K Number
DEN190028

Validate with FDA (Live)

Date Cleared
2020-04-16

(328 days)

Product Code
Regulation Number
862.3245
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MyCare Psychiatry Clozapine Assay Kit is intended for the in vitro quantitative measurement of clozapine in adult human serum using automated clinical chemistry analyzers. Measurements obtained can be used to aid in the management of individuals prescribed clozapine for treatment-resistant schizophrenia. This assay should be used in conjunction with other clinical and laboratory findings and results from this test alone should not be used to make treatment decisions.

Device Description

The MyCare Psychiatry Clozapine Assay Kit is a homogenous two reagent nanoparticle agglutination assay used for detection of clozapine in human serum. It is based on competition between drug and drug-conjugates for binding to drug-specific antibodies covalently bound to nanoparticles. The extent of particle aggregation can be followed spectrophotometrically on clinical chemistry analyzers. This aggregation is measured at a wavelength of around (b)(4) by automated clinical chemistry analyzers.

The assay contains:

  • Reagent 1 (R1) reaction buffer that contains drug-conjugate in a buffered protein . solution
  • . Reagent 2 (R2) - nanoparticle reagent that contains clozapine-specific monoclonal antibody bound to nanoparticles in a buffered solution
AI/ML Overview

Here's a breakdown of the acceptance criteria and the study proving the MyCare Psychiatry Clozapine Assay Kit meets them, based on the provided text:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria/Study GoalReported Device Performance / Outcome
Precision/ReproducibilityTo evaluate precision according to CLSI EP05-A3. (Demonstrate consistent results under repetitive conditions).Achieved acceptable CVs (Coefficient of Variation) for within-run and total precision across various clozapine concentrations in serum pools, spiked serum, and controls. Total CVs ranged from 4.0% to 7.8%.
Linearity/Reportable RangeTo evaluate linearity across the measuring range of the assay following CLSI EP06-A. (Demonstrate accurate measurement across a range of concentrations). Claimed range: 68-1500 ng/mL.Supports claimed measuring range of 68 to 1500 ng/mL. For the worst performing lot, linear regression showed: Slope = 0.920 (0.910 to 0.930), Intercept = 4.1 (1.4 to 6.8), R = 0.9985.
TraceabilityTraceability to a certified reference standard. (Ensure measurements are linked to accepted standards).MyCare Psychiatry Calibrator Kit is traceable to a certified USP clozapine reference standard. Master calibrators verified by LC-MS/MS.
Detection LimitTo determine Limit of Blank (LOB), Limit of Detection (LOD), and Limit of Quantitation (LOQ) according to CLSI EP17-A2. (Establish the lowest concentration the device can reliably detect and quantify).LOB determined to be (b)(4) (specific value redacted, but determined and reported). LOD determined to be 39 ng/mL. LOQ determined to be 68 ng/mL.
Analytical SpecificityPotentially Cross-Reacting Substances: Evaluate cross-reactivity with clozapine metabolites (Clozapine-N-oxide, 8-Hydroxy-8-deschloro-clozapine, Norclozapine) in the presence of clozapine. (Ensure the device measures clozapine specifically, without significant interference from related compounds).No cross-reactivity exceeding ±10% was observed. Clozapine-N-oxide: 3%, 8-Hydroxy-8-deschloro-clozapine: 9%, Norclozapine: 1%.
Endogenous & Exogenous Interfering Substances: Evaluate interference from various endogenous substances (Bilirubin, Hemolysate, Human IgG, Human Serum Albumin, Rheumatoid Factor, Triglycerides) and a wide range of exogenous drugs. (Ensure other common substances in blood or co-administered drugs do not significantly affect results).None of the tested endogenous or exogenous substances were found to lead to clinically significant interference (> ±10%) for the device at the tested concentrations. (Extensive list of compounds tested and reported as having no significant interference).
Method Comparison StudyTo compare results from the MyCare Psychiatry Clozapine Assay Kit with a validated clozapine LC-MS/MS method using clinical samples. (Demonstrate agreement with a gold standard method).Deming regression analysis showed: y = 1.037 (0.939 to 1.135) x - 27.8 (-78.3 to +22.7); R = 0.9269. Accuracy (difference within LC-MS/MS) was 30% within 10% for 68-350 ng/mL range, 34% within 10% for 350-1000 ng/mL range, and 50% within 10% for 1000-1500 ng/mL range.

Important Note: The document consistently contains redacted information, marked as (b)(4), particularly for specific numerical values (e.g., sample sizes, number of replicates, actual LOB values). This means some of the specific "Reported Device Performance" details are generalized based on the description of the study rather than exact numbers.

Study Design Details

  1. Sample sizes used for the test set and the data provenance:

    • Precision Study:
      • N=480 for each of the 9 samples/controls tested across all reagent lots and analyzers (9 * 480 = 4320 total measurements for precision).
      • Provenance: Clinical samples from patients taking clozapine therapy for serum pools; spiked serum samples; control materials. Data is likely retrospective or a mix, prepared as described from clinical samples and spiked.
    • Linearity Study:
      • 11 different clozapine levels, each tested in 5 replicates, using 3 lots of reagents and 2 analyzers. (11 levels * 5 replicates * 3 lots * 2 analyzers = 330 measurements).
      • Provenance: Samples prepared by pooling human serum specimens and spiking with clozapine.
    • Detection Limit Studies:
      • LOB: Each of the "n=" (redacted) serum samples tested in duplicate on 3 different days on 2 different analyzers for "n=" (redacted) runs, using "lots" (redacted) of reagents.
      • LOD: Each of the "n=" (redacted) human serum samples tested in duplicate on 3 different days on "n=" (redacted) different analyzers for "n=" (redacted) runs, using "lots" (redacted) of reagents.
      • LOQ: Each of the 4 individual normal human serum samples tested "n=" (redacted) times on "b)(4)" (redacted) different days on "analyzer" (singular, redacted), using "lots" (redacted) of reagents.
      • Provenance: Normal human serum samples, some spiked with clozapine.
    • Analytical Specificity (Cross-reactivity & Interference):
      • Cross-reactivity: (b)(4) (redacted) replicates per compound.
      • Interference: Each sample tested at a minimum of 5 runs.
      • Provenance: Human serum pools for interference studies.
    • Method Comparison Study:
      • (b)(4) (redacted) samples were used for Deming regression.
      • Counted samples in ranges provided: 50 + 71 + 2 = 123 unique samples or measurements explicitly tabulated, though the total for Deming regression is redacted.
      • Provenance: Clinical samples collected at six different U.S. clinical sites from patients receiving clozapine therapy. This indicates prospective collection from diverse U.S. clinical settings.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    • This is a quantitative assay (measuring clozapine concentration), not an imaging or diagnostic interpretation task that typically requires expert readers for ground truth.
    • The ground truth for the method comparison study was established by a "validated clozapine LC-MS/MS method." LC-MS/MS (Liquid Chromatography-Mass Spectrometry/Mass Spectrometry) is a highly accurate and sensitive analytical technique often considered a "gold standard" for precise quantification of small molecules in biological samples. It does not involve human expert interpretation in the same way as, for example, a radiologist reading an MRI.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. As a quantitative assay compared to an analytical gold standard (LC-MS/MS), there's no "adjudication" in the sense of reconciling differing human interpretations. The LC-MS/MS results serve as the objective reference.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a device for measuring a chemical concentration, not an AI-assisted diagnostic tool that would typically involve human readers.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, implicitly. The entire performance evaluation (precision, linearity, detection limits, specificity, and method comparison) assesses the standalone analytical performance of the assay kit (the "algorithm" equivalent in this context, though it's a biochemical assay, not software) on an automated clinical chemistry analyzer, without human interpretation as part of its measurement function. Human input is for sample preparation, loading, and instrument operation, but not for the actual clozapine measurement itself.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The primary ground truth used for the method comparison study was a validated clozapine LC-MS/MS method. This is an analytical gold standard.
    • For other analytical performance characteristics (precision, linearity, detection limits), the ground truth was established by preparation of samples with known concentrations (e.g., spiking clozapine into serum).
  7. The sample size for the training set:

    • This is an in vitro diagnostic (IVD) assay kit, not a machine learning algorithm that requires a "training set" in the conventional sense. The "development" of such a kit involves chemical formulation, optimization of reagents, and validation of the measurement principles, not data-driven model training.
    • Therefore, there is no discrete "training set" of patient data as might be found for an AI/ML device.
  8. How the ground truth for the training set was established:

    • Not applicable, as there's no "training set" for this type of device. The "ground truth" during the development of the assay would have been established by precisely preparing solutions of known clozapine concentrations and optimizing the assay's chemical reactions and detection parameters to accurately reflect those concentrations.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR MyCare Psychiatry Clozapine Assay Kit DECISION SUMMARY

A. DEN Number:

DEN190028

B. Purpose for Submission:

De Novo request for evaluation of class III designation of the MyCare Psychiatry Clozapine Assay Kit

C. Measurand:

Clozapine (8-chloro-11-(4-methyl-1-piperazinyl)-5H-dibenzo [b,e] [1,4] diazepine)

D. Type of Test:

Homogeneous nanoparticle agglutination immunoassay

E. Applicant:

Saladax Biomedical, Inc.

F. Proprietary and Established Names:

MyCare Psychiatry Clozapine Assay Kit

G. Regulatory Information:

    1. Regulation section:
      21 CFR 862.3245
    1. Classification:
      Class II (Special Controls)
    1. Product code:
      QKT

{1}------------------------------------------------

    1. Panel:
    • 91- Toxicology

H. Intended Use:

    1. Intended use(s):
      See indications for use
    1. Indication(s) for use:
      The MyCare Psychiatry Clozapine Assay Kit is intended for the in vitro quantitative measurement of clozapine in adult human serum using automated clinical chemistry analyzers. Measurements obtained can be used to aid in the management of individuals prescribed clozapine for treatment-resistant schizophrenia. This assay should be used in conjunction with other clinical and laboratory findings and results from this test alone should not be used to make treatment decisions.
    1. Special conditions for use statement(s):
      For prescription use only For In Vitro Diagnostic Use Only
    1. Special instrument requirements:
      The assay was validated on the Beckman Coulter AU480 Clinical Analyzer

I. Device Description:

The MyCare Psychiatry Clozapine Assay Kit is a homogenous two reagent nanoparticle agglutination assay used for detection of clozapine in human serum. It is based on competition between drug and drug-conjugates for binding to drug-specific antibodies covalently bound to nanoparticles. The extent of particle aggregation can be followed spectrophotometrically on clinical chemistry analyzers. This aggregation is measured at a wavelength of around (b)(4) by automated clinical chemistry analyzers.

The assay contains:

  • Reagent 1 (R1) reaction buffer that contains drug-conjugate in a buffered protein . solution
  • . Reagent 2 (R2) - nanoparticle reagent that contains clozapine-specific monoclonal antibody bound to nanoparticles in a buffered solution

{2}------------------------------------------------

J. Standard/Guidance Document Referenced (if applicable):

  • . CLSI EP05-A3 Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline - Third Edition, October 2014.
  • CLSI EP06-A Evaluation of the Linearity of Quantitative Measurement Procedures: A ● Statistical Approach; Approved Guideline, April 2003.
  • CLSI EP17-A2 Evaluation of Detection Capability for Clinical Laboratory Measurement ● Procedures; Approved Guideline - second edition, June 2012.
  • CLSI EP25-A Evaluation of Stability of In vitro Diagnostic Reagents; Approved ● Guideline, September 2009.

K. Test Principle:

Drug-conjugates serve as binding partners to antibodies selective for clozapine which are covalently attached to the surface of nanoparticles. In the absence of free clozapine, this reaction creates large aggregates, resulting in a solution that scatters incident light and leads to an increase in the observed absorption of the solution. When incubated with a sample containing clozapine, the free clozapine in the sample binds to the reagent antibodies, and those antibody-nanoparticle conjugates are no longer available to aggregate, resulting in less scattering of incident light and lower observed absorption of the solution. Thus, maximum absorption occurs with low levels of drug, and minimum absorption occurs with high levels of drug.

L. Performance Characteristics (if/when applicable):

    1. Analytical performance:
    • a. Precision/Reproducibility:

A precision study was conducted according to recommendations in CLSI EP05-A3 using 4 human serum sample pools, which were prepared from clinical samples from patients taking clozapine therapy, serum pools spiked with clozapine, and 3 levels of controls. Each sample was assessed in replicates of (b)(4) a day, for (b)(4) days. Samples were assessed using three lots of reagents, calibrators, and controls on two Beckman Coulter AU480 Clinical Analyzers. The results are summarized below.

Sample Results Across All Reagent Lots and Analyzers

SampleNMean(ng/mL)Within - RunTotal
CVCV
Serum pool 14801484.6%5.6%
Serum pool 24803383.3%4.0%

{3}------------------------------------------------

SampleNMeanWithin - RunTotal
(ng/mL)CVCV
Serum pool 34805773.3%4.0%
Serum pool 44809263.5%4.8%
Spiked Serum 1480986.0%7.8%
Spiked Serum 24801,0944.4%6.6%
Control 14801564.5%5.4%
Control 24804743.7%4.4%
Control 34809453.8%4.5%

b. Linearity/assay reportable range:

A study was conducted to evaluate linearity across the measuring range of the assay following the CLSI EP06-A guideline. Samples with 11 different clozapine levels were evaluated: 55, 132, 169, 352, 491, 588, 821, 1003, 1214, 1467, and 1554 ng/mL. Two samples were prepared by pooling human serum specimens that were then spiked with clozapine to low and high clozapine concentrations, which were then mixed to produce samples with the concentrations tested. Each sample was tested in five replicates using three lots of reagents, calibrators, and controls and two Beckman Coulter AU480 clinical analyzers. The observed values were plotted against the expected values and linear regression analysis was performed. The linear regression results from the worst performing lot of three lots are shown below.

ClaimedRangeSlopeInterceptR
68-1500 ng/mL0.920(0.910 to0.930)4.1(1.4 to 6.8)0.9985

The 95% confidence intervals are shown for slope and intercept.

These results support the claimed measuring range of 68 to 1500 ng/mL for clozapine.

  • Traceability, Stability, Expected values (controls, calibrators, or methods): C.

Traceability

The MyCare Psychiatry Calibrator Kit is traceable to a certified USP clozapine reference standard. Master calibrators are prepared by (b) (4) of this

{4}------------------------------------------------

certified reference material in buffer and verified by LC-MS/MS.

  • d. Detection limit:
    Detection limits were assessed according to recommendations in CLSI EP17-A2.

The limit of blank (LOB) was determined using the non-parametric "classical approach" using "individual normal(b) (4) serum samples containing no clozapine. Each of the serum samples were tested in duplicate on three different days on two different analyzers for n="", using" lots of reagents, calibrators and controls. The
95th percentile of each n = " data set was calculated, and the highest 95th percentile value from among the three n = data sets was determined. The sponsor determined the LOB to be (b) (4) (4) (4) (4) (4) (4) 1) 1) (4) 1) 1) (4) 1) 1) 1 (4) 1) 1 (4) 1) 1 (4) 1 1) 1 (4) 1 1) 1 (4) 1 1) 1 (4) 1 1) 1 (4) 1 1) 1 (4) 1 1 1 (4) 1 1 1 (4) 1 1 1 (

The limit of detection (LOD) was determined using the variant (non-parametric analysis) "classical approach" with a beta of (b) (4) individual normal human serum samples spiked with clozapine at 15, 20, 25, 30, and 35 ng/mL. Each of the serum samples were tested in duplicate on three different days on (0K4) different analyzers for n=""", using " lots of reagents, calibrators and controls. The LOD was determined using the highest median value observed at the lowest spiking level for which each of the three n = "" data sets associated with that spiking level had(b) (4) of results below the LOB. The sponsor determined the LOD to be 39 ng/mL.

The limit of quantitation (LOO) was determined by 4 individual normal human serum samples spiked with clozapine at " targeted concentrations: 40, 50, 60, 70 and 8 ng/mL. Each of the serum sample was tested n = "on(b) (4) different days on analyzer, using lots of reagents/calibrators/controls, for n=000 (results combined across all lots). The LOQ was defined as the specimen concentration where the total allowable error was ≤ 35% when calculated according to the Westgard model. The sponsor determined the LOO to be 68 ng/mL.

  • e. Analytical specificity:

Potentially Cross-Reacting Substances

Cross-reactivity was evaluated by adding each potentially cross-reacting clozapine metabolite into a human serum pool prepared from multiple individual human serum samples. The sponsor tested the three metabolites in the presence of clozapine at 350 ng/mL. Testing was conducted using one lot of reagents, one lot of calibrators, and two analyzers. Samples were tested in (0) (4) replicates. The formula used for calculating percent cross reactivity was:

(b) (4)

The results are summarized below. The sponsor did not observe cross-reactivity

{5}------------------------------------------------

exceeding ±10%:

CompoundTested at (ng/mL)% Cross-reactivity
Clozapine-N-oxide2503%
8-Hydroxy-8-deschloro-clozapine1009%
Norclozapine2,7001%

Endogenous and Exogenous Interfering Substances:

Interference studies were conducted based on the CLSI guideline EP07-A2. Potentially interfering substances (test sample) or solvent (control sample) were spiked into human serum sample pools with clozapine at 350 and 600 ng/mL. Each sample was tested at a minimum of 5 runs using one lot of reagent on two analyzers. The percentage difference between the measurement of clozapine in the test sample and the control sample was calculated. The sponsor considered a percent difference between test samples and control samples of > ±10% to be clinically significant interference. None of the substances in the following tables (endogenous or exogenous substances) were found to lead to clinically significant interference (> ±10%) for the device.

Interfering SubstanceConcentration Tested
Bilirubin18.18 mg/dL
Hemolysate1,050 mg/dL
Human IgG (HIgG)12.5 g/dL
Human SerumAlbumin (HSA)10.9 g/dL
Rheumatoid Factor (RF)508 IU/mL
Triglycerides2,585.87 mg/dL

Endogenous Substances

Exogenous Interferences

CompoundTested at(ng/mL)CompoundTested at(ng/mL)
Acetaminophen200,000Acetazolamide60,000
Acetylsalicylic acid500,000Albuterol1,000
Alendronate sodium1,000Alpha - tocopherol130,000
Alprazolam2,000AmantadineHydrochloride10,000
CompoundTested at(ng/mL)CompoundTested at(ng/mL)
Amikacin sulfate144,000Amiloride HCldihydrate500
Amisulpride1,200Amitriptyline1,000
Amlodipine besylate100S (+)-amphetamine1,000
Amoxapine2,900Amoxicillin80,000
Aripiprazole1,400L-ascorbic acid60,000
Asenapine500Atomoxetine7,900
Atorvastatin calcium800Baclofen3,000
Benztropine600Betamethasone400
Biotin3,600Biperiden300
Blonanserin100Brexpiprazole1,000
Bromperidol100Budesonide50
Bupropion3,000Buspirone200
Caffeine108,000Calcium carbonate315,000
Cannabidiol100Cannabinol100
Carbamazepine45,000Cariprazine50
L-Carnosine100,000Cefalexin200,000
Celecoxib8,800Cetirizinedihydrochloride4,400
8-chlorotheophylline3,000ChlorpromazineHCl3,300
Cimetidine30,000Ciprofloxacin12,000
Citalopram HBr5,500Clindamycin51,000
Clonazepam300Clotiapine500
Clotrimazole50Codeine2,000
Cortisol300(-)-Cotinine2,000
Cyclosporin A9,000Desloratadine600
Desvenlafaxine800Dextro-methorphan1,000
Diazepam30,000DiphenhydramineHCl6,000
Divalproex Sodium400,000Docosahexaenoicacid ethyl ester150,000
Donepezil50,000Doxycycline HCl35,000
Droperidol200D-Serine100,000
Duloxetine200Erythromycin138,000
Escitalopram200Estradiol10
CompoundTested at (ng/mL)CompoundTested at (ng/mL)
Eszopiclone300Ethanol10,000,000
Famotidine2,500Fenofibrate50,000
Fentanyl600Fluoxetine HCl4,000
Fluticasone propionate50Fluvoxamine2,000
Folic acid15Furosemide60,000
Galantamine200Gentamicin sulfate30,000
Glyburide2,000Haloperidol1,000
Heparin sodium salt50 U/mLHydrochlorothiazide6,000
Hyoscine (Scopolamine HBr)100Hyperforin (St. John's Wort)200
Hypericin (St. John's Wort)100Ibuprofen500,000
Iloperidone100Imipramine700
Indinavir sulfate400Lactulose10,000
Lamivudine10,500Lamotrigine42,000
Lansoprazole9,400Levonorgestrel100
Lisinopril dihydrate350Lithium carbonate250,000
Lorazepam1,000Lovastatin500
Loxapine300Lurasidone400
Meclizine dihydrochloride500Metformin40,000
Methotrimeprazine600Methylphenidate HCl350
Metoclopramide HCl500Metoprolol tartrate5,000
Metronidazole123,000Midazolam3,800
Milnacipran10,000Mirtazapine900
Mometasone furoate50Morphine7,800
Naltrexone200Naproxen sodium500,000
Nateglinide30,000Nefazodone HCl6,000
Nicotine1,000Nicotinic acid27,900
Nordiazepam5,000Nortriptyline1,200
Olanzapine400Omeprazole8,400
Oxazepam5,000Oxcarbazepine105,000
Oxycodone500Paliperidone60
Pantothenic acid1,800Paroxetine1,200
Penicillin V42,000Perazine1,400
CompoundTested at (ng/mL)CompoundTested at (ng/mL)
Perlapine150Perphenazine100
Phenobarbital690,000Phentermine500
Phenytoin60,000Pimozide100
Pipamperonedihydrochloride1,200Potassium EDTA1,000
Pravastatin sodium300Prednisolone3,000
Pregabalin22,500Procyclidine1,900
Promethazine1,200R,R-(-)-pseudoephedrine10,000
S,S-(+)-pseudoephedrine10,000Pyridoxine HCl100
Quetiapine2,800Quinidine15,000
Raloxifene50Ranitidine10,500
Retinol4,000Riboflavin200
Rifampicin65,000Risperidone200
Rosuvastatin calcium200Salicylic acid500,000
Sarcosine1,500Sertindole300
Sertralinehydrochloride1,000Simvastatin1,700
Sodium benzoate400,000Sodium fluoride900
Spironolactone600Sulfamethoxazole400,000
Sulpiride50,000Temazepam5,000
Terbinafine9,000Theophylline60,000
Thiamine HCl500Topiramate75,000
Trazodone HCl14,700Triamcinoloneacetonide300
Triamterene9,000Triazolam40
Valproic acid500,000Vancomycin HCl120,000
Varenicline50Venlafaxine HCl700
Vitamin B1250Vitamin D2200
Vitamin K150Warfarin75,000
Ziprasidone600Zolpidemhemitartrate5,000
Zonisamide120,000Zopiclone200
Zuclopenthixol300

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

  • f. Assay cut-off:
    Not applicable

    1. Comparison studies:
    • a. Method comparison study:

A method comparison study was conducted using the MyCare Psychiatry Clozapine Assay Kit on the Beckman Coulter AU480 by testing clinical samples that were collected at six different U.S. clinical sites from patients receiving clozapine therapy. Results from the MyCare Psychiatry Clozapine Assay Kit were compared to results from a validated clozapine LC-MS/MS method. Deming regression analysis was performed with (b)(4) samples, resulting in the following line equation (with 99% confidence intervals shown) from a representative lot:

y = 1.037 (0.939 to 1.135) x - 27.8 (-78.3 to +22.7); R = 0.9269

A summary of the device accuracy (compared to the LC-MS/MS comparator method) at different concentration ranges throughout the claimed measuring interval is shown below.

Difference Range Between the Serum Clozapine Levelby LC/MS-MS and the MyCare Psychiatry ClozapineAssay
Concentration ofClozapine(ng/mL)Within10%(n/N)Within15%(n/N)Within20%(n/N)Within30%(n/N)Within40%(n/N)
68-35030%(15/50)42%(21/50)60%(30/50)78%(39/50)84%(42/50)
350-100034%(24/71)56%(40/71)75%(53/71)93%(66/71)96%(68/71)
1000-150050% (1/2)50% (1/2)100% (2/2)N/AN/A
  • b. Matrix comparison:
    Not applicable

    1. Clinical studies:
    • a. Clinical Sensitivity:

Not applicable

{10}------------------------------------------------

  • b. Clinical specificity:
    Not applicable

  • c. Other clinical supportive data (when a. and b. are not applicable):
    The sponsor provided a discussion of the clinical use of therapeutic drug monitoring for clozapine in patients with treatment-resistant schizophrenia, as provided in the U.S. clinical practice guideline, "The American Psychiatric Association Practice Guideline for the Treatment of Patients with Schizophrenia" (2020).

    1. Clinical cut-off:
      Not applicable.
    1. Expected values/Reference range:
      There is no definitive therapeutic range for clozapine, as this range may be different for each patient.

N. Proposed Labeling:

The labeling supports the decision to grant the De Novo request for this device.

O. Identified Risks to Health and Mitigation Measures

Identified Risks to HealthMitigation Measures
Incorrect test resultsCertain design verification and validationactivitiesCertain labeling information
Incorrect interpretation of test resultsCertain design verification and validationactivitiesCertain labeling information

P. Benefit/Risk Determination

Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

Summary of the Assessment of Benefit

There is currently no available FDA cleared or approved device for determining Clozapine

{11}------------------------------------------------

blood concentrations. Toxic side effects from Clozapine include seizures, myocarditis, agranulocytosis, and other severe adverse effects, and Clozapine currently requires a REMS (Risk Evaluation and Mitigation Strategy) to ensure the benefits of the drug outweigh the risk of severe neutropenia. The availability of the test may aid in the management of patients with treatment-resistant schizophrenia under various clinical scenarios as described in the American clinical practice guideline, "The American Psychiatric Association Practice Guideline for the Treatment of Patients with Schizophrenia" (2020) (e.g. "while the dose of clozapine is being titrated, it is useful to obtain blood levels of clozapine").

Summary of the Assessment of Risk

In the setting of poor or incomplete drug response, a falsely low clozapine level may increase the risk of a prescriber falsely concluding that the outcome is due to sub-therapeutic levels which may prompt an unnecessary increase in drug dose and increase the risk of adverse events such as seizures. In the setting of potential toxicity, a falsely low level may delay clozapine dose de-escalation, increasing the risk of persistent adverse events and severe adverse events. In the setting of poor or incomplete drug response, a falsely high level may increase the risk of a prescriber incorrectly assuming that the patient was receiving an adequate dose and continue a suboptimal dose or conclude that the patient was not responding to clozapine and prescribe alternate therapies that may be less effective for the patient, increasing the risk of continued psychological distress from persistent symptoms, poorer social and occupational functioning, hospitalization, or transfer to residential settings. Additionally, if a patient's clozapine level is falsely elevated, a prescriber may not fully investigate the possibility of suboptimal dosing, which could result in a missed opportunity to identify and rectify conditions that could lead to suboptimal dosing or to consider treatment alternatives that are more acceptable to the patient. Falsely elevated levels may lead to a prescriber deciding to decrease the dose or discontinue medication, even if the patient were tolerating the medication and had experienced benefit, which could increase the risk of relapse, functional impairment, or transition to higher levels of care.

Summary of the Assessment of Benefit-Risk

General controls are insufficient to mitigate the risks of the device. However, the probable clinical benefits outweigh the probable risks for the assay, considering the mitigation of the risks provided for in the special controls. Device design verification and validation, including precision, method comparison, interference, and cross-reactivity studies will help ensure that the device functions as intended and mitigate the risk of falsely high test results. A limitation statement conveying that results from the assay alone should not be used in making treatment decisions will be included in the labeling, as a mitigation against the risk of incorrect interpretation of results. Overall, the probable benefits outweigh the probable risks of incorrect test results or incorrect interpretation of test results for the proposed indications for use, in light of the special controls and general controls.

{12}------------------------------------------------

Q. Conclusion

The De Novo request for the MyCare device is granted and the device is classified under the following and subject to the special controls identified in the letter granting the De Novo request:

Product Code: QKT Device Type: Clozapine test system Class: II (special controls) Regulation: 21 CFR 862.3245

§ 862.3245 Clozapine test system.

(a)
Identification. A clozapine test system is a device intended to measure clozapine in human specimens. Measurements obtained by this device are used in monitoring levels of clozapine to ensure appropriate therapy in patients with treatment-resistant schizophrenia.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Design verification and validation must include the following:
(i) Precision study data that demonstrates precision that is clinically appropriate, as determined by FDA, for the clozapine test system. Precision studies must include a minimum of three samples containing different concentrations of clozapine including near medical decision points and throughout the expected therapeutic range of clozapine. Samples near the medical decision points must be clinical specimens collected from patients taking clozapine;
(ii) Method comparison data that demonstrates accuracy that is clinically acceptable, as determined by FDA, for the clozapine test system;
(iii) Data from studies that demonstrate that the device is free from clinically significant interference, as determined by FDA, from commonly co-administered medications that are used in patients with treatment-resistant schizophrenia; and
(iv) Data from studies that demonstrate that the device is free from clinically significant cross-reactivity, as determined by FDA, from major circulating metabolites found in the intended use population.
(2) The labeling required under § 809.10 of this chapter must include a limiting statement conveying that the assay should only be used in conjunction with information available from clinical evaluations and other diagnostic procedures and that results from the assay alone should not be used in making treatment decisions.