K Number
DEN200002

Validate with FDA (Live)

Device Name
Tangible Boost
Date Cleared
2020-09-24

(255 days)

Product Code
Regulation Number
886.5919
Type
Direct
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Tangible Boost is a monthly treatment to restore the Tangible Hydra-PEG coating and maintain the wettability of Tangible Hydra-PEG treated fluorosilicone acrylate rigid gas permeable lenses. Tangible Boost is intended for prescription (Rx) use only.

Device Description

Tangible Boost is a monthly treatment specifically for RGP CLs made from fluorosilicone acrylate and manufactured with a Hydra-PEG coating. Tangible Boost is intended for prescription (Rx) use only, and for home use. Tangible Boost is not intended for lens disinfection and must be used in conjunction with a compatible multipurpose solution, as described in the labeling.

Tangible Boost restores the Tangible Hydra-PEG coating and maintains the wettability of Tangible Hydra-PEG treated fluorosilicone acrylate lenses. The device consists of two sterile, preservative-free, unit dose solutions to be combined in the single use, disposable Tangible Boost barrel style lens case (included in the device packaging). Patients who are allergic to any ingredient (e.g. Tangible Hydra-PEG) should not use this device.

AI/ML Overview

The provided text describes a De Novo classification request for the Tangible Boost device, a hydrophilic re-coating solution for RGP contact lenses. The acceptance criteria and the study proving the device meets these criteria are primarily outlined through biocompatibility testing, shelf-life/sterility assessments, performance (bench) testing, and a single clinical study.

Here's an analysis based on the provided text:

Acceptance Criteria and Reported Device Performance

The acceptance criteria are primarily defined by the "Acceptance Criteria" column in the tables for biocompatibility testing and implicitly by the successful "Results" obtained in bench and clinical studies.

Table of Acceptance Criteria and Reported Device Performance (Synthesized from text)

Test/Evaluation AreaPurposeAcceptance CriteriaReported Performance
Biocompatibility (Solutions)Evaluate potential for cellular toxicity, sensitization, ocular irritation, acute oral toxicity, and ocular biocompatibility.Non-cytotoxic, Non-sensitizer, Non-irritant, Non-toxic (for acute oral and ocular biocompatibility).Pass for all tests (Cytotoxicity, Sensitization, Ocular irritation, Acute Oral Toxicity, Ocular Biocompatibility). Results demonstrated acceptable performance.
Biocompatibility (Packaging)Evaluate potential for cellular toxicity, ocular irritation, and acute systemic toxicity.Non-cytotoxic, Non-irritant, Non-toxic (for acute systemic toxicity).Pass for all tests (Cytotoxicity, Ocular irritation, Acute Systemic Toxicity). Results demonstrated acceptable performance.
Shelf-life/SterilityDemonstrate sterility, maintain packaging integrity, and ensure device functionality over proposed shelf-life; ensure Boost does not hinder disinfection.Sterility maintained throughout shelf-life; packaging integrity maintained; bacteriostatic and fungistatic within 30-minute treatment; does not hinder ability of compatible daily cleaning solutions to disinfect.Sterilized by filtration and aseptically filled; sterilizing filtration, SIP, environmental monitoring, and aseptic filling were validated. Disinfection efficacy testing demonstrated Boost does not hinder daily cleaning solutions. Demonstrated to be bacteriostatic and fungistatic within 30 minutes. Shelf life of sterile barrier packaging established for 2 years (accelerated) and 12 months (real-time). Transit testing performed.
Performance - BenchLens/solution compatibility: Demonstrate compatibility with RGP contact lenses and care products. Contact angle/wettability: Demonstrate ability to maintain wettability. Thin film Ellipsometry/coating thickness: Demonstrate ability to restore Hydra-PEG coating. Shelf-life stability: Evaluate functional performance (wettability, coating thickness) and solution parameters.Compatibility demonstrated per ISO 11981:2009. Reduced contact angle compared to untreated control. Increased coating thickness to close to original value, effective over 12 months. Functional performance and solution parameters maintained through proposed shelf-life.Lens/solution compatibility per ISO 11981:2009 was performed. Data indicates Boost treatment reduces contact angle of lenses compared to untreated control (over 12 months). Data demonstrates Boost treatment increased coating thickness to close to original value and continued to be effective over 12 monthly treatments. Shelf-life stability testing was performed for functional performance and solution parameters.
Human Factors/Usability (HF/U)Assess patient's ability to properly use the device and identify unforeseen use errors.No unforeseen use errors or unacceptable residual risks identified. Users understand key instructions (usage frequency, disinfection, no direct eye contact), confirmed by reference material assessment.No unforeseen use errors or unacceptable residual risks identified. All participants understood they needed to disinfect lenses after Boost treatment and not place Boost solution directly in eyes. All knew treatment is monthly.
Clinical Study (Safety)Evaluate adverse events, slit lamp findings (e.g., corneal staining), and contact lens discontinuations. Demonstrate no increase in adverse events compared to control, no increase in corneal staining, and no discontinuations due to device.No serious adverse events (SAEs). Similar overall AE incidence to control. No increase in corneal staining for Boost group vs. placebo. No discontinuations.No reported SAEs. Similar overall AE incidence (19% Boost, 20% control); 2 AEs possibly related (1 mild allergic-type, 1 asymptomatic keratitis). No increase in corneal staining for Boost group compared to placebo group. No subjects discontinued from the study.
Clinical Study (Effectiveness)Evaluate lens performance (visual acuity (VA), RGP lens fit) and secondary endpoints (non-invasive tear break-up time (TBUT), subjective symptoms, preference). Demonstrate no reduction in VA, no decline in lens fit, maintained baseline TBUT.No statistically significant change/reduction in VA. No decline in lens fit; changes consistent with day-to-day wear. Maintained baseline TBUT (no worsening). Subjective patient-reported outcomes (CLDEQ, VAS, preference) did not show clinically/statistically significant reduction in discomfort or greater preference vs. placebo.No statistically significant change in VA for test and control groups. No decline in Lens Fit observed; changes not at higher rate for Boost group compared to placebo. Maintained baseline TBUT for both groups. CLDEQ, VAS, and preference results did not demonstrate clinically significant meaningful reduction in discomfort nor greater patient preference compared to placebo.

Study Details for Device Performance Evaluation:

This section addresses specific questions regarding the studies. Based on the provided text, the primary study demonstrating performance is the Clinical Performance Study, supplemented by various non-clinical/bench studies.

1. Sample sizes used for the test set and the data provenance:

  • Clinical Performance Study:

    • Sample Size: 30 subjects (20 test group, 10 control group).
    • Data Provenance: Not explicitly stated as to the country of origin, but implied to be a single-site study. It was a prospective, randomized, subject and investigator masked, placebo-controlled study conducted for 3 months.
  • Human Factors/Usability (HF/U) Study:

    • Sample Size: 15 laypersons.
    • Data Provenance: Not explicitly stated (retrospective/prospective or geographical origin). Implied to be prospective.
  • Bench and Biocompatibility Studies: Specific sample sizes for these tests (e.g., number of lenses, cells, animals) are generally not provided in the summary, other than "representative RGP contact lenses" for bench testing. The provenance is likely lab-based (pre-clinical).

2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Clinical Study:
    • The study mentions "in the opinion of the investigator" for inclusion criteria and "investigator" for assessing conditions. The qualifications of the investigator(s) are not specified beyond their role.
    • For slit lamp findings, VA measurement, and lens fit assessment, these would be medically trained professionals (e.g., ophthalmologists, optometrists), but the exact number of experts, their specific qualifications, or their role in establishing "ground truth" (e.g., independent adjudication) is not detailed.
  • Bench and Biocompatibility Studies: These are laboratory tests with defined protocols (e.g., ISO standards), so "expert" interpretation in the sense of human readers for medical imaging is less applicable. The "ground truth" is derived directly from the test results against established criteria.
  • HF/U Study: "Reference material assessment" was used, suggesting observation and participant feedback. No specific "experts" for ground truth establishment are noted beyond the study conductors.

3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

  • The text does not explicitly mention an adjudication method for the clinical study's primary or secondary endpoints (VA, lens fit, TBUT, subjective symptoms, AEs, corneal staining).
  • For the AEs, it provides a summary table and qualitative assessment (e.g., "possibly related"), but not a formal adjudication process by independent experts.

4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was NOT done. This device is a re-coating solution, not an AI or imaging diagnostic device. Therefore, the concept of "human readers improving with AI assistance" does not apply.

5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No, this is not an AI algorithm. This question is not applicable to the Tangible Boost device.

6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Clinical Study:
    • Direct Clinical Observation/Measurement: Visual acuity (measured), RGP lens fit (assessed by investigator), non-invasive tear break-up time (TBUT) (measured), corneal staining (scored).
    • Patient-Reported Outcomes (PROs): CLDEQ (questionnaire), VAS for comfort, preference questions.
    • Adverse Event Reporting: Based on subject reports and investigator assessment.
    • There is no mention of "expert consensus" or "pathology" in the context of ground truth establishment for clinical endpoints. The outcomes data (e.g., VA, lens fit) are considered the ground truth for evaluating the device's clinical performance.
  • Bench/Biocompatibility Studies: The ground truth is established by the defined methodologies and acceptance criteria of the ISO/ASTM standards being followed (e.g., cytotoxicity determined by cell viability, contact angle by measurement).

7. The sample size for the training set:

  • This question is not applicable as the device is not an AI/machine learning product and therefore does not have a "training set" in that sense.

8. How the ground truth for the training set was established:

  • This question is not applicable as there is no "training set" for this device.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR TANGIBLE BOOST

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Hydrophilic re-coating solution. A hydrophilic re-coating solution is a home use device intended to restore the hydrophilic coating of rigid gas permeable (RGP) contact lenses using reactive coating components.

NEW REGULATION NUMBER: 21 CFR 886.5919

CLASSIFICATION: Class II

PRODUCT CODE: QMM

BACKGROUND

DEVICE NAME: Tangible Boost

SUBMISSION NUMBER: DEN200002

DATE DE NOVO RECEIVED: January 13, 2020

Tangible Science, Inc. CONTACT: 740 Broadway Redwood City, CA 94063

INDICATIONS FOR USE

Tangible Boost is a monthly treatment to restore the Tangible Hydra-PEG coating and maintain the wettability of Tangible Hydra-PEG treated fluorosilicone acrylate rigid gas permeable lenses. Tangible Boost is intended for prescription (Rx) use only.

LIMITATIONS

Use of this device is limited to Hydra-PEG coated fluorosilicone acrylate rigid gas permeable (RGP) contact lenses (CLs).

The sale, distribution, and use of Tangible Boost is restricted to prescription use in accordance with 21 CFR 801.109.

Patient instructions are required for the proper use of the device. Labeling must include instructions on how to correctly use the device.

{1}------------------------------------------------

PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

The device restores Hydra-PEG coating (which naturally diminishes from the contact lens surface over time) and maintains wettability of Hydra-PEG coated fluorosilicone acrylate RGP CLs.

The device was evaluated in a 3-month clinical performance study in which subjects received new RGP CLs and performed monthly Boost treatment after 1 and 2 months of lens wear. Subjects also disinfected their lenses daily with a compatible multipurpose solution.

Clinical study subjective patient reported outcomes did not demonstrate clinically significant meaningful reduction in contact lens discomfort nor did they demonstrate greater patient preference, as compared to the placebo control group.

DEVICE DESCRIPTION

Tangible Boost is a monthly treatment specifically for RGP CLs made from fluorosilicone acrylate and manufactured with a Hydra-PEG coating. Tangible Boost is intended for prescription (Rx) use only, and for home use. Tangible Boost is not intended for lens disinfection and must be used in conjunction with a compatible multipurpose solution, as described in the labeling.

Tangible Boost restores the Tangible Hydra-PEG coating and maintains the wettability of Tangible Hydra-PEG treated fluorosilicone acrylate lenses. The device consists of two sterile, preservative-free, unit dose solutions to be combined in the single use, disposable Tangible Boost barrel style lens case (included in the device packaging). Patients who are allergic to any ingredient (e.g. Tangible Hydra-PEG) should not use this device.

As shown in Figure 1 below, to use Tangible Boost, the patient combines two tubes containing the Boost solutions into the Boost case. The patient rubs their lenses with compatible multipurpose solution for 20 seconds. Multipurpose solutions that are compatible with this device are Menicon Unique pH® and Boston SIMPLUS® Multi-Action Solution. Then, the patient places their lenses into the Boost case lens compartments and soaks the lenses in the combined Boost solution for 30 minutes.

After soaking, patients dispose of the used solution in the Boost case. Patients then rinse their lenses with the compatible multipurpose solution, by submerging the lenses for 20 seconds. The patient then removes the lenses from the Boost case and discards the case. After rinsing, the patient disinfects their lenses in a compatible multipurpose solution overnight, using a standard lens case. Only after this overnight disinfection step are the lenses ready to be worn.

Please refer to the labeling documents for a complete list of warnings, precautions and contraindications.

{2}------------------------------------------------

Step 1COMBINE both Tangible Boost solutions into Tangible Boost Case.
Step 2CLEAN lenses by rubbing in a compatible multipurpose cleaning anddisinfecting solution for 20 seconds.
Step 3SOAK lenses in Tangible Boost solution for 30 minutes.
Step 4DISPOSE of Tangible Boost solution and fill Boost case with a compatiblemultipurpose solution, and submerge lenses for 20 seconds to RINSE.
Step 5DISINFECT in a compatible multipurpose solution overnight using a standard lens case.Dispose of single use Boost case. Do NOT put lenses in eyes prior to disinfectionwith multipurpose solution.Menicon Unique pH® and Boston SIMPLUS® Multi-Action Solution are compatible multipurpose solutions.

Figure 1. Tangible Boost

Figure 1 shows Tangible Boost instructions for use. These instructions are excerpted from the device labeling.

SUMMARY OF NONCLINICAL/BENCH STUDIES

BIOCOMPATIBILITY/MATERIALS

The Tangible Boost solutions were assessed (Table 1) for cytotoxicity, sensitization, ocular irritation, acute oral toxicity, and ocular biocompatibility endpoints. The biocompatibility testing was performed in accordance with the FDA guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", International Standard Organization (ISO) 10993-1 - Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, - Part 5: Tests for in vitro cytotoxicity, - Part 10: Tests for irritation and skin sensitization, and -Part 11: Tests for systemic toxicity, and ISO 9394- Ophthalmic optics - Contact lenses and contact lens care products-Determination of biocompatibility by ocular study with rabbit eyes. Results demonstrated acceptable performance.

TestPurposeMethodAcceptanceCriteriaResults
Evaluate the potential for11 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1Carlos Categorial Collection Comercial Comercial Comercial Concessione of Cardinal Comercial Concessiones of Cardinal Concessionalism Concessiones of Carder Comercial ConcessMEM Elution w/ L-000 1 6 - T 1 - 1 1Non-

929 Mouse Fibroblast

Cells (ISO 10993-5)

Table 1. Biocompatibility Testing on Tangible Boost Solutions
---------------------------------------------------------------------

cellular toxicity of the

device.

Cytotoxicity

Pass

cytotoxic

{3}------------------------------------------------

TestPurposeMethodAcceptanceCriteriaResults
SensitizationEvaluate the sensitizationcapacity of the device.(b) (4)Maximization (ISO10993-10)Non-sensitizerPass
Ocular irritationEvaluate the potential ofthe device to induce ocularirritation.Ocular irritation (ISO10993-10)Non-irritantPass
Acute OralToxicityEvaluate the potential fororal systemic toxicity.Acute Oral Toxicity(ISO 10993-11)Non-toxicPass
OcularBiocompatibilityEvaluate the ocularbiocompatibility.Ocular toxicity (ISO9394)Non-toxicPass

All biocompatibility testing was conducted in accordance with the provisions of 21 CFR 58, Good Laboratory Practice for Nonclinical Laboratory Studies.

The Tangible Boost packaging was assessed (Table 2) for cytotoxicity, ocular irritation and acute systemic toxicity testing. The biocompatibility testing was performed in accordance with International Standard Organization (ISO) 10993-1 - Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process, - Part 5: Tests for in vitro cytotoxicity, - Part 10: Tests for irritation and skin sensitization, and - Part 11: Tests for systemic toxicity. Results demonstrated acceptable performance.

TestPurposeMethodAcceptanceCriteriaResults
Cytotoxicity:PackagingEvaluate the potential forcellular toxicity of theTangible Boost packaging.MEM Elution w/ L-929 Mouse FibroblastCells (ISO 10993-5)Non-cytotoxicPass
Ocular irritation:PackagingEvaluate the potential ofthe Tangible Boostpackaging to induce ocularirritation.Ocular irritation (ISO10993-10)Non-irritantPass
Acute SystemicToxicity :PackagingEvaluate the potential forocular systemic toxicity ofthe Tangible Boostpackaging.Acute OcularToxicity (ISO 10993-11)Non-toxicPass

Table 2. Biocompatibility Testing on Tangible Boost Packaging

{4}------------------------------------------------

All biocompatibility testing was conducted in accordance with the provisions of 21 CFR 58, Good Laboratory Practice for Nonclinical Laboratory Studies.

Biocompatibility testing demonstrated that the Tangible Boost solutions and packaging do not illicit a toxicological response. It is important to note that the Boost Case was previously cleared under K991206.

SHELF LIFE/STERILITY

Tangible Boost is sterilized by filtration and is then aseptically filled into single-use vials. The sterilizing filtration process was validated following recommendations in ANSI/AAMI/ISO 13408-2 Aseptic Processing of Health Care Products - Part 2: Sterilizing Filtration. Steam is used for sterilization in place (SIP) of the sterile circuit and the success of the SIP cycles is confirmed with biological indicators. Environmental monitoring and the aseptic filling process were validated following recommendations in the FDA guidance document Sterile Drug Products Produced by Aseptic Processing - Current Good Manufacturing Practice. Disinfection efficacy testing was performed following recommendations in ISO14729:2001 Ophthalmic Optics - Contact Lens Care Products - Microbiological Requirements and Test Methods for Products and Regimens for Hygienic Management of Contact Lenses to demonstrate that Tangible Boost does not hinder the ability of daily cleaning solutions to disinfect. Tangible Boost was demonstrated to be bacteriostatic and fungistatic within the 30minute treatment time frame following recommendations in the FDA guidance document Guidance for Industry - Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products. Sterility testing will be performed for lot release for each batch of Tangible Boost in accordance with USP <71>.

Shelf life of the sterile barrier packaging was established for 2 years under accelerated aging conditions and 12 months under real time conditions following recommendations in the FDA guidance document Guidance for Industry - Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products using the vacuum decay container closure integrity test (ASTM F2338 Standard Test Methods for Nondestructive Detection of Leaks in Packages by Vacuum Decay Method). Transit testing was performed to evaluate the ability of Tangible Boost to withstand possible stressors during shipment following recommendations in ASTM D4332 Standard Practice for Conditioning Containers. Packaging Components for Testing and ASTM 4169 Standard Practice for Performance Testing of Shipping Containers and Systems. The transit events simulated in the testing included manual handling, stacking (compression), loose load, low pressure, vehicle vibration, and concentrated impact, as recommended by ASTM D4169.

PERFORMANCE TESTING - BENCH

The following chemistry/materials testing was performed for the Tangible Boost Solution on representative fluorosilicone acrylate rigid gas permeable (RGP) contact lenses :

  • . Lens/solution compatibility testing per ISO 11981:2009 was performed to demonstrate the compatibility of representative RGP contact lenses and associated care products (compatible care products are specified in the labeling)

{5}------------------------------------------------

  • Contact angle/wettability testing to demonstrate the ability of the device to maintain the wettability of compatible RGP lenses. See Figure 2.
  • . Thin film Ellipsometry/coating thickness testing to demonstrate the ability of the device to restore the Hydra-PEG coating on compatible RGP lenses. See Figure 3.
  • Shelf-life stability testing to evaluate the functional performance (wettability and coating thickness) and solution parameters of the boost solution though the duration of the proposed shelf-life

(b) (4)

Figure 2. Contact Angle Measurement

Contact angle of Tangible Hydra-PEG (HPT) coated fluorosilicone acrylate contact lenses was measured over 12 months with or without monthly Tangible Boost treatment. The darker point/lines are advancing contact angles and lighter points/lines are the receding angles. The data indicates that as compared to untreated control, Tangible Boost treatment reduces the contact angle of the lenses.

{6}------------------------------------------------

Image /page/6/Figure/0 description: The image shows the text '(b) (4)' in the upper left corner. The text is small and black. The rest of the image is a solid gray color.

Figure 3. Ellipsometry/Coating Thickness

Delta values were measured by an ellipsometer. The difference in delta values between the coated and uncoated surface (Delta) is proportional to the coating thickness. The thickness of the coating over time was measured across multiple cycles of simulated wear and Tangible Boost treatments or control (no Tangible Boost treatment). This data demonstrates that Tangible Boost treatment increased coating thickness to close its original value and that Tangible Boost treatment continued be effective over 12 monthly treatments.

HUMAN FACTORS/USABILITY TESTING

To assess the ability of a patient to properly use the two-component Tangible Boost system, a human factors/usability (HF/U) study was performed according to the Human Factors guidance document, Applying Human Factors and Usability Engineering to Medical Devices. Testing was conducted with 15 laypersons from male and female users from all age groups to assess all tasks and to monitor for unforeseen use errors that could lead to serious harm.

The results of the HF/U study did not identify any unforeseen use errors or unacceptable residual risks. The reference material assessment of the HF/U study confirmed that all participants understood that they needed to disinfect the lenses after Boost treatment, and that they should not place the Boost solution directly in their eyes. All participants also demonstrated that they knew the treatment is intended to be monthly, as listed on the outer box.

{7}------------------------------------------------

The human factors/usability testing results demonstrated that the test participants understand the key information such as usage frequency, disinfection requirement, no direct eye contact with solution.

SUMMARY OF CLINICAL INFORMATION

A clinical study was conducted to evaluate safety and effectiveness of Tangible Boost to replenish Hydra-PEG coating on RGP contact lenses. It was a 3-month, subject and investigator masked, randomized, placebo-controlled study. Subjects were fitted with Hydra-PEG treated fluorosilicone acrylate RGPs (corneal or scleral). After 1 and 2 months of lens wear, subjects treated their lenses with Tangible Boost or saline placebo. Care regime utilized Unique pH multipurpose solution (dispensed for the duration of the trial) and subjects wearing scleral lenses also received Purilens saline as a filling solution. Sample size was 30 subjects; 20 test and 10 control.

Primary safety endpoints included assessment of Adverse Events and slit lamp findings including corneal staining, as well as contact lens discontinuations. Primary effectiveness endpoints included assessment of lens performance as evidenced by no reduction VA and RGP lens fit. Secondary efficacy endpoints included non-invasive tear break-up time (TBUT) and several patient-reported assessments of subjective symptoms and preference.

Study VisitsThere was a minimum of five study visits:
1. Baseline
2. Lens dispensing visit
3. Tangible Boost dispensing visit
4. 1-day follow-up
5. 60-day follow-up
Inclusion criteria• Willing and able to sign the informed Consent form
• Written documentation has been obtained in accordance with therelevant country and local privacy requirements, where applicable
• Male or female
• 18 years of age and older prior to the initial visit
• Established scleral or corneal rigid contact lens wearer
• In the opinion of the investigator, the subject has the ability to followstudy instructions
• In the opinion of the investigator, the subject has the ability tocomplete all study procedures and visits
Exclusion criteria• Has never worn either scleral or corneal contact lenses before
• Does not possess a usable pair of spectacles
• Has any ocular disease or abnormality that would affect the wearingof contact lenses
• Is aphakic (i.e., missing their natural lens inside the eye)

Table 3. Clinical Study Schedule Summary and Inclusion/Exclusion Criteria

{8}------------------------------------------------

• Is currently participating in any other type of eye-related clinical or research study
• Is pregnant or nursing as reported by the subject
• Has a condition or is in a situation which, in the investigator's opinion, may put the subject at significant risk, may confound study outcomes, or may significantly interfere with the subject's participation in the study
• Has a known and self-reported allergy to components in the device
• Has had previous ocular surgery within the past 12 weeks
• Is unable to consent
• Is a prisoner

Primary Efficacy Endpoints:

  • Visual Acuity (VA): No decline in VA throughout the study, endpoint was met. o No statistically significant change in VA for test and control group observed during pre-treatment, 1 day post-treatment and final visit follow-up.
    (b) (4)

Figure 4. Visual Acuity

Visual acuity was measured on a logMAR scale prior to treatment, 1 day after treatment, and at the final follow-up visit.

Lens Fit: No decline in Lens Fit throughout the study, endpoint was met. Changes in lens fit were not observed at higher rate for Boost group compared to placebo (amount consistent with changes expected day-to-day or based on number of hours of wear prior to measurement).

{9}------------------------------------------------

Table 4. Movement

BoostPlacebo
ScleralCornealScleralCorneal
No change(b) (4)
Movement improved
Increased movement
causing destabilization
Loss of movement
leading to poor fit
Other

Table 5. Centration

BoostPlacebo
ScleralCornealScleralCorneal
No change(b) (4)
Centration improved
Change of centration,still adequate
Decenteredinadequately
Other
Total

Table 6. Apical Clearance

BoostPlacebo
ScleralCornealScleralCorneal
No change(b) (4)
Clearance increased,still adequate
Clearance decreased,still adequate
Inadequately lowclearance
Excess clearance
Total

Table 7. Limbal Clearance

BoostPlacebo
ScleralCornealScleralCorneal

{10}------------------------------------------------

No change(b) (4)
Clearance increased,still acceptable
Clearance decreased,still adequate
Inadequate clearance
Excess clearance
Total

Table 8. Landing Zone

BoostPlacebo
ScleralCornealScleralCorneal
No change(b) (4)
Developing ofblanching
Development ofbearing
Improvement ofblanching/bearing
Improvement in lensedge lift
Development of lensedge lift
Inadequate edge fit
Total

Secondary Effectiveness Endpoints:

  • Tear Film Break-Up Time (TBUT): maintained baseline TBUT (no worsening) o for both Boost and Placebo subject groups at the timepoints shown in the graphbelow.

{11}------------------------------------------------

Figure 5. Tear Film Break-Up Time (TBUT)

Non-invasive keratography tear break-up time was used to evaluate the average tear film break-up time. Measurements were taken at V3 (prior to treatment, affer 1 month in Hydra-PEG lenses), V4, (1 day after first Boost treatment), and V5 (final follow-up).

  • Subjective Symptoms: ●
    • 0 Contact Lens Dry Eye Questionnaire (CLDEQ): Patient-reported changes in symptoms assessed by the CLDEQ questionnaire were not clinically significant over the course of study.

(b) (4)

{12}------------------------------------------------

Image /page/12/Figure/0 description: The image shows the characters '(b) (4)' in the upper left corner of the image. The characters are in black font. The background of the image is gray.

Figure 6. Contact Lens Dry Eye Questionnaire (CLDEQ)

CLDEQ survey results at the pre-treatment visit 3 (baseline in Hydra-PEG lenses), 1 day after first Boost treatment (visit 4), and final follow-up (visit 5).

  • Visual analog scales (VAS) for lens comfort score: Patient-reported O changes in comfort were not clinically significant nor statistically significant over the timepoints shown in the plots below.
    (b) (4)

Figure 7. Visual analog scales (VAS)

VAS scores at the pre-treatment visit (visit 3), 1 day after first Boost treatment (visit 4), and final follow-up (visit 5).

{13}------------------------------------------------

  • Preference: Subjects were asked whether they preferred lenses before о Boost treatment, after treatment, or no preference. Patient-reported changes in comfort were not clinically significant over the course of study. Additionally, the noted differences were not statistically significant.
    Table 9. Boost preference at visit 4 (1 day after first Boost treatment) and visit 5 (final follow-up)
Visit 4Visit 5
BoostPlaceboBoostPlacebo
pre-Boost(b)(4)
post-Boost
No preference

Safety Endpoints:

  • Adverse Events (AEs): There were no reported serious adverse events (SAEs), 0 and similar overall AE incidence (19% Boost group, 20% control group).
    • . 1 AE possibly related to Boost - mild allergic-type response to CLs after Boost treatment, reaction similar to the patient's initial response to Hydra-PEG coated lenses indicating possible allergy or sensitivity to Hydra-PEG coating. Patients with Hydra-PEG allergy would be expected to discontinue wearing Hydra-PEG CLs. making them ineligible for Boost Rx. Patients with allergy or sensitivity to Tangible Hydra-PEG shouldn't use Boost.
    • . 1 AE possibly related to Tangible Boost but unlikely - keratitis, was asymptomatic and resolved by next visit without treatment (this type of reaction noted to be relatively common with scleral lenses due to trapping of tear solution under lens bowl), patient went on to use Boost a second time without adverse outcome.

{14}------------------------------------------------

SubjectTreatmentArmDescriptionRelated toTangible Boosttreatment?
01UnassignedDeveloped corneal hydrops (common withadvanced keratoconus) prior to lens dispense orassignment to treatment group.No
12BoostDeveloped upper respiratory infection, no ocularsymptoms. Prescribed Keflex by outside doctor.No
05BoostMild allergic-type response (grittiness, redness,discomfort) to lenses after Boost treatment. UsedProgent to strip Hydra-PEG coating, discontinueduse of Boost, and used artificial tears for 1-2 days.No additional treatment was necessary. Patientreported that these symptoms were also presentupon initial receipt of the Hydra-PEG lenses, butthat symptoms resolved after ~1.5 weeks.Possibly, butsymptoms werepresent prior toBoosttreatment.
19PlaceboMild swelling and hyperemia of inferior nasalconjunctiva. Instructed to discontinue lens wear for3-4 days.No
09PlaceboSubject reported for visit with red sore left eye.Reported getting little sleep the past few days andstated that this has occurred in the past. Subjectused artificial tears to relieve symptoms.No
32BoostDeveloped contact lens-associated acute red eye(CLARE) after lens dispense, which was resolvedprior to being seen by study investigator 2 daysafter event. Exhibited preservative toxicity-typecorneal sensitivity reaction. Subject was educatedabout rinsing the lens bowl prior to insertion andissue resolved. Occurred prior to Boost dispense.No
25BoostDeveloped asymptomatic keratitis which resolvedwithout treatment. Subject was monitored, andcondition was resolved by the next study visit.Possibly related

Table 10. Summary of Adverse Events

  • Corneal Staining: No increase in corneal staining for Boost group compared to o placebo group

{15}------------------------------------------------

(b) (4)

Figure 8. Corneal Staining

Corneal staining scores at pretreatment visit (visit 3), 1 day after first Boost treatment (visit 4), and final follow-up (visit 5). Corneal staining was evaluated on a scale of 6101in each of 5 regions. This graph shows the total of the scores from the 5 regions.

  • Discontinuations: No subjects were discontinued from the study. 0

Pediatric Extrapolation

In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.

LABELING

The device labeling is adequate and meets the requirements of 21 CFR 801.109.

Package insert includes device description, Indications for Use with requirement for prescription use only, contraindications, warnings, precautions, potential adverse reactions, and instructions for use including compatible multipurpose solutions for post-Boost and daily disinfection of lenses.

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of a hydrophilic recoating solution and the measures necessary to mitigate these risks.

Identified RisksMitigation Measures
Adverse events leading to eye irritation(redness, burning, stinging, discomfort, pain), infection, keratitis, corneal ulcer, loss of visual acuity, or allergic reactionClinical performance testingHuman factors evaluationLabeling
Adverse tissue reactionBiocompatibility evaluation

Table 11 - Identified Risks to Health and Mitigation Measures

{16}------------------------------------------------

Lens solution compatibility testing
Coating effectiveness testing
Labeling
InfectionSterility testing and validation
Disinfection solution compatibility
testing
Shelf life testing
Labeling
Use error/ improper device use leading toClinical performance testing
eye irritation (redness, burning, stinging,Human factors evaluation
discomfort, pain), infection, keratitis,Coating performance testing
corneal ulcer, loss of visual acuityLabeling

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the hydrophilic re-coating solution is subject to the following special controls:

    1. Clinical performance testing must evaluate device safety as assessed by adverse events, slit lamp findings, and maintenance of visual acuity.
    1. The patient contacting components of the device and packaging components must be demonstrated to be biocompatible.
    1. Performance testing must demonstrate the sterility of the device.
    1. Use-related risk analysis must be performed to determine if a self-selection study and human factors validation study must be conducted to demonstrate that users can correctly use the device based solely on reading the directions for use.
    1. Performance data must support the shelf-life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
    1. Performance testing must demonstrate compatibility with each lens and solution labeled for use with the device.
    1. Performance testing must demonstrate the ability of the device to restore the coating of compatible lenses.
    1. Labeling must include the following:
    • a. Instructions on how to correctly use the device, including instructions to use fresh components for each use:
    • b. Descriptions of compatible contact lenses;
    • Descriptions of compatible care solutions; C.

{17}------------------------------------------------

  • d. A warning that if patients are not sure of their lens material, they should contact their health care provider prior to use: and
  • e. A precaution against use with lenses that have not been demonstrated to be compatible with the device.

BENEFIT-RISK DETERMINATION

The risks and benefits of the Tangible Boost device are based on non-clinical performance data as well data collected in a clinical study. The device demonstrated an acceptable safety profile in the clinical study, with no SAEs and similar overall AE incidence (2 possibly device-related, a mild allergic-type reaction to Hydra-PEG and a transient asymptomatic keratitis consistent with scleral lens wear), no increase in corneal staining, and no lens discontinuations. Risks associated with use of the Boost device (or mis-use) include adverse reactions leading to eve irritation (redness, burning, stinging, discomfort, pain), infection, keratitis, corneal ulcer, loss of visual acuity, or allergic reaction.

The probable benefits of the device are based on non-clinical laboratory and bench data, as well as data collected in a clinical study. Boost treatment restores Hydra-PEG coating and maintains contact lens wettability as demonstrated by wetting angle data (bench testing). In the clinical study, compared to a placebo control group, the device did not result in reduction of visual acuity, change in lens fit, or worsening of tear break-up time. Clinical study subjective patientreported outcomes did not demonstrate clinically significant meaningful reduction in contact lens discomfort nor did they demonstrate greater patient preference, as compared to the placebo control group.

Patient Perspectives

Patient preference information was provided as described in Table 9 above. The information indicated no clinically meaningful or statistically significant difference in preference in terms of lens comfort with use of Tangible Boost solution.

Benefit/Risk Conclusion

In conclusion, given the available information above, the probable benefits outweigh the probable risks for Tangible Boost. The device provides probable benefits, and the probable risks can be mitigated by use of general controls and the identified special controls.

CONCLUSION

The De Novo request for Tangible Boost is granted and the device is classified as follows:

Product Code: QMM Device Type: Hydrophilic re-coating solution Class: II Regulation Number: 21 CFR 886.5919

§ 886.5919 Hydrophilic re-coating solution.

(a)
Identification. A hydrophilic re-coating solution is a home use device intended to restore the hydrophilic coating of rigid gas permeable (RGP) contact lenses using reactive coating components.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must evaluate device safety as assessed by adverse events, slit lamp findings, and maintenance of visual acuity.
(2) The patient contacting components of the device and packaging components must be demonstrated to be biocompatible.
(3) Performance testing must demonstrate the sterility of the device.
(4) Use-related risk analysis must be performed to determine if a self-selection study and human factors validation study must be conducted to demonstrate that users can correctly use the device based solely on reading the directions for use.
(5) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life.
(6) Performance testing must demonstrate compatibility with each lens and solution labeled for use with the device.
(7) Performance testing must demonstrate the ability of the device to restore the coating of compatible lenses.
(8) Labeling must include the following:
(i) Instructions on how to correctly use the device, including instructions to use fresh components for each use;
(ii) Descriptions of compatible contact lenses;
(iii) Descriptions of compatible care solutions;
(iv) A warning that if patients are not sure of their lens material, they should contact their health care provider prior to use; and
(v) A precaution against use with lenses that have not been demonstrated to be compatible with the device.