K Number
DEN200005

Validate with FDA (Live)

Date Cleared
2021-12-03

(668 days)

Product Code
Regulation Number
878.4880
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Klox Biophotonic LumiHeal™ System is indicated to provide blue light and fluorescent light energy for use on post-surgical incisions for scar management. The System is intended to be used in FST I-IV female patients 22 years and over.

Device Description

The Klox Biophotonic LumiHeal System is a device which consists of a blue light emitting Multi-LED Light device (KT-L Lamp) with emitted wavelengths of 440-460nm and a topical photoconverter gel (LumiHeal Gel). When the gel is illuminated by the LED device, it will emit fluorescence with blue, green, yellow, and orange wavelengths between 400mm to 625mm. The fluorescence mixed with the excitation blue light is used for scar management (i.e., reducing the appearance of acute post-surgical incisions).

The KT-L Lamp (Figure 1) has a timer and a distance sensor with which the system set the illumination time at 5 min and distance between the lamp and the wound area at 5cm. For the KT-L Lamp interface, press the Distance Verification Button to measure the distance. Display Screen should read approximately 50mm. Press the Time Display Button to return display to Timer Mode. During the illumination, it is possible to toggle between the Timer Mode or Measure Mode by pressing either the Time Display Button or the Distance Verification Button. Time remaining during illumination will be displayed in minutes (min). The lamp comes equipped with a power cable, 2 pairs of protective eyeglasses, and a user manual. The device specification is shown in the Table 1.

The LumiHeal Gel is provided as a two-component gel (Figure 2), specifically Jar A (25g) and Jar B (2.5g). The two components of the LumiHeal Gel are intended to be mixed immediately prior to use. Jar A is the carrier gel and Jar B is the chromophore gel. The LumiHeal Gel is applied just before use and is intended to remain in contact with the surface of wound for 5 minutes during the light exposure and application. After the 5-minute application period, exposure to the KT-L Lamp is discontinued and the LumiHeal Gel is removed.

The purpose of the LumiHeal Gel is to facilitate conversion of the non-coherent blue light wavelength from the KT-L Lamp into blue, green, yellow and orange wavelength light between 400nm to 625mm at the skin surface.

AI/ML Overview

Acceptance Criteria and Device Performance for Klox Biophotonic LumiHeal System

This response synthesizes the acceptance criteria and study findings for the Klox Biophotonic LumiHeal System, as described in the provided regulatory document.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state "acceptance criteria" for effectiveness in a quantitative manner (e.g., "device must achieve X% improvement in Y score"). Instead, it describes general expectations for safety and performance (mitigation of risks) and then reports the study's effectiveness findings. Thus, the acceptance criteria are inferred from the stated benefits and the demonstrated performance.

Acceptance Criteria (Inferred from Risks & Benefits)Reported Device Performance
Safety:Safety:
1. Biocompatibility: Patient-contacting components must be biocompatible (prevent adverse tissue reactions).Biocompatibility assessment was adequate for both light-exposed and non-exposed photoconverter gel. Conclusion: "risk of a clinically significant biocompatibility concern is low." (Section: BIOCOMPATIBILITY/MATERIALS)
2. Sterility: Patient-contacting components must be sterile to prevent infection.LumiHeal Gel Jar B is terminally sterilized via autoclaving. Preservative effectiveness testing per USP <51> for real-time aged Jar A and packaging integrity testing were performed. "The test gels met the acceptance criteria for each test." (Section: SHELF LIFE/STERILITY)
3. Shelf Life: Photoconverter gel must maintain sterility and functional performance over its shelf life.Shelf life testing performed (preservative effectiveness, packaging integrity); "The test gels met the acceptance criteria for each test." (Section: SHELF LIFE/STERILITY) Performance after simulated 5 years of service life met specifications. (Section: PERFORMANCE TESTING - BENCH, last bullet point)
4. Electrical & Thermal Safety: Device must be electrically/thermally safe and electromagnetically compatible.The KT-L lamp passed all relevant portions of IEC 60601-1 and IEC 60601-1-2 testing. (Section: ELECTROMAGNETIC CAPABILITY & ELECTROMAGNETIC SAFETY) Evaluation of heat dissipation showed skin temperatures remained "within a safe temperature range (<43ºC) throughout the illumination for the indicated skin types (up to Type IV on the Fitzpatrick skin scale)". (Section: PERFORMANCE TESTING - BENCH, first bullet point)
5. Software Safety: Software errors should not cause injury.Software is considered a "minor level of concern." All elements of software information provided for minor LOC devices were found to contain sufficient detail to provide reasonable assurance of proper operation. (Section: SOFTWARE)
6. Use Error: Labeling and design should mitigate use errors.Usability testing performed, leading to modifications of labeling and lamp design to enable intended operation by users. (Section: HUMAN FACTORS/USABILITY TESTING)
7. Device Performance (Bench): Device must perform as intended under anticipated conditions of use (e.g., light spectrum, power intensity, heat dissipation, gel properties, mechanical functions).Numerous bench tests were conducted, demonstrating: verification and validation of spectrum/power intensity, heat dissipation within safe limits, biophotonic properties of the gel (absorbance, fluorescence, photobleaching), photonic parameters (peak wavelength, power density), radiant fluence delivered, gel viscosity, complete gel removal, and mechanical functions. Performance after simulated 5 years of service life met specifications. Packaging integrity also verified. (Section: PERFORMANCE TESTING - BENCH)
Effectiveness:Effectiveness:
1. Reduction in Appearance of Acute Post-Surgical Incisions: The device should show an improvement in scar appearance.Blinded expert review of photographs using POSAS Observer scale showed "at least a 1 point improvement in each sub-category (Vascularity, Pigmentation, Thickness, Relief and Surface area)" for the LumiHeal system. (Section: Effectiveness - POSAS- Blinded Evaluators Using Photographs) The mean 1 point improvement was also seen in the control, Silicone group, indicating improvement comparable to an existing scar management method. (Section: BENEFIT-RISK DETERMINATION) Patients' POSAS scores and overall opinion also showed a trend toward improvement. (Section: POSAS- Patient - Total Score and Overall Opinion)
2. Comparable or superior to standard of care (implicitly suggested by comparative study design).The study compared the LumiHeal System to Silicone sheets (Cica-Care® Silicone Sheeting). The POSAS Observer sub-categories' evolution was considered similar between LumiHeal and Silicone, except for pigmentation which showed a trend to be lower in the Silicone group. This suggests comparable effectiveness for most evaluated scar characteristics. No explicit superiority criteria were stated, but comparability to a known effective modality serves as a benchmark. Patient satisfaction with LumiHeal was demonstrated. (Section: Patient Perspectives)

2. Sample Sizes and Data Provenance

  • The document states "102 subjects were enrolled and evaluated in the study, with [Tost] to follow-up by the study completion." (Section: Effectiveness - POSAS- Blinded Evaluators Using Photographs). However, a preceding section states, "A total of [number] subjects were enrolled and [number] patients completed the study." and later, "J (patients (92.9%) received all applications as planned in the protocol." This indicates the number of subjects who completed the study is slightly less than the initial enrollment. The exact number of patients who were analyzed for effectiveness from the 102 subjects mentioned is not explicitly stated as a separate "test set" size after follow-up.
  • Data Provenance: The study was conducted at a "single center in two locations" (Section: SUMMARY OF CLINICAL INFORMATION). The demographics table (Table 3) indicates that the majority of participants (93.0%) were White, suggesting the data is likely from a Western country, possibly Canada given the sponsor information. It was a prospective clinical study.

3. Number of Experts and Qualifications for Ground Truth

  • Number of Experts: For the blinded review of photographs, the document refers to "experts" in the plural, but does not specify the exact number.
  • Qualifications of Experts: The document refers to them as "experts" (Section: Effectiveness - POSAS- Blinded Evaluators Using Photographs). It does not provide specific qualifications (e.g., "radiologist with 10 years of experience" or "board-certified plastic surgeon"). However, the nature of the assessment (POSAS on surgical scars) implies they would be clinicians with relevant experience, likely dermatologists or plastic surgeons.

4. Adjudication Method for the Test Set

  • The document states, "a blinded review by experts was performed." (Section: Effectiveness - POSAS- Blinded Evaluators Using Photographs). It describes that for each image, "the blinded experts had to score each criterion of the POSAS Observer scale." It does not specify an adjudication method such as 2+1 or 3+1 if there were multiple experts. It implies that their individual scores were aggregated, possibly averaged, or simply reported without adjudication if a single expert scored each case.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Yes, a comparative study was done, though it's structured as a within-subject comparison against a control rather than a pure MRMC study of "human readers with vs. without AI." Patients served as their own controls, with one breast receiving LumiHeal and the other receiving Silicone sheets.
  • Effect Size of Human Readers Improve with AI vs. without AI assistance: This specific type of MRMC study (human reader with/without AI assistance) was not performed. The study compared the device (LumiHeal) directly against a standard of care (Silicone sheets) for scar management. The "AI" component is the LumiHeal system itself, providing therapeutic light, not an AI assisting human interpretation. The experts were involved in assessing the outcome of the treatment, not in making diagnoses or interpretations assisted by AI.
  • Improvement: The study found that both LumiHeal and Silicone groups showed "at least a 1 point improvement in each sub-category" of the POSAS Observer scale. The evolution of sub-categories was "similar in all sub-groups combined analysis, except for the pigmentation which showed a trend to be lower in the Silicone group." This indicates the device achieved comparable effectiveness to the control.

6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

  • No, a standalone study (in the context of an AI algorithm reading medical images) was not done. The Klox Biophotonic LumiHeal System is a therapeutic device that applies light energy. Its "performance" is its effect on the human body, not its ability to analyze data independently. Human experts were used to assess the outcome of the device's application.

7. Type of Ground Truth Used

  • Expert Consensus (Blinded Evaluators using Photographs): The primary effectiveness ground truth was established by "blinded experts" (clinicians) who scored photographs of the scars using the Physician Observer Scar Assessment Scale (POSAS). This is a well-established subjective clinical assessment tool.
  • Patient Self-Assessment: Patient-reported outcomes (Patient POSAS, Patient's Overall ease of use and satisfaction questionnaire) also contributed to the assessment of effectiveness and benefit, reflecting patient perspective on scar appearance and treatment experience.

8. Sample Size for the Training Set

  • Not applicable. The Klox Biophotonic LumiHeal System is a physical light therapy device, not an AI algorithm that learns from data. Therefore, there is no "training set" in the context of machine learning. The clinical study described is for validation and effectiveness assessment, not for training a model.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As a physical device, there is no "training set" or "ground truth for a training set" for this product in the typical machine learning sense. The device is designed based on known biophotonic principles, and its safety and effectiveness are then validated through non-clinical and clinical studies.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR KLOX BIOPHOTONIC LUMIHEAL SYSTEM

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Phototherapy device for reducing the appearance of acute post-surgical incisions. This device consists of a light emitting device and a photoconverter gel and is intended to employ light energy for reducing the appearance of acute post-surgical incisions. This classification does not include products which contain drugs or biologics.

NEW REGULATION NUMBER: 21 CFR 878.4880

CLASSIFICATION: Class II

PRODUCT CODE: OPE

BACKGROUND

DEVICE NAME: Klox Biophotonic LumiHeal System

SUBMISSION NUMBER: DEN200005

DATE DE NOVO RECEIVED: February 4, 2020

SPONSOR INFORMATION:

Klox Technologies Inc. 275 boul. Armand Frappier Laval, H7V 4A7 Canada

INDICATIONS FOR USE

The Klox Biophotonic LumiHeal System is indicated as follows:

The Klox Biophotonic LumiHeal™ System is indicated to provide blue light and fluorescent light energy for use on post-surgical incisions for scar management. The System is intended to be used in FST I-IV female patients 22 years and over.

LIMITATIONS

The sale, distribution, and use of the Klox Biophotonic LumiHeal System are restricted to prescription use in accordance with 21 CFR 801.109.

{1}------------------------------------------------

The device should not be used by people taking drugs or products or with conditions known to induce severe photosensitivity reactions.

The device should not be used by people with known skin hypersensitivity.

The device should not be used by women who are pregnant or breast-feeding.

Safety and effectiveness in pediatric population (<22 years old) have not been evaluated.

Safety and effectiveness in patients with Fitzpatrick Skin Type V-VI have not been evaluated.

Safety and effectiveness for patients with hypertrophic and keloid scars have not been evaluated.

PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS. AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

The Klox Biophotonic LumiHeal System is a device which consists of a blue light emitting Multi-LED Light device (KT-L Lamp) with emitted wavelengths of 440-460nm and a topical photoconverter gel (LumiHeal Gel). When the gel is illuminated by the LED device, it will emit fluorescence with blue, green, yellow, and orange wavelengths between 400mm to 625mm. The fluorescence mixed with the excitation blue light is used for scar management (i.e., reducing the appearance of acute post-surgical incisions).

The KT-L Lamp (Figure 1) has a timer and a distance sensor with which the system set the illumination time at 5 min and distance between the lamp and the wound area at 5cm. For the KT-L Lamp interface, press the Distance Verification Button to measure the distance. Display Screen should read approximately 50mm. Press the Time Display Button to return display to Timer Mode. During the illumination, it is possible to toggle between the Timer Mode or Measure Mode by pressing either the Time Display Button or the Distance Verification Button. Time remaining during illumination will be displayed in minutes (min). The lamp comes equipped with a power cable, 2 pairs of protective eyeglasses, and a user manual. The device specification is shown in the Table 1.

{2}------------------------------------------------

Figure 1. KT-L Lamp and User Interface

Image /page/2/Figure/1 description: The image shows a medical device called KLOX. The device is on a wheeled stand and has an adjustable arm with a light at the end. The control panel has a display screen, a time display button, a start/stop button, and a distance verification button.

Table 1. KT-L Lamp Device Specification

Device Technology Description:Specifications
Number of LEDs446nm LEDs: 40; 415nm LEDs: 6
Power density55-129 (mW/cm2)
Distance from light source to wound5cm
Illuminated area7.5 x 15cm
Duration of use5min
Fluency16.5-38.7 J/cm2

The LumiHeal Gel is provided as a two-component gel (Figure 2), specifically Jar A (25g) and Jar B (2.5g). The two components of the LumiHeal Gel are intended to be mixed immediately prior to use. Jar A is the carrier gel and Jar B is the chromophore gel. The LumiHeal Gel is applied just before use and is intended to remain in contact with the surface of wound for 5 minutes during the light exposure and application. After the 5-minute application period, exposure to the KT-L Lamp is discontinued and the LumiHeal Gel is removed.

{3}------------------------------------------------

Figure 2. LumiHeal Gel

Image /page/3/Picture/1 description: The image shows two white jars with labels on them, along with two petri dishes. One petri dish is empty, while the other contains a small amount of orange liquid. The jars are positioned behind the petri dishes, with the larger jar on the left and the smaller jar on the right. The background is a plain white surface.

The purpose of the LumiHeal Gel is to facilitate conversion of the non-coherent blue light wavelength from the KT-L Lamp into blue, green, yellow and orange wavelength light between 400nm to 625mm at the skin surface. The composition, as well as the function of each of the ingredients in the LumiHeal Gel are provided in Table 2 below.

Table 2. LumiHeal Gel Ingredients

CHEMICAL NAMECAS NumberFunctionCONCENTRATION in individual Jar w/w
(b)(4)

*Note: pH adjusted by adding sodium hydroxide until pH between 4.80 and 5.10 is obtained

{4}------------------------------------------------

SUMMARY OF NONCLINICAL/BENCH STUDIES

BIOCOMPATIBILITY/MATERIALS

The KT-L lamp is a non-patient contacting component. The LumiHeal Gel is applied to the patient's post-surgical incisions area for an administration period of 5 minutes. The administration is repeated twice a week for a consecutive 8 weeks. The patient-contacting component is the photoconverter gel and the patient contact classification is a surface device, breached or compromised skin contact, prolonged duration (>24 hours to 30 days). Per the patient contact classification and Table A.1 of the FDA biocompatibility guidance entitled, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1 : Evaluation and testing within a risk management process "", the following biocompatibility endpoints assessments are recommended: cvtotoxicity. irritation, sensitization, systemic toxicity (acute, and subacute/sub-chronic), materialmediated pyrogenicity, and implantation. The biocompatibility assessment was adequate for both the light-exposed and non-exposed photoconverter gel. The conclusion was reached that the risk of a clinically significant biocompatibility concern is low.

SHELF LIFE/STERILITY

The KT-L lamp is provided non-sterile. The LumiHeal Gel is provided as a twocomponent gel: Jar A as the carrier gel and Jar B as the chromophore gel. Jar A contains ingredients which are provided non-sterile. Urea peroxide is included as a preservative in Jar A. Jar B is terminally sterilized in an ISO 13485 certified sterilization facility via autoclaving.

To establish its shelf life, the following testing was performed: preservative effectiveness testing, per USP <51> for the real-time aged, final finished Jar A. Additionally, packaging integrity testing for the duration of the proposed shelf-life was provided, using edge dip dye application method per ASTM F1929-12. The test gels met the acceptance criteria for each test.

ELECTROMAGNETIC CAPABILITY & ELECTROMAGNETIC SAFETY

The following Electrical Safety and Electromagnetic Compatibility testing has been performed:

  • . IEC 60601-1:2005 + A1 2012 Medical Electrical Equipment - Part 1: General requirements for basic safety and essential performance with NRTL Deviations USA
  • . IEC 60601-1-2:2014 4th edition Medical Electrical Equipment -Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

The KT-L lamp passed all relevant portions of the testing.

SOFTWARE

{5}------------------------------------------------

The KT-L Lamp comprises custom embedded software (firmware) to control the LEDs and Human-Machine Interface functions of the lamp. The Agency considers the software to be a minor level of concern (LOC) because inadvertent software errors are unlikely to cause any injury to the patient or operator.

All elements of software information corresponding to minor LOC devices as outlined in. FDA's guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005) were provided and contain sufficient detail to provide reasonable assurance that the software will operate in a manner described in the specifications.

PERFORMANCE TESTING - BENCH

Bench testing was conducted on the KT-L Lamp (the "lamp") and the LumiHeal Gel (the "gel") to demonstrate that the Klox Biophotonic LumiHeal System (the "system") performs as expected under the anticipated conditions of use. The following bench testing was conducted to demonstrate the device performance characteristics:

  • . Evaluation of heat dissipation following the system application: The temperatures of skin was measured after a 2mm layer of gel was spread on the skin of the hand and illuminated for 5 min. The skin temperatures can remain within a safe temperature range (<43ºC) throughout the illumination for the indicated skin types (up to Type IV on the Fitzpatrick skin scale).
  • . Evaluation of the biophotonic properties of the gel: The absorbance, fluorescence spectra and photobleaching of the gel were measured before and after illumination with the blue light lamp. The distance with efficient photobleaching, the best time for illumination, and the gel thickness with least % residual fluorescence were determined.
  • . Verification of photonic parameters for the lamp: the peak wavelength and the power density of the blue light were measured and met the specification.
  • . Validation of method for power intensity measurement on the lamp: the method to measure power density values from the lamp was validated to ensure the accuracy of the measurement.
  • Determination of radiant fluence delivered by the system: the average radiant . fluence of fluorescence and transmitted blue light through the gel by the lamp were determined and the photoconverting function of the gel was verified.
  • Determination of the lowest and highest viscosity of the gel: the minimum and the . maximum viscosity were determined to ensure that the gel is able to be easily spread. to stay in place and to be easily removed after light exposure.
  • . Verification of the removal of gel in a wound model after application: the complete removal of photoconverter gel after clinical procedures was verified.
  • . Verification of the mechanical functions of the lamp: the mechanical performance of the lamp was tested and met the specifications.

{6}------------------------------------------------

  • Verification of the performance of the lamp after the simulated service life: the . performance including illumination power output, interface functions, mechanical functions met specifications after the simulated 5 years of service life.
  • Verification of packaging and transport on the lamp; the packaging of the KT-L . lamp resisted the testing and preserved the integrity.

HUMAN FACTORS/USABILITY TESTING

Usability testing was performed to demonstrate that the lamp design and associated labeling are sufficient to enable intended operation of the device by each intended user populations (i.e., nurses and medical practitioners), Intended users were asked to perform the critical tasks under simulated use conditions and address the questions. Modifications of the labeling and the lamp design were followed based on usability testing result.

SUMMARY OF CLINICAL INFORMATION

Overview

A clinical study (the "Study") was conducted to evaluate the safety and effectiveness of the Klox Biophotonic LumiHeal System when applied to acute post-surgical incisions.

The study was conducted at a single center in two locations. Patients were randomized into one of the following six administration schedules. The product was administered during the proliferative phase of the wound.

    1. Initiated at Day post-surgery, weekly
    1. Initiated Day post surgery, Told weekly
    1. Initiated Day post surgery with double (two consecutive) application | a week
    1. Initiated at Day post surgery. Firm weekly
    1. Initiated at Day " post surgery, Town weekly
    1. Initiated at Day post surgery with double (two consecutive) application for a week

The product was administered for weeks on one breast and the other with Silicone sheets (Cica-Care Silicone Sheeting) for to weeks. The administration site was randomly allocated.

Subsequently, all patients were followed for an additional | week post administration.

Study Methodology

Administration of the product were conducted by one of four plastic surgeons that were trained on the study protocol and procedures.

The patients underwent bilateral breast reduction surgery and immediately after surgery were provided post- operative instructions. The surgical incision was cleaned and dried prior to application. A thin layer of LumiHeal Gel was applied on the wound and illuminated with the KT-L Lamp for 5 minutes, at

{7}------------------------------------------------

~2 inches (~5 cm). For patients receiving two consecutive applications during the visit, once the first illumination of the whole breast incision had been performed, the used gel was gently removed and a second application was performed right away, followed by illumination. A priori to administration to protect from the blue light from the lamp, eye protectors (goggles) had to be worn by the patient and the physician. If needed, the breast incision could be divided into two to three areas depending on the size of the incisions. Following application, the site was then wiped with a moist towel or gauze, and then rinsed with saline solution. During the administration period, approximately weeks post-surgery, all patients were invited to perform breast massage with | every day, twice per day for four weeks.

Patient Population

A total of subjects were enrolled and patients completed the study. Every patient had the LumiHeal System administered to one breast and the second one received Silicone sheets.

Accordingly, all patients who had completed the trial were included in the safety and effectiveness analyses.

Subjects enrolled in the study included women (100%) over the age of 23. The study included no subject with Fitzpatrick Skin Type (FST) V and VI.

All Subjects
Number
Age (years)
Mean (standard deviation)47.5 (11.7)
Minimum, Median, Maximum31, 46, 65
N(%)
Sex
Male00
Female0100
Race
American Indian or Alaska Native02.4
Asian00
Black or African American00
White093.0
Others04.6
Fitzpatrick Skin Type
I016.7
II042.9
III033.3
IV07.1
>V00

Table 3. Summary of Demographic Information

{8}------------------------------------------------

Endpoints

Safety

Safety information was collected throughout the trial during study visits.

Effectiveness

Table 4 presents the Study Endpoints. The results of the study are provided in Tables 10-15.

Primary objectivesEvaluation of the safety and tolerability of the subject device compared with theones of Silicone Sheets (Cica-Care® Silicone Sheeting) for surgical wounds viaassessment of adverse events, serious adverse events, device incidents, and ratesof incision healing complications at every visit.
Secondary objectivesEvaluation of the effectiveness of the LumiHeal System compared with one ofSilicone Sheets as assessed via:
Physician Observer Scar Assessment Scale (POSAS) at Weeks (b)(4) and (b) post-surgery
Patient Observer Scar Assessment Scale (POSAS) at Weeks (b)(4)(b) and (b) , post-surgery
Vancouver Scar Assessment Scale (Observer) at Weeks (b)(4)and (b) post-surgery
Patient's self-assessment of ease of wound management at Weeks (b)(4)(b) and (b) post-surgery

Table 4. Study Endpoints

The effectiveness assessments scales used in this study included:

  • POSAS (Observer Scar Assessment Scale)- Blinded Evaluators using Photographs at . Week compared to Last Study Visit. The POSAS (observer) is validated for surgical incisions using photography.
  • POSAS -Patient, completed by the patients .
  • Patients' Overall ease of use and satisfaction questionnaire .

Results

Safety

Compliance to the Study was considered high as only four visits were missed, out of the 420 required according to the Study protocol. Overall, 99.0% of Study visits planned by the protocol were received during the period. J (patients (92.9%) received all applications as planned in the protocol. The mean number of study applications was 12.5, and the median was 14.0 (minimum of and maximum of (applications).

{9}------------------------------------------------

A total of fourteen adverse events were reported. They concerned eleven patients, representing 26.2% of the total number of patients in the ITT population. There were no specific adverse event trends to report. More specifically, the incidence of the events were as follows: infections (7.1%, breast cellulitis, mastitis, viral upper respiratory infection), injury/procedural complications (4.8% wound dehiscence), reproductive system/breast disorders (9.5%), and skin/subcutaneous disorders (7.1%).

The number of adverse events (number of patients and number of events) was comparable between the different groups, except for the group which underwent administration [0] days post-surgery, this a week, in which no adverse event was observed.

Eight adverse events were assessed as mild: one case of viral upper respiratory tract infection, three cases of breast discharge or discomfort, and four cases of skin irritations (eczema, erythema, pruritus or rash). Six other adverse events were reported as moderate in terms of intensity. They were primarily expected after a breast reduction surgery: infection (mastitis and cellulitis- Silicone and Klox groups), wound/breast dehiscence (Klox and Silicone groups), and breast hematoma (Silicone group).

No events were reported as severe.

None of these adverse events was considered as related to the study application (LumiHeal™) according to the investigators.

Four adverse events (breast discomfort, erythema and pruritus) were considered as related to Silicone sheets.

Effectiveness

POSAS- Blinded Evaluators Using Photographs

As planned in the clinical protocol, a blinded review by experts was performed. 1021 subjects were enrolled and evaluated in the study, with Tost to follow-up by the study completion.

For cach image, the blinded experts had to score each criterion of the POSAS Observer scale: Vascularity, Pigmentation, Thickness, Surface area, Relief, Overall opinion. The total score was also calculated. This evaluation had to be made for Week Wisit Mand at last study visit.

The mean results demonstrate there was at least a I point improvement in each sub-category for the subject device and control, Silicone. Amongst the different results, the evolution of each POSAS Observer sub-category (Vascularity, Pigmentation, Thickness, Relief and Surface area) was considered as similar in all sub-groups combined analysis, except for the pigmentation which showed a trend to be lower in the Silicone group, as shown on Figure 3.

Figure 3. Blinded Review - Evaluation of POSAS Observer scoring changes between visit [] and last study visit - Mean scores for each sub-category - All sub-groups combined

{10}------------------------------------------------

Image /page/10/Figure/0 description: The image is a bar graph that compares the mean of different delta values for two treatments, Klax and Silicone. The y-axis represents the mean, ranging from -2.00 to 0.00, while the x-axis represents the treatment type. For each treatment, there are five bars representing different delta values: Vase, Paga, Thick, Relief, and Surf. The graph shows the mean values and error bars for each delta value under each treatment.

POSAS- Patient - Total Score and Overall Opinion

Similar to the POSAS Observer Non-blinded Investigators, the total score of the POSAS Patient is obtained by the addition of each of the sub-scores (Thickness, Irregularity, Color, Stiffness, Pain and Itching), except Overall opinion.

The baseline mean scores were comparable between LumiHeal (0) (4) SD (b) (4) and Silicone (b) (4) (SD (b) (4). There was a trend toward improvement throughout the study with lower scores at Week [ The LumiHeal score was (b) (4) a decrease of (b) (4) and the Silicone score was (b) (4) a decrease of (b) (4)

The Patient Overall Opinion score was comparable between LumiHeal, 610101 and Silicone, (0) (4) SD (0) at baseline and there was a trend toward improvement over the course of the study. At week the the LumiHeal score was (014 a decrease of 0010 and the Silicone score was (b) (4) a decrease of (b) (4)

Pediatric Extrapolation

In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.

LABELING

Device labeling includes an instruction for use for the KT-L lamp, an instruction for use for the LumiHeal Gel, LumiHeal Gel box labeling. The instruction for use for the KT-L lamp includes

{11}------------------------------------------------

description of the optical specification and warnings for eye safety. The instruction for use for the LumiHeal Gel includes information of gel shelf-life and instructions on the gel preparation, conjunction use with the lamp and its removal.

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of a phototherapy device for reducing the appearance of acute post-surgical incisions and the measures necessary to mitigate these risks.

Identified Risks to HealthMitigation Measures
Adverse tissue reactionBiocompatibility evaluation
InfectionSterility testingShelf life testingLabeling
Thermal damage and ocularinjuryNon-clinical performance testingThermal safety testingLabeling
Shock or burns from electricalmalfunction orelectromagnetic interferencewith other devicesElectrical safety testingElectromagnetic compatibility (EMC) testingSoftware verification, validation, and hazard analysis
Use error that may result ininjuryLabelingSoftware verification, validation, and hazard analysis

Table 11. Identified Risks to Health and Mitigation Measures

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act. the phototherapy device for reducing the appearance of acute post-surgical incisions is subject to the following special controls:

  • Non-clinical performance testing must demonstrate that the device performs as (1) intended under anticipated conditions of use. Testing must include the following:
    • Verification and validation testing of the spectrum and power intensity of the (i) light source:
    • (ii) Heat dissipation from the area following device application; and
    • Biophotonic properties of the photoconverter gel. including radiant fluence (iii) (transmitted light and fluorescence) delivered through the photoconverter gel by the device.
  • (2) The patient-contacting components of the device must be demonstrated to be biocompatible.
  • Performance data must evaluate the sterility of the patient-contacting components of (3) the device.

{12}------------------------------------------------

  • (4) Performance data must support the shelf life of the photoconverter gel by demonstrating continued sterility and functional performance over the identified shelf life.
  • (5) Performance testing must demonstrate the electromagnetic compatibility, electrical safety, and thermal safety of the device in the intended use environment.
  • (6) Software verification, validation, and hazard analysis must be performed for any software components.
  • (7) Labeling must include the following:
    • A summary of the device technical specifications, including light wavelength, (i) irradiance and application area;
    • (ii) Warnings for ensuring eye safety, including use of protective eyeglasses used for both the operator and the patient; and
    • A shelf life for the photoconverter gel. (iii)

BENEFIT-RISK DETERMINATION

The risks of the device are based on nonclinical laboratory studies as well as data collected in a clinical study described above.

A total of fourteen adverse events were reported. They concerned eleven patients, representing 26.2% of the total number of patients in the ITT population. There were no specific adverse event trends to report. None of these adverse events was considered as related to the study application (LumiHeal™) according to the investigators. Four events (breast discomfort, erythema and pruritus) were considered as related to Silicone sheets.

The probable benefits of the device are also based on nonclinical laboratory studies as well as data collected in a clinical study as described above.

The mean results of effectiveness assessments from the blinded evaluators using POSAS scale demonstrate there was at least a 1 point improvement in each sub-category (Vascularity, Pigmentation, Thickness, Relief and Surface area) for the subject device. And the mean I point improvement was also seen in the control, Silicone group. The validation data was provided for using the POSAS scale on photographs of closed surgical incisions. Therefore, a benefit has been demonstrated with the use of the device on scars from closed surgical incisions.

PATIENT PERSPECTIVES

Patient perspectives considered for the Klox Biophotonic LumiHeal System during the review included:

  • . Patients were asked at several timepoints during the administration and follow-up periods if they were satisfied with the application with LumiHeal and Silicone, and on the appearance of the wound/scar. The questions were as follows:
    • How satisfied are you with the care received (LumiHeal™ or Silicone)? .
    • How satisfied are you with the steps required (LumiHeal™ or Silicone)? .
    • . How easy was the treatment to receive / manage (LumiHeal™ or Silicone)?

{13}------------------------------------------------

  • . How satisfied are you with the appearance of your wound (LumiHeal™ or Silicone)?
    The results showed patients were satisfied with the LumiHeal application option.

BENEFIT/RISK CONCLUSION

In conclusion, given the available information above, for the following indication statement:

The Klox Biophotonic LumiHeal™ System is indicated to provide blue light and fluorescent light energy for use on post-surgical incisions for scar management. The System is intended to be used in FST I-IV female patients 22 years and over.

The probable benefits outweigh the probable risks for the Klox Biophotonic LumiHeal System. The device provides benefits and the risks can be mitigated by the use of general controls and special controls.

CONCLUSION

The De Novo request for the Klox Biophotonic LumiHeal System is granted and the device is classified as follows:

Product Code: QPE Device Type: Phototherapy device for reducing the appearance of acute post-surgical incisions Regulation Number: 21 CFR 878.4880 Class: II

N/A