The VENTANA MMR IHC Panel is a qualitative immunohistochemistry (IHC) test intended for use in the light microscopic assessment of mismatch repair (MMR) proteins (MLH1, PMS2, MSH2, and MSH6) and BRAF V600E proteins in formalin-fixed, paraffin-embedded colorectal cancer (CRC) tissue sections. The OptiView DAB IHC Detection Kit is used with MLH1, MSH2, MSH6 and BRAF V600E, and the OptiView DAB IHC Detection Kit with OptiView Amplification Kit is used for PMS2 detection. The VENTANA MMR IHC Panel is for use on the VENTANA BenchMark ULTRA instrument. The VENTANA MMR IHC Panel includes VENTANA anti-MLH1 (M1) Mouse Monoclonal Primary Antibody, VENTANA anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody, and VENTANA anti-BRAF V600E (VE1) Mouse Monoclonal Primary Antibody. The VENTANA MMR IHC Panel is indicated in patients diagnosed with colorectal cancer (CRC) to detect mismatch repair (MMR) proteins deficiency as an aid in the identification of probable Lynch syndrome and to detect BRAFV600E protein as an aid to differentiate between sporadic CRC and probable Lynch syndrome. Results from the Ventana MMR IHC Panel should be interpreted by a qualified pathologist in conjunction with histological examination, relevant clinical information, and proper controls. The clinical performance of this device to guide treatment of MMR deficient patients has not been established.
Device Story
The VENTANA MMR IHC Panel is an automated immunohistochemistry test for FFPE colorectal cancer tissue. It utilizes five primary antibodies (anti-MLH1, anti-PMS2, anti-MSH2, anti-MSH6, and anti-BRAF V600E) to detect target protein expression. The device runs on the VENTANA BenchMark ULTRA instrument using OptiView DAB detection (with amplification for PMS2). A qualified pathologist performs visual examination of stained slides under a light microscope. The output—presence or absence of nuclear staining for MMR proteins and cytoplasmic staining for BRAF V600E—aids in identifying probable Lynch syndrome versus sporadic CRC. The device is used in clinical pathology laboratories. Results assist clinicians in determining the need for further genetic counseling and testing. It is not a standalone diagnostic.
Clinical Evidence
Clinical validity was established by comparing IHC results to a validated DNA sequencing panel for Lynch syndrome variants and BRAF V600E mutations in 119 evaluable CRC specimens. The study demonstrated an overall agreement (OPA) of 94.1% (PPA 77.8%, NPA 97.0%). Analytical performance included precision studies (within-day, between-day, between-instrument, lot-to-lot, and reader) showing high concordance (typically 100% OPA). Literature review of 154 papers supported the clinical utility of the 4-marker MMR panel and BRAF V600E testing for Lynch syndrome screening.
Technological Characteristics
The device consists of five ready-to-use monoclonal primary antibodies (MLH1, PMS2, MSH2, MSH6, BRAF V600E) for use on the VENTANA BenchMark ULTRA automated staining platform. Detection utilizes OptiView DAB IHC Detection Kit (with OptiView Amplification Kit for PMS2). The system is designed for FFPE tissue sections. It is an automated, instrument-based IHC assay. Quality control requires internal positive controls (lymphocytes, fibroblasts, or normal epithelium).
Indications for Use
Indicated for patients diagnosed with colorectal cancer (CRC) to detect mismatch repair (MMR) protein deficiency as an aid in identifying probable Lynch syndrome and to detect BRAF V600E protein as an aid to differentiate between sporadic CRC and probable Lynch syndrome.
Regulatory Classification
Identification
Lynch syndrome test systems are in vitro diagnostic tests for use with tumor tissue to identify previously diagnosed cancer patients at risk for having Lynch syndrome.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include the following information, as appropriate:
(i) A detailed description of all test components, including all provided reagents, and required but not provided, ancillary reagents.
(ii) A detailed description of instrumentation and equipment, including illustrations or photographs of non-standard equipment or manuals.
(iii) Detailed documentation of the device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software.
(iv) A detailed description of quality controls including appropriate positive and negative controls that are recommended or provided.
(v) Detailed specifications for sample collection, processing, and storage.
(vi) A detailed description of methodology and assay procedure.
(vii) A description of the assay cut-off (
*i.e.,* the medical decision point between positive and negative results) or other relevant criteria that distinguishes positive and negative results, or ordinal classes of marker expression, including the rationale for the chosen cut-off or other relevant criteria and results supporting validation of the cut-off.(viii) Detailed specification of the criteria for test result interpretation and reporting.
(ix) Detailed information demonstrating the performance characteristics of the device, including:
(A) Data from an appropriate study demonstrating clinical accuracy using well-characterized clinical specimens representative of the intended use population (
*i.e.,* concordance to Deoxyribonucleic Acid (DNA) sequencing results of the Lynch syndrome associated genes or method comparison to the predicate device using samples with known alterations in genes representative of Lynch syndrome). Pre-specified acceptance criteria must be provided and followed.(B) Appropriate device reproducibility data investigating all sources of variance (
*e.g.,* for distributed tests, data generated using a minimum of three sites, of which at least two sites must be external sites). Each site must perform testing over a minimum of 5 nonconsecutive days evaluating a sample panel that spans the claimed measuring range, and includes the clinical threshold. Pre-specified acceptance criteria must be provided and followed.(C) Data demonstrating reader reproducibility, both within-reader and between-reader, assessed by three readers over 3 nonconsecutive days at each site, including a 2 week washout period between reads, as appropriate.
(D) Device precision data using clinical samples spanning the measuring range and controls to evaluate the within-lot, between-lot, within-run, between run, and total variation.
(E) Analytical specificity studies including as appropriate, western blots, peptide inhibition, testing in normal tissues and neoplastic tissues, interference by endogenous and exogenous substances, and cross-reactivity and cross contamination testing.
(F) Device analytical sensitivity data generated by testing an adequate number of samples from individuals with the target condition such that prevalence of the biomarker in the target population is established.
(G) Device stability data, including real-time stability and in-use stability, and stability evaluating various storage times, temperatures, and freeze-thaw conditions, as appropriate.
(H) The staining performance criteria assessed must include overall staining acceptability, background staining acceptability, and morphology acceptability, as appropriate.
(I) Appropriate training requirements for users, including interpretation manual, as applicable.
(J) Identification of risk mitigation elements used by the device, including a description of all additional procedures, methods, and practices incorporated into the instructions for use that mitigate risks associated with testing.
(2) The device's § 809.10(b) of this chapter compliant labeling must include a detailed description of the protocol, including the information described in paragraphs (b)(1)(i) through (viii) of this section, as appropriate, and a detailed description of the performance studies performed and the summary of the results, including those that relate to paragraph (b)(1)(ix) of this section, as appropriate.
Related Devices
K213348 — BOND MMR Antibody Panel · Leica Biosystems Newcastle, Ltd. · Feb 21, 2023
P210001 — VENTANA MMR RxDx Panel · Ventana Medical Systems, Inc. (Roche Tissue Diagnostics) · Aug 17, 2021
P250004 — MMR IHC Panel pharmDx (Dako Omnis) · Agilent Technologies, Inc. · Aug 15, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
# EVALUATION OF AUTOMATIC CLASS III DESIGNATION FOR Ventana MMR IHC Panel DECISION SUMMARY
### A. De Novo Number:
DEN170030
### B. Purpose for Submission:
De novo request for evaluation of automatic class III designation of the VENTANA Mismatch Repair Immunohistochemistry (MMR IHC) Panel.
### C. Measurand:
Mismatch repair proteins MLH1, PMS2, MSH6 and V600E mutated BRAF protein.
# D. Type of Test:
Immunohistochemistry
# E. Applicant:
Ventana Medical Systems, Inc.
# F. Proprietary and Established Names:
VENTANA MMR IHC Panel
### G. Regulatory Information:
- 1. Regulation section:
21 CFR 864.1866
- 2. Classification:
Class II (Special Controls)
- 3. Product code:
PZJ
- 4. Panel:
{1}------------------------------------------------
88- Pathology
### H. Indications for use:
### 1. Indications for use:
The VENTANA MMR IHC Panel is a qualitative immunohistochemistry (IHC) test intended for use in the light microscopic assessment of mismatch repair (MMR) proteins (MLH1, PMS2, MSH2, and MSH6) and BRAF V600E proteins in formalin-fixed, paraffin-embedded colorectal cancer (CRC) tissue sections. The OptiView DAB IHC Detection Kit is used with MLH1, MSH2, MSH6 and BRAF V600E, and the OptiView DAB IHC Detection Kit with OptiView Amplification Kit is used for PMS2 detection. The VENTANA MMR IHC Panel is for use on the VENTANA BenchMark ULTRA instrument. The VENTANA MMR IHC Panel includes VENTANA anti-MLH1 (M1) Mouse Monoclonal Primary Antibody, VENTANA anti-PMS2 (A16-4) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH2 (G219-1129) Mouse Monoclonal Primary Antibody, VENTANA anti-MSH6 (SP93) Rabbit Monoclonal Primary Antibody, and VENTANA anti-BRAF V600E (VE1) Mouse Monoclonal Primary Antibody.
The VENTANA MMR IHC Panel is indicated in patients diagnosed with colorectal cancer (CRC) to detect mismatch repair (MMR) proteins deficiency as an aid in the identification of probable Lynch syndrome and to detect BRAFV600E protein as an aid to differentiate between sporadic CRC and probable Lynch syndrome.
Results from the Ventana MMR IHC Panel should be interpreted by a qualified pathologist in conjunction with histological examination, relevant clinical information, and proper controls.
The clinical performance of this device to guide treatment of MMR deficient patients has not been established.
- 2. Special conditions for use statement(s):
Intended for in vitro diagnostic (IVD) use.
Prescription use only.
- 3. Special instrument requirements:
Ventana BenchMark Ultra instrument
### I. Device Description:
The Ventana MMR IHC panel is comprised of five primary antibodies used to detect the MMR proteins MLH1, PMS2, MSH2 and MSH6 and mutated BRAF V600E protein in CRC tissue specimens. The primary antibodies are used in combination with individually optimized detection reagents and in conjunction with ancillary reagents common to all
{2}------------------------------------------------
immunohistochemistry test systems in order to complete specimen testing. The MMR IHC panel and BRAF V600E are optimized to run on the VENTANA BenchMark Ultra platform with OptiView DAB detection kit or in the case of PSM2 antibody the OptiView DAB detection Kit with the OptiView Amplification Kit. The presence or absence of target proteins is determined by visual examination of the specimen slide under light microscope by a qualified pathologist.
The Ventana MMR IHC panel antibodies are packaged as individual products in single ready to use reagent dispensers. The MMR IHC panel test is run on five separate CRC tissue slides and stained on BenchMark Ultra instrument. The primary antibody reagents are listed in Table 1 below.
| Primary Antibody | Antibody Concentration |
|-------------------------------|------------------------|
| VENTANA anti-MLH1(M1) | ~1 µg/mL |
| VENTANA anti-PMS2 (A16-4) | ~1 µg/mL |
| VENTANA anti-MSH2 (G219-1129) | ~20 µg/mL |
| VENTANA anti-MSH6 (SP93) | ~1 µg/mL |
| VENTANA anti-BRAF V600E (VE1) | ~12 µg/mL |
### Table 1. VENTANA MMR IHC Panel
Detection and ancillary reagents required but not provided with Ventana MMR IHC panel are listed below:
- . OptiView DAB IHC detection Kit containing the following components
- o OptiView peroxidase Inhibitor
- o OptiView HQ universal Linker
- OptiView HRP Multimer o
- OptiView H2O2 O
- O OptiView DAB
- o OptiView Copper
- OptiView Amplification kit .
- O OptiView Amplification (0.003% HQ conjugated tyramide complex)
- O OptiView H2O2
- O OptiView Multimer
- Hematoxylin II .
- Bluing Reagent .
- Reaction Buffer (10x) .
- EZ Prep Reagent (10x) .
- . ULTRA Cell Conditioning (CC1) Pre Dilute
- ULTRA Liquid Cover Slip (LCS) (Pre-dilute) .
### Control Tissue:
MMR antibodies: Pre-qualified CRC tissue with a MMR status of intact may be used as a
{3}------------------------------------------------
positive system-level control for MMR antibodies to detect the intact protein. Alternatively, pre-qualified normal colon tissue fixed and processed in the same manner as the patient tissue can also be used as a positive system-level control. Normal colon will stain positive for all antibodies in the MMR IHC panel. Since the MLH1, PMS2, MSH2, and MSH6 proteins are expressed in all tissues, a normal negative tissue control does not exist for these biomarkers. For a negative system level control, CRC with loss of an MMR protein can be used as an appropriate tissue control for mismatch repair protein deficiency status. However, lymphocytes, fibroblast and epithelial cells should exhibit staining and serve as positive internal control cells in CRC samples with MMR protein deficiency (dMMR).
BRAF V600E: A case of CRC positive for BRAF V600E mutated protein by VENTANA anti-BRAF V600E (VE1) IHC is an appropriate positive control. A negative system-level control is achieved with the fibroblasts and lymphocytes in normal adjacent epithelium.
### Instrumentation and Software:
The MMR IHC panel tests are fully automated. The MMR IHC panel antibodies are for use on the BenchMark ULTRA instrument using Ventana System Software (VSS) software version 12.3 or earlier.
### J. Standard/Guidance Document Referenced (if applicable):
CLSI guideline I/LA28-A2: Quality Assurance for Design Control and Implementation of ImmunohistochemistryAssays; Approved Guideline - Second Edition
Guidance for Submission of Immunohistochemistry Applications to the FDA. 1998
# K. Test Principle:
The MMR IHC panel is an immunohistochemistry test system used to stain FFPE colorectal cancer (CRC) specimens to detect expression of the MMR proteins -MLH1, PMS2, MSH2 and MSH6- and the BRAF V600E mutated protein. The 5 antibodies of MMR IHC panel have individualized staining protocols that are created using available staining parameters provided in staining procedures in the VSS software that drives the BenchMark ULTRA automated staining platform. The panel test is run individually on 5 separate tissue sections and the test process involves sequential application of specific primary antibodies against the panel protein, followed by detection reagents and chromogen deposition for visualization of the target protein expression.
Briefly the assay steps are as follows 1) anti MMR/ BRAF V600E antibody binds to the epitope in the target protein; 2) a Haptenated (HQ) secondary antibody binds to the primary antibody; 3) a tertiary horseradish peroxidase (HRP)-labeled antibody directed against HQ binds to the HQlabeled secondary antibody: and 4) the resulting complex is visualized with hydrogen peroxide and DAB, due to the formation of a visible brown precipitate at the antigen site. The PMS2 test uses the OptiView amplification in addition of to the OptiView DAB detection system for signal amplification by addition of hydrogen peroxide and Tyramide-HQ. The specimen slide is then counterstained with hematoxylin and cover slipped. Results are interpreted using a light
{4}------------------------------------------------
microscope by a pathologist.
### L. Interpretation of Results
### 1. Staining Interpretation
Clinical status for MMR proteins (MLH1, PMS2, MSH2 and MSH6) is assigned by a trained pathologist based on their evaluation for the presence of specific nuclear staining in the tumor. A clinical status of "Intact" is assigned to cases with unequivocal nuclear staining in viable tumor cells in the presence of acceptable internal positive controls (nuclear staining in Ivmphocytes, fibroblasts or normal epithelium in the vicinity of the tumor). A clinical status of "Loss" is assigned to cases with unequivocal loss of nuclear staining or focal weak equivocal nuclear staining in the viable turnor cells in the presence of internal positive controls.
If unequivocal nuclear stain is absent in internal positive controls and/or backeround staining interferes with interpretation, the assay should be considered unacceptable and repeated. Punctate nuclear staining of tumor cells should be considered negative.
BRAF V600E test interpretation is based on the evaluation of presence of immunohistochemical staining in cytoplasm of CRC neoplastic cells. Tumors with unequivocal cytoplasmic staining of any intensity in viable tumor cells are considered to be positive for BRAF V600E mutations.
Interpretation for MMR proteins and the BRAF V600E status is detailed in Table 2 below
| Clinical Status | Description |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intact MMR protein<br>Expression | Unequivocal nuclear staining in viable tumor cells, in the presence<br>of internal positive controls (nuclear staining in lymphocytes,<br>fibroblasts or normal colonic epithelium in the vicinity of the<br>tumor) |
| Loss of MMR Protein<br>Expression | Unequivocal loss of nuclear staining or focal weak equivocal<br>nuclear staining in the viable tumor cells in the presence of<br>internal positive controls |
| Note: If unequivocal nuclear stain is absent in internal positive controls and/or background<br>staining interferes with interpretation, the assay should be considered unacceptable and<br>repeated. Punctate nuclear staining of tumor cells should be considered negative (Loss). In<br>cases with focal tumor cell staining, the intensity of the nuclear staining should be at least that<br>of the internal positive controls along with the confluent /continuous staining of the nuclei in a<br>few epithelial glands or nests for the case to be given a Clinical Status of Intact. In the<br>absence of these conditions, a Clinical Status of Loss is given to the case. | |
| Positive for BRAF<br>V600E mutation | Unequivocal cytoplasmic staining of any intensity in viable<br>tumor cells above background. |
| | | Table 2. Interpretation of staining for VENTANA MMR IHC Panel antibodies | |
|--|--|--------------------------------------------------------------------------|--|
| | | | |
{5}------------------------------------------------
| Clinical Status | Description |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Negative for BRAF<br>V600E mutation | No staining or Equivocal cytoplasmic staining in viable tumor<br>cells.<br>Note: Nuclear staining, weak to strong staining of isolated<br>viable tumor cells/or small tumor clusters should be considered<br>negative. |
- 2. Result Conclusions
- Detection of all four proteins in the tumor indicates normal (i.e., Intact) mismatch repair . status.
- Loss of MLH1 expression is accompanied with the loss of its heterodimer partner, PMS2. . Loss of MLH1 may be either sporadic or evidence of probable Lynch syndrome. In sporadic occurrences of CRC, expression of the MLH1 gene may be suppressed by hypermethylation of its promoter. The presence of the BRAF V600E mutation in CRC cases is tightly linked to MLH1 promoter hypermethylation and loss of MLH1 protein. Loss of MLH1 and positive for BRAF V600E mutated proteins status are likely the result of a sporadic occurrence. Tumors negative for BRAF V600E mutated protein status is suggestive of Lynch syndrome.
- . The loss of PMS2, in the presence of intact MLH1 expression, or loss of MSH2 or MSH6 expression designates the tumor as Loss of the respective MMR protein and is consistent with probable Lynch syndrome.
- All individuals with suspected Lynch syndrome should be referred for genetic counseling . and further genetic testing to confirm the presence of the suspected germline mutation.
### M. Performance Characteristics (if/when applicable):
- 1. Analytical performance:
- a. Precision: Assay precision was evaluated for each of the 5 panel antibodies individually using an identical study design. The following precision parameters were evaluated in the precision studies
- . Within Day
- Between day .
- Between Instrument .
- Lot to Lot .
- . Reader Precision
#### Within Day Precision: i.
Within-day precision was evaluated using 10 CRC cases consisting of 5 cases with intact protein status and 5 cases with loss status for the panel proteins except MSH2 study that included 6 MSH2 intact and 5 MSH2 loss cases. Five (5) slides were stained using each panel antibody and one slide of each case was stained with Negative Control (Monoclonal). All slides were run on the BenchMark ULTRA using the OptiView DAB detection kit except for PSM2 slides which were run using the OptiView Amplification kit. Data was obtained from 50 total observations (10
{6}------------------------------------------------
cases X 5 replicates x 1 instrument) for each panel antibody. Within-day precision was 100%. The study met pre-specified acceptance criteria of ≥85% lower bound confidence interval. Results for individual panel antibodies are shown in Table 3 below.
| | Agreement | | | |
|--------------------------------|-----------|-------|-------|--------------|
| Antibody | | | | |
| | Type | n/N | % | 95% CI |
| anti MLH1 (M1) antibody | PPA | 25/25 | 100.0 | (86.7,100.0) |
| | NPA | 25/25 | 100.0 | (86.7,100.0) |
| | OPA | 50/50 | 100.0 | (92.9, 100) |
| anti-PMS2 (A16-4) antibody | PPA | 25/25 | 100.0 | (86.7,100.0) |
| | NPA | 25/25 | 100.0 | (86.7,100.0) |
| | OPA | 50/50 | 100.0 | (92.9,100.0) |
| anti MSH6 (SP93) antibody | PPA | 25/25 | 100.0 | (86.7,100.0) |
| | NPA | 25/25 | 100.0 | (86.7,100.0) |
| | OPA | 50/50 | 100.0 | (92.9,100.0) |
| anti-MSH2 (G219-1129) Antibody | PPA | 30/30 | 100.0 | (88.6,100.0) |
| | NPA | 25/25 | 100.0 | (86.7,100.0) |
| | OPA | 55/55 | 100.0 | (93.5,100.0) |
| anti BRAF V600E (VE1) antibody | PPA | 25/25 | 100.0 | (86.7,100.0) |
| | NPA | 25/25 | 100.0 | (86.7,100.0) |
| | OPA | 50/50 | 100.0 | (92.9,100.0) |
### Table 3. Within day Precision
#### ii. Between Dav Precision:
Between-day repeatability was evaluated using the same CRC cases as within-day testing. Two slides were stained using each primary antibody and one slide of each case was stained with Negative Control (Monoclonal) across 5 nonconsecutive days. For Day 1, the first two slides from each case from Within-day precision study were used. Data was obtained from 100 total observations (10 cases X 2 replicates X 5 days) for each of the panel antibody (Note: MSH2 evaluation was conducted with 6 intact and 5 loss status cases for a total of 110 observations). Between-day precision was observed to be 100%. The study met pre-specified acceptance criteria.
Results for individual panel antibody are shown in Table 4 below.
| Antibody | Agreement | | | |
|-------------------------|-----------|---------|-------|---------------|
| | Type | n/N | % | 95% CI |
| anti-MLH1 (M1) antibody | PPA | 50/50 | 100.0 | (92.9, 100.0) |
| | NPA | 50/50 | 100.0 | (92.9, 100.0) |
| | OPA | 100/100 | 100.0 | (96.3, 100) |
| anti-PMS2 (A16-4) | PPA | 50/50 | 100.0 | (92.9,100.0) |
### Table 4. Between day Precision
{7}------------------------------------------------
| | NPA | 50/50 | 100.0 | (92.9,100.0) |
|----------------------------|-----|---------|---------|--------------|
| | OPA | 100/100 | 100.0 | (96.3,100.0) |
| anti MSH-6 (SP93) antibody | PPA | 50/50 | 100.0 | (92.9,100.0) |
| | | NPA | 50/50 | 100.0 |
| | | OPA | 100/100 | 100.0 |
| Anti-MSH2 (G219-1129) | PPA | 60/60 | 100.0 | (94.0,100.0) |
| | | NPA | 50/50 | 100.0 |
| | | OPA | 110/110 | 100.0 |
| BRAF V600E | PPA | 50/50 | 100.0 | (92.9,100.0) |
| | | NPA | 50/50 | 100.0 |
| | | OPA | 100/100 | 100.0 |
#### Between instrument: iii.
Instrument to instrument reproducibility was evaluated using the same CRC cases as Within-day testing. Two slides were stained using panel antibody and one slide of each case was stained with Negative Control (Monoclonal) across 3 different ULTRA instruments. Data was obtained from 60 total observations (10 cases X 2 replicates X 3 instruments) for each panel antibody (Note: MSH2 evaluation was conducted with 6 intact and 5 loss status cases for a total of 66 observations). The study met prespecified acceptance criteria. Results for individual panel antibody are shown in Table 5 below.
| | Agreement | | | |
|-----------------------------------|-----------|-------|-------|--------------|
| Day | Type | n/N | % | 95% CI |
| All ULTR MLH1 (M1) antibody As | PPA | 30/30 | 100.0 | (88.6,100.0) |
| | NPA | 30/30 | 100.0 | (88.6,100.0) |
| | OPA | 60/60 | 100.0 | (94.0,100.0) |
| anti-PMS2 (A16-4) | PPA | 30/30 | 100.0 | (88.6,100.0) |
| | NPA | 30/30 | 100.0 | (88.6,100.0) |
| | OPA | 60/60 | 100.0 | (94.0,100.0) |
| anti-MSH2 (G219-1129)<br>antibody | PPA | 36/36 | 100.0 | (90.4,100.0) |
| | NPA | 30/30 | 100.0 | (88.6,100.0) |
| | OPA | 66/66 | 100.0 | (94.5,100.0) |
| anti MSH6 (SP93) antibody | PPA | 30/30 | 100.0 | (88.6,100.0) |
| | NPA | 30/30 | 100.0 | (88.6,100.0) |
| | OPA | 60/60 | 100.0 | (94.0,100.0) |
| BRAF V600E | PPA | 30/30 | 100.0 | (88.6,100.0) |
Table 5. Between Instrument Precision
{8}------------------------------------------------
| antibody | | NPA | 30/30 | 100.0 | (88.6,100.0) |
|----------|--|-----|-------|-------|--------------|
| | | OPA | 60/60 | 100.0 | (94.0,100.0) |
#### Lot to lot Precision: iv.
The study evaluated Lot to lot precision with 3 final lots of the 5 panel antibodies with 10 qualified CRC cases consisting of 5 cases Intact for the panel protein expression and 5 cases Loss. All cases were run in triplicate for each lot test and one slide of each case was stained with the negative control reagent. All slides were run on the BenchMark ULTRA using OptiView DAB detection kit. Data was obtained from 90 total observations (10 cases X 3 replicates X 3 lots). Overall. 90 out of 90 slides were evaluable for MLH1, MSH2, MSH6 and BRAF V600E antibodies, while 3 PMS2 stained slides could not be evaluated due to unacceptable background scores >0.5 and therefore could not be evaluated. The study demonstrated an overall percent agreement of 100% for each of the panel antibodies. Results for individual panel antibody are shown in Table 6 below. The results met the pre-specified acceptance criteria.
| Antibody | Agreement | | | |
|---------------------------|-----------|-------|-------|---------------|
| | Type | n/N | % | 95% CI |
| MLH1 (M1) antibody | PPA | 45/45 | 100.0 | (92.1, 100.0) |
| | NPA | 45/45 | 100.0 | (92.1, 100.0) |
| | OPA | 90/90 | 100.0 | (95.9, 100.0) |
| anti-PMS2 (A16-4) | PPA | 44/44 | 100.0 | (92.9,100.0) |
| | NPA | 43/43 | 100.0 | (92.9,100.0) |
| | OPA | 87/87 | 100.0 | (96.3,100.0) |
| anti MSH6 (Sp93) antibody | PPA | 45/45 | 100.0 | (92.1, 100.0) |
| | NPA | 45/45 | 100.0 | (92.1, 100.0) |
| | OPA | 90/90 | 100.0 | (95.9, 100.0) |
| Anti-MSH2 (G219-1129) | PPA | 45/45 | 100.0 | (92.1, 100.0) |
| | NPA | 45/45 | 100.0 | (92.1, 100.0) |
| | OPA | 90/90 | 100.0 | (95.9, 100.0) |
| BRAF V600E | PPA | 45/45 | 100.0 | (92.1, 100.0) |
| | NPA | 45/45 | 100.0 | (92.1, 100.0) |
| | OPA | 90/90 | 100.0 | (95.9, 100.0) |
Table 6. Between Lot Precision
#### v. Reader Precision:
Between-Reader and Within-Reader precision were assessed by evaluating concordance of marker status across 3 readers and among individual readers using 20 cases of CRC. These 20 CRC specimens consisted of varying number of cases with loss or intact MMR protein status and BRAF V600E positive or negative status as shown in Table 7. Each reader scored all 20 cases in two rounds that were separated by a two week wash out period. Scores were analyzed for agreement between and within readers. Between reader precision ranged 97.5 to 100%. The results are shown
{9}------------------------------------------------
in Table 8. Within reader precision ranged 98.3 to 100%. Within-reader precision is shown in Table 9.The studies met the pre-specified acceptance criteria.
| Marker | # of Cases with<br>Intact Status | # of Cases with<br>Loss Status |
|------------|----------------------------------|--------------------------------|
| MLH1 | 11 | 9 |
| PMS2 | 13 | 7 |
| MSH2 | 12 | 8 |
| MSH6 | 11 | 9 |
| BRAF V600E | 10 | 10 |
Table 7. Case Distribution across Markers in Reader Precision
| Marker | Agreement | | | |
|------------|-----------|---------|--------|--------------|
| Type | n/N | % | 95% CI | |
| MLH-1 | PPA | 66/66 | 100.0 | (94.5,100.0) |
| | NPA | 53/54 | 98.1 | (90.2,99.7) |
| | OPA | 119/120 | 99.2 | (95.4,99.9) |
| PMS2 | PPA | 78/78 | 100.0 | (95.3,100.0) |
| | NPA | 42/42 | 100.0 | (91.6,100.0) |
| | OPA | 120/120 | 100.0 | (96.9,100.0) |
| MSH6 | PPA | 66/66 | 100.0 | (94.5,100.0) |
| | NPA | 53/54 | 98.1 | (90.2,99.7) |
| | OPA | 119/120 | 99.2 | (95.4,99.9) |
| MSH2 | PPA | 72/72 | 100.0 | (94.9,100.0) |
| | NPA | 48/48 | 94.4 | (92.6,100) |
| | OPA | 120/120 | 97.5 | (96.9,100) |
| BRAF V600E | PPA | 60/60 | 100.0 | (94.0,100.0) |
| | NPA | 60/60 | 100.0 | (94.0,100.0) |
| | OPA | 120/120 | 100.0 | (96.9,100.0) |
Table 8. Between Reader Precision by marker
### Table 9. Within reader precision
| Marker | Agreement | | | |
|--------|-----------|-------|------|---------------|
| | Type | n/N | % | 95% CI |
| MLH-1 | APA | 66/67 | 98.5 | (88.0,100.0) |
| | ANA | 52/53 | 98.1 | (83.3,100.0) |
| | OPA | 59/60 | 98.3 | (85.0,100.0) |
| PMS2 | APA | 72/72 | 100 | (95.3100.0) |
| | ANA | 48/48 | 100 | (91.2,100.0) |
| | OPA | 60/60 | 100 | (94.0,100.0) |
| MSH6 | APA | 66/67 | 98.5 | (88.0, 100.0) |
| | ANA | 52/53 | 98.1 | (83.3,100.0) |
| | OPA | 59/60 | 98.3 | (83.3,100.0) |
| MSH2 | APA | 72/72 | 100 | (94.9100.0) |
{10}------------------------------------------------
| | ANA | 48/48 | 100 | (92.3,100.0) |
|------------|-----|-------|-------|--------------|
| | OPA | 60/60 | 100 | (94.0,100.0) |
| BRAF V600E | APA | 60/60 | 100.0 | (93.9,100.0) |
| | ANA | 60/60 | 100.0 | (93.9,100.0) |
| | OPA | 60/60 | 100.0 | (94.0,100.0) |
### b. Reproducibility:
The reproducibility of the Ventana MMR IHC panel was assessed at 3 sites with 40 archival FFPE CRC tissue specimens. For the 4 antibodies against MMR protein in the panel (anti-MLH1, anti-PMS2, anti-MSH2, and anti-MSH6), 6 CRC tissue specimens (3 intact and 3 loss cases) were included in the study, resulting in 24 cases total for these 4 antibodies. In addition, for the anti-BRAF V600E antibody, 16 CRC tissue specimens -8 positive and 8 negative cases- with adequate tumor content (at least 50 viable tumor cells) were included in the study. Multiple tissue sections were cut from each case. Three (3) external clinical sites stained all cases with the designated antibody and the appropriate negative reagent control (NRC) antibody on each of 5 non-consecutive days spanning a period of at least 20 days. Cases were stained on a BenchMark ULTRA instrument in a different randomized order each day.
Specimen sets consisting of H&E, NRC, and antibody stained slide from all four antibodies in the MMR IHC panel staining were combined and randomized into a dayspecific reading set in order to minimize recall bias within the study. The case slide triplets for the anti-BRAF V600E antibody were combined into a separate, randomized, day-specific reading set.
Two pathologists at each site independently evaluated the reading set for the MMR IHC panel to determine the status (intact or loss) for each case. The same two pathologists at each site independently evaluated the separate reading set for BRAF V600E to determine the BRAF V600E status (positive or negative) for each case. The pathologists also evaluated all of the case slides for staining artifacts and staining failures affecting their ability to interpret the slides.
There were 720 observations for the four MMR markers (4 markers * 6 cases per marker * 5 days* 2 readers * 3 sites =720 observations). There were 480 observations for the BRAF marker (16 cases*5days*2 readers* 3 sites= 480 observations). When pooling the observations of all five markers, there were 1200 observations.
MMR-intact cases and BRAF V600E-positive cases were used to calculate PPA, and MMR-loss cases and BRAF V600E-negative cases were used to calculate NPA. Point estimates for PPA, NPA, and OPA were and their 95% confidence intervals (CI) for calculated by using the generalized linear mixed model (GLMM) approach. Modal case reference status for calculation PPA, NPA and OA was derived on the most often observed status of 30 observations.
{11}------------------------------------------------
The overall performance of the Ventana MMR IHC panel plus the anti-BRAF V600E antibody was to be considered acceptable if both the PPA and NPA rates across all observations exhibited a lower bound of the 2-sided 95% confidence interval (LBCI) > 85%, when using the modal result for each case as the reference for that case.
Summary of pooled agreement statistics between modal case reference status and individual observation are summarized in Table 10. PPA was 99.8% for all proteins and NPA was 98.9% for all proteins. There were 9 discordant cases, one each for MSH2 and MSH6 four for PMS2 and 3 for BRAF V600E status. Additionally, pair wise comparison for between site, between day and between readers was assessed by individual panel maker and 95% Confidence intervals calculated with Wilsons' score method. The data is summarized in Tables 11-15.
| Inter-<br>Laboratory<br>Reproducibility | Clinical Status | Agreement | | | |
|-----------------------------------------|-----------------|-----------|-----------|------|--------------|
| | | Type | n/N | % | 95% CI |
| All Proteins | Intact/Positive | PPA | 598/600 | 99.8 | (98.7,100.0) |
| | Loss/Negative | NPA | 593/600 | 98.9 | (97.4, 99.5) |
| | Total | OPA | 1191/1200 | 99.4 | (98.6, 99.7) |
Table 10. Inter laboratory Reproducibility for all proteins
| Table 11. Pairwise Agreement rates for Ventana MLH-1 (M1) Antibody | | | |
|--------------------------------------------------------------------|--|--|--|
|--------------------------------------------------------------------|--|--|--|
| Inter-Laboratory<br>Reproducibility<br>MLH1 (M1) Ab | | Agreement | | | |
|-----------------------------------------------------|--------|-----------|---------|-------|--------------|
| | | Type | n/N | % | 95% CI |
| Between-Site<br>(3 sites) | | APA | 360/360 | 100.0 | (98.9,100.0) |
| | | ANA | 360/360 | 100.0 | (98.9,100.0) |
| | | OPA | 360/360 | 100.0 | (98.9,100.0) |
| Between-<br>Day<br>(5 non-<br>consecutive<br>days) | Site A | APA | 120/120 | 100.0 | (96.9,100.0) |
| | | ANA | 120/120 | 100.0 | (96.9,100.0) |
| | | OPA | 120/120 | 100.0 | (96.9,100.0) |
| | Site B | APA | 120/120 | 100.0 | (96.9,100.0) |
| | | ANA | 120/120 | 100.0 | (96.9,100.0) |
| | | OPA | 120/120 | 100.0 | (96.9,100.0) |
| | Site C | APA | 120/120 | 100.0 | (96.9,100.0) |
| | | ANA | 120/120 | 100.0 | (96.9,100.0) |
| | | OPA | 120/120 | 100.0 | (96.9,100.0) |
| Between-Reader | | APA | 90/90 | 100.0 | (95.9,100.0) |
{12}------------------------------------------------
| Inter-Laboratory<br>Reproducibility<br>MLH1 (M1) Ab<br>(2 pathologists per<br>site) | Agreement | | | |
|-------------------------------------------------------------------------------------|-----------|-------|-------|--------------|
| | Type | n/N | % | 95% CI |
| | ANA | 90/90 | 100.0 | (95.9,100.0) |
| | OPA | 90/90 | 100.0 | (95.9,100.0) |
Table 12. Pairwise Agreement rates for Ventana PMS2 (A16-4) Antibody
| Inter-Laboratory<br>Reproducibility<br>PMS2(A16-4) Ab | Type | n/N | Agreement<br>%<br>95% CI |
|-------------------------------------------------------|------------|---------|--------------------------|
| Between-Site<br>(3 sites) | APA | 344/360 | 95.6<br>(90.7,100.0) |
| | ANA | 344/360 | 95.6<br>(90.7,100.0) |
| | OPA | 344/360 | 95.6<br>(91.1,100.0) |
| Between-<br>Day<br>(5 non-<br>consecutive<br>days) | APA | 120/120 | 100.0<br>(96.9,100.0) |
| | Site A ANA | 120/120 | 100.0<br>(96.9,100.0) |
| | OPA | 120/120 | 100.0<br>(96.9,100.0) |
| | APA | 120/120 | 100.0<br>(96.9,100.0) |
| | Site B ANA | 120/120 | 100.0<br>(96.9,100.0) |
| | OPA | 120/120 | 100.0<br>(96.9,100.0) |
| | APA | 104/120 | 86.7<br>(69.2,100.0) |
| | Site C ANA | 104/120 | 86.7<br>(69.2,100.0) |
| | OPA | 104/120 | 86.7<br>(73.3,100.0) |
| Between-Reader<br>(2 pathologists per<br>site) | APA | 90/90 | 100.0<br>(95.9,100.0) |
| | ANA | 90/90 | 100.0<br>(95.9,100.0) |
| | OPA | 90/90 | 100.0<br>(95.9,100.0) |
Table 13. Pairwise Agreement rates for Ventana anti MSH2 (G219-1129) Antibody
| Inter-Laboratory<br>Reproducibility<br>MSH2<br>(G219-1129) Ab | Agreement | | | | |
|---------------------------------------------------------------|-----------|---------|---------|--------------|--------------|
| | Type | n/N | % | 95% CI | |
| Between-Site<br>(3 sites) | APA | 360/364 | 98.9 | (96.8,100.0) | |
| | ANA | 352/356 | 98.9 | (96.6,100.0) | |
| | OPA | 356/360 | 98.9 | (96.7,100.0) | |
| Between-Day | Site A | APA | 120/120 | 100.0 | (96.9,100.0) |
| | | ANA | 120/120 | 100.0 | (96.9,100.0) |
{13}------------------------------------------------
| Inter-Laboratory<br>Reproducibility<br>MSH2<br>(G219-1129) Ab | | Agreement | | | |
|---------------------------------------------------------------|--------|-----------|---------|-------|--------------|
| | | Type | n/N | % | 95% CI |
| (5 non-<br>consecutive<br>days) | Site B | OPA | 120/120 | 100.0 | (96.9,100.0) |
| | | APA | 120/120 | 100.0 | (96.9,100.0) |
| | | ANA | 120/120 | 100.0 | (96.9,100.0) |
| | Site C | OPA | 120/120 | 100.0 | (96.9,100.0) |
| | | APA | 120/124 | 96.8 | (90.9,100.0) |
| | | ANA | 112/116 | 96.6 | (88.9,100.0) |
| | | OPA | 116/120 | 96.7 | (90.0,100.0) |
| Between-Reader<br>(2 pathologists per<br>site) | | APA | 90/91 | 98.9 | (96.8,100.0) |
| | | ANA | 88/89 | 98.9 | (96.6,100.0) |
| | | OPA | 89/90 | 98.9 | (96.7,100.0) |
Table 14. Pairwise Agreement rates for Ventana MSH6 (SP93) Antibody
| Inter-Laboratory<br>Reproducibility<br>MSH6 (SP93) AB | Agreement<br>Type | n/N | % | 95% CI | |
|-------------------------------------------------------|-------------------|---------------------------|---------|--------------|--------------|
| | | Between-Site<br>(3 sites) | APA | 360/364 | 98.9 |
| ANA | 352/356 | | 98.9 | (96.6,100.0) | |
| OPA | 356/360 | | 98.9 | (96.7,100.0) | |
| Between-<br>Day<br>(5 non-<br>consecutive<br>days) | Site A | APA | 120/120 | 100.0 | (96.9,100.0) |
| | | ANA | 120/120 | 100.0 | (96.9,100.0) |
| | | OPA | 120/120 | 100.0 | (96.9,100.0) |
| | Site B | APA | 120/124 | 96.8 | (90.9,100.0) |
| | | ANA | 112/116 | 96.6 | (88.9,100.0) |
| | | OPA | 116/120 | 96.7 | (90.0,100.0) |
| | Site C | APA | 120/120 | 100.0 | (96.9,100.0) |
| | | ANA | 120/120 | 100.0 | (96.9,100.0) |
| | | OPA | 120/120 | 100.0 | (96.9,100.0) |
| Between-Reader<br>(2 pathologists per<br>site) | APA | 90/91 | 98.9 | (96.8,100.0) | |
| | ANA | 88/89 | 98.9 | (96.6,100.0) | |
| | OPA | 89/90 | 98.9 | (96.7,100.0) | |
| Table 15. Pairwise Agreement rates for Ventana anti BRAF V600E (VE1) | | |
|----------------------------------------------------------------------|--|--|
| Antibody | | |
| Inter-Laboratory | Agreement |
|------------------|-----------|
|------------------|-----------|
{14}------------------------------------------------
| Reproducibility<br>BRAF V600E<br>(VE1) Ab | | Type | n/N | % | 95% CI |
|----------------------------------------------------|--------|------|---------|-------|--------------|
| Between-Site<br>(3 sites) | | APA | 960/972 | 98.8 | (97.2,100.0) |
| | | ANA | 936/948 | 98.7 | (97.0,100.0) |
| | | OPA | 948/960 | 98.8 | (97.1,100.0) |
| Between-<br>Day<br>(5 non-<br>consecutive<br>days) | Site A | APA | 320/320 | 100.0 | (98.8,100.0) |
| | Site A | ANA | 320/320 | 100.0 | (98.8,100.0) |
| | Site A | OPA | 320/320 | 100.0 | (98.8,100.0) |
| | Site B | APA | 320/320 | 100.0 | (98.8,100.0) |
| | Site B | ANA | 320/320 | 100.0 | (98.8,100.0) |
| | Site B | OPA | 320/320 | 100.0 | (98.8,100.0) |
| | Site C | APA | 320/332 | 96.4 | (92.0,100.0) |
| | Site C | ANA | 296/308 | 96.1 | (90.4,100.0) |
| | Site C | OPA | 308/320 | 96.3 | (91.3,100.0) |
| Between-Reader<br>(2 pathologists per<br>site) | | APA | 242/243 | 99.6 | (98.8,100.0) |
| | | ANA | 236/237 | 99.6 | (98.7,100.0) |
| | | OPA | 239/240 | 99.6 | (98.8,100.0) |
### b. Linearity/assay reportable range:
Not applicable
- c. Traceability (controls, calibrators, or methods), Stability, Expected values:
- i. Controls: See section I for description of controls.
- ii. Stability: Product expiration dating is based on testing 3 lots of MMR antibodies in accelerated stability studies.
- a) Reagent Stability Studies:
MLH1, MSH2, MSH6, PMS2: The stability of MMR panel antibodies was tested by subjecting 3 production equivalent lots of MMR antibodies to heat stress in an accelerated stability model. Staining performance for the antibodies was assessed after subjecting the antibodies to 243 hours at 45°C or 531 hours at 37°C to support a stability claim of 24 months. Interim stability after antibody storage at 45°C and 243 hours (12 month equivalent) was also assessed. The staining was performed using an OptiView DAB IHC Detection Kit on the BenchMark ULTRA with 3 tissue blocks, including 2 intact CRC (colorectal carcinoma) cases and 1 dMMR CRC case. The staining intensities were compared with case slides stained at baseline (0 hours) on a 0-4 scale and determined that the staining intensity did not vary more than 1.0 point and background by 0.5 points when compared to staining at base line (0 hours of stress) on a 0-4 scale .
{15}------------------------------------------------
Study met acceptance criteria for MLH-1, MSH2 and MSH6 antibodies to support expiration dating of 24 months for these 3 antibodies. Real time stability studies in support of the 24 month claim are on-going.
PMS2 antibody passed accelerated stability testing on storage at 37°C for 287 hours to support a 12 month dating and at 30°C for 603 and 1081.5 hours. Results from real time stability for one lot of PMS2 antibody was provided to support stability claims based on accelerated stability testing at lower temperatures (300C instead of 37℃ and 45°C). One lot of antibody was tested for intended storage condition (2-8 ℃); ship stress (15℃ and 30℃ for 192 hours) and cold ship stress (-20°C, 192 hours) at 4 and 6 months with 6 PMS2 Intact and 3 Loss cases and tonsil controls specimens. The PMS2 antibody lot passed the 6 month time point with PMS2 specific staining within 1 point and background staining within 0.5 point when compared to baseline on a 0-4 point intensity scale. The data supports a stability claim of 12 months for PMS2. Real time stability studies in support of the claim are on-going
BRAF V600E: Real time stability testing was conducted using 3 Stability Master Lots (SMLs). Each SML consisted of a 1 lot of anti-BRAF V600E (VE1) antibody and 1 lot of OptiView detection kit. Real time stability was assessed for the 3 stability master lots beginning of the study (Day 0) and subsequently assessed for intended storage, ship stress and freeze thaw cold ship stability for a total of 37 months with testing at 3, 4, 6, 8, 9, 12, 14, 18, 20, 24, 26 and 37 months. Intended storage (2-8 ℃) was tested through 37 months, heated ship stress (30℃ for 192 hours) was tested to 26 months and Freeze thaw cold ship stress (-20 C for 192 hours) for 37 months. Initial assessment of stability was performed with thyroid papillary carcinoma cases, and was subsequently confirmed with 6 CRC cases. Testing included ship stress, freeze thaw and cold ship stability parameters. Data supports expiration dating of 24 months for the anti BRAF V600E antibody.
- b) Cut slide Stability: Specimen stability with storage time and temperature was assessed on cut sections stored prior to staining for each of the panel proteins in separate studies. Studies for each of the MMR panel protein used 1 normal colon and 2 CRC cases with expression of MMR (Intact) in tumors. For BRAF V600 E section stability 1 WT and 3 BRAF V600E mutation positive cases were included. Slides were stored at 5±3°C or at 30±5°C. Two (2) slides were stained with Ventana MMR antibody and 1 slide with negative reagent control at different time points starting from the beginning of the experiment (Day 0) up to six months (Week 26). Cut slides for
{16}------------------------------------------------
PMS2 included 5 PMS 2 intact cases and stability was tested to 8 weeks. The anti-MMR panel antibody stained slides were evaluated by pathologist for MMR protein status (Intact/Loss), stain intensity and background. The negative reagent control stained slides were also evaluated. The study required that the stain intensity will not vary by more than 1.0 point when compared to Day 0 scores for a minimum of four weeks after storage at 5±3°C and at 30±5°C. Table 16 lists specimen stability as cut sections stored at 5±3°C and at 30±5°C for each of the panel markers. The specimen block stability claim is the same as the cut-section stability claim.
| Marker | Cut section stability* |
|------------|------------------------|
| MLH1 | 26 Weeks |
| PMS2 | 8 weeks |
| MSH2 | 26 Weeks |
| MSH6 | 26 Weeks |
| BRAF V600E | 26 Weeks |
Table 16. Cut Section Stability by marker
* When Stored at 5±3°C and at 30±5°C
- d. Detection Limit: Not applicable
- Analytical specificity: e.
Analytical specificity was addressed in two separate studies for each of the MMR panel antibodies. The first addressed antibody specificity and the second immunoreactivity in normal and neoplastic tumor specimen.
- Western Blot and IHC: Western blots analyses were conducted to i. demonstrate that the antibodies specifically detect the proteins of predicted molecular weight for each of the 5 Ventana MMR panel antibodies using cell lines with known MMR loss or intact status. Cell lines used in the study were matched pair of human cell line HD PAR595 that expressed wild type MMR proteins or were engineered with frame shift knockouts for PMS2 and MLH1 and complete knockout for MSH6 and MSH2. BRAFV600E containing cell lines were engineered to express moderate and high levels of the V600E protein. Western Blots confirmed presence of reactive bands at expected molecular weighs for each of the 5 panel markers.
IHC tests using the same cell lines formalin-fixed. paraffin-embedded conducted to assess nonspecific binding in the context of use. The results of the IHC with engineered cell lines was consistent with expected reactivity.
The combined results from western blots and cell line IHC demonstrated
{17}------------------------------------------------
specific antibody reactivity for each of the 5 markers included in the Ventana MMR IHC panel.
- ii. Immunoreactivity: Immunoreactivity testing to demonstrate Ventana MMR panel and BRAFV600 E antibodies (M1) staining across multiple cases of normal and tumor tissue types was performed on commercially available tissues and tissue arrays were obtained for Tour of Body (TOB) and Tour of Tumor (TOT) studies. Note: Mismatch repair proteins are present in all actively proliferating cells. For all tissues, positive/negative MMR staining was determined for tissue specific elements in the presence of positive staining in normal control cells (lymphocytes, fibroblasts and epithelial cells). For all tissues, BRAF V600E positive/negative staining was determined for tissue specific elements and such cases should not be considered as positive for BRAF V600E Clinical Status. The summary of staining results with the panel antibodies is shown in Table 17 and Table 18.
| | Positive/Total Cases | | | | |
|-------------------|----------------------|-------|------|------|------------|
| Tissue | MLH1 | PMS2- | MSH2 | MSH6 | BRAF V600E |
| Adrenal Gland | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Bladder | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Bone Marrow | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Ovary | 5/5 | 4/4 | 5/5 | 5/5 | 0/3 |
| Breast | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Cerebellum | 3/3 | 3/3 | 3/3 | 3/3 | 1/3* |
| Cerebrum | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Cervix | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Colon | 3/3 | 3/3 | 3/3 | 3/3 | 5/12* |
| Endometrium | 3/3 | 3/3 | 3/3 | 2/3 | 0/3 |
| Esophagus | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Heart | 3/3 | 2/3 | 1/3 | 3/3 | 0/3 |
| Hypophysis | 3/3 | 3/3 | 3/3 | 3/3 | 3/3** |
| Intestine | 3/3 | 3/3 | 3/3 | 3/3 | 2/4* |
| Kidney | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Liver | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Lung | 4/4 | 3/3 | 3/3 | 4/4 | 0/3 |
| Lymph node | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Mesothelium | 4/4 | 2/3 | 3/3 | 3/3 | 0/3 |
| Pancreas | 3/3 | 3/3 | 3/3 | 3/3 | 2/3* |
| Parathyroid Gland | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Peripheral Nerve | 5/5 | 4/4 | 5/5 | 5/5 | 0/5 |
| Prostate | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Skeletal Muscle | 3/3 | 2/3 | 3/3 | 3/3 | 0/3 |
Table 17. Tour of Body Immunoreactivity (Normal Tissue)
{18}------------------------------------------------
| Skin | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
|-----------------------|-----|-----|-----|-----|------|
| Spleen | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Stomach | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Testis | 3/3 | 3/3 | 3/3 | 3/3 | 2/3* |
| Thymus | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
| Thyroid | 4/4 | 4/4 | 3/3 | 4/4 | 0/3 |
| Tongue/Salivary Gland | 3/3 | 3/3 | 2/3 | 3/3 | 0/3 |
| Tonsil | 3/3 | 3/3 | 3/3 | 3/3 | 0/3 |
For BRAF V600E, * Weak cytoplasmic and nuclear staining in Purkinje cells of cerebellum, smooth muscle and epithelial cells of normal colon, glandular cells of intestine, acinar structures of pancreas, and interstitial cells of testis. ** Moderate staining observed in neuroendocrine cells in hypophysis.
| | MLH1-<br>positive /<br>total cases | PMS2-<br>positive /<br>total cases | MSH2-<br>positive /<br>total cases | MSH6-<br>positive /<br>total cases | BRAF<br>V600E-<br>positive /<br>total cases |
|---------------------------------------------------------------------|------------------------------------|------------------------------------|------------------------------------|------------------------------------|---------------------------------------------|
| Pathology | positive /<br>total cases | positive /<br>total cases | positive /<br>total cases | positive /<br>total cases | positive /<br>total cases |
| Glioblastoma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Atypical meningioma | 1/1 | n.e.* | 1/1 | 1/1 | 0/1 |
| Malignant ependymoma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Malignant<br>oligodendroglioma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Serous adenocarcinoma<br>(ovary) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Adenocarcinoma (ovary) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Islet cell carcinoma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Adenocarcinoma of<br>pancreas | n.e. | n.e. | 1/1 | n.e | 0/1 |
| Seminoma | 2/2 | 2/2 | 2/2 | 2/2 | 0/2 |
| Thyroid medullary<br>carcinoma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Thyroid papillary<br>carcinoma | 1/1 | 1/1 | 1/1 | 1/1 | 3/3 |
| Intraductal carcinoma<br>(breast) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Intraductal carcinoma with<br>early infiltrate (breast) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Invasive ductal carcinoma<br>(breast) | 1/1 | 1/1 | 1/1 | 1/1 | 1/1 |
| Diffuse B-cell lymphoma | 1/1 | n.e. | 1/1 | 1/1 | 0/1 |
| Lung small cell<br>undifferentiated carcinoma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| | MLH1-<br>positive /<br>total cases | PMS2-<br>positive /<br>total cases | MSH2-<br>positive /<br>total cases | MSH6-<br>positive /<br>total cases | BRAF<br>V600E-<br>positive /<br>total cases |
| Pathology | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Lung squamous cell<br>carcinoma | | | | | |
| Neuroendocrine carcinoma<br>(esophagus) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Adenocarcinoma<br>(esophagus) | 1/1 | n.e. | 1/1 | 1/1 | 0/1 |
| Signet ring carcinoma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Adenocarcinoma (small<br>intestine) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Stromal sarcoma (small<br>intestine) | 1/1 | 1/1 | 1/1 | 1/1 | 1/1 |
| Adenocarcinoma (colon) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Interstitialoma (abdominal<br>cavity) | 1/1 | n.e. | n.e. | n.e. | 0/1 |
| Adenocarcinoma (rectum) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Moderate malignant<br>interstitialoma (rectum) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Hepatocellular carcinoma | n.e. | n.e. | n.e. | 1/1 | 0/1 |
| Hepatoblastoma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Clear cell carcinoma<br>(kidney) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Adenocarcinoma (Prostate,<br>Gleason grade:4, Gleason<br>score:4+5) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Adenocarcinoma (Prostate) | 1/1 | n.e. | 1/1 | 1/1 | 0/1 |
| Leiomyoma (uterus) | 1/1 | n.e. | 1/1 | n.e. | n.e. |
| Adenocarcinoma (uterus) | 1/1 | n.e. | 1/1 | 1/1 | 0/1 |
| Clear cell carcinoma<br>(endometrium) | n.e. | n.e. | n.e. | 1/1 | n.e. |
| Squamous cell carcinoma<br>(cervix) | 2/2 | 1/1 | 2/2 | 2/2 | 0/2 |
| Embryonal<br>rhabdomyosarcoma of left<br>leg | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Squamous cell carcinoma<br>of chest wall | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Neurofibroma | 1/1 | n.e. | n.e. | n.e. | 0/1 |
| Neuroblastoma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Pathology | MLH1-<br>positive /<br>total cases | PMS2-<br>positive /<br>total cases | MSH2-<br>positive /<br>total cases | MSH6-<br>positive /<br>total cases | BRAF<br>V600E-<br>positive /<br>total cases |
| Malignant mesothelioma | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Diffuse B-cell lymphoma<br>of lymph node | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Diffuse B-cell lymphoma<br>of right thigh | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Hodgkin's lymphoma left<br>groin | 1/1 | 1/1 | 1/1 | 1/1 | 1/1 |
| Transitional cell carcinoma<br>of bladder | n.e. | n.e. | n.e. | n.e. | 0/1 |
| Low grade malignant<br>leiomyosarcoma (bladder) | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Osteosarcoma of right<br>femur | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
| Spindle cell<br>rhabdomyosarcoma | 1/1 | n.e. | n.e. | 1/1 | 0/1 |
| Moderate malignant<br>leiomyosarcoma of left<br>buttock | 1/1 | 1/1 | 1/1 | 1/1 | 0/1 |
Table 18. Ventana MMR IHC Panel Staining in Tumor Tissues
{19}------------------------------------------------
{20}------------------------------------------------
Note: Mismatch repair proteins are present in all actively proliferating cells. For all tissues, MMR positive/negative staining was determined for tumor cells in the presence of positive staining in normal control cells (lymphocytes, fibroblasts and epithelial cells).
For BRAF V600E, for all tissues, BRAF V600E positive/negative staining was determined for tumor cells.
*n.e .: non-evaluable either due to tissue loss or lack of internal control staining.
- f. Assay cut-off:
No Assay cut off is employed in the assessment of MMR status or in the assessment of BRAF V600E protein status in FFPE CRC tissue.
- 2. Comparison studies:
- a. Method comparison with predicate device:
Not Applicable
- b. Matrix comparison:
The device is only validated for formalin fixed paraffin embedded (FFPE) colorectal cancer tissue.
{21}------------------------------------------------
### 3. Clinical Performance:
The clinical validity of the Ventana MMR IHC panel was determined in a study that assessed agreement between test results obtained from the Ventana MMR IHC panel and a DNA sequencing panel validated for detecting pathogenic lynch syndrome variants and BRAF V600E mutations in colorectal carcinoma (CRC) specimens. The DNA sequencing panel was used to detect variants in the MMR genes MLH-1. MSH2. PMS2 and MSH6, along with EPCAM, and BRAF gene that are associated with MMR deficiency in CRC. The purpose of the study was to estimate the ability of the panel to correctly aid in the identification of patients needing additional Lynch syndrome genetic testing.
A sequential series of CRC specimens were procured and enriched with a second set of specimens with known Lynch syndrome variants due to the rarity of the variants. Specimens from the two studies were combined and randomized for testing with Ventana MMR IHC panel. Concordance (PPA, NPA and OPA) between the two methods was calculated contingent on the DNA sequencing results. MMR status (Intact/Loss) was stratified by BRAF V600E status and DNA sequencing results were stratified by presence or absence of known pathogenic mutations.
CRC cases were procured and assessed for quality (e.g., presence of tumor and internal control cells) prior to use in the study. Specimens were required to have 50% tumor content to meet specimen requirements for DNA sequencing. Of the sequential CRC cases, 7 cases were excluded from the specimen set due to insufficient viable tumor (i.e., adequate cellularity or lack of tumor content), 3 cases due to misclassification as CRC, and 1 due to clerical error. Following review, 111 sequential cases meeting the study criteria were enrolled into the study. A total of 15 CRC cases with confirmed loss status for MMR protiens by IHC brought the total to 126 specimens. Assessment of the demographic data associated with the study specimens determined that it was consistent intended use population.
Results: Of 126 specimens tested by the IHC panel and DNA sequencing, 7 cases were excluded from final analysis due failure of DNA sequencing. Of the remaining 119, one failed IHC testing. The point estimates for overall agreement between Ventana MMR IHC panel and DNA sequencing were 77.8% PPA, 97.0% NPA and 94.1% OPA. The comparisons of IHC and sequencing status for all specimens with MMR IHC results stratified by BRAFV600E status is summarized in Table 19, Table 20A and Table 20B summarizes results by sequential and enrichment study sets respectively.
### Table 19. Agreement between VENTANA MMR IHC Panel Results and DNA Sequencing Results: All Specimens
| | DNA Sequencing Results | | | |
|-----------------------|------------------------|------------------------------|---------|-------|
| MMR IHC Panel Results | Pathogenic<br>Mutation | No<br>Pathogenic<br>Mutation | Invalid | Total |
{22}------------------------------------------------
| MMR IHC Panel Results | | DNA Sequencing Results | | | |
|-----------------------|---------------------------|------------------------|------------------------------|--------------|-------|
| | | Pathogenic<br>Mutation | No<br>Pathogenic<br>Mutation | Invalid | Total |
| MMR Loss | BRAF<br>V600E<br>Positive | 1 | 19 | 0 | 20 |
| | BRAF<br>V600E<br>Negative | 14 | 3 | 1 | 18 |
| MMR Intact | BRAF<br>V600E<br>Positive | 0 | 3 | 1 | 4 |
| | BRAF<br>V600E<br>Negative | 2 | 76 | 5 | 83 |
| Invalid | | 1 | 0 | 0 | 1 |
| Total | | 18 | 101 | 7 | 126 |
| Agreement | | | | | |
| Type | | n/N | % | 95% CI | |
| PPA | | 14/18 | 77.8 | (54.8, 91.0) | |
| NPA | | 98/101 | 97.0 | (91.6, 99.0) | |
| OPA | | 112/119 | 94.1 | (88.4, 97.1) | |
Note: Invalids are defined as failure to produce results by IHC and/or DNA sequencing. Only Invalids resulting from a failure by IHC are included in the analysis. 95% CIs were calculated using the (Wilson) Score method.
The association between the test results and the final diagnosis with respect to Lynch Syndrome is an estimate because the study was enriched with Lynch syndrome positive cases.
| Table 20A. Agreement between VENTANA MMR IHC Panel and DNA | |
|------------------------------------------------------------|--|
| Sequencing Results: Sequential Cohort | |
| Sequential Study Set | | | | | | | |
|-------------------------------|-----------------|------------------------|---------------------------|---------|----|--|--|
| VENTANA MMR IHC Panel Results | | DNA Sequencing Results | | | | | |
| | | Pathogenic<br>Mutation | No Pathogenic<br>Mutation | Invalid | | | |
| | BRAF<br>V600E + | 1 | 18 | 0 | 19 | | |
| MMR Loss | BRAF<br>V600E - | 4 | 2 | 0 | 6 | | |
{23}------------------------------------------------
| Sequential Study Set | | | | | |
|-------------------------------|-----------------|------------------------|---------------------------|---------|-------|
| VENTANA MMR IHC Panel Results | | DNA Sequencing Results | | | |
| | | Pathogenic<br>Mutation | No Pathogenic<br>Mutation | Invalid | Total |
| MMR Intact | BRAF<br>V600E + | 0 | 3 | 1 | 4 |
| | BRAF<br>V600E - | 1 | 76 | 5 | 82 |
| Invalid | | 0 | 0 | 0 | 0 |
| Total | | 6 | 99 | 6 | 111 |
| Agreement | | | | | |
| Type | n/N | % | 95% CI | | |
| PPA | 4/6 | 66.7 | (30.0, 90.3) |…
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.