K Number
DEN170056

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-11-02

(403 days)

Product Code
Regulation Number
878.4550
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The AiBiomed Parathyroid Detection System (PTeye) is an adjunctive tool intended to aid in the identification of parathyroid tissue by confirming parathyroid tissue already visually located by the surgeon.

Device Description

The Parathyroid Detection (PTeye) System aids surgeons in differentiating parathyroid tissue during common surgical procedures. The handheld probe assembly includes one glass fiber-optic element that emits non-ionizing radiation at 785 nm in the near IR range (NIR) and one fiber optic detector element that collects and transmits the fluorescence emitted by the tissue to a photo detector. The Parathyroid Detection (PTeye) System consists of the following components: 1. A console that includes: An LED display that indicates if the laser is on. A display for visual feedback. A speaker for auditory feedback. 2. A handheld fiber-optic probe assembly that interfaces into the console unit using two unique connectors. One connector plugs into the laser output and the other plugs into the photo detector input (fluorescent signal). 3. A foot pedal attached by a cable to the rear of the unit, used to control power to the laser and initiate data collection. 4. An external power supply and power cord that plugs into the power supply. Tissue detection is based on the ratio of the fluorescent response of parathyroid to thyroid tissue; with the assumption that the fluorescence of thyroid tissue is much lower than parathyroid. During surgery, five thyroid data points are collected by touching thyroid tissue with the probe assembly. The system calculates a baseline median for the thyroid tissue based on those points. The baseline value establishes a reference point for distinguishing parathyroid tissue, which produces a higher level of fluorescence. Once the baseline thyroid value has been calculated, a new operational screen will display to support continuous parathyroid search mode. To operate in this mode, the foot switch must be pressed in order to activate the laser. When laser emission is taking place, the LASER ON LED at the front of the system will illuminate. Responses indicating parathyroid tissue are communicated to the user through a bar graph, a detection percentage, a detection ratio, and audio feedback.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Parathyroid Detection (Model PTeye) System, based on the provided text:


Acceptance Criteria and Device Performance for PTeye System

Acceptance CriteriaReported Device Performance
Primary Effectiveness Endpoint: Ability of the PTeye to accurately identify parathyroid glands (PG detection rate).Overall Performance (combined histology and expert surgeon corroboration, excluding r-SHPT patients): - Accuracy: 96.3% - Kappa: 0.90 - Sensitivity: 93.6% - Specificity: 97.1% - Positive Predictive Value (PPV): 91.5% - Negative Predictive Value (NPV): 97.9% - False Positive Rate: 2.9% - False Negative Rate: 6.4%
Safety Endpoint 1: Safe use as determined by a lack of (serious) adverse events.No adverse events, serious adverse events (SAEs), or unanticipated adverse device effects (UADEs) related to the procedure or during the study period were reported for the 81 patients whose data were used for effectiveness analysis. No clinical issues or device complaints were reported.
Safety Endpoint 2: The addition of no more than 5 minutes to the total procedure time during normal use of the device.Individual measurements once the PTeye device is set up take approximately 2 seconds. The total additional procedure time for baseline and subsequent measurements (5 thyroid, 4 extra thyroid, 8 PG, 2 fat, 2 trachea, 2 muscle measurements per patient) was approximately one minute, which is under the 5-minute pre-defined cut-off.
In vivo testing: Characterize the ability of the device to detect autofluorescent signals from tissues or structures consistent with the indications for use.- Normalized NIRAF intensity from PGs was significantly higher than non-PG tissues (p < 0.01). - Normalized NIRAF intensity for PG was about 5.4 times higher than thyroid. - Muscle, fat, and trachea showed little to no NIRAF intensity. - No significant difference in normalized NIRAF intensity between healthy and diseased thyroid glands (p=0.96). - Diseased PGs exhibited lower normalized NIRAF intensity than healthy PGs (p=0.00012).
Biocompatibility: Patient-contacting components must be demonstrated to be biocompatible.Passed biocompatibility requirements for cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity. Limulus Amebocyte Lysate (LAL) evaluation passed (less than 20 EU/device).
Electromagnetic compatibility and electrical, mechanical and thermal safety: Performance testing must demonstrate this.Passed IEC 60601-1 (general requirements for basic safety), IEC 60601-1-2 (EMC), and IEC 60601-1-6 (Usability).
Software verification, validation, and hazard analysis: Must be performed.Software documentation was provided in sufficient detail to provide reasonable assurance the software will operate as described. The software is considered a moderate level of concern. Cybersecurity information was also provided.
Sterility of patient-contacting components: Performance testing must demonstrate this.Sterilization method (EtO) validated per ISO 11135:2014, with a Sterility Assurance Level (SAL) of 10-6. Sterilant residuals were under acceptable limits.
Shelf-life of device components provided sterile: Performance testing must support continued sterility and package integrity over the labeled shelf life.Shelf-life for the optical probe was evaluated after accelerated aging equivalent to one year. The cartridge met acceptance criteria for visual inspection, peel test, and bubble leak testing.
Laser and light safety for eye, tissue and skin: Performance testing must demonstrate this.Passed IEC 60825-1: Safety of laser products Part 1.
Labeling requirements: Instructions for use, detection performance characteristics, and shelf life for sterile components.User manual and instructions for use includes device technical parameters, instructions, relevant clinical study findings, detection performance characteristics, shelf life for sterile components, disposal/cleaning, and warnings for secondary hyperparathyroidism/parathyroid cysts, and not using as a locator. Precaution for rare disease states (tertiary hyperparathyroidism, MEN2A etc.) due to small sample size.
Effect of surgical and ambient light on device performance.Passed (Bench study). The PTeye achieved its performance accuracy without needing to switch off ambient OR lights.
Influence of blood on NIRAF intensity of PG and thyroid.Ex vivo: Normalized NIRAF of PG specimens with blood was lower than without blood but not statistically significant (p=0.53). PG NIRAF was significantly higher than thyroid with and without blood. In vivo: No significant difference in PG NIRAF with and without blood (p=0.95). PG NIRAF was significantly higher than thyroid, regardless of blood (p=0.005).
Effect of probe-to-tissue contact pressure on tissue NIRAF measurements.No notable difference in tissue NIRAF measurements with mild, moderate, or high probe pressure for both PG and thyroid specimens. PG NIRAF levels remained consistently and significantly higher than thyroid, regardless of contact pressure (p=2.2×10^-10).

Study Details:

  1. Sample sizes used for the test set and the data provenance:

    • Test Set (Effectiveness Data): 81 patients (40 from Site A (Vanderbilt), 41 from Site B (Ohio State University Medical Center)).
    • Additional exclusion: 2 patients with secondary hyperparathyroidism (r-SHPT) were excluded from the final effectiveness analysis due to known limitations of the device in these cases, resulting in data from 79 patients for core effectiveness analysis.
    • Total tissues measured: 181 PGs and 546 non-PG tissues (194 thyroid, 116 fat, 119 neck muscle, and 117 trachea).
    • Data Provenance: Prospective, multi-center clinical study conducted in the USA (Vanderbilt University Medical Center, Nashville, TN; Ohio State University Medical Center).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Experts: 4 surgeons (high-volume endocrine surgeons).
    • Qualifications: Perform >25 thyroid surgeries per year and >15 parathyroid surgeries per year.
      • Surgeon #1: 16 years of experience as a practicing endocrine surgeon.
      • Surgeon #2: 4 years of experience as a practicing endocrine surgeon.
      • Surgeon #3: 16 years of experience as a practicing endocrine surgeon.
      • Surgeon #4: 4 years of experience as a practicing endocrine surgeon.
  3. Adjudication method for the test set:

    • Ground truth was established by a combination of:
      • Histological examination: For excised tissues (frozen section or Hematoxylin-Eosin stained). 68 PGs (37.6% of all PG measurements) were confirmed with histology.
      • Expert surgeon's visual assessment: For in-situ tissues not excised, the surgeon's visual identification was used, but only those with "high" or "moderate" confidence were included. "Low confidence" assessments were excluded unless histological validation was obtained. The study was single-blinded, meaning surgeons were blinded to the PTeye device output during their initial visual assessment.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Yes, a comparative effectiveness study was done comparing the PTeye's performance to human visual assessment. However, it was not an MRMC study designed to show how much human readers improve with AI assistance (human-in-the-loop). Instead, it compared the PTeye's standalone performance to the surgeons' standalone visual assessment.
    • Comparison of PTeye vs. Surgeon's Visual Determination (validated by histology for excised PGs, excluding r-SHPT patients):
      • Overall PG detection rate (validated by histology):
        • Surgeons: 93.3% (56/60)
        • PTeye: 93.3% (56/60)
      • Overall non-PG tissue identification (excised, assumed as PG, validated by histology):
        • Surgeons: 0.0% (0/5) correctly identified as non-PG.
        • PTeye: 80.0% (4/5) correctly identified as non-PG.
    • (Effect size of human improvement with AI vs. without AI assistance): This specific "human-in-the-loop" effect size (synergy or improvement) was not reported or evaluated in this study, as the surgeons were blinded to the PTeye's output for their initial assessment. The study focused on the PTeye's standalone accuracy compared to visual assessment.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Yes, the primary effectiveness endpoint measures the standalone performance of the PTeye system in identifying parathyroid glands. The overall accuracy, sensitivity, specificity, PPV, and NPV figures (e.g., 96.3% accuracy) represent the PTeye's performance independently of the surgeon's real-time interpretation of its output (as the surgeon was blinded). The study also directly compared this standalone PTeye performance against the surgeons' standalone visual assessment.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • A combined ground truth was used:
      • Histopathology/Pathology: For excised tissues (frozen section or Hematoxylin-Eosin stained tissue section analysis by pathologists). This was considered the "gold standard" when available.
      • Expert Surgeon Consensus/Opinion: For in-situ tissues not excised, the visual assessment of expert surgeons, expressed with "high" or "moderate" confidence, served as the ground truth.
  7. The sample size for the training set:

    • The document does not explicitly state the sample size for a training set. The study described is a clinical study to support the safety and effectiveness of the device, likely using a pre-trained model. If the device's internal model was developed using a training set, that information is not provided in this regulatory document.
  8. How the ground truth for the training set was established:

    • As the training set sample size is not explicitly provided, the method for establishing its ground truth is also not described in this document.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR PARATHYROID DETECTION (MODEL PTEYE) SYSTEM

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Autofluorescence detection device for general surgery and dermatological use. An autofluorescence detection device for general surgery and dermatological use is an adjunct tool that uses autofluorescence to detect tissues or structures. This device is not intended to provide a diagnosis.

NEW REGULATION NUMBER: 21 CFR 878.4550

CLASSIFICATION: Class II

PRODUCT CODE: ODF

BACKGROUND

DEVICE NAME: Parathyroid Detection (Model PTeye) System

SUBMISSION NUMBER: DEN170056

DATE OF DE NOVO: September 27, 2017

CONTACT: AiBiomed. Corp. 107 West Gutierrez Street Santa Barbara, CA 93101

INDICATIONS FOR USE

The AiBiomed Parathyroid Detection System (PTeye) is an adjunctive tool intended to aid in the identification of parathyroid tissue by confirming parathyroid tissue already visually located by the surgeon.

LIMITATIONS

The sale, distribution, and use of the Biomed Parathyroid Detection System (PTeye) are restricted to prescription use in accordance with 21 CFR 801.109.

The AiBiomed Parathyroid Detection System (Model PTeye) has no contraindications.

Due to limitations in parathyroid detection of autofluorescence by the PTeye System in certain disease states, the device is not recommended for use in patients with secondary hyperparathyroidism and in patients with parathyroid cysts.

{1}------------------------------------------------

The PTeye System is intended to be an aid in the identification of parathyroid tissue and not as a parathyroid locator. The use of this device has not been evaluated as a parathyroid tissue locator.

Due to a small sample size, limited clinical data is available regarding the safety and effectiveness of the PTeye System for rare disease states such as tertiary hyperparathyroidism, concomitant thyroid-parathyroid diseases, malignant parathyroid diseases, or other circumstances when prophylactic thyroidectomies are performed in individuals at high-risk for certain diseases such as MEN2A.

The PTeye requires the probe to be in direct contact with the tissue of interest for proper signal recording as its detection depth is only a few millimeters. If parathyroid tissue is covered by fat or other tissues, the fat and tissues must be manually moved out the way and/or the probe maneuvered around these tissues to make direct contact with area of interest.

PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

The Parathyroid Detection (PTeye) System aids surgeons in differentiating parathyroid tissue during common surgical procedures. The handheld probe assembly includes one glass fiber-optic element that emits non-ionizing radiation at 785 nm in the near IR range (NIR) and one fiber optic detector element that collects and transmits the fluorescence emitted by the tissue to a photo detector.

Image /page/1/Picture/6 description: The image shows a medical device with a screen displaying numerical values. The device is connected to a probe via a cable. There is also a black box-shaped object next to the device. The image also includes a diagram of a similar device, highlighting its various components such as the screen display, laser connector, detector connector, and power switch.

The Parathyroid Detection (PTeye) System consists of the following components: 1. A console that includes:

  • An LED display that indicates if the laser is on. ●
  • A display for visual feedback. ●
  • A speaker for auditory feedback. ●

{2}------------------------------------------------

The console is multiple patient reusable device.

    1. A handheld fiber-optic probe assembly that interfaces into the console unit using two unique connectors. One connector plugs into the laser output and the other plugs into the photo detector input (fluorescent signal). Fiber optic is sterile single use device.
    1. A foot pedal attached by a cable to the rear of the unit, used to control power to the laser and initiate data collection.
    1. An external power supply and power cord that plugs into the power supply.

The single use ethylene oxide (EtO) sterilized fiber-optic probe will also be sold separately.

Tissue detection is based on the ratio of the fluorescent response of parathyroid to thyroid tissue; with the assumption that the fluorescence of thyroid tissue is much lower than parathyroid. During surgery, five thyroid data points are collected by touching thyroid tissue with the probe assembly. The system calculates a baseline median for the thyroid tissue based on those points. The baseline value establishes a reference point for distinguishing parathyroid tissue, which produces a higher level of fluorescence.

Once the baseline thyroid value has been calculated, a new operational screen will display to support continuous parathyroid search mode. To operate in this mode, the foot switch must be pressed in order to activate the laser. When laser emission is taking place, the LASER ON LED at the front of the system will illuminate. Responses indicating parathyroid tissue are communicated to the user through a bar graph, a detection percentage, a detection ratio, and audio feedback. The highest percentage (100%), full bar graph, and the highest frequency beeps indicate 2.5-times the median of the 5 thyroid measurements taken during the baseline stage. The on-screen bar graph and detection percentage, and detection ratio correlate directly with the audio. When using the probe on non-parathyroid tissue, the display shows mostly yellow and a low detection ratio. When using the probe on parathyroid tissue, the display feedback shows primarily green and a high detection ratio.

Image /page/2/Picture/8 description: The image shows a medical device with a digital display. The display shows a detection percentage of 71% and a detection level of 296. There is also a reading of 2.0 on the right side of the screen. The device has labels for 'Detector Input,' 'Laser Aperture,' and 'Laser On,' along with a power switch.

{3}------------------------------------------------

Table 1: PTeye Regulatory Information
Device Trade/Proprietary NameParathyroid Detection (Model PTeye) System
Device Common NameParathyroid Autofluorescence Detection Device
Device ClassClass II
Classification Regulation878.4550
Product CodeQDF

Table 1: PTeve Requiatory Information

SUMMARY OF NONCLINICAL/BENCH STUDIES

BIOCOMPATIBILITY/MATERIALS

The PTeve fiber-optic assembly will be in direct contact with organ tissue. The sterile PTeye fiber-optic assembly was tested per FDA guidelines for biocompatibility and ISO 10993-1 according to FDA guidance document, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". The testing showed that the optical probe passed the biocompatibility requirements for cytotoxicity, sensitization, irritation or intracutaneous reactivity, acute systemic toxicity, and material-mediated pyrogenicity.

The PTeye fiber-optic assembly was tested for Limulus Amebocyte Lysate (LAL) evaluation for three product lots on the final product, post-sterilization. The results show that the test article met the requirement of less than 20 EU/device. In addition, the LAL endotoxin testing will be conducted on 3 samples as part of each product lot release criteria.

STERILITY/PACKAGING/ SHELF LIFE

The PTeye optical probe is a sterile, single use component of the device system. The cartridge is sterilized using EO sterilization and sterilant residuals were quantified and under the acceptable limits for EO and ethylene chlorohydrin (ECH). The sterilization method was validated per ISO 11135:2014 (Sterilization of health care products --Ethylene oxide: "Requirements for development, validation and routine control of a sterilization process for medical devices"). The Sterility Assurance Level (SAL) for the optical probe is 10-6.

The primary packaging consists of one probe assembly seated and secured inside an inner tray which is heat sealed to a Tyvek lid. The sealed primary packaged probe is placed inside a poly/Tyvek pouch which is heat sealed. The sealed outer pouch is placed inside the bottom half of the unit box. The top half of the unit box covers the bottom half of the unit box and the seals of the unit box are secured with tamper proof seals.

{4}------------------------------------------------

The shelf-life of the optical probe was evaluated after accelerated aging equivalent to one year. The cartridge was evaluated by visual inspection, peel test, and bubble leak testing. The test article met the acceptance criteria for each test.

ELECTROMAGNETIC COMPATIBILITY AND ELECTRICAL SAFETY

The following Electrical/Mechanical/Thermal Safety, and electromagnetic compatibility (EMC):

  • IEC 60601-1: 2005 +CORR.1:2006+CORR.2: 2007 + AM1:2012 or IEC 60601-. 1:2012 with US deviations, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
  • . IEC 60601-1-2: 2007 (Edition 3), Medical electrical equipment Part 1-2 – General requirements for basic safety and essential performance - Electromagnetic compatibility.
  • IEC 60601-1-6: Collateral Standard: Medical electrical equipment Part 1-6 - General requirements for basic safety and essential performance - Usability.

LASER/LIGHT SAFETY

The following laser safety testing has been performed:

  • IEC 60825-1: Safety of laser products Part 1: Equipment classification and . requirements

MAGNETIC RESONANCE (MR) COMPATIBILITY

Device is not compatible for Magnetic resonance (MR) environment.

SOFTWARE

The software controls the PTeye system hardware including the laser firing after pressing the foot-pedal. It calculates the background level based on five data points from thyroid tissue and determines the median values of those points. It then calculates the ratio of autofluorescence signal from the background level to demonstrate if the tissue is parathyroid or not.

The software GUI includes an on-screen bar graph, detection percentage and detection ratio which correlate directly with the audio feedback. When using the probe on nonparathyroid tissue, the display shows mostly yellow and a low detection ratio. When using the probe on parathyroid tissue. the display feedback shows primarily green and a high detection ratio. The audio feedback for non-parathyroid tissue sounds with slow beep frequency (one beep per second). As the ratio of autofluorescence to background increases, the beeps are emitted more rapidly.

{5}------------------------------------------------

Device does not have wireless capability or a communication port.

The agency considers the software to be a moderate level of concern (LOC) because inadvertent software errors could result in injury to the patient or delay in procedure time.

All of the elements of software and cybersecurity information as outlined in FDA's guidance documents "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005) and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" (issued June 14, 2013) were provided.

Overall, the software documentation included in the De Novo request is in sufficient detail to provide reasonable assurance that the software will operate in a manner described in the specifications.

TestPurposeAcceptance CriteriaResults
Short term variationTo demonstrate the influence of shortPassed
in laser intensityterm variation in laser intensity on the
device performance during a surgery
Change in laserTo demonstrate the influence of thePassed
intensity due to agingdevice ageing and degradation (like
or differentdecrease in laser intensity over time or
environmentaldue to different environmental
conditioncondition) on the performance of the
device
Effect of surgical andTo demonstrate the influence of thePassed
ambient light on thesurgical and ambient light on the
device performanceperformance of the device

PERFORMANCE TESTING - BENCH

SUMMARY OF CLINICAL INFORMATION

A. Study Design

A clinical study was conducted to support the safety and effectiveness of the AiBiomed Parathyroid Detection System (PTeve) to aid in the identification of parathyroid (PG) tissue during thyroid and parathyroid surgical procedures.

The single blinded non-randomized study was conducted at two centers, Vanderbilt and Ohio State University Medical Center and included tissue measurements during thyroid and parathyroid surgical procedures. Measurements were taken intraoperatively by surgeons who were blinded to PTeye device output. During the surgical procedure, five thyroid data points were initially collected by touching the probe tip to the thyroid tissue. The system used these 5 points to calculate a baseline near infrared autofluorescence (NIRAF) median value and this established the reference baseline for each individual patient. If no thyroid tissue was present due to a previous thyroidectomy/thyroid ablation. baseline NIRAF median was alternatively obtained on neck muscle or trachea for that particular patient. Subsequent tissue NIRAF measurements in the patient were then

{6}------------------------------------------------

normalized to this NIRAF baseline for obtaining the detection ratio. Upon visualizing a tissue of interest, the surgeon first stated the degree of confidence in having identified tissue as parathyroid gland with high, moderate or low confidence, based solely on visual inspection of the tissue-in-situ and without relying on the PTeye device. This information was recorded for assessing the performance of the PTeye as compared to the surgeon's visual assessment. The surgeon then placed the probe of the PTeye on the suspect tissue site and pressed the foot-pedal. resulting in tissue NIRAF intensity and detection ratio being displayed only to the study investigator and not surgeon in real-time.

The study was originally designed to compare the performance of the PTeye to a prior prototype parathyroid detection system. Of the original 133 patients enrolled, 82 patients were tested with the PTeye in its final design. There are technological differences between the PTeye and the prototype device, including ambient light interference and peak intensity to determine baseline (NIRAF), which may alter the device effectiveness results. Therefore, effectiveness results only for the PTeye final design were considered. However, for the safety results, both treatment groups are included in the summary below. Of note, there were no reported surgery or device related adverse events for the 51 patients whose results have been excluded from the final effectiveness analysis.

Subjects enrolled in the study included both men (23.5%) and women (76.5%) over the age of 18.

DemographicVariablesSite ASite BOverall
Number ofpatients40(38- excluding 2patients withsecondaryhyperparathyroidism)4181(79- excluding 2patients withsecondaryhyperparathyroidism)
Gender
Female28 (70.0%)34 (82.9%)62 (76.5%)
Male12 (30.0%)7 (17.1%)19 (23.5%)
Race
Caucasian34 (85.0%)39 (95.1%)73 (90.1%)
Non-Caucasian6 (15.0%)2 (4.9%)8 (9.9%)
Age (years)54.3 ± 15.852.3 ± 16.453.8 ± 19.9
BMI (kg/m²)29.3 ± 6.231.6 ± 9.230.5 ± 7.9

Table 1: Summary of Demographic Information

Clinical Inclusion and Exclusion Criteria

Inclusion Criteria:

To be eligible for study enrollment, a subject was required to satisfy each of the following criteria.

    1. Adults (18-99 years of age) scheduled to undergo parathyroid or thyroid surgery.

{7}------------------------------------------------

    1. Willing to sign the informed written consent form.

Exclusion Criteria:

A subject was not eligible to participate if they met any of the following exclusion criteria.

    1. Pregnant
    1. Unsuitable for study participation in the opinion of the Investigator- attending surgeon.

The study was initially designed to evaluate the performance of the PTeye in differentiating between parathyroid gland and non-parathyroid gland tissue once a potential candidate tissue was surgically exposed during the procedure. This was subsequently corroborated by the surgeon's visual identification for in situ parathyroid glands using an unvalidated confidence scale (low, medium, or high) and with histological examination for the excised parathyroid gland tissues.

Primary effectivenessPerformance measured by the ability of the PTeye to accurately identify
endpointsparathyroid glands [PG detection rate].
Secondary effectivenessIntra-patient and inter-patient variability of NIRAF in thyroid and PGs.
endpoints
Effect of thyroid and PG pathology on intraoperative parathyroid
Identification.
In-vivo and ex-vivo effect of blood on NIRAF intensity of PG and
thyroid with the PTeye system to assess if a hemorrhagic surgical field
would affect parathyroid identification.
Ex-vivo effect of probe-to-tissue contact pressure on PG fluorescence
intensity.
Safety EndpointSafe use as determined by a lack of (serious) adverse events.
The addition of no more than 5 minutes to the total procedure time
during normal use of the device.

Table 2: Study Endpoints

Table 3: Subject Accountability

All SubjectsFinal PTeye System
Enrolled subjectsN = 133N = 82
Completed subjects-82
Discontinued subjects51 (prototype device)3
Reason for discontinuation
Protocol deviationa1
Previously identified disease stateb2

a Protocol deviation due to communication error between study coordinator and surgeon regarding the timing of depressing the PTeye foot pedal leading to incorrect baseline NIRAF measurements.

b Patients with secondary hyperparathyroidism were determined to exhibit in prior studies and thus, were excluded from the final effectiveness performance assessment.

{8}------------------------------------------------

The ability of the PTeye to accurately identify parathyroid glands [PG detection rate] included the following assessments using objective histology and subjective expert surgeon opinion:

PTeve Performance Data Analysis:

    1. Sensitivity- Number of true positives. (as determined by PTeye and validated by histology or surgeon's visual identification with high/moderate confidence) divided by actual positives (PG sites – total number of positives as determined by histology or surgeon's visual identification with high/moderate confidence)
    1. Specificity: Number of true negatives, (as determined by PTeye and validated by histology or surgeon's visual identification with high/moderate confidence) divided by actual negatives (non-PG sites - total number of negatives as determined by histology or surgeon's visual identification with high/moderate confidence)
    1. PPV: Number of true positives, (as determined by PTeye and validated by histology or surgeon's visual identification with high/moderate confidence) divided by number of device positives (total number of positives as determined by PTeye alone)
    1. NPV: Number of true negatives, (as determined by PTeye and validated by histology or surgeon's visual identification with high/moderate confidence) divided by number of device negatives (total number of negatives as determined by PTeye alone)
    1. False positive rate: Rate of device positive measurements when tested on actual negatives (number of negatives validated by histology or surgeon's visual identification with high/moderate confidence).
    1. False negative rate: Rate of device negative measurements when tested on actual positives (number of positives validated by histology or surgeon's visual identification with high/moderate confidence).

Based on these parameters, overall accuracy of the PTeye in PG identification and associated kappa values were accordingly calculated.

Statistical significance of NIRAF intensities between thyroid, parathyroid gland, fat, muscle and trachea were determined using a two-tailed Student's t-test for unequal variance, with an alpha (level of significance) of 0.01. The same statistical approach was adopted to determine if there was a significant difference between NIRAF measured from normal and diseased thyroid and parathyroid glands.

Comparison of surgeon's visual determination versus the PTeye as validated with histology-based gold standard:

The performance accuracy of the participant surgeons in differentiating between parathyroid gland (PG) and non-PG tissues relying on their visual skills were compared to that of the PTeye. This was performed in those cases when in vivo measurements were performed on tissues that were later excised for histological validation via frozen section or Hematoxylin-Eosin stained tissue section analysis by the pathologists that could serve as the gold standard. Due to lack of histological validation of in-situ tissues, these were not considered for comparing performance accuracy between the surgeons and the PTeye. In addition, comparison of surgeon versus PTeye

{9}------------------------------------------------

were evaluated for each investigational site/study center. All participant surgeons at both centers were high-volume surgeons (who perform >25 thyroid surgeries per year and >15 parathyroid surgeries per vear).

Assessing variability and associated factors within patient data acquired with the PTeye:

Following data acquisition and analysis, the report also investigated: (i) the distribution of demographic variables at both study centers including: age (18- 99 years of age), gender (male or female), race (Caucasian or non-Caucasian), body-mass index (BMI) at time of surgery, (ii) Intra-patient and inter-patient variability of NIRAF in thyroid and PGs respectively and (iii) the effect of thyroid and parathyroid disease on intraoperative PG identification.

Influence of blood on NIRAF of thyroid and PG:

A. Ex vivo Validation:

The effect of blood on the NIRAF intensity of PG and thyroid was assessed ex vivo with the PTeye in order to determine if a hemorrhagic surgical field would affect PG identification. Three fresh frozen specimens each, of normal thyroid and PG adenoma were obtained from the NIH funded Co-operative Human Tissue Network (Vanderbilt University Medical Center, Nashville, TN). After thawing the specimens, at least six NIRAF measurements were obtained from each specimen ex vivo. To simulate a hemorrhagic surgical field, 0.5 cc of heparinized murine blood was introduced on to the specimen surface. NIRAF intensity of each specimen was measured with the PTeye and normalized to the thyroid NIRAF and grouped into four categories: (i) thyroid without blood (n, (ii) thyroid with blood, (iii) PG without blood and (iv) PG with blood, with each group consisting of 18 NIRAF measurements. Statistical significance was determined using a two-tailed Student's t-test for unequal variance, with an alpha level of significance, alpha of 0.01.

B. In vivo validation:

The influence of a hemorrhagic surgical field on PG identification with the PTeye was also tested in vivo using 4 PGs from 3 patients. The surgeon upon identifying a tissue thought to be the PG with high/moderate confidence, obtained PTeye measurement from the tissue. In these three patients -1-2 measurements were taken when the PG was covered with blood, which routinely occurs during the surgical procedure. After the measurement, the surgeon suctioned the blood away, rinsed the tissue with saline and repeated the PTeye measurement on the same tissue site. Detection ratios were grouped for three categories: (i) PG with blood and (iii) PG without blood, with each group consisting of at least 6 NIRAF measurements. Statistical significance between (i) thyroid and PG without blood and (ii) thyroid and PG with blood were determined by a 2-tailed Student's t-test for unequal variance, with an alpha level of significance of 0.01.

Effect of probe-to-tissue contact pressure on tissue NIRAF measurements with the PTeye:

{10}------------------------------------------------

The effect of probe-to-tissue contact pressure on tissue NIRAF intensity was examined in vitro on excised fresh frozen human PG (n=3) and thyroid (n=2) tissues. NIRAF measurements with the PTeve were collected where the user reported the probe contact pressures qualitatively to be 'mild', 'moderate', and 'high'. Differences in PG detection ratios were correlated with the degree of probe-to-tissue contact pressure.

Schedule of assessments:

There were no formal follow-up assessments or visits for the patient: subjects exit the study at the conclusion of standard recovery from the surgical procedure.

Safety definitions and reporting requirements:

All adverse events (AEs), regardless of seriousness, severity, or relationship to the study device, were to be recorded in the Case Report Forms (CRF) by the investigators and/or study coordinators and reported to corresponding IRBs at the two centers and to the Principal Investigator. The evaluation was to include a determination of the seriousness and severity of the event, whether the event or the severity of the event was anticipated or unanticipated, and the relationship of the event to the study device.

A serious adverse event (SAE) is defined according to ISO 14155:2003 as any adverse event that:

  • Led to a death
  • Led to a serious deterioration in the health of the subject that ●
  • Resulted in a life-threatening illness or injury ●
  • . Resulted in a permanent impairment of a body structure or a body function
  • Required in-patient hospitalization or prolongation of existing hospitalization
  • . Resulted in medical or surgical intervention to prevent permanent impairment to body structure or a body function

An unanticipated adverse device effect (UADE) is defined per 21 CFR 812.3 as any serious adverse effect on health or safety or any life-threatening problem or death caused by, or associated with, a device, if that effect, problem, or death was not previously identified in nature, severity, or degree of incidence in the investigational plan or application (including a supplementary plan or application), or any other unanticipated serious problem associated with a device that relates to the rights, safety, or welfare of subjects.

SAE and UADE were to be reported by the investigators to the IRB and to AiBiomed within 24 hours. Serious AEs and UADEs were documented on the Serious Adverse Event Form.

Safety and Effectiveness Results

    1. Safety Results

{11}------------------------------------------------

At the end of patient follow-up period of two weeks, no adverse events related to the procedure or during the study period were reported. There were no adverse events, SAE's, or UADE's reported in the 81 patients considered for effectiveness data analysis. No clinical issues or adverse events related to the intra-operative use of the device description of procedural difficulties, or device complaints were reported at the closure of the study.

Individual measurements once the PTeye device is set up take approximately 2 seconds. (5) thyroid baseline measurements, (4) extra thyroid measurements, (8) PG measurements, (2) fat measurements, (2) trachea measurements, and (2) muscle measurements per patients took approximately one minute of additional total procedure time during normal use of the device, this was under the 5-minute pre-defined cut-off.

2. Effectiveness Results

Primary endpoint parathyroid identification rate:

Measurements were obtained using the final design of the PTeye on 181 PGs and 546 non-PG tissues (194 thyroid, 116 fat. 119 neck muscle, and 117 trachea) in 81 patients. Individual performance data was populated for all 81 patients (40 from site A and 41 from site B). Of the 181 PGs measured, fluorescence measurements of 68 PGs (68/181, 37.6%) were confirmed with histology. The remaining 113 PGs could not be confirmed with histology and were validated based on high or medium confidence of the surgeon's visual assessment. Tissues identified with low confidence by the surgeon were excluded from further analysis unless histological validation was obtained from those tissues.

A. Normalized NIRAF Intensity:

Testing with the PTeve system vielded 362 measurements from 181 PGs and 546 measurements from non-PG locations in 81 patients undergoing parathyroidectomy and/or thyroidectomy.

Figure 1: Normalized NIRAF Intensity to Tissues:

{12}------------------------------------------------

Image /page/12/Figure/0 description: This bar graph shows the tissue NIRAF normalized to thyroid NIRAF. The x-axis shows the different tissues, including thyroid, parathyroid, muscle, fat, and trachea. The y-axis shows the NIRAF values, with parathyroid having the highest value of approximately 3.5, while the other tissues have values less than 1.

NIRAF measured with the PTeye on different neck tissues normalized to NIRAF of the thyroid. Error bar – Standard Error. **p-value <0.01 for normalized NIRAF of PGs compared to the non-PG tissues - thyroid, fat, muscle and trachea.

As seen above, normalized NIRAF intensity measured from PGs was significantly higher than that of the non-PG tissues in the neck such as thyroid, fat, muscle and trachea (pvalue = 1.21 × 10-4). Overall, the normalized NIRAF intensity for PG tissue measured with the PTeye was about 5.4 times higher than that of the measured thyroid, while muscle, fat and trachea showed little to no NIRAF intensity.

B. Accuracy of PG identification based on histology/surgeon validation:

The PTeye could successfully identify PGs with 92.3% sensitivity (167/181) and 97.3% specificity (531/546). PG identification with the PTeye system had an overall accuracy of 96% (kappa = 0.89) with 2.7% false positive rate and 7.7% false negative rate. The PPV and NPV for this system were 91.8% and 97.4% respectively. The performance accuracy of PTeye in identifying PG tissue was achieved without needing to switch off ambient OR lights. Corresponding information from the surgeon assessment and histology (when available) from the same tissue in the same patient is presented below.

Table 4: PTeye Performance Based on Histology and Expert Surgeon Corroboration

{13}------------------------------------------------

All Diagnoses (with histology)Histology+ for ParathyroidHistology- for Parathyroid
PTeye+ for Parathyroid611
PTeye- for Parathyroid74
All Diagnoses (with surgeon's eye)Expert Surgeon+ for ParathyroidExpert Surgeon- for Parathyroid
PTeye+ for Parathyroid10614
PTeye- for Parathyroid7527
All Diagnoses (combined)Gold Standard+ for ParathyroidGold Standard- for Parathyroid
PTeye+ for Parathyroid16715
PTeye- for Parathyroid14531
Sensitivity: 89.7%; Specificity: 80.0%;PPV: 98.4%; NPV: 36.4%;False Positive Rate: 20.0%; False Negative Rate: 11.3%Parathyroid glands (with surgeon's eye for validation):Accuracy: 96.8%; k: 0.89;Sensitivity: 93.8%; Specificity: 97.4%;PPV: 88.3%; NPV: 98.7%;False Positive Rate: 2.6%; False Negative Rate: 6.2%Overall PerformanceAccuracy: 96.0%; Kappa: 0.89;Sensitivity: 92.3%; Specificity: 97.3%;PPV: 91.8%; NPV: 97.4%;False Positive Rate: 2.7%; False Negative Rate: 7.7%Histology and surgeon's assessment information that validates PTeye performance indifferentiating between PG and non-PG tissues in 81 patients at sites A & B - including 2 r-

surgeon's visual assessment with high-moderate degree of confidence for in-situ tissues

The performance of the PTeye improves further upon exclusion of renal-induced secondary hyperparathyroidism (r-SHPT) patients. in whom the PGs exhibit NIRAF very irregularly as found in prior studies by McWade et al. Surgery (2016). Based on McWade et al results, the labeling excludes using the PTeye in patients with r-SHPT. Removing data from the r-SHPT patients, led to the PTeye identifying PGs with an even higher sensitivity at 93.6% (162/173), while a 97.1% specificity (510/525) was obtained. As a result, an overall accuracy of 96.3% (kappa = 0.90) was achieved with 2.9% false positive rate and 6.4% false negative rate. The PPV and NPV for this system were 91.5% and 97.9% respectively.

Table 5: PTeye Performance Based on Histology and Expert Surgeon Corroboration (excluding secondary hyperparathyroid patients)

{14}------------------------------------------------

All Diagnoses (with histology)Histology+ for ParathyroidHistology-for Parathyroid
PTeye+ for Parathyroid561
PTeye- for Parathyroid44
Expert Surgeon+ for ParathyroidExpert Surgeon- for ParathyroidAll Diagnoses (with surgeon's eye)
PTeye+ for Parathyroid10614
PTeye- for Parathyroid7506
All Diagnoses (combined)Gold Standard+ for ParathyroidGold Standard-for Parathyroid
PTeye+ for Parathyroid16215
PTeye- for Parathyroid11510
Parathyroid glands (with histological validation):Accuracy: 92.3%; Kappa: 0.58;Sensitivity: 93.3%; Specificity: 80.0%;PPV: 98.3%; NPV: 50.0%;False Positive Rate: 20.0%; False Negative Rate: 6.7%Parathyroid glands (with surgeon's eye for validation):Accuracy: 96.7%; k: 0.89;Sensitivity: 93.8%; Specificity: 97.3%;PPV: 88.3%; NPV: 98.6%;False Positive Rate: 2.7%; False Negative Rate: 6.2%Overall Performance:Accuracy: 96.3%; Kappa: 0.90;Sensitivity: 93.6%; Specificity: 97.1%;PPV: 91.5%; NPV: 97.9%;False Positive Rate: 2.9%; False Negative Rate: 6.4%Histology and surgeon's assessment information that validates PTeye performance in

C. Performance of the surgeon's eye versus the PTeye System when correlated with histological validation

The four surgeons at the two study sites are high- volume surgeons who perform more than 25 thyroidectomies and more than 15 parathyroidectomies each year. At study site A (Vanderbilt), surgeon #1 has 16 years of experience as a practicing endocrine surgeon while surgeon #2 has 4 years of the same. Similarly, at study site B (Ohio State), surgeon #3 and surgeon #4 have 16 and 4 years respectively of experience as endocrine surgeons.

At site A. the surgeons could correctly identify 95.2% of the excised PGs (40 out of 42) in comparison to 92.9% detected with the PTeye (39 out of 42), when cross validated with histology. In contrast, the PTeye performed better at site B by correctly detecting 94.4% of the excised PGs (17 out of 18), in comparison to 88.9% detected by the surgeons (16 out of 18). Overall the detection rate for PG tissues when validated with histology remained the same for surgeons versus the PTeye at 93.3% as seen below. The PTeye however, outperformed the surgeons at both study sites in correctly identifying non-PG tissues which were excised for histological validation. The PTeye was accurate in determining that the

{15}------------------------------------------------

tissue was not a PG in 80% of the specimens (4 out of 5), while the surgeons were incorrect for all the non-PG specimens (0 out of 5).

Diagnoses(with histology)Excised PGsExcised non-PGs tissue (excised assumed as PG)*
SurgeonsPTeyeHistologySurgeonsPTeyeHistology
Site A403942034
Percentage40/42 (95.2%)39/42 (92.8%)0/4 (0.0%)3/4 (75.0%)
Site B161718011
Percentage16/18 (88.9%)17/18 (94.4%)0/1 (0.0%)1/1 (100.0%)
Site A + Site B5656600155
Percentage56/60 (93.3%)56/60 (93.3%)0/5 (0.0%)4/5 (80.0%)

Table 6: Performance of the surgeons versus that of the PTeye when validated with histology for excised PGs - excluding data from 2 r-SHPT patients

Table 7: Individual surgeon contribution to the number of surgeries and tissue readings

Surgeon withexperience.(16 or 4 years)SiteA. VanderbiltB. Ohio StateNumber ofsurgeriesNumber ofPG'sNumber ofPGreadingsNumber ofnon PGreadings
Surgeon AExperience: 16 yearsVanderbilt UniversityMedical Center226412897
Surgeon BExperience: 4 yearsVanderbilt UniversityMedical Center184896119
Surgeon CExperience: 16 yearsOhio State UniversityMedical Center3863126305
Surgeon DExperience: 4 yearsOhio State UniversityMedical Center361225
Total81181362546

Secondary Effectiveness Endpoints:

A. Intra-patient and inter-patient variability of NIRAF in PG and thyroid tissues:

Among the 81 patients studied with the PTeye, no PG tissues were visually identified by the surgeon in Patient #46 and #88 for the purpose of device testing. There were no thyroid NIRAF measurements in patient #36. #106 and #133. as these patients had no thyroid remnants due to previous thyroid interventions. Baseline NIRAF was performed on muscle

{16}------------------------------------------------

or trachea in these patients. While the baseline thyroid NIRAF was obtained for patient #73, additional thyroid measurements were not obtained due to lack of time.

Data analysis of the NIRAF from PG and thyroid tissues across these patients revealed that intra-patient variability of normalized NIRAF in thyroid averaged at 28.6%, while that of the PGs was 34%. When the inter-patient variability was analyzed as the deviation of individual patient NIRAF from the mean calculated across all patients expressed as a percentage, the inter-patient variability for normalized NIRAF in thyroid averaged at 74%. In comparison, the mean inter- patient variability for normalized NIRAF in PGs was about 1.5 times higher at 112%. Despite the higher intra-patient and inter-patient variability, normalized NIRAF intensities for PGs were consistently higher than that of the thyroid glands. It can also be noted that PG NIRAF are notably lower compared to the thyroid in the two r-SHPT subjects, i.e. patient #39 and #48 - as seen in the figure below.

{17}------------------------------------------------

Image /page/17/Figure/0 description: The image shows four bar graphs comparing the NIRAF normalized to thyroid tissue for parathyroid and thyroid tissues at two different sites. Site A includes data from 40 patients, while Site B includes data from 41 patients. Each graph displays the NIRAF values for individual patients, with the x-axis representing patient numbers and the y-axis representing the NIRAF values. The parathyroid values are consistently higher than the thyroid values across all patients and both sites.

Figure 2: Intra-patient and inter-patient variability of tissue NIRAF measured with the PTeye in PG and thyroid tissues

Intra-patient and inter-patient variability of tissue NIRAF measured with the PTeye in PG and thyroid tissues. Since tissue NIRAF normalized to thyroid NIRAF for PGs in Patient #83 had values that exceeded 30, the y-axes was truncated at an upper limit of 10 to depict intra-patient and inter-patient variability in tissue NIRAF among all 81 patients. Error bars - Standard error

Since tissue NIRAF normalized to thyroid NIRAF for PGs in Patient #83 had values that exceeded 30, the y-axes was truncated at an upper limit of 10 to depict intra-patient and interpatient variability in tissue NIRAF among all 81 patients. Error bars - Standard error

{18}------------------------------------------------

B. Effect of thyroid and/or PG disease on NIRAF intensity assessed by the PTeye:

As shown below, patients assessed with the PTeye system included: (i) 46 cases of diseased thyroid with normal PG, (ii) 31 cases of diseased PG with normal thyroid, (iii) 3 cases of diseased thyroid and diseased PG and (iv) 1 case of normal thyroid and PG (admitted for prophylactic thyroidectomy). The diseased thyroid cases comprised of (i) non-toxic benign thyroid diseases (n=23), (ii) toxic benign thyroid diseases (n=8) and malignant thyroid conditions (n=15). Cases of diseased PGs were predominantly primary hyperparathyroidism (n=28), with two renal failure induced-secondary hyperparathyroidism (r-SHPT) and one tertiary hyperparathyroidism cases, (note that the occurrence of the latter two diseases are very rare). PG rate was found to be the lowest for renal failure induced secondary hyperparathyroidism (r-SHPT) patients at 62.5%. Therefore r-SHPT patients were excluded from the indications for use of the PTeye.

Thyroid/Parathyroid diseaseSite A(# of Patients)Site B(# of Patients)Total(# of patients)Parathyroid glandidentification rate# of PG glands, (% identified)
Benign thyroid disease-non-toxic(solitary nodule/multinodular goiter/Hashimoto'sthyroiditis)12112335/40(87.5%)
Benign thyroid disease-toxic(Grave's Disease/solitary nodule/multinodulargoiter/thyroiditis)26817/18(94.4%)
Malignant thyroid diseaseбರಿਹ ਦ38/39(97.4%)
Primary hyperparathyroidismਹਵ122861/65(93.9%)
Secondary hyperparathyroidism2025/8(62.5%)
Tertiary hyperparathyroidism1013/3(100.0%)
Concomitant thyroid-parathyroid disease0335/5(100.0%)
Prophylactic thyroidectomy for MEN2A1013/3(100.0%)

Table 8: Distribution of different disease groups tested with PTeye and PG identification rate for each disease group.

As observed in the figure below, there was no significant difference seen in the normalized NIRAF intensity between healthy and diseased thyroid glands (p=0.96). On the other hand, diseased PGs exhibited a lower normalized NIRAF intensity as compared to healthy PGs with demonstrable statistical significance (p=0.00012). However, PG tissue exhibited significantly higher normalized NIRAF intensity as compared to the thyroid gland, irrespective of whether the thyroid or PG was normal or diseased (p <0.00001)

Figure 3: Variation of normalized NIRAF intensity between healthy thyroid, diseased thyroid, healthy PG and diseased PG states tested with the PTeve

{19}------------------------------------------------

Image /page/19/Figure/0 description: The image is a bar graph comparing tissue NIRAF normalized to thyroid NIRAF in healthy and diseased thyroid and parathyroid tissues. The y-axis represents the tissue NIRAF normalized to thyroid NIRAF, ranging from 0 to 5. The x-axis represents the different tissue types: healthy thyroid, diseased thyroid, healthy parathyroid, and diseased parathyroid. The graph shows that healthy parathyroid tissue has the highest NIRAF value, around 4, while healthy and diseased thyroid tissues have NIRAF values below 1.

Variation of normalized NIRAF intensity between healthy thyroid, diseased thyroid, healthy PG and diseased PG states tested with the PTeve. Error bar - Standard Error. **p-value <0.001 for normalized NIRAF of PG compared to thyroid - regardless of whether healthy or diseased, **p-value <0.001 for normalized NIRAF of healthy PG compared to diseased PG.

Performance of the PTeye in Cases of Renal Failure Induced Secondary Hyperparathyroidism (r-SHPT)

Normalized NIRAF intensity of the PGs in r-SHPT patients were frequently found to be lower than 1.2 (the threshold set for PG identification), in contrast with healthy and other types of diseased PGs. In the 2 r-SHPT patients evaluated with the PTeye, only 62.5% of the PGs (5 out of 8) had a normalized NIRAF intensity greater or equal to 1.2. This was also in agreement with the findings reported in an earlier study reported by McWade et al. (Surgery, 2016). Sensitivity of PG identification with the PTeye improved to 93.6% (162/173 PGs), if patients with r-SHPT were excluded from performance analysis. The finding suggests that a device such as PTeye in its present design may not be useful in aiding PG identification in r-SHPT cases.

C. Influence of hemorrhagic field on tissue NIRAF measurements with the PTeye:

Effect of Blood on NIRAF on thyroid and PG specimen's ex vivo

The influence of blood on intraoperative PG identification with PTeye was studied with an ex vivo simulated experiment using thyroid and PG specimens covered with heparinized murine blood. Figure 12 represents data from an ex vivo experiment that shows the effect of blood on NIRAF measured from thyroid and/or PG tissues using the PTeye. The findings indicate that normalized NIRAF intensity of thyroid tissue was lower in the presence of blood as compared to that of the same thyroid specimens when blood was washed away (p = 0.007). In comparison, normalized NIRAF of PG specimens with blood was lower than that of PG specimens without blood but not statistically significantly (p=0.53). Moreover, the PG specimens was found to have normalized

{20}------------------------------------------------

NIRAF intensity significantly higher than the thyroid with and without blood - regardless of the presence of blood on the PG itself.

Image /page/20/Figure/1 description: The image is a bar graph comparing tissue NIRAF normalized to thyroid NIRAF for thyroid and parathyroid specimens with and without blood. The y-axis represents the normalized NIRAF values, ranging from 0 to 9. The x-axis shows four categories: "Thyroid without blood", "Thyroid with blood", "Parathyroid without blood", and "Parathyroid with blood". The graph indicates that the parathyroid specimens have significantly higher NIRAF intensity compared to the thyroid specimens, with p-values less than 0.01.

Figure 4: Influence of blood on normalized NIRAF intensity ex vivo

Effect of Blood on NIRAF on thyroid and PG specimens in vivo

The same findings were observed upon investigating the influence of blood in vivo as observed in the figure below. Based on the normalized NIRAF measurements obtained from three (3) patients with two measurements per PG (with and without blood), there is no significant difference in the NIRAF of PGs in presence and absence of blood even in an in vivo setting (p=0.95). Further, the PGs in vivo were found to have normalized NIRAF intensity significantly higher than the thyroid, regardless of the presence of blood on the PG (p=0.005).

{21}------------------------------------------------

Image /page/21/Figure/0 description: The image is a title for a figure. The title reads, "Figure 5: Influence of blood on normalized NIRAF intensity in vivo". The title is written in a bold, sans-serif font. The figure number is 5.

Influence of blood on normalized NIRAF intensity measured in vivo on PGs in 3 patients with the PTeye (2 measurements per parathyroid without and with blood). Error bar – Standard Error. ** denotes statistical significance between blue and red bar, with p-value <0.01 for NIRAF intensity of PG compared to thyroid.

D. Effect of probe-to-tissue contact pressure on NIRAF measurements with the PTeye:

Upon assessing the influence of probe contact pressure on tissue NIRAF measurement with the PTeye, no notable difference in the tissue NIRAF measurements was observed with probe pressure (mild, moderate or high). This applied to both PG as well as thyroid specimens as seen in the figure below. It was also noted that PG NIRAF levels stayed consistently and significantly higher than that of the thyroid, regardless of the probe contact pressure for either tissue type (p=2.2×10-10)

{22}------------------------------------------------

Image /page/22/Figure/0 description: The image is a bar graph that shows the influence of probe pressure on normalized NIRAF intensity measured ex vivo on thawed frozen human PG and thyroid specimens with the PTeye. The x-axis shows the different types of tissues, including Parathyroid A, Parathyroid B, Parathyroid C, Thyroid A, and Thyroid B. The y-axis shows the tissue NIRAF normalized to thyroid NIRAF. The graph shows that the NIRAF intensity is highest for Parathyroid A and lowest for Thyroid A and Thyroid B. The error bars represent the standard error, and the ** indicates a p-value <0.001 for NIRAF intensity of PG compared to thyroid.

Figure 6: Influence of probe pressure on normalized NIRAF intensity measured ex vivo

Pediatric Extrapolation

In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.

LABELING

The device user manual and instructions for use include a description of the device technical parameters and instructions for use for the device. The user manual also contains relevant findings from the clinical study with the detection performance characteristics of the device when used as intended. The document also states the shelf life for any sterile components as well as the necessary measures to properly dispose of any single use items and clean the reusable components of the device.

The user manual includes a warning for use with patients with secondary hyperparathyroidism or parathyroid cysts. Additionally, there is a warning not to use the PTeye device as a parathyroid tissue locator. Lastly, there is a precaution stating that due to a small sample size, limited clinical data is available regarding the safety and effectiveness of the PTeye System for rare disease states such as: tertiary hyperparathyroidism, concomitant thyroid-parathyroid diseases, malignant parathyroid diseases, or other circumstances when prophylactic thyroidectomies are performed in individuals at high-risk for certain diseases such as MEN2A.

{23}------------------------------------------------

Labeling for this device is in accordance with the special controls listed below.

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of the autofluorescence detection device for general surgery and dermatological use and the measures necessary to mitigate these risks.

Identified Risks to HealthMitigation Measures
Electrical, mechanical, or thermal hazardsleading to user injury or discomfortElectromagnetic compatibility testing
Electrical, mechanical and thermal safety testing
Software verification, validation, and hazard analysis
Labeling
Tissue, skin burn, or eye injury due tolight and laser exposureLight and laser exposure safety testing
Labeling
Infection and cross contaminationSterilization validation
Shelf life testing
Labeling
Adverse tissue reactionBiocompatibility evaluation
False identification of target tissues orstructures leading to errors in patientmanagement (e.g., removal of healthytissue or not removing diseased tissue)In vivo performance testing
Software verification, validation, and hazard analysis
Labeling

Table 9 - Identified Risks to Health and Mitigation Measures

{24}------------------------------------------------

SPECIAL CONTROLS:

In combination with the general controls of the FD&C Act, the autofluorescence detection device for general surgery and dermatological use is subject to the following special controls:

  • (1) In vivo testing under the anticipated conditions of use must characterize the ability of the device to detect autofluorescent signals from tissues or structures consistent with the indications for use.
  • (2) The patient-contacting components of the device must be demonstrated to be biocompatible.
  • (3) Performance testing must demonstrate the electromagnetic compatibility and electrical, mechanical and thermal safety of the device.
  • (4) Software verification, validation, and hazard analysis must be performed.
  • (5) Performance testing must demonstrate the sterility of patient-contacting components of the device.
  • (6) Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
  • (7) Performance testing must demonstrate laser and light safety for eye, tissue and skin.
  • (8) Labeling must include the following:
    • (i) Instructions for use;
    • The detection performance characteristics of the device when used as (ii) intended: and
    • (iii) A shelf life for any sterile components.

BENEFIT/RISK DETERMINATION

Risks:

The risks of the device are based on nonclinical laboratory studies as well as data collected in a clinical study described above.

No device or procedure related adverse events (AEs), serious adverse events (SAEs), or unanticipated adverse device effects (UADEs) were observed in the clinical study. This is likely due to limitations of the pivotal study design. Not all end users (non-endocrine surgeonspecialists) were tested and no additional human factors testing was performed. Due to the single blinded study design (surgeons blinded to device output) we do not know if device use aided in the identification of parathyroid tissue (e.g. surgeon left tissue behind that was healthy parathyroid tissue or removed uncertain tissue due to device identification/confirmation as parathyroid tissue). The study demonstrated the risks surgeons and patients would have using current techniques of: visual identification of parathyroid gland tissue, intra-operative parathyroid hormone measurement, and frozen histology. The pivotal clinical study was a comparative study to determine the sensitivity/specificity/accuracy of the subject device in identifying parathyroid tissue compared to expert surgeon opinion and histology. Given the clinical study design limitations, there is a moderate degree of uncertainty for the risks of device use. The potential risks of incorrect parathyroid gland identification during parathyroid and thyroid surgery are well-known, namely:

  • . Accidental removal of parathyroid tissue during surgical procedures (4-20%)

{25}------------------------------------------------

  • Accidental injury to parathyroid tissue during surgical procedures
  • . Hypocalcemia(hypoparathyroidism) after thyroid/parathyroid surgery (20-30%)
  • Increased operative time due to prolonged identification/confirmation of parathyroid glands (use of frozen histology or PTH blood testing)

The degree to which these risks will be experienced by both end user and patient using the PTeye device is uncertain given the moderate degree of risk uncertainty due to the lack of supportive data due to the study design.

Benefits:

The probable benefits of the device are based on nonclinical laboratory studies as well as data collected in a clinical study described above.

The magnitude of the clinical benefit has been not established based on the study design and data provided by the sponsor. There is potential for clinical benefit due to device use, namely:

  • Prevent accidental removal of parathyroid tissue during surgical procedures (4-20%) ●
  • Prevent accidental injury to parathyroid tissue during surgical procedures ●
  • Reduce the risk of hypocalcemia (hypoparathyroidism) after thyroid/parathyroid surgery (20-30%)
  • Reduce Operative time with positive identification of parathyroid (no or reduced frozen or PTH blood testing)

The pivotal study represents a comparative study, there is a lack of pre-specified: endpoints and success criteria. Robustness of the data is moderate due to the single blinded study design and quantitative histological confirmation of tissues in the pivotal study. The ability to accurately identify parathyroid tissue using the subject device. PTeve, has been established from the pivotal study data. The methods to confirm device ability in detecting parathyroid tissue are a combination of an unvalidated, subjective, four endocrine expert surgeons' opinion/confidence scale (low, medium, and high) in correctly identifying parathyroid tissue and histology confirmed specimens in 37.6% (68/181) of the parathyroid measurements (81 patients, 181 parathyroid gland measurements) used to determine true positive and true negative tissue identification rates.

The usability of the device to detect parathyroid tissue is limited to the testing of expert endocrine surgeons. The device is indicated for use only once a probable parathyroid is visualized by a surgeon seeking confirmation (identification aide) and should not be applied for localizing a 'missing' parathyroid (i.e. not to be used as a parathyroid finder or locator). A surgeon not familiar with identifying parathyroid tissue may not benefit from device use and this is compounded by the study design limited to expert endocrine surgeon testing to support parathyroid identification rates of the device itself. If the surgeon (end-user) does not benefit from device use, the patient will not benefit from device use (lower operative time, reduce parathyroid injury or accidental removal rates, reduce post-operative hypocalcemia rates). The data does support the feasibility of device use to identify parathyroid tissue as well as an expert surgeon 93.3% in most disease states (not including secondary hyperparathyroidism) and has potential for clinical benefit (for both surgeon and patient alike).

{26}------------------------------------------------

Risks of device use, and study limitations are mitigated with warnings for use in patients with: secondary hyperparathyroidism, parathyroid cysts, or for using the PTeye device to locate parathyroid tissue instead of as an adjunctive aid in parathyroid identification of visually confirmed parathyroid tissue. Additionally, device use risks and study design limitations are further mitigated with the addition of a precaution stating that "Due to a small sample size, limited clinical data is available regarding the safety and effectiveness of the PTeye System for rare disease states such as: tertiary hyperparathyroidism, concomitant thyroid-parathyroid diseases, malignant parathyroid diseases, or other circumstances when prophylactic thyroidectomies are performed in individuals at high-risk for certain diseases such as MEN2A."

Lastly, the risks of device use can be mitigated so that the benefits outweigh the risks of device use have not been fully established or understood from the limited testing design and human factors provided in the submission. However, the moderate uncertainty risk level can be mitigated if using the subject device as an adjunct to current techniques of parathyroid identification (visual inspection, frozen section histology, and intra-operative parathyroid blood level measurements). Adjunctive medical devices are defined as: Therapeutic or diagnostic products used in conjunction with but not required by another medical assessment or intervention and not intended to be a sole therapy or stand-alone diagnostic. Using the PTeye device as an adjunct diagnostic device in conjunction with current techniques may offer potential clinical benefit to the end-user and patient while mitigating device use risks to levels already experienced by patients undergoing thyroid or parathyroid surgery.

Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

Benefit/Risk Conclusion

In conclusion, given the available information above, for the following indication statement:

The AiBiomed Parathyroid Detection System (Model PTeye) is an adjunctive tool intended to aid in the identification of parathyroid tissue by confirming parathyroid tissue already visually located by the surgeon.

The probable benefits outweigh the probable risks for the Parathyroid Detection System (PTeye). The device provides benefits and the risks that can be mitigated using general controls and the identified special controls.

{27}------------------------------------------------

CONCLUSION

The De Novo request for the Parathyroid Detection System (PTeye) is granted and the device is classified under the following:

Product Code: QDF Device Type: Autofluorescence detection device for general surgery and dermatological use Class: II

Regulation: 21 CFR 878.4550

REFERENCES

    1. McWade MA , Paras C, White LM, Phay JE, Solórzano CC, Broome JT, Mahadevan-Jansen A. Label-free intraoperative parathyroid localization with near-infrared autofluorescence imaging. A.J. Clin Endocrinol Metab. 2014 Dec:99(12):4574-80.
    1. Paras C , Keller M, White L, Phay J, Mahadevan-Jansen A. Near-infrared autofluorescence for the detection of parathyroid glands. J. Biomed Opt. 2011 Jun;16(6):067012.
    1. McWade Melanie A., Paras C. White LM, Phay JE, Mahadevan-Jansen A, Broome JT, A novel optical approach to intraoperative detection of parathyroid glands. Surgery. 2013 Dec;154(6):1371-7.
    1. McWade MA, Sanders ME, Broome JT, Solórzano CC, Mahadevan-Jansen A. Establishing the clinical utility of autofluorescence spectroscopy for parathyroid detection. Surgery. 2016 Jan:159(1):193-202
    1. De Leeuw F. Breuskin I. Abbaci M. Casiraghi O. Mirghani H. Ben Lakhdar A. Laplace-Builhé C, Hartl D. Intraoperative Near-infrared Imaging for Parathyroid Gland Identification by Auto-fluorescence: A Feasibility Study. World J Surg. 2016 Sep:40(9):2131-8
    1. Ladurner R, Sommerey S, Arabi NA, Hallfeldt KK, Stepp H, Gallwas JK. Intraoperative near-infrared autofluorescence imaging of parathyroid glands. Surg Endosc. 2016 Nov 14.
    1. Thomas G. Sanders M. Solórzano C. Mahadevan-Jansen A. Evaluating a clinical prototype for label-free rapid intra-operative parathyroid detection. Unpublished manuscript.

§ 878.4550 Autofluorescence detection device for general surgery and dermatological use.

(a)
Identification. An autofluorescence detection device for general surgery and dermatological use is an adjunct tool that uses autofluorescence to detect tissues or structures. This device is not intended to provide a diagnosis.(b)
Classification. Class II (special controls). The special controls for this device are:(1) In vivo testing under anticipated conditions of use must characterize the ability of the device to detect autofluorescent signals from tissues or structures consistent with the indications for use.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance testing must demonstrate the electromagnetic compatibility and electrical, mechanical, and thermal safety of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) Performance testing must demonstrate the sterility of patient-contacting components of the device.
(6) Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(7) Performance testing must demonstrate laser and light safety for eye, tissue, and skin.
(8) Labeling must include the following:
(i) Instructions for use;
(ii) The detection performance characteristics of the device when used as intended; and
(iii) A shelf life for any sterile components.