K Number
DEN190026

Validate with FDA (Live)

Date Cleared
2020-05-01

(352 days)

Product Code
Regulation Number
886.5305
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The iTEAR100 Neurostimulator is an electromechanical nerve stimulator device, indicated for temporary use (up to 30 days) to increase acute tear production during vibratory stimulation of the external nasal nerve in adults, under prescription of an eye care provider.

Device Description

iTEAR100 Neurostimulator is a hand-held, portable electromechanical actuator intended to acutely increase tear production through vibratory stimulation of the external nasal nerve in patients. The device is battery-operated with a single vibratory tip. As the effective tip is pushed against the tissue, the beam deflects inward until it is deflected fully into the device. The device is comprised of a direct current vibration motor, vibrating beam(cantilever), a control circuit with micro SD card, tactile switch.

The motor is attached to a cantilever beam made from acrylonitrile butadiene styrene (ABS), a control circuit to charge the battery and deliver power to the motor, and a housing also made from ABS. The motor and cantilever beam move together to produce a linear vibration at the tip of the cantilever (oscillating tip). A lithium-ion rechargeable battery is supplied inside the device, and is not removable or replaceable. A micro USB port on the device allows for charging. The device is locked by firmware from use after a predetermined number of stimulation days has been triggered.

AI/ML Overview

This document describes the iTEAR100 Neurostimulator, an electromechanical tear stimulator. While the document contains details on clinical studies and safety, it does not provide acceptance criteria in the typical format of a medical device performance study, nor does it present device performance against specific, pre-defined quantitative acceptance criteria like those often seen for AI/ML devices or diagnostic tools.

Instead, the document focuses on demonstrating the device's effectiveness (increase in tear production) and safety through two clinical studies. The "acceptance criteria" for this device appear to be primarily qualitative and related to its intended use and safety profile rather than specific performance metrics (e.g., sensitivity, specificity, or accuracy for a diagnostic device).

Therefore, I will interpret "acceptance criteria" based on the established "Special Controls" and the "Benefit-Risk Conclusion" outlined in the document, as these represent what the FDA has deemed necessary for market clearance. The "reported device performance" will be derived from the clinical study results.

Here's an analysis based on the provided text:


Acceptance Criteria and Device Performance for iTEAR100 Neurostimulator

1. Table of Acceptance Criteria and Reported Device Performance

As this device is an electromechanical stimulator and not a diagnostic or AI/ML-driven device with traditional performance metrics, the acceptance criteria are interpreted from the "Special Controls" and "Benefit-Risk Determination" sections of the De Novo request. The reported device performance is drawn directly from the clinical study summaries.

Acceptance Criteria (Derived from Special Controls & Benefit-Risk)Reported Device Performance (from Clinical Studies)
Clinical Performance: 1. Evaluate tear production under anticipated conditions of use. 2. Evaluate all adverse events (including tissue damage, pain, headache, discomfort). 3. Demonstrate probable benefits outweigh probable risks.1. Tear Production (CLP-007 Study - Sham-Controlled): - Statistically significant increase in Schirmer's test scores in the active group (mean within-subject change pre- to post-stimulation = 21.7mm) compared to sham (mean within-subject change pre- to post-stimulation = -0.5mm). - Between-group difference of 22.2mm (SD 2.61) favoring the active device. Tear Production (CLP-002 Study - Longitudinal): - Stimulated Schirmer's test score was greater than unstimulated at each visit during 30-day follow-up. - Trend of decreased effectiveness (tear production) over time was observed but mechanism not identified. 2. Adverse Events (Safety): - CLP-002 (180-day follow-up): - 2 events definitely related (headache, sneezing, ticking sensation; intermittent nose soreness). - 7 events possibly related (Table 5). - Most common related events: headache and dizziness. - 7/9 related AEs were mild, 1 moderate, 1 severe. - One serious unanticipated AE (continuous nausea, dizziness, lightheaded, headache for 30 days) possibly related, led to subject discontinuation. - 4 subjects experienced VA loss (attributed to disease fluctuations/lack of correction/dry eye, resolved). - CLP-007 (Single visit, sham-controlled): - No device-related adverse events or serious adverse events reported. - One unrelated near-syncopal episode. - No significant changes in visual acuity, corneal edema, facial neurological exams, or hemodynamic parameters. 3. Benefit-Risk Conclusion: - Probable benefits (statistically significant acute tear production increase) outweigh probable risks (few, mostly self-limited non-serious AEs; one serious AE resolved without treatment) through clinical data.
Non-Clinical Performance: 1. Demonstrate device performs as intended. 2. Assess mechanical output specifications (vibration amplitude/frequency, pressure/force, acoustic levels). 3. Validate mechanical safety safeguards. 4. Conduct use life testing.1-4. Bench Testing & Biocompatibility (refer to "SUMMARY OF NONCLINICAL/BENCH STUDIES"): - Acoustic Levels: Passed (criteria (b)(4)). - Safeguards related to pressure/force: Passed. - Vibrational amplitude/frequency: Passed (criteria (b)(4)). - Thermal performance: Passed. - Use Life: Passed ("no change in mechanical parameters or physical characteristics" after X hours simulating (b)(4) bilateral treatments). - Biocompatibility: Acceptable performance for cytotoxicity, skin irritation, skin sensitization per ISO 10993-5 and ISO 10993-10. - EMC/Electrical Safety: Passed conforming to ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62133.
Software Performance: 1. Software verification, validation, and hazard analysis. 2. Address potential hazards with satisfactory results. 3. "Moderate" level of concern addressed.Software Documentation & Testing: - Submission contained all elements per "Moderate" level of concern FDA guidance. - Adequate documentation (specifications, architecture, development environment, traceability, revision history, unresolved anomalies, cybersecurity). - Hazard analysis performed to characterize software risks (malfunction, measurement errors, cybersecurity). - V&V testing performed with satisfactory results.
Biocompatibility: 1. All patient-contacting components demonstrated to be biocompatible.Biocompatibility Evaluation: - Tested per ISO 10993-5:2009 (MEM elution cytotoxicity) and ISO 10993-10:2010 (intracutaneous skin irritation, guinea pig maximization skin sensitization). - Results demonstrated acceptable performance.
Electrical Safety, Thermal Safety, EMC: 1. Performance data demonstrating compliance.Electrical, Thermal, EMC Testing: - Conformed to ES 60601-1:2005/(R)2012, IEC 60601-1-2 Edition 4.0 2014-02, IEC 60601-1-6 Edition 3.1 2013-10, IEC 62133:2012.
Labeling: 1. Physician and patient labeling includes specified information.Labeling Adequacy: - Labeling is adequate and meets 21 CFR 801.109 requirements. - Summarizes clinical trial results, probable benefit, identified risks, prescription-only use, patient training, etc.

(Note: Criteria with "(b)(4)" indicate redacted information in the original document, likely specific numerical thresholds that were proprietary.)

2. Sample Sizes Used for the Test Set and Data Provenance

The primary "test sets" for effectiveness and safety were the cohorts in the two clinical studies:

  • CLP-002 Study:
    • Sample Size: 108 subjects enrolled. 101 subjects completed 30 days of follow-up. 80 subjects agreed to continue, 58 reached the 180-day endpoint.
    • Data Provenance: Multi-center (8 US sites). Prospective, open-label, single-arm study.
  • CLP-007 Study:
    • Sample Size: 60 subjects enrolled (28 in Active group, 32 in Sham group). Subject accountability was 100% for assessments.
    • Data Provenance: Multi-center (sites not specified, but context implies US). Prospective, double-masked, randomized, sham-controlled, single-visit clinical trial.

There is no mention of specific "test sets" in the context of an algorithm's performance on a separate dataset, as this is not an AI/ML device.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This device is not a diagnostic tool where "ground truth" is established by expert interpretation of images or other data. The ground truth for effectiveness is objective physiological measurement (Schirmer's test score, a direct measure of tear production). Safety data (adverse events) were reported by subjects and assessed by investigators at the study sites.

Therefore, the concept of "experts establishing ground truth for a test set" with qualifications like "radiologist with 10 years of experience" does not directly apply here. The expertise lies in the clinical investigators (eye care providers) performing and interpreting the Schirmer's test and monitoring for adverse events.

4. Adjudication Method for the Test Set

Given the nature of the device and the objective measurement of tear production via the Schirmer's test, a formal "adjudication method" like "2+1" or "3+1" (common in image interpretation studies) is not described or implied.

  • Quantitative Endpoint (Schirmer's): The Schirmer's test is a standardized measurement.
  • Adverse Events: Adverse events were reported and assessed by the study investigators. There's no mention of an independent, blinded adjudication committee for adverse events, which is sometimes done in larger trials but not explicitly stated here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, What was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

An MRMC study was not applicable as the iTEAR100 Neurostimulator is a therapeutic device, not an AI/ML diagnostic or assistive tool for human readers. The clinical studies focused on the direct physiological effect of the device (tear production) and its safety, not on improving human interpretive performance.

6. If a Standalone (i.e., Algorithm Only Without Human-in-the-loop Performance) was Done

This is not applicable. The iTEAR100 Neurostimulator is a human-operated electromechanical device. Its performance is inherently "standalone" in the sense that it directly causes tear production. There is no underlying algorithm performing a diagnostic task that could be evaluated without human interaction.

7. The Type of Ground Truth Used

The "ground truth" for the primary effectiveness endpoint (tear production) was obtained through direct physiological measurement using the Schirmer's test. This test objectively quantifies tear volume (length of wetting on a filter paper strip).

For safety, the "ground truth" for adverse events involved patient self-reporting, clinical observation by investigators, and medical assessment.

8. The Sample Size for the Training Set

This concept is not applicable to the iTEAR100 Neurostimulator. This is a medical device, not an AI/ML algorithm that requires a "training set" in the computational sense. The device's design and mechanical parameters are established through engineering and bench testing, not through data training. The clinical studies are for validation of safety and effectiveness in humans, not for "training."

9. How the Ground Truth for the Training Set was Established

As mentioned in point 8, the concept of a "training set" for this device does not apply.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR ITEAR100 NEUROSTIMULATOR

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Electromechanical tear stimulator. An electromechanical tear stimulator is a nonimplantable device intended to increase tear production via mechanical stimulation.

NEW REGULATION NUMBER: 21 CFR 886.5305

CLASSIFICATION: Class II

PRODUCT CODE: QKV

BACKGROUND

DEVICE NAME: iTEAR100 Neurostimulator

SUBMISSION NUMBER: DEN190026

DATE DE NOVO RECEIVED: May 15, 2019

Olympic Ophthalmics, Inc. CONTACT: 400 NW Gilman Blvd #1370 Issaquah, WA 98027

INDICATIONS FOR USE

The iTEAR100 Neurostimulator is an electromechanical nerve stimulator device, indicated for temporary use (up to 30 days) to increase acute tear production during vibratory stimulation of the external nasal nerve in adults, under prescription of an eye care provider.

LIMITATIONS

The sale, distribution, and use of the iTEAR100 Neurostimulator is restricted to prescription use in accordance with 21 CFR 801.109.

Patient training is required on the proper use of the device before home use.

The safety and effectiveness of the iTEAR100 Neurostimulator for the treatment of dry eye disease or for the improvement in dry eye symptoms have not been established.

{1}------------------------------------------------

The device increases tear production during neurostimulation, i.e., tearing was assessed only during stimulation.

The clinical study was not designed to evaluate any changes in nerve sensitivity.

Clinical study results demonstrate a trend of decreased effectiveness (tear production) over time. The mechanism for this decrease has not been identified and was not analyzed as part of the study.

PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

iTEAR100 Neurostimulator is a hand-held, portable electromechanical actuator intended to acutely increase tear production through vibratory stimulation of the external nasal nerve in patients. The device is battery-operated with a single vibratory tip. As the effective tip is pushed against the tissue, the beam deflects inward until it is deflected fully into the device. The device is comprised of a direct current vibration motor, vibrating beam(cantilever), a control circuit with micro SD card, tactile switch.

The motor is attached to a cantilever beam made from acrylonitrile butadiene styrene (ABS), a control circuit to charge the battery and deliver power to the motor, and a housing also made from ABS. The motor and cantilever beam move together to produce a linear vibration at the tip of the cantilever (oscillating tip). A lithium-ion rechargeable battery is supplied inside the device, and is not removable or replaceable. A micro USB port on the device allows for charging. The device is locked by firmware from use after a predetermined number of stimulation days has been triggered.

The frequency of operation without any force applied is approximately 270 Hz. The amplitude of movement is approximately 0.5 mm without force applied. In the optimal force range, the cantilever tip is depressed approximately 1 mm and the frequency is approximately 250 Hz. Below is a list of stimulation parameters for the device:

ParameterControl - In airWith Force
Frequency270 Hz250 Hz
Amplitude0.6 mmNot tested
Acceleration (g)505
Tip Retraction (maximum force) (100%)<6NN/A
Acoustic<35 db<35 db

{2}------------------------------------------------

Image /page/2/Picture/0 description: A hand is holding a white, triangular-shaped device with a blue oscillating tip. The device has several black dots on its surface. The text "Oscillating Tip" is written next to the device, with a blue arrow pointing to the tip. The device appears to be in a white box.

Figure 1. iTEAR100 Neurostimulator

The device activates tear production through stimulation of the nasolacrimal reflex. Stimulation mechanically activates external nasal nerve and initiates the nasolacrimal reflex, resulting in tear secretion. To produce the intended effect, the vibratory tip of the device should be applied to the lateral aspect of the nose for several seconds.

Image /page/2/Picture/3 description: The image shows two diagrams of a human face. The diagram on the left shows a side view of a person's face with a blue circle around the nose. The diagram on the right shows a frontal view of a person's face with the nerves labeled. The nerves labeled are the infratrochlear nerve, supratrochlear nerve, external branch of anterior ethmoidal nerve, and infraorbital nerve.

Figure 2 is a schematic showing correct use of the device

SUMMARY OF NONCLINICAL/BENCH STUDIES

BIOCOMPATIBILITY/MATERIALS

All patient contacting portions (all outer surfaces) of the device were assessed for cytotoxicity, skin irritation, and skin sensitization potential. Evaluation of these biocompatibility endpoints for devices that contact intact skin is consistent with recommendations in the June 2016 FDA Guidance, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process." These biocompatibility tests were performed per ISO 10993-5:2009 (MEM elution cytotoxicity) and ISO 10993-10:2010 (intracutaneous skin irritation, guinea pig maximization skin sensitization). For intracutaneous skin irritation and guinea pig maximization testing, both polar and non-polar

{3}------------------------------------------------

test article extracts were evaluated. For the intracutaneous skin irritation testing, observations were recorded to 72 hours after injection. For the guinea pig maximization testing, observations were made out to 72 hours after the challenge phase. Results demonstrated acceptable performance.

SHELF LIFE/STERILITY

The iTEAR100 Neurostimulator is provided non-sterile. Cleaning and maintenance instructions of the stimulator components of the device are included in the labeling.

ELECTROMAGNETIC CAPABILITY (EMC) AND ELECTRICAL SAFETY

Testing for the iTear100 device conformed to the following electromagnetic compatibility, electrical, mechanical, and thermal safety standards:

  • ES 60601-1:2005/(R)2012. A1:2012. C1:2009/(R)2012 and A2:2010/(R)2012 o (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety And Essential Performance (IEC 60601-1:2005, MOD)
  • IEC 60601-1-2 Edition 4.0 2014-02, Medical Electrical Equipment Part 1-2: General o Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements and Tests 5-89
  • . 60601-1-6 Edition 3.1 2013-10 Medical Electrical Equipment - Part 1-6: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Usability
  • . IEC 62133:2012 (Second Edition) Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications

PERFORMANCE TESTING - BENCH

The sponsor has provided the test methods and results on multiple samples of their device for the following verification tests.

This testing is intended to address the risks of physical injury to the local muscles, nerves and skin due to repeated stimulation. Excessive acceleration, force, amplitude and frequency could all cause injury and or abrasions.

TestPurposeMethodAcceptance CriteriaResults
Acoustic LevelsTo assess the acoustic outputSound level meter in acoustic isolation chamber(b)(4)passed
Safeguards related to pressure/force against tissueTo assess force output performance when in operationLoad cell was lowered onto vibrating tip of device until tip depressed in housingpassed

{4}------------------------------------------------

Vibrationalamplitude/ frequencyTo assess the motion ofthe applicator tip ofdeviceUse strobe to determinefrequency and directmeasurement of amplitude(b)(4)passed
ThermalperformanceTo determine operatingtemperature of deviceThermocouples wereapplied to device andtemperature measured withdevice in operationpassed
Use Life *To assess fatigue of thecantileverCantilever deflectionloading measured beforeand after life cycle testpassed
  • The sponsor has performed fatigue testing on their device for " " hours which simulates (b)(4) bilateral treatments. This resulted in no change in mechanical parameters or physical characteristics of the device.

SOFTWARE

All components of the device are controlled/monitored by software, which is responsible for the functionality, user interface, safety checks and performance accuracy. The software is considered to be a moderate level of concern (LOC) because inadvertent software errors could result in injury to the patient.

The following device functions of the iTEAR100 Neurostimulator are controlled by the software:

    1. Software detects the state of the switch and operate the vibration motor.
    1. When in use the software modulates the motor on/off signal creating a stutter in the motor motion. This stuttering serves as a treatment duration indicator for the user.
    1. Logs use data to memory SD Card.
    1. Detect battery level.
    1. Provide USB detect for charging. This disables the unit from being used when connected to USB.
    1. LED control. LED is flashed signaling battery state.
    1. Disable device after 30 Day of use

The submission contained all the elements of software documentation corresponding to the "Moderate" level of concern, as outlined in the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Adequate documentation describing the software/firmware, software specifications, architecture design, software development environment, traceability, revision level history, unresolved anomalies and cybersecurity provide the foundation that the software will operate in a manner as described in the specifications. Hazard analysis was performed to characterize software risks including device malfunction, measurement related errors and cybersecurity. The submission included verification and validation (V&V) testing to address the potential hazards with satisfactory results.

SUMMARY OF CLINICAL INFORMATION

{5}------------------------------------------------

Two clinical studies. CLP-OO2 and CLP-OO7. were completed with the iTEAR100 Neurostimulator™ and are discussed below.

CLP-002 Study

CLP-OO2 was a prospective, open-label, single-arm, multi-center (8 US sites), 2-stage study. The first stage of this pivotal trial included 30 days of follow-up while using the device at home and included all enrolled subjects. After 30 days, subjects were given the option to participate in an additional 150 days of follow-up while using the device (180 days of follow-up in total). The subject was provided a take-home device to perform the study treatment per the instructions for use. It was recommended that subjects use the study device on each side of the nose for 30 seconds on each side twice per day for the 30-day follow-up period. In the extended follow up, subjects were asked to use the device at least once every other day for the 150 days in the extended follow-up period.

Below are the inclusion and exclusion criteria: Inclusion Criteria

    1. Twenty-one (21) years of age or older
    1. Schirmer test with anesthetic of (b)(4)

in at least one eve

    1. Ability to produce tears upon training with (b)(4) change in Schirmer score compared to baseline in at least one eye (the technique for this assessment is provide below)
    1. In the opinion of the investigator, subject is in good general health and free of any condition that could impair study participation or ocular evaluation
    1. Subject is willing and able to give written informed consent and commits to comply with study requirements

Exclusion Criteria

    1. Sjogren's syndrome or other rheumatologic condition
    1. Intraocular surgery within 6 months of Visit 1
    1. Intraocular or periocular injection within 6 months of Visit 1
    1. Used intranasal neurostimulation within two months of Visit 1 or plans to use it during the study
    1. Lid function abnormalities
    1. Any acute infectious or non-infectious ocular condition of the anterior or posterior segments in either eye within 30 days of Visit 1
    1. Diseases/conditions of ocular surface associated with clinically significant scarring/destruction of conjunctiva/cornea
    1. Subject has any of the following conditions:
    • · History of facial nerve palsy
    • · History of neuromuscular disorder
    • · Uncontrolled ocular or systemic disease
    • · Other clinically significant local skin condition (e.g., skin infection) at target treatment site
    1. Participation in any clinical trial with a new active substance or a new device within 30 days of Visit 1 (with the exception of the devices to be used in the study described herein).

{6}------------------------------------------------

Study Visits were performed at Baseline (Day 0). Follow-up Call (day 3). Day 30. Dav 90, Day 180. Study Visits were performed at Baseline (Day 0), Follow-up Call (day 3), Day 14, Day 30, Day 90, Day 180. Clinical assessments included:

  • o Corrected distance visual acuity (Baseline, Day 30, Day 180)
  • Corneal and conjunctival staining (Baseline, Day 14, Day 30, Day 90, Day 180) ●
  • o Schirmer test (Baseline, Day 14, Day 30, Day 90, Day 180)
  • o Neurological exam (Baseline, Day 30, Day 180)
  • o Adverse event recording (all visits)

The primary endpoint was statistically significant improvement from baseline to Day 30 on the Schirmer Index (SI). Briefly, anesthesia was applied. followed by placement of the Schirmer strip in the subject's eye and measurement of the distance a subject's tears traveled along it. After 5 minutes, iTEAR100 stimulation was applied for 30 seconds on each side (with the strip in place), and the was re-measured 5 minutes afterward. The SI is the difference between the 2 measurements (pre- and post- stimulation). Subjects were allowed to maintain medication regimens if the regimens were considered to be stable and unlikely to change over the course of 180 days. Subjects were chosed on their ability at day 0 to respond to neurostimulation and the company recommends a trial in the office prior to writing a prescription. The study enrolled 108 subjects, of which 101 subjects completed the 30 days of follow-up. Eighty (80) subjects agreed to continue in the extended follow-up phase of the study, of which 58 reached the final 180-day endpoint. Demographic data and subject accountability data for the study are provided below:

Table 1- Demographic Characteristics

n
SexF82
M26
Age<306
30-5018
50-7060
>7024
EthnicityHispanic/Latino5
NotHispanic/Latino101
NA2
RaceWhite87
Black2
Other3
NA16
All108

{7}------------------------------------------------

Table 2 - Subject Accountability

Main StudyOptional ExtendedFollow-up:At Day 30, 80subjects chose toenroll in theextended follow-up
BaselineDay 3(call)Day 14Day 30Day 90Day 180
Theoretical1081081081088080
Death000000
Withdrawal(cumulative)004669
Lost toFollow Up(cumulative)0011713
Missing012000
Expected1081081041027471
ActualA10810397885140
ActualB1081071011016758
% Follow UpActualA100%95.4%89.8%81.5%63.8%50.0%
% Follow UpActualB100%99.1%93.5%93.5%83.8%72.5%

Expected is equal to Theoretical minus Deaths minus Withdrawals

Actual is all subjects with data available for primary endpoint and within window

is all subjects with data available for primary endpoint

Follow-up Actual is equal to ActualA divided by Theoref

Follow-up Actual® is equal to Actual® divided by Theoretical

The stimulated Schirmer's test score was greater than the unstimulated test score at each visit during the 30-day follow-up period (Table 3). There was also an increase in the pre-stimulation Schirmer score during the course of the study.

{8}------------------------------------------------

Image /page/8/Figure/0 description: The image is a bar graph that shows pre-stimulation and post-stimulation values at baseline, day 14, and day 30. The graph also shows the difference between pre- and post-stimulation. At baseline, the pre-stimulation value is around 6, and the post-stimulation value is around 28. At day 30, the pre-stimulation value is around 9, and the post-stimulation value is around 19.

Table 3 - Change in tear production (Schirmer score)

Safety information for the entire 180-day follow-up period is provided below.

Table 4 - Ophthalmic Adverse Events

EventDescriptionDaytoAEDuration(days)SeverityImpactofDeviceAction Taken
Inflammation due to Schirmer strip1314MildNo changeDrug Rx
Blurred vision1232*MildNo ChangeNew Contact Lenses
Eye Floater921MildNo ChangeNone
Corneal Medicamentosa9160MildNo changeDecrease artificial tear usage
Corneal Neovascularization18021ModerateNo changeNone
Dry Eye Pain219MildNo changeArtificial Tears
Eyelid irritation due to Schirmer strip01MildNo changeNone
Stye on lower lid OD1MildNo changeDrug Rx
Vitreous Hemorrhage OS6416MildDiscontinuedNone
Severe itching and sensitive eyes papillae OU58ModerateDiscontinuedDrug Rx
Eye infection OU557MildNo changeNone
Eye infection OU748MildNo changeNone
Eye infection OU950MildNo changeNone
Corneal Abrasion due to Schirmer strip23ModerateNo changeDrug Rx
Sjogren's syndrome30OngoingMildNo changeReferred to rheumatologist

*The cause of the blurry vision was poor contact lens fit and the days to resolution represents the time of the phone call inquiry that was made outside the study follow-up period, since the subject could not recall the exact date of the new lens fitting.

{9}------------------------------------------------

There were 2 events that were definitely related to the device: one event described as headache, sneezing, and ticking sensation (onset at 1 day from study enrollment); the other was described as intermittent nose soreness (onset at 85 days from study enrollment). There were 7 events that were possibly related to the device (Table 5).

EventDescriptionDaytoAEDuration(days)SeverityImpactofDeviceAction Taken
Inflammation dueto Schirmer strip1314MildNo changeDrug Rx
Blurred vision1232*MildNo ChangeNewContactLenses
Eye Floater921MildNo ChangeNone
CornealMedicamentosa9160MildNo changeDecreaseartificialtearusage
CornealNeovascularization18021ModerateNo changeNone
Dry Eye Pain219MildNo changeArtificial Tears
Eyelid irritationdue to Schirmerstrip01MildNo changeNone
Stye on lower lidOD1MildNo changeDrug Rx
VitreousHemorrhage OS6416MildDiscontinuedNone
Severe itching andsensitiveeyespapillae OU58ModerateDiscontinuedDrug Rx
Eye infection OU557MildNo changeNone
Eye infection OU748MildNo changeNone
Eye infection OU950MildNo changeNone
Corneal Abrasiondue to Schirmerstrip23ModerateNo changeDrug Rx
Sjogren'ssyndrome30OngoingMildNo changeReferredtorheumatologist

Table 5 - Adverse Events Possibly Related to Device

*The cause of the blurry vision was poor contact lens fit and the days to resolution represents the time of the phone call inquiry that was made outside the study follow-up period, since the subject could not recall the exact date of the new lens fitting.

Headache and dizziness were the most common related events reported in the study. Seven (7) of the 9 related adverse events were considered mild, 1 was considered moderate, and 1 was considered severe. Most related events were transient and resolved within approximately 1 week. After a single device treatment, one subject experienced continuous nausea, dizziness,

{10}------------------------------------------------

lightheaded and headache for 30-day duration. The subject was instructed to seek neurological evaluation, stop device use, and discontinue from the study. This adverse event was considered to be a serious unanticipated adverse event, possibly related to the device. Four subjects experienced a loss of 2 lines or more in visual acuity during treatment. For 3 of the 4 subjects, loss of visual acuity was attributed to fluctuations in the disease, and/or the fact that subjects were asked not to use artificial tears or other treatments when follow-up visits were scheduled. Pinhole correction and manifest refraction was not attempted or used for any of these subjects. The subjects' vision returned to baseline or 20/20 as fast as 2 weeks for 1 subject and within 180 days for the other 2 subjects. The last subject experienced a loss of visual acuity when the 30 day follow-up visit was performed without correction, but demonstrated no change in visual acuity when the exam was performed with correction, at 180 days. Therefore, this subject's change in visual acuity was attributed to a lack of correction at the 30 day visit. All 4 subjects' vision returned to baseline during the follow-up period. The investigators believed that the alterations in vision were due to temporary worsening of dry eye disease, however, causality could not be definitively determined due to limitations in the study design.

CLP-007 Study

CLP-007 was a multi-center, nonsignificant-risk, prospective, double-masked, randomized, sham controlled, single visit clinical trial that enrolled 60 subjects. The sham group received an iTEAR100 Neurostimulator device which looked identical to and made noise similar to a fully functional device but had a tip that did not vibrate. All subjects received cross-over treatment with the active device at the conclusion of the study procedures. Regardless of whether they were using the sham or active device, subjects were instructed to apply the device tip against the skin of the nose at the junction of the nasal cartilage and the nasal bone. The application duration was 30 seconds or less for each side of the nose.

The inclusion and exclusion criteria are provided below:

Inclusion Criteria

    1. 18 years of age or older;
    1. Willing and able to provide an English language written informed consent;
    1. Able to safely use the study device and be free of any condition that, in the opinion of the investigator, could impair study participation.

Exclusion Criteria

    1. Presence of clinically significant nasal/facial skin conditions (e.g., infection, ulceration, wound):
    1. Under arrest or was otherwise in custody;
    1. Presence of any other condition, which in the judgment of the principal investigator would prevent a potential subject from safely completing the study or tolerating device use, such as mental illness, dementia, severe agitation, etc.).

No study-mandated tests were required for enrollment. The schedule of assessments and procedures for the study is shown in the table below.

{11}------------------------------------------------

Table 6 - Clinical Assessments

Schedule
AssessmentScreening/Day0a
Eligibility screening & ICX
OSDI questionnaireX
Demographic informationX
Focused DES historyX
Baseline nasal skin exam & facial neurological examX
Baseline visual acuity (OS)X
Baseline slit lamp examination (SLE) with emphasis on comeal edema (OS)X
Baseline blood pressure (BP), heart rate (HR), pulse oximetryX
Baseline Schirmer's test (OD)X
Baseline meibomian gland expression score (OS)X
Randomization & experimental device training/treatmentbX
Repeat Schirmer's test (OD) with new Schirmer strip during device treatmentb,cX
Repeat BP, HR, pulse oximetry during Schirmer's test (at 2.5 & 5 min)bX
Repeat nasal skin exam & facial neurological examX
Repeat visual acuity (OS)X
Repeat SLE with emphasis on corneal edema (OS)X
Repeat meibomian gland expression score (OS)X
Cross-over treatment with fully active iTEAR device & qualitative assessment ofresponseX
PI usability surveyX
Subject usability surveyX
Record adverse eventsX

a Screening and Day 0 were the same

b Second investigator to perform; initial investigator was out of the room.

· Device treatment begins immediately after placement of Schirmer's test strip

The primary endpoint of this study was the mean within-subject change in the Schirmer's test. score post vs. pre neurostimulation in the active treatment vs. sham groups in the primary analysis population (i.e., subjects with baseline Schirmer's test score ≤10mm). The replacement Schirmer technique was used for this study, under which a Schirmer strip is applied to the subject's eye to assess pre-stimulation tear production, and then a second strip is inserted to assess post-stimulation tear production.

{12}------------------------------------------------

The safety of the device was evaluated by the incidence of device-related adverse events and serious adverse events, as well as changes in the following measures:

  • Best corrected Snellen distance visual acuity (VA); .
  • . Ophthalmic exam with emphasis to determine the presence of corneal edema;
  • . Nasal skin examination and facial neurologic examination;
  • . Hemodynamic parameters (blood pressure, heart rate, pulse oximetry).

Below are the study demographics:

Table 8 - Baseline Demographics

SummaryªRandomized Treatment
Active (N=28)Sham (N=32)All Subjects (N=60)
Sex, n (%)16 (57.1%)24 (75.0%)40 (66.7%)
Female16 (57.1%)24 (75.0%)40 (66.7%)
Male12 (42.9%)8 (25.0%)20 (33.3%)
Age (Years)
Mean (SD)49.7 (15.99)50.0 (16.25)49.9 (16.00)
Median515051
Min, Max24, 7921, 7721, 79
Ethnicity, n (%)
Hispanic or Latino3 (10.7%)7 (21.9%)10 (16.7%)
Not Hispanic or Latino25 (89.3%)25 (78.1%)50 (83.3%)
Race, n (%)
Asian5 (17.9%)7 (21.9%)12 (20.0%)
Black2 (7.1%)1 (3.1%)3 (5.0%)
Caucasian18 (64.3%)21 (65.6%)39 (65.0%)
Other3 (10.7%)3 (9.4%)6 (10.0%)
Subjects using EyeTreatment, n (%)21 (75.0%)25 (78.1%)46 (76.7%)

Each randomized subject underwent all study assessments, thus subject accountability was 100%. In the primary analysis population, there was a statistically significant increase in Schirmer's test scores in the active group (mean within-subject change pre- to post-stimulation = 21.7mm) compared to sham (mean within-subject change pre- to post-stimulation = - 0.5mm) with a between group difference of 22.2mm (SD 2.61) favoring the active device, as shown in in the figure below. The bars included in the figure show the standard deviation for the mean change reported, which indicates the degree of variability for the with-in subject change scores.

{13}------------------------------------------------

Image /page/13/Figure/0 description: The image is a bar graph titled "Mean Change in Schirmer From Baseline". The y-axis is labeled "Schirmer Change Post-stimulation (mm)" and ranges from -10 to 35. There are two bars, one labeled "Active, 21.7" which has a value of approximately 21.7, and the other labeled "Sham, -0.5" which has a value of approximately -0.5.

Figure 9 - Schirmer's Test Score Change in Active Compared to Sham Groups

There were no device-related adverse events or serious adverse events reported in this study. One unrelated adverse event involved a near-syncopal episode that occurred following the administration of topical proparacaine anesthetic to the subject's right eye prior to placement of the baseline Schirmer's test strip. The subject was randomized but had not received treatment with either a sham or an active study device. She was terminated from any further study participation and was monitored by the investigator at the site until her symptoms had resolved. Additionally, there were no observed significant changes in post vs. pre-test visual acuity, corneal edema, facial neurologic examinations, or hemodynamic parameters.

Pediatric Extrapolation

In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.

LABELING

The professional and patient labeling are adequate and meet the requirements of 21 CFR 801.109. The labels summarize the clinical trial results that characterized the probable benefit and the identified risks of the device. including tissue damage, pain. headache, and discomfort. Both guides contain requirements for use by prescription only and proper patient training, Indications for Use, contraindications, device description, technical parameters, warnings, precautions, potential complications, instructions for use (including an explanation of all userinterface components and information regarding proper device placement), recommended stimulation schedule, instructions for device maintenance/cleaning, summary of clinical trials, information related to electromagnetic compatibility, language to direct end users to contact the device manufacturer and MedWatch if they experience any adverse events with this device, expected service life, disposal & replacement, environmental operating conditions, electrical specifications, and symbols & markings.

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of an electromechanical tear stimulator and the measures necessary to mitigate these risks.

{14}------------------------------------------------

Identified RisksMitigation Measures
Tissue damage due tooverstimulation/understimulationor mechanical injury, devicebreakageClinical performance testingNon-clinical performance testingSoftware verification, validation, and hazard analysisLabeling
Adverse tissue reactionBiocompatibility evaluationLabeling
Electrical shock or burnElectrical, thermal, and mechanical safety testingSoftware verification, validation, and hazard analysisLabeling
Interference with other devicesElectromagnetic compatibility (EMC) testingSoftware verification, validation, and hazard analysisLabeling
Pain, headache, or discomfortClinical performance testingNon-clinical performance testing
Insufficient tear productionClinical performance testing

Table 9 - Identified Risks to Health and Mitigation Measures

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the electromechanical tear stimulator is subject to the following special controls:

    1. Clinical performance testing under anticipated conditions of use must evaluate tear production and all adverse events, including tissue damage, pain, headache, and discomfort.
    1. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be conducted:
    • a. An assessment of mechanical output specifications including vibration amplitude and frequency, pressure and force, and acoustic (noise level) properties;
    • b. Mechanical safety testing to validate safeguards related to the pressure aspects of the device; and
    • c. Use life testing.
    1. Performance data must demonstrate the electrical safety, thermal safety, and electromagnetic compatibility (EMC) of all electrical components of the device.
    1. All patient-contacting components of the device must be demonstrated to be biocompatible.
    1. Software verification, validation and hazard analysis must be performed.
    1. Physician and patient labeling must include:
    • a. A detailed summary of the device's technical parameters;

{15}------------------------------------------------

  • Instructions for use, including an explanation of all user-interface components and b. information regarding proper device placement;
  • Information related to electromagnetic compatibility classification: C.
  • d. Instructions on how to clean and maintain the device;
  • A summary of the clinical performance testing conducted with the device; e.
  • Language to direct end users to contact the device manufacturer and MedWatch if f. they experience any adverse events with this device; and
  • g. Information on how the device operates and the typical sensations experienced during treatment.

BENEFIT-RISK DETERMINATION

The risks of the device are based on non-clinical laboratory data, as well as data collected in clinical trials described above. The device exhibited an acceptable safety profile for up to 30 day use in the clinical studies which were conducted. One device-related serious adverse event was observed; this event resolved without treatment. Non-serious device-related adverse events were few in number and all were self-limited.

The probable benefits of the device are also based on nonclinical laboratory data, as well as data collected in clinical trials. Compared to a sham control, the device was shown to provide a statistically significant increase in acute tear production during a single stimulation. Results from an additional clinical study suggest that a temporary increase in acute tear production may be seen in response to repeated device stimulation up to 30 days.

Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

Benefit/Risk Conclusion

In conclusion, given the available information above, the probable benefits outweigh the probable risks for the iTEAR100 Neurostimulator. The device provides benefits, and the risks can be mitigated by use of general controls and the identified special controls.

CONCLUSION

The De Novo request for the iTEAR100 Neurostimulator is granted and the device is classified as follows:

Product Code: OKV Device Type: Electromechanical tear stimulator Class: II Regulation Number: 21 CFR 886.5305

§ 886.5305 Electromechanical tear stimulator.

(a)
Identification. An electromechanical tear stimulator is a non-implantable device intended to increase tear production via mechanical stimulation.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing under anticipated conditions of use must evaluate tear production and all adverse events, including tissue damage, pain, headache, and discomfort.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following must be conducted:
(i) An assessment of mechanical output specifications, including vibration amplitude and frequency, pressure and force, and acoustic (noise level) properties;
(ii) Mechanical safety testing to validate safeguards related to the pressure aspects of the device; and
(iii) Use life testing.
(3) Performance data must demonstrate the electrical safety, thermal safety, and electromagnetic compatibility (EMC) of all electrical components of the device.
(4) All patient-contacting components of the device must be demonstrated to be biocompatible.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Physician and patient labeling must include:
(i) A detailed summary of the device's technical parameters;
(ii) Instructions for use, including an explanation of all user-interface components and information regarding proper device placement;
(iii) Information related to electromagnetic compatibility classification;
(iv) Instructions on how to clean and maintain the device;
(v) A summary of the clinical performance testing conducted with the device;
(vi) Language to direct end users to contact the device manufacturer and MedWatch if they experience any adverse events with this device; and
(vii) Information on how the device operates and the typical sensations experienced during treatment.