K Number
DEN170047

Validate with FDA (Live)

Manufacturer
Date Cleared
2018-04-03

(204 days)

Product Code
Regulation Number
866.3985
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Unyvero LRT Application is a qualitative nucleic acid multiplex test intended for the simultaneous detection and identification of nucleic acid sequences from the following microorganisms and antibiotic resistance markers in endotracheal aspirates from adult hospitalized patients with suspected lower respiratory tract infections.

The Unyvero LRT Application performed on the Unyvero System is indicated as an aid in the diagnosis of lower respiratory tract infection in adult hospitalized patients with signs and symptoms of lower respiratory infection: results should be used in conjunction with other clinical and laboratory findings. As tracheal aspirates commonly contain colonizing microorganisms, detection of Unyvero LRT microbial targets does not indicate that the microorganism is the cause of the disease. Unyvero positive results do not rule out co-infection with microorganisms not detected by the Unyvero LRT Application. Negative results do not preclude lower respiratory infection, as the causative agent may be a microorganism not detected by this test.

A negative result for any antibiotic resistance marker does not indicate that detected microorganisms are susceptible to applicable antimicrobial agents. Detected resistance markers cannot be definitively linked to specific microorganisms, and may be present in organisms that are not detected by the Unyvero LRT Application such as organisms present as colonizing or normal flora.

Microbiology cultures of aspirates should be performed to obtain isolates for species identification and antimicrobial susceptibility testing, to differentiate quantities of identified microorganisms as well as normal flora present in the specimen and to identify potential microorganisms not targeted by the Unyvero LRT Application.

Device Description

The Unyvero LRT Application is a qualitative test that includes specimen processing, genomic bacterial DNA isolation and purification, multiplex PCR and array hybridization and detection. The Unyvero LRT Application performed using the Unyvero System detects specific nucleic acid sequences from microorganisms and resistance markers in tracheal aspirates collected from patients with signs and symptoms of lower respiratory infection.

The Unyvero LRT Application consists of the following components:

  • Unyvero LRT Cartridge: Contains DNA isolation and purification reagents, a DNA isolation column, eight separate PCR chambers with eight corresponding detection arrays. The Cartridge also contains fluorescently-labeled primers, hybridization and wash buffers and oligonucleotide probes for detection of targeted PCR products using array hybridization technology.
  • Unyvero T1 Sample Tube: Contains glass beads and buffers to lyse bacteria and liquefy the sample.
  • Unyvero T1 Sample Tube Cap (with Internal Control): Contains proteinase K and a synthetic internal control gene for process monitoring. The T1 Sample Tube Cap seals the Unyvero Sample Tube after which the internal control is combined with each patient specimen. The internal control DNA sequence does not have significant homology to targeted sequences and is amplified independently in each of the eight PCR chambers and the amplified internal control product is hybridized on each array.
  • Unyvero M1 Master Mix: Contains reagents for DNA amplification.
  • Unyvero T1 Transfer Tool: The Transfer tool can be used to transfer viscous specimens from the primary sample container to the Unyvero Sample Tube.

The Unyvero System consists of the following components:

  • Unyvero Lysator: The Lysator lyses the specimen and can process up to four specimens simultaneously in four separate slots.
  • Unyvero Analyzer: The Analyzer automates DNA purification, amplification and detection. Each Analyzer can simultaneously process up to two Unyvero Cartridges with each slot available using random access.
  • Unyvero Cockpit: The Cockpit provides the main user interface for the Unyvero System, guides the user through the steps to run the Unvvero LRT Application and automatically generates and displays test results. The Cockpit is equipped with a high-resolution touch screen and a barcode reader.
  • Unyvero Sample Tube Holder: The Sample Tube holder holds the Sample Tube securely while the specimen is transferred into the Sample Tube.
AI/ML Overview

Here's a breakdown of the acceptance criteria and study proving the device meets them, based on the provided text:

Device: Unyvero Lower Respiratory Tract (LRT) Application and Unyvero System (Qualitative nucleic acid amplification assay)

Purpose: Simultaneous detection and identification of nucleic acid sequences from specified microorganisms and antibiotic resistance markers in endotracheal aspirates from adult hospitalized patients with suspected lower respiratory tract infections.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in a quantitative table format (e.g., "PPA must be >90%"). However, it describes the performance observed in the validation studies and then concludes that this performance is "acceptable" in conjunction with various limitations and recommended practices. The implicit acceptance criteria appear to be the demonstrated performance characteristics from the clinical studies.

Below is a table summarizing the reported device performance for organism detection (compared to a composite comparator, which is generally the most stringent comparison) and resistance marker detection (with software masking). The document doesn't provide a single, overarching table of "acceptance criteria" but rather details performance metrics across various studies.

Table of Reported Device Performance (Key Clinical Metrics)

Target (Type)Performance Metric (vs. Composite Comparator)Value (95% CI)Notes
Microorganism Targets(Prospective Study, comparison to Composite Comparator unless otherwise noted)
Acinetobacter spp.PPA95.8% (79.8 - 99.3%)Also 100% PPA vs. Culture for numerous, moderate, abundant categories; 37.0% PPV vs. Culture (likely due to DNA vs. viable organism)
Chlamydia pneumoniaePPAna (0/0 positive cases)100.0% NPA (99.4 - 100.0%) for 603 cases
Citrobacter freundiiPPA16.7% (3.0 - 56.3%)Very low PPA in prospective study; complemented by 86.0% PPA in contrived study.
Enterobacter cloacae complexPPA94.4% (74.2 - 99.0%)
Escherichia coliPPA97.4% (86.5 - 99.5%)
Haemophilus influenzaePPA88.2% (65.7 - 96.7%)
Klebsiella oxytocaPPA77.8% (45.3 - 93.7%)Complemented by 89.3% PPA in contrived study.
Klebsiella pneumoniaePPA90.9% (76.4 - 96.9%)
Klebsiella variicolaPPA100.0% (34.2 - 100.0%)Low number of positive cases (2/2 prospectively); complemented by 96.4% PPA in contrived study.
Legionella pneumophilaPPA100.0% (34.2 - 100.0%)Low number of positive cases (2/2 prospectively); complemented by 100.0% PPA in contrived study.
Moraxella catarrhalisPPA52.2% (33.0 - 70.8%)Notably low PPA in prospective study; 96.0% PPA in contrived study. High FP rate against culture.
Morganella morganiiPPA85.7% (48.7 - 97.4%)Complemented by 95.6% PPA in contrived study.
Mycoplasma pneumoniaePPA100.0% (34.2 - 100.0%)Low number of positive cases (2/2 prospectively); complemented by 100.0% PPA in contrived study.
Proteus spp.PPA96.0% (80.5 - 99.3%)
Pseudomonas aeruginosaPPA89.4% (81.1 - 94.3%)
Serratia marcescensPPA87.5% (69.0 - 95.7%)
Staphylococcus aureusPPA91.6% (85.2 - 95.4%)
Stenotrophomonas maltophiliaPPA89.3% (78.5 - 95.0%)
Streptococcus pneumoniaePPA62.5% (38.6 - 81.5%)Notably low PPA in prospective study.
Antibiotic Resistance Markers(Prospective Study, comparison to molecular comparator with software masking applied for host microorganism)
ctx-MPPA93.8% (71.7 - 98.9%)
kpcPPA100.0% (61.0 - 100.0%)
ndmPPAna (0/0 positive cases)
oxa-23PPA85.7% (48.7 - 97.4%)
oxa-24PPA100.0% (34.2 - 100.0%)Low number of positive cases (2/2)
oxa-48PPAna (0/0 positive cases)
oxa-58PPAna (0/0 positive cases)
temPPA100.0% (67.6 - 100.0%)This is for tem detected when H. influenzae is also detected. Overall tem PPA (without masking) was 100.0% (93.4 - 100.0%) for 54/54 cases.
vimPPA100.0% (34.2 - 100.0%)Low number of positive cases (2/2)
mecAPPA91.5% (81.6 - 96.3%)This is for mecA detected when S. aureus is also detected. Overall mecA PPA (without masking) was 87.1% (80.1 - 91.9%) for 108/124 cases.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set (Clinical Studies):
    • Prospective Study: 603 evaluable tracheal aspirate specimens.
      • Provenance: Multi-center study at nine clinical sites in the United States.
      • Nature: Prospectively collected.
    • Retrospective (Archived) Study: 369 evaluable archived specimens (211 from US study sites, 158 from other US or European sites tested in-house at Curetis).
      • Provenance: US and European sites.
      • Nature: Previously frozen, retrospective.
    • Contrived Clinical Study: Ranged from 21 to 50 specimens for each microorganism or resistance marker.
      • Provenance: Prepared and tested at three sites in the United States and in-house at Curetis.
      • Nature: Contrived specimens (spiked into natural tracheal aspirate matrices).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts

The document does not specify the number or qualifications of experts involved in establishing the ground truth for the clinical studies.

  • For "typical" microorganisms in the prospective study, the primary ground truth was standard-of-care (SoC) tracheal aspirate culture. For the composite comparator, it also included independent and validated multiplexed PCR assays with bi-directional sequencing.
  • For "atypical" microorganisms, the ground truth was two independent and validated multiplex PCR assays with bi-directional sequencing.
  • For antibiotic resistance markers, the ground truth was validated multiplex PCR assays followed by bi-directional sequencing.
  • For comparative phenotypic analysis of mecA, phenotypic antimicrobial susceptibility (AST) testing results from established culture methods were used.

The validation of these comparator methods (LoDs, inclusivity) is mentioned, implying expert oversight, but specific expert count or detailed qualifications are not provided.

4. Adjudication Method for the Test Set

The document describes the "Composite Comparator" method as an adjudication of sorts:

  • For 'typical' microorganisms: A specimen was considered positive if culture was positive OR if the validated comparator PCR and follow-up bi-directional sequencing was positive. Any specimen negative by both culture and PCR was considered negative. This is a form of "consensus" or "best clinical practice" ground truth, where molecular confirmation is used to bolster culture results, or establish positivity where culture might miss (e.g., non-viable organisms, difficult to culture).
  • For 'atypical' microorganisms: Specimens positive by either of the two PCR/sequencing assays were considered positive. Those negative by both were considered negative. This implies a "molecular consensus" approach.
  • For 'false positives' against culture: False positive LRT results were analyzed with molecular assays (PCR/bi-directional sequencing) using sample DNA extracts to confirm presence or absence of microorganisms. This acts as a re-adjudication/confirmation step for discordant results.

There is no mention of human expert adjudication panels in the way it might be done for image-based AI devices (e.g., 2+1 radiologist review). The adjudication is inherently built into the definition of the reference/comparator method, often relying on molecular methods to resolve discrepancies or provide more sensitive ground truth than traditional culture.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done for this device. This device is a diagnostic assay for molecular detection, not an AI-assisted diagnostic aid for human readers/interpreters in the typical MRMC study design (e.g., radiology AI). The study proves the standalone performance of the device.

6. If a Standalone (algorithm only without human-in-the-loop performance) was done

Yes, the study primarily evaluates the standalone performance of the Unyvero LRT Application. The reported PPA and NPA values for specific targets are measures of the device's accuracy in detecting nucleic acid sequences in the specimens, independent of human interpretation of the device's raw output. The device itself (Unyvero System with LRT Application) automates DNA purification, amplification, and detection, and automatically generates and displays results. The human involvement is in sample preparation and loading, and interpreting the final qualitative "positive/negative" result the device provides, rather than interpreting raw data alongside or without AI assistance.

7. The Type of Ground Truth Used

The ground truth used varied depending on the target:

  • For 'typical' microorganisms:
    • Primary Reference: Standard-of-care (SoC) tracheal aspirate culture.
    • Composite Comparator: A combination of SoC culture PLUS independent and validated multiplexed PCR assays with bi-directional sequencing. This represents a more sensitive and comprehensive "true presence" ground truth, especially for organisms that may be difficult to culture or present as non-viable DNA.
  • For 'atypical' microorganisms: Two independent and validated multiplex PCR assays with bi-directional sequencing.
  • For antibiotic resistance markers: Validated multiplex PCR assays followed by bi-directional sequencing.
  • For genotypic linkage and phenotypic correlation: Molecular sequencing of cultured isolates for the presence of the resistance marker gene, and phenotypic antimicrobial susceptibility testing (AST) results from those isolates.

8. The Sample Size for the Training Set

The document does not explicitly mention a "training set" in the context of machine learning or AI models. Given that this is a PCR-based diagnostic assay, it's developed through traditional molecular assay design and optimization, not typically through supervised machine learning with distinct training and test sets as seen in AI imaging. The "development" or "optimization" of the assay (e.g., primer design, LoD determination, inclusivity/exclusivity testing) would constitute its "training" or optimization phase, but no sample size for this phase is provided. The data presented relates to the validation of the final assay.

9. How the Ground Truth for the Training Set Was Established

As noted above, there's no explicitly defined "training set" in the machine learning sense. The ground truth for the analytical studies and clinical validation studies (which would inform refinement if developmental iterations were considered "training") was established as follows:

  • Analytical Studies (LoD, Reproducibility, Inclusivity, Exclusivity, Interference): Samples were contrived by spiking known concentrations of well-characterized reference strains (ATCC, NCTC, DSM, JMI, Micromyx, NRZ, UCLA, RKI, clinical isolates) into artificial respiratory matrix (ARM) or PBS. The "ground truth" here is the known identity and concentration of the spiked microorganisms/resistance markers.
  • Clinical Studies (Prospective and Retrospective): As detailed in point 7, ground truth was established by:
    • Standard-of-care (SoC) tracheal aspirate culture.
    • Validated multiplex PCR assays with bi-directional sequencing.
    • A composite of SoC culture and PCR/sequencing.
    • Molecular sequencing of isolates and phenotypic AST.

This means the "ground truth" was established by conventional microbiological and molecular diagnostic methods, which were considered the gold standard for comparing the performance of the new Unyvero system.

{0}------------------------------------------------

EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR Unyvero LRT Application and Unyvero System DECISION SUMMARY

A. DEN Number:

DEN170047

  • B. Purpose for Submission: De Novo request for evaluation of automatic class III designation for the Unyvero Lower Respiratory Tract (LRT) Application and the Unyvero System.
  • C. Measurand: DNA sequences of the following organisms and antibiotic resistance markers:
Microorganism TargetsAntibiotic Resistance Marker Targets
Acinetobacter spp.ctx-M ( bla CTX-M, subgroup 1 only)
Chlamydia pneumoniaekpc ( bla KPC)
Citrobacter freundiimecA
Escherichia colindm ( bla NDM)
Enterobacter cloacae complexoxa-23 ( bla OXA-23)
Haemophilus influenzaeoxa-24 ( bla OXA-24)
Klebsiella oxytocaoxa-48 ( bla OXA-48)
Klebsiella pneumoniaeoxa-58 ( bla OXA-58)
Klebsiella variicolatem ( bla TEM)
Legionella pneumophilavim ( bla VIM)
Moraxella catarrhalis
Morganella morganii
Mycoplasma pneumoniae
Proteus spp.
Pseudomonas aeruginosa
Serratia marcescens
Staphylococcus aureus
Stenotrophomonas maltophilia
Streptococcus pneumoniae

D. Type of Test:

Qualitative nucleic acid amplification assay

E. Applicant:

Curetis GmbH

F. Proprietary and Established Names:

Unyvero Lower Respiratory Tract (LRT) Application Unyvero System

G. Regulatory Information:

{1}------------------------------------------------

1. Regulation section:

21 CFR 866.3985

    1. Classification:
      Class II (Special Controls)
    1. Product code:
      QBH
    1. Panel:
      83-Microbiology

H. Indications for use:

    1. Indications for use:
      The Unyvero LRT Application is a qualitative nucleic acid multiplex test intended for the simultaneous detection and identification of nucleic acid sequences from the following microorganisms and antibiotic resistance markers in endotracheal aspirates from adult hospitalized patients with suspected lower respiratory tract infections.
MicroorganismAssociated antibiotic resistance marker
Acinetobacter spp. actx-Mb, kpc, ndm, oxa-23, oxa-24, oxa-58, vim
Chlamydia pneumoniae-
Citrobacter freundiictx-Mb, kpc, ndm, oxa-48, vim
Enterobacter cloacae complex cctx-Mb, kpc, ndm, oxa-48, vim
Escherichia colictx-Mb, kpc, ndm, oxa-48, vim
Haemophilus influenzaetem
Klebsiella oxytocactx-Mb, kpc, ndm, oxa-48, vim
Klebsiella pneumoniae dctx-Mb, kpc, ndm, oxa-48, vim
Klebsiella variicolactx-Mb, kpc, ndm, oxa-48, vim
Legionella pneumophila-
Moraxella catarrhalis-
Morganella morganiictx-Mb, kpc, ndm, oxa-48, vim
Mycoplasma pneumoniae-
Proteus spp. ectx-Mb, kpc, ndm, oxa-48, vim
Pseudomonas aeruginosactx-Mb, kpc, ndm, vim
Serratia marcescensctx-Mb, kpc, ndm, oxa-48, vim
Staphylococcus aureusmecA
Stenotrophomonas maltophilia-
Streptococcus pneumoniae-

ª Acinetobacter spp. includes: A. baumannii, A. calcoaceticus, A. haemolyticus, A. junii, A.

{2}------------------------------------------------

lwoffii, A. nosocomialis, A. parvus A. pittii, (detected by LRT Application) and A. ursingii (not detected by LRT Application). b ctx-M1 subgroup. & Enterobacter cloacae complex includes: E. asburiae, E. cloacae, E. hormaechei, E. kobei, E. ludwigii, and E. xiangfangensis. d Klebsiella pneumoniae includes two variants: K. pneumoniae (variant 1), and K. quasipneumoniae (variant 2) e Proteus spp. includes P. hauseri, P. mirabilis, P. penneri and P. vulgaris

The Unyvero LRT Application performed on the Unyvero System is indicated as an aid in the diagnosis of lower respiratory tract infection in adult hospitalized patients with signs and symptoms of lower respiratory infection: results should be used in conjunction with other clinical and laboratory findings. As tracheal aspirates commonly contain colonizing microorganisms, detection of Unyvero LRT microbial targets does not indicate that the microorganism is the cause of the disease. Unyvero positive results do not rule out co-infection with microorganisms not detected by the Unyvero LRT Application. Negative results do not preclude lower respiratory infection, as the causative agent may be a microorganism not detected by this test.

A negative result for any antibiotic resistance marker does not indicate that detected microorganisms are susceptible to applicable antimicrobial agents. Detected resistance markers cannot be definitively linked to specific microorganisms, and may be present in organisms that are not detected by the Unyvero LRT Application such as organisms present as colonizing or normal flora.

Microbiology cultures of aspirates should be performed to obtain isolates for species identification and antimicrobial susceptibility testing, to differentiate quantities of identified microorganisms as well as normal flora present in the specimen and to identify potential microorganisms not targeted by the Unyvero LRT Application.

    1. Special conditions for use statement(s):
    • . For in vitro diagnostic use only
    • . For prescription use only

Limitations:

  • . Antimicrobial resistance can occur via mechanisms other than the resistance markers detected by the Unyvero LRT Assay. Negative results for the LRT resistance markers do not indicate antimicrobial susceptibility of detected organisms.
  • . Detected resistance markers cannot be definitively linked to specific microorganism(s), and may be present in organisms that are not detected by the Unyvero LRT assay, including colonizing flora. Standard culture of tracheal aspirates and subsequent testing of the isolated microorganism(s) is necessary to definitively link antimicrobial resistance with a specific microorganism.
  • Detection of antibiotic resistance markers in a specimen may not correlate with phenotypic gene expression. Resistance marker results should be used in

{3}------------------------------------------------

conjunction with final culture and antimicrobial susceptibility testing results.

  • . Detected microorganisms may be indicative of colonizing or normal respiratory flora and may not be the causative agent of pneumonia.
    (b) (4)

  • The LRT Application is a qualitative test and does not provide a quantitative ● value for the microorganisms or antibiotic resistance markers detected in the specimen. Compared to standard of care culture methods, the Unyvero LRT Application is not able to assess the presence or amount of normal flora in the specimen or whether the detected microorganism/antibiotic resistance marker is present in predominating amounts as compared to other microorganisms in the specimen. The LRT Application also does not assess if detected microorganisms are present at clinically relevant concentrations, or are present as colonizing flora.

  • . When three or more microorganisms are detected for a specimen, the user is advised to wait for microbiology culture results to verify the predominant microorganism(s) and to assess for the presence of normal respiratory flora.

  • Bacterial nucleic acids may be present in the specimen independent of organism ● viability. Detection of organism nucleic acid target(s) does not imply that the corresponding organisms are the causative agents for clinical symptoms.

  • . A negative result for the 'atypical' microorganisms (C. pneumoniae, L. pneumophila, and M. pneumoniae) does not exclude the presence of this microorganism in the patient specimen. A positive result should be evaluated in the overall context of the patient's clinical condition and other laboratory results being part of the standard-of-care routine.

  • . Due to the small number of positive prospective and archived specimens for certain microorganisms and antibiotic resistance markers, performance characteristics for C. pneumoniae, C. freundii, K. oxytoca, K. variicola, L. pneumophila, M. catarrhalis, M. morganii, M. pneumoniae, and antibiotic resistance markers ctx-M, kpc, ndm, oxa-23, oxa-24, oxa-48, oxa-58, and vim were established primarily using contrived clinical specimens.

  • . Based on in-silico analysis and inclusivity wet testing, certain clinically relevant species for some LRT panel microorganism analytes are not detected or detected with reduced sensitivity: Acinetobacter ursingii detection is predicted at reduced sensitivity for the Acinetobacter spp. assay, Enterobacter asburiae detection is predicted at reduced sensitivity for the Enterobacter cloacae complex assay.

  • Subtypes included in ctx-M subgroups ctx-M2, ctx-M8, ctx-M9 (ctx-M14), ctx-. M25 and ctx-M45 were not evaluated and are not predicted to be detected by the LRT assay based on in-silico analysis.

  • . The mecC variant is not detected by Unyvero LRT and for mecC positive Staphylococcus aureus strains, the assay will generate negative results for mecA.

  • . Based on in-silico analysis and exclusivity wet testing. the following LRT panel microorganism assays are expected to cross-react with closely related clinically relevant species: Citrobacter freundii (cross-reactive to C. braakii, Kluyvera georgiana), Escherichia coli (cross-reactive to E. albertii, E. fergusonii and

{4}------------------------------------------------

Shigella spp. (S. dysenteriae, S. sonnei, S. boydii)), Haemophilus influenzae (cross-reactive to H. haemolyticus), Klebsiella oxytoca (cross-reactive to K. michiganensis).

4. Special instrument requirements:

The Unyvero LRT Application is performed on the Unyvero System which includes the Unyvero Lysator, Unyvero Analyzer and Unyvero Cockpit.

I. Device Description:

The Unyvero LRT Application is a qualitative test that includes specimen processing, genomic bacterial DNA isolation and purification, multiplex PCR and array hybridization and detection. The Unyvero LRT Application performed using the Unyvero System detects specific nucleic acid sequences from microorganisms and resistance markers in tracheal aspirates collected from patients with signs and symptoms of lower respiratory infection.

The Unyvero LRT Application consists of the following components:

  • Unyvero LRT Cartridge: Contains DNA isolation and purification reagents, a DNA . isolation column, eight separate PCR chambers with eight corresponding detection arrays. The Cartridge also contains fluorescently-labeled primers, hybridization and wash buffers and oligonucleotide probes for detection of targeted PCR products using array hybridization technology.
  • Unyvero T1 Sample Tube: Contains glass beads and buffers to lyse bacteria and liquefy the sample.
  • . Unyvero T1 Sample Tube Cap (with Internal Control): Contains proteinase K and a synthetic internal control gene for process monitoring. The T1 Sample Tube Cap seals the Unyvero Sample Tube after which the internal control is combined with each patient specimen. The internal control DNA sequence does not have significant homology to targeted sequences and is amplified independently in each of the eight PCR chambers and the amplified internal control product is hybridized on each array.
  • . Unyvero M1 Master Mix: Contains reagents for DNA amplification.
  • . Unyvero T1 Transfer Tool: The Transfer tool can be used to transfer viscous specimens from the primary sample container to the Unyvero Sample Tube.

The Unyvero System consists of the following components:

  • . Unyvero Lysator: The Lysator lyses the specimen and can process up to four specimens simultaneously in four separate slots.

{5}------------------------------------------------

  • Unyvero Analyzer: The Analyzer automates DNA purification, amplification and detection. Each Analyzer can simultaneously process up to two Unyvero Cartridges with each slot available using random access.
  • Unyvero Cockpit: The Cockpit provides the main user interface for the Unyvero System, guides the user through the steps to run the Unvvero LRT Application and automatically generates and displays test results. The Cockpit is equipped with a high-resolution touch screen and a barcode reader.
  • Unyvero Sample Tube Holder: The Sample Tube holder holds the Sample Tube securely while the specimen is transferred into the Sample Tube.

Other materials required but not provided:

  • Pipette capable of dispensing 180uL
  • DNase/RNase free, aerosol resistant pipette tips
  • 1mL Luer-Lock svringe ●

J. Standard/Guidance Document Referenced:

  • CLSI EP07-A2. Interference Testing in Clinical Chemistry: Approved Guideline, 2nd ● edition.
  • CLSI EP17-A2. Evaluation of Detection Capability for Clinical Laboratory ● Measurement Procedures; Approved Guideline - 2nd edition.
  • CLSI M100, Performance Standards for Antimicrobial Susceptibility Testing, 26th ● Edition.

K. Test Principle:

The Unyvero LRT Application is a qualitative PCR-based assay that detects DNA sequences of microorganisms and associated resistance markers in tracheal aspirate specimens. The assay includes specimen processing (lysis), DNA extraction and isolation, multiplex PCR with eight parallel multiplex endpoint PCR reactions, and qualitative detection of amplified DNA products using hybridization arrays.

A tracheal aspirate specimen is first pipetted into the Unyvero Sample Tube and closed with the Unyvero Sample Tube Cap. Closing the sample tube automatically adds the lysis reagent and the internal control gene template to the specimen. The sample tube which fits into the Unyvero Lysator only if closed with the Unyvero sample tube cap is then placed on the Lysator. After the specimen is lysed in the Lysator, the Sample Tube and Master Mix are loaded into the Unyvero LRT Cartridge which is then inserted into the position assigned by the Unyvero Analyzer for automated processing and analysis.

In the Unyvero LRT Cartridge, the remainder of the testing steps are automated by the Unyvero Analyzer. The lysed specimen is further processed and then transferred onto a

{6}------------------------------------------------

(b) DNA purification column for nucleic acid (b) (4) DNA is transferred to a chamber, where mixing with the Master Mix takes place. This mixture is distributed into eight separate PCR reaction chambers each containing multiple primer pairs. After amplification, PCR products are hybridized to the corresponding array probes. Each array has been manufactured (b) (4) (b) (4)

(b) (4)

Result data are transferred

to the Unyvero Cockpit for visualization and result printout. A test run is completed after approximately 4.5 hours, and results for panel microorganisms and associated antibiotic resistance markers are displayed on the Unyvero Cockpit screen.

L. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Reproducibility:

The reproducibility of the Unyvero LRT Application was assessed using a representative panel of samples prepared in artificial respiratory matrix (ARM, adapted from Dinesh S. D.1) which was used as a surrogate for tracheal aspirate matrix. In a separate analytical study, the Unyvero LRT Application was shown to demonstrate equivalent performance for samples prepared in ARM matrix and samples prepared in natural tracheal aspirate matrix (See Section L.2.b below).

Samples for the reproducibility study were prepared with whole-organism preparations for the following LRT panel analytes: E. coli/tem. P. geruginosa, K. pneumoniae, M. morganii, S. maltophilia and S. aureus/mecA. Sample #1 (REPRO 1) was inoculated with each organism at moderate positive concentrations (~5 x LoD) and Sample #2 (REPRO 2) was inoculated with each organism at low positive concentrations (~1.7 x LoD). For the mec A target, the REPRO 1 and REPRO2 panel members had concentrations of 12.5 x LoD and 4.2 x LoD respectively. For the tem target, the REPRO1 and REPRO2 panel members had concentrations of 6.2 and 2.1 x LoD respectively. Test runs were independently performed by three different operators. Each operator was assigned one of three Unyvero systems, each consisting of 1 Unyvero Cockpit, 2 Unyvero Lysators and at least 4 Unyvero Analyzers. A total of 90 replicate samples were tested for each panel member with each operator testing samples in triplicate on a minimum of five testing days. Positive and negative controls were also tested daily by each operator. The organisms and concentrations used to prepare study samples are presented in Table 1 below. Study results are presented in Table 2.

Table 1: Microorganisms for Reproducibility Study ATCC Strain | LoD (CFU/mL) | Repro1/x-fold LoD | Repro2/x-fold LoD Analytes

1 Dinesh, Artificial Sputum Medium. Protocol Exchange (2010)

{7}------------------------------------------------

E. coli352187.50E+043.75E+055.01.28E+051.7
K. pneumoniae V27006035.00E+042.50E+055.08.50E+041.7
M. morganii258305.00E+052.50E+065.08.50E+051.7
P. aeruginosa101455.00E+042.50E+055.08.50E+041.7
S. aureus335915.00E+062.50E+075.08.50E+061.7
S. maltophilia136373.00E+041.50E+055.05.10E+041.7
mec A335912.00E+062.50E+0712.58.50E+064.3
tem352186.00E+043.75E+056.31.28E+052.1

Table 2: Reproducibility study results, stratified by target, concentration and operator

REPRO1 (5x LoD)REPRO 2 (1.7x LoD)
analyteuserx-foldLoDPos.[%]95 % CIx-foldLoDPos.[%]95 % CI
E. coliATCC 35218user 1530/30100.088.7 - 100.01.729/29100.088.3 - 100.0
user 230/30100.088.7 - 100.029/3096.783.3 - 99.4
user 329/3096.783.3 - 99.429/3096.783.3 - 99.4
all89/9098.994.0 - 99.887/8997.892.2 - 99.4
K. pneumoniaevariant 2ATCC 700603user 1521/2972.454.3 - 85.31.714/3046.730.2 - 63.9
user 225/2986.269.4 - 94.515/3050.033.2 - 66.8
user 324/3080.062.7 - 90.519/2965.547.3 - 80.1
all70/8879.570.0 - 86.748/8953.943.6 - 63.9
mecA(from: S. aureusATCC 33591)user 112.530/30100.088.7 - 100.04.225/2986.269.4 - 94.5
user 227/3090.074.4 - 96.529/3096.783.3 - 99.4
user 329/3096.783.3 - 99.429/3096.783.3 - 99.4
all86/9095.689.1 - 98.383/8993.386.1 - 96.9
M. morganiiATCC 25830user 1528/2996.682.8 - 99.41.7a21/3070.052.1 - 83.3
user 227/2993.178.0 - 98.123/3076.759.1 - 88.2
user 328/3093.378.7 - 98.224/2982.865.5 - 92.4
all83/8894.387.4 - 97.568/8976.466.6 - 84.0
P. aeruginosaATCC 10145user 1530/30100.088.7 - 100.01.730/30100.088.7 - 100.0
user 230/30100.088.7 - 100.028/3093.378.7 - 98.2
user 330/30100.088.7 - 100.030/30100.088.7 - 100.0
all90/90100.095.9 - 100.088/9097.892.3 - 99.4
S. aureusATCC 33591user 1528/3093.378.7 - 98.21.730/30100.088.7 - 100.0
user 230/30100.088.7 - 100.026/3086.770.3 - 94.7
user 330/30100.088.7 - 100.029/3096.783.3 - 99.4
all88/9097.892.3 - 99.485/9094.487.6 - 97.6
S. maltophiliaATCC 13637user 1530/30100.088.7 - 100.01.729/3096.783.3 - 99.4
user 228/3093.378.7 - 98.228/3093.378.7 - 98.2
user 330/30100.088.7 - 100.029/3096.783.3 - 99.4
all88/9097.892.3 - 99.486/9095.689.1 - 98.3
tem(from: E. coliATCC 35218)user 16.229/3096.783.3 - 99.42.129/29100.088.3 - 100.0
user 229/3096.783.3 - 99.422/3073.355.6 - 85.8
user 330/30100.088.7 - 100.026/3086.770.3 - 94.7
all88/9097.892.3 - 99.477/8986.577.9 - 92.1

ª M. morganii was also evaluated at 0.6x LoD in PBS matrix in an initial reproducibility study. Results for this panel member were acceptable based on the microorganism concentration tested (e.g., positivity: 77/88, 87.5% detection for samples prepared at below LoD concentrations).

{8}------------------------------------------------

Testing of reproducibility samples generated acceptable results for all analytes except for K. pneumoniae and M. morganii for which positivity rates were lower than expected for both the 5x LoD and 1.7x LoD panel members. The reason for these unexpected results was likely due to inaccurate quantitation of organism stocks used for sample preparation. It was noted that testing by different operators/sites generated equivalent results for these analytes (i.e., similarly lower than expected percent positivity at each testing site).

To further evaluate the reproducibility of the Unyvero LRT assay for detection of these two organism targets, ten additional sample replicates for both K. pneumoniae and M. morganii were prepared using freshly grown and quantitated organism suspensions with final sample concentrations of 5x LoD. For both microorganisms, a positivity rate of 90% was observed with 9/10 samples generating positive results.

Out of 14 positive quality control runs with samples containing moderate positive analyte concentrations (~8x LoD). one quality control sample was false negative for K. pneumoniae and one sample was false negative for M. morganii.

Three false positive results were observed during the study; two false positive results in reproducibility test sample runs (S. pneumoniae, vim) and one false positive in a positive control run (M. catarrhalis).

Partially invalid results (one PCR chamber only) were observed in 8/180 (4.4%) reproducibility test runs, 2/14 (1.4%) negative quality control runs, and 0/14 positive quality control runs.

Results from the reproducibility study were further evaluated based on Unyvero signal output levels (RBU) for each target and shown in Table 3 below.

{9}------------------------------------------------

analyteuserx-foldLoDAvgSignal% CVRangex-foldLoDAvgSignal% CVRange
user 15(b) (4)b) (4)(b) (4)1.7(b) (4)(b) (4)(b) (4)
E. coliuser 2
ATCC 35218user 3
all
K. pneumoniaeuser 15(b) (4)(b) (4)1.7b(b) (4)(b) (4)
variant 2user 2
ATCC 700603user 3
all
mecAuser 112.5(b) (4)(b) (4)4.2(b) (4)(b) (4)
(from:user 2
S. aureususer 3
ATCC 33591)all
user 15(b) (4)(b) (4)1.7(b) (4)(b) (4)
M. morganiiuser 2
ATCC 25830user 3
all
user 15(b) (4)(b) (4)1.7(b) (4)(b) (4)
P. aeruginosauser 2
ATCC 10145user 3
all
user 15(b) (4)(b) (4)1.7(b) (4)(b) (4)
S. aureususer 2
ATCC 33591user 3
all
user 15(b) (4)(b) (4)1.7(b) (4)(b) (4)
S. maltophiliauser 2
ATCC 13637user 3
all
temuser 16.2(b) (4)(b) (4)2.1(b) (4)(b) (4)
(from: E. coliuser 2
ATCC 35218)user 3
all

Table 3: Reproducibility study, Quantitative analysis of Unyvero signal levels

  • b. Linearity/assay reportable range:
    Not applicable.

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):
    Internal Control: An internal control consisting of a synthetic DNA sequence without homology to LRT microorganism or resistance marker target sequences is separately processed in each of the eight Unyvero LRT Cartridge PCR chambers. The internal control allows for assessment of DNA purification, amplification, array hybridization, and detection steps. The internal control monitors for the presence of PCR inhibitors

{10}------------------------------------------------

in the specimens and enables the system to detect any failures in the testing process that could potentially result in an incorrect test result. Results generated from Individual PCR chambers are considered valid if successful amplification and detection is completed for either the internal control or any microorganism/resistance marker target in the multiplex PCR reaction.

External Controls

Positive and negative controls were run daily at each of the testing sites during the clinical study. Testing included (b) (4)

External controls are not provided with the Unyvero LRT Application: however. testing of external positive and negative controls are recommended in the assay labeling. Controls may consist of previously characterized positive samples or negative samples spiked with well characterized microorganisms. Previously characterized negative samples may be used as negative controls. External controls should be used in accordance with local, state, and/or federal regulations, accreditation requirements and individual laboratory's quality control policies, as applicable.

Specimen Stability

The Unyvero LRT instructions for use indicates that it is acceptable to store tracheal aspirate specimens at 2-8°C for up to 24 hours prior to testing. To evaluate effects on assay performance for 24 hour refrigerated storage prior to starting an LRT test, an analysis was performed for prospectively tested clinical tracheal aspirate specimens that were tested with the LRT application at various times of storage up to 24 hours. Both qualitative performance (compared to culture) and semi-quantitative Unyvero LRT signal levels for positive analytes were assessed between groups stored for different time periods. Results for specimen which were stored for times close to the 24-hour sample storage time limit did not show a significant difference in clinical performance or a significant difference in assay signal as compared to specimens that were tested immediately after collection.

The sample storage recommendations for the LRT application are consistent with those for traditional culture2 which include storage and transport of tracheal aspirate specimens at 2-8°C for up to 24 hours. The analysis of the clinical specimen results together with the same recommendation for tracheal aspirate culture support the claim

2 Baron, E. J. et al. A guide to utilization of the microbiology laboratory for diagnosis of infectious diseases: 2013 recommendations by the Infectious Diseases Society of America (IDSA) and the American Society for Microbiology (ASM)(a). Clin. Infect. Dis. Off. Publ. Infect. Dis. Soc. An. 57, (2013).

{11}------------------------------------------------

of refrigerated storage of tracheal aspirates for up to 24 hours prior to testing with the Unyvero LRT Application.

d. Detection limit:

A study was conducted to determine the Limit of Detection (LoD) of the Unyvero LRT Application for each targeted analyte. Samples were prepared in Artificial Respiratory Matrix (ARM) which was used as a surrogate for tracheal aspirate matrix. Samples prepared in ARM matrix were shown to generate equivalent results to samples prepared in natural tracheal aspirate matrix (See Section L.2.b below).

For the LoD study, samples were inoculated with multiple organisms (i.e., multispiked). To mitigate potential for competitive inhibition, sample compositions were designed to ensure that only one or two analytes were amplified in each multiplex PCR chamber.

LoD estimates were initially determined by testing small numbers of replicates at multiple different organism concentrations. The LoD for each analyte was then confirmed by testing 20 replicates at the estimated LoD concentration. The final LoD is defined as the lowest concentration (reported as CFU/mL, IFU/mL, or CCU/mL) of sample for which ≥95% of sample replicates generate positive results. Tables 4 and 5 below summarize the confirmed LoDs for each LRT panel microorganism and antibiotic resistance marker.

{12}------------------------------------------------

LRT Panel MicroorganismID of Tested ReferenceStrainLoD Concentration[CFU/mL, or other as indicated infootnote]
Acinetobacter spp.ATCC 19606 ( A.baumannii )1x105
Chlamydia pneumoniae (in IFU/mL) aATCC VR-22821.5x104
Citrobacter freundiiATCC 80902x105
Escherichia coliATCC 117757.5x104
Enterobacter cloacae complexATCC 13047 ( E. cloacae )5x106
Haemophilus influenzaeATCC 333912x106
Klebsiella oxytocaATCC 131822x105
Klebsiella pneumoniae Variant 1ATCC 138834.5x105
Klebsiella pneumoniae Variant 2ATCC 7006035x104
Klebsiella variicolaATCC BAA-8301x105
Legionella pneumophilaATCC 331522x106
Moraxella catarrhalisATCC 252388x105
Morganella morganiiATCC 258305x105
Mycoplasma pneumoniae (in CCU/mL) bATCC 290851x105
Proteus mirabilisATCC 299061x105
Proteus vulgarisATCC 299056x105
Pseudomonas aeruginosaATCC 101455x104
Serratia marcescensATCC 138801x105
Staphylococcus aureusATCC 126005x106
Stenotrophomonas maltophiliaATCC 136373x104
Streptococcus pneumoniaeATCC 496195x105

Table 4: Limit of detection for LRT panel microorganisms

a IFU: inclusion-forming units

b CCU: color-changing units

LRT Panel Antibiotic Resistance MarkerID of Tested Reference StrainLoD Concentration [CFU/mL]
ctx-MNRZ-00751 ( K. pneumoniae )1x105
kpcNRZ-00281 ( E. coli )5x105
mecAATCC 33591 ( S. aureus )2x106
ndmJMI 50067 ( E. coli )5x104
oxa-23NCTC 13301 ( A. baumannii )2x107
oxa-24NCTC 13302 ( A. baumannii )5x105
oxa-48NCTC 13442 ( K. pneumoniae )2x106
oxa-58NCTC 13305 ( A. baumannii )7.5x105
temaNCTC 13351 ( E. coli )6x104
vimDSM-24600 ( P. aeruginosa )5x104

Table 5: Limit of detection for I RT panel antibiotic resistance markers

a Although the LRT Application reports tem only with H. influenzae, LoD was determined with an E. coli strain positive for tem. Note that inclusivity testing was successfully performed with two tem positive H. influenzae strains.

e. Analytical Reactivity (Inclusivity)

The inclusivity study of the LRT Application was assessed using samples inoculated at low positive concentrations with well-characterized bacterial isolates. Most strains were evaluated in samples prepared at <3x LoD. Testing was performed in duplicate

{13}------------------------------------------------

and if 100% detection was not observed during initial testing, additional samples prepared with higher organism concentrations were evaluated.

Analyte LoDs were established using ARM as a surrogate for aspirate matrix. Inclusivity testing was performed with contrived samples prepared in PBS. Analytical testing of PBS and ARM matrices demonstrated that the LoDs for the two matrices are comparable for most analytes; however, for a few analytes, PBS and ARM LoDs differed by a factor between 0.3-3. The differences in LoD between the two matrices for these few analytes was not considered significant and did not raise concerns that the stated reactivity would differ when testing clinical specimens.

Inclusivity/Microorganism Targets:

Inclusivity testing for samples prepared with microorganism concentrations at <3x LoD generated 100% detection for the LRT microorganism targets except for the following microorganisms for which one or more strains were not detected or were detected only at higher than LoD concentrations (also bolded in table below):

  • One of four strains of Proteus mirabilis (detected at 5x LoD and not detected at ● lower concentrations). Sequencing of the isolate revealed sequence mismatches in a detection probe.
  • . One of three strains of Proteus vulgaris (detected at 5x LoD and not detected at lower concentrations). Sequencing of the isolate revealed sequence mismatches in a detection probe.
  • . One of six strains of Moraxella catarrhalis (detected at 3.8x LoD and not detected at lower concentrations)
  • . One of seven strains of Serratia marcescens (1/2 replicates detected at 5x LoD and 0/2 replicates detected at 10x LoD)
  • Two of six strains of Stenotrophomonas maltophilia (<100% detection at all concentrations tested):
    • Strain DSM-50173 was positive in 3/4 replicates prepared at 1.7 x LoD । and 1/2 replicates prepared at 6.8x LoD; sequencing of this strain revealed multiple mismatches for one internal probe.
    • -Strain DSM-21874 was positive in 2/4 replicates prepared at 1.7 x LoD and 0/2 replicates at 6.8x LoD; sequencing of this strain did not reveal any mismatches to primers or probes. Testing was repeated for this strain at 1x LoD (0/2), 1.7x (0/2) and at 3.3x LoD for which 2/2 sample replicates generated the expected positive result.

The following limitation is included in the Unyvero LRT labeling:

  • . Based on in-silico analyses, some strains may either not be detected or be detected with reduced sensitivity due to variations in targeted sequences for the following microorganisms: C freundii. K. oxytoca. L. pneumophilia. M. morganii. Based on wet testing for Proteus spp., and S. marcescens and based

{14}------------------------------------------------

on in-silico/wet testing for M. catarrhalis and S. maltophilia some strains may be detected with reduced sensitivity due to variations in targeted sequences.

Results from inclusivity testing are shown in Table 6.

StrainStrain IDTest Conc.[CFU/mL]LoDfactora# Positive/# Tests
Acinetobacter baumanniiATCC 196068x1040.8x2/2
Acinetobacter baumanniiNCTC 133058x1040.8x2/2
Acinetobacter baumanniiNCTC 133018x1040.8x2/2
Acinetobacter baumanniiMicromyx 63348x1040.8x2/2
Acinetobacter baumanniiMicromyx 61538x1040.8x2/2
Acinetobacter baumanniiMicromyx 61491.5x1051.5x4/4
Acinetobacter baumanniiMicromyx 44108x1040.8x2/2
Acinetobacter baumanniiJMI 497551x1051x2/2
Acinetobacter baumanniiNRZ-004498x1040.8x2/2
Acinetobacter baumanniiNRZ-005188x1040.8x2/2
Acinetobacter baumanniiUCLA A48x1040.8x2/2
Acinetobacter baumanniiUCLA A58x1040.8x2/2
Acinetobacter calcoaceticusATCC 230558x1040.8x2/2
Acinetobacter lwoffiiATCC 153098x1040.8x2/2
Acinetobacter haemolyticusATCC 179068x1040.8x2/2
Chlamydia pneumoniaeATCC VR-13104.5x104 b[IFU/mL]3x2/2
Chlamydia pneumoniaeATCC VR-22824.5x104[IFU/mL]3x2/2
Chlamydia pneumoniaeATCC 535924.5x104[IFU/mL]3x2/2
Citrobacter freundiiATCC 80902x1051x6/6
Citrobacter freundiiATCC 438643x1051.5x2/2
Citrobacter freundiiNCTC 85812x1051x2/2
Citrobacter freundiiNCTC 97505x1052.5x2/2
Citrobacter freundiiNRZ-004525x1052.5x2/2
Citrobacter freundiiUCLA C12x1051x2/2
Citrobacter freundiiUCLA C56x1053x2/2
Enterobacter cloacaeATCC 130471.5x1073x2/2
Enterobacter cloacaeATCC 233553x1060.6x2/2
Enterobacter cloacaeATCC 491416x1050.1x2/2
Enterobacter cloacaeATCC BAA-24686x1050.1x2/2
Enterobacter cloacaeJMI 462391.5x1073x2/2
Enterobacter cloacaeNRZ-002395x1061x2/2
Enterobacter cloacae ssp. dissolvensATCC 233735x1061x2/2
Enterobacter hormaecheiATCC 491625x1061x2/2
StrainStrain IDTest Conc.[CFU/mL]LoDfactora# Positive/# Tests
Enterobacter asburiaeATCC 359531.5x1073x2/2
Escherichia coliATCC 117751x1051.3x2/2
Escherichia coliATCC 259221x1051.3x2/2
Escherichia coliATCC 352181x1051.3x4/4
Escherichia coliATCC BAA-25231x1051.3x2/2
Escherichia coliNCTC 133515x1040.7x2/2
Escherichia coliNCTC 134761x1051.3x2/2
Escherichia coliJMI 500676x1040.8x2/2
Escherichia coliNRZ-001761x1051.3x2/2
Escherichia coliNRZ-002226x1040.8x2/2
Escherichia coliNRZ-002811.8x1052.4x2/2
Haemophilus influenzae (serotype a)ATCC 90062x1061x2/2
Haemophilus influenzae (serotype c)ATCC 90076x1063x2/2
Haemophilus influenzae (serotype b)ATCC 102116x1063x2/2
Haemophilus influenzae (serotype b)ATCC 492476x1063x2/2
Haemophilus influenzae(non-typeable/non-capsulated)ATCC 333916x1063x2/2
Haemophilus influenzae (serotype b)ATCC 497666x1063x2/2
Haemophilus influenzae (serotype b)NCTC 84682x1061x2/2
Klebsiella oxytocaATCC 131828x1040.4x6/6
Klebsiella oxytocaATCC 438638x1040.4x2/2
Klebsiella oxytocaATCC 87241.6x1050.8x2/2
Klebsiella oxytocaATCC 491318x1040.4x2/2
Klebsiella oxytocaNCIMB 128194x1052x2/2
Klebsiella oxytocaNRZ-220608x1040.4x2/2
Klebsiella pneumoniaeATCC 138834x1050.9x4/4
Klebsiella pneumoniaeNCTC 134391x1062.2x2/2
Klebsiella pneumoniaeNCTC 134406x1051.3x2/2
Klebsiella pneumoniaeNCTC 134424x1050.9x2/2
Klebsiella pneumoniaeNCTC 134434x1050.9x2/2
Klebsiella pneumoniaeMicromyx 46538x1051.8x4/4
Klebsiella pneumoniaeMicromyx 46761x1062.2x2/2
Klebsiella pneumoniaeJMI 498316x1051.3x2/2
Klebsiella pneumoniaeJMI 497679x1052x2/2
Klebsiella pneumoniaeNRZ-000024x1050.9x2/2
Klebsiella pneumoniaeNRZ-001031.5x1063.3x2/2
Klebsiella pneumoniaeNRZ-002239x1052x4/4
Klebsiella pneumoniaeNRZ-002496x1051.3x2/2
Klebsiella pneumoniaeNRZ-004729x1052x2/2
Klebsiella pneumoniaeNRZ-007519x1052x2/2
Klebsiella pneumoniae variant II ( K. quasipneumoniae )ATCC 7006036x1051.3x2/2
StrainStrain IDTest Conc.[CFU/mL]LoDfactora# Positive/# Tests
Klebsiella variicolaATCC BAA-8303x1053x2/2
Klebsiella variicolaclinical strain 13.9x1053.9x2/2
Klebsiella variicolaclinical strain 22.6x1052.6x2/2
Klebsiella variicolaclinical strain 31.5x1051.5x2/2
Klebsiella variicolaclinical strain 42.4x1052.4x3/3
Klebsiella variicolaclinical strain 51.5x1051.5x2/2
Legionella pneumophila (serotype 1)ATCC 331528x1050.4x2/2
Legionella pneumophila (serotype 2)ATCC 331546x1063x2/2
Legionella pneumophila (serotype 3)ATCC 331554x1050.2x2/2
Legionella pneumophila (serotype 6)ATCC 332152x1061x2/2
Legionella pneumophila (serotype 8)ATCC 350964x1050.2x2/2
Legionella pneumophila (serotype 10)ATCC 432834x1050.2x2/2
Legionella pneumophilaUCLA L14x1050.2x2/2
Legionella pneumophilaUCLA L54x1050.2x2/2
Legionella pneumophilaUCLA L64x1050.2x2/2
Moraxella catarrhalisATCC 252382x1062.5x0/2
Moraxella catarrhalisATCC 436174x1050.5x2/2
Moraxella catarrhalisATCC 81764x1050.5x2/2
Moraxella catarrhalisATCC 252404x1050.5x2/2
Moraxella catarrhalisATCC 232464x1050.5x2/2
Moraxella catarrhalisATCC 491432x1062.5x2/2
Morganella morganiiATCC 80191x1050.2x2/2
Morganella morganiiATCC 258295x1051x2/2
Morganella morganiiATCC 258301x1050.2x2/2
Morganella morganii spp. siboniiATCC 499481.5x1063x2/2
Mycoplasma pneumoniaeATCC 290853x105 CCU/mLc3x2/2
Mycoplasma pneumoniaeATCC 293433x105 CCU/mLc3x2/2
Mycoplasma pneumoniaeATCC 154923x106 CFU/mL3x2/2
Mycoplasma pneumoniaeATCC 155313x106copies/mL3x2/2
Proteus mirabilisATCC 124534x1040.4x2/2
Proteus mirabilisATCC 141533x1053x2/2
Proteus mirabilisATCC 259334x1054x0/2
Proteus mirabilisATCC 259335x1055x d2/2
Proteus mirabilisATCC 299063x1053x4/4
Proteus vulgarisATCC 63802.4x1064x0/2
Proteus vulgarisATCC 63803x1065x d2/2
Proteus vulgarisATCC 84274x1040.1x2/2
Proteus vulgarisATCC 299051.8x1063x4/4
Proteus hauseriATCC 7008265x104e0.1x2/2
Proteus penneriATCC 335195x104e0.1x2/2
Pseudomonas aeruginosaATCC 101451x1052x2/2
StrainStrain IDTest Conc.[CFU/mL]LoDfactora# Positive/# Tests
Pseudomonas aeruginosaATCC 278531x1052x2/2
Pseudomonas aeruginosaDSM-246005x1041x4/4
Pseudomonas aeruginosaNCTC 134372x1040.4x4/4
Pseudomonas aeruginosaMicromyx 25621x1052x4/4
Pseudomonas aeruginosaNRZ-001962x1040.4x2/2
Pseudomonas aeruginosaNRZ-039615x1041x2/2
Pseudomonas aeruginosaUCLA P204x1040.8x2/2
Serratia marcescensATCC 81003x1053x2/2
Serratia marcescensATCC 138803x1053x2/2
Serratia marcescensATCC 147563x1053x2/2
Serratia marcescensATCC 153653x1053x2/2
Serratia marcescensATCC 27117 f2x1052x1/2
3x1053x1/2
5x1055x1/2
1x10610x0/2
Serratia marcescensATCC 438612x1052x2/2
Serratia marcescens ssp. sakuensisDSM-171743x1053x2/2
Staphylococcus aureusIDEXX VB9624558x1061.6x2/2
Staphylococcus aureusIDEXX VB99813538x1061.6x2/2
Staphylococcus aureusIDEXX VB9690398x1061.6x2/2
Staphylococcus aureusATCC BAA-23128x1061.6x2/2
Staphylococcus aureusNCTC 124936x1061.2x2/2
Staphylococcus aureusATCC 335916x1061.2x2/2
Staphylococcus aureusDSM-170916x1061.2x2/2
Staphylococcus aureusATCC 126008x1061.6x2/2
Staphylococcus aureusATCC 292138x1061.6x2/2
Staphylococcus aureusATCC 433001.5x1073x2/2
Staphylococcus aureusRKI 07-031658x1061.6x2/2
Staphylococcus aureusRKI 01-006948x1061.6x2/2
Staphylococcus aureusRKI 09-001878x1061.6x2/2
Staphylococcus aureusRKI 08-024928x1061.6x2/2
Stenotrophomonas maltophiliaATCC 136365x1041.7x2/2
2x1040.7x2/2
3x1041x1/2
Stenotrophomonas maltophiliaATCC 13637 g5x1041.7x5/6
1x1053.3x1/2
2x1056.7x1/2
Stenotrophomonas maltophiliaATCC 176662x1040.7x2/2
Stenotrophomonas maltophiliaATCC 491302x1040.7x2/2
Stenotrophomonas maltophiliaDSM-50173 g[ATCC 17444]2x1040.7x1/4
3x1041x0/1
5x1041.7x3/4
1x1052x1053.3x6.7x0/21/2
StrainStrain IDTest Conc.[CFU/mL]LoDfactora# Positive/# Tests
Stenotrophomonas maltophiliaDSM-21874 g[NCIMB 9528]2x1043x1045x1041x1052x1050.7x1x1.7x3.3x6.7x1/40/22/40/20/2
Streptococcus pneumoniaeATCC 334001.5x1062.5x1063x5x
Streptococcus pneumoniae(serotype 19F)ATCC 496191.5x1063x2/2
Streptococcus pneumoniae(serotype 3)ATCC 63031.5x1062.5x1063x5x1/22/2
Streptococcus pneumoniae(serotype 5)ATCC 63052x1050.4x2/2
Streptococcus pneumoniaeATCC 491501.5x1063x2/2
Streptococcus pneumoniae(serotype 1)ATCC 63012x1051x1060.4x2x1/22/2
Streptococcus pneumoniaeATCC 100152x1050.4x2/2
Streptococcus pneumoniae(serotype 2)ATCC 273361.5x1063x2/2
Streptococcus pneumoniae(serotype 9V)DSM-118652x1050.4x2/2
Streptococcus pneumoniae(serotype 23F)DSM-118661.5x1063x2/2
Streptococcus pneumoniae(serotype 6B)DSM-118671.5x1063x2/2

Table 6: Microorganism targets, Inclusivity strains

{15}------------------------------------------------

{16}------------------------------------------------

{17}------------------------------------------------

{18}------------------------------------------------

Analyte LoDs were established using "artificial respiratory medium" (ARM) as sample matrix surrogate for aspirates. Inclusivity testing was performed by contrived sample matrix. Test concentrations are given as multiples of claimed ARM LoDs. PBS and ARM LoDs are comparable for most analytes PBS and ARM LoDs differ by a factor between 0.3-3.

b Concentration of C. pneumoniae strains was determined as IFU (inclusion-forming units) / mL.

For Mycoplasma pneumoniae inclusivity testing different source materials were used: strain cultures determined in CCU (color changing units) / mL or CFU / mL, or a genomic DNA extract (in copies / mL); test concentrations are referenced against an LoD of 1x103 CCUmL that correlates approximately to 1x106 CFU / mL or 1x106 copies / mL.

Sequencing of one P. mirabilis and one P. vulgaris strain with reduced performance revealed single mismatches for internal probes.

C Test concentrations for P. hauseri and P. penneri are shown as LoD multiples as determined for P. vulgaris.

One strain did not perform consistently at concentrations close to the LoD; sequencing attempts have failed.

8 Strain ATCC 13637 (type strain) was used to establish LoD. During inclusivity testing, single failures at concentrations above LoD were observed, although no probe mismatches were present. Inclusivity testing was repeated for this strain at 1x ARM LoD and 2/2 positive replicates were obtained.

Strain DSM-50173 showed reduced sensitivity and sequencing revealed multiple mismatches for one internal probe.

Strain DSM-21874 showed reduced sensitivity, however, sequencing did not reveal any mismatches to primers or probes. Testing was repeated for this strain at 1x LoD (0/2), 1.7x (0/2) and 3.3x LoD (2/2).

To supplement inclusivity testing for each LRT panel microorganism target, in-silico analysis was performed to assess Unyvero LRT primer and probe sequences for the predicted detection of microorganism strains with applicable sequences available in the GenBank database (search performed January 2018).

Results from the in-silico analyses identified strain entries that are predicted to be

{19}------------------------------------------------

detected when the organism is present at LoD concentrations (match of relevant primer and hybridization probe sequences), predicted to be detected with reduced sensitivity (typically, single relevant primer or probe sequence mismatches; detection likely at higher than LoD concentrations only) or predicted to be not detected at clinically relevant concentrations (multiple relevant primer mismatches in primer and probe sequences).

Table 7 lists microorganisms that were evaluated in inclusivity testing that are also predicted to be detected at LoD concentrations based on in-silico analysis. Table 8 lists microorganisms evaluated in inclusivity testing for which one or more strain entries are predicted to be detected with reduced sensitivity.

Additional microorganisms were evaluated by in-silico analysis only. For these microorganisms, Table 9 lists those that are predicted to be detected when present at LoD concentrations and Table 10 lists microorganisms that are predicted to be detected with reduced sensitivity for one or more strain entries.

The following language was included in the device labeling along with results from the in-silico analyses performed:

  • . In-silico analysis results were provided as supplementary data. The results are not intended to be a surrogate for wet testing and do not assure that specific strains will be detected.
  • . The performance of the Unyvero LRT Application has not been established for those microorganism species that were evaluated by in-silico analysis only.
MicroorganismMicroorganism
Acinetobacter baumanniiMycoplasma pneumoniae
Acinetobacter calcoaceticusProteus mirabilis
Acinetobacter lwoffiiProteus vulgaris
Acinetobacter haemolyticusProteus hauseric
Chlamydia pneumoniaeProteus penneric
Enterobacter cloacaeaPseudomonas aeruginosa
Escherichia coliSerratia marcescens
Klebsiella pneumoniaebStaphylococcus aureus
Klebsiella variicolaStreptococcus pneumoniaed

Table 7: Microorganisms with reference strains detected at or near LoD concentrations in inclusivity wet testing that are predicted to be detected at LoD based on in-silico analysis for all strain entries

ª Including E. cloacae ssp. dissolvens

b Including K. pneumoniae variant 2 (K. quasipneumoniae)

6 No Genbank entries available, BLAST search was performed using the whole genome shotgun (wgs) database d In-silico analysis includes serotype 7F

{20}------------------------------------------------

Table 8: Microorganisms with reference strains detected at or near LoD concentrations in the Inclusivity study and are predicted to be detected with reduced sensitivity for one or more strain entries

Microorganism# entries predicted to be detectedat LoD# entries predicted to be detected withreduced sensitivity
Citrobacter freundii113
Enterobacter asburiae-9
Enterobacter hormaechei11a1
Klebsiella oxytoca83
Legionella pneumophila3136b
Moraxella catarrhalis114
Morganella morganii15
Stenotrophomonas maltophilia172

a In-silico analysis includes subspecies. oharae and steigerwaltii

8 Inclusivity wet testing included three strains predicted with reduced sensitivity by in-silico analysis (ATCC 33152, ATCC 33154); all strains were detected at concentrations at LoD.

Table 9: Microorganisms predicted to be detected at LoD based on in-silico analysis only

Microorganism
Acinetobacter nosocomialisAcinetobacter oleivorans
Acinetobacter pittiiAcinetobacter schindleri
Acinetobacter juniiEnterobacter kobei
Acinetobacter parvusEnterobacter ludwigii
Acinetobacter lactucaeEnterobacter xiangfangensis

Table 10: Microorganisms predicted to be detected with reduced sensitivity for one or more strain entries based on in-silico analysis only

Microorganism# entries predicted to be detected atLoD# entries predicted to be detected withreducedsensitivity
Acinetobacter ursingii-9 a,b
Acinetobacter soli-1
Acinetobacter guillouiae-1

a No Genbank entries available, BLAST search was performed using the whole genome shotgun (wgs) database

b Tests with reference strain DSM-16037 were negative at 107 CFU/mL

Based on the results from the Inclusivity study and in-silico analyses for microorganism targets, the following limitation is included in the Unyvero LRT labeling:

  • . Based on in-silico analyses, some strains may either not be detected or be detected with reduced sensitivity due to variations in targeted sequences for

{21}------------------------------------------------

the following microorganisms: C freundii, K. oxytoca, L. pneumophilia, M. catarrhalis, M. morganii. Based on wet testing for Proteus spp. and S. marcescens and based on in-silico/wet testing for M. catarrhalis and S. maltophilia some strains may be detected with reduced sensitivity due to variations in targeted sequences.

Inclusivity/Resistance Marker Targets:

Inclusivity testing for resistance marker targets included testing of samples prepared with well-characterized strains determined to carry resistance markers targeted by the Unyvero LRT Application. Samples were prepared at near LoD concentrations (<3x LoD) for each resistance marker and tested in duplicate for each strain. Results are shown in Table 11 below. When phenotypic antibiotic susceptibility testing results (AST) were available, the corresponding information is also shown in the table (e.g., Carbapenema = resistant to Carbapenems, Carbapenem = susceptible to Carbapenems).

Inclusivity testing resulted in detection of targeted resistance markers for most strains evaluated. For the following microorganism/resistance markers, one or more strains were not detected or were detected only at higher than LoD concentrations (also bolded in Table 11 below):

  • One strain carrying ctx-M (E. cloacae) detected at 6x LoD but not detected . at lower concentrations (one of eight ctx-M carrying strains tested)
  • . Four of eight strains carrying vim:
    • -Two strains carrying vim-1 (E. cloacae, K. pneumoniae) not detected at any concentration tested.
      • -One strain carrying vim-1 (C. freundii) detected in one of two replicates at 4x LoD and detected in both replicates at 10x LoD.
      • -One strain carrying vim-10 (P. aeruginosa) detected in one of two replicates at 2x LoD and 2/2 replicates detected at 8x LoD.

Table 11: Resistance marker targets, Inclusivity strains

MarkerSubgroupHost StrainStrain IDTest Conc.[CFU/mL]LoD a# Positive/# Tests
ctx-Mctx-M3Klebsiella pneumoniaeNRZ-007516x1040.6x4/4
ctx-Mctx-M15Klebsiella pneumoniaeNRZ-002496x1040.6x2/2
ctx-MNAKlebsiella pneumoniaeNCTC 134431x1051x2/2
ctx-MNAEnterobacter cloacaeJMI 462391x1056x1051x6x0/22/2
ctx-MNAEscherichia coliJMI 500676x1040.6x2/2
ctx-MNAKlebsiella pneumoniaeNRZ-000026x1044x1050.6x4x1/22/2
ctx-MNAKlebsiella pneumoniaeJMI 497676x1040.6x2/2
ctx-MNAKlebsiella pneumoniaeNRZ-004721.2x1051.2x2/2
kpckpc-2Escherichia coliNRZ-002818x1051.6x2/2
MarkerSubgroupHost StrainStrain IDTest Conc.[CFU/mL]LoD a# Positive/# Tests
kpckpc-2Klebsiella pneumoniaeNRZ-001036x1040.1x2/2
kpckpc-3Escherichia coliNRZ-002228x1051.6x2/2
kpckpc-3Klebsiella pneumoniaeNRZ-002231x1050.2x2/2
kpckpc-3Klebsiella pneumoniae(CarbapenemR)Micromyx 46534x1050.8x2/2
kpckpc-3Klebsiella pneumoniae(CarbapenemR)Micromyx 46768x1051.6x2/2
ndmndm-1Acinetobacter baumanniiJMI 497551x1052x2/2
ndmndm-1Enterobacter cloacae(ImipenemR, ErtapenemR)ATCC BAA-24681x1052x2/2
ndmndm-1Enterobacter cloacaeJMI 462391x1052x2/2
ndmndm-1Escherichia coliJMI 500676x1041.2x2/2
ndmndm-1Klebsiella pneumoniaeJMI 497676x1041.2x2/2
ndmndm-1Klebsiella pneumoniaeNCTC 134431.5x1053x2/2
ndmndm-1Klebsiella pneumoniaeJMI 498311x1052x2/2
oxa-23oxa-23Acinetobacter baumanniiNCTC 133012x1071x3/3
oxa-23NAAcinetobacter baumannii(CarbapenemR)Micromyx 44102x1071x2/2
oxa-23NAAcinetobacter baumannii(CarbapenemR)Micromyx 61482x1071x2/2
oxa-23NAAcinetobacter baumannii(CarbapenemR)Micromyx 61492x1071x2/2
oxa-23NAAcinetobacter baumannii(CarbapenemR)Micromyx 61532x1071x2/2
oxa-23NAAcinetobacter baumannii(CarbapenemR)Micromyx 63342x1071x2/2
oxa-23NAAcinetobacter baumannii(ImipenemR, MeropenemR)UCLA A52x1071x2/2
oxa-24oxa-25Acinetobacter baumanniiNCTC 133025x1040.1x4/4
oxa-24oxa-72Acinetobacter baumanniiNRZ-004495x1040.1x4/4
oxa-24NAAcinetobacter baumannii(ImipenemR MeropenemR)UCLA A45x1040.1x3/4
oxa-24NAAcinetobacter baumannii(ImipenemR MeropenemR)clinical strain 15x1040.1x2/2
oxa-24NAAcinetobacter baumannii(ImipenemR, MeropenemR)clinical strain 25x1040.1x2/2
oxa-48oxa-48Escherichia coli(ErtapenemR)ATCC BAA-25232x1061x2/2
oxa-48oxa-48Escherichia coliNRZ-001762x1061x2/2
oxa-48oxa-48Klebsiella pneumoniaeNRZ-000022x1061x2/2
oxa-48oxa-48Klebsiella pneumoniaeNCTC 134422x1061x2/2
oxa-48oxa-162Escherichia coliNRZ-003613x1061.5x2/2
oxa-48oxa-162Klebsiella pneumoniaeNRZ-004725x1062.5x2/2
oxa-48oxa-232Klebsiella oxytocaNRZ-220602x1061x2/2
oxa-58oxa-58Acinetobacter baumanniiNRZ-005184x1050.5x2/2
oxa-58oxa-58Acinetobacter baumanniiNCTC 133054x1050.5x2/2
MarkerSubgroupHost StrainStrain IDTest Conc.[CFU/mL]LoD a# Positive /# Tests
temtem-1Escherichia coliATCC 352181x1051.6x2/2
temtem-3Escherichia coli (ESBL)NCTC 133515x1040.8x2/2
temNACitrobacter freundiiATCC 438641.8x1053x2/2
temNAEnterobacter cloacaeJMI 462391x1051.7x2/2
temNAKlebsiella pneumoniaeJMI 497676x1041x2/2
temNAEscherichia coliNRZ-002811.8x1052.5x1053x4x1/22/2
temNAHaemophilus influenzae(AmpicillinR, CefinaseR)clinical strain 11x1051.6x5/5
temNAHaemophilus influenzae(CefinaseR)clinical strain 21x1051.6x5/5
vimvim-1Citrobacter freundiiNRZ-00452 b2x1055x1054x10x1/22/2
vimvim-1Pseudomonas aeruginosa(CeftazidimeR, ImipenemR)DSM-246005x1041x4/4
vimvim-1Enterobacter cloacaeNRZ-00239 b5x1041x1051x2x0/20/4
vimvim-1Klebsiella pneumoniaeNCTC 134395x1041x2/2
vimvim-1Klebsiella pneumoniaeNCTC 13440 b4x1058x0/2
vimvim-10Pseudomonas aeruginosaNCTC 13437 b1x1054x1052x8x1/22/2
vimNAPseudomonas aeruginosaUCLA P203x1040.6x2/2
vimNAPseudomonas putida(CarbapenemR)Micromyx 16124x1040.8x2/2
vimNAPseudomonas aeruginosa(CarbapenemR)Micromyx 25621x1052x4/4
mecASCCmecIStaphylococcus aureusRKI 07-031654x1062x2/2
mecASCCmecIIStaphylococcus aureusRKI 01-006944x1062x2/2
mecASCCmecIIIStaphylococcus aureus(MethicillinR)ATCC 335913x1061.5x2/2
mecASCCmecIVStaphylococcus aureusRKI 09-001874x1062x2/2
mecASCCmecVStaphylococcus aureusRKI 08-024924x1062x1/2
mecANAStaphylococcus aureus(MethicillinR, CefoxitinR)NCTC 124936x1050.3x2/2
mecANAStaphylococcus aureus(MethicillinR)DSM-170913x1061.5x2/2

{22}------------------------------------------------

{23}------------------------------------------------

4 Analyte LoDs were established using "artificial respiratory medium" (ARM) as sample matix surrogate for aspirates. Inclusivity was performed by contrived samples using PBS as sample matrix. Test concentrations are given as multiples of claimed ARM LoDs. PBS and ARM LoDs are comparable for most analytes PBS and ARM LoDs differ by a factor between 0.3 3. b For three vim-1 strains and one vim-10 strain positive results were only obtained at 8-10x LoD or were negative at applied concentrations. For strain NCTC 13437 (vin-10) sequencing did not show any primer or probe mismatches. For strains NRZ-00239, NRZ-00452, NCTC 13440 (all vim-1) sequencing revealed mismatches for all internal probes.

To supplement inclusivity testing for specific resistance marker variants and subgroups, in-silico analysis was performed with Unyvero LRT primer and probe sequences compared to sequences available in the GenBank database. Variants predicted to be detected at LoD (match of relevant primer and probe sequences), variants predicted to be detected with reduced sensitivity (typically, single relevant

{24}------------------------------------------------

mismatches of primer or probe sequences; detection likely at higher than LoD concentrations only), and variants predicted to be not detected at clinically relevant concentrations (multiple relevant mismatches in primer and probe sequences) are listed in Table 12.

The following statements are included in the assay labeling:

  • In-silico analysis results were provided as supplementary data. Results are not . intended to be a surrogate for wet testing and do not assure that specific resistance marker variants will be detected by the assay.
  • . The performance of the Unyvero LRT Application has not been established for those resistance marker variants that were evaluated by in-silico analysis only.
Resistance Marker:SubgroupVariants predicted at LoDVariants predicted atreduced sensitivityVariants not predicted
ctx-M:ctx-M1 subgroup1, 3, 10, 11, 15, 22, 23, 28-30, 32-34, 36, 37, 42, 52-55, 57, 58, 61, 66, 69, 71,72, 79, 80, 83, 88, 101, 103,108, 109, 114, 116, 117,132, 136, 138, 139, 142,144, 150, 155-158, 162-164, 166, 167, 170, 172,173, 175-177, 179-184,186, 188, 189, 19012, 60, 62, 64, 68, 82, 96,107, 133, 169-
ctx-M:ctx-M2ctx-M8ctx-M9ctx-M25ctx-M45subgroups--2, 4-9, 13, 14, 16-21, 24-27,31, 35, 38-41, 43-51, 56,59, 63, 65, 67, 73-78, 81,84-87, 89-95, 97-100, 102,104-106, 110-113, 115,121-126, 129-131, 134,137, 141, 147, 148, 152,159-161, 165, 168, 171,174, 185, 191
kpc1-32--
ndm1, 3-19, 212-
oxa:oxa-2323, 27, 49, 73, 134, 146,165-171, 225, 239, 366,398, 422, 423, 435, 440,469, 481, 482, 483, 565103, 133-
oxa:oxa-2424-26, 33, 40, 72, 139, 160,207, 437--
oxa:oxa-4848, 48b, 162, 163, 181, 199,232, 244, 245, 247, 252,370, 405, 416, 438, 439,484, 505, 514, 515, 517,519, 538, 546, 566, 567204, 54754, 436, 535
oxa:oxa-5858, 96, 97, 164, 397, 467512420

Table 12: In-silico predicted detection, resistance marker variants

{25}------------------------------------------------

Resistance Marker:SubgroupVariants predicted at LoDVariants predicted atreduced sensitivityVariants not predicted
tem1-4, 6, 8-12, 15-17, 19, 20-22, 24, 26, 28-30, 32-36,40, 43, 45, 47-49, 52-55,57, 60, 63, 67, 68, 70-72,76-88, 90-96, 101, 102,104-116, 120-139, 141-150,152-160, 162, 164, 166-169, 171, 176-177, 181-199, 201, 204-217, 219,220, 224-22897, 98, 99, 151, 163178
vim1-4, 6, 8-12, 14, 15-20, 23,24, 26-29, 30, 31, 33-37,39-46, 48, 50-52, 54, 555, 25, 38, 497, 13, 47

Although the Unyvero LRT Application can detect multiple tem variants as shown in Table 13 above, positive or negative tem results are reported by the assay only if H. influenzae is concurrently detected in the specimen. It is noted that the tem-1 variant is carried by H. influenzae while other gram negative rods species can carry tem-1 or other tem variants. If the source of positive tem result is not H. influenzae and H. influenzae is not detected, the tem results will not be reported by the LRT software (i.e., it will be masked). However, if the source of a positive tem result is another microorganism and H. influenzae is also detected in the specimen, the tem result will be reported.

The following limitation is included in the Unyvero LRT Application labeling:

  • Because the tem gene is ubiquitous in members of the Enterobacteriaceae. . positive LRT results for tem may be due to the presence of Enterobacteriaceae in the specimen.
    In summary, based on the results from wet-testing and in-silico analyses for targeted resistance markers, the following limitation is included in the Unyvero LRT labeling:

  • . Wet testing was not performed for all known resistance marker types and/or subtypes. Based on in-silico analyses and inclusivity wet testing, some antibiotic resistance marker variants may either not be detected or detected with reduced sensitivity due to variations in targeted sequences for ctx-M1 subgroup, ndm. oxa-23. oxa-48. oxa-58. vim.
    The potential for false negative results for targeted resistance markers is further mitigated by inclusion of the following limitation in the Unyvero LRT Application labeling and report:

  • Antimicrobial resistance may occur via multiple mechanisms other than the . resistance markers detected by the Unyvero LRT assay. Negative results for LRT resistance markers do not indicate antimicrobial susceptibility of

{26}------------------------------------------------

detected organisms.

Cross-Reactivity/Exclusivity: f.

A study was conducted to evaluate the potential for cross-reactivity (exclusivity) of Unyvero LRT Application targets with closely related microorganisms as well as commensal microorganisms that are commonly present in the respiratory tract. Study samples were prepared with microorganisms at high concentrations (~107 CFU/mL) and testing was performed in duplicate.

In addition to testing of exclusivity samples, in-silico (BLAST) analysis was used to evaluate the potential for cross-reactivity for any microorganism strain entries with applicable sequences available in the Gen-Bank database.

The following cross-reactivity with Unyvero LRT targets was either observed in the evaluation of test samples or predicted based on in-silico analysis.

  • . Cross-reactivity with the Citrobacter freundii target is predicted with Citrobacter braakii and Kluyvera georgiana.
  • . Cross-reactivity with the Enterobacter cloacae complex target is predicted with Enterobacter soli, Enterobacter mori, and Enterobacter nickellidurans.
  • Cross-reactivity with the E. coli target is predicted with Shigella dysenteriae, . Shigella boydii, Shigella flexneri, Shigella sonnei, Escherichia albertii and Escherichia fergusonii.
  • . Cross-reactivity with the Haemophilus influenzae target was demonstrated with Haemophilus haemolyticus and Haemophilus parainfluenzae. Note that wet- testing of one Haemophilus parainfluenzae strain generated expected negative results.
  • . Cross-reactivity with the Klebsiella oxytoca target is predicted with Klebsiella michagenensis.
  • Cross-reactivity with the Staphylococcus aureus target is predicted with . Staphylococcus argenteus and Staphylococcus simiae.
  • . Cross-reactivity with the Stenotrophomonas maltophilia target is predicted with Stenotrophomonas nitritireducens, Stenotrophomonas daejeonensis, Stenotrophomonas acidaminiphila, Stenotrophomonas koreensis and Stenotrophomonas rhizophila. All Stenotrophomonas species predicted to be cross-reactive are not associated with human respiratory infection.

Based on in-silico analysis, no cross-reactivity is predicted for the following LRT panel targets: Acinetobacter spp., C. pneumoniae, K. pneumoniae, K. variicola, L. pneumophila, M. catarrhalis, M. morganii, M. pneumoniae, Proteus spp., P. aeruginosa, S. marcescens, ctx-M, kpc, ndm, oxa-23, oxa-24, oxa-48, oxa-58, tem, vim, and mecA.

The following limitation regarding cross-reactivity of the Unyvero LRT Application is include in device labeling:

{27}------------------------------------------------

  • . Based on in-silico analysis and exclusivity wet testing, the following LRT panel microorganism assays are expected to cross-react with closely related clinically relevant species: Citrobacter freundii (cross-reactive to C. braakii, Khyvera georgiana;), Escherichia coli (cross-reactive to E. albertii, E. fergusonii and Shigella spp. (S. dysenteriae, S. sonnei, S. flexneri, S. boydii)), Haemophilus influenzae (cross-reactive to H. haemolyticus and H. parainfluenzae), Klebsiella oxytoca (cross-reactive to K. michiganensis).
    Results of in-silico analysis and wet-testing for near-neighbor microorganism strains is shown in Table 13.

No cross reactivity was observed in testing of the commensal microorganism strains listed in Table 14.

Close Neighbor StrainCross-Reactivity Prediction(in-silico analysis)Wet TestingResultat 107CFU/mL,Strain IDCross-Reactionsobserved in ClinicalStudy(N = number ofspecimens)
Citrobacter freundiiDetection predicted at higherthan LoD concentrations b--
Citrobacter braakiiDetection predicted at higherthan LoD concentrations--
Kluyvera georgianaDetection predicted at higherthan LoD concentrations--
Citrobacter koseriDetection not predictednegativeATCC 27156-
Enterobacter cloacae complex---
Enterobacter soli aDetection predicted at LoD--
Enterobacter mori aDetection predicted at LoD--
Enterobacter nickellidurans aDetection predicted at LoD--
Escherichia coliDetection predicted at LoD--
Shigella dysenteriae aDetection predicted at LoD--
Shigella boydii aDetection predicted at LoD--
Shigella flexneri aDetection predicted at LoD--
Shigella sonnei aDetection predicted at LoD--
Escherichia albertiiDetection predicted at LoD--
Escherichia fergusoniiDetection predicted at LoD--
Haemophilus influenzae---
Haemophilus haemolyticusDetection predicted at higherthan LoD concentrationspositiveATCC 333902
Haemophilus parahaemolyticusDetection not predictednegativeATCC 10014-
Haemophilus parainfluenzaeDetection not predictednegativeATCC 333921
Aggregatibacter actino-mycetemcomitansDetection not predictednegativeATCC 33384-
Close Neighbor StrainCross-Reactivity Prediction(in-silico analysis)Wet TestingResultat 107CFU/mL,Strain IDCross-Reactionsobserved in ClinicalStudy(N = number ofspecimens)
Aggregatibacter aphrophilusDetection not predictednegativeATCC 19415-
Klebsiella oxytoca--
Klebsiella michiganensisDetection predicted at LoD/Detection predicted at higherthan LoD concentrations c--
Staphylococcus aureus--
Staphylococcus argenteus aDetection predicted at LoD--
Staphylococcus simiae aDetection predicted at LoD--
CNS:S. epidermidisS. capitisS. lugdunensisS. haemolyticusS. saprophyticusDetection not predictednegativeATCC 51625ATCC 27840ATCC 43809ATCC 29970ATCC 15305-
Stenotrophomonas maltophilia
Stenotrophomonas spp.,(environmental/soil microorganisms):aS. nitritireducensS. daejeonensisS. acidaminiphilaS. koreensisS. rhizophilaDetection predicted at LoD--
Xanthomonas spp. aDetection predicted at higherthan LoD concentrations---
Pseudoxanthomonas spp. aDetection predicted at higherthan LoD concentrations---
Streptococcus pneumoniae
other Streptococcus sp.:S. agalactiaeS. anginosusS. dysgalactiaeS. gordoniiS. intermediusS. mitisS. mutansS. oralisS. parasanguinisS. pseudopneumoniaeS. pyogenesS. salivariusS. sanguinisS. vestibularisDetection not predictednegativeATCC 13813ATCC 33397ATCC 43078ATCC 10558ATCC 27335ATCC 49456ATCC 25175ATCC 35037ATCC 15912ATCC BAA-960ATCC 12344ATCC 7073ATCC 10556-

Table 13: Exclusivity testing: In-silico prediction, wet-testing of exclusivity samples, and cross-reactivity observed in the clinical study

{28}------------------------------------------------

ª No clinical relevance for respiratory infections

b Strains are predicted to be detected at higher than LoD concentrations due to primer and probe mismatches

< Few strains are predicted to be detected at higher than LoD concentrations

{29}------------------------------------------------

respiratory flora strainstrain IDrespiratory flora strainstrain ID
Actinomyces odontolyticusATCC 17929Granulicatella adiacensATCC 49175
Aspergillus fumigatusATCC 204305Kingella kingaeATCC 23330
Candida albicansATCC 90028Lactobacillus acidophilusATCC 4356
Candida dubliniensisDSM-13268Micrococcus luteusATCC 4698
Candida glabrataATCC 2001Mycobacterium bovisclinical isolate
Candida kruseiATCC 24210Mycoplasma oraleATCC 23714
Candida parapsilosisATCC 22019Neisseria lactamicaATCC 23970
Candida tropicalisATCC 750Neisseria siccaATCC 29193
Cardiobacterium hominisATCC 15826Pantoea agglomeransATCC 27155
Eikenella corrodensATCC 23834Peptostreptococcus stomatisDSM-17678
Enterococcus faecalisATCC 29212Porphyromonas gingivalisATCC 33277
Enterococcus faeciumATCC 35667Prevotella buccalisATCC 35310
Fusobacterium nucleatumATCC 25586Raoultella planticolaATCC 33531

Table 14: Exclusivity testing. Commensal respiratory flora (No cross reactivity observed)

g. Assay Cut-off

The Unyvero LRT Application is comprised of eight individual multiplex PCR assays and hybridization arrays located in separate reaction chambers in the LRT Cartridge. After hybridization of PCR products, fluorescent array signals are captured with a fluorescent camera system over a defined temperature range. After correction for background values, a signal threshold is applied for probes of individual analytes. In addition to the signal thresholds, individual cutoffs are defined to reduce background signals for certain analytes that are commonly part of the host flora of healthy individuals.

h. Interfering substances:

An interfering substances study was conducted to evaluate potential inhibitory effects on the performance of the Unyvero LRT Application with substances that may be present in lower respiratory specimens. The substances evaluated included respiratory medications, antibiotics, sample storage media, sample liquefying agent (LRT lysis buffer reagent), blood, mucin and human DNA. Each substance was evaluated in samples at a (b) (4) concentration as recommended in the CLSI guideline, 'Interference Testing in Clinical Chemistry'. Samples were prepared with pools of six representative microorganisms targeted by the LRT assay: E. coli, K. pneumoniae, M. morganii, P. aeruginosa, S. aureus/mecA+, and S. maltophilia. Each microorganism was spiked into test samples at concentrations close to LoD. Test samples were evaluated in duplicate and compared to results from control samples prepared with microorganisms only.

A qualitative analysis of study results was performed. A substance was considered as a potential interferent if reduced positivity was observed in samples spiked with the

{30}------------------------------------------------

substance in comparison to control samples. A quantitative analysis was also conducted for which trends in assay signal were assessed as compared to control samples.

No interference was observed for any substance evaluated at the tested concentrations. Substances evaluated and their respective test concentrations are presented in Table 15.

Interfering SubstanceTest concentrationInterference
Reference/ControlPBS (no interferents added)N/AN/A
respiratory drugsGuaifenesin$1.5 \times 10^{-2}$ Mno
Dextromethorphan$3.7 \times 10^{-6}$ Mno
Acetyl-Cysteine$1 \times 10^{-2}$ Mno
Salbutamol$4 \times 10^{-6}$ Mno
Carbocystein$2.8 \times 10^{-3}$ Mno
Ambroxol$8 \times 10^{-4}$ Mno
Beclomethason$7 \times 10^{-4}$ Mno
Theophyllin$2.2 \times 10^{-4}$ Mno
antibioticsAmpicillin$1.5 \times 10^{-4}$ Mno
Cefuroxime$1.4 \times 10^{-3}$ Mno
Erythromycin$8.2 \times 10^{-5}$ Mno
Ciprofloxacin$3 \times 10^{-5}$ Mno
Amikacin$1.4 \times 10^{-4}$ Mno
Imipenem$5 \times 10^{-4}$ Mno
Clindamycin$8.9 \times 10^{-5}$ Mno
Trimethoprim$1.4 \times 10^{-4}$ Mno
Sulfamethoxazole$1.6 \times 10^{-3}$ Mno
inhalation agent (NaCl)sodium chloride5% w/vno
lysis bufferlysis buffer DTT, 90% v/vlysis buffer: 90% v/v(final conc. in lysistube) or 80% v/v (foradded sample), DTT:40 mM (final conc. inlysis tube) or 35 mM(for added sample)no
sample components/ARM matrixcomponentsEDTA blood100% v/vno
human placenta DNA1 µg/µLno
fish sperm DNA4 µg/µLno
mucin (pig stomach, type II)20 mg/mLno

Table 15: Interfering substances study

h. Competitive Interference:

To evaluate the potential for competitive inhibition between targeted microorganisms, various combinations of microorganisms were evaluated in contrived samples. Study samples were prepared with low positive analytes at near LoD concentrations. together with high positive analytes at ~107 CFU or higher. Control samples containing analytes at low concentrations were also tested.

{31}------------------------------------------------

The following analytes were selected for evaluation for competitive inhibition based on the prevalence of each microorganism in tracheal aspirate specimens. Two sample pools were prepared with multiple targeted microorganisms at either low or high concentrations in each sample as shown in Table 16 and 17.

Low Positive Analytes (tested nearLoD)High Positive Analytes
E. coli/temP. aeruginosa
K. pneumoniaeS. aureus/mecA
S. maltophiliaA. baumanii
Proteus spp.

Table 16: Pool 1 (High positive/Low positive microorganism pool)

Table 17: Pool 2 (High positive/Low positive microorganism pool)
Low PositiveAnalytesHigh PositiveAnalytes
P. aeruginosaE. coli/tem
S. aureus/mecAK. pneumoniae
A. baumaniiS. maltophilia
P. mirabilis

Six replicates were evaluated for each of the two test samples and each of the two controls.

Except for Proteus spp., all low positive microorganisms listed in Tables 14 and 15 were detected in all test replicates. For Proteus spp., only 4/6 replicates were positive for samples containing low positive Proteus concentrations combined with other microorganisms in high concentrations. These results are acceptable as Proteus spp. was present in the sample pool at a concentration of 0.5x LoD. In addition, the low positive control was positive for only 2/6 replicates, indicating that the false negative results for Proteus spp. were not likely due to competitive inhibition.

i. Carry-over:

A study was conducted to evaluate the potential for sample to sample carry-over during the Unyvero LRT Application testing process. The study comprised alternating high positive and negative contrived samples tested on the same Unyvero test system. The system setup included 1 Unyvero cockpit, 1 Lysator and 2 Analyzers. The test set included a total of 20 (5 on each analyzer slot) positive and negative cartridge runs. Prior to test initiation, additional negative cartridge runs per slot were performed to confirm the absence of any contaminants in test materials and devices. Samples were prepared in ARM matrix with positive samples containing seven representative microorganisms at

{32}------------------------------------------------

high concentrations (~107 CFU/mL). Results for all 48 test runs were fully valid with no false positive results observed for negative ARM samples or for negative controls.

  • i. Fresh versus frozen study:
    A fresh versus frozen study was conducted to assess the impact of freezing tracheal aspirate specimen on the performance of the LRT assay. Aspirate specimens that were tested prospectively (fresh) during the LRT clinical study were re-tested after having been exposed to prolonged storage since initial testing. Testing included a total of 30 specimens containing low and high concentrations of representative analytes.

Specimens were chosen to include specimens with initial positive results covering multiple different pathogen and resistance markers as well specimen results covering a representative signal range as observed during prospective LRT clinical testing. Specimens included in the study were positive for both single and multiple target analytes. Due to multi-detections, the set of 30 samples was initially positive for a total of 73 LRT analytes (pathogens and antibiotic resistance markers combined).

Of the 30 test runs performed, 29 runs were completely valid and one was partially valid. Results of the study showed overall positivity of 97% for all analytes (71 of 73 expected positives) and no trend toward higher or lower assay signals between fresh and frozen test results.

Results of the study demonstrated that the freeze-thaw process and prolonged frozen storage did have a significant effect on analyte positivity or on average assay signal intensities.

    1. Comparison studies:
    • a. Method comparison

N/A

  • b. Matrix comparison:
    Due to the challenge of obtaining large volumes of natural negative tracheal aspirate matrix, Artificial Respiratory Matrix (ARM) was used to prepare samples for LoD, reproducibility and other analytical studies. A separate matrix equivalency study was performed to compare assay performance for samples prepared in ARM and samples prepared in pooled natural tracheal aspirate matrix. Tracheal aspirate test samples were prepared using pools of 4-6 individual tracheal aspirates determined to be negative for all LRT panel analytes. Both ARM and natural aspirate matrices were spiked with microorganism concentrations near the LoD for nine representative LRT analytes (between 104 - 106 CFU/mL). Matrix equivalence was evaluated based on qualitative analysis (percent positivity) and quantitative analyses (comparison of mean assay signal).

{33}------------------------------------------------

As shown in Table 18 below, positivity rates for ARM Matrix were slightly lower for all analytes overall than for the natural aspirate matrix, suggesting that the ARM matrix may be somewhat more challenging (i.e., for some analytes, the LoD may be slightly higher in ARM than in natural tracheal aspirate matrix). It was noted however that Unyvero LRT assay signal levels for positive Unyvero LRT samples were equivalent between the two matrix types for each LRT microorganisms. The study results support the use of ARM for preparation of samples for the reproducibility study and other analytical studies.

Target AnalyteARM MatrixAspirate Matrix
S. marcescens4/66/6
E. coli6/66/6
K. pneumoniae6/65/6
vim6/66/6
M. morganii5/66/6
S. aureus5/66/6
tem6/66/6
P. aeruginosa4/66/6
S. maltophilia5/66/6
TotalPositive/Expected47/54 = 87%53/54 = 98%

Table 18: Results from Matrix Comparison

3. Clinical Studies:

Prospective Study:

Clinical performance of the Unyvero LRT Application performed on the Unyvero System was evaluated in a multi-center study at nine clinical sites in the United States. A total of 860 tracheal aspirate specimens were prospectively collected from patients with signs and symptoms of lower respiratory infection and were tested with the LRT assay within 24 hours of specimen collection. Specimens excluded from the performance analyses included 38 specimens that did not meet the specimen inclusion/exclusion criteria, 161 specimens that generated non-reportable results (e.g., fully invalid results or instrument failures) and 58 specimens that generated partially invalid results (invalid results in one or more PCR chambers). Altogether, a total number of 603 evaluable prospectively tested specimens were included in the performance analyses. Gram stains (quality screening) were performed for the majority of specimens tested in the study.

Reference/comparator methods used in the prospective study (Table 19) included standard of care (SoC) tracheal aspirate culture and validated comparator PCR assays. All positive comparator PCR results were followed with bi-directional sequencing. Validation of each comparator PCR assay included demonstration of similar LoDs and inclusivity to the Unyvero LRT Application.

{34}------------------------------------------------

Table 19: Prospective Study Reference/Comparator Methods

Prospective Study Reference/Comparator Methods
a) SoC (culture)
b) Composite Comparatorfor 'typical'1 microorganisms SoC (culture) Independent and validated multiplexed PCR assay for which any positive PCR result is followed by bi-directional sequencing. One comparator PCR per microorganism target for 'atypical'2 microorganisms Two independent and validated multiplex PCR assays for which any positive PCR result is followed by bi-directional sequencing
c) Multiplexed PCR assays followed by bi-directional sequencing (for antibiotic resistance markers). One comparator PCR target per resistance marker.

H. influenzae, K. oxytoca, K. pneumoniae, K. variicola, M. catarrhalis, M. morganii. Proteus sp., P. aeruginosa. S. marcescens. S. aureus. S. maltophilia. S. pneumoniae.

2 Atypical' analytes: C. pneumoniae, L. pneumophila, M. pneumoniae

For each 'typical' microorganism target, clinical performance of the Unyvero LRT Application was evaluated in comparison to culture. In addition, clinical performance was evaluated as compared to a composite comparator of culture and PCR/bi-directional sequencing. For the composite comparator, the specimen was considered positive for a microorganism target if culture was positive or if the validated comparator PCR and follow-up bi-directional sequencing was positive. Any specimen that was negative by both culture and PCR was considered negative for the microorganism.

For each 'atypical' microorganism target (Chlamydia pneumoniae, Mycoplasma pneumoniae and Legionella pneumophila), clinical performance of the Unyvero LRT Application was evaluated in comparison to a composite of two validated PCR assays for which all positive PCR results were followed by bi-directional sequencing. Specimens that were positive for a targeted microorganism by either of the two PCR/sequencing assays were considered positive by the composite comparator. Specimens that were negative for both PCR assays were considered negative for the analyte.

Clinical performance of the Unyvero LRT Application for 'typical' microorganism targets as compared to culture are shown in Table 20. Performance for both 'typical' and 'atypical' microorganism targets as compared to their respective composite comparator methods are shown in Table 21.

{35}------------------------------------------------

TPFNFP bTNPPA [%](95% CI)NPA [%](95% CI)PPV [%](95% CI)NPV [%](95% CI)Semi-quantitativeculture resultTPFNPPA [%]
Acinetobacter spp.10017576100.0(72.3 - 100.0)97.1(95.5 - 98.2)37.0(21.5 - 55.8)100.0(99.3 - 100.0)Acinetobacter spp.rare20100.0
Citrobacter freundii13259725.0(4.6 - 69.9)99.7(98.8 - 99.9)33.3(6.1 - 79.2)99.5(98.5 - 99.8)Acinetobacter spp.few00na
Enterobacter cloacaecomplex141658293.3(70.2 - 98.8)99.0(97.8 - 99.5)70.0(48.1 - 85.5)99.8(99.0 - 100.0)Acinetobacter spp.moderate60100.0
Escherichia coli2312255795.8(79.8 - 99.3)96.2(94.3 - 97.5)51.1(37.0 - 65.0)99.8(99.0 - 100.0)Acinetobacter spp.numerous20100.0
Haemophilus influenzae8016579100.0(67.6 - 100.0)97.3(95.7 - 98.3)33.3(18.0 - 53.3)100.0(99.3 - 100.0)Citrobacter freundiirare010.0
Klebsiella oxytoca421158666.7(30.0 - 90.3)98.2(96.7 - 99.0)26.7(10.9 - 52.0)99.7(98.8 - 99.9)Citrobacter freundiifew010.0
Klebsiella pneumoniae a2131756287.5(69.0 - 95.7)97.1(95.3 - 98.2)55.3(39.7 - 69.9)99.5(98.5 - 99.8)Citrobacter freundiimoderate10100.0
Klebsiella variicola a202599100.0(34.2 - 100.0)99.7(98.8 - 99.9)50.0(15.0 - 85.0)100.0(99.4 - 100.0)Citrobacter freundiinumerous010.0
Moraxella catarrhalis1212c58833.3(6.1 - 79.2)98.0(96.5 - 98.9)7.7(1.4 - 33.3)99.7(98.8 - 99.9)Enterobacter cloacae complexrare00na
Morganella morganii109593100.0(20.7 - 100.0)98.5(97.2 - 99.2)10.0(1.8 - 40.4)100.0(99.4 - 100.0)Enterobacter cloacae complexfew60100.0
Proteus spp.1011957390.9(62.3 - 98.4)96.8(95.0 - 97.9)34.5(19.9 - 52.7)99.8(99.0 - 100.0)Enterobacter cloacae complexmoderate6185.7
Pseudomonas aeruginosa6452051492.8(84.1 - 96.9)96.3(94.3 - 97.6)76.2(66.1 - 84.0)99.0(97.8 - 99.6)Enterobacter cloacae complexnumerous20100.0
Serratia marcescens11014578100.0(74.1 - 100.0)97.6(96.1 - 98.6)44.0(26.7 - 62.9)100.0(99.3 - 100.0)Escherichia colirare30100.0
Staphylococcus aureus8123948197.6(91.6 - 99.3)92.5(89.9 - 94.5)67.5(58.7 - 75.2)99.6(98.5 - 99.9)Escherichia colifew4180.0
Stenotrophomonasmaltophilia2512755096.2(81.1 - 99.3)95.3(93.3 - 96.8)48.1(35.1 - 61.3)99.8(99.0 - 100.0)Escherichia colimoderate80100.0
Streptococcuspneumoniae72658877.8(45.3 - 93.7)99.0(97.8 - 99.5)53.8(29.1 - 76.8)99.7(98.8 - 99.9)Escherichia colinumerous80100.0
TPFNFPTNPPA [%](95 % CI)NPA [%](95 % CI)PPV [%](95 % CI)NPV [%](95 % CI)Haemophilus influenzaerare00na
Acinetobacter spp.231457595.8(79.8 - 99.3)99.3(98.2 - 99.7)85.2(67.6 - 94.1)99.8(99.0 - 100.0)Haemophilus influenzaefew10100.0
Chlamydiapneumoniaeb000603na100.0(99.4 - 100.0)na100.0(99.4 - 100.0)Haemophilus influenzaemoderate20100.0
Citrobacter freundii15259516.7(3.0 - 56.3)99.7(98.8 - 99.9)33.3(6.1 - 79.2)99.2(98.1 - 99.6)Haemophilus influenzaenumerous50100.0
Enterobacter cloacaecomplex171358294.4(74.2 - 99.0)99.5(98.5 - 99.8)85.0(64.0 - 94.8)99.8(99.0 - 100.0)Klebsiella oxytocarare020.0
Escherichia coli371855797.4(86.5 - 99.5)98.6(97.2 - 99.3)82.2(68.7 - 90.7)99.8(99.0 - 100.0)Klebsiella oxytocafew30100.0
Haemophilusinfluenzae152857788.2(65.7 - 96.7)98.6(97.3 - 99.3)65.2(44.9 - 81.2)99.7(98.7 - 99.9)Klebsiella oxytocamoderate00na
Klebsiella oxytoca72858677.8(45.3 - 93.7)98.7(97.4 - 99.3)46.7(24.8 - 69.9)99.7(98.8 - 99.9)Klebsiella oxytocanumerous10100.0
Klebsiellapneumoniaea303756290.9(76.4 - 96.9)98.8(97.5 - 99.4)81.1(65.8 - 90.5)99.5(98.5 - 99.8)Klebsiella pneumoniaerare010.0
Klebsiella variicola a202599100.0(34.2 - 100.0)99.7(98.8 - 99.9)50.0(15.0 - 85.0)100.0(99.4 - 100.0)Klebsiella pneumoniaefew80100.0
Legionellapneumophila b200601100.0(34.2 - 100.0)100.0(99.4 - 100.0)100.0(34.2 - 100.0)100.0(99.4 - 100.0)Klebsiella pneumoniaemoderate9190.0
Moraxella catarrhalis1211c157952.2(33.0 - 70.8)99.8(99.0 - 100.0)92.3(66.7 - 98.6)98.1(96.7 - 99.0)Klebsiella pneumoniaenumerous4180.0
Morganella morganii61459285.7(48.7 - 97.4)99.3(98.3 - 99.7)60.0(31.3 - 83.2)99.8(99.1 - 100.0)Klebsiella variicolarare00na
Mycoplasmapneumoniae b201600100.0(34.2 - 100.0)99.8(99.1 - 100.0)66.7(20.8 - 93.9)100.0(99.4 - 100.0)Klebsiella variicolafew10100.0
Proteus spp.241557396.0(80.5 - 99.3)99.1(98.0 - 99.6)82.8(65.5 - 92.4)99.8(99.0 - 100.0)Klebsiella variicolamoderate10100.0
Pseudomonasaeruginosa769851089.4(81.1 - 94.3)98.5(97.0 - 99.2)90.5(82.3 - 95.1)98.3(96.7 - 99.1)Klebsiella variicolanumerous00na
Serratia marcescens213457587.5(69.0 - 95.7)99.3(98.2 - 99.7)84.0(65.3 - 93.6)99.5(98.5 - 99.8)Moraxella catarrhalisrare00na
Staphylococcus aureus109101147391.6(85.2 - 95.4)97.7(96.0 - 98.7)90.8(84.3 - 94.8)97.9(96.2 - 98.9)Moraxella catarrhalisfew00na
Stenotrophomonasmaltophilia506254589.3(78.5 - 95.0)99.6(98.7 - 99.9)96.2(87.0 - 98.9)98.9(97.6 - 99.5)Moraxella catarrhalismoderate020.0
Streptococcuspneumoniae106358462.5(38.6 - 81.5)99.5(98.5 - 99.8)76.9(49.7 - 91.8)99.0(97.8 - 99.5)Moraxella catarrhalisnumerous10100.0
Morganella morganiirare00na
Morganella morganiifew00na
Morganella morganiimoderate10100.0
Morganella morganiinumerous00na

Table 20: Prospective study, comparison to reference culture

a As K. variicola is often reported by culture as K. pneumoniae, DNA extracts for culture positive K. pneumoniae samples were sequenced. For two of 26 K. pneumoniae positive samples a sequencing result for K. variicola was obtained. Strains identified for both samples were confirmed by sequencing of provided isolates and K. variicola was assigned as reference identity.

Specimens with false positive LRT results were analyzed with molecular assays (PCR/bi-directional sequencing) using sample DNA extracts for presence or absence of microorganisms was confirmed in 16 of 17 cases for Acinetobacter spp., 0 of 2 cases for C. freundii, 5 of 6 cases for E. cloacae complex, 21 of 22 cases for E. coli, 14 of 16 cases for H. influenzae, 7 of 11 cases for K. oxytoca, 15 of 17 cases for K. variicola, 12 of 12 cases for M. variicola, 12 of 12 cases for M. catarrhalis, 6 of 9 cases for M. morganii, 18 of 19 cases for Proteus spp., 16 of 20 cases for P. aeruginosa, 13 of 14 cases for S. marcescens, 36 of 39 cases for S. aureus, 27 of 27 cases for S. maltophilia, and 6 of 6 cases for S. pneumoniae.

6 11/12 FP results for M. catarrhalis when compared to culture were confirmed by the molecular assay (PCR/bi-directional sequencing). Clinical relevance of such findings however has not been established.

{36}------------------------------------------------

Table 21: Prospective study, comparison to composite comparator

a As K. variicola is often reported by culture as K. pneumoniae, DNA extracts for culture positive K. pneumoniae samples were sequenced. For two of 26 K. pneumoniae positive samples a sequencing result for K. variicola was obtained. Strain identities for both samples were confirmed by sequencing of provided isolates and K. variicola was assigned as reference identity.

b . Atypical' microorganisms C. pneumophila and M. pneumoniae were compared to two independent molecular tests (PCR/bi-directional sequencing) as composite comparator.

6 9/11 FN results when composite comparator were only reported positive by the molecular comparator assay (PCR/bi-directional sequencing) for M. catarrhalis.

{37}------------------------------------------------

For prospective specimens with positive culture results, positive percent agreement (PPA) for the Unyvero LRT Application was calculated based on semi-quantitative reference culture results as shown in Table 22.

Table 22: Prospective study, comparison to semi-quantitative culture results
Semi-

{38}------------------------------------------------

Semi-quantitativeculture resultTPFNPPA [%]
Proteus spp.rare1150.0
few30100.0
moderate30100.0
numerous30100.0
Pseudomonas aeruginosarare7277.8
few13286.7
moderate280100.0
numerous160100.0
Serratia marcescensrare10100.0
few40100.0
moderate40100.0
numerous20100.0
Staphylococcus aureusrare2250.0
few130100.0
moderate350100.0
numerous300100.0
Stenotrophomonas maltophiliarare00na
few60100.0
moderate110100.0
numerous8188.9
Streptococcus pneumoniaerare00na
few00na
moderate6185.7
numerous1150.0

For the 603 prospectively tested specimens included in the performance analyses, the LRT assay detected at least one microorganism in 312 specimens (51.7%) and culture reported at least one microorganism in 236 specimens (39.1%).

Multi-detections were reported by the LRT assay for 125 specimens (20.7%) and reported by culture for 62 specimens (10.3%) (Table 23).

Numbers of Detected MicroorganismsAspirate Specimens
LRT# specimensLRT[%]SoC (Culture)# specimensSoC (Culture)[%]
029148.336760.9
any positive31251.723639.1
118731.017428.9
27512.4559.1
3254.161.0
4172.810.2
530.500.0
640.700.0
810.200.0

Table 23: Numbers of targeted microorganisms as reported by LRT or Culture

{39}------------------------------------------------

For 227 specimens, both the LRT Application and culture reported at least one LRT panel microorganism (Table 24). The LRT Application was negative for nine specimens with positive culture results. For 282 specimens, both the LRT Application and culture reported a negative result (no growth or normal/mixed flora result). For 85 specimens, the LRT Application reported a positive result while culture was negative. Of the 85 LRTpositive/culture negative specimens, culture results were reported as normal flora (65), no growth (18) or presence/absence of flora not reported (2).

Positive CultureNegative Culture
LRT ResultMicroorganism(s) ReportedNo GrowthNormal/mixed FloraNAa
any positive by LRT22718652
negative by LRT91051698

Table 24: Comparison of positive and negative culture results to LRT

a presence or absence of flora not reported

Tables 25 and Table 26 include details of multi-detections observed during the prospective clinical study for 'typical' LRT analytes. Multi-detections reported by the LRT Application (N=122) and compared to culture are shown in Table 23. Multidetections reported by culture (N=62) and compared to Unyvero LRT are shown in Table 24.

Table 23 lists the multi-detection results generated by the LRT Application and identifies any discordant results (microorganism targets identified by the LRT application and negative by culture). For example, both E. coli and S. qureus were detected by LRT in five specimens and both E. coli and S. aureus were reported by culture in only four of the five cases. Similarly, four specimens were positive for both P. aeruginosa and S. aureus by LRT; two of the corresponding cultures were negative for P. aeruginosa and three were negative for S. aureus.

In Table 24, multi-detections generated by culture are compared to LRT assay results and identifies any discordant results (culture positive/LRT negative). For example, culture identified five specimens that grew both S. aureus and K. pneumoniae. In four cases the LRT assay also identified both microorganisms; however, in one case, the LRT assay detected only S. aureus, but not K. pneumoniae.

{40}------------------------------------------------

Table 25: Multi-detections of 'typical' microorganisms detected by LRT Application (N=122) as compared to culture (excludes three specimens that were positive for 'atypical' microorganisms -see table footnote)

Multi-detections in Aspirate Specimens by LRTaNumber of Specimens with Multi-Detections by LRTNumber of Specimens with Discordant ResultsFalse Positive Analytes:Microorganisms negative by SoC culture (LRT Positive/SoC Negative)(N) = Number of Specimens
P. aeruginosa, S. maltophilia97S. maltophilia (7)
E. coli, S. aureus51E. coli (1); S. aureus (1)
P. aeruginosa, S. aureus43P. aeruginosa (2); S. aureus (3)
S. aureus, S. maltophilia42S. aureus (1); S. maltophilia (2)
H. influenzae, S. aureus43H. influenzae (3)
K. pneumoniae, S. aureus30
E. cloacae complex, E. coli31E. cloacae complex (1)
E. coli, P. aeruginosa32E. coli (1); P. aeruginosa (1)
Proteus spp., S. aureus32Proteus spp. (2)
M. catarrhalis, S. aureus32M. catarrhalis (2); S. aureus (1)
H. influenzae, Proteus spp., S. aureus22H. influenzae (2); Proteus spp. (2)
E. coli, K. pneumoniae22E. coli (2); K. pneumoniae (2)
K. pneumoniae, S. maltophilia21S. maltophilia (1)
Proteus spp., P. aeruginosa, S. aureus22Proteus spp. (2); P. aeruginosa (1); S. aureus (2)
E. cloacae complex, K. oxytoca, S. maltophilia22E. cloacae complex (2); K. oxytoca (2); S. maltophilia (2)
K. oxytoca, S. maltophilia21K. oxytoca (1); S. maltophilia (1)
Proteus spp., P. aeruginosa22Proteus spp. (2); P. aeruginosa (1)
Acinetobacter spp., S. aureus22Acinetobacter spp. (2)
S. marcescens, S. aureus22S. marcescens (2); S. aureus (1)
E. coli, K. pneumoniae, Proteus spp.21E. coli (1)
E. cloacae complex, E. coli, S. aureus11E. cloacae complex (1); E. coli (1)
H. influenzae, P. aeruginosa, S. marcescens11H. influenzae (1); P. aeruginosa (1); S. marcescens (1)
Acinetobacter spp., Proteus spp., P. aeruginosa, S. marcescens, S. maltophilia11Acinetobacter spp. (1); Proteus spp. (1); S. marcescens (1); S. maltophilia (1)
E. cloacae complex, K. variicola, M. catarrhalis11M. catarrhalis (1)
E. coli, H. influenzae, K. pneumoniae, M. morganii11H. influenzae (1); M. morganii (1)
H. influenzae, K. pneumoniae11H. influenzae (1)
Acinetobacter spp., E. coli, H. influenzae, P. aeruginosa, S. aureus, S. maltophilia11Acinetobacter spp. (1); E. coli (1); H. influenzae (1); S. maltophilia (1)
E. cloacae complex, E. coli, K. oxytoca, P. aeruginosa11E. cloacae complex (1); E. coli (1); K. oxytoca (1); P. aeruginosa (1)
Acinetobacter spp., K. pneumoniae11Acinetobacter spp. (1); K. pneumoniae (1)
Multi-detections in Aspirate Specimens by LRTApplication aNumber of Specimenswith Multi-Detections byLRTNumber of Specimenswith Discordant ResultsFalse Positive Analytes:Microorganisms negative by SoCculture (LRT Positive/SoC Negative)(N) = Number of Specimens
M. morganii, Proteus spp., P. aeruginosa, S. aureus11M. morganii (1); S. aureus (1)
M. morganii, S. marcescens11S. marcescens (1)
Acinetobacter spp., E. cloacae complex, P. aeruginosa11Acinetobacter spp. (1)
E. coli, P. aeruginosa, S. marcescens11E. coli (1)
K. variicola, S. maltophilia11K. variicola (1); S. maltophilia (1)
K. pneumoniae, S. marcescens, S. aureus11K. pneumoniae (1); S. marcescens (1); S. aureus (1)
Acinetobacter spp., P. aeruginosa, S. aureus, S. maltophilia11Acinetobacter spp. (1); S. aureus (1)
E. coli, H. influenzae, S. aureus11H. influenzae (1)
Acinetobacter spp., K. pneumoniae, P. aeruginosa, S. maltophilia11Acinetobacter spp. (1); K. pneumoniae (1)
Acinetobacter spp., M. morganii, Proteus spp., P. aeruginosa11Acinetobacter spp. (1); M. morganii (1)
K. pneumoniae, M. morganii, Proteus spp., P. aeruginosa, S. marcescens, S. maltophilia11K. pneumoniae (1); M. morganii (1); S. maltophilia (1)
Acinetobacter spp., M. catarrhalis, M. morganii, Proteus spp., P. aeruginosa, S. marcescens, S. maltophilia, S. pneumoniae11M. catarrhalis (1); M. morganii (1); Proteus spp. (1); S. marcescens (1)
E. cloacae complex, Proteus spp.11Proteus spp. (1)
E. coli, S. maltophilia10
M. morganii, P. aeruginosa11M. morganii (1)
E. coli, M. catarrhalis, P. aeruginosa11E. coli (1); M. catarrhalis (1); P. aeruginosa (1)
M. catarrhalis, P. aeruginosa, S. aureus11M. catarrhalis (1); P. aeruginosa (1)
S. marcescens, S. maltophilia10
K. pneumoniae, P. aeruginosa, S. aureus11S. aureus (1)
E. coli, S. marcescens, S. aureus, S. maltophilia11E. coli (1)
K. pneumoniae, M. morganii, S. aureus11M. morganii (1)
Acinetobacter spp., Proteus spp., P. aeruginosa, S. maltophilia11Proteus spp. (1); P. aeruginosa (1); S. maltophilia (1)
C. freundii, K. oxytoca11C. freundii (1); K. oxytoca (1)
P. aeruginosa, S. marcescens, S. maltophilia11S. marcescens (1); S. maltophilia (1)
E. coli, P. aeruginosa, S. maltophilia11S. maltophilia (1)
P. aeruginosa, S. aureus, S. maltophilia11S. aureus (1); S. maltophilia (1)
E. cloacae complex, S. aureus10
Acinetobacter spp., E. coli, P. aeruginosa, S. marcescens, S. aureus, S. maltophilia11Acinetobacter spp. (1); E. coli (1); S. aureus (1); S. maltophilia (1)
Multi-detections in Aspirate Specimens by LRTApplication aNumber of Specimenswith Multi-Detections byLRTNumber of Specimenswith Discordant ResultsFalse Positive Analytes:Microorganisms negative by SoCculture (LRT Positive/SoC Negative)(N) = Number of Specimens
Acinetobacter spp., Proteus spp., P. aeruginosa, S.aureus11Acinetobacter spp. (1); Proteus spp. (1);P. aeruginosa (1)
K. oxytoca, P. aeruginosa11K. oxytoca (1); P. aeruginosa (1)
E. coli, K. pneumoniae, P. aeruginosa, S. aureus11E. coli (1); K. pneumoniae (1)
Acinetobacter spp., E. coli, K. pneumoniae, Proteusspp., P. aeruginosa, S. aureus11Acinetobacter spp. (1); E. coli (1);Proteus spp. (1); S. aureus (1)
Acinetobacter spp., S. marcescens, S. maltophilia11Acinetobacter spp. (1)
E. coli, Proteus spp., P. aeruginosa, S. marcescens11E. coli (1)
K. oxytoca, S. aureus11K. oxytoca (1)
K. oxytoca, P. aeruginosa, S. maltophilia10
Proteus spp., P. aeruginosa, S. marcescens, S.maltophilia11Proteus spp. (1); S. marcescens (1); S.maltophilia (1)
E. coli, H. influenzae10
Acinetobacter spp., K. pneumoniae, S. aureus, S.maltophilia11K. pneumoniae (1)
Acinetobacter spp., M. morganii, Proteus spp., S.aureus11Acinetobacter spp. (1); M. morganii (1)
H. influenzae, S. pneumoniae10
Acinetobacter spp., E. coli, K. pneumoniae, P.aeruginosa11Acinetobacter spp. (1); E. coli (1); K.pneumoniae (1)
K. pneumoniae, K. variicola11K. pneumoniae (1); K. variicola (1)
P. aeruginosa, S. marcescens10
K. oxytoca, K. pneumoniae, M. morganii, P.aeruginosa, S. marcescens11K. oxytoca (1); M. morganii (1); P.aeruginosa (1); S. marcescens (1)
Acinetobacter spp., E. coli, P. aeruginosa, S.marcescens, S. maltophilia11Acinetobacter spp. (1); E. coli (1); S.marcescens (1); S. maltophilia (1)
M. catarrhalis, Proteus spp., P. aeruginosa, S.marcescens11M. catarrhalis (1); Proteus spp. (1); P.aeruginosa (1); S. marcescens (1)
H. influenzae, M. catarrhalis11M. catarrhalis (1)
K. pneumoniae, P. aeruginosa11K. pneumoniae (1); P. aeruginosa (1)
C. freundii, K. oxytoca, K. pneumoniae, S. aureus11C. freundii (1); K. pneumoniae (1)
Acinetobacter spp., Proteus spp., S. aureus11Proteus spp. (1)
K. variicola, S. aureus11S. aureus (1)

{41}------------------------------------------------

{42}------------------------------------------------

a Multidetections 'atypical' analytes: This table does not include three specimens with multi-detections including 'atypical' microorganisms:

M. pneumoniae - two specimens positive for M. pneumoniae with another analyte (one H. influenzae, and one . S. pneumoniae), of which one specimen was FP for H. influenzae and one specimen was FP for S. pneumoniae as compared to culture. Both specimens were concordant for M. pneumoniae as compared to the comparator

. L. pneumophila - one specimen positive for L. pneumophila/S. aureus detection concordant with culture and L. pneumophila concordant with the composite comparator)

{43}------------------------------------------------

Table 26: Multiple 'typical' microorganisms reported by culture (N=62), as compared to LRT Application

Multi-Detections in Aspirate Specimens reportedby SoC cultureNumber of Specimens withMulti-Detections by SoCcultureTotal Number ofSpecimens with DiscordantResultsFalse Negative Analytes:Microorganisms negative by LRT(SoC positive/LRT negative) (N) =number of specimens
K. pneumoniae, S. aureus51K. pneumoniae (1)
E. coli, S. aureus50
P. aeruginosa, S. maltophilia40
P. aeruginosa, S. aureus30
E. coli, P. aeruginosa31P. aeruginosa (1)
P. aeruginosa, S. marcescens30
Proteus spp., P. aeruginosa, S. marcescens20
Proteus spp., S. aureus20
S. marcescens, S. maltophilia20
K. oxytoca, S. aureus21K. oxytoca (1)
Proteus spp., P. aeruginosa20
K. pneumoniae, P. aeruginosa20
S. aureus, S. maltophilia20
E. cloacae complex, E. coli20
E. cloacae complex, K. pneumoniae11K. pneumoniae (1)
C. freundii, Proteus spp.11C. freundii (1); Proteus spp. (1)
K. oxytoca, S. maltophilia10
Acinetobacter spp., P. aeruginosa, S. maltophilia, S. pneumoniae10
H. influenzae, S. aureus10
E. coli, K. pneumoniae, Proteus spp.10
Acinetobacter spp., S. aureus, S. maltophilia10
E. cloacae complex, S. aureus10
Acinetobacter spp., P. aeruginosa11P. aeruginosa (1)
K. pneumoniae, Proteus spp.10
M. catarrhalis, S. aureus10
S. marcescens, S. aureus, S. maltophilia10
E. coli, H. influenzae10
H. influenzae, S. pneumoniae10
Acinetobacter spp., S. aureus10
E. cloacae complex, P. aeruginosa10
C. freundii, P. aeruginosa11C. freundii (1)
K. oxytoca, P. aeruginosa, S. maltophilia10
E. coli, K. pneumoniae10
Multi-Detections in Aspirate Specimens reported by SoC cultureNumber of Specimens with Multi-Detections by SoC cultureTotal Number of Specimens with Discordant ResultsFalse Negative Analytes: Microorganisms negative by LRT (SoC positive/LRT negative) (N) = number of specimens
K. pneumoniae, S. maltophilia10
E. coli, S. maltophilia10
C. freundii, S. pneumoniae11C. freundii (1)
M. catarrhalis, S. pneumoniae11M. catarrhalis (1); S. pneumoniae (1)

{44}------------------------------------------------

Retrospective Clinical Study, Archived Specimens

For the retrospective clinical study, 266 previously frozen tracheal aspirate specimens were tested at US study sites with the Unyvero LRT Application. Four specimens were excluded for not meeting specimen inclusion criteria. 32 specimens were excluded due to non-reportable results and 19 specimens with reportable results were excluded due to partially valid results. The remaining 211 evaluable US study specimens were supplemented with 158 specimens collected at other US or European sites and tested inhouse at Curetis. Altogether, results from a total number of 369 evaluable archived specimens were included in the performance analyses. All specimens were selected based on positive standard of care results which included culture for 'typical' analytes and other test methods for 'atypical' analytes. All historical positive results were confirmed with validated PCR/bi-directional sequencing assays prior to inclusion in the study.

Tables 27 and 28 include results for the retrospective clinical study for 'typical' and 'atypical' analytes respectively. Note that a standard reference method for the 'atypical' analytes was not applied to all archived specimens; therefore, only positive percent agreement could be calculated for these analytes.

{45}------------------------------------------------

PositivityTP/(TP+FN)PPA[%](95% CI)Negativity bTN/(TN+FP)NPA[%](95% CI)
Acinetobacter spp.18/18100.0(82.4 - 100.0)344/350c98.3(96.3 - 99.2)
Citrobacter freundii2/2100.0(34.2 - 100.0)354/36098.3(96.4 - 99.2)
Enterobacter cloacae complex24/2596.0(80.5 - 99.3)329/33697.9(95.8 - 99.0)
Escherichia coli37/3897.4(86.5 - 99.5)311/32994.5(91.5 - 96.5)
Haemophilus influenzae21/2487.5(69.0 - 95.7)326/33796.7(94.3 - 98.2)
Klebsiella oxytoca17/17100.0(81.6 - 100.0)337/34797.1(94.8 - 98.4)
Klebsiella pneumoniae a28/2996.6(82.8 - 99.4)321/33895.0(92.1 - 96.8)
Klebsiella variicola a9/9100.0(70.1 - 100.0)356/35899.4(98.0 - 99.8)
Moraxella catarrhalis9/9100.0(70.1 - 100.0)336/34597.4(95.1 - 98.6)
Morganella morganii1/1100.0(20.7 - 100.0)346/35398.0(96.0 - 99.0)
Proteus spp.29/3096.7(83.3 - 99.4)325/33896.2(93.5 - 97.7)
Pseudomonas aeruginosa48/5292.3(81.8 - 97.0)286/31092.3(88.7 - 94.7)
Serratia marcescens34/3597.1(85.5 - 99.5)326/33397.9(95.7 - 99.0)
Staphylococcus aureus72/7892.3(84.2 - 96.4)259/28690.6(86.6 - 93.4)
Stenotrophomonas maltophilia33/3497.1(85.1 - 99.5)298/32890.9(87.2 - 93.5)
Streptococcus pneumoniae21/2391.3(73.2 - 97.6)335/34198.2(96.2 - 99.2)

Table 27: Archived study performance, 'typical' microorganisms

4 As K. variicola is often reported by culture as K. pneumoniae, DNA extracts for culture positive K. pneumoniae specimens were sequenced. For nine of 38 K. pneumoniae positive specimens a sequencing result for K. variicola was obtained and K. variicola was assigned as reference identity.

b Specimens with FP results obtained in 211 US study specimens were analyzed by molecular tests (PCR bi-directional sequencing). Sequencing confirmed the presence of microorganisms in FP specimens as follows: Acinetobacter spp. 6 of 6, C. freundii 3 of 5, E. cloacae complex 6 of 7, E. coli 13 of 13, H. influenzae 4 of 6 (2 non-confirmed specimens were identified as H. haemolyticus), K. oxytoca 8 of 8, K. pneumoniae 13 of 13, K. variicola 2 of 3, M. morganii 6 of 7, M. pneumoniae 0 of 1, Proteus spp. 10 of 10, P. aeruginosa 15 of 16, S. aureus 13 of 13, S. maltophilia 19 of 19, S. pneumoniae 1 of 1.

E Note that for supplementary specimens partially valid results were included, therefore, the total number of data points for each microorganism analyte may differ.

{46}------------------------------------------------

PositivityTP/(TP+FN)PPA[%](95 % CI)
Chlamydia pneumoniae0/0na
Legionella pneumophila2/2100.0(34.2 - 100.0)
Mycoplasma pneumoniae0/0na

Table 28: Archived study performance, 'atypical' microorganisms

Clinical Study, Contrived Specimens:

For low prevalence microorganism and resistance marker analytes, the prospective and archived studies were supplemented with contrived specimens. Each contrived specimen was prepared in a unique natural tracheal aspirate specimen matrix. All aspirate specimen matrices were prescreened to ensure that they were negative for all Unyvero LRT analytes prior to use in the study. Specimens were spiked with pools of microorganisms at two concentrations: low positive (2x LoD or lower) and moderate positive (typically 3-10x LoD). Microorganisms evaluated in the contrived study were C. pneumoniae, C. freundii, K. oxytoca, K. variicola, L. pneumophila, M. catarrhalis, M. morganii, and M. pneumoniae. LRT Application resistance markers evaluated in the contrived study were ctx-M, oxa-23, oxa-24, oxa-48, oxa-58, kpc, vim, and ndm. Testing of contrived specimens was performed at three sites in the United States as well as in-house at Curetis.

Up to five different strains for each analyte were used to prepared contrived specimens with the total number of specimens ranging from 21 to 50 for each microorganism or resistance marker.

{47}------------------------------------------------

Table 29 below shows positivity rates (number of positive LRT results / number of expected positive results, PPA) and negativity rates (number of negative LRT results / number of expected negative results, NPA) observed in the study.

For other LRT panel microorganisms not evaluated in the contrived study, the following false positive results were observed: 1 of 292 for Proteus spp. (NPA: 99.7%, 95% CI: 98.1% – 99.9%); 4 of 216 for P. aeruginosa (NPA: 98.1%, 95% CI: 95.3% – 99.3%); 2 of 297 for S. marcescens (NPA: 99.3%, 95% CI: 97.6% - 99.8%); 2 of 258 for S. aureus (NPA: 99.2%, 95% CI: 97.2% – 99.8%); and 1 of 291 for S. pneumoniae (NPA: 99.7%, 95% Cl: 98.1% - 99.9%). Additional false positive results were observed for E. cloacae complex: (21 of 266, NPA: 92.1% 95% CI: 88.2% - 94.8%); for E. coli (6 of 205, NPA: 97.1%, 95% CI: 93.8% – 98.7%); and S. maltophilia (10 of 317, NPA: 96.8%, 95% CI: 94.3% - 98.3%) with false positive results linked to contamination of test materials for 19 of 21 (E. cloacae complex), 5 of 6 (E. coli), and 10 of 10 (S. maltophilia) positive results.

Results from the contrived study are presented in Table 29.

{48}------------------------------------------------

Table 29: Contrived specimen testing

Analytestrain IDs (# tests) aConcentration[CFU/mL],(x LoD)Positivity(# positive/# expected)PPA [%](95% CI)Negativity(# negative/# notexpected)NPA [%](95% CI)
1.5 x 104 (1x)14/14100.0(78.5 - 100.0)
Chlamydia pneumoniaeATCC VR2282 (in IFU/ml) (21) b, c4.5 x 104 (3x)7/7100.0(64.6 - 100.0)
total21/21100.0(84.5 - 100.0)303/303100.0(98.7 - 100.0)
4 x 105 (2x)21/2584.0(65.3 - 93.6)
Citrobacter freundiiATCC 8090 (20), ATCC 43864 (20), NRZ-00452 (10)1 x 106 (5x)22/2588.0(70.0 - 95.8)
total43/5086.0(73.8 - 93.1)236/23799.6(97.6 - 99.9)
Klebsiella oxytoca8 x 104 (0.4x)12/1485.7(60.1 - 96.0)
ATCC 13182 (6), ATCC 49131 (6),ATCC 43863 (6), NCIMB 12819 (5),NRZ-22060 (5)4 x 105 (2x)13/1492.9(68.5 - 98.7)
total25/2889.3(72.8 - 96.3)294/29599.7(98.1 - 99.9)
2 x 105 (2x)12/1392.3(66.7 - 98.6)
Klebsiella variicolaATCC BAA-830 (28)1 x 106 (10x)15/15100.0(79.6 - 100.0)
total27/2896.4(82.3 - 99.4)295/29699.7(98.1 - 99.9)
Legionella pneumophila4 x 106 (2x)24/24100.0(86.2 - 100.0)
ATCC 33154 (9), ATCC 33215 (10),ATCC 35096 (10), ATCC 43283 (10),ATCC 33155 (10)1 x 107 (5x)25/25100.0(86.7 - 100.0)
total49/49100.0(92.7 - 100.0)247/247100.0(98.5 - 100.0)
1.6 x 106 (2x)24/2596.0(80.5 - 99.3)
Moraxella catarrhalisATCC 25238 (10), ATCC 43617 (20),ATCC 8176 (10), ATCC 25240 (10)5 x 106 (6x)24/2596.0(80.5 - 99.3)
total48/5096.0(86.5 - 98.9)245/245100.0(98.5 - 100.0)
Morganella morganii1 x 106 (2x)24/24100.0(86.2 - 100.0)
ATCC 25830 (20), ATCC8019 (10),ATCC 25829 (9), DSM-46262 (10)3 x 106 (6x)23/2592.0(75.0 - 97.8)
total47/4995.6(83.3 - 98.9)245/245100.0(98.5 - 100.0)
Mycoplasma pneumoniae2 x 105 (2x)25/25100.0(86.7 - 100.0)
ATCC 29085 (20), ATCC 29343 (20),(in CCU/mL),5 x 105 (5x)25/25100.0(86.7 - 100.0)
ATCC 15492 (10) (in CFU/mL) dtotal50/50100.0(92.9 - 100.0)274/274100.0(98.6 - 100.0)
2 x 105 (2x)23/2592.0(75.0 - 97.8)
Analytestrain IDs (# tests) aConcentration[CFU/mL],(x LoD)Positivity(# positive/# expected)PPA [%](95 % CI)Negativity(# negative/# notexpected)NPA [%](95 % CI)
(75.0 - 97.8)
ctx-MNRZ-00751, NRZ-00002, NRZ-00249,(K. pneumoniae), JMI 46239 (E. cloacae),JMI 50067 (E. coli) (10 each)6 x 105 (6x)24/2596.0(80.5 - 99.3)
total47/5094.0(83.8 - 97.9)183/192 e95.3(91.3 - 97.5)
kpc1 x 106 (2x)23/2495.8(79.8 - 99.3)
NRZ-00222 (10), NRZ-00281 (10) ( E. coli ),NRZ-00223 (9), Micromyx 4653 (10),Micromyx 4676 ( K. pneumoniae ) (10)3 x 106 (6x)25/25100.0(86.7 - 100.0)
total48/4998.0(89.3 - 99.6)245/245100.0(98.5 - 100.0)
ndmJMI 50067 (6) ( E. coli ), NCTC 13443 (6),JMI 49831 (6) ( K. pneumoniae ), JMI 49755(5) ( A. baumannii ), JMI 46239 (6) ( E. cloacae )1 x 105 (2x)13/1492.9(68.5 - 98.7)
5 x 105 (10x)15/15100.0(79.6 - 100.0)
total28/2996.6(82.8 - 99.4)237/263 e90.1(85.9 - 93.2)
oxa-23Micromyx 4410 (6), Micromyx 6148 (6),Micromyx 6149 (4), Micromyx 6153 (6),UCLA A5 (6) ( A. baumannii ) (6 each)1 x 107 (0,5x)13/1492.9(68.5 - 98.7)
2 x 107 (1x)14/14100.0(78.5 - 100.0)
total27/2896.4(82.3 - 99.4)139/14099.3(96.1 - 99.9)
oxa-24NCTC 13302, NRZ-00449, UCLA A4, twoclinical isolates ( A. baumannii ) (10 each)6 x 104 (0,1x)25/25100.0(86.7 - 100.0)
3 x 105 (0,6x)25/25100.0(86.7 - 100.0)
total50/50100.0(92.9 - 100.0)90/90100.0(95.9 - 100.0)
oxa-48NRZ-00176 (19), ATCC BAA-2523 (10) ( E. coli ), NCTC 13442 (10), NRZ-00002 (10)( K. pneumoniae )4 x 106 (2x)24/24100.0(86.2 - 100.0)
1 x 107 (5x)24/2596.0(80.5 - 99.3)
total48/4998.0(89.3 - 99.6)205/205100.0(98.2 - 100.0)
oxa-58NCTC 13305 (18), NRZ-00518 (12)( A. baumannii )4 x 105 (0,5x)15/15100.0(79.6 - 100.0)
1 x 106 (1.3x)15/15100.0(79.6 - 100.0)
total30/30100.0(88.7 - 100.0)138/138100.0(97.3 - 100.0)
vimNRZ-00452 (10) ( C. freundii ), NRZ-00239(20) ( E. cloacae ),DSM-24600 (19)( P. aeruginosa )1 x 105 (2x)22/2491.7(74.2 - 97.7)
3 x 105 (6x)25/25100.0(86.7 - 100.0)
total47/4995.9(86.3 - 98.9)234/234100.0(98.4 - 100.0)

{49}------------------------------------------------

a total of 50 tests (for C. freundii, L. pneumophila, M. pneumoniae, cx-M, kpc, oxa-24, oxa-48. vin) or 30 tests (for K. oxytoca, K. variicola, M. morganii, ndm, oxa-23, oxa-58) or C. pneumoniae) was performed. Missing results were due to invalid test results.

{50}------------------------------------------------

0 Numbers in parentheses indicate number of tests performed for an individual strain

C IFU: inclusion-forming units

4 CCU: color-changing units, concentration used for ATCC 15492: 1 x 106 and 3 x 10° CFU/mL (1 CCU/mL was estimated to be equivalent to 10 CFU/mL)

° Eight positive ctx-M results and 24 positive ndm results were linked to contamination in test materials

Clinical Performance, Resistance Marker Targets

Performance characteristics for LRT Application antibiotic resistance marker targets were evaluated in the prospective study (603 aspirate specimens) and supplemented with results from contrived specimens (results shown in Table 29 above).

To assess the performance of the Unyvero LRT Application for detection of each resistance marker target, positive and negative percent agreement was calculated as compared to results of validated multiplex PCR assays followed by bi-directional sequencing.

It is noted that antibiotic resistance marker results are only reported by the LRT Application if at least one corresponding host microorganism is simultaneously detected. If an applicable microorganism is not detected in the specimen, positive resistance marker results are masked on the results screen (i.e., the result is masked and the report indicates N/A regardless whether the resistance marker is detected or not detected). Evaluation of assay performance for detection of resistance markers by the LRT Application was performed both with and without application of masking/reporting rules.

Table 30 includes performance of the LRT assay for detection of resistance marker targets as observed in the prospective study and compared to PCR/bi-directional sequencing. Analysis includes all positive resistance marker results (i.e., without software masking).

{51}------------------------------------------------

Table 30: Prospective Study, Resistance marker performance as compared to molecular comparator assays (multiplex PCR followed by bi-directional sequencing), without application of software masking

PositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M15/1693.8(71.7 - 98.9)584/58799.5(98.5 - 99.8)
kpc7/7100.0(64.6 - 100.0)596/596100.0(99.4 - 100.0)
ndm0/0na603/603100.0(99.4 - 100.0)
oxa-236/785.7(48.7 - 97.4)594/59699.7(98.8 - 99.9)
oxa-242/366.7(20.8 - 93.9)600/600100.0(99.4 - 100.0)
oxa-480/0na602/60399.8(99.1 - 100.0)
oxa-580/0na603/603100.0(99.4 - 100.0)
tem54/54100.0(93.4 - 100.0)537/54997.8(96.2 - 98.7)
vim2/2100.0(34.2 - 100.0)601/601100.0(99.4 - 100.0)
mecA108/12487.1(80.1 - 91.9)453/47994.6(92.2 - 96.3)

Table 31 shows LRT performance in the prospective study for each targeted resistance marker based on comparison to molecular comparator assays (PCR/bi-directional sequencing) for only those specimens in which an applicable LRT microorganism target was detected by the LRT assay (i.e., results shown are after application of software masking). Based on the reporting rules for the Unyvero LRT assay, performance for ctx-M, blaxec, blands and blaym includes specimens that were positive by the Unyvero LRT Application for a targeted microorganism of the Enterobacteriaceae, Acinetobacter spp. and/or P. geruginosa. For oxa-48. performance includes only those specimens that were positive for one or more of the Enterobacteriaceae targets. For oxa-23, oxa-24 and oxa-58. the performance includes only those specimens that were positive for Acinetobacter spp. For blaTEM, performance includes only those specimens that were positive for H. influenzae. For mecA, performance includes only those specimens that were positive for S. aureus.

{52}------------------------------------------------

Table 31: Prospective Study, Resistance marker performance as compared to molecular comparator assays (multiplex PCR followed by bi-directional sequencing), with application of masking.

PositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)Not reported(masked)
ctx-M15/1693.8(71.7 - 98.9)177/17998.9(96.0 - 99.7)408
kpc6/6100.0(61.0 - 100.0)189/189100.0(98.0 - 100.0)408
ndm0/0na195/195100.0(98.1 - 100.0)408
oxa-236/785.7(48.7 - 97.4)18/2090.0(69.9 - 97.2)576
oxa-242/2100.0(34.2 - 100.0)25/25100.0(86.7 - 100.0)576
oxa-480/0na138/13999.3(96.0 - 99.9)464
oxa-580/0na27/27100.0(87.5 - 100.0)576
tem8/8100.0(67.6 - 100.0)16/16100.0(80.6 - 100.0)579
vim2/2100.0(34.2 - 100.0)193/193100.0(98.0 - 100.0)408
mecA54/5991.5(81.6 - 96.3)53/6186.9(76.2 - 93.2)483

Tables 32-41 include performance for resistance marker targets for each applicable microorganism detected by the LRT assay. Performance for each resistance marker target is evaluated as compared to PCR/bi-directional sequencing. Each table includes only the subset of specimens that are positive by LRT for the specified microorganism target; therefore, the results shown are after application of software masking.

{53}------------------------------------------------

Table 32: Prospective Study, Resistance marker performance as compared molecular comparator assays, stratified for LRT positive samples for Acinetobacter spp. (N=27). Note that detection of each resistance marker cannot be definitively linked to Acinetobacter spp.

Acinetobacterspp.PositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M5/5100.0(56.6 - 100.0)21/2295.5(78.2 - 99.2)
kpc2/2100.0(34.2 - 100.0)25/25100.0(86.7 - 100.0)
ndm0/0na27/27100.0(87.5 - 100.0)
oxa-236/785.7(48.7 - 97.4)18/2090.0(69.9 - 97.2)
oxa-242/2100.0(34.2 - 100.0)25/25100.0(86.7 - 100.0)
oxa-580/0na27/27100.0(87.5 - 100.0)
vim1/1100.0(20.7 - 100.0)26/26100.0(87.1 - 100.0)

Table 33: Prospective Study, Resistance marker performance as compared to molecular comparator assays, stratified for LRT positive samples for Citrobacter freundii (N=3)

CitrobacterfreundiiPositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M0/0na3/3100.0(43.9 - 100.0)
kpc0/0na3/3100.0(43.9 - 100.0)
ndm0/0na3/3100.0(43.9 - 100.0)
oxa-480/0na3/3100.0(43.9 - 100.0)
vim0/0na3/3100.0(43.9 - 100.0)

{54}------------------------------------------------

Table 34: Prospective Study, Resistance marker performance as compared to molecular comparator assays, stratified for LRT positive samples for Enterobacter cloacae complex (N=20). Note that detection of each resistance marker cannot be definitively linked to E. cloacae complex.

EnterobactercloacaecomplexPositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M0/0na20/20100.0(83.9 - 100.0)
kpc1/1100.0(20.7 - 100.0)19/19100.0(83.2 - 100.0)
ndm0/0na20/20100.0(83.9 - 100.0)
oxa-480/0na20/20100.0(83.9 - 100.0)
vim0/0na20/20100.0(83.9 - 100.0)

Table 35: Prospective Study, Resistance marker performance as compared to molecular comparator assays, stratified for LRT positive samples for K. oxytoca (N=15). Note that detection of each resistance marker cannot be definitively linked to K. oxytoca.

KlebsiellaoxytocaPositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M0/0na14/1593.3(70.2 - 98.8)
kpc0/0na15/15100.0(79.6 - 100.0)
ndm0/0na15/15100.0(79.6 - 100.0)
oxa-480/0na15/15100.0(79.6 - 100.0)
vim0/0na15/15100.0(79.6 - 100.0)

{55}------------------------------------------------

Table 36: Prospective Study, Resistance marker performance as compared to molecular comparator assays, stratified for LRT positive samples for K. pneumoniae (N=38). Note that detection of each resistance marker cannot be definitively linked to K. pneumoniae.

KlebsiellapneumoniaePositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M8/8100.0(67.6 - 100.0)29/3096.7(83.3 - 99.4)
kpc2/2100.0(34.2 - 100.0)36/36100.0(90.4 - 100.0)
ndm0/0na38/38100.0(90.8 - 100.0)
oxa-480/0na38/38100.0(90.8 - 100.0)
vim0/0na38/38100.0(90.8 - 100.0)

Table 37: Prospective Study, Resistance marker performance as compared to molecular comparator assays, stratified for LRT positive samples for K. variicola (N=4). Note that detection of each resistance marker cannot be definitively linked to K. variicola.

KlebsiellavariicolaPositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M0/0na4/4100.0(51.0 - 100.0)
kpc0/0na4/4100.0(51.0 - 100.0)
ndm0/0na4/4100.0(51.0 - 100.0)
oxa-480/0na4/4100.0(51.0 - 100.0)
vim0/0na4/4100.0(51.0 - 100.0)

{56}------------------------------------------------

Table 38: Prospective Study, Resistance marker performance as compared to molecular comparator assays, stratified for LRT positive samples for Morganella morganii (N=10). Note that detection of each resistance marker cannot be definitively linked to Morganella morganii.

MorganellamorganiiPositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M2/2100.0(34.2 - 100.0)8/8100.0(76.6- 100.0)
kpc2/2100.0(34.2 - 100.0)8/8100.0(67.6 - 100.0)
ndm0/0na10/10100.0(72.3 - 100.0)
oxa-480/0na9/1090.0(59.6 - 98.2)
vim1/1100.0(20.7 - 100.0)9/9100.0(70.1 - 100.0)

Table 39: Prospective Study, Resistance marker performance as compared to molecular comparator assays, stratified for LRT positive samples for Proteus spp. (N=29). Note that detection of each resistance marker cannot be definitively linked to Proteus spp.

Proteusspp.PositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M3/3100.0(43.9 - 100.0)26/26100.0(87.1 - 100.0)
kpc1/1100.0(20.7 - 100.0)28/28100.0(87.9 - 100.0)
ndm0/0na29/29100.0(88.3 - 100.0)
oxa-480/0na29/29100.0(88.3 - 100.0)
vim1/1100.0(20.7 - 100.0)28/28100.0(87.9 - 100.0)

{57}------------------------------------------------

Table 40: Prospective Study, Resistance marker performance as compared to molecular comparator assays, stratified for LRT positive samples for Pseudomonas aeruginosa (N=84). Note that detection of each resistance marker cannot be definitively linked to Pseudomonas aeruginosa.

PseudomonasaeruginosaPositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M9/9100.0(70.1 - 100.0)75/75100.0(95.1 - 100.0)
kpc3/3100.0(43.9 - 100.0)81/81100.0(95.5 - 100.0)
ndm0/0na84/84100.0(95.6 - 100.0)
oxa-480/0na83/8498.8(93.6 - 99.8)
vim2/2100.0(34.2 - 100.0)82/82100.0(95.5 - 100.0)

Table 41: Prospective Study, Resistance marker performance as compared to molecular comparator assays, stratified for LRT positive samples for Serratia marcescens (N=25). Note that detection of each resistance marker cannot be definitively linked Serratia marcescens.

SerratiamarcescensPositivityTP / (TP+FN)PPA[%](95 % CI)NegativityTN / (TN+FP)NPA[%](95 % CI)
ctx-M1/1100.0(20.7 - 100.0)24/24100.0(86.2 - 100.0)
kpc4/4100.0(51.0 - 100.0)21/21100.0(84.5 - 100.0)
ndm0/0na25/25100.0(86.7 - 100.0)
oxa-480/0na25/25100.0(86.7 - 100.0)
vim0/0na25/25100.0(86.7 - 100.0)

{58}------------------------------------------------

For the prospective study, additional analyses of clinical performance for detection of resistance marker targets was conducted in combination with results from microorganism detection. For each marker, two different 3x3 tables were generated; one 3x3 table with Comparator 'A', reflecting culture as the reference method for LRT microorganism targets and one 3x3 table with Comparator 'B', reflecting the composite comparator method (culture plus PCR/bi-directional sequencing) for LRT microorganism targets. For the resistance markers, the comparator method was PCR/bi-directional sequencing for all analyses.

Agreement rates were determined for the following resistance marker/microorganism combinations following the Unyvero LRT reporting rules:

  • tem: H. influenzae
  • ctx-M, kpc, vim: All Enterobacteriaceae targets, Acinetobacter spp., and P. . aeruginosa combined
  • oxa-48: all Enterobacteriaceae targets combined .
  • oxa-23. oxa-24: Acinetobacter spp. .
  • mecA: S. aureus .

Numbers of available culture isolates, results of linkage analysis (confirmation that the host microorganism strain is the source of the antibiotic resistance marker determined by PCR/bi-directional sequencing of culture isolates), and the number of multi-detection samples (more than one host microorganism detected by the composite comparator (comparator A) or SoC/culture (comparator B) are indicated as footnotes to each of the agreement tables. Agreement tables are not presented for ndm and oxa-58 as there were no positive results for these analytes by either the LRT or comparator assays.

Results are shown in Tables 42 through 49, with data presented in two tables for each resistance marker (one with Comparator A and one with Comparator B).

{59}------------------------------------------------

H. influenzaectemComposite Comparator Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+aOrg+/Res+bOrg+/Res-Org-total
UnyveroResultOrg+/Res+6028
Org+/Res-09615
Org-11577579
total710585602
Haemophilus influenzae/temrate [%]positivity95 % CI
Agreement (Org+/Res+)85.76/748.7 - 97.4
Agreement (Org+/Res-)90.09/1059.6 - 98.2
Agreement (Org-)98.6577/58597.3 - 99.3

Table 42: Agreement rates for tem between LRT and comparator methods A and B, (single) host microorganism: H. influenzae

4 isolate linkage: 3 culture positive cases, 2 isolates [2 of 2 confirmed]

b multi-detection specimens: not applicable (tem reported only with H. influenzae)

& note that other LRT panel microorganisms (Enterobacter spp., P. aeruginosa) could be the source of tem

B

H. influenzaectemCulture Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+bOrg+/Res-Org-total
Org+/Res+a3058
UnyveroResultOrg+/Res-051116
Org-00579579
total35595603
Haemophilus influenzae/temrate [%]positivity95 % CI
Agreement (Org+/Res+)1003/343.9 - 100.0
Agreement (Org+/Res-)1005/556.6 - 100.0
Agreement (Org-)97.3579/59595.7 - 98.3

ª Isolate linkage: 2 isolates [2 of 2 confirmed]

b Multi-detection specimens: not applicable (tem reported only with H. influenzae)

6 Note that other LRT panel microorganisms (Enterobacter spp., P. aeruginosa could be the source of tem

{60}------------------------------------------------

Corr. Host Microorganismctx-MComposite Comparator Result for Host Microorganism -PCR/Seq for Antibiotic Resistance Marker
Org+/Res+bOrg+/Res-Org-total
UnyveroResultOrg+/Res+a151117
Org+/Res-115620177
Org-016392408
total16173413602
Corr. host microorganism /ctx-Mrate [%]positivity95 % CI
Agreement (Org+/Res+)93.815/1671.7 - 98.9
Agreement (Org+/Res-)90.2156/17384.8 - 93.8
Agreement (Org-)94.9392/41392.4 - 96.7

Table 43: Agreement rates for ctx-M between LRT and comparator methods A and B, host microorganisms: Enterobacteriaceae, Acinetobacter spp., P. aeruginosa.

a At least one isolate available for 10 specimens [linkage confirmed for 5 of 10 specimens]

b Multi-detection specimens (two or more corresponding LRT host microorganisms): 9 of 16

B

Corr. Host Microorganismctx-MCulture Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+ bOrg+/Res-Org-total
UnyveroResultOrg+/Res+ a110617
Org+/Res-112849178
Org-010398408
total12138453603
Corr. host microorganism /ctx-Mrate [%]positivity95 % CI
Agreement (Org+/Res+)91.711/1264.6 - 98.5
Agreement (Org+/Res-)92.8128/13887.2 - 96.0
Agreement (Org-)87.9398/45384.5 - 90.6

a At least one isolate available for 10 specimens [linkage confirmed for 5 of 10 specimens]

b Multi-detection specimens (two or more corresponding LRT host microorganisms): 6 of 12

{61}------------------------------------------------

ACorr. Host MicroorganismkpcComposite Comparator Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+ bOrg+/Res-Org-total
Org+/Res+ a6006
UnyveroResultOrg+/Res-016721188
Org-016392408
total6183413602
Corr. host microorganism / kpcrate [%]positivity95 % CI
Agreement (Org+/Res+)100.06/661.0 - 100.0
Agreement (Org+/Res-)91.3167/18386.3 - 94.5
Agreement (Org-)94.9392/41392.4 - 96.7

Table 44: Agreement rates for kpc between LRT and comparator methods A and B, host microorganisms: Enterobacteriaceae, Acinetobacter spp., P. aeruginosa.

a Sample linkage: at least one isolate available for 5 specimens [linkage confirmed for 4 of 5 specimens] b Multi-detection specimens (two or more corresponding LRT host microorganisms): 4 of 6

B

Corr. Host MicroorganismkpcCulture Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+ bOrg+/Res-Org-total
Org+/Res+ a6006
UnyveroOrg+/Res-013455189
ResultOrg-010398408
total6144453603
Corr. host microorganism /kpcrate [%]positivity95 % CI
Agreement (Org+/Res+)1006/661.0 - 100.0
Agreement (Org+/Res-)93.1134/14487.7 - 96.2
Agreement (Org-)87.9398/45384.5 - 90.6

a Sample linkage: at least one isolate available for 5 specimens [linkage confirmed for 4 of 5 specimens] b Multi-detection specimens (two or more corresponding LRT host microorganisms): 2 of 6

{62}------------------------------------------------

Corr. Host MicroorganismvimComposite Comparator Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+bOrg+/Res-Org-total
UnyveroResultOrg+/Res+a2002
Org+/Res-017121192
Org-016392408
total2187413602
Corr. host microorganism/ vimrate [%]positivity95 % CI
Agreement (Org+/Res+)100.02/234.2 - 100.0
Agreement (Org+/Res-)91.4171/18786.6 - 94.7
Agreement (Org-)94.9392/41392.4 - 96.7

Table 45: Agreement rates for vim between LRT and comparator methods A and B, host microorganisms: Enterobacteriaceae, Acinetobacter spp., P. aeruginosa.

a Sample linkage: at least 1 isolate available for 2 specimens [linkage confirmed for 2 of 2 specimens] b Multi-detection specimens (two or more corresponding LRT host microorganisms): 1 of 2

B

Corr. Host MicroorganismvimCulture Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+bOrg+/Res-Org-total
UnyveroResultOrg+/Res+a2002
Org+/Res-013855193
Org-010398408
total2148453603
Corr. host microorganism/vimrate [%]positivity95 % CI
Agreement (Org+/Res+)100.02/234.2 - 100.0
Agreement (Org+/Res-)93.2138/14888.0 - 96.3
Agreement (Org-)87.9398/45384.5 - 90.6

a Sample linkage: at least 1 isolate available for 2 specimens [linkage confirmed for 2 of 2 specimens] b Multi-detection specimens (two or more corresponding LRT host microorganisms): 1 of 2

{63}------------------------------------------------

Table 46: Agreement rates for oxa-48 between LRT and comparator methods A and
B, host microorganisms: Enterobacteriaceae.
ACorr. Host Microorganismoxa-48Composite Comparator Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+bOrg+/Res-Org-total
UnyveroResultOrg+/Res+a0011
Org+/Res-011919138
Org-010453463
total0129473602
Corr. host microorganism/oxa-48rate [%]positivity95 % CI
Agreement (Org+/Res+)na0/0na
Agreement (Org+/Res-)92.2119/12986.3 - 95.7
Agreement (Org-)95.8453/47393.6 - 97.2

ª Sample linkage: N/A, no culture positive specimens, no isolates

b Multi-detection specimens (two or more corresponding host microorganisms): none

B

Corr. Host Microorganismoxa-48Culture Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+ bOrg+/Res-Org-total
UnyveroResultOrg+/Res+ a0011
Org+/Res-08059139
Org-08455463
total088515603
Corr. host microorganism/oxa-48rate [%]positivity95 % CI
Agreement (Org+/Res+)na0/0na
Agreement (Org+/Res-)90.980/8883.1 - 95.3
Agreement (Org-)88.3455/51585.3 - 90.8

^a Sample linkage: N/A, no culture positive specimens, no isolates

ª Sample linkage: N/A, no culture positive specimens, no isolates
b Multi-detection specimens (two or more corresponding host microorganisms): none

{64}------------------------------------------------

AAcinetobacter spp.oxa-23Composite Comparator Result for Host MicroorganismPCR/Seq for Antibiotic Resistance Marker
Org+/Res+bOrg+/Res-Org-total
Org+/Res+a6208
UnyveroResultOrg+/Res-113519
Org-01575576
total716580603
Acinetobacter spp./oxa-23rate [%]positivity95 % CI
Agreement (Org+/Res+)85.76/748.7 - 97.4
Agreement (Org+/Res-)81.313/1657.0 - 93.4
Agreement (Org-)99.1575/58098.0 - 99.6

Table 47: Agreement rates for oxa-23 between LRT and comparator methods A and B, (single) host microorganism: Acinetobacter spp.

a Isolate linkage: 2 culture positive specimens, 2 isolates [2 of 2 confirmed]

b Multi-detection specimens: not applicable (oxa-23 reported only with Acinetobacter spp.)

Acinetobacter spp.oxa-23Culture Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+ bOrg+/Res-Org-total
UnyveroResultOrg+/Res+ a2068
Org+/Res-171119
Org-00576576
total37593603
Acinetobacter spp./oxa-23rate [%]positivity95 % CI
Agreement (Org+/Res+)66.72/320.8 - 93.9
Agreement (Org+/Res-)100.07/764.6 - 100.0
Agreement (Org-)97.1576/59395.5 - 98.2

4 Isolate linkage: 2 isolates [2 of 2 confirmed]

B

b Multi-detection specimens: not applicable (oxa-23 reported only with Acinetobacter spp.)

{65}------------------------------------------------

Acinetobacter spp.Composite Comparator Result for Host Microorganism –PCR/Seq for Antibiotic Resistance Marker
oxa-24Org+/Res+ bOrg+/Res-Org-total
Org+/Res+ a2002
UnyveroOrg+/Res-020525
ResultOrg-01575576
total221580603
Acinetobacter spp./oxa-24rate [%]positivity95 % CI
Agreement (Org+/Res+)100.02/234.2 - 100.0
Agreement (Org+/Res-)95.220/2177.3 - 99.2
Agreement (Org-)99.1575/58098.0 - 99.6

Table 48: Agreement rates for oxa-24 between LRT and comparator methods A and B. host microorganisms: Enterobacteriaceae. Acinetobacter spp., P. geruginosa.

alisolate linkage: 1 culture positive specimen, 1 isolate [1 of 1 confirmed]

b Multi-detection specimens: not applicable (oxa-24 reported only with Acinetobacter spp.)

B

Acinetobacter spp.oxa-24Culture Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+bOrg+/Res-Org-total
UnyveroResultOrg+/Res+a1012
Org+/Res-091625
Org-00576576
total19593603
Acinetobacter spp./oxa-24rate [%]positivity95 % CI
Agreement (Org+/Res+)100.01/120.7 - 100.0
Agreement (Org+/Res-)100.09/970.1 - 100.0
Agreement (Org-)97.1576/59395.5 - 98.2

ª Isolate linkage: 1 isolate [1 of 1 confirmed]

b Multi-detection specimens: not applicable (oxa-24 reported only with Acinetobacter spp.)

{66}------------------------------------------------

S. aureusmecAComposite Comparator Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+ bOrg+/Res-Org-total
UnyveroResultOrg+/Res+ a537262
Org+/Res-544958
Org-46473483
total6257484603
S. aureus/mecArate [%]positivity95 % CI
Agreement (Org+/Res+)85.553/6274.7 - 92.2
Agreement (Org+/Res-)77.244/5764.8 - 86.2
Agreement (Org-)97.7473/48496.0 - 98.7

Table 49. Agreement rates for mecA between LRT and comparator methods A and B, (single) host microorganism: S. aureus.

a Isolate linkage: 48 culture positive specimens, 36 isolates [29 of 36 confirmed] b Multi-detection specimens: not applicable (mecA only reported with S. aureus)

S. aureusmecACulture Result for Host Microorganism +PCR/Seq for Antibiotic Resistance Marker
Org+/Res+ bOrg+/Res-Org-total
UnyveroResultOrg+/Res+ a4441462
Org+/Res-4292558
Org-02481483
total4835520603
S. aureus/mecArate [%]positivity95 % CI
Agreement (Org+/Res+)91.744/4880.4 - 96.7
Agreement (Org+/Res-)82.929/3567.3 - 91.9
Agreement (Org-)92.5481/52089.9 - 94.5

4 Isolate linkage: 36 isolates [29 of 36 confirmed]

B

b Multi-detection specimens: not applicable (mecA reported only with S. aureus)

{67}------------------------------------------------

For the prospective study, clinical performance of the Unyvero LRT Application for detection of mecA was also compared to phenotypic antimicrobial susceptibility (AST) testing results for all S. aureus isolates recovered from the reference culture. Culture results were the reference method S. aureus detection and cefoxitin and/or oxacillin AST results were used as the phenotypic reference method for mecA. Results are shown in Table 50 below.

S. aureusmecACulture Result for S. aureus and Cefoxitin/Oxacillin ASTresults
SA+/Res+(MRSA)SA+/Res-(MSSA)SA-total
UnyveroResultSA+/mecA+4071461
SA+/mecA--1312557
SA-20481483
total4338520601
S. aureus/mecArate [%]positivity95 % CI
Agreement (SA+/Res+)93.040/4381.4-97.6
Agreement (SA+/Res-)81.631/3866.6-90.8
Agreement (SA-)92.5481/52089.9-94.5

Table 50: Agreement rates for S. aureus mecA between LRT and phenotypes of corresponding S. aureus strains.

A known limitation for detection of LRT resistance marker targets directly from tracheal aspirate specimens is that detection of markers cannot be definitively linked to corresponding detected microorganisms. For example, on-panel microorganisms present in the specimen that are not detected by LRT or off-panel microorganisms could serve as the source of a detected marker. In addition, tracheal aspirate specimens commonly contain multiple microorganisms, of which more than one microorganism could be the source of a detected resistance marker.

For the prospective and archived U.S. specimen cohorts, an analysis of genotypic 'linkage' as well as phenotypic agreement was performed to evaluate the relationship between detection of resistance markers by the LRT Application directly from tracheal aspirate specimens, the presence of the marker in corresponding culture isolates (based on PCR/bi-directional sequencing), and agreement with phenotypic AST results of reference culture isolates. Included in the analysis were specimens that were positive for targeted resistance markers by the LRT assay and were also 'true positive' for applicable LRT microorganism targets (i.e., positive by LRT and positive by culture for applicable microorganism targets). The analysis also included only those specimens for which isolates were available for confirmatory genotypic testing. For 'true positive' specimens, positive results for antibiotic resistance marker targets were compared to individual available isolates from SoC/culture positive specimens for:

{68}------------------------------------------------

    1. Genotypic Linkage: Genotypic linkage of positive LRT antibiotic resistance marker results to sequencing of cultured isolates (presence of the antibiotic resistance marker in the genome of one (or more) host microorganism strains isolated from a specific specimen, determined by PCR/bi-directional sequencing). Note that only a subset of isolates was available for this analysis. Confirmed genotypic linkage is defined as a positive LRT result for a resistance marker where a corresponding culture isolate is also positive by sequencing).
    1. Phenotypic Analysis: Agreement of positive LRT resistance marker results to associated phenotypic antimicrobial susceptibility testing (AST) results. AST results were collected for the antimicrobials listed in Table 51. Analysis of AST agreement with positive LRT results is presented in Tables 53-59 only for the subset of specimens to include those LRT positive specimens demonstrating genotypic linkage of the detected resistance marker.

For evaluation of phenotypic agreement of positive LRT resistance marker results. AST results for applicable antibiotics were collected and evaluated for all applicable reference culture isolates (Table 51). AST results were reported as MIC values or zone diameters (for Kirby-Bauer tests). AST results were determined using breakpoints listed in CLSI guidance M100S (Performance Standards for Antimicrobial Susceptibility Testing. 26th Edition 2016). "Intermediate" AST results were regarded as "resistant" and any strain was regarded as "resistant" if at least one of the corresponding drug AST results were "intermediate" or "resistant". Any strain was regarded as "susceptible" if AST results were susceptible for all applicable tested antibiotics.

{69}------------------------------------------------

Antibiotic resistancemarkerAssociated resistanceAST assay
temPenicillinsAmpicillin, Cefinase
ctx-M3rd Generation Cephalosporins andCefepimeEnterobacteriaceae:Cefotaxime, Ceftazidime, CeftriaxoneAcinetobacter spp.:Cefotaxime, Ceftazidime, Ceftriaxone,CefepimeP. aeruginosa :Ceftazidime, Cefepime
kpc, ndmCarbapenemsEnterobacteriaceae:Meropenem, Ertapenem, ImipenemAcinetobacter spp./P. aeruginosa :Meropenem, Imipenem
oxa-48Enterobacteriaceae: Meropenem, Ertapenem,Imipenem
oxa-23, oxa-24, andoxa-58Acinetobacter spp.:Meropenem, Imipenem
mecAOxacillinS. aureus :Oxacillin, Cefoxitin

Table 51: AST assays used for evaluation of antibiotic resistance markers detected by LRT

Evaluation of genotypic linkage and phenotypic agreement for resistance markers detected by the LRT Application was performed on a per specimen basis. For many specimens, more than one isolate with the potential to carry a detected resistance marker was recovered from the reference culture.

For specimens that were positive for LRT antibiotic resistance markers, the overall number and percent of specimens with confirmed genotypic linkage as observed in the archived (N=185) and prospective (N=603) studies combined is shown in Table 52. Results are not presented for ndm and oxa-58 (no positive LRT results) and for oxa-48 (one positive LRT result, no isolate available).

{70}------------------------------------------------

Resistance MarkerDetectedLRT MicroorganismTargetNumber ofspecimens withdemonstratedlinkage/Number ofTrue positive LRTMicroorganismResultsPercentage ofSpecimens withConfirmedLinkage
temH. influenzae1/250%
ctx-MEnterobacteriaceae, P.aeruginosa, and/orAcinetobacter spp.9/1656.3%
kpcEnterobacteriaceae, P.aeruginosa, and/orAcinetobacter spp.4/580%
vimEnterobacteriaceae, P.aeruginosa, and/orAcinetobacter spp.2/366.7%
oxa-23Acinetobacter spp.4/4100%
oxa-24Acinetobacter spp.1/1100%
mecAS. aureus37/4778.7%

Table 52: Performance. Linkage Analysis

Further details of the genotypic linkage and phenotypic agreement analyses are shown in Tables 53-59 with 'true positive' specimens presented in individual table rows. An additional column is added to each table listing specimens with positive resistance markers results and corresponding microorganism targets detected by LRT but not recovered by culture. It is noted that these microorganisms may be the source of the detected resistance marker; however, due to the lack of isolates, linkage and agreement with AST results cannot be evaluated.

Analysis of phenotypic agreement for the subset of specimens with positive LRT Application resistance marker results showed 100% agreement (i.e., resistance) for all positive specimens with culture isolates demonstrating genotypic linkage (i.e., corresponding culture isolate is positive for the resistance marker).

The following microorganism abbreviations are used in Tables 53-59: Acinetobacter spp. (Aci), Enterobacter cloacae complex (Ecl), Escherichia coli (Eco), Haemophilus influenzae (Hae), Klebsiella oxytoca (Kox), Klebsiella pneumoniae (Kpn), Morganella morganii (Mor), Proteus spp. (Pro), Pseudomonas aeruginosa (Pse), Serratia marcescens (Ser), Staphylococcus aureus (Sau).

{71}------------------------------------------------

Table 53: Linkage analysis for LRT positive specimens for tem that are true positive for H. influenzae and have available isolates. Agreement of AST results for specimens with confirmed linkage

True Positive HostMicroorganisms a# SpecimensIsolate LinkageAgreement of ASTresult for linkedmicroorganisms[%] b(95 % CI)
allarch.prosp.notconf.conf.Linkage[%](95 % CI)
tem2 d02111/250.0(9.4-90.6)1/1100.0(20.7-100.0)
Hae*10101R
Hae, [Eco*] c10110-

a True positive: SoC positive H. influenzae detected by LRT, *: indicates microorganisms for a certain specimen with confirmed presence of tem ("linkage") by PCR/bi-directional sequencing, "[Org.]": indicates other true positive microorganisms (Enterobacteriaceae, Acinetobacter spp., P. aeruginosa) that could potentially be the source of tem.

b R: resistant to Penicillin, S: susceptible to Penicillin, na: no AST data available.

C Note that tem was linked to an E. coli isolate isolated from this specimen and the H. influenzae isolate also isolated from this specimen was susceptible to cefinase.

d In total, four H. influenzae true positive specimens (one archived, three prospective) were

reported positive for tem by LRT; H. influenzae isolates were available for two cases.

Table 54: Linkage analysis for LRT positive specimens for ctx-M that are true positive for applicable LRT microorganisms and have available isolates. Agreement of AST results for specimens with confirmed linkage

True Positive HostMicroorganismsa# SpecimensIsolate LinkageLinkage[%](95 % CI)Agreement ofAST result forlinkedmicroorganismsb[%](95 % CI)Specimens that areLRT positive/culturenegative foradditional applicablemicroorganismsc
allarch.prosp.notconf.conf.
ctx-M16d610799/1656.3(33.2-76.9)7/7100.0(64.6-100.0)
Eco*, Pse20202R (1), na (1)0 of 2
Kpn*21102R (1), na (1)1 of 2
Eco*10101R0 of 1
Kpn*, Pse11001R1 of 1
(Kpn), Pse, Pro*11001R0 of 1
Kpn, Pro*11001R1 of 1
Pse, Pro*10101R1 of 1
Pse41340-4 of 4
(Kpn), Pse10110-0 of 1
Mor, Pse, Ser11010-1 of 1
(Pse), (Pro), Ser10110-1 of 1

4 True positive: SoC positive microorganisms detected by LRT, *: indicates microorganisms for a certain specimen with confirmed presence of cts-M ("linkage") by PCR bi-directional sequencing, "(Org.)": indicates microorganisms for which no isolate was collected and linkage analysis could not be performed.

b R: resistant to Third Generation Cephalosporins, na: no AST data

{72}------------------------------------------------

Additional LRT positive culture negative results for applicable targeted microorganisms that could potentially be the source of detected antibiotic resistance markers.

In total, 23 specimens were reported positive for ex-M by LRT (12 archived, 11 prospective); isolates were available for 16 cases.

Table 55: Linkage analysis for LRT positive specimens for kpc that are true positive for applicable LRT microorganisms and have available isolates. Agreement of AST results for specimens with confirmed linkage

True Positive HostMicroorganisms a# SpecimensIsolate LinkageLinkage[%](95 % CI)Agreement ofAST result forlinkedmicroorganisms[%] b(95 % CI)Specimens that areLRTpositive/culturenegative foradditionalapplicablemicroorganisms c
allarch.prosp.notconf.conf.
kpc5 d05144/580.0(37.6-96.4)3/3100.0(43.9-100.0)
Ecl*10101na0 of 1
Kpn*10101R0 of 1
Mor*10101R1 of 1
Pse*10101R1 of 1
(Pse), (Pro), Ser10110-1 of 1

a True positive: SoC positive microorganisms detected by LRT, *: indicates microorganisms for a certain specimen with confirmed presence of kpc ("linkage") by PCR bi-directional sequencing, "(Org.)": indicates microorganisms for which no isolate was collected and linkage analysis could not be performed.

b R: resistant to Carbapenems, na: no AST data available

C Additional LRT positive/culture negative for applicable targeted microorganisms that could potentially be the source of detected antibiotic resistance markers.

a In total, nine specimens were reported positive for kpc by LRT (three archived, six prospective); isolates were available for five cases.

Table 56: Linkage analysis for LRT positive specimens for vim that are true positive for applicable LRT microorganisms and have available isolates. Agreement of AST results for specimens with confirmed linkage

True Positive Host Microorganismsa# SpecimensIsolate LinkageLinkage [%] (95 % CI)Agreement of AST result for linked microorganismsb [%] (95 % CI)Specimens that are LRT positive/culture negative for additional applicable microorganismsc
vim3d12122/366.7 %(20.8-93.9)2/2100.0(34.2-100.0)
Pse*10101R0 of 1
Pse*, (Pro)10101R1 of 1
Kpn, (Pro)11010-1 of 1

a True positive: SoC positive microorganisms detected by LRT, *: indicates microorganisms for a certain specimen with confirmed

{73}------------------------------------------------

presence of vim ("linkage") by PCR/bi-directional sequencing, "(Org.)": indicates microorganisms for which no isolate was collected and linkage analysis could not be performed.

b R: resistant to Carbapenems, na: no AST data available.

6 Additional LRT positive culture negative results for applicable targeted microorganisms that could potentially be source of detected antibiotic resistance markers.

In total, three specimens (one archived, two prospective) were reported positive for vim by LRT, isolates were available for all cases.

Table 57: Linkage analysis for LRT positive specimens for oxa-23 that are true positive for Acinetobacter spp. and have available isolates. Agreement of AST results for specimens with confirmed linkage

True Positive HostMicroorganisms a# SpecimensIsolate LinkageAgreement ofAST result forlinkedmicroorganisms
allarch.prosp.notconf.conf.Linkage[%](95 % CI)[%] b(95 % CI)
oxa-234c22044/4100.0(51.0-100.0)3/3100.0(43.9-100.0)
Aci*42204R (3), na (1)

a True positive: SoC positive Acinetobacter spp. detected by LRT, *: indicates microorganisms for a certain specimen with confirmed presence of oxa-23 ("linkage") by PCR/bi-directional sequencing.

b R: resistant to Carbapenems, S: susceptible to Carbapenems, na: no AST data available

C In total, eight specimens were reported positive for oxa-23 by LRT (six archived, two prospective); isolates were available for four cases.

Table 58: Linkage analysis for LRT positive specimens for exa-24 that are true positive for Acinetobacter spp. and have available isolates. Agreement of AST results for specimens with confirmed linkage

True Positive HostMicroorganisms a# SpecimensIsolate LinkageLinkage[%]95 % CIAgreement of ASTresult for linkedmicroorganisms[%] b(95 % CI)
allarch.prosp.notconf.conf.
oxa-241 c01011/1100.0(20.7-100.0)1/1100.0(20.7-100.0)
Aci*10101R

a True positive: SoC positive Acinetobacter spp. detected by LRT, *: indicates microorganisms for a certain specimen with confirmed presence of oxa-24 ("linkage") by PCR/bi-directional sequencing.

b R: resistant to Carbapenems, S: susceptible to Carbapenems, na: no AST data available.

In total, one prospective specimen with available isolate was reported positive for oxa-24 by LRT.

{74}------------------------------------------------

Table 59: Linkage analysis for LRT positive specimens for mecA that are true positive for S. aureus and have available isolates. Agreement of AST results for specimens with confirmed linkage.

True Positive HostMicroorganisms a# SpecimensIsolate LinkageLinkage[%](95 % CI)Agreement of ASTresult for linkedmicroorganisms[%] b(95 % CI)
allarch.prosp.notconf.conf.
mecA47 c1136103737/4778.7(65.1-88.0)36/36100.0(90.4-100.0)
Sau*37829037R (36), na (1)
Sau1037100-

4 True positive: SoC positive S. aureus detected by LRT, *. indicates microorganisms for a certain specimen with confirmed presence of mecA ("linkage") by PCR/bi-directional sequencing.

b R: resistant to oxacillin. S: susceptible to oxacillin. na: no AST data available.

6 In total, 81 specimens were reported positive for mecA by LRT (33 archived, 48 prospective); isolates were available for 47 cases.

Summary of Clinical Study:

The clinical study design for evaluation of the Unyvero LRT assay did not include evaluation of patient outcome and therefore the potential impact of test results on patient care is unknown. The assay labeling and assay report include multiple limitations and warnings for the laboratory and clinician, most importantly stating that results from this assay must be used in conjunction with results from culture.

Microorganism Targets ('typical' bacteria): Clinical performance of the Unvvero LRT Application for detection of 'typical' microorganisms in tracheal aspirate specimens was evaluated by comparing LRT results to traditional semi-quantitative culture as well as to a composite comparator consisting of culture plus a molecular comparator (PCR followed by bidirectional sequencing). For most microorganism targets, assay sensitivity and specificity compared to culture were greater than 90% and 95% respectively.

The analysis and interpretation of tracheal aspirate cultures can be somewhat subjective, with the judgement of the laboratory technologist playing a critical role in determining the final culture results. Current recommendations for reporting semi-quantitative results for tracheal aspirate cultures are based on the relative quantities of all potential pathogens as well as the amount of normal respiratory flora that grow on culture plates.

LRT false positive results when compared to culture are not unexpected as the Unyvero LRT assay does not distinguish between viable and non-viable organisms, does not quantify the amount of DNA present and reports detected microbial DNA without providing information on the presence or amount of normal respiratory flora.

Although the LRT assay generated a significant number of false positive results compared to culture, most of these results were confirmed to be positive by PCR and sequencing. demonstrating that the targeted bacterial DNA was present in the specimen and the Unyvero

{75}------------------------------------------------

LRT assay correctly detected the presence of microbial DNA targets. For many false positive results, culture results were reported as mixed flora or normal respiratory flora and many specimens were positive by the LRT assay for two or more targeted microorganisms as shown in tables 25 above.

False positive results for 'typical' bacterial analytes can be mitigated by limitations provided in the assay labeling and test report informing the laboratory and clinician that detected microorganisms may be from colonizing flora and may not be the causative agent of disease. Most importantly, false positive and false negative Unvyero results can be mitigated by the requirement that concomitant culture is performed for all tracheal aspirate specimens tested with the Unyvero LRT Application.

Microorganism targets ('atypical' bacteria): Performance for the 'atypical' analytes was demonstrated primarily with contrived specimens with Unyvero LRT results showing acceptable agreement to expected results. The following limitation is included in the package insert:

  • . A negative result for the 'atypical' microorganisms (C. pneumoniae, L. pneumophila, and M. pneumoniae) does not exclude the presence of this microorganism in the patient specimen. A positive result should be evaluated in the overall context of the patient's clinical condition and other laboratory results being part of the standard-of-care routine.
    Resistance marker targets: The Unyvero LRT assay demonstrated acceptable performance for detection of targeted resistance markers when compared to validated PCR and sequencing assays. The assay masking rules applied by the assay software allow reporting of positive resistance marker results only for specimens with concurrently detected microorganisms that have the potential to carry the detected marker. However, despite these rules, detected resistance markers cannot always be definitively be linked to the concurrently detected microorganism as shown in Table 52 above. False positive results for resistance markers can be mitigated by the requirement for concomitant culture and subsequent AST testing. False negative results for resistance markers are mitigated by the following limitation included in the assay report:

  • . Antimicrobial resistance may occur via multiple mechanisms other than the resistance markers detected by the Unyvero LRT assay. Negative results for LRT resistance markers do not indicate antimicrobial susceptibility of detected organisms.
    Note: Greater than (b) a additional prospectively collected specimens were evaluated in the pivotal clinical study for the Unyvero LRT Application and results from this testing were submitted in the De Novo application. Results for this specimen cohort were reviewed and considered as additional supportive evidence in the assessment of Unyvero LRT Application performance.

4. Clinical cut-off:

Not applicable

{76}------------------------------------------------

5. Expected values/Reference range:

Of the 603 evaluable tracheal aspirate specimens in the prospective study, the LRT Application reported 312 specimens with at least one positive LRT panel microorganism, including 125 specimens with multi-detections for two or more microorganisms. Reference culture testing reported 236 specimens with at least one LRT panel microorganism, including 62 specimens with two or more LRT panel microorganisms reported. Expected values (number of positive results for each microorganism reported by the Unyvero LRT Application) are presented in Table 60 for all prospectively tested specimens as well as for specimens positive by LRT for multiple microorganisms. Table 61 includes the numbers of positive resistance marker results reported by the LRT assay in the prospective study.

Expected Values for allSpecimens(N= 603)Expected Values for Multi-detection Specimens(N= 125)
# specimens[%]# specimens[%]
Acinetobacter spp.274.52016.0
Chlamydia pneumoniae00.000.0
Citrobacter freundii30.521.6
Enterobacter cloacae complex203.3118.8
Escherichia coli457.53225.6
Haemophilus influenzae244.01512.0
Klebsiella oxytoca152.5118.8
Klebsiella pneumoniae386.32520.0
Klebsiella variicola40.743.2
Legionella pneumophila20.310.8
Moraxella catarrhalis132.297.2
Morganella morganii101.7108.0
Mycoplasma pneumoniae30.521.6
Proteus spp.294.82520.0
Pseudomonas aeruginosa8413.95443.2
Serratia marcescens254.12016.0
Staphylococcus aureus12019.95745.6
Stenotrophomonas maltophilia528.63931.2
Streptococcus pneumoniae132.232.4

Table 60: Expected values of LRT panel microorganism targets, prospective study

{77}------------------------------------------------

Expected Values(N= 603)
# specimens[%]
ctx-M172.8
kpc61.0
ndm00.0
oxa-2381.3
oxa-2420.3
oxa-4810.2
oxa-5800.0
tem81.3
vim20.3
mecA6210.3

Table 61: Expected values of LRT panel antibiotic resistance markers as determined by the LRT Application for the prospective aspirate study

Table 62 includes the numbers of positive results for resistance markers and concurrently detected microorganism targets as observed in the prospective clinical study. Results are stratified by the number of resistance marker targets detected.

Table 62: Expected Values of resistance markers and microorganism targets,
prospective study
MicroorganismResistanceMarker# Specimens
No resistance marker reported513
negative (no microorganisms detected)-291
Pseudomonas aeruginosa-30
Staphylococcus aureus-29
Stenotrophomonas maltophilia-13
Escherichia coli-11
Streptococcus pneumoniae-10
Klebsiella pneumoniae-9
Pseudomonas aeruginosa, Stenotrophomonasmaltophilia-9
Enterobacter cloacae complex-8
Haemophilus influenzae-7
Acinetobacter spp.-6
Serratia marcescens-5
Moraxella catarrhalis-4
Proteus spp.-4
Enterobacter cloacae complex, Escherichia coli-3
Klebsiella oxytoca-3
Moraxella catarrhalis, Staphylococcus aureus-3
Pseudomonas aeruginosa, Staphylococcus aureus-3
Staphylococcus aureus, Stenotrophomonasmaltophilia-3
Enterobacter cloacae complex, Klebsiella oxytoca,Stenotrophomonas maltophilia-2
Escherichia coli, Klebsiella pneumoniae-2
MicroorganismResistance Marker# Specimens
Escherichia coli, Klebsiella pneumoniae, Proteus spp.-2
Escherichia coli, Pseudomonas aeruginosa-2
Escherichia coli, Staphylococcus aureus-2
Haemophilus influenzae, Proteus spp., Staphylococcus aureus-2
Haemophilus influenzae, Staphylococcus aureus-2
Klebsiella oxytoca, Stenotrophomonas maltophilia-2
Klebsiella pneumoniae, Stenotrophomonas maltophilia-2
Proteus spp., Pseudomonas aeruginosa-2
Proteus spp., Staphylococcus aureus-2
Acinetobacter spp., Enterobacter cloacae complex, Pseudomonas aeruginosa-1
Acinetobacter spp., Klebsiella pneumoniae, Saureus, Stenotrophomonas maltophilia-1
Acinetobacter spp., Moraxella catarrhalis, Morganella morganii, Proteus spp., Pseudomonas aeruginosa, Serrratia marcescens, Stenotrophomonas maltophilia, Streptococcus pneumoniae-1
Acinetobacter spp., Proteus spp., Pseudomonas aeruginosa, Serrratia marcescens, Stenotrophomonas maltophilia-1
Acinetobacter spp., Serrratia marcescens, Stenotrophomonas maltophilia-1
Citrobacter freundii-1
Citrobacter freundii, Klebsiella oxytoca-1
Escherichia coli, Haemophilus influenzae, Klebsiella pneumoniae, Morganella morganii-1
Escherichia coli, Haemophilus influenzae, Staphylococcus aureus-1
Escherichia coli, Moraxella catarrhalis, Pseudomonas aeruginosa-1
Escherichia coli, Proteus spp., Pseudomonas aeruginosa, Serratia marcescens-1
Escherichia coli, Pseudomonas aeruginosa, Serratia marcescens-1
Escherichia coli, Stenotrophomonas maltophilia-1
Enterobacter cloacae complex, Escherichia coli, Klebsiella oxytoca, Pseudomonas aeruginosa-1
Enterobacter cloacae complex, Klebsiella variicola, Moraxella catarrhalis-1
Enterobacter cloacae complex, Proteus spp.-1
Enterobacter cloacae complex, Staphylococcus aureus-1
Haemophilus influenzae, Klebsiella pneumoniae-1
Haemophilus influenzae, Moraxella catarrhalis-1
Haemophilus influenzae, Pseudomonas aeruginosa, Serratia marcescens-1
Klebsiella oxytoca, Klebsiella pneumoniae, Morganella morganii, Pseudomonas aeruginosa, Serratia marcescens-1
Klebsiella oxytoca, Pseudomonas aeruginosa-1
Klebsiella oxytoca, Pseudomonas aeruginosa, Stenotrophomonas maltophilia-1
MicroorganismResistanceMarker# Specimens
Klebsiella pneumoniae, Klebsiella variicola-1
Klebsiella pneumoniae, Morganella morganii, Staphylococcus aureus-1
Klebsiella pneumoniae, Pseudomonas aeruginosa-1
Klebsiella pneumoniae, Serratia marcescens, Staphylococcus aureus-1
Klebsiella variicola, Stenotrophomonas maltophilia-1
Legionella pneumophila-1
Legionella pneumophila, Staphylococcus aureus-1
Moraxella catarrhalis, Proteus spp., Pseudomonas aeruginosa, Serratia marcescens-1
Mycoplasma pneumoniae-1
Mycoplasma pneumoniae, Streptococcus pneumoniae-1
Proteus spp., Pseudomonas aeruginosa, Serratia marcescens, Stenotrophomonas maltophilia-1
Proteus spp., Pseudomonas aeruginosa, Staphylococcus aureus-1
Pseudomonas aeruginosa, Serratia marcescens-1
Pseudomonas aeruginosa, Serratia marcescens, Stenotrophomonas maltophilia-1
Serratia marcescens, Staphylococcus aureus-1
Serratia marcescens, Stenotrophomonas maltophilia-1
One resistance marker reported78
Staphylococcus aureusmecA34
Klebsiella pneumoniae, Staphylococcus aureusmecA3
Acinetobacter spp., Staphylococcus aureusmecA2
Escherichia coli, Staphylococcus aureusmecA2
Acinetobacter spp., Morganella morganii, Proteus spp., Staphylococcus aureusmecA1
Citrobacter freundii, Klebsiella oxytoca, Klebsiella pneumoniae, Staphylococcus aureusmecA1
Escherichia coli, Klebsiella pneumoniae, Pseudomonas aeruginosa, Staphylococcus aureusmecA1
Escherichia coli, Serratia marcescens, Staphylococcus aureus, Stenotrophomonas maltophiliamecA1
Enterobacter cloacae complex, Escherichia coli, Staphylococcus aureusmecA1
Klebsiella variicola, Staphylococcus aureusmecA1
Moraxella catarrhalis, Pseudomonas aeruginosa, Staphylococcus aureusmecA1
Morganella morganii, Proteus spp., Pseudomonas aeruginosa, Staphylococcus aureusmecA1
Proteus spp., Pseudomonas aeruginosa, Staphylococcus aureusmecA1
Proteus spp., Staphylococcus aureusmecA1
Pseudomonas aeruginosa, Staphylococcus aureusmecA1
Serratia marcescens, Staphylococcus aureusmecA1
Staphylococcus aureus, Stenotrophomonas maltophiliamecA1
Klebsiella pneumoniaectx-M3
Escherichia colictx-M2
Acinetobacter spp., Escherichia coli, Klebsiella pneumoniaectx-M1
MicroorganismResistanceMarker# Specimens
Acinetobacter spp., Klebsiella pneumoniaectx-M1
Escherichia coli, Pseudomonas aeruginosactx-M1
Escherichia coli, Pseudomonas aeruginosa, Stenotrophomonas maltophiliactx-M1
Klebsiella oxytocactx-M1
Enterobacter cloacae complexkpc1
Klebsiella pneumoniaekpc1
Morganella morganii, Serratia marcescenskpc1
Acinetobacter spp., Proteus spp., Pseudomonas aeruginosa, Stenotrophomonas maltophiliaoxa-231
Acinetobacter spp., Pseudomonas aeruginosa, Staphylococcus aureus, Stenotrophomonas maltophiliaoxa-231
Acinetobacter spp.oxa-241
Morganella morganii, Pseudomonas aeruginosaoxa-481
Haemophilus influenzaetem2
Haemophilus influenzae, Streptococcus pneumoniaetem1
Haemophilus influenzae, Mycoplasma pneumoniaetem1
Escherichia coli, Haemophilus influenzaetem1
Haemophilus influenzae, Staphylococcus aureustem1
Pseudomonas aeruginosa, Staphylococcus aureus, Stenotrophomonas maltophiliavim1
Two resistance markers reported
Escherichia coli, Staphylococcus aureusctx-M, mecA1
Klebsiella pneumoniae, Pseudomonas aeruginosa, Staphylococcus aureusctx-M, mecA1
Klebsiella pneumoniae, Morganella morganii, Proteus spp., Pseudomonas aeruginosa, Serratia marcescens, Stenotrophomonas maltophiliactx-M, kpc1
Acinetobacter spp., Klebsiella pneumoniae, Pseudomonas aeruginosa, Stenotrophomonas maltophiliactx-M, oxa-241
Acinetobacter spp., Escherichia coli, Pseudomonas aeruginosa, Serratia marcescens, Stenotrophomonas maltophiliakpc, oxa-231
Acinetobacter spp., Proteus spp., Staphylococcus aureusmecA, oxa-231
Acinetobacter spp., Proteus spp., Pseudomonas aeruginosa, Staphylococcus aureusmecA, oxa-231
Haemophilus influenzae, Staphylococcus aureusmecA, tem1
Three resistance markers reported
Acinetobacter spp., Escherichia coli, Klebsiella pneumoniae, Proteus spp., Pseudomonas aeruginosa, Staphylococcus aureusctx-M, mecA, oxa-231
Acinetobacter spp., Escherichia coli, Haemophilus influenzae, Pseudomonas aeruginosa, Staphylococcus aureus, Stenotrophomonas maltophiliactx-M, mecA, tem1
Acinetobacter spp., Morganella morganii, Proteus spp., Pseudomonas aeruginosactx-M, oxa-23, vim1
Acinetobacter spp., Escherichia coli, Pseudomonas aeruginosa, Serratia marcescens, Staphylococcus aureus, Stenotrophomonas maltophiliakpc, mecA, oxa-231

{78}------------------------------------------------

{79}------------------------------------------------

{80}------------------------------------------------

{81}------------------------------------------------

M. Instrument Name:

Unyvero System

N. System Descriptions:

    1. Modes of Operation:
      Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?

Yes

Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?

Yes __________________________________________________________________________________________________________________________________________________________________________

    1. Software:
      FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes ______________ or No _____________________________________________________________________________________________________________________________________________________

Level of Concern Moderate

Software Description

A detailed description of the Unyvero System software was provided and included the function of each Unyvero System software component. A summary of the features of the System component software is shown in Table 63.

{82}------------------------------------------------

Component of UnyveroSystemDescription and Function of System Component
Unyvero LysatorThe Lysator software is responsible for specimen lysis andcommunicates with the Cockpit software. The Lysator software managesinternal mechanical/electrical function of the Lysator instrument.
Unyvero AnalyzerThe Analyzer software controls the integrated mechanical, electronic andoptical elements to execute a test run within the Unyvero LRT Cartridge.The Analyzer software is responsible for managing run analysisworkflow, generation of test results and communicating with theUnyvero Cockpit software. The Analyzer software monitors internalmechanical/electrical functions of the Analyzer instrument.
Unyvero CockpitThe Unyvero Cockpit software provides the main user interface for theoverall Unyvero System. The Cockpit software is responsible formanaging communication between all Unyvero System components(Cockpit, Lysator, Analyzer), management of analysis workflow, andpresentation and storage of test results.
Entire Unyvero SystemSoftware for each component is involved in management ofcommunication between components.

Table 63: Unyvero System

Device Hazard Analysis:

A risk analysis and corresponding risk management plan was provided for the Unyvero LRT Application (i.e., assay) and Unyvero System. The risk analysis included potential risks to the patient. Risks identified for the patient were mitigated to Low/Moderate risk levels with assay and instrument controls, use of barcodes for process control, verification and validation procedures and detailed instructions for use. The risk analysis also included potential risks to the operator. Risks identified for the operator were mitigated to Low/Moderate risk levels through labeling, verification and validation, manufacturing quality control measures, and EMC testing. The Device Hazard Analysis for the Unyvero LRT Application and Unyvero System was acceptable.

Architecture Design Chart: A detailed structure of the software used in the Unyvero System was provided.

Software Requirements Specification (SRS): SRS documentation was provided describing requirements and specifications for each of the software components of the Unyvero System was described.

Traceability Analysis: Documentation of traceability matrix that links all product requirements, functional specifications, and verification and validation testing for the complete Unyvero system was provided.

Software Development Environment Description: A description of Unyvero software development environment was provided and was acceptable.

Verification and Validation Testing:

The sponsor provided adequate documentation of verification and validation (V & V) testing covering all software/instrument components of the Unyvero System. V & V

{83}------------------------------------------------

testing of Unyvero System software was successfully completed at the individual component and system integration levels. The normal operation and user interface of all Unyvero software components were also tested and verified.

Revision Level History:

The firm provided a software revision level history that detailed the updates to the system software corresponding to each version.

Unresolved Anomalies: All major residual risks and unresolved anomalies were properly mitigated. Any remaining anomalies did not present major concerns for safety and efficacy for either the user or the patient.

EMC Testing: The Lysator, Analyzer, and Cockpit components of the Unyvero System were subjected to EMC testing. Testing was conducted according to acceptable standards and no EMC issues were observed.

3. Specimen Identification:

Specimen Identification information can be manually entered or automatically entered using the integrated barcode reader.

4. Specimen Sampling and Handling:

Before starting the test, the user scans the clinical identification (barcode) from the primary specimen container using the built-in barcode reader of the Unvvero Cockpit or the information may be entered manually on the cockpit on-screen keyboard.

The specimen is initially vortexed and then manually pipetted into the Unyvero Sample Tube. If the specimen is viscous, a Unyvero T1 Sample Transfer Tool may be used to facilitate pipetting the specimen into the Sample Tube.

After the specimen is placed in the Sample Tube, the user then places the Unvvero Cap on the Sample Tube. scans the Sample Tube barcode and places it in the Unyvero Lysator. After processing on the Lysator is finished, the user places the Sample Tube and thawed Mastermix into the Unyvero LRT Cartridge, scans the Cartridge barcode and places it into the indicated position in the Unyvero Analyzer as per Unyvero software instructions. The Unyvero software then instructs the user to start the test which is fully automated until completion.

5. Calibration:

Calibration is not required by the user.

    1. Quality Control:

{84}------------------------------------------------

See section L1(c) for information on internal and external controls.

O. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Not applicable

  • P. Proposed Labeling: The labeling is sufficient and satisfies the requirements of 21 CFR parts 801 and 809 as well as the Special Controls for this type of device.

Q. Identified Risks to Health and Mitigation Measures:

Identified RisksMitigation Measures
Incorrect identification or lack ofidentification of a pathogenic microorganismby the device can lead to improper patientmanagementGeneral Controls and Special Controls (1),(2), (3) and (4)
Failure to correctly interpret test resultsGeneral Controls and Special Controls (1),(2)(iii), (2)(iv), (2)(v), (2)(vi), (2)(vii),(2)(viii), and (3)
Failure to correctly operate the instrumentGeneral Controls and Special Controls (1),(2)(i), (4)(ii), (4)(iii) and (4)(iv)

{85}------------------------------------------------

R. Benefit/Risk Analysis:

Summary
Summary ofthe Benefit(s)The Unyvero LRT Application is the first multiplex PCR assay to detect and identify nucleic acids from bacteria and antimicrobial resistance markers directly from tracheal aspirate specimens. The Unyvero LRT Application can provide bacterial identification results and antimicrobial resistance marker results in approximately 4.5 hours compared to traditional bacterial culture for which final bacterial identification and antimicrobial susceptibility testing can take several days. The performance of the Unyvero LRT Application demonstrated acceptable performance for detection of assay targets in clinical specimens. Although sensitivity/PPA did not exceed 95% for many analytes for the primary efficacy endpoint and specificity/NPA demonstrated a relatively high rate of false positive results, clinical performance is mitigated by use of the assay in conjunction with traditional culture and interpretation by healthcare providers. Detection of resistance markers by the Unyvero LRT assay correlated with phenotypic antimicrobial resistance in cultured isolates that carried each antimicrobial resistance marker. Positive antimicrobial resistance marker results from the LRT assay may identify patients for which broad empiric therapy may be necessary.
Summary ofthe Risk(s)False positive results and false negative results are the primary risks associated with use of the Unyvero LRT Application. A false positive result may lead to unnecessary antimicrobial therapy, and potential adverse drug reactions, such as allergic reactions, C. difficile colitis and/or increased antimicrobial resistance. A false negative result may result in a delay of effective antimicrobial therapy, with subsequent worsening of infection and associated increase in morbidity or mortality. Misinterpretation of antimicrobial resistance genes could result in unnecessary broad-spectrum antimicrobial therapy or a delay to effective therapy, which could lead to potential adverse drug reactions or increased morbidity or mortality.
Summary ofOtherFactorsNone.

{86}------------------------------------------------

ConclusionsDo theprobablebenefitsoutweigh theprobablerisks?The probable benefits of the Unyvero LRT assay outweigh the potential risks in light ofthe listed special controls and applicable general controls. The Unyvero LRT assay is thefirst multiplex PCR diagnostic device to detect and identify bacterial nucleic acids andantimicrobial resistance genes directly from tracheal specimens and is likely to benefitpatients by more rapidly diagnosing tracheitis or ventilator-associated pneumonia andidentifying highly resistance bacterial infections. The clinical performance observed incomparison to traditional bacterial culture and/or validated PCR assays indicated that theUnyvero LRT assay could provide potential benefits to patients by rapid and accuratediagnosis of lower respiratory tract infections. The proposed special controls will ensurethat errors will be uncommon, and potential errors are further mitigated by currentlaboratory practices, which include standard of care bacterial culture, other diagnostics,and product labeling.
----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

S. Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

T. Conclusion:

The information provided in this de novo submission to classify this device into class II under regulation 21 CFR 866.3985. FDA believes that the stated special controls, and applicable general controls, including design controls, provide reasonable assurance of the safety and effectiveness of the device type. The device is classified under the following:

Product Code:QBH
Device Type:Device to detect and identify microorganisms and associated resistancemarker nucleic acids directly in respiratory specimens
Class:II (special controls)
Regulation:21 CFR 866.3985
  • (a) Identification. A device to detect and identify microorganisms and associated resistance marker nucleic acids directly in respiratory specimens is an in vitro diagnostic device intended for the detection and identification of microorganisms and associated resistance markers in respiratory specimens collected from patients with signs or symptoms of respiratory infection. The device is intended to aid in the diagnosis of respiratory infection in conjunction with clinical signs and symptoms and other laboratory findings. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist other than those detected by the device.
    (b) Classification. Class II (special controls). The special controls for this device are:

{87}------------------------------------------------

  • (1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of what the device detects, the type of results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
  • (2) The 21 CFR 809.10(b) labeling must include:
    • (i) A detailed device description, including all device components, control elements incorporated into the test procedure, instruments, ancillary reagents required but not provided, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens.
    • (ii) Performance characteristics from analytical studies, including but not limited to limit of detection, inclusivity, reproducibility, cross reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, and linearity, as applicable.
    • (iii) A limiting statement that the device is intended to be used in conjunction with clinical history, signs and symptoms, and results of other diagnostic tests, including culture and antimicrobial susceptibility testing.
    • (iv) A detailed explanation of the interpretation of test results for clinical specimens and acceptance criteria for any quality control testing.
    • (v) A limiting statement that negative results for microorganisms do not preclude the possibility of infection, and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
    • If applicable, a limiting statement that detected microorganisms may not be the (vi) cause of lower respiratory tract infection and may be indicative of colonizing or normal respiratory flora.
    • (vii) If applicable, a limiting statement that detection of resistance markers cannot be definitively linked to specific microorganisms and that the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora.
    • (viii) If applicable, a limiting statement that detection of antibiotic resistance markers may not correlate with phenotypic gene expression.
  • The 21 CFR 809.10(b) labeling and any test report generated by the device must (3) include a limiting statement that negative results for resistance markers do not indicate susceptibility of detected microorganisms.

{88}------------------------------------------------

  • (4) Design verification and validation must include:
    • Performance characteristics from clinical studies that include prospective (i) (sequential) samples and, if appropriate, additional characterized samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from an FDA accepted reference method and/or FDA accepted comparator method, as appropriate. Results from the clinical studies must include the clinical study protocol (including predefined statistical analysis plan, if applicable), clinical study report, and results of all statistical analyses.
    • A detailed device description including the following: (ii)
      • (A)Thorough description of the assay methodology including, but not limited to, primer/probe sequences, primer/probe design, and rationale for target sequence selection, as applicable.
      • (B) Algorithm used to generate a final result from raw data (e.g., how raw signals are converted into a reported result).
    • (iii) A detailed description of device software, including, but not limited to, validation activities and outcomes.
    • As part of the risk management activities, an appropriate end user device training (iv) program must be offered as an effort to mitigate the risk of failure from user error.

§ 866.3985 Device to detect and identify microorganisms and associated resistance marker nucleic acids directly in respiratory specimens.

(a)
Identification. A device to detect and identify microorganisms and associated resistance marker nucleic acids directly from respiratory specimens is an in vitro diagnostic device intended for the detection and identification of microorganisms and associated resistance markers in respiratory specimens collected from patients with signs or symptoms of respiratory infection. The device is intended to aid in the diagnosis of respiratory infection in conjunction with clinical signs and symptoms and other laboratory findings. These devices do not provide confirmation of antibiotic susceptibility since mechanisms of resistance may exist other than those detected by the device.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The intended use for the 21 CFR 809.10 labeling must include a detailed description of what the device detects, the type of results provided to the user, the clinical indications appropriate for test use, and the specific population(s) for which the device is intended.
(2) The 21 CFR 809.10(b) labeling must include:
(i) A detailed device description, including all device components, control elements incorporated into the test procedure, instrument requirements, ancillary reagents required but not provided, and a detailed explanation of the methodology, including all pre-analytical methods for processing of specimens.
(ii) Performance characteristics from analytical studies, including, but not limited to, limit of detection, inclusivity, reproducibility, cross reactivity, interfering substances, competitive inhibition, carryover/cross contamination, specimen stability, and linearity, as applicable.
(iii) A limiting statement that the device is intended to be used in conjunction with clinical history, signs and symptoms, and results of other diagnostic tests, including culture and antimicrobial susceptibility testing.
(iv) A detailed explanation of the interpretation of test results for clinical specimens and acceptance criteria for any quality control testing.
(v) A limiting statement that negative results for microorganisms do not preclude the possibility of infection, and should not be used as the sole basis for diagnosis, treatment, or other patient management decisions.
(vi) If applicable, a limiting statement that detected microorganisms may not be the cause of lower respiratory tract infection and may be indicative of colonizing or normal respiratory flora.
(vii) If applicable, a limiting statement that detection of resistance markers cannot be definitively linked to specific microorganisms and that the source of a detected resistance marker may be an organism not detected by the assay, including colonizing flora.
(viii) If applicable, a limiting statement that detection of antibiotic resistance markers may not correlate with phenotypic gene expression.
(3) The 21 CFR 809.10(b) labeling and any test report generated by the device must include a limiting statement that negative results for resistance markers do not indicate susceptibility of detected microorganisms.
(4) Design verification and validation must include:
(i) Performance characteristics from clinical studies that include prospective (sequential) samples and, if appropriate, additional characterized samples. The study must be performed on a study population consistent with the intended use population and compare the device performance to results obtained from an FDA accepted reference method and/or FDA accepted comparator method, as appropriate. Results from the clinical studies must include the clinical study protocol (including predefined statistical analysis plan, if applicable), clinical study report, and results of all statistical analyses.
(ii) A detailed device description including the following:
(A) Thorough description of the assay methodology including, but not limited to, primer/probe sequences, primer/probe design, and rationale for target sequence selection, as applicable.
(B) Algorithm used to generate a final result from raw data (e.g., how raw signals are converted into a reported result).
(iii) A detailed description of device software, including, but not limited to, validation activities and outcomes.
(iv) As part of the risk management activities, an appropriate end user device training program must be offered as an effort to mitigate the risk of failure from user error.