the effects of an additional treatment month.

Participants in the On Stimulants cohort were stable on a stimulant medication but inadequately managed by that stimulant. Participants in the No Stimulants cohort were stable without any stimulant (or other ADHD medication) for at least 30 days before baseline. Neither cohort was on non-ADHD medications.

During the first treatment month (Days 1 through 28) participants in the On Stimulants cohort received stimulant plus EndeavorRx, and participants in the No Stimulants cohort received EndeavorRx only. During the 1-month pause (Days 29 through 56) between treatment phases, participants in the On Stimulants cohort remained on stimulant, and participants in No Stimulants cohort remained off ADHD medication. During the second treatment month (Days 57 through 84), participants in the On Stimulants cohort received stimulant plus EndeavorRx, and participants in the No Stimulants cohort received EndeavorRx.

Inclusion and exclusion criteria included: children between the age of 8 to 14 with an ADHD diagnosis of inattentive or combined subtype per DSM V, MINI-KID and ADHD related impairment based on the ADHD Impairment Rating Scale, IRS Total ≥ 3). Participants had to be either stable on (On Stimulants group) or off ADHD medication (No Stimulants group), and not present other significant comorbid psychiatric diagnoses. Nonverbal IO was required to be >80.

Enrollment

A total of 236 participants were screened for inclusion in this study, 130 were enrolled in the On Stimulants group, and 76 in the No Stimulants group.

Accountability

195/206 (95%) completed the study through the primary endpoint (1st month). Overall, 179/206 (87%) participants completed the full 3-month study, and 27 (13.1%) participants discontinued the study. A total of 124 and 71 participants completed the study through day 28 in the On Stimulants and No Stimulants cohorts, respectively.

Demographics

The overall mean (SD) age of participants was 10.6 (1.77) years. Overall, the majority of participants were white(b) (4) and male (b) (4)

{6}------------------------------------------------

Table 2. STARS-Adjunct: Demographics / Baseline Characteristics

On StimulantsOff StimulantsTotal
N(b) (4)
Age Mean (SD)
Sex (Male)
Race (White)

Results

Safety

37 (18%) participants experienced a device-related AE across the two treatment months. The most common device-related AEs experienced by participants were decreased frustration tolerance (27 [13.1%] participants), headache (4 [1.9%] participants), and irritability (3 [1.5%] participants). All device-related AEs were either mild or moderate in severity. No serious device-related AEs occurred during this trial. There were 3 participants who discontinued treatment due to a treatment-related adverse event. All 3 of these participants experienced decreased frustration tolerance.

Effectiveness

The primary endpoint was the change in ADHD Impairment Rating Scale (IRS), which is a parent or clinician rating of ADHD-specific impairment across domains such as social functioning, academic progress, and self-esteem, including an overall impairment. The results of this study demonstrate that after the first treatment month (day 28), IRS overall severity score was significantly improved for both the On Stimulants (-0.7, p<0.001) and No Stimulants (-0.5, p<0.001) groups compared to baseline (Figure 4). The ADHD-RS Total. ADHD-RS Inattentive, and ADHD-RS Hyperactive subscales and the CGI-I were also significantly improved for both cohorts compared to baseline at day 28. We found that the symptoms and impairments as measured with the IRS, ADHD-RS and CGI-I remained significantly improved from baseline after the 1-month treatment pause. An additional month of treatment led to further improvement in impairment and symptom measures (IRS: Figure 5, ADHD-RS: Figure 6 and CGI-I not shown; all pvalues <0.001 relative to baseline).

{7}------------------------------------------------

Image /page/7/Figure/0 description: This bar graph shows the IRS overall improvement with and without stimulants. The y-axis shows the IRS overall improvement (Mean +/- SE) from -0.8 to 0.0. The x-axis shows three categories: No Stimulants, On Stimulants, and Both Cohorts. The IRS overall improvement for No Stimulants is about -0.5, for On Stimulants is about -0.7, and for Both Cohorts is about -0.6. The p-value is less than 0.001 for all three categories.

Figure 3. STARS-Adjunct IRS Improvement from Baseline to Day 28

ITT Population: No Stimulants n=76, On Stimulants n=130, Both Cohorts n=206

Image /page/7/Figure/3 description: The image is a title for a figure. The title reads "Figure 4. STARS-Adjunct IRS Improvement over Two Treatment Months". The title is written in a clear, legible font, and it is centered on the image.

Image /page/7/Figure/4 description: The image is a line graph titled "IRS Overall Improvement". The x-axis represents days, with labels BL, 28, 56, and 84. The y-axis represents IRS Overall Improvement (Mean +/- SE), ranging from 3.5 to 5.5. There are three lines on the graph, representing different groups: STARS-Adjunct (On Stimulants), STARS-Adjunct (No Stimulants), and STARS-Adjunct (Both Cohorts).

ITT Population: No Stimulants n=76, On Stimulants n=130, Both Cohorts n=206

{8}------------------------------------------------

Figure 5. STARS-Adjunct ADHD-RS Total Improvement over Two Treatment Months ADHD-RS Total Improvement

Image /page/8/Figure/1 description: This image shows a line graph comparing the ADHD-RS Total Improvement (Mean +/- SE) over time for three different groups: STARS-Adjunct (On Stimulants), STARS-Adjunct (No Stimulants), and STARS-Adjunct (Both Cohorts). The x-axis represents the days (BL, 28, 56, 84), and the y-axis represents the ADHD-RS Total Improvement. The graph indicates that all three groups experienced improvement over time, with the 'No Stimulants' group showing slightly higher improvement at certain time points. The graph is divided into two treatment cycles.

ITT Population: No Stimulants n=76, On Stimulants n=130, Both Cohorts n=206

The TOVA ACS and TOVA API performance was significantly correlated to academic performance measures of TOSREC and MFaCTS at each time point throughout the study and an improvement in the TOVA ACS and TOVA API was related to an improvement in both of these academic performance measures (Figures 7 and 8).

Image /page/8/Figure/4 description: The image is a bar graph that shows the TOSREC change from baseline to day 28. The y-axis represents the TOSREC change from baseline to day 28 (Mean +/- SE), and the x-axis represents the ACS change. The bar graph shows that the TOSREC change from baseline to day 28 is significantly higher in the ACS change >= 0.9 group (p < 0.001) compared to the ACS change < 0.9 group (p = 0.800).

Figure 6. TOSREC Following 1st Month of EndeavorRx Treatment

ITT Population: ACS/API Change < 0.9 n=129, ACS/API Change ≥ 0.9 n=71

{9}------------------------------------------------

Image /page/9/Figure/0 description: This image is labeled Figure 7 and shows the acronym MFaCTS. The figure is titled, "MFaCTS Following 1st Month of EndeavorRx Treatment." The text is written in a clear, sans-serif font and is centered on the page.

Image /page/9/Figure/1 description: This image is a bar graph that shows the change from baseline to day 28 in MFACTS scores. The x-axis shows two groups: ACS Change < 0.9 and ACS Change >= 0.9. The y-axis shows the MFACTS change from baseline to day 28. The p-value for the ACS Change < 0.9 group is 0.217, and the p-value for the ACS Change >= 0.9 group is 0.005.

ADHD Proof Of Concept (POC)

Background and Trial Design

Patient population and duration of device use:

Children in ADHD POC used EndeavorRx for 4 weeks.

Overview of trial design

Akili Interactive conducted an initial proof-of-concept study . ADHD POC, to assess the feasibility, acceptability and efficacy of EndeavorRx in 8 to 12-year old children with ADHD. This open-label study compared 40 children with ADHD to 40 Healthy Controls. The ADHD group had a diagnosis of ADHD and were not taking ADHD medications. The study was conducted at 3 sites in the US. The study regimen began with a day 0 in-clinic assessment. Participants were then sent home and instructed to complete approximately 25 minutes of EndeavorRx per day for 5 days per week for 4 weeks. A final inclinic assessment was completed on day 28.

Study objectives were: 1) To assess treatment compliance and acceptability of an at-home intervention: and 2) To explore whether participants demonstrated improvements in attention function, as measured by TOVA, following the intervention period.

Inclusion and exclusion criteria included: Children 8-12 years old, the ADHD group required a diagnosis of ADHD, with ADHD-RS Total score of ≥24 at baseline, was not on any ADHD medication and had no comorbid psychiatric diagnosis. Healthy Controls had an ADHD-RS Total score ≤13.

Enrollment

87 participants were screened for participation in the trial. Of these, 3 failed to meet screening criteria. 40 children were in the ADHD group and 44 consisted of Healthy Controls. 4 Healthy Controls were terminated early due to insufficient gameplay. The final analyzed sample consisted of 40 children in the ADHD group and 40 in the Control group (children completing 4 weeks of at-home play and returning to the clinic on Day 28). The safety population included all participants randomized to EndeavorRx.

6 Naomi O. Davis, Jeffrey Bower, and Scott H. Kollins, "Proof-of-Concept Study of an at-Home, Engaging, Digital Intervention for Pediatic ADHD," PLoS One 13, no. 1 (2018): e0189749, https://doi.org/10.1371/journal.pone.0189749.

{10}------------------------------------------------

Accountability

Eighty-four percent of treatment sessions were completed. The ADHD group completed an average of (b)(4) hours of intervention), the non-ADHD group completed an average of (b)(4) hours of intervention) of the required at-home sessions.

Demographics

The ADHD group mean age was 10.4 years and 2440 (b) (4) were male. The Healthy Controls had
similar characteristics (mean age 10.5 years and 15) (4) male). The ADHD group had ADHD symptom scores (ADHD-RS Total: (b) (4)

Statistical Analysis

The means of the pre-test scores were compared to the post-test scores using a (b)(4) (b)(4)

Results

Safety

A total of 9 adverse events were reported over all study phases; however, none of those adverse events were judged by the investigator as related to EndeavorRx.

Effectiveness

Significant improvements were observed on the TOVA API in the ADHD group (mean of improvement 1.43, p=0.03). There was no significant change for the Healthy Control group (0.39, p=0.3) (Figure 9).

Image /page/10/Figure/11 description: This bar graph compares the change from baseline to day 28 between AKL-T01 ADHD and AKL-T01 Healthy Control groups. The y-axis represents the change from baseline to day 28 (Mean +/- SE). The AKL-T01 ADHD group has a value of approximately 1.4, while the AKL-T01 Healthy Control group has a value of approximately 0.4. The p-value for the AKL-T01 ADHD group is 0.03, while the p-value for the AKL-T01 Healthy Control group is not significant (ns).

Figure 8. TOVA API Change in ADHD POC Study

ITT Population: AKL-T01 n=40, AKL-T01 Healthy Control n=40

The following additional studies have been performed in smaller populations with largely similar results:

{11}------------------------------------------------

The ASD Pilot was a randomized, double-blind, controlled study in 19 children, ages 9 to 15 years old, who were diagnosed with Autism Spectrum Disorder (ASD) and comorbid ADHD, 11 were randomized to EndeavorRx and 8 to digital control. The EndeavorRx group numerically improved in the TOVA API (1.86, p=0.12) while the Control group worsened (-0.82, p=ns). The EndeavorRx group improved significantly in ADHD symptoms. For example, on the ADHD-RS Total, the group had -6.72 change (p=0.003) from baseline. Both groups had high compliance with their intervention. There was one nonserious adverse event of decreased frustration tolerance in the EndeavorRx group.

The SPD Pilot® investigated the effects of EndeavorRx in children ages 8 to 12 years old diagnosed with Sensory Processing Disorder (SPD) and comorbid ADHD (SPD+ADHD, n=20), SPD only (n=13), and Healthy Controls (n=24). All groups received EndeavorRx for 4 weeks.

There was significant improvement irrespective of group in TOVA measures (RT Mean H1 and RT Var H1), but only the SPD+ADHD group showed a significant improvement in parent-reported ADHDinattentive symptoms (-4.5, p<0.001), which correlated with EEG measures. There were 8 children lost to follow-up. No treatment-related adverse events were reported.

In sum, all five studies of the device demonstrate safety, and support the effectiveness and favorable risk 910,11/benefit profile for use in improving the inattention component of ADHD.

LABELING

The EndeavorRx Instructions for Use are consistent with the clinical data and cover all the hazards and other clinically relevant information that may impact use of the device. The labeling is sufficient and satisfies the requirements of 21 CFR § 801.109 for prescription devices. Although there are no contraindications, it is noted in the labeling that EndeavorRx may not be appropriate for patients with photo-sensitive epilepsy, color blindness or physical limitations that restrict use of a mobile device. The labeling indicates that for these circumstances the healthcare provider should be consulted.

The following precautionary statements are included within the device labeling:

  • . Please follow all your mobile device manufacturer's instructions for the safe operation of your mobile device. For example, this may include: appropriate volume settings, proper battery charging, not operating the device if damaged, and proper device disposal. Contact your mobile device manufacturer for any questions or concerns that pertain to your device
  • If your child experiences frustration, emotional reaction, dizziness, nausea, headache, eye-strain, ● or joint pain while playing EndeavorRx, pause the treatment. If the problem persists contact your child's healthcare provider. If your child experiences seizure, stop the treatment and contact your child's healthcare provider.
  • Federal law restricts this device to sale by or on the order of a physician. ●
  • Summary of Side Effects / Adverse Events.

&#x27; Benjamin Yerys et al., "Brief Report: Pilot Study of a Novel Interactive Digital Treatment to Improve Cognitive Control in Children with Autism Spectum Disorder and Co-Occurring ADHD Symptoms," Journal of Autism and Developmental Disorders 49, no. 4 (2019): 1727–37, https://doi.org/10.1007/s10803-018-3856-7.

് Joaquin A. Anguera et al., "A Pilot Study to Determine the Feasibility of Enhancing Cognitive Abilities in Children with Sensory Processing Dysfunction," PloS One 12, no. 4 (2017): e0172616, https://doi.org/10.1371/journal.pone.0172616.

9 Ferrie C. D et al., "Video game induced seizures, "J Neurol Neurosurg Psychiatry 1994 Aug; 57 (8): pp925-931

10 Christine L. Mathews, Holly E. R. Morrell & Jon E. Molle (2019) Video gameaddiction, ADHD symptomatology, and video game reinforcement, The American Journal of Drug and Alcohol Abuse, 45:1, 67-76, DOI: 10.1080/00952990.2018.1472269

11 Becker S. P et al., "Nighttime Media Use in Adolescents with ADHD: Links to Sleep Problems and Internalizing Symptoms," Sleep Med. 2018 November ; 51: 171-178. doi:10.1016/j.sleep.2018.06.021.

{12}------------------------------------------------

RISKS TO HEALTH

Table 3 identifies the risks to health that may be associated with digital therapy device for Attention Deficit Hyperactivity Disorder and the measures necessary to mitigate these risks.

Table 3. Identified Risks to Health and Mitigation Measures
Identified RiskMitigation Measures
Ineffective treatment leading to worsening oruncontrolled symptomsClinical performance testing
Labeling
Device software failure leading to delayed accessSoftware verification and validation
Labeling
Treatment results in frustration, emotionalreaction, dizziness, nausea, headache, eye-strain,or joint painLabeling
Clinical performance testing
Software verification, validation, and hazardanalysis
Treatment results in seizureLabeling
Clinical performance testing
Treatment results in screen addictionLabeling
Clinical performance testing
Treatment results in decreased sleep qualityLabeling
Clinical performance testing

Table 3. Identified Risks to Health and Mitigation Measures

{13}------------------------------------------------

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the digital therapy device for Attention Deficit Hyperactivity Disorder is subject to the following special controls:

  1. Clinical performance testing must demonstrate and document the following under the labeled conditions for use. which include considerations for the ability of the device to:
  • a. Use a validated measure to evaluate effectiveness of device to provide therapy for ADHD or any of its individual symptoms; and
  • Capture all adverse events. b.
  1. Software must be described and provided in a clear and detailed manner to include all features and functions of the software implementing the digital therapy. Software verification, validation, and hazard analysis must also be provided.

  2. The labeling must include the following items:

  • Patient and physician labeling must include instructions for use, including images that a. demonstrate how to interact with the device;
  • Patient and physician labeling must list the minimum operating system (OS) requirements that b. support the software of the device:
  • Patient and physician labeling must include a warning that the digital therapy device is not C. intended for use as a standalone therapeutic device;
  • Patient and physician labeling must include a warning that the digital therapy device does not d. represent a substitution for the patient's medication; and
  • Physician labeling must include a summary of the clinical performance testing conducted with the e. device.

BENEFIT-RISK DETERMINATION

Across five clinical studies of over 600 children with ADHD, EndeavorRx showed a general improvement in attention as well as areas of improvement in other symptoms associated with of ADHD. The totality of the evidence demonstrated clinical benefit in attention, as measured by the TOVA, academic performance measures, and other assessment tools, in children with ADHD with a demonstrated attention issue. Across all studies, 34.5% of EndeavorRx treated participants moved into the normative range on at least one objective measure of attention, a clinically meaningful improvement in attention. Improvements in ADHD symptoms and impairment favored EndeavorRx over control, including clinically meaningful response rates. The STARS-Adjunct study demonstrated improvements in clinical symptoms and impairment measures seen after 1 month treatment with EndeavorRx, remained stable after a one-month treatment pause and then further increased with a second month of treatment, independent of medication status. Children who improved in objective attention (TOVA API) also showed improvement in math and reading performance measures.

The risks associated with EndeavorRx are minimal: Of 538 participants using EndeavorRx, 50 (9.3%) experienced treatment-related adverse events. No serious adverse events were reported. All adverse events were generally transient. Only three events led to device discontinuation, and no subject reported lasting or irreversible effects after discontinuation.

Therefore, the probable benefits of the EndeavorRx outweigh the probable risks in light of the listed special controls and the general controls.

Patient Perspectives

Patient perspectives considered for the EndeavorRx during the review include:

{14}------------------------------------------------

    1. Market research studies involving caregivers, physicians and health insurers to understand their perspective when considering medication for ADHD children, challenges with the current paradigm, and the need for alternative additional treatment options.
      ~90% of caregivers in one study involving caregivers of children currently or formerly on medication, reported they would request a non-drug prescription digital treatment with the profile of EndeavorRx from their physician if it was available. The profile reflected a 6%, non-serious side effect profile (including frustration and headache) and efficacy related to attention.

Parents expressed their desire for the product for multiple reasons in two market research studies: 75% of caregivers in the Armature-Caregiver study thought it would be an effective treatment overall, and most respondents believed it would strengthen their child's cognitive function (>82%, Armature-Caregiver study) and improve concentration and focus (>83%, Armature-Caregiver study) among other issues. In the GfK-Caregiver study, over 80% of caregivers thought it would help with their child's ADHD. Finally, eight in ten parents in the GfK-Caregiver study believed the treatment is easy to use, and half believe it will be extremely easy to use.

Image /page/14/Figure/3 description: The image shows two pie charts comparing the opinions of parents and physicians regarding AKL-T01. The first pie chart shows the percentage of parents who would request AKL-T01 from their doctor, with the majority saying yes. The second pie chart shows the percentage of physicians who think AKL-T01 fills a need for their patients, with the vast majority saying yes.

    1. A post-treatment survey of parents and their child to assess how treatment impacted their attention was conducted in the ASD-Pilot study.
      For parents, the following question was asked, "Do you think playing the app improved your child's ability to pay attention?" For children the following question was asked, "Do you think playing the app improved your ability to pay attention?" For both the parent and child questions, the answer options were "yes" or "no." Seventy-three percent (73%) of children who received EndeavorRx reported that it improved their attention, versus 50% in the Control group who received the digital control. In addition, 64% of parents believed that EndeavorRx improved their child's attention in real life, versus 50% in the Control group. The survey revealed that 63.6% of parents reported that time playing EndeavorRx was very worthwhile for their child, versus 50% in the Control group, and 90.9% would want their child to continue to play (selected 'yes' or 'maybe') versus 62.5% in the Control group.

Benefit/Risk Conclusion

In conclusion, given the available information above, for the following indication statement:

{15}------------------------------------------------

EndeavorRx is a digital therapeutic, indicated to improve attention function as measured by computer-based testing in children ages 8-12 years old with primarily inattentive or combinedtype ADHD, who have a demonstrated attention issue. Patients who engage with EndeavorRx demonstrate improvements in a digitally assessed measure (TOVA) of sustained and selective attention, and may not display benefits in typical behavioral symptoms, such as hyperactivity. EndeavorRx should be considered for use as part of a therapeutic program that may include: clinician-directed therapy, medication, and/or educational programs, which further address symptoms of the disorder.

the probable benefits outweigh the probable risks for the EndeavorRx. The device provides benefits and the risks can be mitigated by the use of general and the identified special controls.

CONCLUSION

The De Novo request for the EndeavorRx is granted and the device is classified as follows:

Product Code: QFT

Device Type: Digital therapy device for Attention Deficit Hyperactivity Disorder

Regulation Number: 21 CFR 882.5803

Class: II

§ 882.5803 Digital therapy device for Attention Deficit Hyperactivity Disorder.

(a)
Identification. A digital therapy device for Attention Deficit Hyperactivity Disorder (ADHD) is a software intended to provide therapy for ADHD or any of its individual symptoms as an adjunct to clinician supervised treatment.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate and document the following under the labeled conditions for use, which include considerations for the ability of the device to:
(i) Use a validated measure to evaluate effectiveness of device to provide therapy for ADHD or any of its individual symptoms; and
(ii) Capture all adverse events.
(2) Software must be described and provided in a clear and detailed manner to include all features and functions of the software implementing the digital therapy. Software verification, validation, and hazard analysis must also be provided.
(3) The labeling must include the following items:
(i) Patient and physician labeling must include instructions for use, including images that demonstrate how to interact with the device;
(ii) Patient and physician labeling must list the minimum operating system (OS) requirements that support the software of the device;
(iii) Patient and physician labeling must include a warning that the digital therapy device is not intended for use as a standalone therapeutic device;
(iv) Patient and physician labeling must include a warning that the digital therapy device does not represent a substitution for the patient's medication; and
(v) Physician labeling must include a summary of the clinical performance testing conducted with the device.